-
Contents
-
Table of Contents
-
Troubleshooting
-
Bookmarks
Related Manuals for Philips Avalon FM20
Summary of Contents for Philips Avalon FM20
-
Page 1
In stru ctio ns fo r Use Avalon Fetal Monitor FM20/30, FM40/50 Rel ease G.0 with Software Rev ision G.02.x x Patient Monitori ng… -
Page 3: Table Of Contents
Safety Tests 3 Basic Operation Supported Measurements Avalon FM20 and FM30 Avalon FM40 and FM50 Cordless Monitoring Getting to Know Your Avalon FM20/FM30 Getting to Know Your Avalon FM40/FM50 Operating and Navigating Operating Modes Automatic Screen Layouts Settings Checking Your Monitor Revision…
-
Page 4
5 Alarms Alarm Mode Visual Alarm Indicators Audible Alarm Indicators Acknowledging Alarms Acknowledging Disconnect INOPs Pausing or Switching Off Alarms Alarm Limits Reviewing Alarms Latching Alarms Testing Alarms Alarm Behavior at Power On 6 Patient Alarms and INOPs Alarm Messages Technical Alarm Messages (INOPs) 7 Admitting and Discharging Admit/Discharge on the Monitor… -
Page 5
Fetal Movement Profile Troubleshooting Additional Information Testing Ultrasound Transducers 11 Monitoring Twin FHRs Important Considerations Monitoring Twins Externally Monitoring Twins Internally Cross-Channel Verification Separating FHR Traces Troubleshooting 12 Monitoring Triple FHRs Important Considerations Monitoring Triplets Cross-Channel Verification Separating FHR Traces «Standard»… -
Page 6
Testing DECG Mode 17 Monitoring Noninvasive Blood Pressure Introducing the Oscillometric Noninvasive Blood Pressure Measurement Preparing to Measure Noninvasive Blood Pressure Starting and Stopping Measurements Enabling Automatic Mode and Setting Repetition Time Choosing the Alarm Source Assisting Venous Puncture Calibration Troubleshooting 18 Monitoring SpO2 Selecting an SpO2 Sensor… -
Page 7
Cleaning and Disinfecting Cleaning and Disinfecting Monitoring Accessories Sterilizing 24 Maintenance Inspecting the Equipment and Accessories Inspecting the Cables and Cords Maintenance Task and Test Schedule Storing Recorder Paper Cleaning the Printhead Disposing of the Monitor 25 Accessories and Supplies Information on Latex Transducers Fetal Accessories… -
Page 9: Introduction
Introduction Who this Book is For This book is for trained healthcare professionals using the Avalon FM20, FM30, FM40 and FM50 fetal/maternal monitors. It describes how to set up and use the monitor and transducers. Familiarize yourself with all instructions including warnings and cautions before starting to monitor patients. Read and keep the Instructions for Use that come with any accessories, as these contain important information about application and care and cleaning that is not repeated in this book.
-
Page 10: Confirm Fetal Life Before Using The Monitor
1 Introduction Confirm Fetal Life Before Using the Monitor Fetal monitoring technology available today is not always able to differentiate a fetal heart rate (FHR) signal source from a maternal heart rate (MHR) source in all situations. Therefore, you should confirm fetal life by independent means before starting to use the fetal monitor, for example, by palpation of fetal movement or auscultation of fetal heart sounds using a fetoscope, stethoscope, or Pinard stethoscope.
-
Page 11: Introducing The Avalon Family Of Fetal Monitors
1 Introduction Introducing the Avalon Family of Fetal Monitors The Avalon family of fetal monitors consists of the Avalon FM20, FM30, FM40 and FM50. While the FM20/FM30 and the FM40/FM50 have different form factors, the method of operation is very similar for all monitors.
-
Page 12
1 Introduction Indications for Use Avalon Fetal/Maternal Monitor FM20: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, oxygen saturation, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas, in private households and during transports in healthcare facilities. -
Page 13: Installation
For a list of conventions used in this guide, see “Basic Operation” on page 19. Not all accessories and supplies may be available in all geographies. Please contact your local Philips sales representative for details of availability. Installation Checklist Use this checklist to document your installation.
-
Page 14: Unpacking And Checking The Shipment
2 Installation Step Task Check Box when Task Done Check/set paper speed (see “Choosing Paper Speed” on page 48) Perform System Test as necessary (see the Service Guide) For monitors with the battery option (#E25) chosen, confirm that the battery can be charged, and that the monitor can be powered by the battery.
-
Page 15: Mounting The Monitor
2 Installation Claims for Damage If the shipping cartons are damaged, contact the carrier. If any of the equipment is damaged, contact both the carrier and your local Philips service organization for repair or replacement arrangements. Repacking Retain the original packing carton and material, in case you need to return equipment to Philips for service.
-
Page 16: Connecting The Monitor To Ac Mains
2 Installation Connecting the Monitor to AC Mains FM20/30 The monitor is an electrical Class II device in which the protection against electric shock does not rely on basic insulation and a protective earth conductor but on double and/or reinforced insulation. FM40/50 The monitor is an electrical Class I device.
-
Page 17: How And When To Carry Out The Test Blocks
2 Installation How and When to Carry Out the Test Blocks The following table defines which test and inspection blocks need to be performed, and when they are required. Test Block Test or Inspection to be Performed Test Block Required for Which Events? Visual Inspect the monitor, transducers and cables for any…
-
Page 18: Safety Tests
2 Installation Safety Tests Details of the safety tests and procedures required after an installation or an exchange of system components are described in your monitor’s Service Guide. These safety tests are derived from international standards but may not be sufficient to meet local requirements. WARNING •…
-
Page 19: Basic Operation
Basic Operation This chapter gives you an overview of the monitor and its functions. It tells you how to perform tasks that are common to all measurements (such as entering data, switching a measurement on, changing some monitor settings, and setting up the recorder). The alarms section gives an overview of alarms. The remaining sections tell you how to perform individual measurements, and how to care for and maintain the equipment.
-
Page 20: Avalon Fm20 And Fm30
This section outlines the capabilities of your monitor. Avalon FM20 The Avalon FM20 fetal/maternal monitor provides a solution for external fetal monitoring applications, and optional non-invasive maternal vital signs. You can monitor fetal heart rates (FHRs) externally using ultrasound, uterine activity and maternal pulse using an external Toco transducer, and the maternal heart rate (MHR) via maternal ECG electrodes, and optionally, non-invasive blood pressure.
-
Page 21: Avalon Fm40 And Fm50
The Avalon FM30 fetal/maternal monitor offers a solution for both external and internal fetal monitoring applications, and optional non-invasive maternal vital signs. The Avalon FM30 shares all the features and capabilities of the Avalon FM20. In addition, you can monitor one FHR internally via direct fetal electrocardiogram (DECG), uterine activity internally using…
-
Page 22: Cordless Monitoring
3 Basic Operation Avalon FM50 The Avalon FM50 fetal/maternal monitor offers a solution for both external and internal fetal monitoring applications, and non-invasive maternal vital signs. The Avalon FM50 shares all the features and capabilities of the Avalon FM40. In addition, you can monitor one FHR internally via direct fetal electrocardiogram (DECG), and uterine activity internally using an intra-uterine pressure (IUP) catheter together with a Toco transducer or patient module.
-
Page 23: Getting To Know Your Avalon Fm20/Fm30
3 Basic Operation Getting to Know Your Avalon FM20/FM30 Overview Touchscreen Display (tilt and fold) Power LED Paper Drawer Paper Drawer release Connectors (see Left Side view) Right Side On/Off Switch Power Connector with Battery Option On/Standby Switch MSL Connector…
-
Page 24: Getting To Know Your Avalon Fm40/Fm50
3 Basic Operation Rear 10 Display Release 11 Carrying Handle 12 Built-in Stand Left Side 13 SpO Socket (optional) 14 Noninvasive Blood Pressure Socket (optional) 15 Fetal Sensor Sockets — each socket accepts any fetal transducer, an Avalon CTS Cordless Fetal Transducer System base station (connected via the interface cable M2731- 60001), or event marker…
-
Page 25
3 Basic Operation Rear Reserved for future use: protective earth intended for use in system installations. Equipotential Grounding Point Power Cord Connector Loudspeaker Slot 01 for optional LAN / RS232 system interface (for connection to an obstetrical information and surveillance system) Slot 02 for optional interfaces: Either dual PS/2 system… -
Page 26
3 Basic Operation Ultrasound Transducer (M2736A) Cable — connects to any of the four Fetal Sensor Sockets on the monitor Note that the The M2736AA US transducer is identical to the M2736A US transducer, including all specifications. Toco Transducer with ECG/IUP capability (M2735A) transducer only) Connector — for connecting ECG/IUP adapter cables (M2735A Toco Butterfly Belt Clip (shown… -
Page 27: Operating And Navigating
3 Basic Operation Patient Module for ECG/IUP (M2738A) Connector — for connecting ECG/IUP adapter cables (same as for Toco transducer) 10 Cable — connects to any of the four Fetal Sensor Sockets on the monitor Operating and Navigating Your monitor has a touchscreen. Everything you need to operate the monitor, other than to turn it on and off, is contained on its screen.
-
Page 28
3 Basic Operation Date and time Bed label (when connected to a Philips OB TraceVue system) Fetal heart sound volume adjust/indicator Alarm volume adjust/indicator Numeric/measurement values Fetal Trace Recorder — status indicator Fetal recorder is On Fetal recorder is Off (when Paper Save Mode is off ) -
Page 29
3 Basic Operation Status line — shows status and prompt messages Signal quality indicator Good / full acceptable / medium Poor / no signal Measurement label (a cordless measurement from a connected Avalon CTS system is indicated by symbol) NST timer, if configured (default is Off) Keys The monitor has three different types of keys. -
Page 30
3 Basic Operation SmartKeys SmartKeys are configurable graphical keys, located at the bottom of the main screen. They give you fast access to functions. The selection of SmartKeys available on your monitor depends on your monitor configuration and on the options purchased. Main Setup Start/ Stop — enter main setup… -
Page 31
3 Basic Operation Start/ Stop Stop All — stop all noninvasive blood pressure measurements — start/stop manual noninvasive blood pressure measurement — start auto series — stop current automatic measurement within series Start NBP Stop NBP — start manual noninvasive blood — stop manual noninvasive blood pressure measurement pressure measurement… -
Page 32: Operating Modes
3 Basic Operation Selecting Screen Elements Select a screen element to tell the monitor to carry out the actions linked to the element. You access most screen elements by touching that element directly. For example, select the FHR1 Setup FHR1 Start/ Stop numeric to call up the menu, or select the…
-
Page 33: Automatic Screen Layouts
3 Basic Operation Automatic Screen Layouts Your monitor’s preconfigured screen layouts define how measurement information is arranged on screen. The monitor automatically applies the correct screen layout for the measurements you are monitoring. No user action is required. Connecting or disconnecting transducers, or switching the noninvasive blood pressure measurement on or off, results in an automatic adjustment of the screen layout.
-
Page 34
3 Basic Operation CAUTION This resets all settings to factory defined values, but be aware that some values will differ from those with which the monitor was originally shipped from the factory (recorder speed and paper scale type will need to be corrected, for instance). After loading the Factory Default, please check the settings and, if necessary, change them to the settings you normally use. -
Page 35: Checking Your Monitor Revision
Store Date, Time Select to change the date and time. If connected to a Philips OB TraceVue system, the monitor uses the OB TraceVue system date and time, including daylight saving time changes. WARNING Changing the date and time while the monitor is connected to an OB TraceVue system can result in a mismatch in the time and date between the monitor and the OB TraceVue system.
-
Page 36
3 Basic Operation Switching On: FM20/FM30 FM20/30 • Connect the monitor to AC mains and switch the monitor on. • The green power-on LED comes on. Selftest: OK • The monitor performs a self-test as it starts up. , the serial number, and revisions for Auto Start the software and firmware are printed on the fetal trace paper (if recorder is configured… -
Page 37
3 Basic Operation Lift the display forwards. You hear a click as the first position engages. If you want to tilt the display further, lift the display further forwards until you reach the desired angle. To fold the display: Pull the display forwards as far as it will go. Then push it all the way back until it clicks shut. -
Page 38
3 Basic Operation Fastening Belts and Transducers You can use more than one belt if, for example, you are monitoring uterine activity and FHR simultaneously. There are two basic ways to fasten belts and transducers: • Belts with button fixings. •… -
Page 39
3 Basic Operation Using Belt with Velcro Fixings Insert one end of the belt between the belt guides on one side of the butterfly belt clip, and secure with the velcro fixing. Insert the other end of the belt between the belt guides on the other side of the butterfly belt clip, adjust for the correct tension, then secure with the velcro fixing. -
Page 40
3 Basic Operation Connecting a Transducer to the Monitor socket Noninvasive blood pressure socket Fetal sensor sockets You can plug a fetal transducer, a ECG/IUP patient module, an Avalon CTS Cordless Fetal Transducer System interface cable (M2731-60001, red connector), or an external event marker into any of the four fetal sensor sockets marked or «Fetal Sensors»… -
Page 41
3 Basic Operation When you connect a transducer or sensor: • The appropriate measurement is shown on the display. For fetal measurements using an Avalon CTS system, the symbol appears additionally next to the measurement label, indicating that the measurement is being made by a cordless transducer. Measurement label Cordless measurement symbol •… -
Page 42
3 Basic Operation • The recorder prints an annotation showing the date, time, paper speed, and monitoring mode. It repeats this every 10 minutes. Checking/Setting Paper Scale Scale Type Internat’l Fetal You can check the paper for USA, or for other geographies) in the Recorder menu. -
Page 43
3 Basic Operation A — Protrusion holds paper guide in closed position Release the paper guide from one side of the holder..then remove the paper guide. -
Page 44
Refitting is a reversal of the removal procedure. Loading Paper: FM20/FM30 CAUTION Using recorder paper that is not approved by Philips can result in accelerated paper fading and can damage the thermal line printhead. This type of damage is not covered by warranty. FM20/30… -
Page 45
Off ” on page 49 for details). Loading Paper: FM40/FM50 CAUTION Using recorder paper that is not approved by Philips can result in accelerated paper fading and can damage the thermal line printhead. This type of damage is not covered by warranty. FM40/50… -
Page 46
3 Basic Operation A — Protrusion holds paper guide in closed position Prepare to place the new pack of paper in the tray with the bottom side down. The bottom side is indicated by the word STOP printed on the final page of the new pack. Unfold the top page of the pack and position the uterine activity scale on the right. -
Page 47
3 Basic Operation Now close the paper guide. Start/ Stop Start Recordng Press the recorder SmartKey or the SmartKey to switch on the recorder. Annotations of trace information are printed on the trace paper (see “Switching the Recorder On and Off ” on page 49 for details). Paper-Out Indication Each pack of paper has 150 pages. -
Page 48
3 Basic Operation Choosing Paper Speed You can choose a paper speed of 1, 2, or 3 centimeters per minute (cm/min). The default setting is 3 cm/min. The ACOG technical bulletin on FHR monitoring states that «accurate pattern recognition is difficult if not impossible at 1 cm/min and that 1 cm/min is only recommended for more economic screening. -
Page 49
3 Basic Operation Using the Paper Guide To tear off the trace paper after monitoring using the paper guide: If the recorder is running (the «recorder on» status indicator is displayed), turn off the Start/ Stop Stop Recordng recorder by selecting the fetal recorder SmartKey or the SmartKey. -
Page 50
3 Basic Operation When you switch on: • The «recorder on» status indicator is displayed in the bottom right-hand corner of the screen. • The paper advances quickly for 2 cm and then returns to the set speed. • Whenever the recorder is switched on, a trace header is printed vertically on the trace paper, containing the following: Selftest: OK –… -
Page 51
3 Basic Operation The monitor prints the time, date, paper speed and monitoring modes in the trace header when first switched on, in a periodic time stamp every ten minutes after, and if the monitoring modes change. The time stamp begins with the symbol. -
Page 52
3 Basic Operation To mark an event on the trace paper you can: Set Marker • Either select the SmartKey • Or press the button on the remote event marker. The remote event marker is connected to the monitor via any fetal transducer socket. A small arrow is printed on the heart rate scale on the trace paper. -
Page 53: Starting Monitoring
3 Basic Operation MD Notified Sitting On Back Left Lateral Ambulating Tocolytic Given Membranes Ruptured Amniotomy Amniotic Fluid Clear Amniotic Fluid Not Clear Oxytocin Urinary Catheter Micro Blood Analysis Up to two notes can be printed directly, and the monitor can temporarily store up to a further two notes, and these are printed after the first two have been recorded.
-
Page 54: Switching The Monitor To Standby
3 Basic Operation Switching the Monitor to Standby To switch the monitor to Standby: Either Monitor Standby Select the SmartKey Main Setup Enter the menu using the SmartKey Monitor Standby Select Pressing any key or selecting any field on the screen will resume monitoring. After Monitoring Discharge the patient.
-
Page 55: Troubleshooting
3 Basic Operation Troubleshooting Problem Possible Causes Solutions Light or no trace. Wrong paper. Use recommended paper. Dirty printhead. Clean printhead. See “Cleaning the Printhead” on page 177. FM20/30 only: Paper misaligned due to Shut the drawer fully, pushing evenly with drawer not being correctly shut.
-
Page 56
3 Basic Operation… -
Page 57: Using Batteries (With Fm20/30 Battery Option)
Using Batteries (with FM20/30 Battery Option) You can switch between battery-powered and mains-powered (AC) operation without interrupting monitoring. The monitor is connected to the AC mains power via the external power supply. NOTE The Battery Option is not available for the FM40/FM50.
-
Page 58: External Power Supply M8023A
4 Using Batteries (with FM20/30 Battery Option) External Power Supply M8023A The external power supply M8023A (option #E25) allows you to operate the fetal monitor from an AC (alternating current) power source of 100 V to 240 V (± 10%) and 50/60 Hz (± 5%). If this option is used, then the M8023A (option #E25) power supply is included for FM20/30.
-
Page 59
4 Using Batteries (with FM20/30 Battery Option) Battery LED Colors If the monitor is connected to If the monitor is running on mains power, this means battery power, this means battery power is > 90% Green battery charging (battery power Yellow <… -
Page 60: Battery Status Window
4 Using Batteries (with FM20/30 Battery Option) Battery Status Indicator Battery Malfunction Indicator Alternates with the battery gauge on the The red ! flashes. Critical battery Main Screen. situation or malfunction. Check in the Battery Statuswindow Check in the Battery Status window to see which malfunction to see which status symbol is displayed to indicator is displayed, or refer to…
-
Page 61: Optimizing Battery Performance
4 Using Batteries (with FM20/30 Battery Option) Battery status symbols Battery malfunction symbols battery is empty (red) incompatible battery battery not charging as the temperature (red) battery malfunction is above or below the specified range battery requires maintenance (red) battery has no power left charging stopped to protect the battery (red) battery temperature too high (red) battery is missing, insert battery…
-
Page 62
Battery Safety Information WARNING Use only Philips batteries part number M4605A. Use of a different battery may present a risk of fire or explosion. Do not open batteries, or dispose of them in fire, or cause them to short circuit. They may ignite, explode, leak or heat up, causing personal injury. -
Page 63: Alarms
Alarms The alarm information here applies to all measurements. Measurement-specific alarm information is contained in the sections on individual measurements. The fetal monitor has three alarm levels: red, yellow, and INOP. Red and yellow alarms are patient alarms. A red alarm indicates high priority, such as a potentially life threatening situation (for example, SpO below the desaturation alarm limit).
-
Page 64: Visual Alarm Indicators
The audible alarm indicators of your monitor are configurable. In the monitor’s Configuration Mode, you can change the alarm sound to suit the different alarm standards valid in different countries. Standard Philips Alarms • Red alarms: A high pitched sound is repeated once a second.
-
Page 65
5 Alarms • Yellow alarms: A lower pitched sound is repeated every two seconds. • INOPs: an INOP tone is repeated every two seconds. ISO/IEC Standard 9703-2 Audible Alarms • Red alarms: A high pitched tone is repeated five times, followed by a pause. •… -
Page 66: Acknowledging Alarms
5 Alarms Acknowledging Alarms Silence To acknowledge all active alarms and INOPs, select the . This switches off the audible alarm indicators. A check mark beside the alarm message indicates that the alarm has been acknowledged. If the condition that triggered the alarm is still present after the alarm has been acknowledged, the alarm message stays on the screen with a check mark symbol beside it.
-
Page 67
5 Alarms Alarms Off Select the SmartKey. Main Setup Select Alarms Select Alarms Off Select To Switch Individual Measurement Alarms On or Off This applies to alarm mode Select the measurement numeric to enter its setup menu. Alarms Select to toggle between The alarms off symbol is shown beside the measurement numeric. -
Page 68: Alarm Limits
5 Alarms Alarm Limits The alarm limits you set determine the conditions that trigger yellow and red limit alarms. WARNING Be aware that the monitors in your care area may each have different alarm settings, to suit different scenarios. Always check that the alarm settings are appropriate before you start monitoring. Viewing Individual Alarm Limits (Alarm Mode «All»…
-
Page 69: Latching Alarms
5 Alarms Review Alarms Window Alarms On SpO₂ NON-PULSAT. Alarms Silenced ** FHR1 HIGH ** SpO₂ LOW Review Alarms window contains a list of up to 300 of the most recent alarms and INOPs with date and time information. If configured to do so, each alarm is shown with the alarm limit active Review Alarms when the alarm was triggered and the maximum value measured beyond this limit.
-
Page 70
5 Alarms then audible latching can only be set to . The following table shows the possible combinations for latching settings: Possible Combinations for Alarm Latching Settings Visual Latching Setting Audible Latching Setting Red and Yellow Red and Yellow Red and Yellow Red and Yellow Alarm Latching Behavior Alarm Condition… -
Page 71: Testing Alarms
5 Alarms Testing Alarms In general, to test the functioning of visible and audible alarms, do the following: Enable the alarm. Set the alarm limits. Measure or simulate the parameter that is out of range, or signal loss. Verify that the visible and audible alarms are working. As an example, to test the FHR alarms: Connect the US transducer to a fetal sensor socket.
-
Page 72
5 Alarms… -
Page 73: Patient Alarms And Inops
Patient Alarms and INOPs This chapter lists alarms and technical alarms (INOPs) for the fetal monitors, alphabetically, irrespective of their priority. INOPs start “Technical Alarm Messages (INOPs)” on page 76. Alarm Messages Fetal alarms are identified by either «FHR» or «DFHR». All other alarms without these identifiers refer to maternal parameters.
-
Page 74
6 Patient Alarms and INOPs Alarm Message From Condition Indication *** EXTREME BRADY MECG The maternal heart rate obtained Numeric flashes, red alarm from the maternal ECG has fallen message, alarm tone. ***BRADY xxx < yyy below the extreme bradycardia limit. xxx denotes the lowest measured value, and yyy is the extreme bradycardia limit. -
Page 75
6 Patient Alarms and INOPs Alarm Message From Condition Indication ** Pulse HIGH The pulse rate has exceeded the high Numeric flashes and high alarm limit. limit is highlighted, yellow alarm message, alarm tone. ** Pulse LOW The pulse rate has dropped below Numeric flashes and low limit the low alarm limit. -
Page 76: Technical Alarm Messages (Inops)
Check that there is no paper jam, that the print drawer is properly shut, that the paper is loaded with the grid facing INOP tone upwards, and that the correct Philips paper is being used. Check Settings Monitor If this INOP appears, check the monitor and patient settings before you resume monitoring.
-
Page 77
6 Patient Alarms and INOPs INOP Message, Indication Source What to do CHECK TI CONFIG Monitor If this INOP appears, the monitor can be used normally, except for the Trace Interpretation feature. Contact your service personnel. Check Touch Input Monitor Perform a visual and functional check of the touch input device. -
Page 78
6 Patient Alarms and INOPs INOP Message, Indication Source What to do FetRec CHK CONF. Recorder Speed Scale Type Recorder and/or settings are set to «Unknown» and need to be set to the correct values in INOP tone Configuration Mode before the recorder can be operated. FHR1 EQUIP MALF There is a problem with the FHR hardware. -
Page 79
6 Patient Alarms and INOPs INOP Message, Indication Source What to do NBP MEASURE FAILED Noninvasive Check that you are using the correct cuff size and placement, blood pressure and that the correct patient category is selected. Try restarting Numeric is replaced by a -?- the measurement. -
Page 80
6 Patient Alarms and INOPs INOP Message, Indication Source What to do SpO₂ NOISY SIGN. Excessive patient movement or electrical interference is causing Numeric is replaced by a -?- irregular pulse patterns. Try to reduce patient movement or to INOP tone. relieve the cable strain on the sensor. -
Page 81
6 Patient Alarms and INOPs INOP Message, Indication Source What to do Toco UNPLUGGED Toco Reconnect the Toco transducer to the monitor. Check all connections are sound. INOP tone Unsupported LAN Monitor There is a problem with the communication to the network and central monitoring is currently not possible. -
Page 82
6 Patient Alarms and INOPs… -
Page 83: Admitting And Discharging
Admitting and Discharging The fetal monitor can store basic patient demographic information used to identify patients. Admit/Discharge on the Monitor This section describes how you admit and discharge patients when using the monitor as a stand-alone device (that is, when not used with a obstetrical information and surveillance system such as OB TraceVue).
-
Page 84: New Patient Check
7 Admitting and Discharging Editing Patient Information To edit the patient information after a patient has been admitted, select the patient name field on the Patient Demographics main screen of the fetal monitor to open the window, and make the required changes.
-
Page 85: Ob Tracevue: Via Lan
7 Admitting and Discharging OB TraceVue: via LAN Both the monitor from which you are transferring a patient and the monitor to which you are transferring her must be switched on and connected to the OB TraceVue network during the patient transfer.
-
Page 86
7 Admitting and Discharging… -
Page 87: Non-Stress Test Timer
Non-Stress Test Timer The non-stress test ( ) timer shows the elapsed time for the non-stress test. The timer counts up to the time you set for the NST. Setting NST Autostart/Autostop NST Autostart You can set the recorder so that it starts automatically ( ) when the NST timer is started, NST Autostop and stops automatically (…
-
Page 88: Accessing The Nst Setup Pop-Up Keys
8 Non-Stress Test Timer Accessing the NST Setup Pop-up Keys You control and set up the NST timer (for example, start, stop, or clear the timer, and set the run time) using a selection of pop-up keys that you access via any one of three possible routes: Timer •…
-
Page 89
8 Non-Stress Test Timer Run Time Setup The run time can be set from 10 to 60 minutes. To set the run time, you first need to enter the menu: Setup NST To enter the menu: Either Main Setup a. Enter the menu using the SmartKey . -
Page 90
8 Non-Stress Test Timer… -
Page 91: Non-Stress Test Report
Non-Stress Test Report It is generally accepted that a non stress test ( ) allows you to assess fetal well-being. The monitor’s NST report process uses fetal ultrasound (but not DECG) heart rate traces and the Maternal Toco Trace, generating a printed report when criteria are met and the test can be considered reassuring. The American term Non Stress Test (NST) is used for antepartum testing.
-
Page 92: Nst Report Status Window
9 Non-Stress Test Report Report Recording • . Choose from: Manual Record Report – — press the pop up key to trigger a manual request. After Recorder Stop – — report is recorded as soon as recorder becomes idle Immediately –…
-
Page 93
9 Non-Stress Test Report Example NST Report Field Field Content Report Title, with FHR NST Report for FHR1 on 12 Oct 2009 label and date Product Information Product DE53102345 G.01.70, OB A.04.24, Toco DE52401090, FHR1 DE00002345 A.05.26 Patient Information Rogers, Alice Age: 27 Gestational Age: Week 34, Day 5 Start time, end time,… -
Page 94: Nst Reassurance Criteria
9 Non-Stress Test Report NST Reassurance Criteria The patient is monitored for a user-definable period of time (10 — 60 minutes in steps of 5 minutes). The test is considered reassuring when the following criteria are met: • The fetal heart rate is valid at least 90% (this is configurable) of the specified time span. •…
-
Page 95: Monitoring Fhr And Fmp Using Ultrasound
Monitoring FHR and FMP Using Ultrasound To monitor a single FHR externally, you use an ultrasound transducer attached to a belt around the mother’s abdomen.The ultrasound transducer directs a low-energy ultrasound beam towards the fetal heart and detects the reflected signal. Your monitor can also detect fetal movements and print the fetal movement profile ( ) on the trace.
-
Page 96: Limitations Of The Technology
10 Monitoring FHR and FMP Using Ultrasound Limitations of the Technology All tissues moving towards or away from the transducer generate Doppler echoes. Therefore, the resulting signal that is provided to the monitor’s speaker, and for further fetal heart signal processing, can contain components of the beating fetal heart wall or valves, fetal movements, fetal breathing or hiccup, maternal movements such as breathing or position changes, and pulsating maternal arteries.
-
Page 97: Cross-Channel Verification
10 Monitoring FHR and FMP Using Ultrasound Cross-Channel Verification To reduce the possibility of mistaking the MHR for FHR, it is recommended that you monitor both maternal and fetal heart rates (see “Monitoring Maternal Heart / Pulse Rate” on page 155). The monitor’s cross-channel verification (CCV) facility can help by automatically detecting when the same heart rate is being recorded by different transducers.
-
Page 98: Preparing To Monitor
Apply a thin layer of ultrasound gel to the underside of the transducer. CAUTION Using ultrasound gel not approved by Philips may reduce signal quality and may damage the transducer. This type of damage is not covered by warranty. •…
-
Page 99: Selecting Fetal Heart Sound
10 Monitoring FHR and FMP Using Ultrasound WARNING Periodically compare the mother’s pulse with the signal coming from the monitor’s loudspeaker to ensure that you are monitoring fetal heart rate. Do not mistake a doubled or elevated MHR for FHR. Note that when applied to the patient, the ultrasound transducer may warm slightly (less than 1°C/ 1.8°F above ambient temperature).
-
Page 100: Fetal Movement Profile
10 Monitoring FHR and FMP Using Ultrasound Select the required volume from the volume scale. Alarm Volume Fetal Movement Profile The Fetal Movement Profile (FMP) parameter detects fetal movements via an ultrasound transducer connected to the monitor. Only the fetus monitored on the FHR1 channel is monitored for FMP. Once you have enabled FMP (see “Switching FMP On and Off”…
-
Page 101
10 Monitoring FHR and FMP Using Ultrasound FMP Statistics FMP statistics are printed every ten minutes. FMP started here FMP enabled The FMP statistics are presented as two percentage figures: • The first figure shows the percentage of detected fetal movements in the previous ten minutes (see B above). -
Page 102: Troubleshooting
10 Monitoring FHR and FMP Using Ultrasound Troubleshooting Problem Possible Causes Solutions Erratic trace. Fetal arrhythmia. Consider monitoring FHR using DECG after the rupture of membranes. Obese patient. Erratic display. Transducer position not optimal. Reposition transducer until signal quality indicator shows a good signal (at least half- full).
-
Page 103: Additional Information
10 Monitoring FHR and FMP Using Ultrasound Problem Possible Causes Solutions FHR1 EQUIP MALF FHR2 EQUIP MALF FHR3 EQUIP MALF INOP (see “Patient Alarms and INOPs” on displayed. page 73). FHR1 SIGNAL LOSS FHR2 SIGNAL LOSS FHR3 SIGNAL LOSS INOP displayed. FHR1 UNPLUGGED FHR2 UNPLUGGED FHR3 UNPLUGGED…
-
Page 104
It is strongly recommended that the maternal pulse oximeter or maternal ECG be employed to make use of the monitor’s Cross-Channel Verification (CCV) feature, especially during the second stage of labor or when the maternal pulse is elevated over 100 bpm. The Philips Avalon fetal monitor offers maternal pulse oximetry (SpO ) and maternal ECG for maternal pulse detection and the creation of a maternal heart rate pattern plotted on the same recorder as the FHR pattern. -
Page 105
10 Monitoring FHR and FMP Using Ultrasound “Noisy/Erratic” signals: With mixed or weak signals the tracing may reveal very brief episodes of erratic recorded traces. These represent the autocorrelation algorithm finding brief sequences of apparent and persistent heartbeats amidst a mixed or weak signal. These erratic recorded traces are commonplace, especially in association with fetal or maternal movement. -
Page 106
10 Monitoring FHR and FMP Using Ultrasound Audio Output The audio output from the device is an aural representation of movement that, in most cases, permits accurate auscultation of the FHR corresponding to the FHR displayed on the monitor and rate pattern depicted on the trace recording. -
Page 107
10 Monitoring FHR and FMP Using Ultrasound fetuses are monitored, the Avalon will identify and label apparent CCV between either two fetuses or a fetus and the mother. FHR1 Pulse FHR1 Pulse Avalon CCV numeric display and recorded trace showing coincident alerts When monitoring the maternal ECG, a beat-to-beat maternal heart rate trace is printed alongside the FHR recorded trace. -
Page 108
10 Monitoring FHR and FMP Using Ultrasound Half-Counting Baseline Rate Baseline Variability Moderate Accelerations Present Decelerations Not apparent Contractions Minimal Artifact Half-counting, noise, drop out Comment Reassuring tracing. The half-count at 4-5 minutes into the tracing may simulate a fetal deceleration, but the abruptness and the lack of any compensatory changes when the normal rate returns suggests that this is half-counting. -
Page 109
10 Monitoring FHR and FMP Using Ultrasound Maternal-Switching (Maternal Insertion) Baseline Rate 170 — Tachycardia Baseline Variability Moderate Accelerations Unable to determine Decelerations Absent Contractions Absent Artifact Maternal insertion, noise Comment The fetus has an elevated baseline rate of about 170 bpm with minimal to moderate variability. -
Page 110
10 Monitoring FHR and FMP Using Ultrasound Noisy/Erratic Signal and Dropout Contractions Minimal Artifact Noisy signal, drop-out Comment Reassuring tracing. Note that there is episodic drop out of the signal with discontinuity of the fetal tracing. Remediation Either improving the position of the transducer or the application of a fetal scalp electrode will reduce the amount of artifact in the tracing. -
Page 111
10 Monitoring FHR and FMP Using Ultrasound Selection of Literature References on Artifacts ___________________________________________________________ Fetal Monitoring, A Multidisciplinary Approach, Sixth edition Susan M. Tucker, Lisa A. Miller, David A. Miller Mosby Inc. 2009, ISBN 978-0-323-05670-0 ___________________________________________________________ Signal ambiguity resulting in unexpected outcome with external fetal heart rate monitoring By Duncan R. -
Page 112: Testing Ultrasound Transducers
If any of the following tests fail, repeat the test using another transducer. If the second transducer passes the tests, confirming that the first transducer is defective, contact your service personnel. If the second transducer also fails the tests, contact your Philips Service Engineer or Response Center. To test an ultrasound transducer: Switch on the monitor and the recorder.
-
Page 113: Monitoring Twin Fhrs
Monitoring Twin FHRs You can monitor twin FHRs externally using two ultrasound transducers. It is not possible to monitor twins externally using cordless ultrasound transducers. Additionally, you can monitor twin FHRs throughout labor and delivery after rupture of the FM30/50 membranes by monitoring one twin externally using ultrasound and the other internally using DECG.
-
Page 114: Monitoring Twins Externally
11 Monitoring Twin FHRs for FM40/50. • The trace recorded for FHR1 is thicker (darker) than that recorded for FHR2. This ensures that the two heart rates are easily distinguishable. The thickness of the recorded trace can be changed in Configuration Mode.
-
Page 115: Monitoring Twins Internally
11 Monitoring Twin FHRs Monitoring Twins Internally FM30/50 Monitor one twin using the procedures described in the Chapter “Monitoring FHR and FMP Using Ultrasound” on page 95. Monitor the second twin using the procedures described in the Chapter “Monitoring FHR Using DECG” on page 137. Example of a screen showing twin monitoring using a combination of US and DECG (The fetal heart rate monitored via DECG is labeled «DFHR1″/»DFHR2″/»DFHR3″…
-
Page 116: Separating Fhr Traces
11 Monitoring Twin FHRs Separating FHR Traces To help you to interpret traces with similar baselines, you can separate the baselines by an offset of 20 bpm by switching trace separation on. For details of the offset, see “Determining the Separation Order”…
-
Page 117
11 Monitoring Twin FHRs When Trace Separation is On When trace separation is turned on, the recorder prints a dotted line labeled with the two FHRs at the top, and at the bottom. Examples of the two methods (Standard, Classic) for determining the trace separation order are provided here. -
Page 118
11 Monitoring Twin FHRs Only the FHR1 trace is offset. The numerical FHR value displayed on the monitor remains unchanged. Subtract 20 from the recorded trace for FHR1 to obtain the true FHR1 value. For example, if the recorded trace shows 160, then the true FHR is 140. Switching Trace Separation On and Off Connect three ultrasound transducers to the monitor to measure FHR. -
Page 119: Troubleshooting
11 Monitoring Twin FHRs Troubleshooting Common problems that may occur when monitoring FHR using ultrasound are listed in “Monitoring FHR and FMP Using Ultrasound” on page 95. See also “Monitoring FHR Using DECG” on page 137 for common problems you might encounter when monitoring FHR directly. The following problem may occur when monitoring twins.
-
Page 120
11 Monitoring Twin FHRs… -
Page 121: Monitoring Triple Fhrs
Monitoring Triple FHRs If your monitor is equipped with the triplets option, it carries the label. You can monitor triple FHRs externally using three ultrasound transducers. Triplets monitoring is not possible using the Avalon CTS Cordless Fetal Transducer System. Refer to the appropriate preceding chapters for contra-indications and other information about the measurement methods you have chosen.
-
Page 122: Monitoring Triplets
12 Monitoring Triple FHRs for FM40/50. • The trace recorded for the FHR3 is thicker (darker) than that recorded for FHR1, which is thicker than that for FHR2. This ensures that the three heart rates are easily distinguishable. The thickness of the recorded trace can be changed in Configuration Mode.
-
Page 123: Separating Fhr Traces
12 Monitoring Triple FHRs Separating FHR Traces To help you to interpret traces with similar baselines, you can separate the baselines by an offset of 20 bpm by switching trace separation on. For details of the offset, see “Determining the Separation Order”…
-
Page 124: When Trace Separation Is Off
12 Monitoring Triple FHRs • The label for FHR1 is annotated with and the FHR3 label is annotated with Trace Separation Classic The following trace shows triplets with on, and using separation order. The traces for FHR1 and FHR3 are offset. The numerical FHR values displayed on the monitor remain unchanged.
-
Page 125: Troubleshooting
12 Monitoring Triple FHRs Troubleshooting Common problems that may occur when monitoring FHR using ultrasound are listed in “Monitoring FHR and FMP Using Ultrasound” on page 95. The following problem may occur when monitoring triplets. Problem Possible Cause Solution More than one transducer is recording Reposition one or more ultrasound is printed repeatedly.
-
Page 126
12 Monitoring Triple FHRs… -
Page 127: Fetal Heart Rate Alarms
Fetal Heart Rate Alarms Fetal heart rate (FHR) alerting can give both audible and visual warning of a non-reassuring fetal condition. Your monitor must be configured to alarm mode to enable the FHR alerting (see “Alarms” on page 63). Changing Alarm Settings When you do any of the following actions for any FHR measurement channel, this applies for all active FHR measurements, both ultrasound and DECG: •…
-
Page 128: Changing Signal Loss Delay
13 Fetal Heart Rate Alarms Changing Alarm Delays You can change the alarm delays in Configuration Mode. Connect either an ultrasound or a DECG transducer to a free socket on the monitor. Enter the Setup Menu for a connected FHR measurement. High Delay To change the high alarm limit delay time, select and select the delay time (in seconds)
-
Page 129: Monitoring Uterine Activity Externally
Monitoring Uterine Activity Externally You can measure uterine activity externally using a Toco transducer. You can also use a Toco Toco MP transducer for the same purpose, although they also have wider (ECG/IUP and Pulse) capabilities. The external Toco transducer measures the frequency, duration and relative strength of contractions, but not their absolute intensity.
-
Page 130: External Toco Monitoring
14 Monitoring Uterine Activity Externally External Toco Monitoring Prepare for Toco monitoring using the list below. The standard procedures in use in your facility determine the sequence of actions. • Fasten the abdominal transducer belt around the patient. • Connect the Toco transducer to a free socket on the monitor. The Toco baseline is automatically reset.
-
Page 131: Troubleshooting
14 Monitoring Uterine Activity Externally Troubleshooting External Toco Monitoring Problem Possible Causes Solutions Quality of the trace deteriorates The belt is incorrectly fastened and is The belt must be tight enough to ensure or the Toco baseline varies. too slack or too tight or the belt has lost good contact between the patient’s skin its elasticity.
-
Page 132: Testing Toco Transducers
If any of the following tests fail, repeat the test using another transducer. If the second transducer passes the tests, confirming that the first transducer is defective, contact your service personnel. If the second transducer also fails the tests, contact your Philips Service Engineer or Response Center. To test a Toco transducer: Switch on the monitor and the recorder.
-
Page 133: Monitoring Uterine Activity Internally
Monitoring Uterine Activity Internally FM30/50 You can monitor intrauterine pressure (IUP) using an intrauterine catheter together with a patient module or a Toco transducer, after rupture of the membranes and the cervix is sufficiently dilated. What You Need Illustration 1 shows the complete connection chain from the IUP catheter to the fetal monitor using the patient module: A Disposable Koala IUP Catheter B Reusable Koala IUP Adapter Cable (9898 031 43931)
-
Page 134: Internal (Iup) Monitoring
15 Monitoring Uterine Activity Internally C Toco Transducer (M2735A) Internal (IUP) Monitoring Read the instructions that accompany the intrauterine catheter and the adapter cable before you start monitoring. Zero the monitor when instructed. WARNING Do not catheterize if placenta previa is diagnosed or if uterine bleeding from an undetermined source is present.
-
Page 135: Troubleshooting
15 Monitoring Uterine Activity Internally Troubleshooting Internal (IUP) Monitoring Problem Possible Causes Solutions No change in pressure during Dry environment or possible extra- Please refer to catheter Instructions for Use. contraction. ovular placement of sensor tip. Only pressure peaks can be Zero adjustment is incorrect.
-
Page 136
15 Monitoring Uterine Activity Internally… -
Page 137: Monitoring Fhr Using Decg
Monitoring FHR Using DECG FM30/50 This chapter describes how to monitor a single fetal heart rate via direct ECG (DECG), using a spiral fetal scalp electrode in the intrapartum period. Read and adhere to the instructions that accompany the fetal scalp electrode, the DECG adapter cable, and the attachment electrode.
-
Page 138
16 Monitoring FHR Using DECG Illustration 1 shows the complete connection chain from the fetal scalp electrode to the fetal monitor using the Toco transducer. A Fetal Scalp Electrode, single spiral (9898 031 37631) B Fetal Scalp Electrode, double spiral, Europe only, not for USA (9898 031 37641) C DECG Adapter Cable (9898 031 37651) D Pre-gelled Attachment Electrode (9898 031 39771) E Toco+ Transducer (M2735A) -
Page 139: Making Connections
16 Monitoring FHR Using DECG Making Connections WARNING Follow the instructions supplied with each of the monitoring accessories you are using. Prepare for DECG monitoring using the list below. The standard procedures in use in your facility determine the sequence of actions. •…
-
Page 140: Suppressing Artifacts
16 Monitoring FHR Using DECG being monitored, both waves are displayed, with the DECG wave above the MECG wave. The MECG MECG wave is labeled Check the artifact suppression setting and change it if necessary (see “Suppressing Artifacts” on page 140). Measurement label (DFHR1) Measurement label (Toco) Additional measurement label…
-
Page 141: Troubleshooting
16 Monitoring FHR Using DECG Troubleshooting Problem Possible Cause Solutions DFHR1 EQUIP MALF See “Patient Alarms and INOPs” on DFHR2 EQUIP MALF page 73. DFHR3 EQUIP MALF DFHR1 LEADS OFF Spiral electrode detached at connector. Reconnect the fetal scalp electrode. DFHR2 LEADS OFF DFHR3 LEADS OFF Poor or no contact between leg…
-
Page 142: Testing Decg Mode
16 Monitoring FHR Using DECG Testing DECG Mode Refer to the monitor’s Service Guide.
-
Page 143: Monitoring Noninvasive Blood Pressure
Monitoring Noninvasive Blood Pressure This fetal monitor uses the oscillometric method for the noninvasive blood pressure measurement. In adult mode, the blood pressure measurements determined with this device comply with the American National Standard for Electronic or Automated Sphygmomanometers (ANSI/AAMI SP10-1992) in relation to mean error and standard deviation, when compared to intra-arterial or auscultatory measurements (depending on the configuration) in a representative sample population.
-
Page 144: Preparing To Measure Noninvasive Blood Pressure
Plug the air tubing into the red NBP connector. Avoid compression or restriction of pressure tubes. Air must pass unrestricted through the tubing. Make sure that you are using a Philips-approved correct sized cuff and that the bladder inside the cover is not folded or twisted.
-
Page 145
17 Monitoring Noninvasive Blood Pressure A wrong cuff size, and a folded or twisted bladder, can cause inaccurate measurements. The width of the cuff should be in the range from 37% to 47% of the limb circumference. The inflatable part of the cuff should be long enough to encircle at least 80% of the limb. -
Page 146: Starting And Stopping Measurements
17 Monitoring Noninvasive Blood Pressure The NBP timestamp will normally show the completion time of the NBP measurement. Only under the following conditions the timestamp shows the beginning of the measurement: Auto Sequence • when in mode, and • the monitor is configured to synchronize the measurements in a measurement series to an «easy- Repeat Time to-document»…
-
Page 147: Choosing The Alarm Source
17 Monitoring Noninvasive Blood Pressure NOTE Be aware that a combination of a recorder speed of less than 3 cm/min and a repetition time of less than five minutes can result in not all noninvasive blood pressure measurements being recorded on the fetal trace.
-
Page 148: Calibration
17 Monitoring Noninvasive Blood Pressure Calibration NBP is not user-calibrated. Cuff-pressure transducers must be verified at least once every two years by a qualified service professional, and calibrated, if necessary. See the Service Guide for details. Troubleshooting Problem Possible Causes Solutions Cuff will not inflate.
-
Page 149: Monitoring Spo2
FM30/40/50 The pulse oximetry measurement (SpO ) is intended for use with maternal patients. Philips pulse oximetry uses a motion-tolerant signal processing algorithm, based on Fourier Artifact Suppression Technology (FAST). It provides two measurements: • Oxygen saturation of arterial blood (SpO ) — percentage of oxygenated hemoglobin in relation to the sum of oxyhemoglobin and deoxyhemoglobin (functional arterial oxygen saturation).
-
Page 150: Connecting Spo2 Cables
CAUTION Extension cables: Do not use more than one extension cable (M1941A). Do not use an extension cable with Philips reusable sensors or adapter cables with part numbers ending in -L (indicates «long» cable version). Electrical Interference: Position the sensor cable and connector away from power cables, to avoid…
-
Page 151: Measuring Spo2
18 Monitoring SpO2 Measuring SpO2 During measurement, ensure that the application site: – has a pulsatile flow, ideally with a signal quality indicator of at least medium. – has not changed in its thickness (for example, due to edema), causing an improper fit of the sensor.
-
Page 152: Spo2 Signal Quality Indicator
18 Monitoring SpO2 SpO2 Signal Quality Indicator The SpO numeric is displayed together with a signal quality indicator (if configured and enough space is available) which gives an indication of the reliability of the displayed values. The level to which the triangle is filled shows the quality of the signal; the indicator below shows a medium signal quality, the signal quality is at a maximum when the triangle is completely filled.
-
Page 153: Setting Up Tone Modulation
18 Monitoring SpO2 Alarm Delays There is a delay between a physiological event at the measurement site and the corresponding alarm at the monitor. This delay has two components: • The general system delay time is the time between the occurrence of the physiological event and when this event is represented by the displayed numerical values.
-
Page 154
18 Monitoring SpO2… -
Page 155: Monitoring Maternal Heart / Pulse Rate
Monitoring Maternal Heart / Pulse Rate You can monitor the maternal heart/pulse rate using one of four sources: • Maternal heart rate (MHR) via MECG electrodes • (pulse rate) • Maternal pulse from Toco MP transducer (pulse rate) • NBP (pulse rate) Maternal heart / pulse rates derived from MECG, SpO and Toco MP are continuous measurements, and are compared against the FHR for cross-channel verification.
-
Page 156: Mhr From Mecg Electrodes
19 Monitoring Maternal Heart / Pulse Rate beat MECG heart rate trace or an ultrasound heart rate trace calculation, having switched to the maternal rate with no significant delay. Note that Maternal Pulse from Toco has an averaging of 4 seconds and an overall delay of between 6 and 8 seconds.
-
Page 157
19 Monitoring Maternal Heart / Pulse Rate Applying Electrodes To derive the MHR (when you do not want to view the MECG waveform), you can place the electrodes just below the outer end of the clavicle near each shoulder. 1 MECG Electrodes Making Connections WARNING Follow the instructions supplied with each of the monitoring accessories you are using. -
Page 158: Monitoring Mecg Wave
19 Monitoring Maternal Heart / Pulse Rate Monitoring MECG Wave WARNING The fetal/maternal monitor is not a diagnostic ECG device. In particular, the display of fetal/maternal ECG is intended only for evaluating signal quality for fetal/maternal heart rate as derived from the ECG waveform.
-
Page 159: Pulse Rate From Spo2
19 Monitoring Maternal Heart / Pulse Rate Viewing the Waveform on the Screen DFHR1 parameter Toco parameter FHR2 parameter Numeric/measurement values For the FM30/50, the MECG wave is displayed automatically on the screen, labeled MECG. If DECG is also being monitored (FM30/50), and the DECG wave is configured to On, both waves are displayed, with the…
-
Page 160: Adjusting The Heart Rate / Pulse Alarm Limits
19 Monitoring Maternal Heart / Pulse Rate Adjusting the Heart Rate / Pulse Alarm Limits To adjust the pulse alarm limits: Setup SpO₂ Pulse (SpO₂) Setup Pulse (SpO₂) In the menu, select . This opens the menu. Pulse (SpO₂) Pulse (SpO₂) Ensure .
-
Page 161: Printing The Ecg Waveform
Printing the ECG Waveform FM30/50 You can print the ECG wave onto the trace paper. If you are monitoring both DECG and MECG, MECG both waves will be printed. The start of the wave recording is annotated above the wave with DECG 25 mm/sec for Maternal ECG, with…
-
Page 162
20 Printing the ECG Waveform DECG MECG Paper speed DECG and MECG waveforms When the recorder is on, there are two choices for printing the ECG wave: Separate • : This recording mode gives you a six-second ECG strip on the fetal trace paper in fast printout mode. -
Page 163
20 Printing the ECG Waveform The following trace shows both the DECG and MECG waveforms: Paper speed Date To make your choice: Main Setup Enter the menu by selecting the SmartKey Fetal Recorder Fetal Recorder Select to enter the menu. ECG Wave Separate Overlap… -
Page 164
20 Printing the ECG Waveform… -
Page 165: Paper Save Mode For Maternal Measurements
Paper Save Mode for Maternal Measurements Your monitor’s recorder features a Paper Save Mode, where maternal vital signs are recorded using less paper than during a normal trace recording. When Paper Save Mode is enabled, and if the recorder is stopped, it will start automatically to print data from maternal measurements as they occur, and then stop again to save paper.
-
Page 166
21 Paper Save Mode for Maternal Measurements… -
Page 167: Recovering Data
Note that the data in the memory is cleared when a software upgrade is performed. CAUTION Only use Philips paper. Using paper other than Philips paper may result in the failure to recover traces. Recovering Traces on Paper The monitor is able to recover traces by printing them out at a high speed from the monitor’s backup memory.
-
Page 168: Recovering Traces On An Ob Tracevue System
22 Recovering Data • The change back to a real-time recording from a trace recovery printout prompts the recording to restart. A new vertical trace header annotation consisting of the time, date and paper speed is printed, letting you see where the trace recovery printout ends, and where the real-time trace continues.
-
Page 169
22 Recovering Data The speed of the printout depends on the configured recorder speed and on the amount of trace data available. The fetal trace printed from the trace data contains all data from the real-time trace, with the exception of the maternal heart rate, the pulse numeric and the ECG wave. Information for scale type, trace separation and recorder speed are not stored in the trace memory, but is applied when the stored recording starts. -
Page 170
22 Recovering Data… -
Page 171: Care And Cleaning
Warranty does not cover damage caused by using unapproved substances or methods. Philips makes no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection. Consult your hospital’s Infection Control Officer or Epidemiologist. For comprehensive details on cleaning agents and their efficacy refer to “Guideline for Disinfection and…
-
Page 172
Cleaning and Disinfecting Clean and disinfect the Avalon FM20, FM30, FM40 and FM50 fetal monitors and the transducers M2734A, M2734B, M2735A, M2736A, and M2738A (including ECG adapter cables) after each use. Clean equipment before disinfecting. For other accessories, see “Cleaning and Disinfecting Monitoring Accessories”… -
Page 173
23 Care and Cleaning Recommended Cleaning Agents We recommend that you use one of the following disinfectants: Product Name Product Type Ingredients Isopropanol liquid Isopropanol 80% Bacillol® AF liquid, spray 100 g concentrate contains: Propan-1-ol 45.0 g; Propan-2-ol 25.0 g; Ethanol 4.7 g. Bacillol®25 liquid Ethanol 100 mg/g… -
Page 174
23 Care and Cleaning… -
Page 175
In case of problems: If you discover a problem with any of the equipment, contact your service personnel, Philips, or your authorized supplier. Electric shock hazard: Do not open the monitor housing. Refer all servicing to qualified service personnel. -
Page 176
Ensure that these tasks are carried out as indicated by the monitor’s maintenance schedule, or as specified by local laws, whichever comes sooner. Contact a Philips-qualified service provider if your monitor needs a safety or performance test. Clean and disinfect equipment to decontaminate it before testing or maintaining it. -
Page 177
24 Maintenance – Products containing solvent-based adhesives such as (but not exclusively) laminating film, transparent film or labels sensitive to pressure. To ensure long lasting legibility and durability of thermal printouts, store your documents separately in an air-conditioned place and use: •… -
Page 178
24 Maintenance Monitor: • There is no metal molded into the plastic parts and no metal sprays on the plastic. • All plastic parts with a weight greater than 10g (0.35 ounces) are marked with the ISO code for identification. •… -
Page 179
All accessories listed for the fetal monitor may not be available in all geographies. You can order parts, accessories and supplies from Philips supplies at www.medical.philips.com or consult your local Philips representative for details. All accessories and supplies listed here are reusable, unless indicated otherwise. -
Page 180
ECG reusable legplate adapter cable M1362B (QwikConnect Plus) QwikConnect Plus Solution ECG leg attachment electrode for DECG M1349A (NOT compatible with New Philips legplate adapter cable DECG Solution accessories) DECG fetal scalp electrode: single spiral, 15133E worldwide availability DECG fetal scalp electrode: double 15133D spiral, Europe only. -
Page 181
25 Accessories and Supplies DECG Accessories: Component Compatibility Use the following pictorial guide to check component compatibility for DECG accessories. Do NOT mix accessories from the New Philips DECG Solution (marked ) with those from the QwikConnect Plus Solution (marked… -
Page 182
25 Accessories and Supplies Adult/Pediatric Multi-Patient Comfort Cuffs and Disposable Cuffs Maternal Patient Limb Circumference Bladder Disposable Reusable cuff Tubing Category Width cuff Part No. Part No. Adult (Thigh) 42.0 — 54.0 cm 20.0 cm M1879A M1576A M1598B (1.5 m) Large Adult 34.0 — 43.0 cm 16.0 cm… -
Page 183
Tyco Healthcare. Some sensors may not be available in all countries. Do not use more than one extension cable with any sensors or adapter cables. Do not use an extension cable with Philips reusable sensors or adapter cables with part numbers ending in -L (indicates «Long» version). -
Page 184
Use only on adult patients with FM30/40/50. M1194AN Special Edition (SE) Ear sensor (1.5 m cable) for patients more than 40 kg. Philips disposable sensors. Not available in the USA. M1904B Identical to OxiMax MAX-A Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter cable… -
Page 185
25 Accessories and Supplies Product Number Description Comments Philips disposable sensors. Available worldwide. M1131A Adult/Pediatric finger sensor (0.45 m cable) Requires M1943A (1.0 m) or M1943AL (3.0 m) adapter cable Use only on adult patients with FM30/40/50. M1133A Adult/Infant/Neonatal (0.9 m cable) for patients >… -
Page 186
25 Accessories and Supplies Masimo SpO2 Accessories Product Description Philips Part Number Number MASIMO LNOP reusable sensors (No adaptor cable required) LNOP DC-I Adult Finger Sensor (> 30 kg) 989803140321 LNOP DC-IP Pediatric Finger Sensor (10 — 50 kg) 989803140331… -
Page 187
Extension / Adapter Cables Product Description Comments Number M1941A Extension cable (2 m) For use with Philips reusable sensors and adapter cables. M1943A Adapter cable (1.1 m cable) Adapter cable for Philips/ Nellcor disposable sensors. M1943AL Adapter cable (3 m cable) -
Page 188
25 Accessories and Supplies… -
Page 189
Specifications and Standards Compliance The monitors are intended to monitor a mother and her fetus(es), which from an electrical safety point of view, are one person. Environmental Specifications The monitor may not meet the given performance specifications if stored and used outside the specified temperature and humidity ranges. -
Page 190
26 Specifications and Standards Compliance Sensors Operating Temperature Range 0°C to 37°C (32°F to 98.6°F) WARNING Explosion Hazard: Do not use in the presence of flammable anesthetics, such as a flammable anesthetic mixture with air, oxygen or nitrous oxide. Use of the devices in such an environment may present an explosion hazard. -
Page 191
26 Specifications and Standards Compliance Transducers (M2734A/M2734B/M2735A/M2736A/M2738A) Dimensions and M2734A&B/35/36A Size (diameter) 83 mm/3.27 in Weight Weight (without cable) < 220 g/7.8 oz. M2738A Maximum size mm/ 50 x 28 x 135 (2.0 x 1.1 x 5.3 in) (in): width x height x depth Cable length 2.5 m… -
Page 192
26 Specifications and Standards Compliance Battery Specifications Performance Specifications Battery Operating Time Basic monitoring configuration: >2 hours (with new, fully charged (Display Brightness: 70%, Recorder: «On» at 3cm/ battery) min, NBP: Auto Mode at 15min, 2 US Transducers, 1 Toco+ with MECG, 1 Patient Module with DECG) Charge Time When monitor is off: approx. -
Page 193
26 Specifications and Standards Compliance Fetal / Maternal Specifications Performance Specifications Ultrasound Measurement Method Ultrasound Pulse Doppler Measurement Range 50 to 240 bpm Resolution Display 1 bpm Printer 1/4 bpm Jitter @ 200 bpm ≤ 3 bpm Display Update Rate 1 / second US Intensity Average output power… -
Page 194
26 Specifications and Standards Compliance Performance Specifications Auto Offset Correction 3 seconds after connecting the transducer, the Toco value is set to 20 units Auto Zero Adjust Toco value is set to zero following a negative measurement value for 5 seconds Maternal Pulse from Toco Emitted Light Energy ≤… -
Page 195
26 Specifications and Standards Compliance Performance Specifications Filter Bandwidth 0.8 to 80 Hz Inop Auxiliary Current (Leads Off Detection) < 100 µA Input Signal Range DECG 20 µVpp to 6 mVpp MECG 150 µVpp to 6 mVpp Dielectric Strength 1500 Vrms Defibrillator Protection None ESU Protection… -
Page 196
26 Specifications and Standards Compliance MECG Alarm Specifications Range Adjustment MECG Alarm Limits High Range: 31 to 240 1 bpm steps (30 to 40 bpm) 5 bpm steps (40 to 240 bpm) Default: 120 bpm Low Range: 30 to 235 Default: 50 bpm Tachycardia Difference to high limit: 0 to 50 bpm… -
Page 197
26 Specifications and Standards Compliance Fetal / Maternal Defaults Settings FHR (Ultrasound/DECG) Alarms On/Off Default Default Color for FHR Numeric Orange Toco Default color for Toco numeric Green Default IUP Scale Unit mmHg Default color for IUP numeric Green Maternal Heart Rate Default Color for MECG Numeric (MHR) Measurement Noninvasive Blood Pressure… -
Page 198
26 Specifications and Standards Compliance Measurement Validation: In adult mode, the blood pressure measurements determined with this device comply with the American National Standard for Electronic or Automated Sphygmomanometers (ANSI/AAMI SP10 — 1992) in relation to mean error and standard deviation, when compared to intra-arterial or auscultatory measurements (depending on the configuration) in a representative population. -
Page 199
M1191T, M1192T, M1194A, M1194AN, M1196A, M1196T = 3% values and the reference (70% to 100%) values Philips Disposable Sensors with M1943A(L): M1131A, M1901B, M1903B, M1904B = 3% (70% to 100%) M1133A, M1134A = ±2% (70% to 100%) NellcorPB® Sensors with M1943A(L):… -
Page 200
26 Specifications and Standards Compliance Alarm Specifications Range Adjustment Delay 50 to 100% 1% steps (0, 1, 2, 3,… 30) + 4 seconds Desat 50 to Low alarm limit 1% steps Pulse 30 to 300 bpm 1 bpm steps (30 to 40 bpm) max. -
Page 201
26 Specifications and Standards Compliance Recorder Specifications Built-in Thermal Array Fetal Trace Recorder Mechanism Thermal Array Recorder Paper & Printing Type Standard Z-fold paper Standard Speeds (real-time traces) 3 cm/min, 2 cm/min, 1cm/min Fast Print Speed (stored traces) Max. 20 mm/s Print speed is variable and depends on the print load ECG Wave Print Speed… -
Page 202
26 Specifications and Standards Compliance Recorder Default Settings Setting Choice Default Auto Start Confirmed Stop Bridge Paperout Paper Save Mode NST Autostart NST Autostop Trace Separation Separation Order Standard Classic Standard Intensity 1..5 Recorder Symbols Symbol Description Alarm is on (printed next to measurement label) Upper and lower alarm limit (printed next to measurement label) FMP detection is on Beginning of the date/time annotation… -
Page 203
26 Specifications and Standards Compliance Recorder Symbols Symbol Description Trace separation -20 bpm (in label) Trace separation Off (in trace) Trace separation +20 bpm (in trace) Trace separation -20 bpm (in trace) Trace separation +20 bpm and -20 bpm (in trace) Coincidence of heart rates is detected Marker Special wave, with different speed and scale (for example, fast printout of… -
Page 204
26 Specifications and Standards Compliance M4605A Battery Specifications Battery Type Smart Battery 10.8 V, 6000 mAh, Lithium Ion Safety complies with UL 2054 Electromagnetic Compatibility complies with the requirements for FCC Type B computing Device, and EN 61000-4-2 and EN 61000-3-2 (EMC) Communication Standard complies with the SMBus specification v 1.1… -
Page 205
Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 71034 Boeblingen Germany Visit our website at: www.philips.com © Copyright 2003 — 2011. Koninklijke Philips Electronics N.V. All Rights Reserved. Trademark Acknowledgement Oxisensor II, Oxi-Cliq , and OxiMax are trademarks of Tyco Healthcare Group LP, Nellcor Puritan Bennett Division. -
Page 206
26 Specifications and Standards Compliance Electromagnetic Compatibility (EMC) The device and its accessories, listed in the accessories section, comply with the following EMC standards: • EN/IEC 60601-1-2: 2001+A1:2004 Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment. You must operate your monitoring equipment according to the EMC information provided in this book. -
Page 207
26 Specifications and Standards Compliance Reducing Electromagnetic Interference CAUTION The device should not be used adjacent to, or stacked with, other equipment unless otherwise specified. The product and associated accessories can be susceptible to interference from continuous, repetitive, power line bursts, and other RF energy sources, even if the other equipment is compliant with EN 60601-1-2 emission requirements. -
Page 208
In the table below, the term «device» refers to the Avalon FM20/30/40/50 fetal monitor together with its accessories. The table gives details of the electromagnetic emissions, and how these are classified, for the device, and the electromagnetic environments in which the device is specified to technically function. -
Page 209
26 Specifications and Standards Compliance Electromagnetic Immunity The monitor is suitable for use in the specified electromagnetic environment. The user must ensure that it is used in the appropriate environment as described below. Table 2 — Guidance and Manufacturer’s Declaration: Electromagnetic Immunity Immunity Test IEC 60601-1-2 Compliance Level… -
Page 210
26 Specifications and Standards Compliance Table 3 — Guidance and Manufacturer’s Declaration: Electromagnetic Immunity Conducted RF Immunity Test EN/IEC 61000-4-6 Electromagnetic Environment Guidance: Recommended Separation Distance IEC 60601-1-2 Test Level Compliance Level over 150 kHz to 80 MHz (in Meters, at Frequency Range Tested) for Ultrasound and ECG Measurements… -
Page 211
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. If you require further information or assistance, please contact Philips Support. Recommended Separation Distances from Other RF Equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. -
Page 212
26 Specifications and Standards Compliance Table 5 — Separation Distance (d) in Meters According to Frequency of Transmitter at IEC 60601-1-2 Test Compliance Level 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz Rated Maximum Output Power (P) of Transmitter (in Watts) 0.01 W… -
Page 213
26 Specifications and Standards Compliance FM40/50 When power is lost — no power is available from the AC power source or from the battery — before the monitor is put into Standby, a beeper will sound. The tone can be silenced by pressing the On/Standby and FM20/ switch. -
Page 214
26 Specifications and Standards Compliance Symbols on the System This caution/warning symbol indicates that you should consult the Instructions for Use (this document), and particularly any warning messages. This symbol indicates that you should consult the Instructions for Use (this document). Power-On/Off Switch — FM20/FM30 without Battery Option Power-On/StandBy Switch — FM40/FM50 and FM20/30 with Battery Option Power-On LED… -
Page 215
26 Specifications and Standards Compliance Analog interface indicator for connection to any analog video display (VGA resolution). Video IP 67 Ingress Protection code according to IEC 60529. The IUP/ECG patient module (M2738A) is rated IP 67 (protection against dust, access to hazardous parts, and the effects of continuous immersion in water to a depth of 0.5 meter for 30 minutes). -
Page 216
26 Specifications and Standards Compliance… -
Page 217
SpO2 accessories cleaning monitor actions disposal of electronic waste Philips sensors (disposable) cleaning method disposal of monitor Philips sensors (reusable) cleaning monitoring accessories dyshemoglobins transducer accessories general cleaning requirements… -
Page 218
interference from RF equipment Standby starting monitoring internal monitoring factory defaults switching on intrauterine pressure. See IUP noninvasive blood pressure factory monitoring mode intravascular dyshemoglobins defaults monitoring triplets and FHR intravascular dyshemoglobins effects on recorder factory defaults (SpO2) monitoring twins and FHR SpO2 factory defaults ISO/IEC standard 9703-2 mounting information… -
Page 219
Philips sensors (reusable) signal quality reviewing alarms window paper SpO2 alarms default paper speed standards loading paper EMC standards safety paper out INOP safety standards maintenance interval paper speed standards compliance safety standards changing Standby default safety tests… -
Page 220
fixing transducer to belt IUP, zeroing Toco transducer touch tone volume Toco+ transducer warnings transducer belt clip transducer specifications transducer testing Toco ultrasound triplets cross-channel verification external monitoring importance of monitoring MHR monitoring triplets offsetting baselines troubleshooting DECG MECG troubleshooting FHR twins cross-channel verification external monitoring… -
Page 221
Part Number 453564290161 Printed in Germany 08/11 *453564290161*…
- About
- Blog
- Projects
- Help
-
Donate
Donate icon
An illustration of a heart shape - Contact
- Jobs
- Volunteer
- People
Bookreader Item Preview
texts
Philips Avalon FM20 FM30 Service Guide
Philips Avalon FM20 FM30 Service Guide
- Addeddate
- 2020-05-20 17:41:02
- Classification
- Clinical;Infant and Fetal Care;Fetal Monitor and Transducer;Philips Fetal Monitor and Transducer;Philips Avalon Series;Philips Avalon FM20 FM30 FM40 FM50
- Identifier
- manual_Philips_Avalon_FM20_FM30_Service_Guide
- Identifier-ark
- ark:/13960/t7rp1z09r
- Ocr
- ABBYY FineReader 11.0 (Extended OCR)
- Page_number_confidence
- 94.92
- Ppi
- 600
- Scanner
- Internet Archive Python library 1.9.0
comment
Reviews
There are no reviews yet. Be the first one to
write a review.
SIMILAR ITEMS (based on metadata)
Download
Table of Contents
Add to my manuals
Share
URL of this page:
HTML Link:
Bookmark this page
Manual will be automatically added to «My Manuals»
Print this page
- Manuals
- Brands
- Philips Manuals
- Monitor
- Avalon FM20
- Instructions for use manual
Fetal monitor
Hide thumbs
Also See for Avalon FM20:
- Service manual (261 pages)
1
2
Table Of Contents
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
-
page
of
221/
221 -
Contents
-
Table of Contents
-
Troubleshooting
-
Bookmarks
Table of Contents
Advertisement
5 Alarms
Alarm Mode
Visual Alarm Indicators
Audible Alarm Indicators
Acknowledging Alarms
Acknowledging Disconnect INOPs
Pausing or Switching Off Alarms
Alarm Limits
Reviewing Alarms
Latching Alarms
Testing Alarms
Alarm Behavior at Power On
6 Patient Alarms and INOPs
Alarm Messages
Technical Alarm Messages (INOPs)
7 Admitting and Discharging
Admit/Discharge on the Monitor
New Patient Check
OB TraceVue: via LAN
OB TraceVue: via RS232
8 Non-Stress Test Timer
Setting NST Autostart/Autostop
Viewing the NST Timer
Timer Expiry Notification
Accessing the NST Setup Pop-up Keys
9 Non-Stress Test Report
Setting Up an NST Report
NST Report Status Window
NST Reassurance Criteria
Non-Reassuring Report
Non-Reactive NST Test
10 Monitoring FHR and FMP Using Ultrasound
Technical Description
Limitations of the Technology
Misidentification of MHR as FHR
Cross-Channel Verification
What You Need
Cordless Monitoring — Important Considerations
Preparing to Monitor
Selecting Fetal Heart Sound
Changing the Fetal Heart Sound Volume
4
63
63
64
64
66
66
66
68
68
69
71
71
73
73
76
83
83
84
85
85
87
87
87
87
88
91
91
92
94
94
94
95
95
96
96
97
97
97
98
99
99
Table of Contents
Previous Page
Next Page
- 1
- 2
- 3
- 4
- 5
- 6
- 7
- 8
Advertisement
Table of Contents
Troubleshooting
-
Troubleshooting
55 -
Troubleshooting
102 -
Troubleshooting
119 -
Troubleshooting
125 -
Troubleshooting
131 -
Troubleshooting
135 -
Troubleshooting
141 -
Calibration
148 -
Adjusting the Heart Rate / Pulse Alarm Limits
160
Related Manuals for Philips Avalon FM20
-
Monitor Philips Avalon FM20 Service Manual
(261 pages)
-
Monitor Philips FM-2 Service Manual
Antepartum fetal monitor (110 pages)
-
Monitor Philips PHILIPS 190G6 User Manual
Multitainment monitor (91 pages)
-
Monitor Philips 226V3LSB/00 Service Manual
21.5″lcd color monitor chassis: meridian 3 (94 pages)
-
Monitor Philips 170C6FS/00 Service Manual
17 inch tft lcd colour monitor (78 pages)
-
Monitor Philips 223V5LSB/00 Service Manual
21.5″lcd color monitor (143 pages)
-
Monitor Philips Brilliance 499P9H User Manual
(39 pages)
-
Monitor Philips 190CW9FB/69 Service Manual
(88 pages)
-
Monitor Philips BDL4771V User Manual
(54 pages)
-
Monitor Philips 278E8 User Manual
(32 pages)
-
Monitor philips 193V5 User Manual
Tft-lcd monitor (35 pages)
-
Monitor Philips 10BDL3051T User Manual
(43 pages)
-
Monitor Philips Signage Solutions BDL4677XH/00 Brochure & Specs
Signage solutions videowall display 46″, direct led backlight, full hd, 700cd/m² (3 pages)
-
Monitor Philips Q-Line 55BDL3010Q User Manual
Signage solutions (35 pages)
-
Monitor Philips 220VW9FB/97 Service Manual
(91 pages)
Related Products for Philips Avalon FM20
- Philips HeartStart FR3 Series
- Philips FM20/30
- Philips FM40/50
- Philips Fetal Double-Spiral Electrode
- Philips Avalon CL
- Philips FM242
- Philips FM24
- Philips FM 242 (FTV 2.1)
- Philips FM23 AC
- Philips FM24 AB
- Philips FM24SS130B/00
- Philips FM-2
- Philips FMXXFD20B
- Philips Swarovski FM01SW40/00
- Philips FM16FD02B/00
- Philips FM01FD10B/00
This manual is also suitable for:
Avalon fm40Avalon fm30Avalon fm50
Table of Contents
Фетальный монитор Avalon FM-20 ф.Philips |
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
||||||
|
Detail Specifications: 1523/1523899-avalon_fm20.pdf file (04 Nov 2022) |
Accompanying Data:
Philips FM20 Medical Equipment, Monitor PDF Service Manual (Updated: Friday 4th of November 2022 01:26:45 AM)
Rating: 4.8 (rated by 67 users)
Compatible devices: BDS4621/27, Brilliance 200P6IG, 160VW9FB/78, 271B8QJKEB/01, HNP7190T, 60PP950299, 107S61, 107P40.
Recommended Documentation:
Text Version of Service Manual
(Ocr-Read Summary of Contents, UPD: 04 November 2022)
-
135, 127 1Index A accessories 100 active settings 108 alarm behavior troubleshooting 48 configuration implications 121 settings monitor settings 121 tones troubleshooting 48 altitude range monitor 7 transducers 8 analyzer,safety 33 assembly contents bottom housing 98 display 101 drawer 100 paper sensor 100 power supply 99 stepper motor 100 top cover 99 top cover housing 100 transducer cable asse…
-
124, 12 Configuration Settings Appendix Measurement-Related Settings 116 Configuring Toco Main Setup —> Measurements —> Toco Configuring IUP Main Setup —> Measurements —> IUP Configuring DFHR (DECG) Main Setup —> Measurements —> DFHR DFHR Configuration Implications Your monitor must be configured to alarm mode All to enable the FHR alarms. High Limit/Low …
-
93, Removing the SpO2 Assembly 8 Disassembly and Reassembly 85 Removing the SpO 2 Assembly 1 Remove the top cover assembly (see page 57). The SpO 2 assembly is identified in the next picture. 2 Remove the two screws holding the SpO 2 assembly. 3 Remove the flat brown cable from the rear of the SpO 2 sensor socket. (Power supply shown removed) SpO 2 assembly Screw l…
-
51, Philips FM20 Troubleshooting Guide 7Troubleshooting 43 Checks with the Instrument Switched On, AC connected The green power LED lights for about 1.5 seconds after switching on, and then goes out, and remains unlit doing normal operation. The location of the green LED is shown in the following photograph: Individual Parameter INOPs If you see any of the following parameter INOPs: try exchanging the…
-
127, Philips FM20 Measurement-Related Settings 12 Configuration Settings Appendix 119 Configuring SpO 2 Main Setup —> Measurements —> SpO 2 SpO 2 Configuration Implications SpO 2 The On/Off state of the SpO 2 measurement cannot be preconfigured. SpO 2 is automatically switched On when an SpO 2 sensor is connected to the monitor. Average The SpO 2 numeric represents an average …
-
7, Philips FM20 v Assembly and Kit Contents 98 Bottom Housing Assembly (M2703-64101) 98 Power Supply Assembly (M2703-60001) 99 Top Cover Assembly (M2703-60052) 99 Top Cover Housing (M2703-64102) 100 Stepper Motor Assembly (M2703-60004) 100 Paper Sensor Assembly (M2703-60003) 100 Drawer Assembly (M2703-64651) 100 Display Assembly (M2703-64503) 101 Transducer Cable Assembly (M2735-64201) 101 Small Parts…
-
118, 11 Understanding Configuration Loading the User Defaults 110 2 Select Factory Default from the pop-up keys at the bottom of the screen. 3 Select Confirm to load the settings stored in the User Defaults. When you load the Factory Default, note that: • Global Settings are not reset. • Hardware Settings are not reset. •Paper Scale Type and Recorder Speed settings are not reset. These are…
-
54, Philips FM20 7 Troubleshooting Troubleshooting Guide 46 Screen is Blank The information listed in this table is only valid if the boot phase has completed without error. See Boot Phase Failures table for a description of the boot phase. Touchscreen Not Functioning No Test Sound issued or INOP Speaker Malfunct. issued Speaker cable disconnected Check speaker connections.…
-
76, 8 Disassembly and Reassembly Recorder Disassembly 68 Removing the Recorder Chassis 1 Remove the top cover assembly (see page 57). 2 Slide open the paper drawer, to gain access to the two countersunk screws (in the following photographs, we have removed the drawer assembly). 3 Remove the two countersunk screws. 4 Turn over the top cover assembly and place it top down on a cloth or other soft …
-
58, 7 Troubleshooting Troubleshooting Guide 50 The INOP Check Paper is issued. The drawer is open and there is paper on the paper sensor. Ensure the paper is loaded correctly, and close the drawer. Paper jam. Open the drawer, remove paper, tear off scrumpled section of paper and re-load, or load a new pack of paper. Close the drawer. Paper sensor dirty. Clean paper sensor (see Chapter …
-
109, Assembly and Kit Contents 9Parts 101 Display Assembly (M2703-64503) Transducer Cable Assembly (M2735-64201) Lever Stop 1 Latch 1 Drawer Assembly Contents Qty Display Assembly Contents Qty Housing, Bottom 1 Housing, Top 1 Clamp 1 Pin 2 Display Holder, lefthand 1 Display Holder, righthand 1 Backlight Tube 2 Hinge 2 Chassis Guide 1 Cable Guide, rear 1 Board Holder …
Recommended Instructions:
10-FLX2-200-VOIP, F27683GYJX1AK2, 55345-rf, WIRELESS LAN MODULE GN-1050, IC-735, HP8143
-
ENGLISHFRANÇAISDEUTSCHESPAÑOLITALIANOSVENSKAPORTUGUÊSNORSKDANSKSUOMIPOLSKIEΛΛHNIKATÜRKÇECESKY JAZYKMAGYAR0459VenaTech®LPBrachial Introducer System (Antecubital)[For US market only: Vena Cava FilterBrachial Introducer System (Antecubital)](1)REF. 04439985B!enfrdeesitsvptnodafipleltrcshuF1944 Ed. 02 (03/2014)F1944_Ed02_couv-new_1944 Ed.02 Couverture 06/03/14 15:32 Page2 …
VenaTech LP 8
-
©2019 Micropoint Biotechnologies Co., Ltd. Všechna práva vyhrazena.Vytištěno v Číně. P/N 630-00182 Rev. A3CSVložte testovací proužek do přístroje a ponechte jej v něm, dokud se jamka na vzorku nezarovná s tečkou na přístroji. Poté s pomocí čtečky naskenujte čárový kód, při skenování dávejte pozor, aby oblast snímání pokrývala čárov …
qLabs ElectroMeter Q-3 Plus 10
-
PN 3984515 January 2011, Rev. 3, 3/16 © 2011-2016 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies. ProSim™ 8 Vital Signs Simulator …
ProSim 8 Series 30
Additional Information:
Operating Impressions, Questions and Answers:
[Page 1] Philips FM20 FETAL MONITORING SERVICE GUIDE Obstetrical Care Avalon Fetal Monitor FM20 FM30 |
[Page 2] Philips FM20 S Printed in Germany 07/05 *M2703-9000A* Part Number M2703-9000A 4512 610 10021 |
[Page 3] Philips FM20 i 1Table of Contents 1 Introduction 1 Who Should Read This Guide 1 What to Do Next 1 Repair Strategy 2 Manufacturer’s Information 2 Passwords 3 Warnings and Cautions 3 2 Site Preparation 5 Introduction 5 Site Planning 5 Roles and Responsibilities 5… |
[Page 4] Philips FM20 ii 4 Theory of Operation 15 Monitor Hardware Overview 15 Power Supply 16 Connector Block 16 Bus Master Board 16 Main CPU Board 16 Fetal Recorder (Thermal Printer Unit) 16 Recorder Adapter Board 17 Thermal Line Printhead (TLPH) 17 Paper Sensor 17 Ste… |
[Page 5] Philips FM20 iii Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): DECG Mode 28 Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): MECG Mode 29 Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): IUP Mode 30 Performance Assuran… |
[Page 6] Philips FM20 iv Alarm Behavior 48 Fetal Recorder 48 LAN / RS232 51 Tr an sd u c ers 5 2 Status Log 53 Troubleshooting with the Support Tool 54 Troubleshooting the Individual Measurements or Applications 54 8 Disassembly and Reassembly 55 Introduction 55 Serial Nu… |
[Page 7] Philips FM20 v Assembly and Kit Contents 98 Bottom Housing Assembly (M2703-64101) 98 Power Supply Assembly (M2703-60001) 99 Top Cover Assembly (M2703-60052) 99 Top Cover Housing (M2703-64102) 100 Stepper Motor Assembly (M2703-60004) 100 Paper Sensor Assembly (M27… |
[Page 8] Philips FM20 vi NBP Configuration Implications 120 Monitor-Related Settings 121 Configuring Alarms 121 Alarm Settings Configuration Implications 121 Configuring the NST Timer 122 NST Timer Configuration Implications 122 Configuring Fetal Recorder Settings 122 Rec… |
[Page 9] Philips FM20 1 1 1Introduction This Service Guide contains technical details for the Avalon FM20 and FM30 Fetal/Maternal Monitors. It provides a technical foundation to support effective troubleshooting and repair. It is not a comprehensive, in-depth explanatio… |
[Page 10] Philips FM20 1 Introduction Repair Strategy 2 Repair Strategy The Service Support Tool software helps you to determine whether a fault is a hardware or software problem. The main replaceable parts are: • unit exchange for the transducers •replacement of – … |
[Page 11] Philips FM20 Passwords 1 Introduction 3 Passwords In order to access different modes within the monitor a password may be required. The passwords are listed below. Monitoring Mode: No password required Configuration Mode: 71034 Demo Mode: 14432 Service Mode: 134… |
[Page 12] Philips FM20 1 Introduction Warnings and Cautions 4 |
[Page 13] Philips FM20 5 2 2Site Preparation Introduction This section describes the procedures you should follow to plan and prepare a site for an Avalon FM20/FM30 fetal monitor installation. • Site planning. • Roles and responsibilities for local and Philips personn… |
[Page 14] Philips FM20 2 Site Preparation Introduction 6 • Pull cables, install conduit, install wallboxes. Philips Personnel • Provide the customer with the safety, environmental and power requirements. •Assemble mounts, as necessary. • Provide requirements for ca… |
[Page 15] Philips FM20 Site Requirements 2 Site Preparation 7 It is the customer’s responsibility to provide and install wallboxes to house faceplates. The customer must notify the Philips installation coordinator of which size is to be used. Procedures for Philips P… |
[Page 16] Philips FM20 2 Site Preparation Site Requirements 8 Safety Requirements (Customer or Philips) The monitor is an electrical Class II device in which the protection against electric shock does not rely on basic insulation and a protective earth conductor but on do… |
[Page 17] Philips FM20 Site Requirements 2 Site Preparation 9 WARNING • Do not use a device in the patient vicinity if it does not comply with IEC-60601-1. The whole installation, including devices outside of the patient vicinity, must comply with IEC-60601-1-1. Any no… |
[Page 18] Philips FM20 2 Site Preparation Site Requirements 10 Input Devices The following tables describes the input devices which can be connected to the monitor via the optional PS/2 interface. Product Option Number Part Number 12NC Part Number Description M8024A #A0… |
[Page 19] Philips FM20 11 3 3Installation Instructions The information contained in this chapter, in addition to that given in the Instructions for Use, should enable the monitor to be installed ready for use (the preparation and planning should be adhered to as specifie… |
[Page 20] Philips FM20 3 Installation Instructions Repackaging for Shipment or Storage 12 Repackaging for Shipment or Storage If the instrument is to be shipped to a Philips Sales/Support Office, securely attach a label showing the name and address of the owner, the instr… |
[Page 21] Philips FM20 Connecting the Monitor to Non-Medical Devices 3 Installation Instructions 13 Connecting the Monitor to Non-Medical Devices Connect the monitor to an obstetrical surveillance system, such as OB TraceVue, via the optional system interface. For cabling… |
[Page 22] Philips FM20 3 Installation Instructions Checking and Setting Line Frequency 14 PS/2 Keyboard/Mouse Switch off the monitor before connecting any PS/2 compatible device. Connect the PS/2 connector to the PS/2 Interface board in the monitor at the slot indicated by… |
[Page 23] Philips FM20 15 4 4Theory of Operation This chapter describes the functional operation of the monitor and the transducers. It incorporates features of the mechanical design, indicating the physical relationship of the assemblies and components. Monitor Hardware … |
[Page 24] Philips FM20 4 Theory of Operation Power Supply 16 The monitor consists of nine main functional components: • Power supply M2703-60001 • Connector Block (1253-8415) • Bus Master Board (M2703-66460) • Main CPU Board (M2703-66450) • Fetal Recorder (Therma… |
[Page 25] Philips FM20 LCD Display and Touchscreen 4 Theory of Operation 17 •Recorder Adapter Board •Thermal Line Printhead (TLPH) •Paper Sensor •Stepper Motor Recorder Adapter Board Recorder signals are handled by the Recorder Adapter Board (M2703-66430), connecte… |
[Page 26] Philips FM20 4 Theory of Operation Input/Output Interface Boards 18 Input/Output Interface Boards There are two optional interface boards available: • LAN/RS232 Interface Board, used for connecting to a PC running the Support Tool and to a surveillance and doc… |
[Page 27] Philips FM20 Transducer Types 4 Theory of Operation 19 Transducer Types Transducers consist of the following types: •US •Toco •Toco + (includes DECG/MECG and IUP capability) They all share the same power supply, and analog-to-digital conversion circuitries… |
[Page 28] Philips FM20 4 Theory of Operation Ultrasound Transducer Frontend 20 Ultrasound Transducer Frontend The ultrasound frontend is a pulsed Doppler system with a 1.0 MHz ultrasound frequency, and a pulse repetition rate of 3 kHz. Seven ultrasound crystals are used a… |
[Page 29] Philips FM20 21 5 5Interfaces There are two interface boards available as options for the Avalon fetal monitors: • LAN / RS232 system interface •Dual PS/2 interface The interfaces are “plug-and-play” boards, and fit into dedicated slots on the underside o… |
[Page 30] Philips FM20 5Interfaces Dual PS/2 Interface 22 |
[Page 31] Philips FM20 23 6 6Testing and Maintenance This chapter contains the testing and maintenance procedures to ensure the proper functioning of the monitor and accessories, covering preventive maintenance, performance assurance and safety. Carry out the procedures a… |
[Page 32] Philips FM20 6 Testing and Maintenance When to Perform Test Blocks 24 When to Perform Test Blocks This table tells you when to perform specific test blocks. See page 25 for test details. Table 2: When to perform test blocks Service Event Test Block(s) Required -… |
[Page 33] Philips FM20 Test Reporting 6 Testing and Maintenance 25 Test Reporting Authorized Philips personnel report test result back to Philips to add to the product development database. Hospital personnel, however, do not need to report results. How to Carry Out the T… |
[Page 34] Philips FM20 6 Testing and Maintenance Preventive Maintenance Procedures 26 Preventive Maintenance Procedures The preventive maintenance tasks are restricted to the noninvasive blood pressure measurement calibration. Carry out the noninvasive blood pressure meas… |
[Page 35] Philips FM20 Other Regular Tests 6 Testing and Maintenance 27 5 The ultrasound transducer contains seven piezoelectric crystals. Basic functioning of each can be verified by holding a flat bottomed pencil or similar above each crystal and moving it up and down … |
[Page 36] Philips FM20 6 Testing and Maintenance Other Regular Tests 28 Toco Transducer Electrical Check To test a Toco transducer: 1 Switch on the monitor and the recorder. 2 Connect the transducer to the fetal monitor. 3 Gently apply pressure to the Toco sensor. 4 Check … |
[Page 37] Philips FM20 Other Regular Tests 6 Testing and Maintenance 29 6 EITHER Make a short between the spiral electrode and the reference electrode with your fingers (it is best to wet your fingers first). Use a sterile Fetal Scalp Electrode. CAUTION The tip of the spi… |
[Page 38] Philips FM20 6 Testing and Maintenance Other Regular Tests 30 Result: You should see MECG values displayed on the maternal display or annotated on the recorder trace. If the test results are not as outlined above, repeat the test with another ECG transducer. If … |
[Page 39] Philips FM20 Performance Assurance Tests 6 Testing and Maintenance 31 Performance Assurance Tests Some of the following test procedures must be performed in service mode. To enter service mode select Operating Modes in the main menu. Then select Service Mode and… |
[Page 40] Philips FM20 6 Testing and Maintenance Performance Assurance Tests 32 9 To calibrate the noninvasive blood pressure measurement, select Close Valves off then Calibrate NBP and wait for the instrument to pump up the expansion chamber.Wait a few seconds after pum… |
[Page 41] Philips FM20 Safety Tests 6 Testing and Maintenance 33 1 Connect an adult SpO 2 transducer to the SpO 2 connector. 2 Measure the SpO 2 value on your finger (this assumes that you are healthy). 3 The value should be between 95% and 100%. Safety Tests You are re… |
[Page 42] Philips FM20 6 Testing and Maintenance Safety Tests 34 Electrical Isolation Diagram This diagram gives an overview of the electrical isolation of the monitor. Accessible metal parts are identified (see arrows). Safety Test Procedures Use the test procedures outl… |
[Page 43] Philips FM20 Safety Tests 6 Testing and Maintenance 35 NOTE The protective earth conductor is required for EMC purposes. It has no protective function against electric shock. The protection against electric shock is provided by double and/or reinforced insulatio… |
[Page 44] Philips FM20 6 Testing and Maintenance System Test 36 System Test After mounting and setting up a system, perform system safety tests according to IEC60601-1-1. What is a Medical Electrical System? A medical electrical system is a combination of at least one medi… |
[Page 45] Philips FM20 Touchscreen Calibration 6 Testing and Maintenance 37 WARNING • Do not use additional AC mains extension cords or multiple portable socket-outlets. If a multiple portable socket-outlet without an approved isolation transformer is used, the interrup… |
[Page 46] Philips FM20 6 Testing and Maintenance Disabling/Enabling Touch Operation 38 If the touchscreen is accidentally mis-calibrated by selecting the wrong spot, you must use another input device to re-enter calibration mode. If you have the support tool, you can sele… |
[Page 47] Philips FM20 Checking the Fetal Recorder Offset 6 Testing and Maintenance 39 Checking the Fetal Recorder Offset The easiest way to check the recorder offset is to connect a resting Toco transducer (one that is not under any load) to the monitor and then change t… |
[Page 48] Philips FM20 6 Testing and Maintenance Fetal Recorder Selftest Report 40 Fetal Recorder Selftest Report To verify your printer configuration, or if you doubt the performance of the recorder, you may want to print a test report. To print a selftest report, in Ser… |
[Page 49] Philips FM20 41 7 7Troubleshooting A list of system error messages and troubleshooting information for common problems you may encounter while using the monitor and its accessories is given in the Instructions for Use. This chapter provides a guide for qualifie… |
[Page 50] Philips FM20 7 Troubleshooting Software Revision Check 42 2 Select Product. You see the hardware revision in the pop-up window, along with the serial number, part number, and the software revision. The following table shows which part number corresponds to which… |
[Page 51] Philips FM20 Troubleshooting Guide 7Troubleshooting 43 Checks with the Instrument Switched On, AC connected The green power LED lights for about 1.5 seconds after switching on, and then goes out, and remains unlit doing normal operation. The location of the gree… |
[Page 52] Philips FM20 7 Troubleshooting Troubleshooting Guide 44 Initial Instrument Boot Phase The following table describe the regular initial boot phase of the monitor. If the boot phase does not proceed as described below go to Boot Phase Failures for Troubleshooting … |
[Page 53] Philips FM20 Troubleshooting Guide 7Troubleshooting 45 Boot Phase Failures Symptoms Possible Causes of Failure Failure Isolation and Remedy Green LED does not light up, and no test tone is heard No AC mains connection Check that the power cord is not damaged a… |
[Page 54] Philips FM20 7 Troubleshooting Troubleshooting Guide 46 Screen is Blank The information listed in this table is only valid if the boot phase has completed without error. See Boot Phase Failures table for a description of the boot phase. Touchscreen Not Functioni… |
[Page 55] Philips FM20 Troubleshooting Guide 7Troubleshooting 47 General Monitor INOP Messages INOP Message Possible Causes of Failure Failure Isolation and Remedy CheckInternVoltage Problem with the voltages (5V) in the monitor Remove all I/O boards and put them back in… |
[Page 56] Philips FM20 7 Troubleshooting Troubleshooting Guide 48 Keyboard/Mouse Not Functioning Alarm Tones Alarm Behavior If your monitor did not alarm in the way in which the end user expected, please consult the Instructions for Use for possible setup issues or config… |
[Page 57] Philips FM20 Troubleshooting Guide 7Troubleshooting 49 No paper transport. Poor connection. Check all internal connectors. Paper jam. Open the drawer, remove paper, tear off scrumpled section of paper and re-load, or load a new pack of paper. Close the drawer. … |
[Page 58] Philips FM20 7 Troubleshooting Troubleshooting Guide 50 The INOP Check Paper is issued. The drawer is open and there is paper on the paper sensor. Ensure the paper is loaded correctly, and close the drawer. Paper jam. Open the drawer, remove paper, tear off s… |
[Page 59] Philips FM20 Troubleshooting Guide 7Troubleshooting 51 LAN / RS232 Symptoms Cause of Failure Failure Isolation and Remedy External device (such as a surveillance system like OB TraceVue) not receiving data The LAN/RS232 port is not configured for data export… |
[Page 60] Philips FM20 7 Troubleshooting Troubleshooting Guide 52 Transducers Symptoms Possible Cause Failure Isolation and Remedy Transducer appears not to work, and the transducer Finder LED does not light when you touch the parameter field on the screen. INOP OB E… |
[Page 61] Philips FM20 Troubleshooting Guide 7Troubleshooting 53 Status Log Many events that occur during start-up or regular monitoring are logged in the Status Log. The Status Log can be cleared. Not all entries in the Status Log are errors. You can print the Status Log… |
[Page 62] Philips FM20 7 Troubleshooting Troubleshooting Guide 54 The first column in the log identifies the event class (“C”: caused a cold start, “H”: caused a hot start, “N”: no restart, for information only). Column 3 and 4 identify the event source and ev… |
[Page 63] Philips FM20 55 8 8Disassembly and Reassembly WARNING • Before attempting to open or disassemble the monitor, disconnect it from the AC mains supply. • Energized circuits are accessible with the covers open. Do not work on the monitor with the covers open an… |
[Page 64] Philips FM20 8 Disassembly and Reassembly Serial Numbers 56 Tools Required CAUTION When replacing the front cover, do not over-torque the screws. Excessive torque may damage the plastic screw mountings. You need the following tools: Serial Numbers The serial num… |
[Page 65] Philips FM20 Removing the Top Cover Assembly 8 Disassembly and Reassembly 57 Removing the Top Cover Assembly 1 First fold the display completely flat. 1 Carefully place the monitor upside down. To avoid scratches, place the unit on some cloth or other soft surf… |
[Page 66] Philips FM20 8 Disassembly and Reassembly Removing the Top Cover Assembly 58 2 Remove the four screws securing the top cover assembly to the bottom housing, using a T-10 Torx driver. 3 To gain access to the screw in the rear right hand corner, release the cable … |
[Page 67] Philips FM20 Removing the Display Assembly 8 Disassembly and Reassembly 59 5 Separate the top cover from the bottom housing from the front of the monitor as illustrated. 6 The top cover assembly is now separated from the bottom housing assembly. Removing the Disp… |
[Page 68] Philips FM20 8 Disassembly and Reassembly Removing the Display Assembly 60 3 First release the cable lock by moving it in the direction of the arrows, then disconnect the white recorder ribbon cable from the recorder adapter board. 4 Next, disconnect the silver … |
[Page 69] Philips FM20 Removing the Display Assembly 8 Disassembly and Reassembly 61 5 Turn over the top cover assembly. 6 Remove the two plastic hinge pins on either side of the display housing. Release them by turning the slotted head anti-clockwise with a small flat-bl… |
[Page 70] Philips FM20 8 Disassembly and Reassembly Replacing the Display Assembly 62 Replacing the Display Assembly Replacing the display assembly is the reverse of the removal procedure. 1 With the top cover assembly facing upwards, feed the display ribbon cable through … |
[Page 71] Philips FM20 Replacing the Display Assembly 8 Disassembly and Reassembly 63 5 Reconnect the white recorder ribbon cable to the recorder adapter board, ensuring the cable lock is secure after fitting. 6 Refit the plastic hinge pins. Do not apply excessive force w… |
[Page 72] Philips FM20 8 Disassembly and Reassembly Recorder Disassembly 64 7 Refit the ratchet clip into the slots on the rear of the display housing and the top cover, as a reversal of the removal procedure. Recorder Disassembly The recorder consists of the following ma… |
[Page 73] Philips FM20 Recorder Disassembly 8 Disassembly and Reassembly 65 . Detail showing plastic hooks open |
[Page 74] Philips FM20 8 Disassembly and Reassembly Recorder Disassembly 66 Replacing the Drawer Assembly Before replacing the drawer assembly, refer to the previous photographs, and study the detail showing the position of the plastic hooks when the drawer is fixed in po… |
[Page 75] Philips FM20 Recorder Disassembly 8 Disassembly and Reassembly 67 Final position (hook latched onto metal guide) |
[Page 76] Philips FM20 8 Disassembly and Reassembly Recorder Disassembly 68 Removing the Recorder Chassis 1 Remove the top cover assembly (see page 57). 2 Slide open the paper drawer, to gain access to the two countersunk screws (in the following photographs, we have remo… |
[Page 77] Philips FM20 Recorder Disassembly 8 Disassembly and Reassembly 69 5 First release the cable lock by moving it in the direction of the arrows, then disconnect the white recorder ribbon cable from the recorder adapter board. 6 Next, disconnect the silver display r… |
[Page 78] Philips FM20 8 Disassembly and Reassembly Recorder Disassembly 70 8 Remove the two screws (one on each side) fastening the RFI bracket to the recorder chassis. 9 Partially unscrew the screw nearest the recorder adapter board holding the stepper motor to the recor… |
[Page 79] Philips FM20 Recorder Disassembly 8 Disassembly and Reassembly 71 11 Free the RFI bracket on the stepper motor side by applying a little sideways pressure to the recorder chassis, while at the same time pulling the RFI bracket forwards. . 12Remove the RFI bracke… |
[Page 80] Philips FM20 8 Disassembly and Reassembly Recorder Disassembly 72 13 Lift out the recorder chassis. Removing the Thermal Line Printhead (TLPH) 1 Remove the recorder chassis as described in the section “Replacing the Recorder Chassis” on page 74”. . 2 Turn t… |
[Page 81] Philips FM20 Recorder Disassembly 8 Disassembly and Reassembly 73 3 Remove the two screws holding the TLPH to the TLPH holder. 4 Turn the chassis assembly over again, and carefully remove the TLPH from the holder as shown. Replacing the TLPH The procedure for rep… |
[Page 82] Philips FM20 8 Disassembly and Reassembly Recorder Disassembly 74 Replacing the Recorder Chassis The procedure for replacing the recorder chassis is a reversal of the removal procedure (see “Replacing the Recorder Chassis” on page 74 for the sequence). 1 Ens… |
[Page 83] Philips FM20 Recorder Disassembly 8 Disassembly and Reassembly 75 6 Now tighten the four screws to secure the recorder chassis. 7 Ensure that you reconnect the paper sensor cable and the stepper motor cable to the recorder adapter board. Removing the Paper Senso… |
[Page 84] Philips FM20 8 Disassembly and Reassembly Recorder Disassembly 76 3 First release the cable lock by moving it in the direction of the arrows, then disconnect the white recorder ribbon cable from the recorder adapter board. 4 Next, disconnect the silver display r… |
[Page 85] Philips FM20 Recorder Disassembly 8 Disassembly and Reassembly 77 5 Disconnect the stepper motor the record-sense cable connectors from the recorder adapter board. 6 With all cables disconnected, remove the recorder adapter board. Replacing the Recorder Adapter B… |
[Page 86] Philips FM20 8 Disassembly and Reassembly Recorder Disassembly 78 Removing the Stepper Motor 1 Remove the top cover assembly (see “Removing the Top Cover Assembly” on page 57). 2 Place the top cover assembly top down on a cloth or other soft surface. 3 Discon… |
[Page 87] Philips FM20 Replacing the Stepper Motor 8 Disassembly and Reassembly 79 5 Remove the stepper motor. Replacing the Stepper Motor 1 Refit the stepper motor to the recorder chassis, but do NOT tighten the screws yet! The upper slot on the stepper motor allows fine… |
[Page 88] Philips FM20 8 Disassembly and Reassembly Replacing the Top Cover Assembly 80 4 When in place, tighten the screws, remembering that the upper screw is secured with a small nut. Replacing the Top Cover Assembly 1 Ensure all items are replaced in the top cover ass… |
[Page 89] Philips FM20 Removing the Power Supply Assembly 8 Disassembly and Reassembly 81 5 Holding both assemblies together, carefully place the unit upside down on a soft surface to prevent scratching or other damage. 6 Refit the four screws securing the top cover assem… |
[Page 90] Philips FM20 8 Disassembly and Reassembly Replacing the Power Supply Assembly 82 4 Lift the cable end of the power supply assembly with one hand, while guiding the power socket/on/off switch free of the aperture in the bottom housing, then lift out the power sup… |
[Page 91] Philips FM20 Removing the Noninvasive Blood Pressure Assembly 8 Disassembly and Reassembly 83 2 Disconnect the tubing from the noninvasive blood pressure connector, and the ribbon cable from the main CPU board. 3 Disconnect the grounding cable’s spade connecto… |
[Page 92] Philips FM20 8 Disassembly and Reassembly Replacing the Noninvasive Blood Pressure Assembly 84 5 To separate the noninvasive blood pressure assembly from the noninvasive blood pressure assembly holder, remove the three screws. Replacing the Noninvasive Blood Pre… |
[Page 93] Philips FM20 Removing the SpO2 Assembly 8 Disassembly and Reassembly 85 Removing the SpO 2 Assembly 1 Remove the top cover assembly (see page 57). The SpO 2 assembly is identified in the next picture. 2 Remove the two screws holding the SpO 2 assembly. 3 Remov… |
[Page 94] Philips FM20 8 Disassembly and Reassembly Replacing the SpO2 Assembly 86 4 Lift the side of the SpO 2 assembly nearest the SpO 2 socket, carefully disconnecting the multi-pin connector shown. 5 Remove the SpO 2 assembly. Replacing the SpO 2 Assembly The proc… |
[Page 95] Philips FM20 Removing the Interface Boards 8 Disassembly and Reassembly 87 Removing the Interface Boards The interface boards can be accessed from the underside of the monitor’s housing. 1 Turn the monitor upside down on a non-scratch surface. 2 Release the boa… |
[Page 96] Philips FM20 8 Disassembly and Reassembly Removing the Main CPU Board 88 Removing the Main CPU Board To remove the main CPU board, proceed as follows: 1 Remove the top cover assembly (see page 57). 2 Remove the screws holding the metal shield. 3 Remove the shield… |
[Page 97] Philips FM20 Replacing the Main CPU Board 8 Disassembly and Reassembly 89 5 Remove the main CPU board by lifting it straight up. Replacing the Main CPU Board The procedure to replace the main CPU board is a reversal of the removal procedure. Ensure all the cables… |
[Page 98] Philips FM20 8 Disassembly and Reassembly Exchanging the Transducer Cable 90 Exchanging the Transducer Cable See the “Transducer Cable Assembly (M2735-64201)” on page 101 for items that come with the cable. Important when fitting the screw covers! Do NOT remo… |
[Page 99] Philips FM20 Exchanging the Transducer Cable 8 Disassembly and Reassembly 91 Disconnect the small cable connector, remove the old cable, and fit the new cable (as a reversal of the removal procedure). 3 Remove the sealing gasket from the top cover, and replace i… |
[Page 100] Philips FM20 8 Disassembly and Reassembly Exchanging the Transducer Belt Button 92 Exchanging the Transducer Belt Button CAUTION NEVER immerse a transducer in liquid if the belt button has been removed, or is loose, broken or damaged. 1 Remove the belt button us… |
Download Service manual of Philips FM20 Monitor for Free or View it Online on All-Guides.com.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
FETAL MONITORING
SERVICE GUIDE
Obstetrical Care
Avalon Fetal Monitor
FM20
FM30