INTRODUCTION

1-1 Unpacking and Inspection

Upon removing your new Clamp Multimeter from its packing, you should have the following items:

1. Clamp Multimeter.

2. Test lead set (one black, one red).

3. Carrying case.

4. Instruction manual.

1-2 Meter Safety

Terms as Marked on Equipment

ATTENTION — Refer to manual

DOUBLEINSULATION — Protection Class II

DANGER — Risk of electric shock.

1

loading

Related Manuals for APPA 39MR

Summary of Contents for APPA 39MR

  • Page 1
    INTRODUCTION 1-1 Unpacking and Inspection Upon removing your new Clamp Multimeter from its packing, you should have the following items: 1. Clamp Multimeter. 2. Test lead set (one black, one red). 3. Carrying case. 4. Instruction manual. 1-2 Meter Safety Terms as Marked on Equipment ATTENTION —…
  • Page 2
    Symbols in this Manual This symbol indicates where cautionary or other information is found in the manual.  Battery 1-3 Front Panel Refer to Figure 1 and the following numbered steps to familiarize yourself with the meter’s front panel controls and connectors. 1.
  • Page 3
    3. Function Switch — Rotary switch is used to select Hz, V , V , Ω, A , A function. 4. Trigger — Press the lever to open the transformer jaws. When the pressure on the lever is released, the jaws will close again.
  • Page 4
    Figure 1…
  • Page 5: Specifications

    SPECIFICATIONS 2-1 General Specifications Display : 3 3/4 Digital Liquid Crystal Display ( LCD) with a maximum reading of 3999. Polarity : Automatic polarity indicated. Overrange Indication : “OL” or “-OL” indicated. Range : Autoranging. Measuring Rate : 2 times per second, nominal. Position Error : +/- 1% of Reading.

  • Page 6
    Max/Conductor Size : 51mm diameter or 24 x 60mm busbar. Size : 100mm (W) x 265mm (L) x 42mm (H). Weight : 420 grams (including battery) Accessories : Test leads, battery (installed), manual and carrying case. 2-2 Environmental Conditions: Indoor use Maximum Altitude : 2000 Meter Installation Category…
  • Page 7
    2-3 Electrical Specifications Accuracy is ± (% reading + number of digits) at 23ºC ± 5ºC, less than 80% R.H. (1) Frequency : Hz Over Voltage Range Resolution Accuracy Protection 4KHz AC/DC 2000A ±(0.5% rdg + 3 dgt) for 1 min. 10KHz 10Hz Min.
  • Page 8
    (2) ACV : V Range Resolution Accuracy Overload Protection 400V 0.1V ±(1.2% rdg + 5 dgt) 850V rms 50Hz -500Hz 600V Additional error according to crest factor (CF) : CF from 2 to 3 : + 1.4% CF from 3 to 4 : + 3% Input impedance : 10MΩ…
  • Page 9
    (4) Resistance : Ω Max. Open Overload Range Resolution Accuracy Circuit Voltage Protection 4KΩ 1Ω ±(1.0% rdg + 2 dgt) 3.3V 600V r.m.s. 40KΩ 10Ω Instant Continuity Description : Built — in buzzer sound when resistance is less than 100Ω. (5) DCA : A Range Resolution…
  • Page 10
    Additional error according to remanence : 1% max. of current crest (6) ACA : A Frequency Overload Range Resolution Accuracy Response Protection 0 ~ 400A 0.1A ±(1.5% rdg + 5 dgt) 40 ~ 400 Hz AC 2000A 400 ~ 600A for 1 min.
  • Page 11: Operation

    OPERATION This instrument has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus and has been supplied in a safe condition. This instruction manual contains some information and warnings which have to be followed by the user to ensure safe operation and to retain the instrument in safe condition.

  • Page 12
    6. If the unit is not to be used for a long period of time, remove the battery. 7. Do not forget to turn off after use. THIS INSTRUMENT MUST NOT BE USED ON UNINSULATED CONDUCTORS AT A VOLTAGE GREATER THAN 600V ac/dc.
  • Page 13
    B. Frequency measurement by input terminal (AC voltage) 1. Set the rotary switch to «Hz» position. 2. Connect the black test lead to the “COM” terminal on the bottom of the meter and the red test lead to the «V Ω» terminal.
  • Page 14
    3-4 DC Voltage Measurement 1. Set the rotary switch at «V » position. 2. Connect the black test lead to the “COM” terminal on the bottom of the meter and the red test lead to the «V Ω” terminal. You can now place the test probes on the conductors to make the measurement. To avoid electrical shock, hazard or damage to the meter, do not attempt to measure voltage that might exceed 1000V dc between the common input terminal and earth ground.
  • Page 15
    3-6 DC Current Measurement Set the rotary switch at “A “ position at first. Low current : Press DCA ZERO switch before measuring. High current : 1. Enclose the conductor lead and close the clamp. 2. Remove the clamp and press the DCA ZERO switch . (remnant magnetization compensation) 3.
  • Page 16
    3-7 AC Current Measurement 1. Set the rotary switch at » A » position. 2. Open Spring-loaded clamp by pressing trigger on left side of meter. 3. Position clamp around wire or conductor and release clamp trigger, make sure that the clamp is entirely closed. The clamp must be positioned around only one conductor of a circuit.
  • Page 17: Maintenance

    MAINTENANCE WARNING TO AVOID ELECTRICAL SHOCK REMOVE TEST LEADS BEFORE OPENING THE COVER. 4-1 General Maintenance 1. Repairs or servicing not covered in this manual should only be performed by qualified personnel. 2. Periodically wipe the case with a dry cloth and detergent do not use abrasives or solvents. 4-2 Battery Installation or Replacement The meter is powered by a single 9V battery.

  • Page 18
    Battery Replacement Or Installation Figure 2…

Symbols in this Manual

This symbol indicates where cautionary or other information is found in the manual.

Battery

1-3 Front Panel

Refer to Figure 1 and the following numbered steps to familiarize yourself with the meter’s front panel

controls and connectors.

1. Digital Display — The digital display has a 3 3/4 digit LCD readout (maximum reading 3999) plus auto polarity,

decimal point,

2. Input Terminal — The black test lead is always connected to the «COM» input jack and the red test lead is always

connected to the «V Ω» input jack when measuring voltage or resistance.

* COM input terminal : Reference input connector.

* V Ω input terminal : Positive input connector.

, AC , DC ,

and unit annunciators.

2

APPA Manuals and Guides:

The main types of APPA 39MR instructions: user guide — rules of useing and characteristics, service manual — repair, diagnostics, maintenance, operation manual — description of the main functions of APPA 39MR equipment, etc.

Most of the instructions, that you can see on the site are uploaded by our users. If you have available a manual or document for APPA 39MR, which is currently not on the site or present in a different language version, we ask you to upload your document on website, using the «uploading form» available to all registered users.

Detail Specifications:

1787/1787406-39mr.pdf file (18 Oct 2022)

Accompanying Data:

APPA 39MR Multimeter PDF Manual  (Updated: Tuesday 18th of October 2022 08:57:51 PM)

Rating: 4.5 (rated by 28 users)

Compatible devices: 107N, 30R, 97II, 36II, 605, 98II, 607, HI 98127.

Recommended Documentation:

Text Version of Manual 

(Ocr-Read Summary of Contents, UPD: 18 October 2022)

  • 8, 8 (2) ACV : V Range Resolution Accuracy Overload Protection 400V 0.1V ±(1.2% rdg + 5 dgt) 50Hz -500Hz 600V 1V 850V rms Additional error according to crest factor (CF) : CF from 2 to 3 : + 1.4% CF from 3 to 4 : + 3% Input impedance : 10MΩ // 100pF. Conversion type : True RMS.…

  • 3, APPA 39MR 3 3. Function Switch — Rotary switch is used to select Hz, V, V, Ω, A, A function. 4. Trigger — Press the lever to open the transformer jaws. When the pressure on the lever is released, the jaws will close again. 5. Transformer Jaws — Designed to pick up the current flowing through the conductor. 6. Hand Guard …

  • 10, 10 Additional error according to remanence : 1% max. of current crest (6) ACA : A  Range Resolution Accuracy Frequency Response Overload Protection 0 ~ 400A 0.1A ±(1.5% rdg + 5 dgt) 40 ~ 400 Hz 400 ~ 600A 1A 600 ~ 1000A 1A ±(2.5% rdg + 5 dgt) 40 ~ 200 Hz AC 2000A for 1 min. Additional error according to crest factor (CF) : CF from 2 to …

  • 9, 9 (4) Resistance : Ω Range Resolution Accuracy Max. Open Circuit Voltage Overload Protection 4KΩ 1Ω 3.3V 600V r.m.s. 40KΩ 10Ω ±(1.0% rdg + 2 dgt) Instant Continuity Description : Built — in buzzer sound when resistance is less than 100Ω. (5) DCA : A  Range Resolution Accuracy Overload Protection 0 ~ 400A 0.1A AC 2000A …

  • 15, 15 3-6 DC Current Measurement Set the rotary switch at “A  “ position at first. Low current : Press DCA ZERO switch before measuring. High current : 1. Enclose the conductor lead and close the clamp. 2. Remove the clamp and press the DCA ZERO switch . (remnant magnetization compensation) 3. Enclose the conductor lead and close the clamp to go on with measurements (same current di…

  • 12, 12 6. If the unit is not to be used for a long period of time, remove the battery. 7. Do not forget to turn off after use.  THIS INSTRUMENT MUST NOT BE USED ON UNINSULATED CONDUCTORS AT A VOLTAGE GREATER THAN 600V ac/dc. 3-2 Frequency measurement A. Frequency measurement by clamp on (AC current) 1. Set the rotary switch to «Hz» position. 2. Open spri…

  • 6, APPA 39MR 6 Max/Conductor Size : 51mm diameter or 24 x 60mm busbar. Size : 100mm (W) x 265mm (L) x 42mm (H). Weight : 420 grams (including battery) Accessories : Test leads, battery (installed), manual and carrying…

  • 13, APPA 39MR 13 B. Frequency measurement by input terminal (AC voltage) 1. Set the rotary switch to «Hz» position. 2. Connect the black test lead to the “COM” terminal on the bottom of the meter and the red test lead to the «V Ω» terminal. You can now place the test probes on the conductors to make the measurement. 3-3 AC Voltage Measure…

Recommended Instructions:

TM800W560L, Palmcorder VM-L458, RPW202A, AR-3000, YP-100 — 1, QMD80B

  • 1OPERATING INSTRUCTIONS8 Function Digital Autoranging MultimeterDM6410Rotary SwitchOFFPower off+Battery MeasurementAA400A400ADC Voltage MeasurementAA400A400AAA400A400AAC/DC CurrentAA400A400AAC Voltage Measurement Continuity / Diode MeasurementAA400A400AResistance MeasurementAA400A400ARead this owners manual thoroughly before use and save.IMPORTANT: RECEIVING INSTR …

    DM6410 6

  • PN 3945752 November 2011 Rev. 1, 10/17 © 2011-2017 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies. 28 II Ex True-rms Digital Multimeter …

    28-II 36

  • Instruction ManualHI 93721Iron High Range ISMhttp://www.hannainst.comThis Instrument is inCompliance with the CE DirectivesHanna Instruments reserves the right to modify the design, constructionand appearance of its products without advance notice.WARRANTYWARRANTYWARRANTYWARRANTYWARRANTYHI 93721 is warranted for two years against defects in workman-ship and materials when used for its int …

    HI 93721 2

  • June 2013, Rev. 1, 10/15 © 2013-2015 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies. 106/107 Digital Multimeters Users ManualShop for Fluke products online at:1.888.610.7664www.MyFlukeStore.com …

    106 33

Additional Information:

Popular Right Now:
Operating Impressions, Questions and Answers:

Краткое описание клещей электроизмерительных APPA 39MR

Электроизмерительные клещи APPA 39MR предназначены для кратковременного измерения силы переменного и постоянного тока без разрыва силовой цепи, а также для измерения постоянного и переменного напряжения и сопротивления с помощью выносных щупов. Клещи лектроизмерительные применяются в полевых, цеховых, лабораторных условиях. Основная область применения: электротехника.

APPA 39MR представляют собой многофункциональные портативные электроизмерительный приборы в ударопрочном корпусе. В приборе применен бесконтактный метод измерения силы переменного тока основанный на применении датчиков на эффекте Холла с последующим аналогоцифровом преобразовании входных сигналов.

Отличительные особенности APPA 39MR

  • Постоянный/переменный ток от 0,1 А до 1000 А;
  • Переменное напряжение от 0,1 В до 600 В;
  • Постоянное напряжение от 0,1 В до 1000 В;
  • Сопротивление от 1 Ом до 40 кОм;
  • Измерение частоты от 20 Гц до 10 кГц;
  • Удержание показаний;
  • Пиковое значение;
  • Максимальный диаметр провода 51 мм;
  • Измерение среднеквадратических значений True RMS.

Технические характеристики клещей APPA 39MR

Параметр Значение
Переменное напряжение
Пределы измерений 400; 600 В
Погрешность ±(1,2 % + 5 ед. счета)
Максимальное разрешение 1 мВ
Полоса  частот 50…500 Гц
Измерение ср. кв. значения сигнал произвольной формы
Входной импеданс 10 МОм/100 пФ
Защита входа 850 В
Постоянное Напряжение
Пределы измерений 400; 1000 В
Погрешность ± (0.7 % + 2 е.м.р.)
Максимальное разрешение 1 мВ
Защита входа 1000 В
Переменный ток
Пределы измерений 400; 1000 А
Погрешность ±(1,5 % + 5 ед. счета)
Максимальное разрешение 0,01 А
Полоса частот 40…400 Гц
Измерение ср.кв. значения сигнал произвольной формы
Защита входа 2000 А (< 1 мин)
Постоянный ток
Пределы измерений 400; 1000 А
Погрешность ±(1,0 % + 3 ед. счета)
Максимальное разрешение 0,1 А
Защита входа 2000 А (< 1 мин)
Частота (Hz)
Основная гармоника 100 Гц … 50 кГц
Разрешение 0,01 Гц
Погрешность измерения ± (1,0% + 2 е.м.р.)
Сопротивление
Пределы измерений 4; 40 кОм
Погрешность ±(1,0 % + 2 ед. счета)
Максимальное разрешение 1 Ом
Защита входа 600 В
Прозвон цепи
Порог срабатывания 100 Ом
Индикация Непрерывный звуковой сигнал частотой 2,7 кГц
Защита входа 600 В
Частота и напряжение
Пределы измерений 4; 10 кГц
Погрешность ± (0,5 % + 3 ед. счета)
Макс. разрешение 1 Гц
Мин. входная частота 20 Гц
Чувствительность 6 А; 3 В
Защита входа 2000 А (< 1 мин)
Общие данные
Тип преобразователя Датчик Холла
Максимальная индикация 4000
Скорость измерения 2 изм./с
Автовыключение Через 10 мин
Макс. диаметр провода 51 мм (шина 24 60 мм)
Источник питания 9 В (тип «Крона»)
Срок службы источника питания 40 ч
Условия эксплуатации 0°С…50 °С, отн. влажность не более 80 %
Габаритные размеры 100 x 265 x 42 мм
Масса 420 г

Комплект поставки:

Наименование Количество
1 Клещи токоизмерительные APPA 39MR 1 — шт.
2 Измерительные провода 2 — шт.
3 Источник питания 1 — шт.
4 Транспортная сумка 1 — шт.
5 Руководство по эксплуатации 1 — шт.

Понравилась статья? Поделить с друзьями:
  • Каркам атлас 2 инструкция по применению
  • Вымпел 27 зарядное устройство инструкция пошаговая инструкция
  • Вымпел 27 зарядное устройство инструкция пошаговая инструкция
  • Можно ли уволить за несоблюдение должностной инструкции
  • Система руководства библиотеки