Manuals and User Guides for Schiller AT-102. We have 1 Schiller AT-102 manual available for free PDF download: Service Handbook

AT-102 ECG Recorder Service Handbook

Contents

AT-102

12-Channel ECG Recorder

Service Handbook

SCHILLER AG

Altgasse 68

6341 Baar, Switzerland

Phone: + 41 41 766 42 42

Fax: + 41 41 761 08 80

www.schiller.ch

i

Article Number 2. 540 028

SCHILLER AG 2002

SCHILLER

AT-102 ECG Recorder Service Handbook

AT-102 Service handbook

Article Number 2. 540 028

Release a

November 2002

Associated Documents

Physicians Guide to the SCHILLER Interpretation and Measurement Program

AT-102 User Guide

The SCHILLER sales and service centre network is worldwide. For the address of your local distributor, contact your nearest SCHILLER subsidiary. In case of difficulty a complete list of all distributors and subsidiaries is provided on our internet site:

http://www.schiller.ch

ii

AT-102 ECG Recorder Service Handbook

Contents

Intended Use

The AT-102 is a 12-channel ECG device used for the recording, analysis and evaluation of ECG Recordings. Recordings made with the AT-102 can be used as a diagnostic aid for heart function and heart conditions. The AT-102 is designed for indoor use and can be used for all patients of both sexes, all races, and all ages.

Physician`s Responsibility

The AT-102 ECG Unit is provided for the exclusive use of qualified physicians or personnel under their direct supervision. The numerical and graphical results and any interpretation derived from a recording must be examined with respect to the patient`s overall clinical condition. Patient preparation and the general recorded data quality, which could affect the report data accuracy, must also be taken into account.

It is the responsibility of the physician to make the diagnosis or to obtain expert opinion on the results, and to institute correct treatment if indicated.

FEDERAL LAW IN THE USA RESTRICTS THIS DEVICE TO SALE BY OR

ON THE ORDER OF A PHYSICIAN

iii

Article Number 2. 540 028

SCHILLER AG 2002

SCHILLER

AT-102 ECG Recorder Service Handbook

AT-102

Service Handbook

Contents

Intended Use………………………………………………………………………..

iii

Physician`s Responsibility………………………………………………………

iii

Terms of Warranty……………………………………………………………………………..

x

Disposal Instructions and Battery Care ……………………………………………..

xi

Safety Notices ……………………………………………………………………..

xii

Operational Precautions …………………………………………………………………..

xii

Precautions for Operation with other Devices …………………………………..

xiii

Maintenance Precautions ………………………………………………………………..

xiv

Symbols and Conventions Used in this Service handbook ………..

xv

What’s in this book ………………………………………………………………

xvi

Section 1 Operating Elements ……………………….

1.1

Introduction ……………………………………………………………………….

1.2

Features ………………………………………………………………………………………….

1.2

Operating Philosophy Overview ………………………………………………………

1.3

Initiating Functions or Tasks ……………………………………………………………

1.3

Main Components of the AT-102……………………………………………………….

1.4

Back Panel ……………………………………………………………………………………..

1.5

Power Supply ………………………………………………………………………………….

1.6

Switching On and Off ………………………………………………………………………

1.6

Changing a Mains Fuse …………………………………………………………………..

1.7

Potential Equalisation ……………………………………………………………………..

1.7

Keypad……………………………………………………………………………………………

1.8

LCD Screen ………………………………………………………………………………….

1.10

iv

AT-102 ECG Recorder Service Handbook

Contents

ECG Settings …………………………………………………………………..

1.12

Auto Format 1 and 2 ……………………………………………………………………..

1.13

ECG Printout ………………………………………………………………………………..

1.13

Average Cycles …………………………………………………………………………….

1.14

Rhythm Leads ………………………………………………………………………………

1.14

Measurements, Markings and Interpretation ………………………………….

1.14

Filters …………………………………………………………………………………………..

1.15

Baseline filter………………………………………………………………………………..

1.15

Myogram filter ………………………………………………………………………………

1.16

Mains filter ……………………………………………………………………………………

1.16

Baseline Stabiliser (SCHILLER SBS) ……………………………………………..

1.16

Smoothing Filter (SCHILLER SSF)…………………………………………………

1.16

Interpretation ………………………………………………………………………………..

1.17

Leads …………………………………………………………………………………………..

1.18

Defining Lead Sequence & Printout ………………………………………………

1.18

Stress Settings …………………………………………………………………

1.20

General Settings …………………………………………………………………………..

1.21

Selecting the ERGO Device …………………………………………………………..

1.21

Blood Pressure Entry ……………………………………………………………………

1.21

Selecting the Default Test Protocol………………………………………………..

1.22

Defining the Stage Printout Format ……………………………………………….

1.22

ST Amplitude Lead………………………………………………………………………..

1.22

Defining / Editing Exercise Protocols …………………………………………….

1.23

Factory programmed Treadmill Protocols………………………………………

1.24

Bruce …………………………………………………………………………………………..

1.24

Factory programmed Bicycle Protocols ………………………………………..

1.24

System Settings ……………………………………………………………….

1.25

Unit ………………………………………………………………………………………………

1.26

User Identification (User ID) …………………………………………………………..

1.26

Date and Time ………………………………………………………………………………

1.26

Language …………………………………………………………………………………….

1.27

Startup Screen ……………………………………………………………………………..

1.27

Paper Mode ………………………………………………………………………………….

1.27

Communication…………………………………………………………………………….

1.28

Test and Information……………………………………………………………………..

1.29

Obtaining a printout of all current settings ……………………………………..

1.30

Communications Test …………………………………………………………………..

1.32

Upgrade/ Update Software ……………………………………………………………

1.32

Default Settings…………………………………………………………………………….

1.33

Unit Defaults Table………………………………………………………………………..

1.34

v

Article Number 2. 540 028

SCHILLER AG 2002

SCHILLER

AT — 102 ECG Recorder Service Handbook

Section 2 Functional Overview

……………………… 2.1

Introduction ……………………………………………………………………….

2.2

MK 18 — 1 Main Board ………………………………………………………….

2.4

Power Supply ………………………………………………………………………………….

2.4

Program and ECG Memory ………………………………………………………………

2.4

Thermal Print Head Controller ………………………………………………………….

2.4

Printer Timing ………………………………………………………………………………….

2.5

Paper Mark ……………………………………………………………………………………..

2.5

Power On Reset ………………………………………………………………………………

2.5

Stepper Motor Controller …………………………………………………………………

2.5

ECG Isolated Power Supplies ………………………………………………………….

2.5

ECG Signal ……………………………………………………………………………………..

2.6

Noise Damping ……………………………………………………………………………….

2.6

RS-232 Interface ………………………………………………………………………………

2.6

External Modem ………………………………………………………………………………

2.6

Top Assembly ……………………………………………………………………

2.7

LCD Screen …………………………………………………………………………………….

2.7

Alphanumerical Keyboard ……………………………………………………………….

2.7

Section 3 Fault Finding …………………………………

3.1

Introduction ……………………………………………………………………….

3.2

General Check Procedures………………………………………………………………

3.2

Fault Finding Chart ……………………………………………………………..

3.3

AT-102 Initial Fault Diagnosis Chart (Sheet 1) ……………………………………

3.3

AT-102 Initial Fault Diagnosis Chart (Sheet 2) ……………………………………

3.4

AT-102 Fault Diagnosis Chart (Sheet A — Power Problems) …………………

3.5

AT-102 Fault Diagnosis Chart (Sheet B — Power Problems) ………………..

3.6

AT-102 Fault Diagnosis Chart (Sheet C — General Problems) ………………

3.7

AT-102

Fault Diagnosis Chart (Sheet D — Printer Problems)………………..

3.8

AT-102

Fault Diagnosis Chart (Sheet F — Exercise Mode Problems) ……

3.9

AT-102

Fault Diagnosis Chart (Sheet G — Spirometry Problems) ………

3.10

AT-102

Fault Diagnosis Chart (Sheet H — Communication

(RS) Problems) ………………………………………………………………………………

3.11

vi

AT-102 ECG Recorder Service Handbook

Contents

Functional Check ………………………………………………………………

3.12

Thermal Printer Check……………………………………………………….

3.13

Print Head Alignment and Print Head Tension………………………………..

3.14

Thermal Printer Fault Diagnosis…………………………………………………….

3.14

RS-232 Interface (and Spiro Check) ……………………………………

3.15

Section 4 Module Removal and Replacement .. 4.1

Introduction ……………………………………………………………………….

4.3

Safety Notices ……………………………………………………………………

4.4

Physical Overview ………………………………………………………………

4.5

Exploded View Lower Casing ………………………………………………

4.6

Exploded View Upper Casing ………………………………………………

4.7

Prerequisites, Test Equipment, Tools, and Accessories …………..

4.8

General Prerequisites ………………………………………………………………………

4.8

Part Numbers ………………………………………………………………………………….

4.8

Opening and Closing the Case …………………………………………….

4.9

Top Assembly Removal……………………………………………………………………

4.9

Top Assembly Replacement …………………………………………………………..

4.11

Main Board MK 18 — 1 ………………………………………………………..

4.12

Parts …………………………………………………………………………………………….

4.12

Board Removal …………………………………………………………………………….

4.12

Board Replacement ………………………………………………………………………

4.14

Printer Tray and Thermal printer ………………………………………….

4.15

Thermal Printer Removal ………………………………………………………………

4.15

Thermal Printer Replacement. ……………………………………………………….

4.15

Battery Pack …………………………………………………………………….

4.16

Battery Pack Removal …………………………………………………………………..

4.16

Checks and Tests After Battery Replacement ………………………………..

4.16

Keyboard …………………………………………………………………………

4.17

LCD screen board…………………………………………………………….

4.18

LCD board Removal ……………………………………………………………………..

4.18

vii

Article Number 2. 540 028 SCHILLER AG 2002

SCHILLER

AT-102 ECG Recorder Service Handbook

Section 5 Adjustments, System Upgrades and

Software Updates ………………………………………….

5.1

Introduction ……………………………………………………………………….

5.3

Safety Notices and Conditions ……………………………………………..

5.4

Conditions ………………………………………………………………………………………

5.4

Test Equipment ………………………………………………………………….

5.5

Proprietary Test Equipment/Tools ……………………………………………………

5.5

Main Board MK 14-10 Adjustment Locations ………………………….

5.6

Component Location MK18-1 (Component Side)………………………………

5.6

Battery Charge Voltage ……………………………………………………….

5.7

Precautions and Requirements………………………………………………………..

5.7

Tools and Equipment ………………………………………………………………………

5.7

Procedure ……………………………………………………………………………………….

5.7

Paper Mark Detector Check …………………………………………………

5.8

Tools, Equipment and Material …………………………………………………………

5.8

Procedure ……………………………………………………………………………………….

5.8

ECG Amplifier +2V, -2V and PWM Ramp Time Adjustment ……

5.10

Tools, Equipment and Material ………………………………………………………

5.10

Procedure …………………………………………………………………………………….

5.10

Service Screen…………………………………………………………………

5.12

Upgrading the Unit / Updating the Software …………………………..

5.14

Installing New Software Options (Upgrade) ……………………………………

5.14

Updating the System Software………………………………………………………

5.16

Section 6 Spare Parts ……………………………………

6.1

Ordering Information …………………………………………………………..

6.2

Spare Parts ……………………………………………………………………….

6.3

viii

AT-102 ECG Recorder Service Handbook

Contents

Section 7 Technical Data ……………………………….

7.1

Technical Data ……………………………………………………………………

7.3

System:…………………………………………………………………………………………..

7.3

Safety Standards: ……………………………………………………………………………

7.4

Technical Data for ECG: …………………………………………………………………..

7.4

Technical Data for Spirometry (Option): ……………………………………………

7.5

Standard …………………………………………………………………………………………

7.7

Configurations ………………………………………………………………………………..

7.7

Annex A Glossary…………………………………………

A.1

Introduction ………………………………………………………………………

A.2

Acronyms …………………………………………………………………………

A.3

Annex B AT-102 Circuit Diagrams and

Engineering Drawings ………………………………….

B.1

Index

ix

Article Number 2. 540 028

SCHILLER AG 2002

SCHILLER

AT-102 ECG Recorder Service Handbook

Terms of Warranty

The CardioLaptop AT-102 is warranted against defects in material and manufacture for the duration of one year (as from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement of the defective part. Any liability for subsequent damage is excluded. The warranty is void if unauthorized or unqualified persons attempt to make repairs.

In case of a defect, send the apparatus to your dealer or directly to the manufacturer. The manufacturer can only be held responsible for the safety, reliability, and performance of the apparatus if:

assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by him, and

the SCHILLER AT-102 and approved attached equipment is used in accordance with the manufacturers instructions.

THERE ARE NO EXPRESS OR IMPLIED WARRANTIES WHICH EXTEND BEYOND THE

WARRANTIES HEREINABOVE SET FORTH. SCHILLER MAKES NO WARRANTY OF

MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO

THE PRODUCT OR PARTS THEREOF.

This equipment has been tested and found to comply with the limits for a class A digital device, pursuant to both Part 15 of the FCC (Federal Communications Commission) Rules and the radio interference regulations of the Canadian Department of Communications. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with this instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

x

AT-102 ECG Recorder Service Handbook

Contents

Disposal Instructions and Battery Care

Do not dispose of battery, boards, or components by fire or incinerator — Danger of

Explosion

Do not open the battery casing — Danger of acid burn

Only dispose of the battery, boards, or components in official recycling centres or municipally approved areas. Alternatively, used batteries and components can be returned to SCHILLER AG for disposal.

Units no longer required can be returned to SCHILLER AG for disposal. Alternatively dispose of the unit in municipally approved recycling centres.

xi

Article Number 2. 540 028

SCHILLER AG 2002

SCHILLER

AT-102 ECG Recorder Service Handbook

Safety Notices

Operational Precautions

Before using the unit, ensure that an introduction regarding the unit functions and the safety precautions has been provided by a product representative.

The guidelines for patient electrode placement are provided as an overview only. They are not a substitute for medical expertise.

IEC 601-1-1 states that the patient must remain at least 1.5 metres clear of the AT-102.

When this is not possible an isolation transformer must be installed.

It must be ensured that neither the patient nor the electrodes (including the neutral electrode) come into contact with other persons or conducting objects (even if these are earthed).

This unit is CF classified and defibrillation protected when the original patient cable is used. However, as a safety precaution when possible,

remove electrodes before defibrillation.

Do not touch the unit casing during defibrillation.

If the patient cable should become defective after defibrillation, lead-off is displayed and an acoustic alarm given.

Do not operate the unit if the earth connection is suspect or if the mains lead is damaged or suspected of being damaged.

This product is not designed for sterile use.

This product is not designed for outdoor use.

Do not use this unit in areas where there is any danger of explosion or in the presence of flammable gases such as anaesthetic agents.

Do not operate the unit if the earth connection is suspect or if the mains lead is damaged or suspected of being damaged.

There is no danger when using the ECG unit for a patient with a pacemaker fitted.

The LCD screen assembly is heavy and can cause injury if closed unintentionally.

Ensure fingers are kept clear.

Surface temperature of applied parts must not exceed 41o.

If the display is damaged, a leakage of fluid may occur. do not inhale the vapour from this fluid and avoid contact with mouth and skin. if contact is made, clean contaminated area immediately with fresh water.

xii

AT-102 ECG Recorder Service Handbook

Contents

Safety Notices

Precautions for Operation with other Devices

Use only accessories and other parts recommended or supplied by SCHILLER AG. Use of other than recommended or supplied parts may result in injury, inaccurate information and/or damage to the unit.

Externally connected units must use the same common earth.

Externally connected units must use an original SCHILLER interface cable.

If several units are coupled, there is a danger of summation of leakage currents. When two or more units are coupled together, an isolation transformer must be used in the mains supply.

The AT-102 complies with EMC regulations for medical products which affords protection against emissions and electrical interference. However, special care must be exercised when the unit is used with high frequency equipment.

There is no danger when using the ECG unit simultaneously with electrical stimulation equipment. However, the stimulation units should only be used at a sufficient distance from the electrodes. In case of doubt, the patient should be disconnected from the recorder.

To avoid possible interference from the Ergometer when carrying out an exercise test, it is recommended that both the AT-102 and the Ergometer are connected to the same common ground.

xiii

Article Number 2. 540 028

SCHILLER AG 2002

SCHILLER

AT-102 ECG Recorder Service Handbook

Safety Notices

Maintenance Precautions

BEFORE CARRYING OUTANY MAINTENANCE PROCEDURES, SWITCH THE UNIT

OFF AND DISCONNECT FROM THE MAINS BY REMOVING THE MAINS PLUG.

The unit is protected by double pole / neutral fusing for continued protection against the risk of fire. Replace only with the same fuse type and rating.

Do not use high temperature sterilisation processes (such as autoclaving). Do not use e- beam or gamma radiation sterilisation.

Do not use solvent or abrasive cleaners on either the unit or cable assemblies.

Do not, under any circumstances, immerse the unit or cable assemblies in liquid.

xiv

AT-102 ECG Recorder Service Handbook

Contents

Symbols and Conventions Used in this Service handbook

The following words and symbols mark special messages throughout this guide.

General Warning. Text set off in this manner indicates that failure to follow directions could result in bodily harm or loss of life or failure to follow directions could result in damage to equipment or loss of information.

WARNING:

Specific Warning. Text set off in this manner indicates that failure to follow directions could result in bodily harm or loss of life.

CAUTION:

Text set off in this manner indicates that failure to follow directions could result in damage to equipment or loss of information.

NOTE:

Text set off in this manner presents clarifying information, specific instructions, commentary, sidelights, or interesting points of information.

xv

Article Number 2. 540 028

SCHILLER AG 2002

SCHILLER

AT-102 ECG Recorder Service Handbook

What’s in this book

The service philosophy for the AT-102 is fault finding to module level. The purpose of this book is to provide all the information necessary to enable the service engineer to efficiently locate and replace a faulty module. This book assumes no detailed knowledge of the AT-102 but does require that the service engineer is familiar with standard workshop practices, and to have attended an AT102 service course. The book is divided into the following chapters:

Chapter 1 — Operating Elements

The purpose of this chapter is to provide an easy reference for all the main operator functions and to give a basic introduction to the AT-102 . This chapter gives details of the operator controls with the operation and function of each key briefly explained. The information in this chapter provides a background to the operating functions only. Complete operating information is provided in the SCHILLER AT-102 User Guide.

Chapter 2 — Functional Overview

This chapter provides a functional overview of the AT-102 . The description is supported by functional block diagrams.

Chapter 3 — Fault Diagnosis

This chapter provides a guide to locate a fault to module level. The diagnostics are presented in a logical sequence of fault finding algorithms and procedures. Illustrations are provided to support the text where needed.

Chapter 4 — Module Removal and Replacement

This chapter gives an overview of the physical construction of the AT-102 with the main physical attributes of the unit briefly described. The physical description is supported by illustrations showing the internal location of all modules. Removal and replacement instructions for all removable modules are also provided in this chapter. Each procedure is autonomous with details of tools, jumper settings, adjustments and settings or special requirements that are required before and after replacement. Functional checks that must be carried out after replacing a module are also provided.

xvi

AT-102 ECG Recorder Service Handbook

Contents

Chapter 5 -Adjustments

This chapter provides all adjustments and settings. Also detailed in this chapter are basic functional test procedures that can be performed to check the functioning of the unit.

Chapter 6 — Spare Parts

This chapter provides the part numbers and reordering information for all replaceable modules. Also included in this chapter are details of any special test equipment or special tools required for adjustment or fault finding procedures.

Chapter 7 — Technical Data

The full technical specification of the AT-102 is given in this chapter.

Annex A — Glossary

This chapter explains all the acronyms and signal titles used in this book and in the AT-102 circuit diagrams.

Annex B — Circuit Diagrams & Board layouts

The circuit diagrams and component layouts are provided for all boards. These details are provided for information only.

xvii

Article Number 2. 540 028

SCHILLER AG 2002

SCHILLER

AT-102 ECG Recorder Service Handbook

xviii

AT-102 ECG Service Handbook

Section 1

Operating Elements

Section 1

Operating Elements

This section contains an introduction to the AT-102 and an overview of all external connections. It also gives an overview of the operating philosophy of the AT-102 and an introduction to the basic functions of the unit. An overview of the system settings are given in this section — for full operating details and system setup see the AT-102 User Guide.

1.1

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

Introduction

The SCHILLER AT-102 is a 12-channel ECG unit designed to record, display, and analyse resting ECGs (exercise ECGs can also be recorded). The unit has been extensively researched to give an ergonomic, clear interface that`s easy to use without compromising functionality. The AT-102 has the following features:

Features

Alphanumeric keypad and dedicated soft key interface for easy, user friendly operation.

Storage and transmission facilities for recordings.

Excercise ECG with interface for control of digital ergometers and treadmills. (Option)

Integral full size thermal quality printer with various user defined print format options. External laser or deskjet printer (Option).

ECG Interpretation including measurements and average cycles with automatic and manual printout of the recording. (Option)

Spirometry (Option)

1.2

AT-102 ECG Service Handbook

Section 1

Operating Elements

Introduction

Operating Philosophy Overview

There are broadly four types of data display as follows

Data Acquisition and

In this screen the real-time ECG is displayed. From this

ECG Recording Screen

screen a continuous printout can be initiated and/or an

auto recording can be made. In auto mode 10 seconds of

ECG data is analysed and averaged and the results given

on a printout. The format and data of an auto mode printout

is independent of the screen display.

An auto mode recording can also be stored in the memory

for later print or transmission.

Memory Screen

In this screen stored recordings can be accessed.

Patient Data Screen

Patient data entry via the keypad.

Data Entry and Setup

In these screens all system settings, resting and exercise

ECG settings, and spiro settings are made.

Initiating Functions or Tasks

Most functions and tasks are initiated by the 5 softkeys situated immediately below the LCD. The function of the softkeys varies according to the screen displayed and is displayed on the LCD immediately above the key itself.

During data acquisition, further dedicated function keys are provided to make an auto mode recording (START) and to stop a manual printout (STOP). The top line of the alphanumeric keypad, additionally enables direct settings of lead group, trace speed and sensitivity, filter on/off and other functions, for both the real-time display and (manual) printout.

1.3

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

Introduction

Main Components of the AT-102

1

2

3

4

5

1.Keypad and dedicated function keys

2.Patient cable connector

3.RS-232 for any of the following:

°connection of an ergo device

°connection of a spiro sensor

°connection of a modem or a PC for export of stored recordings

4.Softkey control

5.LCD Display.

1.4

AT-102 ECG Service Handbook

Section 1

Operating Elements

Introduction

Back Panel

1.LPT connector for the connection of an external printer

2.Master Reset

3.Potential equalisation stud

4.Mains connector (with fuse below)

CAUTION:

All externally connected hardware must be approved by SCHILLER. Connection of any hardware not approved by SCHILLER is at the owner`s risk. The unit guarantee may also be invalid.

1.5

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

Introduction

Power Supply

The mains connection is on the rear of the unit.

The power supply voltage is set by the factory for 100-115V (nom. 110V) or 220-240V (nom. 230V) working. The setting is indicated by the indented metal strip on the fuse panel. Contact your dealer if the voltage needs to be changed.

Switching On and Off

The AT-102 is switched on with the green ON key and off with the red OFF key. These keys are situated on the top right of the keypad.

The mains indicator lamp on the keypad is always lit when the unit is connected to the mains supply.

The unit can either be operated from the mains supply or from the built-in rechargeable battery. The power source is indicated on the top line of

the LCD. When mains is connected a mains symbol is displayed. When the unit is running on battery power a battery symbol is displayed

Power Indicator symbol

The internal battery provides power for up to 3 hours. The Battery indicator blinks when the battery capacity is limited.

To recharge the battery, connect the apparatus to the mains supply by means of the supplied power cable. A totally discharged battery requires less than 15 hours to be fully recharged (60% in less than 3 hours, 90% in less than 7 hours). The unit can remain connected to the mains supply without damage to either the battery or the unit.

NOTE

When working from battery power, the unit is automatically switched off after 5 minutes (30 seconds if battery capacity is limited) if no key is pressed.

1.6

AT-102 ECG Service Handbook

Section 1

Operating Elements

Introduction

Changing a Mains Fuse

CAUTION

If it is necessary to change a fuse, always replace with the correct rating i.e. 2x200mAT for 230V, or 2x315mAT for 110V .

To change a fuse press the retaining lug in the middle of the fuse panel (situated below the

mains connector on the back panel). Remove the fuse panel and replace the fuse(s). Click

the fuse panel back in position.

Potential Equalisation

The potential equalisation stud at the rear of the unit can be used to equalise the ground

potential of the AT-102 to that of all mains powered equipment in the vicinity. Use the

hospital or building common ground.

CAUTION:

To avoid possible interference from the Ergometer when carrying out an exercise test, it is recommended that both the AT-102 and the Ergometer are connected to the same common ground.

To prevent the possibility of leakage current when an external printer is connected, always ensure that the mains lead, or the potential equalisation (next to the mains connector), is attached to the AT-102

A yellow/green ground cable is supplied as an option (Article number 2. 310 005).

1.7

Article Number 2. 540 028

SCHILLER AG 2002

Schiller AT-102 Service Handbook

AT-102 Service Handbook

Introduction

Keypad

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

1.8

AT-102 ECG Service Handbook

Section 1

Operating Elements

Introduction

1.Softkeys — the function of these keys changes depending on the screen displayed. The function of these keys is shown on the screen above the keys. If nothing is written above a softkey, it has no function for the current screen.

2.Auto Mode recording (in Auto mode 1).

Press the Function key (17) followed by the AUTO key (2) for auto mode 2.

3.STOP printout / confirm (new) setting

4.Open / Close paper tray (to replace thermal printing paper)

5.The top figures on the number keys `1` and `2 ` (designated < and >), change the lead group displayed on the screen, forward and backward resp.

6.Auto sensitivity key — automatically sets the ECG printout sensitivity ( in AUTO mode only) to the best setting for the signal strength (5mm/mV or 10mm/mV)

7.The top figures on the number keys designated 5, 10, and 20 set the sensitivity of the ECG both on the screen and on the (manual) printout. The sensitivity is 5, 10 or 20 mm / mV.

8.The top figures on the number keys designated 5/10, 25, and 50 set the speed of the ECG both on the screen and on the (manual) printout. The speed on the screen can only be set to 25 or 50 mm/s. The speed of the manual printout can be 5, 10, 25 or 50 mm/s. The 5 and 10 mm/s settings are both on the same key which toggles the two speeds.

9.Inserts a 1mV reference marker on the screen and printout. Recentres the trace.

10.Toggles the QRS beeper ON/ OFF

11.Myogram filter ON / OFF. The cutoff frequency can be user defined in `Setup`.

12.Delete last typed character.

13.Patient data key. Press this key to enter a new patient or modify the data for the current one.

NOTE:

The patient data screen, or the ECG screen is the first screen displayed on inital switch on. This is set for user preference in the system settings (See following).

14. ON / OFF Keys

15. Mains Indicator — lit when mains connected.

16. Press the function key (17) and the UP/DOWN arrows to adjust screen contrast. When entering patient data use the LEFT/RIGHT arrow keys to move the cursor in the data field. Use the UP/DOWN arrow keys to go up/down to the next data entry

17. Function Key (Fn). When pressed before another key, initiates the second function of that key.

For example, second letters on the keypad (è, é, ç, ø @ etc.), are entered by holding

the function key before pressing the letter key.

1.9

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

Introduction

LCD Screen

The display will vary according to the current task being carried out. In all screens however, the top and bottom lines always display the same information: the top line displays system information (time, patient, power source etc.,), and the bottom line always gives the softkey options.

The following is an example of a typical resting ECG screen.

3

1

4

5

6

7

8

9 10

2

Items 1, 2 and 3 are in the same position for all screens.

1.Top line — time, date, patient name, and current power source — mains (), or battery (). When battery capacity is limited the battery symbol flashes.

2.Softkey designation. Pressing the key below the text carries out the function indicated. The options available will change according to the screen displayed.

3.Data acquisition area or data entry area.

1.10

AT-102 ECG Service Handbook

Section 1

Operating Elements

Introduction

Items 4 to 10 are specific for ECG acquisition only:

4.Current Heart Rate (averaged over 4 beats and refreshed every 2 seconds). The HR is also given on a manual printout. Note that with an auto mode printout the HR is averaged over the full 10 seconds of the recording.

5.Electrode connections — when a lead flashes it indicates that the electrode resistance is too high. The electrode(s) must be reapplied.

6.Sensitivity — 5, 10 or 20 mm/mV. Change the sensitivity with the keys 3 (auto), 4, 5 and 6. An `A` in this box indicates that automatic sensitivity is selected (auto mode printout only).

7.Speed — 25 or 50 mm/s. Change the speed with the keys 8 and 9.

8.Lead indication (leads currently displayed on the screen). Change the lead group with the < and > keys on the keypad.

9.Myogram Filter indication — ‘Filter ON’ or ‘Filter OFF’. The filter is applied with the filter key.

Note: the frequency of the filter cutoff is defined in Section 4 Setup.

10.Area for system messages or instructions.

1.11

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

ECG Settings

The AT-102 ECG and system settings are entered by selecting ‘setup’ from the initial

screen:

The following pages detail the programmable ECG parameters.

NOTE:

In units where the interpretation option is not installed, interpretation statements, cannot be displayed.

1.12

AT-102 ECG Service Handbook

Section 1

Operating Elements

ECG Settings

Use the select softkey to select the different settings

Use the Up/Down softkeys to highlight the various options.

Auto Format 1 and 2

Two separate Auto formats can be defined for the AT-102. Use the NEXT softkey to confirm setting and to move onto the next screen (Format 2).

ECG Printout

Press the `SELECT` softkey to choose from the following options:

No Printout

No printout of the ECG given at the end of an auto mode recording (the

recording can be stored in the memory and printed at a later time if

required).

4*3 + 1 Rhythm

Leads are printed in a 4 * 3 format at 25mm/s, with the selected rhythm

lead at the bottom of the page at 25mm/s.

1*12 at 25mm/s

Leads are printed in a 1 * 12 format at 25mm/s — no rhythm lead printed.

1.13

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

ECG Settings

8*5s + 4*10s

Short at 25mm/s Long at 25mm/s Long at 50mm/s

The first 8 leads printed for 5 seconds and the last 4 leads printed for 10 seconds.

Leads are printed in short form (1 sheet) at 25mm/s.

Leads are printed in long form (2 sheets) at 25mm/s.

Leads are printed in long form (2 sheets) at 50mm/s.

Average Cycles

No Printout No printout of average cycles

4*3, 25mm/s + 2 Rhythm Leads are averaged over the entire 10 second recording and printed in 4 groups of 3 leads at 25mm/s, with the two selected rhythm leads at the bottom of the page at 25mm/s.

4*3, 50mm/s + 2 Rhythm Leads are averaged over the entire 10 second recording and printed in 4 groups of 3 at 50mm/s, with the two selected rhythm leads at the bottom of the page at 25mm/s.

2*6, 50mm/s + 2 Rhythm Leads are averaged over the entire 10 second recording and printed in two groups of six at 50mm/s, with the two selected rhythm leads at the bottom of the page at 25mm/s.

Rhythm Leads

Select the rhythm lead 1 and rhythm lead 2 as described above.

Rhythm Lead 1

Select any lead (I, II, III, aVR, aVl, aVF, V1 to V6)

Rhythm Lead 2

Select any lead (I, II, III, aVR, aVl, aVF, V1 to V6)

Measurements, Markings and Interpretation

Measurements

Select yes or no to print a detailed table of measurement results

Markings

Select yes or no to print reference markings on the ECG average

cycle print. A vertical marker shows the beginning and end of P wave

and QRS, and the end of the T wave

Interpretation

Select yes or no to print interpretation statement

Full details of the interpretation option are given in the SCHILLER ECG Measurement and Interpretation booklet (art. No. 2.510 179).

1.14

AT-102 ECG Service Handbook

Section 1

Operating Elements

ECG Settings

Filters

There are five different filters which can be set individually as follows:

Baseline filter

The cutoff frequency of the filter is set on the top line. The cutoff can be 0.05Hz, 0.15Hz or

0.3Hz.

Note:

The set value is the lower limit of the frequency range and is normally set to 0.05 Hz. The settings 0.15 and 0.30 Hz should only be used when absolutely necessary, as the possibility exists that they could affect the original ECG signal, especially the ST segments.

1.15

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

ECG Settings

Myogram filter

The Myogram filter suppresses disturbances caused by strong muscle tremor. The filter is applied by pressing the FILTER key (or programmed on as default when the unit is switched on).

When the Myogram filter is on, `FILTER ON` is displayed in the information box.

The cutoff frequency is user defined at 25Hz or 35Hz. When `off at power up` is selected, the Myogram filter is off when the unit is first switched on.

Note: An ECG recorded in auto mode is stored unfiltered. It is therefore possible to print the stored ECG either with or without passing the myogram filter. Filter ON is indicated on the LCD.

Mains filter

The mains filter is an adaptive digital interference filter designed to suppress ac interference without attenuating or distorting the ECG.

Set the mains filter in accordance with the frequency of your local mains supply.

Baseline Stabiliser (SCHILLER SBS)

The baseline stabiliser greatly reduces the baseline fluctuations without affecting the ECG signal. The purpose of the stabiliser is to keep the ECG signals on the baseline of the printout. This filter is only effective in auto mode printout.The Baseline Stabiliser is applied to a recording (on), or not applied to a recording (off).

Smoothing Filter (SCHILLER SSF)

The smoothing filter (SSF: SCHILLER smoothing filter) is a low pass filter to suppress high frequency artefacts between the QRS complexes. When this filter is switched on, `SSF` is shown on the bottom line of the automatic printout.

1.16

AT-102 ECG Service Handbook

Section 1

Operating Elements

ECG Settings

Interpretation

The interpretation settings enable the user to determine whether or not certain comments

will be added to the interpretation statements on the ECG printout. Furthermore, the

patient’s age can be assumed (<30 or >30). Low or high can also be set for interpretation

sensitivity. Low sensitivity will suppress certain non-specific ECG diagnoses; this may be

advisable when carrying out ECGs for screening.

Sensitivity

High or low sensitivity

Age Assumed to be

Greater than 30 years, or 30 years and under.

NOTE:

The `Patient age assumed to be..` setting is only applicable when patient data has not been entered. When a patient`s date of birth has been entered, this setting is ignored.

Abnormal ECG

`Normal` / `Abnormal` is printed or not printed

Unconfirmed Report

`Unconfirmed Report` is printed or not printed

1.17

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

ECG Settings

Leads

Defining Lead Sequence & Printout

The required settings can be selected as follows:

Lead Sequence

Select between:

Standard lead sequence or

Cabrera lead sequence

Signals

Select between:

Simultaneous — all ECG leads are printed in the same time segment

(in automatic mode only), or

Sequential — each group is a contiguous time segment of

approximately 2.5 or 5 seconds (in automatic mode only).

Auto-Centering

Select between:

On — all ECG traces are centred dynamically for optimal use of paper

width, or

Off — ECG traces are set to a fixed baseline position and may

possibly overlap.

1.18

AT-102 ECG Service Handbook

Section 1

Operating Elements

ECG Settings

The lead group settings allow extra leads to be displayed on the screen when set to `on`.

The following lead groups can be displayed:

Rhythm Lead Group

II, avF, III

Left Posterior (V4-V9)

V4, V5, V6

Right Precordials (V5r)

V1, V2, V3

Right Precordials (V6r)

V1, V2, V3r

Nehb (D, A, J)

D, A, J

/V2, V4, V5

/V7, V8, V9

/V3r, V4r, V5r

/V4r, V5r, V6r (only three leads)

The above leads can also be printed when displayed (only in manual mode)

The lead groups are changed both on the screen and on the manual printout with the lead next/previous keys:

1.19

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

Stress Settings

The AT-102 Stress ECG settings are entered by selecting ‘Setup’ and ‘Stress Settings’ from the initial screen:

The following pages detail the stress settings for the AT-102.

1.20

AT-102 ECG Service Handbook

Section 1

Operating Elements

Stress Settings

General Settings

Selecting the ERGO Device

Ergo devices available for use with the AT-102 are as follows:

Bikes:

The Ergoline digitally controlled exercise bicycle 900 / 911

SECA CT100 mod.545

Treadmills:

The Ergoline digitally controlled treadmill TM435/TM4000ES

RAM 770CE

MTM-1500

Blood Pressure Entry

Select between Off, Auto and Manual. When Manual is selected the BP screen (for entry of blood pressure), is displayed 55 seconds before the end the stage. When the stages are less than one minute long, or a stage is held, then the BP screen is displayed every 2 minutes.

The AUTO function is not available at the time of print.

1.21

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

Stress Settings

Selecting the Default Test Protocol

Two protocols for a bike and two for a treadmill are available for selection when starting a stress test — the two protocols displayed when starting a stress test are defined here.

The user defined protocols (one for a treadmill and one for a bike) are defined by the user (see following page).

Defining the Stage Printout Format

A stage printout is given at the end of every stage. When the stages are less than 2 minutes long, or a stage is held, then a prinout is given every 2 minutes.

The waveforms are averaged to give 4 x 3 plus 1 rhythm lead, or 6 x 2. The rhythm lead printed is the ST lead defined below.

ST Amplitude Lead

Define the lead that is used for ST measurement. This lead is also printed as the rhythm lead if set in the stage printout.

1.22

AT-102 ECG Service Handbook

Section 1

Operating Elements

Stress Settings

Defining / Editing Exercise Protocols

For a bicycle the following can be defined:

Protocol Name — The name defined here appears in the general stress settings (when selecting the two default bike protocol (see previous page)

Base Load — The load in Watts, applied during the warm-up period

Step load — The load increase at every step

Recovery Load — The load applied during the recovery phase

Step Interval — The duration of each step

For a treadmill Protocol, the following can be set:

Protocol Name — As above

Step Interval — The duration of each step

Speed — The individual treadmill speed of each step

Elevation — The individual treadmill elevation of each step

1.23

Article Number 2. 540 028

SCHILLER AG 2002

AT-102 Service Handbook

Stress Settings

Factory programmed Treadmill Protocols

One factory programmed Treadmill protocols is availbale as follows:

Bruce

Stage

Duration

Speed

Elevation

1

3 min

2.7km/h (1.7mph)

10%

2

3 min

4.0km/h (2.5mph)

12%

3

3 min

5.4km/h (3.4mph)

14%

4

3 min

6.7km/h (4.2mph)

16%

5

3 min

8.0km/h (5.0mph)

18%

6

3 min

8.8km/h (5.5mph)

20%

7

3 min

9.6km/h (6.0mph)

22%

Factory programmed Bicycle Protocols

One factory programmed Bicycle protocols is available 252525. This gives a base load (warmup) of 25W (50W), with a step load for both protocols of 25W and a recovery load of 25W.

1.24

Loading…

Напишите нам!

Копия запроса будет направлена на указанную вами почту

Организация

Город

ФИО

E-mail

Рабочий телефон

Должность

Мобильный телефон

Прикрепить файл

Сообщение

Обязательные поля


Нажимая кнопку «Отправить», я даю свое согласие на обработку моих персональных данных, в соответствии с Федеральным законом от 27.07.2006 года №152-ФЗ «О персональных данных», на условиях и для целей, определенных в Согласии на обработку персональных данных

Электрокардиограф — устройство, которое в обязательном порядке должно быть в любом медицинском центре. Аппарат ЭКГ позволяет проводить исследование сердца, которое является ключевым при диагностике различных заболеваний. Одними из наиболее популярных на сегодняшний день являются модели компании Schiller. «Лассамед» напрямую работает с производителем и предлагает вам купить электрокардиографы 12 канальные по выгодной цене.

Особенности устройства

ЭКГ-аппарат 12 канальный Cardiovit AT-102 — это полноценная портативная рабочая станция, которая позволяет проводить все основные тесты в любом месте. Это позволяет использовать ее не только в медицинских центрах, но и на выезде. Cardiovit AT-102 дает возможность проводить стандартный ЭКГ покоя, при этом получая данные одновременно с 12 каналов. В результате специалист видит полную картину работы сердца пациента и существенно экономит время. Для распечатки результата в устройстве предусмотрен встроенный термопринтер. Модель также оснащена памятью, в которой можно хранить результаты 40 последних исследований. В комплекте с Cardiovit AT-102 также идут кабели, обеспечивающие возможность подключения к ПК и внешним принтерам по стандатным протоколам.

Еще одним важным преимуществом модели является возможность выбора источника питания. Кардиограф может работать как от сети, так и от аккумулятора.

Cardiovit AT-102 управляется с интегрированной цифровой и кнопочной клавиатуры. Он имеет два режима работы: ручной, при котором результаты интерпретирует специалист, и автоматический, позволяющий прибору самостоятельно определять грубые нарушения работы сердца.

Чтобы купить 12 канальный электрокардиограф Schiller и другие приборы из нашего каталога, заполните форму в этом разделе, закажите обратный звонок на сайте или свяжитесь с нами по телефону в Москве +7 (495) 946-94-42.

Основные характеристики 12-канального ЭКГ аппарата:

  • 6/12-канальный электрокардиограф
  • Синхронная регистрация 12 отведений

  • ЖК экран
  • Алфавитно-цифровая клавиатура
  • Автоматический и ручной режимы работы
  • Печать отчета на встроенном термопринтере (формат А4)
  • Память на 40 ЭКГ
  • Программа «М» — программа измерений (интервалы, амплитуды, электрические оси, усредненные комплексы)
  • Определение пейсмейкера
  • Работа от сети и аккумулятора
  • Интерфейс для передачи данных
  • Интерфейс для внешнего принтера

Возможности:

Электрокардиограф двенадцатиканальный CARDIOVIT АТ-102 переносная диагностическая рабочая станция: ЭКГ покоя, нагрузки и спирометрия. Стресс-система на базе электрокардиографа CARDIOVIT АТ-102 в настоящее время является одной из самых доступных систем с базовой программой ЭКГ нагрузки.


Дополнительно:

  • Программа «С» — программа анализа (интервалы, амплитуды, электрические оси, усредненные комплексы, интерпретационные сообщения для взрослых и детей)
  • Программа ЭКГ нагрузки
  • Программа тромболизиса
  • Программа SEMA — передача и хранение данных
  • Спирометрия (при подключении спирометрического датчика)
  • Эргометры ERG 911S, 911BP
  • Внешний принтер HP DeskJet 
  • Пациентный кабель с клипсовидными наконечниками
  • Одноразовые электроды для ЭКГ нагрузки
  • Загубники одноразовые пластиковые (10 шт/уп) для датчика SP-250
  • Загубники одноразовые картонные (100 шт/уп) для датчика SP-260
  • Съемные фильтры (100 шт/уп) для датчика SP-260
  • Носовые зажимы
  • Калибровочный насос
  • Пояс для эргометрии
  • Система вакуумной аппликации электродов
  • Тележка

Комплектация поставки 12-канального электрокардиографа:

  • Пациентный кабель
  • Комплект многоразовых электродов для взрослых (6 грудных, 4 конечностных)
  • Регистрирующая бумага АТ-102 (стартовый набор) 
  • Гель (стартовый набор) 

Schiller AT-101

Schiller AT-101 — это 12-канальный электрокардиограф, предназначенный для регистрации, визуализации и анализа ЭКГ покоя. Эргономичный интерфейс прибора обеспечивает простоту управления.

Schiller AT-101

Стандартные характеристики Schiller AT-101

  • Буквенно-цифровая клавиатура и выделенные функциональные клавиши для простого удобного управления.
  • Встроенный высококачественный термопринтер с различными опциями формата печати, определяемыми пользователем.
  • Измерения и усредненные циклы с возможностью автоматической и ручной распечатки регистрации.

Дополнительные возможности Schiller AT-101

  • Внешний принтер
  • Интерпретация ЭКГ
  • Память на 40 регистраций
  • Тромболизис
  • Коммуникационный модуль SCHILLER (SCM) для передачи данных на сервер через:
    • сеть Ethernet
    • модем (опция)

Обзор Schiller AT-101

Экран получения данных и регистрации ЭКГ

В этом экране визуализируется ЭКГ в режиме реального времени. Из этого экрана возможен запуск длительной печати и/или произведена автоматическая регистрация. В автоматическом режиме 10 секунд регистрации ЭКГ анализируются и усредняются и результаты выводятся на печать. Формат и данные для автоматической распечатки не зависят от представления экрана и определяются в экранах настройки. Регистрация, произведенная в автоматическом режиме, также может быть сохранена в памяти для последующей распечатки или передачи.

Экран памяти

В этом экране возможен доступ к сохраненным регистрациям, их распечатка или передача.

Экран данных пациента

Ввод данных пациента через клавиатуру.

Функции

Большинство функций или заданий активируются при помощи 5 выделенных клавиш, расположенных непосредственно под ЖК-дисплеем. Функции выделенных клавиш варьируются и зависят от текущего экрана; активная функция клавиши визуализируется на экране непосредственно над клавишей.

Schiller AT-101 экран

В процессе получения данных доступны дополнительные выделенные клавиши для запуска регистрации ЭКГ в автоматическом режиме (ПУСК) и остановки ручной распечатки (СТОП). Верхний ряд буквенно-цифровой клавиатуры позволяет производить прямые установки группы отведений, скорости и чувствительности кривой, фильтра и других функций, как для экрана в режиме реального времени, так и для распечатки в ручном режиме

Основные компоненты Schiller AT-101

Schiller AT-101 обзор
  1. ЖК экран
  2. Выделенные клавиши
  3. Клавиатура и выделенные функциональные клавиши
  4. Принтер

Задняя панель

Schiller AT101 задняя панель
  1. Разъем для кабеля пациента
  2. Разъем LPT для подсоединения внешнего принтера
  3. RS-232 для подсоединения модема или ПК для экспорта сохраненных регистраций
  4. Сетевой разъем (с предохранителем)
  5. Разъем RJ11 с SCM (опция)
  6. Вывод заземления
  7. Разъем RJ45 для сети Ethernet с SCM (опция)

Клавиатура Schiller AT-101

Экран ЭКГ

1. Выделенные клавиши — функция этих клавиш изменяется в зависимости от представленного экрана. Функция клавиши указывается на экране над клавишей. Если над клавишей не появляется пояснение ее функции, это означает, что клавиша неактивна в этом экране.

2. Регистрации ЭКГ в автоматическом режиме (в автоматическом режиме 1). Нажмите клавишу Регистр (SHIFT), затем клавишу ПУСК (2) для запуска автоматического режима 2.

3. Остановка печати.

Буквенно цифровая клавиатура

Schiller AT101 клавиатура

4. Верхний регистр клавиш «1» и «2» (обозначенный как < и >) изменяет группу отведений на экране на следующую или предыдущую соответственно.

5. Клавиша автоматической чувствительности — автоматически устанавливает чувствительность печати ЭКГ (только в автоматическом режиме) для подбора оптимальной силы сигнала (5мм/мВ или 10мм/мВ)

6. Верхний регистр цифровых клавиш имеет обозначения 5, 10 и 20 для выбора чувствительности ЭКГ на экране и на (ручной) печати. Возможные установки чувствительности 5, 10, 20 мм/мВ

7. Верхний регистр цифровых клавиш имеет обозначения 5/10, 25 и 50, которые являются установками скорости ЭКГ на экране и (ручной) печати. Скорость экрана может быть установлена только на 25 или 50 мм/с. Скорость распечатки в ручном режиме может быть 5, 10, 25 или 50 мм/с. Установки скорости 5 и 10 мм/с выбираются одной клавишей, которая последовательно переключается между этими установками.

8. Вставка справочного маркера 1 мВ на экране и на распечатке. Перецентровка кривой.

9. Включение / выключение бипера QRS.

10. Включение / выключение миографического фильтра. Отсекающая частота устанавливается в экране настройки.

11. Удаление последнего введенного символа.

12. Клавиша данных пациента. Нажмите эту клавишу для ввода данных нового пациента или изменения данных текущего пациента.

13. Клавиша включения /выключения.

14. Индикатор сети — горит, если прибор подключен к сети.

15. Нажмите функциональную клавишу (17) и используйте клавиши вверх / вниз для настройки контрастности экрана. При вводе данных пациента используйте клавиши вправо / влево для передвижения курсора в поле данных. Используйте стрелочки ВВЕРХ/ВНИЗ для перехода к следующей/предыдущей позиции.

16. Клавиша Регистр для выбора заглавных букв; в AT-101 Tele — включение / выключение режима GSM.

17. Функциональная клавиша. При нажатии перед другой клавишей, активирует вторую функцию этой клавиши.

Экран Schiller AT-101

Представление экрана будет варьироваться в зависимости от текущего задания. Однако во всех экранах верхняя и нижняя строки содержат одинаковую информацию: в верхней строке приводится системная информация, а нижняя строка содержит данные о функции выделенных клавиш.

Экран ЭКГ
  1. Верхняя строка — время, дата имя пациента и текущий источник питания — сеть или аккумулятор. Когда мощность аккумулятора снижается, символ мигает.
  2. Область получения данных или область ввода данных.
  3. Обозначение выделенных клавиш. Нажатие клавиши под текстом активирует указанную функцию. Доступные опции изменяются соответственно выбранному экрану.
  4. Текущая ЧСС (усредняется каждые 4 сокращения и обновляется каждые 2 сек.). ЧСС приводится также на распечатке в ручном режиме. Учтите, что в авто режиме ЧСС усредняется за каждые 10 сек. регистрации.
  5. Подсоединение электродов — если индикатор отведения мигает (и раздается акустический сигнал тревоги), это означает, что сопротивление электрода слишком высоко. Электрод должен быть переустановлен.
  6. Чувствительность — 5, 10, 20 мм/мВ. Изменение чувствительности осуществляется при помощи клавиш 3 (авто), 4, 5 и 6. Символ A в этом поле свидетельствует о том, что выбрана установка автоматической чувствительности (активна только при автоматическом режиме печати).
  7. Скорость — 25 или 50 мм/с. Изменение скорости с помощью клавиш 8 и 9.
  8. Индикатор отведений (текущих отведений на экране). Изменить группу отведений можно при помощи клавиш < и > на клавиатуре.
  9. Индикатор миографического фильтра: «Фильтр включен» или «Фильтр выключен». Фильтр активируется при помощи клавиши фильтра.
  10. Область для системных сообщений или инструкций.

Аккумулятор Schiller AT-101

Прибор может работать как от сети, так и от встроенного аккумулятора. Источник питания указывается в верхней строке экрана. Встроенный аккумулятор гарантирует до 3 часов работы.

  • Когда прибор работает от аккумулятора, визуализируется символ батареи.
  • При работе от батареи прибор автоматически выключается через 5 минут (30 секунд, если емкость батареи ограничена), если не нажимается ни одна клавиша.
  • В процессе нормальной эксплуатации обслуживание не требуется. Если прибор не используется свыше 3 месяцев, батарея должна быть заряжена. Заменяйте батарею примерно каждые 4 года (в зависимости от частоты использования), если время работы от батареи снизится до 1 часа и менее.
  • Прибор может оставаться подключенным к сети без риска повреждения батареи или прибора.

Для зарядки полностью разряженной батареи до 90% требуется около 7 часов. Прибор может использоваться в процессе зарядки батареи. Однако в этом случае время зарядки батареи существенно увеличится.

Заправка бумаги в Schiller AT-101

Schiller AT-101 бумага

Прибор поставляется без бумаги в лотке для бумаги. Используйте только оригинальную бумагу SCHILLER. Термобумага чувствительна к жаре, влажности и химическим веществам. Храните бумагу в прохладном и сухом месте.

  1. Нажмите клавишу открытия лотка для бумаги (1).
  2. Откройте крышку лотка для бумаги.
  3. Вставьте бумагу.
  4. Закройте крышку. Убедитесь, что бумага находится точно между направляющими.
  5. Нажмите клавишу СТОП, чтобы передвинуть бумагу в требуемое положение.

Термочувствительная бумага, используемая на AT-101, требует условий, несколько отличающихся от условий хранения обычной бумаги, так как она может реагировать на химикаты и тепло.

При выполнении следующих пунктов можно обеспечить нормальные условия для хранения бумаги:

  1. До использования храните бумагу в фирменной картонной упаковке. Не снимайте картонную упаковку до начала использования бумаги.
  2. Храните бумагу в прохладном, темном и сухом месте.
  3. Не храните бумагу вблизи химикатов, например, стерилизационных растворов.
  4. Не храните бумагу в пластиковых коробках.
  5. Некоторые виды клея могут вступать в реакцию с бумагой — не приклеивайте распечатку на картонное основание при помощи клея.

Скачать инструкцию на Schiller AT-101

Скачать инструкцию и другую документацию на Schiller AT-101 можно здесь.

Руководство пользователя ( user manual ) Schiller AT-101 скачать.

Регистрационное удостоверение Schiller AT-101 скачать.

Так же смотрите сфигмоманометр Schiller BR-102 plus.

Понравилась статья? Поделить с друзьями:
  • Как найти руководство в dota 2
  • Протоиерей никольский руководство к изучению православного богослужения
  • Колонка jbl charge 3 инструкция на русском по применению
  • Иноферт инструкция по применению при планировании беременности форум
  • Как собрать поперечный маятник на детской кроватке инструкция