Easy laser e420 инструкция по применению

  1. Manuals
  2. Brands
  3. Damalini Manuals
  4. Laser Shaft Alignment Tools
  5. Easy-Laser E420
  6. Manual
  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

E420

MANUAL

English

GlobalTestSupply

www.

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

05-0629 Revision 2.0

System version 6.0

loading

Summary of Contents for Damalini Easy-Laser E420

  • Page 1
    E420 MANUAL English GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com 05-0629 Revision 2.0 System version 6.0…
  • Page 2
    GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com…
  • Page 3: Table Of Contents

    CONTENT INTRODUCTION Service and calibration Travelling with your measurement system DISPLAY UNIT Navigation buttons Function buttons Status bar Screen dump LED lights Battery Charge the Display unit Charge measuring units Calculator Measurement file handling Save file File manager Open file as template Copy file to USB memory Barcode Print file (Optional)

  • Page 4
    PROGRAM VALUES Tolerance Zoom Half or Zero set value Live values – colours Automatic recording Calibration check HORIZONTAL Mount the units Bluetooth® Enter distances Measure using Easy Turn™ Measure using 9-12-3 Result and adjust Live values Thermal compensation Tolerance SOFTFOOT VERTICAL Preparations Measure…
  • Page 5: Introduction

    Quality certificate Damalini AB is ISO 9001:2008 certified. Certificate number 900958. Damalini AB confirm, that our products are produced according to applicable national and international regulations and standards. All components are checked before assem- bly and final products are tested in functionality and visually checked before delivery The calibration of the equipment fully complies with ISO9001: 2008 #7.6…

  • Page 6
    Compensation for possible damage due to failure on Easy-Laser product is not in- ® cluded in the warranty. Freight cost to Damalini is not included in the warranty. Note! Before delivery of the product for warranty repair, it is the responsibility of the buyer to backup all data.
  • Page 7: Service And Calibration

    USB memory stick that is delivered with most systems. EasyLink The new version of our database program EasyLink is available on the USB memory stick that is delivered with most systems. You can always download the latest version from damalini.com>download>software. GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com…

  • Page 8: Travelling With Your Measurement System

    The E-series is not compatible with previous analogue units from the D-series. You can however continue to use previous brackets. Disclaimer Damalini AB and our authorized dealers will take no responsibility for damage to machines and plant as a result of the use of Easy-Laser measurement and alignment ®…

  • Page 9: Display Unit

    DISPLAY UNIT Press and hold the On/Off button to reset the Display unit. Battery info See “LED lights” on page 8 On/Off Escape OK button Navigation buttons Numeric buttons Function buttons External power USB A (master) Escape Escape will either return to start view, or close active window (for example file man- ager) and return to previous view.

  • Page 10: Navigation Buttons

    Display unit Navigation buttons To navigate on the screen, use the navigation buttons. The selected icon is marked with a yellow frame. The navigation buttons are also used to move between the icons in a submenu and to change the values in the fields. Escape Navigation buttons OK button…

  • Page 11: Status Bar

    Display unit Status bar The Status bar contains additional information such as warning icon, current time and Bluetooth connection. ® Measurement unit. Change units via Settings. The yellow arrow indicates that there are sub-menus There are also text messages regarding: •…

  • Page 12: Screen Dump

    Display unit Screen dump It is possible to take screen dumps of what is currently displayed on screen. You can e-mail the screen dump or use it for reports. Take a screen dump 1. Press and hold the numeric button period (.) for 5 seconds. 2.

  • Page 13: Battery

    Display unit Battery Select to display the Battery view. This view gives you a good overview of the battery status of all connected equipment. Display unit Detector or Measuring unit Serial number Charging Battery symbols correspond to the LED signals on the Display unit. The E-series is not compatible with units from the D-series.

  • Page 14: Calculator

    Display unit Calculator The calculator is found on the Start view and Control panel ( 1. Select to open the calculator. 2. Use the numerical buttons and function buttons to enter values. 3. Use the button to compute. Select to display sub-menu Use OK button as equal sign (=) Unit converter The unit converter is found on the Start view and Control panel (…

  • Page 15: Measurement File Handling

    Display unit Measurement file handling Save file 1. Select to save your measurement. 2. Enter a file name. The date and time will automatically be added to the file name. The measurements that you save will be available to other users as well. 3.

  • Page 16: Open File As Template

    1. Insert a USB memory. 2. Select the file you want and select 3. A folder is automatically created on the USB memory. The file is saved in the folder Damaliniarchive. Barcode Save file with barcode The barcode scanner is not included in all systems. The first time you measure a machine, you stick a barcode on the machine and save the measurement together with the scanned barcode.

  • Page 17: Print File (Optional)

    Display unit Print file (Optional) Part no. 03-1004 The thermal printer is optional equipment. 1. Save the measurement. To print from a Shaft program, you need to open a saved measurement before you can print a report. 2. Connect the thermal printer and select 3.

  • Page 18: Control Panel

    Display unit Control panel Select to open the Control panel. Some of the settings are personal and will be default next time you start the system. Note! All settings are not available for all systems. Filter Select to open the Filter view. If the laser beam passes through air with varying temperature, this may influence the direction of the laser beam.

  • Page 19: Unit And Resolution

    Display unit Unit and resolution Personal setting Select to open the Units and resolution view. Use the navigation buttons to move between the fields. Set Metric or Imperial and which resolution you want to use. Default is set to 0.01 mm (0.4 mil). The selected unit is shown on the Status bar. Note! It is possible to select 0.0001mm only in the E940 system.

  • Page 20: Language

    Display unit Language Personal setting Select to open the Language view. Default is set to English. Use the navigation buttons to select a language. Press to save changes. Language view User Select to open the Users view. A user account is used for storing your personal settings.

  • Page 21
    Display unit Automatic sleep mode Personal setting Select to open the Automatic sleep mode view. Select how much time before automatic sleep mode. Use the navigation buttons to select. Press OK to save changes. See also “On/Off” on page 5. Information Select to display the information regarding serial number and version of the…
  • Page 22: System Update

    Display unit System update Download update file 1. Go to www.damalini.com > Download > Software > E series Display unit Firmware Update. 2. Download the update file to your PC. 3. Unzip the file. 4. Copy the .elu file to the root of a USB memory.

  • Page 23: License

    Display unit License It is easy to upgrade your Display unit. 1. Contact your Easy-Laser distributor if you wish to upgrade your Display unit. ® 2. An e-mail will be sent to you with information on how to download the update file.

  • Page 24: Bluetooth® Set Up

    Display unit Bluetooth® set up Bluetooth wireless technology makes it possible for Display unit and Detector to ® exchange data without using cables. Some detectors have built-in Bluetooth , others have a separate Bluetooth ® unit that you attach to the detector. Please see Technical data for more information.

  • Page 25
    Display unit Use only one Bluetooth unit ® Many of our systems are delivered with two Measuring units. In some cases you might want to use only one unit together with a laser transmitter. By default both units are set to “Connect ”.
  • Page 26
    GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com…
  • Page 27: Choose Program

    CHOOSE PROGRAM Preparations Before starting a measurement, there are several things that are good to check to ensure a good and accurate measurement. • Ensure a good measurement environment. Strong sunlight, warning lights, vibra- tions and temperature gradients can affect the readings. •…

  • Page 28
    GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com…
  • Page 29: Program Values

    PROGRAM VALUES With the program Values, you can see live readings from the detectors. As default, a target and a table is displayed. Press OK to register values. Live vertical values Detector or Measuring unit Registered values Serial number Use the navigation Unit one (out of two connected) buttons to scroll Target…

  • Page 30: Tolerance

    Program values Tolerance 1. Select to set tolerance. It is possible to set different tolerance in vertical and horizontal direction. 2. Use navigation buttons to move between the fields and to change the tolerance. 3. Press OK. Live values and marking displayed in Live values displayed in red when green when within tolerance.

  • Page 31: Half Or Zero Set Value

    Program values Edge warning When the laser beam is close to the edge, the edge is “lit up” as a warning. It is not possible to register values when you see the edge warning. Half or Zero set value Half value Zero set value Absolute value Select…

  • Page 32: Automatic Recording

    Program values Automatic recording In Values, it is possible to make automatic recording of values. This is very useful when you want to register values during a longer time period for example. 1. Select to start automatic recording. 2. Set Interval. 3.

  • Page 33: Calibration Check

    Program values Calibration check Use the program Values to check if the detector readings are within specified tolerances. Quick check 1. Set the tolerance to 0.01 mm (0.5mil). 2. Select and show targets for both M- and S-unit. 3. Select to zero set value.

  • Page 34
    GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com…
  • Page 35: Horizontal

    HORIZONTAL For horizontally mounted machines. Select between two different measuring methods: EasyTurn Start anywhere on the turn. The three measuring positions can be reg- istered with as little as 20º between positions. By default, the EasyTurn program is shown. See “Measure using Easy Turn™” on page 35. 9-12-3 The measuring positions are registered at positions 9, 12, 3 o’clock.

  • Page 36: Mount The Units

    Horizontal Mount the units 1. Mount the S-unit on the stationary machine and the M-unit on the movable machine. 2. Mount the units facing each other. Make sure they are at the approximately same rotational angle. Laser adjustment knob On/Off button Connection for charging cable Locking knob Hook for the chain…

  • Page 37
    Horizontal Adjust measuring units Place the Measuring units on the rods, make sure they are at the approximately same rotational angle. You need to place the measuring units with an offset, see image. Also make sure that the adjustment knob is adjustable in both directions. Image showing measuring units for E530 system.
  • Page 38: Enter Distances

    Horizontal Enter distances Confirm each distance with Distance between S-unit and M-unit. Measure between the rods. Distance between S-unit and centre of coupling. Distance between M-unit and feetpair one. Distance between feetpair one and feetpair two. Coupling diameter. Optional, select to activate field.

  • Page 39: Measure Using Easy Turn

    Horizontal Measure using Easy Turn™ Preparations Follow the preparations as described in the previous pages. 1. Mount the measuring units. 2. Enter distances, confirm each distance with OK. 3. If needed, perform a rough alignment. 4. If needed, perform a Softfoot check. Measure It is possible to measure with as little as 40º…

  • Page 40: Measure Using

    Horizontal Measure using 9-12-3 Preparations Follow the preparations as described in the previous pages. 1. Mount the measuring units. 2. Enter distances, confirm each distance with OK. 3. If needed, perform a rough alignment. 4. If needed, perform a Softfoot check. Measure 1.

  • Page 41: Result And Adjust

    Horizontal Result and adjust Offset, angle and feet values are clearly displayed. Both horizontal and vertical direc- tion are shown live, which makes it easy to adjust the machine. Values within tolerance are green. Vertical offset and angle Horizontal offset and angle Thermal compensation has been set Feet values.

  • Page 42: Live Values

    Horizontal Live values When reading the values, face the stationary machine from the movable machine. Positions for measuring units as seen from the movable machine. Live values are marked with yellow frame. Face the stationary machine (S) from the movable machine (M). Then 9 o’clock is to the left, as in the measuring programs. Offset and angle values The offset and angle value indicate how well the machine is aligned at the coupling.

  • Page 43
    Horizontal Adjust Adjust the machine if needed. 1. Shim the machine according to the vertical feet values. 2. Adjust the machine sideways according to the live horizontal values. 3. Tighten the feet. 4. Select to remeasure. Position indicator To adjust, you need to place the measuring units in live position (9, 12, 3 or 6 o’clock). Select to show the Position indicator.
  • Page 44: Thermal Compensation

    Horizontal Thermal compensation During normal operation, machinery is influenced of different factors and forces. The most common of these changes is the change in the temperature of the machine. This will cause the height of the shaft to increase. This is called thermal growth. To com- pensate for thermal growth, you enter values for cold condition compensation.

  • Page 45: Tolerance

    Horizontal Tolerance 1. Select . The tolerance window is displayed. 2. Select a tolerance and press GAP values Add user defined tolerance Function buttons Close Tolerance view. See “Control panel” on page 14. Edit user defined tolerance. Delete user defined tolerance. Add new tolerance You can add your own user defined tolerance.

  • Page 46
    Horizontal Tolerance table The rotation speed of the shafts will decide the demands on the alignment. The table on this side can be used as a guidance if no other tolerances is recommended by the manufacturer of the machines. The tolerances is set to the maximum allowed deviation from accurate values, with no consideration to if that value should be zero or compensated for thermal growth.
  • Page 47: Softfoot

    SOFTFOOT Perform a softfoot check to ensure that the machine is resting evenly on all its feet. A softfoot can be angular and/or parallel, see image. Softfoot can be caused by: • Twisted machinery foundations. • Twisted or damaged machinery feet. •…

  • Page 48
    Softfoot Measure softfoot 1. Tighten all feet bolts. 2. Turn the measuring units to 12 o’clock. 3. Adjust laser to the centre of targets. If needed, adjust the units on the rods, then use laser adjustments knobs. Turn the measuring units to 12 o’clock. Adjust laser line to centre of target.
  • Page 49: Vertical

    VERTICAL The program Vertical is used for vertical and/or flange mounted machines. Preparations 1. Mount the M-unit on the movable machine and the S-unit on the stationary machine. 2. Select to open Vertical program. 3. Enter distances. Confirm each distance with If you have a barcode reader, simply scan the barcode and all machine data is read.

  • Page 50: Measure

    Vertical Measure The program Vertical uses the 9-12-3 method. 1. Position the units at 9 o’clock, at bolt number one. Make sure that it is possible to also position the units at 12 and 3 o’clock. 2. Press to register first position. The first position is automatically set to zero. 3.

  • Page 51: Result

    Vertical Result The result is displayed as sideways offset in the coupling and angular error between shafts. Live values in Live values in the 9-3 direction the 6-12 direction Live values The values can be displayed live in two directions: •…

  • Page 52: Adjust Machine

    Vertical Shim result view To view this, you need to enter number of bolts and diameter of bolt circle. 1. Select to open Shim value view. The values are not live. 2. Read values. The highest bolt is calculated as 0.00. Values below zero indicates that the bolt is low and need shimming.

  • Page 53: Technical Data

    TECHNICAL DATA System Easy-Laser® E420 Shaft, Part No. 12-0745 A complete system contains Measuring unit M Measuring unit S Display unit Shaft brackets with chains Set of rods 4×60 mm, 4×120 mm [4.72”, 2.36”] Measuring tape 3 m Power adapter (100–240 V AC) DC split cable for charging DC to USB adapter, for charging Quick reference manual…

  • Page 54: Display Unit E53

    Technical Data Display unit E53 Part no. 12-0748 In the Display unit you are guided through the measurement procedure and can save and analyze the results. Battery info See “LED lights” on page 8 On/Off Escape OK button Navigation buttons Numeric buttons Function buttons Display unit…

  • Page 55: Measuring Units

    Technical Data Measuring units Part no. 12-0746 Part no. 12-0747 On/Off Laser aperture Connector for charger cable See “LED lights” on page 8 The rods have a height of PSD (20 mm/0.78”) 120–180 mm. They can be extended with accessory rods. Made of stainless steel.

  • Page 56
    Technical Data GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com…
  • Page 57: Index

    INDEX Angular misalignment 38 PDF 12 Personal settings 15 Position indicator 37, 39 Bar-code 13 Print 14 Barcode 8 Projector 7 Battery view 9 Bluetooth® 7, 21 Buttons 5, 5 Resolution 16 Rotate coordinates 7 Rough alignment 33 Calibration 3 Calibration check 29 Convert units 11 Screen dump 8…

  • Page 58
    GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com…

  1. Manuals
  2. Brands
  3. Damalini Manuals
  4. Laser Shaft Alignment Tools
  5. Easy-Laser E420

Manuals and User Guides for Damalini Easy-Laser E420. We have 1 Damalini Easy-Laser E420 manual available for free PDF download: Manual

Note for Owners:

Guidesimo.com webproject is not a service center of Damalini trademark and does not carries out works for diagnosis and repair of faulty Damalini Easy-Laser E420 equipment. For quality services, please contact an official service center of Damalini company. On our website you can read and download documentation for your Damalini Easy-Laser E420 device for free and familiarize yourself with the technical specifications of device.

More Laser Shaft Alignment Tools Devices:

  • PRÜFTECHNIK OPTALIGN smart RS5 BT

    1A user’s handbook Dear Customer,Welcome to precision laser shaft alignment using OPTALIGN® smart RS5 BT. This OPTALIGN smart RS5 BT handbook sets out to help the user work through the meas-urement system with the required ease. It is meant to be a quick learning tool that hopefully should make compelling reading.PRÜFTECHNIK 85737 IsmaningGermanyEmail [email protected] …

    OPTALIGN smart RS5 BT Laser Shaft Alignment Tools, 194

  • LUDECA smartALIGN

    1. Know your smartALIGN2. Mount smartALIGN and switch on3.Enter dimensionsThe main operating buttons are the joystick, the menu buttonand the escape button.4. MeasureFrom the start screen, select the dimensions icon. Depress joystickto enter machine dimensions. The rectangular box appears whenthe joystick is pushed forwards or backwards. Proceed to edit di-mension by moving the joystick forwards, …

    smartALIGN Laser Shaft Alignment Tools, 2

  • PRÜFTECHNIK Rotalign Touch

    © Copyright 2015 by PRÜFTECHNIK Condition Monitoring GmbHSafety and general informationSicherheits- und allgemeine HinweiseInformations générales et de sécuritéInformación general y de seguridadVeiligheids- en algemene informatieSäkerhet och allmän informationBezpečnostní a všeobecné informaceInformação geral e de segurança安全信息Printed in Germany DOC 50.601 …

    Rotalign Touch Laser Shaft Alignment Tools, 7

  • PRÜFTECHNIK ROTALIGN ALI 3.500

    (217) 352-9330 | [email protected] | artisantg.com-~ ARTISAN® ~I TECHNOLOGY GROUP Your definitive source for quality pre-owned equipment. Artisan Technology Group Full-service, independent repair center with experienced engineers and technicians on staff. We buy your excess, underutilized, and idle equipment along with credit for buybacks and trade-ins. Custom engineering so your equipment work …

    ROTALIGN ALI 3.500 Laser Shaft Alignment Tools, 164

Recommended Documentation:

Table of Contents for Damalini Easy-Laser E420:

  • Display unit E53 Part no. 12-0748 In the Display unit you are guided through the measurement procedure and can save and analyze the results. Navigation buttons Function buttons Numeric buttons See “LED lights” on page 8 OK button EscapeOn/Off Battery info Display unit Type of display/size VGA 5.7” colour, backlit LED Displayed resolution 0.001 mm / 0.05 thou Internal battery (stationary) Heavy duty Li Ion chargeable Operating time Appro. 30 hours (Normal operating cycle) Temp

  • 24 www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

  • 48 Vertical Shim result view To view this, you need to enter number of bolts and diameter of bolt circle. 1. Select to open Shim value view. The values are not live. 2. Read values. The highest bolt is calculated as 0.00. Values below zero indicates that the bolt is low and need shimming. 3. Select to return to Result view. Note! If you shim the machine, remeasure from position 9 o’clock to update all measurement values. Adjust machine 1. Compare the offset and angular err

  • 11 Display unit Measurement le handling Save le 1. Select and to save your measurement. 2. Enter a le name. The date and time will automatically be added to the le name. The measurements that you save will be available to other users as well. 3. Press to save the le. File manager Select (found on the start view and Control panel) to open saved measurements. The File manager is displayed. Here you can easily when and from which program the le was saved. Press t

  • 27 Program values Edge warning When the laser beam is close to the edge, the edge is “lit up” as a warning. It is not possible to register values when you see the edge warning. Half or Zero set value Note the change of the current range Half value Select to half displayed value. Zero line of the PSD moves halfway towards the laser line. Zero set value Select to zero set displayed value. Zero line of the PSD moves to the laser line. Absolute value Select to return to

  • MANUAL English 05-0629 Revision 2.0 System version 6.0 E420 www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

  • 5 DISPLAY UNIT Press and hold the On/Off button to reset the Display unit. Navigation buttons Function buttons Numeric buttons See “LED lights” on page 8 OK button EscapeOn/Off Battery info USB A (master)External power Escape Escape will either return to start view, or close active window (for example le man- ager) and return to previous view. On/Off When you select sleep mode, the Display unit goes into a low-power mode. When in sleep mode, the Display unit restores itself quickly. In sleep mode, the yellow LED light is ashing. CancelShut downSlee

  • 45 VERTICAL The program Vertical is used for vertical and/or ange mounted machines. Preparations 1. Mount the M-unit on the movable machine and the S-unit on the stationary machine. 2. Select and to open Vertical program. 3. Enter distances. Conrm each distance with . If you have a barcode reader, simply scan the barcode and all machine data is read. See “Measurement le handling” on page 11. Function buttons Leave program. See “Control panel” on page 14. Select to enter diameter of coupling. Forward to measure v

  • 9 Display unit Battery Select to display the Battery view. This view gives you a good overview of the battery status of all connected equipment. Display unit Detector or Measuring unit Serial number Battery symbols correspond to the LED signals on the Display unit. Charging The E-series is not compatible with units from the D-series. Charge the Display unit The Display unit can be used from

  • 14 Display unit Control panel Select and to open the Control panel. Some of the settings are personal and will be default next time you start the system. Filter Select to open the Filter view. If the laser beam passes through air with varying temperature, this may inuence the direction of the laser beam. If measurement values uctuate, this could mean unstable readings. Try to reduce air movements between las

  • 44 Softfoot Measure softfoot 1. Tighten all feet bolts. 2. Turn the measuring units to 12 o’clock. 3. Adjust laser to the centre of targets. If needed, adjust the units on the rods, then use laser adjustments knobs. Turn the measuring units to 12 o’clock. Adjust laser line to centre of target. 4. Press OK. The Softfoot measure view is displayed. The rst bolt is marked with yellow. 5. Loosen and then retighten the rst bolt. 6. Press OK to register value. 7. Register values on all four feet. The result is displayed. 8. S

  • 43 SOFTFOOT Perform a softfoot check to ensure that the machine is resting evenly on all its feet. A softfoot can be angular and/or parallel, see image. Softfoot can be caused by: • Twisted machinery foundations. • Twisted or damaged machinery feet. • Improper amount of shims under machine feet. • Dirt or other unwanted materials under machine feet. Start Softfoot from main shaft menu 1. Select and . 2. Enter distances. 3. Select to continue. Start Softfoot from Horizontal program 1. Select and to open Horizontal program. 2. Enter dista

  • 41 Horizontal Tolerance 1. Select and . The tolerance window is displayed. 2. Select a tolerance and press . Add user dened tolerance GAP values Function buttons Close Tolerance view. See “Control panel” on page 14. Edit user dened tolerance. Delete user dened tolerance. Add new tolerance You can add your own user dened tolerance. 1. Select the row “Add new tolerance”. Press . 2. Enter name and tolerance. 3. Press . The new tolerance is a

  • www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

  • 39 Horizontal Adjust Adjust the machine if needed. 1. Shim the machine according to the vertical feet values. 2. Adjust the machine sideways according to the live horizontal values. 3. Tighten the feet. 4. Select to remeasure. Position indicator To adjust, you need to place the measuring units in live position (9, 12, 3 or 6 o’clock). Select to show the Position indicator. Position indicator Function buttons Toggle button. Show/hide position indicator manually.

Questions, Opinions and Exploitation Impressions:

You can ask a question, express your opinion or share our experience of Damalini Easy-Laser E420 device using right now.

Понравилась статья? Поделить с друзьями:
  • Свечи с содой инструкция по применению
  • Кортексин уколы инструкция по применению внутримышечно взрослым для чего цена
  • Рибоксин таблетки инструкция по применению цена отзывы аналоги таблетки цена
  • Кто осуществляет руководство кабинетом охраны труда ответ на тест
  • Элевит 1 триместр инструкция состав для беременных