Front page
FLOWSIC100 Flare
Description
Installation
Operation
O P E R A T I N G I N S T R U C T I O N S
Related Manuals for SICK FLOWSIC100 Flare
Summary of Contents for SICK FLOWSIC100 Flare
-
Page 1
Front page O P E R A T I N G I N S T R U C T I O N S FLOWSIC100 Flare Description Installation Operation… -
Page 2
SICK Engineering GmbH assumes no liability for the correctness of an unauthorized translation. Please contact the publisher in case of doubt. Legal information Subject to change without notice. © SICK Engineering GmbH. All rights reserved. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 3
Hazard which could result in material damage. Information Symbols Information about use in potentially explosive atmo- spheres. Important technical information for this product Supplementary information Link to information in another place FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V2-5/2018-10 · © SICK Engineering GmbH… -
Page 4: Table Of Contents
Analog output of check cycle……….73 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 5
Connection of relay contacts ……….149 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V2-5/2018-10 · © SICK Engineering GmbH… -
Page 6
Maintaining the MCUP for use zone 2 ……..208 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 7
SICK Commissioning check list …….. -
Page 8
Contents FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 9: Important Information
Important Information FLOWSIC100 Flare Important Information About this document Main hazards Intended use Responsibility of user Transport safety device for retractable sender/receiver units FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 10: About This Document
Service Manual for the FLOWSIC100 Flare measuring system. NOTICE: Always read the Operating Instructions before starting any work! The observance of all safety and warning notes is imperative! FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 11: Main Hazards
SIC100 Flare specified for such use (→ pg. 13, § 1.3.2). Observe the information on → pg. 11, § 1.2.1 during installation work on run- ning equipment («hot tapping» method). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 12: Hazards Due To Electrostatic Discharges
Use appropriate caution when performing maintenance and cleaning work. For example, the surfaces should only be cleaned with a damp cloth. The affected devices will be identified by the manufacturer by a warning label. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 13: Intended Use
Important Information Intended use 1 . 3 Purpose of the device 1.3.1 The FLOWSIC100 Flare measuring system may only be used for measurements of gas velocity, gas volume, mass flows and molecular weight in pipelines. Correct use 1.3.2 General 1.3.2.1 …
-
Page 14
Option low temperature version with minimum temperature -196 °C (not available for EX/EX-RE Zone 1 and CSA) Option minimum temperature -50 °C Limit for FLSE100 Standard temperature +180 °C, limit for FLSE low temperature version +100 °C FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 15
In this case the FLSE100 must be repaired by the manufacturer only. Warranty will be expired. Compliance with the above requirements is the sole responsibility of the user. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 16
Install and operate the MCUP control unit without explosion protection only outside Ex zones. Use the explosion-protected MCUP control unit only according to the respective Ex marking. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 17: Restrictions Of Use
Leak-tightness is to be checked reg- ularly during operation and the seal replaced, as required. Before every re- installation new seals have to be used in the required design. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 18
Important Information Application limitations for use in hazardous area classification Ex zone 1 The following additional regulations are applicable for the use of FLOWSIC100 Flare in applications with Ex zone 1: Ultrasonic probes made of titanium may be used in zone 1 only when risks of ignition arising from impacts or friction on the sensor housing can be ruled out. -
Page 19: Responsibility Of User
1.4.1 Designated users The FLOWSIC100 Flare measuring system may only be installed and operated by skilled technicians who, based on their technical training and knowledge as well as knowledge of the relevant regulations, can assess the tasks given and recognize the hazards involved.
-
Page 20: Safety Information And Protective Measures
Repairs on explosion-protected system components are possible only by the device manufacturer and, to a limited extent, also by others. For detailed information about possible repairs and repair instructions see the service manual. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 21: Transport Safety Device For Retractable Sender/Receiver Units
The probe tube with transducer must be secured with cotter pins. Fig. 1 Transport safety device for retractable sender/receiver unit (shown semitransparent) Retraction flange Probe tube Cotter pin Transducer FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 22
Important Information FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 23: Product Description
Product Description FLOWSIC100 Flare Product Description System basic information System components Computations Check cycle FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 24: System Basic Information
Functional principle 2.1.1 The FLOWSIC100 Flare gas velocity measuring devices operate according to the principle of difference measurement in ultrasonic transit times. On both sides of a pipeline, the sender/receiver units are installed in a certain path angle to the gas flow (→ Fig. 2).
-
Page 25
The volume flow in operating mode is calculated from the gas velocity and the geometric dimensions of the pipeline. → pg. 70, § 2.3.2 describes the calculations for volume flow in standard state, mass flow and molecular weight. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 26: System Configuration
In this way the path failure can be temporarily compensated and the measurement is continued with slightly increased uncertainty. Under such conditions, the measurement system automatically signals «Maintenance request». FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 27
Three individual measuring paths from different measuring points can be connected to measurement one control unit with this variant. Separate processing and output of the measuring results for all three measuring points are performed by the MCUP. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 28: Communication Between Sender/Receiver Units And Control Unit
MCUP Junction box An approved explosion-protected junction box (SICK option) is to be used for cabling in Ex zone 2. Cabling in hazardous areas Ex zone 1 may be carried out only with an approved «Exe» junction box or «Ex-d» junction box (not included in SICK scope of delivery).
-
Page 29: System Overview
● Dry calibrated (3D measured measurement system. Geometric uncertain- ties are minimized.) ● Flow calibrated (Complete measurement system calibrated on a flow test bench.) Both variants reduce the measurement uncertainty (→ pg. 210, § 6.1). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 30
Ex zone 2 α MCUP control unit certified only for Ex zone 2 Connection cable Sender/receiver unit FLSE100-EXS Slave Potentially explosive atmosphere zone 1 or zone 2 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 31: System Components
9 Connection for slave probe (only for FLOWSIC100 EX-S) 4 Pressure balance element (not removable) 10 Ground terminal 5 Connection flange 11 Fixing ring 6 Duct probe 12 Supporting ring FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 32
All relevant safety regulations, standards and guidelines must be considered when using the condensation drain. The use of the condensation drain is the sole responsibility of the plant operator. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 33
All sender/receiver units can be used for internal pipe pressure up to max. 16 bar. If requested by the customer, the sender/receiver units are also available in other versions. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 34
Sealings in retraction mechanism (not for high PTFE temperature versions) Sealings in retraction mechanism (high temperature FFKM, PTFE versions only) PTFE Centering Sensor contour PTFE, conductive Supporting ring FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 35
Flange connection prepared for mounting on counter flange 3″ CL150 RF acc. to ASME B16.5 Flange connection prepared for mounting on counter flange DN80 PN16 form B1 acc. to EN1092-1 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 36
Not for Ex devices Metric cable gland Housing cable entry NPT cable gland Without junction box Junction box With junction box not suitable for armored cables, not for IECEX FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 37
Versions for low-temperature range -196 … +100°C on request For information about permissible sample gas pressure/temperature range, see also → pg. 41, Fig. 7, → pg. 42, Fig. 9, → pg. 43, Fig. 11, → pg. 44, Fig. 13. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 38
Available on request +++: very good ++: good +: conditional good —: not suitable SS: stainless steel group 4 Ti: titanium grade 2 / grade 5 HS: hastelloy FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 39
For retractable ultrasonic transducers only. indicated in the type code Leaking O-rings may only be changed by SICK service. +++: very good ++: good +: conditional good —: not suitable FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 40
Fig. 6 FLSE100-EXS sender/receiver units, non-retractable version Probe with 2-channel electronics (master) Probe without electronics (slave) Modules/parts not named: → pg. 31, Fig. 5 Connection to master probe FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 41
1 Connection for optional venting2 1 Connection for optional venting2 Duct probe RE Duct probe RE 3 Retraction flange4 3 Retraction flange4 Connection to master probe Connection to master probe FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 42
Range for high- temperature version Range for standard temperature ball valve and retraction flange Temperature in °C NOTICE: Range for low temperature version see p. 45, Fig. 14. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 43
Pressure in bar Range for standard temperature version Range for high- temperature version Temperature in °C NOTICE: Range for low temperature version see p. 45, Fig. 14. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 44
Range for high- temperature version Range for standard temperature ball valve and retraction flange Temperature in °C NOTICE: Range for low temperature version see p. 45, Fig. 14. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 45
Product Description Low temperature versions Permissible sample gas pressure/temperature range for sender/receiver units in low temperature Figure 14 version, retractable and non-retractable versions FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 46
Modules/parts not named: 1 Electronic unit in Exd housing → pg. 31, Fig. 5 2 Connection Ex ia to slave probe 3 Connection Ex ia to master probe FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 47
1 Electronic unit in Exd housing4 Connection Ex ia to slave probe 2 Connection for optional venting5 Retraction flange 3 Duct probe RE6 Connection Ex ia to master probe FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 48
FLSE100-EXRE sender/receiver unit, retractable version (guide rods not shown) Permissible gas pressure/temperature range: → pg. 42, Fig. 9 1 Connection for optional venting Modules/parts not named: 2 Retraction flange → pg. 31, Fig. 5 3 Centering FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 49
Permissible gas pressure/temperature range: → pg. 44, Fig. 13 1 Electronic unit in Exd housing4 Retraction flange Modules/parts not named: → pg. 31, Fig. 5 2 Connection for optional venting5 Centering 3 Duct probe RE FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 50
Permissible gas pressure/temperature range: → pg. 44, Fig. 13 Modules/parts not named: 1 Electronic unit in Exd housing4 Retraction flange → pg. 31, Fig. 5 2 Connection for optional venting5 Centering 3 Duct probe RE FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 51
Permissible gas pressure/temperature range: → pg. 42, Fig. 9). NOTICE: Maximum pressure for use of retraction mechanism: 0.5 barg Modules/parts not named: 1 Connection for optional venting → pg. 31, Fig. 5 2 Retraction flange FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 52
Maximum pressure for use of retraction mechanism: 0.5 barg Modules/parts not named: 1 Electronic unit in Exd housing → pg. 31, Fig. 5 2 Connection for optional venting 3 Retraction flange FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 53
Maximum pressure for use of retraction mechanism: 0.5 barg 1 Electronic unit in Exd housing Modules/parts not named: → pg. 31, Fig. 5 2 Connection for optional venting 3 Retraction flange FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 54: Mounting Accessories
● Do not exceed temperature and pressure ranges as listed in → p. 55, 2.2.2.1. Material for use outside this specification is available as option or on request. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 55
● Seals must be chosen in accordance with application conditions (gas tem- perature). ● An optional insulation set with polymer seals and sleeves for bolts is avail- able FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 56
Take into account additional 3 mm ”welding space” for nozzles of type «Long Welding Neck» Nozzle length 235 mm for pipe size 12” Other sealing material on request According to diagrams of pressure/temperature ranges (e.g. → pg. 41, Fig. 7) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 57
@ -10 … +150 °C (acc. EN 1092-1, table 4.4.1-4) Ball valve PN16 DN80 FLSE100-EXPR PN16 DN80 High and low temperature ball valves acc. to DIN on request. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 58: Nozzle Installation Tool
– installation paper strip as tool to determine the exact nozzle position on the pipeline. Fig. 24 Nozzle installation tool Design for FLOWSIC100 EX-S Nozzle Marking Pipe wall Design for FLOWSIC100 EX / EX-RE Design for FLOWSIC100 EX-PR FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 59: Mcup Control Unit
– Service interface USB 1.1 or RS232 – Service interface RS485 (standard for ex-protected MCUP) The parameters are stored reliably even in case of power failure. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 60
230 V a.c. 4″ NPT (3 measur- steel housing, painted steel automatically ing points), flange cable glands controlled plate connection 1): not applicable for USA and Canada FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 61
Alternatively, up to 8 single measured values of a sender/receiver unit can be displayed. Type Display Power (green) Power supply OK Failure (red) Function fault Maintenance Maintenance requirement request (yellow) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 62
Display alarm or fault message Menu Display main menu Fig. 26 LC-Display 120000.00 m³/h 65.04 m/s 334.21 m/s T ak. 18.3 °C 17.6 °C 18.0 °C Menu FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 63
● The pulse output is part of the interface modules to output the measuring values for actual volume, standard volume and mass (connection as «NAMUR» or «Open Collector» → pg. 145, Fig. 76; configuration → pg. 175, § 4.2.6). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 64
Without optional modules, 2 x DI on board Digital input option 1 digital input module (4 inputs per (not yet available) module) 2 digital input modules (not yet available) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 65
1 analog output module without digital input module without digital output module W without digital output module S Modbus RS485 with pulse output module reserve without additional features FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 66: Connection Cables And Cable Glands
5 m – cable glands metric M20, M25, NPT ½», NPT ¾», material brass and stainless steel Other connection cables for MCUP (power supply, outputs MCUP etc.) are not in the SICK scope of delivery.
-
Page 67
● Only analog cable Exi of device type FL100 EX-S is approved for ATEX and CSA versions (interconnection of master and slave unit). ● All other cables offered by SICK are only for ATEX/IECEx versions for use in ATEX zone 1 or zone 2 applications. -
Page 68: Meter Body Option (Spoolpiece)
Meter body option (spoolpiece) 2.2.6 The FLOWSIC100 Flare can also be fitted with an optional spoolpiece to reduce geometric uncertainty of device installation and to simplify assembly. The exact design (nominal diameter, connection, material) always depends on the customer specifications.
-
Page 69: Computations
Calculation and calibration of volume flow 2.3.1 Volume flow in operating state Generally, the devices of the FLOWSIC100 Flare family are used to determine the volume flow in closed pipelines. The volume flow Q is defined by the representative cross-…
-
Page 70: Mass Flow And Molecular Weight Determination
4 … 20 mA ● Pressure transmitter: Absolute pressure type, 1 … 16 bar, 2-wires Smart transmitter, explosion protected, out- put 4 … 20 mA FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 71: Response Time
FLOWSIC100 Flare requires more measured values for an output signal of the same precision. This increases the response time within certain limits as compared with the set time. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 72: Check Cycle
Check cycle 2. 4 The FLOWSIC100 Flare is equipped with an integrated check cycle for automatic functional check of all device components. This check cycle is an approved tool for periodic device inspection according to valid regulations of Continuous Emission Monitoring.
-
Page 73: Span Test
● The output duration of 90 s each time is the standard setting ex factory. It can be changed in the SOPAS ET program. ● The output is suitable for velocity-dependent measured values only (gas velocity, volume flow act., volume flow std.). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 74
Product Description FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 75: Assembly And Installation
Assembly and Installation FLOWSIC100 Flare Assembly and Installation Project planning Assembly Installing non-retractable sender/receiver units Installing the sun roof for the sender/receiver units FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 76: Project Planning
Fitting locations Cable lengths § 2.2. According to Technical Data in → pg. 210, Plan power Operating voltage, supply power requirements § 6.1 Plan adequate cable cross-sections and fuses FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 77: Determining The Measuring And Installation Location
● Uncritical flow inlet conditions requires straight upstream piping > 10 x Di and downstream piping > 5x Di. Manufacturer SICK offers expertise support for an optimal adjustment of the meter for the given inlet and outlet piping conditions.
-
Page 78
Fastening and sealing Bolts, nuts and flange seals used must be suited to the operational conditions and comply with legal material regulations and relevant standards. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 79
Non- Retractable retractable Type inch inch 29.5 FLSE100-EXS 1040 40.9 FLSE100-EX 31.5 FLSE100-EXRE 1250 49.2 1070 42.1 FLSE100-EXPR short 1900 74.8 1200 47.2 FLSE100-EXPR long 2050 80.7 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 80: Further Notes For Project Planning
Isolate the nozzle pipe to reduce dew point underruns (only for low gas temperatures < 100 °C). b) Short inlet and outlet paths (→ pg. 77, Fig. 30) Use the best possible positioning for the measurement path (consult SICK for sup- port). c) Vertical ducts with flow direction from top to bottom The measured value gets a negative sign (output LC display).
-
Page 81: Notes For Installation Of Sender/Receiver Units In 1-Path, 2-Paths Or 3X1 Path
FLSE100-EXPR Mounting direction 75° prevents condensate Assembly nozzle Flow accumulations in direction nozzle tube 1-path measurement Pipeline Flow Assembly nozzle direction 2-path measurement 0.6 •r 0.6 •r FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 82: Preparation Work
Min. gas temperature Max. gas temperature Max. operating pressure at 50 °C Max. operating pressure at T gas max. (variable 06) Flange size Not relevant Not relevant Bar code FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 83: Special Notes For Handling The Meter Body Option
Max. load information can be found on the type plate of the lifting gear. Only use the eye bolts when lifting the meter body. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 84: Assembly
600 Nm (retractable device types). Strong pipe vibrations can cause damages and can lead to dangerous situations. Use a mechanical support for the nozzles welded to the pipeline, e.g. «gusset plates». FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 85
Assembly and Installation Fig. 34 Option mechanical support for nozzle «gusset plates» Gusset NOTICE: The plant operator is responsible for the safety of the system under mechanical burden. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 86: Fitting The Nozzles On The Pipeline (Measuring Systems Without Meter Body)
3) Loosen the fastening, take the strip off and lay it out on a level surface. For 1 path measurements, fold the strip to the overlap line so that the part matching pipe Overlap line circumference (U) is halved. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 87
6a )Measure and note values a and U/2 (needed to calculate the path angle and path length). 6b) Measure and note values a and U (needed to calculate the path angle and path length). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 88
Path 1 Path 2 Take the strip off again and join the additional markings with a line. Marking lines Marking lines FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 89
3) Draw a guide line (1) for the nozzle position(s), mark the crossing points (2) and draw marking points (3) in distance 80 mm (x) from the crossing points. Path 1 Path 2 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 90
Path 1 Path 2 7) Take the strip off again and join the additional markings with a line. Marking lines Marking lines FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 91
Work steps Position the welding aid (1) on the pipeline (2) as shown in the following figure. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 92
Align the nozzle so that the marking lines on nozzle and pipe wall are flush and tighten enough so that the nozzle is pressed against uncoated wire and pipeline surface. Make sure the nozzle remains aligned correctly. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 93
Determining of effective nozzle length (Nozzle) Centering On cross-duct versions, weld the nozzle on the opposite pipeline side in the same man- ner and then determine distance D2. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 94
Comply with all legal, general and company-internal regulations. Only start installation work when all planned measures have been checked and approved by the plant operator. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 95
▸ Keep the ball valve closed and fitted until a sender/receiver unit has been fit- ted. ▸ Secure the ball valve against unintentional activation. ▸ Instruct other persons accordingly. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 96
= Probe nominal length (manufacturer’s specification) D1, D2 = effective nozzle length (→ pg. 93, Fig. 42) = Nozzle offset (→ pg. 86, § 3.3.1.1) = 1 for 1-path measurements, 0.8 for 2-path measurements FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 97
– NL D1 – – 2 S VL – – NL D2 – – 2 S VL – π α — — atan FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 98
Use a suitable protractor to determine and register the actual value. The manufacturer measures the path length for each probe and includes this in the Zero Point protocol in the delivery documents. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 99: Installing The Meter Body (Option) In The Pipeline
Check for correct seating and alignment of the flange gasket after installing the flange bolts, but prior to tightening. Align the meter body of the FLOWSIC100 Flare such that the offsets between inlet pipe, meter body and outlet pipe are minimized.
-
Page 100: Installing The Mcup Control Unit
We therefore recommend fitting the MCUP in a control room (measuring station or similar) to ensure free access to the MCUP. This considerably simplifies communication with FLOWSIC100 Flare in order to set parameters or to locate malfunction or error causes.
-
Page 101
300 (for ATEX zone 2 version) 320 (for CSA CII Div 2 version) > 350 > 550 MCUP in medium size housing Clearance for cables Clearance to open the door FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 102
Size 4 Size 6 11,5 Special housing variant — stainless steel MCUP in Ex housing for zone 1 — stainless steel [all dimensions in mm] Figure 47 Ø18 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 103
Explosion-protected version for CSA Cl I, Div 1 3.3.3.3 MCUP in Ex housing for zone CSA Cl I, Div 1[all dimensions in mm] Fig. 48 Ø B Housing variante Size 4 Size 6 11,5 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 104: Installing The Mcup With Option «Set Fixing Plate For 2-Inch Post
Fix plate (4) to post using round steel bow (1), nut (3) and washer (2) Mount MCUP to fixing plate using screw (8), washer (7), washer plus spring washer (7) + (6) and nut (5). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 105: Installing The Sun Roof Mcup (Option)
Mounting sun roof MCUP on the wall 3.5.1.2 Use „mounting kit sun roof” on the wall. Use the holes marked in Fig. 51. Mounting sun roof to the wall Figure 51 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 106
The appropriate mounting kit for the mounting of the MCUP on the sun roof can be found in Table 3. Mounting MCUP on the sun roof Figure 52 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 107
Size A,B,C,D MCUP Medium Mounting kit MCUP Red (2) Size A,B,C,D MCUP STAHL, Size 4 (with or without terminal compartment) Mounting kit MCUP Green (3) Size A,B,C,D FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 108: Sun Roof Mcup Mounting On 2-Inch Post
Mounting kit post bracket 2-inch 2061076 Mounting kit sun roof 2072663 Mounting kit MCUP Size A 2072665 Mounting kit MCUP Size B,C,D 2072666 Mounting kit MCUP Size E 2072664 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 109
Then fix the mounting plate (1) to the 2-inch post with the „mounting kit post bracket 2-inch”. Finally fix the sun roof (2) on the mounting plate (1) with the “mounting kit sun roof” according to Fig. 55. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 110
Mounting on 2-inch post Figure 55 Mounting MCUP on sun roof 3.5.2.3 Mount the MCUP to the sun roof as outlined on p. 105, § 3.5. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 111: General Notes For Installation Of Sender/Receiver Units
For installation of seals see instructions of sealing manufacturer in → pg. 239, § 6.5. WARNING: Hazards through leaks Operation in leaky condition is not allowed and possibly dangerous. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 112: Installing Non-Retractable Sender/Receiver Units
Venting of any gas with hazard potential (toxic, aggressive, explosive etc.) requires always suitable protective measures and safety precautions for safe venting and is in the responsibility of the plant operator (→ p. 31, 2.2.1) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 113
Installing the nozzle isolation set To prevent danger of galvanic corrosion a «Nozzle isolation set» can be used (option). For installation see instruction in → pg. 239, § 6.5. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 114
Assembly and Installation Fig. 56 Retractable sender/receiver units assembly Pipe wall Nozzle 1 shown without connection for venting, for device type FLSE100-EXS FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 115: Installing The Sun Roof For The Sender/Receiver Units
Bild 57 Overview sun roof 1 Holder 2 Split pins 3 Protective hood 4 Washers 5 Hexagon nuts 6 Round steel bow FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 116: Installing The Sun Roof For The Sender/Receiver Units
Pay attention to the correct alignment of the holder. See adjacent figure. 2 Position the protective hood on the holder. 3 Secure the sun roof with the 4 split pins. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 117: Dimensional Drawings
Assembly and Installation Dimensional drawings 3.7.2 Bild 58 Holder FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 118
Assembly and Installation Bild 59 Protective hood FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 119: Electrical Installation
– Wire cross-sectional area at least 0.5 mm² (AWG20 to AWG16 max.) – Screen with Cu wire mesh The cable specification above represents the SICK standard cables and does not take into account any special requirements for cabling in hazardous areas. The plant operator is responsible, that all used cables comply with valid regulations and guidelines for cabling in hazardous areas at his plant.
-
Page 120
Assembly and Installation Cable glands ● Cable glands for non-ex, ATEX Zone 2 and ATEX Zone 1 are offered by SICK as acces- sories for the following components: – FLSE100 ATEX Zone 1 without junction box – FLSE100 CSA Cl I, Div1 –… -
Page 121
(no termination in delivery state). Solutions for external terminations are available on request. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 122: Requirements For Installation In The Ex Zone
● Replace unused cable glands with the enclosed Exe or Ex-d sealing plugs. ● Connect or safeguard unused wires to ground so that a short circuit is excluded with other conductive parts. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 123
”Exi” or a blue cable cov- ering or with blue shrink sleeves on the cable ends or with the SICK item number, at least on the associated packaging. The technical safety data are shown in the Type Examination Cer- tificate. -
Page 124
Do not disconnect unless power has been removed or the area is known to be non-hazardous. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 125
(Fifth Edition, dated August 15, 2005) ● ANSI/UL 60079-11 (Second Edition, dated March 9, 2007) ● ANSI/UL 60079-15 (First Edition, dated December 2, 2002) ● ANSI/ISA 12.27.01-2003 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 126
The information is given purely in order to check whether a complete board to be replaced matches the requirements (details for exchange see Service Manual). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 127: Connection Of Sender/Receiver Units
Communication between FLSE Master and MCUP (interface 1) IF2: Communication between FLSE Master and FLSE Slave (interface 2) NOTICE: Self-locking terminals for wire sizes 0.5 .. 2.5 mm² (AWG20 … AWG12). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 128
Communication between FLSE Master and MCUP (interface 1) IF2: Communication between FLSE Master and FLSE Slave (interface 2) n.c.: Not connected NOTICE: Self-locking terminals for wire sizes 0.5 .. 1.5 mm² (AWG20 … AWG16). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 129: Mcup Connections
: Wide range, not for Div 2 / zone 2 version : Automatic switching : It may only be used a power supply with safe mains isolation (PELV).The negative terminal is grounded to the housing of device. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 130
Digital input 4 (active low forced to gnd) 5.5 V d.c. max. 1 mA 22, 23 +, — Analog output (20 mA) 20 … 28 V d.c. max. 22 mA Ground FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 131
Input 24 V d.c. 20 … 28 V d.c. P1 service USB Port : normal closed : normal open : use only according to consultation with the manufacturer FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 132
Grounding of sender/receiver units has to be established according to p. 127, § 3.8.4. The equipotential has to be connected in accordance with IEC 60079-14. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 133
Power supply unit I/O module option Display module Terminal connection for power supply 90…250 V AC Slot for interface module option Terminal connection for wiring by customer FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 134
71, 73 A, gnd 72, 73 B, gnd 74, 76 +Us, gnd 75, 76 -Us, gnd 77, 79 imp+, res 1 78, 80 imp-, res 2 Interface 1 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 135
+Us, gnd 85, 86 -Us, gnd Not planned for use P1 service USB Port : normal closed : normal open Use only according to consultation with the manufacturer FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 136: Cabling
Seal cable entries not in use with blind plugs WARNING: Be sure to check the wiring before switching the supply voltage on. Only modify wiring when disconnected from the mains and potential-free. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 137
(available from SICK on request) to the white MCUP. green +24 V DC yellow RS485a ”Point-to-point” wiring for 2-path and 2×1-path RS485b measurement on request. Terminal compartment * MCUP Ex zone 2 only FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 138
(available from SICK on request) to the MCUP. yellow RS485 b ”Point-to-point” wiring for 2-path and 2×1-path measurement on request. brown * MCUP Ex zone 2 only white +24 V DC FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 139
”Point-to-point” wiring for 2-path and 2×1-path measurement on request. +24 V DC FLSE100-EXS analog * RS485a RS485b *: Not for FLSE100-EXPR **: MCUP Ex zone 2 only FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 140
”Point-to-point” wiring for 2-path and 2×1-path measurement on request. Connection cable analog to Exi * *: Not for FLSE100-EXPR **: MCUP Ex zone 2 only FLSE100-EXS analog * FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 141
For Ex zone 1 only with approved junction box Exi or junction box Exd FLSE100-EXS analog *: Not for FLSE100-EXPR 2): MCUP Ex zone 2 only **: MCUP Ex zone 2 only FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 142: Fitting And Connecting The Interface And I/O Module Options
Connect I/O modules using the terminals on the module carrier (→ Fig. 72, Fig. 73, Fig. 74), the Profibus module using the terminals on the module and connect the Ethernet module via a customer provided network cable. ● Terminal assignment AO module FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 143
Digital output module terminal assignment Digital output module Module carrier n.o. com. com. n.o. n.c.^ com. com. n.c. NOTICE: Screw-fixed terminals for wire sizes 0.5 .. 1.5 mm² (AWG20 … AWG16). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 144
Power HART & ImpulsOut part no. : 2050607 part no. : 2050607 gal. 4..20 mA > 250 Ohm → LC input FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 145
= 100 mA 10 k Vc 2V – Vc 2V – ——————— — ≤ ≤ ——————— — 0 1A 0 01A Interface modules mounted on top-hat rail Figure 77 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 146
Assembly and Installation 19” rack mounted interface module Figure 78 Jumper settings for impulse output Figure 79 Jumper set to NAMUR Jumper set to OC NAMUR NAMUR FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 147
Following this, the connection of the modules is represented exemplarily to slot 1. The connection of optional modules (analog and digital type) at the other slots 2-8 is carried out in the same way. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 148
106 b 106 b 107 c 107 c 107 c 108 d 108 d 108 d 109 -gnd 109 -gnd 109 -gnd 110 scr 110 scr 110 scr FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 149: Connection Of External Pressure And Temperature Transmitters
● For MCUP powering the relay contact circuits are connected to earth, which means galvanic isolation is not available. ● Use an external Class2/SELV specified power supply to fulfill isolation from earthing. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 150
Assembly and Installation FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 151: Start-Up And Parameter Settings
Start-up and Parameter Settings FLOWSIC100 Flare Start-up and Parameter Settings Basics Standard start-up Advanced start-up Operating/setting parameters via the optional LC-Display FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 152: Basics
This work should only be carried out by SICK Service. Possible settings are described in the Service Manual. We recommend to use the SICK Commissioning check list (→ pg. 239, § 6.5) for easy parameter setting.
-
Page 153: Installing The Operating And Parameter Program Sopas Et
As an alternative, the driver can also be installed in the Windows control panel using the hardware installation program. NOTICE: Use only USB cabling which is in scope of SICK supply or cable with identical specification. Maximum cable length is 3 m. Longer cable lengths can lead to communication problems.
-
Page 154
Start-up and Parameter Settings Fig. 83 Installing the USB driver FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 155: Connecting To The Device
If required, select the desired language in the “Tools / Language” menu (→ pg. 155, Fig. 84). Confirm the dialog shown with “Yes” to restart SOPAS ET with the changed language. Changing the language Fig. 84 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 156
An IP address specified by the customer is entered at the factory if the address is available when the device is ordered. If not, standard address 192.168.0.10 is entered. To change the IP address, see → pg. 195, § 4.3.5.2. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 157
7 To save the search settings, enter a name and click “Finish”. SOPAS ET starts the device search. The devices found are displayed in the “Device search” area when device search is finished (→ pg. 161, Fig. 95). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 158
3 Select the communication interfaces where the search is to be made and click “Next”. Selecting the communication components Fig. 90 4 Configure the interfaces and click “Next”. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 159
Define the protocol settings in the “Protocol” directory according to → pg. 159, Fig. 91. Defining the protocol settings Fig. 91 Define the timeout settings in the “Timing” directory according to → Fig. 92. Defining the timeout settings Fig. 92 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 160
5 To save the scan settings, enter a name and click “Finish” (→ pg. 158, Fig. 89). SOPAS ET starts the device search. The devices found are displayed in the “Device search” area when device search is finished (→ pg. 161, Fig. 95). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 161: Information On Using The Program
4.1.4 Overview Fig. 95 1 Device search 4 Device search abort 2 Project area 5 Device search result 3 Device search progress 6 Number of devices found FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 162
– The device windows can be opened by a double-click on the respective device file or the context menu (→ pg. 163, Fig. 97). Device selection Fig. 96 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 163
Export Exports the device information and the associated project information and saves them in a *.sopas file. Delete device Deletes the device from the project. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 164: Standard Start-Up
In applications with extremely fluctuating gas compositions, it is recommended to optimize the device parameters in cooperation with the manufacturer. Error message «Parameter Error» is output when equipment data input is not complete. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 165: Entering Application-Specific Parameters In The Sender/Receiver Units
Select the unit of measurement to be used in the «Tools / Options» menu on the over- view page (metric or US standard), → pg. 165, § 100. Selecting the unit of measurements Fig. 100 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 166
● Enter the calibration coefficient for gas velocity Cv_1 («Calibration Coef- ficients» group) as a negative value if FLOWSIC100-EXPR is installed at a vertical pipeline (→ pg. 81, § 3.1.3). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 167: Entering Application Specific Parameters For Single-Path Configuration
For flow calibrated devices the coefficients CC0 … CC4 must be the same as those detailed on the calibration certificate. If necessary, set the parameters in SOPAS ET with values from the calibration certificate. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 168: Configuration Of Application-Specific Parameters For 10″ Pipelines And 2-Path
«Authorized client». The manufacturer is not liable for any disruption caused by incorrect settings made by the user in SOPAS. Fig. 103 Service Login Set SOPAS to maintenance mode. Fig. 104 Maintenance mode page FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 169
ET will then synchronize the MCUP. To complete the configuration: Open the «Linearization parameter» page. Change parameter cc0 to -223.085 and parameter cc1 to 168.508. Press Enter. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 170
SICK support. For flow calibrated 2-path configurations The parameters are individually set by SICK during production of the meter after flow calibration (incl. individual spool piece and transducers). Be careful to not change these parameters, since the individual flow calibration may be affected. -
Page 171
Start-up and Parameter Settings Figure 107 2-path configuration settings (for uncalibrated and dry calibrated 2-path configuration) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 172: Determining The Check Cycle
Hours Start at time of day Minutes Defining a start time-point in hours, minutes and seconds The value measured last is output during check value determination. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 173: Setting The Analog Outputs Parameters
Fig. 109 «Parameter / I/O Configuration / Analog Output Parameters» directory The «Analog output 2(3)» fields only appear when an AO module is plugged in. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 174
Output of results of check cycle: On values On/Off No output of results of check cycle: Off Set the parameters for fields «Analog output 2(3)» analog to field «Analog output 1». FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 175: Setting Parameters For The Pulse Output
Measuring point 1, 2 or 3 output at the pulse output Relation between impulse frequency, volume and impulse factor: m³ Impulse factor [kHz] = •Impulse rating m³ 3600 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 176: Setting The Analog Inputs Parameters
External Modbus Reading in of external pressure values via Modbus registers; for detailed information value see Modbus Specification. NOTICE: All pressure values in SOPAS are «absolute pressure». FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 177
If one of the transmitters fails, the value of the remaining transmitter is used. If both transmitters fail, the value measured last is used as constant value until the function of the transmitters is restored. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 178
If one of the transmitters fails, the value of the remaining transmitter is used. If both transmitters fail, the value measured last is used as constant value until the function of the transmitters is restored. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 179: Defining The Limit Value
When value > 0, the limit value relay switches when the entered Limiting value Value value is exceeded or underflown. Over limit Switching direction Under limit Defines the switching direction FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 180: Logbook Function
Data can be saved in the following ways: ● Saving as a project Data recordings can also be stored separately from device parameters. ● Saving as a device file FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 181
The name of the file to be stored can be chosen freely. It is useful to specify a name with a reference to the measuring point involved (name of the company, equipment designation). Fig. 116 «Project / Save» menu FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 182
«Diagnosis / Protocols» menu The file name and storage location must be specified for export to a PDF file. Fig. 118 Specifying file names and storage location FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 183
Start-up and Parameter Settings Parameter protocol example Fig. 119 FLOWSIC100 Flare Parameter protocol (example) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 184: Starting Normal Measuring Operation
Setting the «View Envelope» function shows the envelope curves of both transducers. The signal patterns should correspond to the displays in Fig. 120 to Fig. 121. Example Fig. 120 Burst form HF signal Type FLSE100-EXS (raw signal) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 185
Start-up and Parameter Settings Fig. 121 Burst form demodulated signal Type FLSE100-EXS (envelope curve) Cursor Cursors of both transducers must be at same position. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 186: Advanced Start-Up
Customer Data» directory. To enter, select device file «MCU-P», enter the Level 1 password and set the measuring system to Maintenance mode (→ pg. 161, § 4.1.4). Fig. 122 «Parameter / Customer Data» directory FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 187: Entering Counter Function Parameters
● Switch to the «Parameter / Data archives» directory. ● Enter the parameters for recording interval, entry depth, write mode and reference to measuring point for all 9 archives. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 188
Measuring point 1, Select measure point to make the settings for the Measuring point Measuring point 2, data archive Measuring point 3 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 189
The retrieved data can be exported as ASCII CSV file. These data can then be used in other applications (e.g. MS Excel). Click «Export data» to export the data and specify the file name and storage location. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 190
Select «Delimited» as the original data type Select file origin «65001 : Unicode (UTF-8)», click «Next» Fig. 126 CSV Import settings step 1 (example: Russian header text) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 191
Choose the option «Semicolon» as the delimiter, click «Next» Fig. 127 CSV Import settings step 2 Ensure settings are as shown in Fig. 128 and click «Finish» Fig. 128 CSV Import settings step 3 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 192
● Molar mass [g/mol] ● Gas density a.c. [kg/m³] ● AGC (Amplifier gain control), average during last recording period ● Performance, average during last recording period ● Meter status FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 193: Entering Process Parameters
– When using the MR-113-03 algorithm: Gas composition values – When using the Hydrocarbon algorithm: Viscosity, compressibility Fig. 129 «Parameter / Measuring Point 1 / Process Values» directory FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 194: Setting Interface Module Parameters
«Parameter / System Configuration» directory NOTICE: Factory setting for counted flow direction is «positive flow». GSD file and measured value assignment are available for the Profibus DP module on request. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 195
Ethernet. Increasing the value in the «Scantimeout» field to 3000 ms can minimize communication problems. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 196: Setting Modbus And Hart® Bus Module Parameters
Open «Parameter / System Configuration» directory and select type «HART» in the «Select the installed module» menu (group «Interface Module — Type Selection»). Fig. 133 «Parameter / System Configuration» directory FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 197: Application-Specific Flow Calibration
«Adjust factor» window in the «Parameter / Measure Point 1 / Measure Point Parameter» directory, field «Measuring Point 1 — Application Parameter (I)” (→ pg. 164, § 4.2). FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 198: Operating/Setting Parameters Via The Optional Lc-Display
Start an action Save Store a changed parameter Toggle between test and graphic display Return from submenu to main menu Meas Display the contrast setting (after 2.5 s) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 199: Menu Structure
Start-up and Parameter Settings Menu structure 4.4.2 Fig. 135 LC-Display menu structure *: Number depends on available AO‘s/AI‘s FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 200: Parameter Setting
Limit Low Path Length 6.2 m Password Installation Angle Min: 1234 Cross Section Area Max: /sensor/1/para/ /sensor/1/para/ /sensor/1/para/ Back Enter Back ∧ → Save Back ∧ → Save FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 201: Using Sopas Et To Modify Display Settings
Assignment of measuring point to first and second measured value bar in the graphic Line 1 to 8 display Value Assignment of measuring variable to the respective measured value bar FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 202
Start-up and Parameter Settings FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 203: Maintenance
Maintenance FLOWSIC100 Flare Maintenance General Recognizing malfunctions Maintaining the sender/receiver units Maintaining the MCUP for use zone 2 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 204
Maintenance FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 205: General
Scheduled maintenance work can be carried out by the plant operator. Only qualified personnel according to Section 1 should be allowed to do the work. If desired, SICK Service or authorized Service support centers can carry out all maintenance work. Any repairs will be made by specialists on-site whenever possible.
-
Page 206: Maintaining The Sender/Receiver Units
Close the ball valve after removing the sender/receiver unit with a blind flange. Take special precautions in the presence of toxic, aggressive, explosive, hot, or other gases posing a health risk (risk of serious injuries)! FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 207: Cleaning The Sender/Receiver Unit
6 months. Reinstall the sender/receiver unit after completion of the work. The work required to replace parts (duct probe, transducers) is listed in the FLOWSIC100 Flare Service Manual. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 208: Maintaining The Mcup For Use Zone 2
Replacement by complete processor boards or complete modules is recommended. To prevent repetitive degradation the user shall clarify the causes and shall take appropriate precautionary measures before restart the equipment. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 209: Specifications
Options for MCUP control unit Password Sealing installation instruction Connection diagrams for MCUP Ex zone 1 — configuration examples with optional modules Wiring diagrams for USA and Canada FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 210: Technical Data
MCUP control unit, zone 1 ATEX II 2 G Ex de IIC T6 MCUP control unit, zone 2 ATEX II 3 G Ex nA IIC T4 Gc FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 211
: Preferred operating range for nominal pipe diameter ≥ 24″, smaller diameters at difficult gas compositions with sound- absorbing components : Operating temperature range of ball valve: standard: -50… +200 °C high temperature: -50 … +350 °C low temperature: — 196 … +100 °C FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 212
: Additional inputs/outputs when using optional modules. : Replacement fuses T 2 A are part of delivery. Ex-protected control units contain a separate note on label of the fuse type. FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 213: Dimensions And Part Numbers
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 214
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 215
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 216
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 217
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 218
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 219: Flowsic100 Ex/Ex-Re Sender/Receiver Units
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 220
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 221
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 2″ CL150 RF acc. to ASME B16.5. DN50 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 222: Flowsic100 Ex-Pr Sender/Receiver Units
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 3″ CL150 RF acc. to ASME B16.5. DN80 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 223
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 3″ CL150 RF acc. to ASME B16.5. DN80 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 224
1 For part nos. see → p. 225, 6.2.4 2 Process connections to fit flange 3″ CL150 RF acc. to ASME B16.5. DN80 PN16 form B1 acc. to EN 1092-1 available as option FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 225: Sender/Receiver Units — Parts List
FLSE100-EXS 330TA218VS RS 82CSAMY NN 1047659 FLSE100-EXS 330TA218VS RS N2CNNNN NN 1047660 Sensor system CSA CI I, Div1/Div2, titanium FLSE100-EXS 330TA218VS RS 83CDATN NN 1047673 FLSE100-EXS 330TA218VS RS N3CNNNN NN 1047674 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 226
Sensor system ATEX Zone 2 with junction box, titanium FLSE100-EXRE 380TA214VS RS 42CSAMY NN 1047696 Sensor system CSA CI I, Div1/Div2, titanium FLSE100-EXRE 380TA214VS RS 43CDATN NN 1047700 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 227
Sensor system ATEX Zone 2 with junction box, titanium FLSE100-EXPR 400TA311VS RS 12CSAMY NN 1047684 Sensor system CSA CI I, Div1/Div2, titanium FLSE100-EXPR 400TA311VS RS 13CDATN NN 1047688 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 228: Connection Cables
Ex-d cable gland metric, M20 brass 5324884 Ex-d cable gland metric M20, stainless steel 5324992 Ex-d cable gland NPT 1/2″, brass 5324915 Ex-d cable gland NPT 1/2″, stainless steel 5325257 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 229
Exe cable gland NPT 1/2″, brass 5325271 Exe cable gland NPT 1/2″, stainless steel 5325272 Exe cable gland NPT 3/4″, brass 5325273 Exe cable gland NPT 3/4″, stainless steel 5325274 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 230: Junction Box For Ex Zone 2 (For Atex Versions Only)
Mounting holes Ground terminal Ø 4.8 Designation Part No. Junction box for connection cable for zone 2 2046562 Fastening set 2D4-1.4571/PA 2031890 (for junction box for connection cable) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 231: Nozzles
Nozzle CL1502Z, prepared for pipe size 60″ 2052007 Nozzle CL1502Z, prepared for pipe size 64″ 2052008 Nozzle CL1502Z, prepared for pipe size 68″ 2052009 Nozzle CL1502Z, prepared for pipe size 72″ 2052010 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 232
Nozzle CL1503Z, prepared for pipe size 72″ 2057133 Nozzle installation tool Designation Part No. Nozzle installation tool EX-S 2050598 Nozzle installation tool EX 2050599 Nozzle installation tool EX-PR 2050601 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 233: Additional Accessories For Retractable Sender/Receiver Units
MCUP without explosion protection Fig. 163 MCUP control unit without explosion protection in compact size housing Designation Part No. Control unit MCUP-WNA1MN00000NRN 1050334 Control unit MCUP-2NA1MN00000NRN 1047618 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 234
Specifications Fig. 164 MCUP control unit in 19″ rack (shown with Display module option) 465.9 482.6 Designation Part No. Control unit MCUP-WNF1MN00000NRN 1047704 Control unit MCUP-2NF1MN00000NRN 1047705 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 235
Designation Part No. Control unit MCUP-WNB3MN00000NRN 1050336 Control unit MCUP-2NB3MN00000NRN 1050335 Control unit MCUP-SNB3MB00000NSN 1050337 Control unit MCUP-2NB3MB00000NSN 1047706 Control unit MCUP-SNB3CE00000NSN 1050340 Control unit MCUP-2NB3CE00000NSN 1050339 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 236
MCUP for ATEX Zone 1 Fig. 166 MCUP in Ex housing for zone 1 STAHL Ø 11 Designation Part No. Control unit MCUP-WND1MA00000NSN 1048164 Control unit MCUP-2ND1MA00000NSN 1047707 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 237
MCUP in Ex housing for CSA Cl I, Div 1 Fig. 167 Ø B Designation Part No. Control Unit MCUP-2NK1CD00000NSN 1069773 Control Unit MCUP-WNK1CD00000NSN 1069771 Control Unit MCUP-WNL1CD00000NSN on request Control Unit MCUP-2NL1CD00000NSN on request FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 238: Options For Mcup Control Unit
Interface Modbus + Impulse (on request) 2050674 Interface HART® BUS + Impulse (on request) 2050608 Password 6. 4 User level Password 0 «Operator» Without 1 «Authorized client» sickoptic FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 239: Sealing Installation Instruction
Specifications Sealing installation instruction 6 . 5 Fig. 168 Sealing installation (designed by ”pikotek”) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 240
Specifications 9A and polymer gasket Fig. 169 Sealing installation (designed by ”pikotek”), bolt torque values for grooved metal gasket B GYLON FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 241: Sick Commissioning Check List
Type key of MCU: Measurement Location: Add. description: Service information Date of this service: Time of this service: Technician name: Technician phone: Technician email: Date of last service: FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 242
Specifications Fig. 171 SICK Commissioning check list, page 2 1 Configuration sensor FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 243
Configure reference data temperature and pressure. Configure the measuring points. Configure reference data temperature and pressure Open page «Analog Input Parameter» and configure the values for pressure and temperature E_49990 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 244
Specifications Fig. 173 SICK Commissioning check list, page 4 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 245
Specifications Fig. 174 SICK Commissioning check list, page 5 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 246
Specifications Fig. 175 SICK Commissioning check list, page 6 Check Diagnostic Check Output / Input FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 247
Specifications Fig. 176 SICK Commissioning check list, page 7 FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 248
Specifications Fig. 177 SICK Commissioning check list, page 8 Miscellaneous MCU FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 249
Specifications Fig. 178 SICK Commissioning check list, page 9 _______________________________________________________________________________ _______________________________________________________________________________ _______________________________________________________________________________ _______________________________________________________________________________ _______________________________________________________________________________ _______________________________________________________________________________ _______________________________________________________________________________ _______________________________________________________________________________ _______________________________________________________________________________ _______________________________________________________________________________ __________________________ ___________________________ _______________ FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 250: Connection Diagrams For Mcup Ex Zone 1 — Configuration Examples With Optional
Connection diagrams for MCUP Ex zone 1 — configuration examples 6. 7 with optional modules Fig. 179 2-path configuration, 1 measuring point, supply voltage 230 V a.c (Drawing 9178258) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 251
Specifications Fig. 180 1-path configuration, 1 measuring point, supply voltage 115 / 230 V a.c (Drawing 9176519) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 252
Specifications Fig. 181 1-path configuration, 3 measuring points, supply voltage 230 V a.c (Drawing 9253809) FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 253: Wiring Diagrams For Usa And Canada
Manufacturer drawing FLSE100-EXS/-EXPR Maßstab: Page E_41943 1 of 6 Ursprung: Verteildatum: SICK Engineering GmbH Ersatz für: Bergener Ring 27 Name Gepr. Ind. Änderung 01458 Ottendorf-Okrilla Ersetzt durch: FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 254
Specifications Fig. 183 Wiring diagram for FLSE100-EXS/FLSE100-EXPR +24V +24V FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 255
Specifications Fig. 184 Wiring diagram for FLSE100-EXS/FLSE100-EXPR +24V +24V FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 256
Specifications Fig. 185 Wiring diagram for FLSE100-EXS/FLSE100-EXPR +24V +24V FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 257
Specifications Fig. 186 Wiring diagram for FLSE100-EXS/FLSE100-EXPR +24V +24V FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 258
Specifications Fig. 187 Materials FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 259: Flse100-Ex/Flse100-Exre
Manufacturer drawing FLSE100-EX/-EXRE Maßstab: Page E_41944 1 of 5 Ursprung: Verteildatum: SICK Engineering GmbH Ersatz für: Bergener Ring 27 Name Gepr. Ind. Änderung 01458 Ottendorf-Okrilla Ersetzt durch: FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 260
Specifications Fig. 189 Wiring diagram for FLSE100-EX/FLSE100-EXRE +24V FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 261
Specifications Fig. 190 Wiring diagram for FLSE100-EX/FLSE100-EXRE +24V FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 262
Specifications Fig. 191 Wiring diagram for FLSE100-EX/FLSE100-EXRE +24V FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 263
Specifications Fig. 192 Materials FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 264: Connection Of External Pressure And Temperature Transmitters
Specifications Connection of external pressure and temperature transmitters 6.8.3 Fig. 193 FLOWSIC100 Flare — loop powered pressure and temperature transmitters MCUP Terminals Terminals FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH…
-
Page 265
Specifications Fig. 194 FLOWSIC100 Flare — external powered pressure and temperature transmitters MCUP Terminals FLOWSIC100 Flare · Operating Instructions · 8013344/11L2/V 2-5/2018-10 · © SICK Engineering GmbH… -
Page 266
E-Mail sick@sick.no United Kingdom Denmark Poland E-Mail sick@sick.dk Finland Romania Vietnam France Russia E-Mail sales.gsg@sick.com Germany Singapore E-Mail sales.gsg@sick.com Hong Kong Slovakia E-Mail ghk@sick.com.hk E-Mail mail@sick-sk.sk Hungary Slovenia India South Africa www.sick.com SICK AG | Waldkirch | Germany | www.sick.com…
Detail Specifications: 842/842688-flowsic100.pdf file (14 Mar 2023) |
Accompanying Data:
SICK FLOWSIC100 Flare Measuring Instruments PDF Operating Instructions Manual (Updated: Tuesday 14th of March 2023 09:12:25 PM)
Rating: 4.9 (rated by 66 users)
Compatible devices: InspectorP64 Series, SIM1012, ALIS, VMS4200, MLG-2 Prime, PowerCEMS100, FLOWSIC100 Flare-XT, RFH620.
Recommended Documentation:
Text Version of Operating Instructions Manual
(Ocr-Read Summary of Contents, UPD: 14 March 2023)
Recommended Instructions:
CV50* ANALOGUE Series, 826W — PowerLite WXGA LCD Projector, 142725, Phaser 6125
-
The information contained in this document supersedes all similar information that may be found elsewhere in this manual. Service – Due to the sophisticated nature of the sensors and associated instrumentation provided by PCB Piezotronics, user servicing or repair is not recommended and, if attempted, may void the factory warranty. …
357A94 13
-
ES-FLOW™ 1xxCUltrasonic Liquid Flow Meter/ControllerDoc. no.: 9.17.153 rev. A Date: 24-06-2021Instruction ManualATTENTIONPlease read this document carefully before installing and operating the product.Not following the guidelines could result in personal injury and/or damage to the equipment. …
ES-FLOW 1 C Series 41
-
Operating instructionsBetriebsanleitungMode d‘emploiManual de instruccionesENDEFRESAnalysegerät zur Ermittlung der Gasqualität von Isoliergasen, Typ GA11Analytic instrument for N2 gasAnalytic instrument for determining the quality of insulated gases, model GA11Analizador para determinar la calidad del gas aislante, modello GA11Instrument d‘analyse pour …
GA11 148
-
WOLFF GmbH & Co. KG | D-74360 Ilsfeld | Ungerhalde 1 | Tel. +49 (0)7062-91556-0 | Fax +49 (0)7062-91556-10 | E-Mail [email protected]-tools.com | Internet www.wolff-tools.de — 1 — � …
GM-200 16
Additional Information:
Operating Impressions, Questions and Answers:
- Manuals
- Brands
- SICK Manuals
- Measuring Instruments
- FLOWSIC100 Flare
- Operating instructions manual
Gas flow rate measuring device
Hide thumbs
Also See for FLOWSIC100:
- Operating instructions manual (266 pages)
- Addendum to operating instructions (88 pages)
,
1
2
Table Of Contents
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
-
page
of
184/
184 -
Contents
-
Table of Contents
-
Bookmarks
Table of Contents
Advertisement
Assembly and Installation
FLOWSIC100
3
Assembly and Installation
Project planning
Assembly
Installation
FLOWSIC100 · Operating Instructions · 8012513/YSA5/V 2-1/2016-07 · © SICK Engineering GmbH
49
Table of Contents
Previous Page
Next Page
- 1
- …
- 46
- 47
- 48
- 49
- 50
- 51
- 52
- 53
Show Quick Links
- Quick Links:
-
Calculating and Calibrating the Volume Flow
Hide quick links:
Advertisement
Table of Contents
Related Manuals for SICK FLOWSIC100
-
Measuring Instruments sick FLOWSIC100 Flare Operating Instructions Manual
For flare gas and steam flow measurements (266 pages)
-
Measuring Instruments SICK FLOWSIC100 Flare Addendum To Operating Instructions
(88 pages)
-
Measuring Instruments SICK FLOWSIC100 Flare-XT Operating Instructions Manual
Ultrasonic mass flow measuring device (188 pages)
-
Measuring Instruments SICK FLOWSIC600 Technical Information
Ultrasonic gas flow meter (262 pages)
-
Measuring Instruments SICK FLOWSIC600 Operating Instructions Manual
Ultrasonic gas flow meter for custody transfer and process applications (100 pages)
-
Measuring Instruments SICK FLOWSIC600 Manual
Ultrasonic gas flow meter (18 pages)
-
Measuring Instruments SICK FLOWSIC600 DRU Operating Instructions Manual
Gas flow meters (102 pages)
-
Measuring Instruments SICK FLOWSIC500 Operating Instructions Manual
Ultrasonic gas flow measuring devices (138 pages)
-
Measuring Instruments SICK FLOWSIC500 Operating Instructions Manual
Ultrasonic gas flow meter with
optional volume conversion (158 pages) -
Measuring Instruments SICK FLOWSIC500 Operating Instructions Manual
Ultrasonic gas flow meter with optional volume conversion (114 pages)
-
Measuring Instruments SICK FLOWSIC500 Operating Instructions Manual
Ultrasonic gas flow meter with optional volume conversion (176 pages)
-
Measuring Instruments SICK FLOWSIC600-XT Service Manual
Ultrasonic gas meter (68 pages)
-
Measuring Instruments SICK FLOWSIC600-XT Operating Instructions Manual
Ultrasonic gas flow meter (154 pages)
-
Measuring Instruments SICK FLOWSIC600-XT Operating Instructions Manual
Ultrasonic gas flow meter (164 pages)
-
Measuring Instruments SICK FLOWSIC60 Operating Instructions Manual
Continuous air flow measurement
for mining applications (71 pages) -
Measuring Instruments SICK FLOWSIC30 Installation, Operation And Maintenance Manual
Ultrasonic gas flow meter for wellhead applications (88 pages)
Related Content for SICK FLOWSIC100
-
NAV200 Assembly, Electrical Installation, Device Replacement
SICK NAV200
-
VICOTEC450 Assembly And Installation
SICK VICOTEC450
-
FLOWSIC200 Assembly And Installation
SICK FLOWSIC200
-
DUSTHUNTER SB30 Assembly And Installation
SICK DUSTHUNTER SB30
-
DUSTHUNTER T Assembly And Installation
SICK DUSTHUNTER T
-
DUSTHUNTER C200 Assembly And Installation
SICK DUSTHUNTER C200
-
FLOWSIC100 Flare Assembly And Installation
sick FLOWSIC100 Flare
-
S100 Standard System Plug Assembly
SICK S100 Standard
-
VMS410 Mounting, Electrical Installation, Maintenance And Replacement
SICK VMS410
-
MOC3ZA Electrical Installation
SICK MOC3ZA
-
FLOWSIC500 Installation
SICK FLOWSIC500
-
WTF9-3 Installation Électrique
SICK WTF9-3
-
deTec2 Core Installation
SICK deTec2 Core
-
FLOWSIC30 Installation
SICK FLOWSIC30
-
WSE16 Electrical Installation
SICK WSE16
-
InspectorP64 Series Electrical Installation
SICK InspectorP64 Series
SICK Manuals and Guides:
The main types of SICK FLOWSIC100 Flare instructions: user guide — rules of useing and characteristics, service manual — repair, diagnostics, maintenance, operation manual — description of the main functions of SICK FLOWSIC100 Flare equipment, etc.
Most of the instructions, that you can see on the site are uploaded by our users. If you have available a manual or document for SICK FLOWSIC100 Flare, which is currently not on the site or present in a different language version, we ask you to upload your document on website, using the «uploading form» available to all registered users.