Hanna hi 96711 руководство по эксплуатации

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

Instruction Manual

HI 96711

Free & Total Chlorine

Fast Tracker

Ion Specific Meter

w w w. h a n n a i n s t . c o m

1

sale@ponpe.com

www.ponpe.com

loading

Summary of Contents for Hanna HI 96711

  • Page 1: Instruction Manual

    Instruction Manual HI 96711 Free & Total Chlorine Fast Tracker Ion Specific Meter w w w. h a n n a i n s t . c o m sale@ponpe.com www.ponpe.com…

  • Page 2: Table Of Contents

    W A R R A N T Y W A R R A N T Y HI 96711 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge.

  • Page 3: Preliminary Examination

    Please examine this Product carefully. Make sure the instrument is not damaged. If any damage has occurred during the shipment, please notify your dealer. This HI 96711 Ion Specific Meter is supplied complete with: • Two Sample Cuvets and Caps •…

  • Page 4: General Description

    GENERAL DESCRIPTION GENERAL DESCRIPTION HI 96711 is an auto diagnostic portable microprocessor meter that benefits from Hanna’s years of experience as manufacturer of analytical instruments. It has an advanced optical system based on a special tungsten lamp and a narrow band interference filter that allows most accurate and repeatable readings.

  • Page 5: Abbreviations

    Immediately after installation of the tags you can start collecting data. Use the Fast Tracker Ion Specific Meter to take measurement and memorize the test result by pressing the Log-on-Demand key. Then, the meter will ask for the tag identification. Simply touching the iButton with the ®…

  • Page 6: Specifications

    SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS Range Free Cl 0.00 to 5.00 mg/L Total Cl 0.00 to 5.00 mg/L Resolution 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L Precision ±0.02 mg/L @ 1.00 mg/L Typical EMC Deviation ±0.01 mg/L Light Source Tungsten lamp with narrow band interference filter @ 525 nm Light Detector…

  • Page 7: Principle Of Operation

    The optical system of Hanna’s HI 96 colorimeters series is based on special subminiature tungsten lamps and narrow-band interference filters to guarantee both high performance and reliable results.

  • Page 8
    A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned and beamed to the sample contained in the cuvet. The optical path is fixed by the diameter of the cuvet. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity I or I.
  • Page 9: Functional Description

    FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION INSTRUMENT DESCRIPTION 1) Serial interface connector. 2) Tag reader connector. 3) Cuvet Holder. 4) Liquid Crystal Display (LCD). KEYPAD DESCRIPTION 5) CAL CHECK bi-functional key, press shortly to validate the actual calibration, or hold the key for 3 seconds to enter calibration.

  • Page 10
    DISPLAY DESCRIPTION 1) Four digit main display. 2) Measurement unit. 3) Three and a half digit secondary display. 4) “TIME” and “DATE”: appear with time (hh:mm) and date (yyyy.mm.dd) presentation. 5) The hourglass icon: appears when the instrument is making an internal check-up. 6) Status information.
  • Page 11: Guide To Display Codes

    Light over range: the cuvet is not inserted correctly and an excess of ambient light is reaching the detector. If the cuvet is properly installed, then contact your dealer or the nearest Hanna Customer Service Center. The meter has lost its configuration. Contact your dealer or the nearest Hanna Customer Service Center.

  • Page 12
    “Error”: the concentration of the used calibration solution is not correct. Repeat the calibration procedure with the right standard solution, and verify it is not expired. If the calibration procedure fails again, contact your dealer or the nearest Hanna Customer Service Center. LOGGING MODE MESSAGES The blinking “READ TAG”…
  • Page 13
    Measurement date: year (yyyy), month (mm), day (dd). Measurement time: hour (hh) and minutes (mm). Delete the last record. The blinking “DEL” indicates that the last record will be deleted if the /DEL key is pressed. Note:To delete a record, a second confirmation is required. Delete all records.
  • Page 14: General Tips For An Accurate Measurement

    GENERAL TIPS FOR AN ACCURATE MEASUREMENT GENERAL TIPS FOR AN ACCURATE MEASUREMENT GENERAL TIPS FOR AN ACCURATE MEASUREMENT GENERAL TIPS FOR AN ACCURATE MEASUREMENT GENERAL TIPS FOR AN ACCURATE MEASUREMENT The instructions listed below should be carefully followed during testing, to ensure best accuracy. •…

  • Page 15: Measurement Procedure

    MEASUREMENT PROCEDURE MEASUREMENT PROCEDURE MEASUREMENT PROCEDURE MEASUREMENT PROCEDURE MEASUREMENT PROCEDURE • Turn the meter on by pressing ON/OFF. When the LCD displays “—-”, the meter is ready. On the secondary LCD “P1” or “P2” will appear, indicating that the free or total chlorine parameter is selected.

  • Page 16
    • Replace the cap and shake gently for 20 seconds (or 2 minutes in case of seawater analysis). • Replace the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove. • Hold READ/TIMER for three seconds. The display will show the hourglass blinking and the count- down prior to measurement.
  • Page 17: Validation Procedure

    V A L I D A T I O N P R O C E D U R E Use the validation procedure to ensure that the instrument is properly calibrated. Warning: Do not validate or calibrate the instrument with standard solutions other than Hanna CAL CHECK Standards, otherwise erroneous results will be obtained.

  • Page 18: Calibration Procedure

    Note: It is possible to interrupt the calibration procedure by pressing CAL CHECK or ON/OFF. Warning: Do not validate or calibrate the instrument with standard solutions other than Hanna CAL CHECK™ Standards, otherwise erroneous results will be obtained. For accurate validation and calibration please perform test at room temperature, 18 to 25°C (64.5 to 77.0°F).

  • Page 19
    • Enter the calibration mode by holding the CAL CHECK key for three seconds. The date of the last calibration appears (e.g. 2004 05 23). “F.CAL” means that the factory calibration is selected. Note: At this moment it is possible to restore Factory Calibration (see instructions, page 20).
  • Page 20: Setup

    • The instrument will show for two seconds the CAL CHECK standard value and then “Stor” to confirm that the new calibration data has been accepted. The meter returns automatically to the measurement mode. Note: If the display shows “Err”, the calibration proce- dure failed.

  • Page 21
    Setting the date The year is displayed on the main display and the month and day on the secondary one. • To set a new date, press LOG/CFM. The last two digits of the year start blinking. • Set the year by using the UP and DOWN keys. •…
  • Page 22
    Setting the time The time is displayed in hours and minutes. • To set a new time, press LOG/CFM and the hour start blinking. • Set the hour by using the UP and DOWN keys. • Press LOG/CFM to switch to minutes. •…
  • Page 23: Logging

    L O G G I N G L O G G I N G L O G G I N G L O G G I N G L O G G I N G LOG DATA The log on demand feature is active after a valid measurement is obtained (no errors).

  • Page 24
    RECALL DATA The log records are organized in chronological order. The first displayed record is the last one stored. • To view logged data, press RCL/SETUP. • Use the UP or DOWN keys to scroll the log memory, record by record. •…
  • Page 25
    Notes:To make log searching easier, the following features are available: It is allowed to scroll from any record panel, except from delete last or delete all. If you scroll between records, the new selected record number is displayed for one second on the secondary LCD, together with “TAG”…
  • Page 26: Tag Installation

    T A G I N S T A L L A T I O N T A G I N S T A L L A T I O N T A G I N S T A L L A T I O N T A G I N S T A L L A T I O N T A G I N S T A L L A T I O N Place the tag near a sampling point.

  • Page 27: Accessories

    To avoid damage or burns, do not put the instrument in microwave ovens. For your own and the instrument safety do not use or store the instrument in hazardous environments. Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice.

  • Page 28
    SALES AND TECHNICAL SERVICE CONTACTS SALES AND TECHNICAL SERVICE CONTACTS SALES AND TECHNICAL SERVICE CONTACTS SALES AND TECHNICAL SERVICE CONTACTS SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel.

Скачать

Страница из 25

1

w w w . h a n n a — d e . c o m

Messgerät entspricht den 

CE-Richtlinien

Bedienungsanleitung

HI 96711

Photometer

Freies und Gesamtchlor

Stand: 04/10

background image

1

w w w . h a n n a i n s t . c o m

Instruction Manual

HI 96711C

Free and Total Chlorine

ISM

Instruction Manual

HI 96711C

Free and Total Chlorine

ISM

w w w . h a n n a i n s t . c o m

1

Dear Customer,

Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com.

TABLE OF CONTENTS

PRELIMINARY EXAMINATION ……………………………………………………………………………………

4

GENERAL DESCRIPTION ………………………………………………………………………………………….

5

ABBREVIATIONS ……………………………………………………………………………………………………

5

SPECIFICATIONS …………………………………………………………………………………………………….

6

PRECISION AND ACCURACY ……………………………………………………………………………………..

6

PRINCIPLE OF OPERATION ………………………………………………………………………………………

7

FUNCTIONAL DESCRIPTION ……………………………………………………………………………………..

9

ERRORS AND WARNINGS ……………………………………………………………………………………..

11

GENERAL TIPS FOR AN ACCURATE MEASUREMENT …………………………………………………….

13

STARTUP

……………………………………………………………………………………………………………

15

RANGE SELECTION ……………………………………………………………………………………………….

15

MEASUREMENT …………………………………………………………………………………PROCEDURE

16

VALIDATION ………………………………………………………………………………………PROCEDURE

18

CALIBRATION …………………………………………………………………………………….PROCEDURE

20

GLP …………………………………………………………………………………………………………………..

23

BATTERY ……………………………………………………………………………………….

MANAGEMENT

24

BATTERY ……………………………………………………………………………………….

REPLACEMENT

24

ACCESSORIES ………………………………………………………………………………………………………

25

WARRANTY …………………………………………………………………………………………………………

26

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895 , USA.

Hanna Instruments HI 96711 User Manual

PRELIMINARY EXAMINATION

Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please notify your Dealer.

Each HI 96711C Ion Selective Meter is supplied complete with:

Two Sample Cuvettes and Caps

CAL CHECK standard cuvettes

9V Battery

Scissors

Cloth for wiping cuvettes

Instruction Manual

Instrument Quality Certificate

Rigid carrying case

Note: Save all packing material until you are sure that the instrument works correctly. Any defective item must be returned in its original packing.

GENERAL DESCRIPTION

The HI 96711 is an auto diagnostic portable microprocessor meter that benefits from Hanna’s years of experience as a manufacturer of analytical instruments. It has the advanced optical system based on a special tungsten lamp and a narrow band interference filter that allows most accurate and repeatable readings. All instruments are factory calibrated and the electronic and optical design minimizes the need of frequent calibration.

With the powerful CAL CHECKTM validation function, you are able to validate good performance of your instrument at any time. The validation procedure is extremely user friendly. Just use the exclusive HANNA ready-made, NIST traceable standards to verify the performance of the instrument and recalibrate if necessary.

All instruments are splash waterproof and the lamp and filter units are protected from dust or dirt by a transparent cup. This makes the instruments fulfill field applications. Display messages aid the user in routine operation. The meter has an auto-shut off feature that will turn off the instrument after 10 minutes of non use in measurement mode or after 1 hour if left in calibration mode.

The meter uses an exclusive positive-locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell. It is designed to fit a cuvette with a larger neck making it easier to add both sample and reagents. The cuvette is made from special optical glass to obtain best results.

The HI 96711 meter measures the free and total chlorine (Cl2) content in water samples in the 0.00 to 5.00 mg/L (ppm) range. The method is an adaptation of the USEPA Method 330.5 for wastewater, and Standard Method 4500-Cl G for drinking water.

The reagents are in powder form and supplied in packets. The amount of reagent is precisely dosed to ensure the maximum repeatability.

ABBREVIATIONS

°C: degree Celsius °F: degree Fahrenheit

USEPA: US Environmental Protection Agency

mg/L: milligrams per liter. mg/L is equivalent to ppm (parts per million) mL: milliliter

mV: millivolts

SPECIFICATIONS

Range

Free Cl2 0.00 to 5.00 mg/L

Total Cl2 0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L

Accuracy

±0.03 mg/L ±3% of reading @ 25°C

Typical EMC Deviation

±0.01 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interference filter @ 525nm.

Method

Adaptation of the USEPA method 330.5 and Standard Method 4500-Cl G.

The reaction between chlorine and the DPD reagent causes a pink tint

in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10′ of non-use in measurement mode;

after 1 hour of non-use in calibration mode;

with last reading reminder.

Dimensions

192 x 104 x 69 mm (7.6 x 4.1 x 2.7″)

Weight

360 g (12.7 oz.).

REQUIRED REAGENTS

Code

Unit

Description

Quantity/test

HI

93701-0

Free Cl2

DPD Powder Reagent

1

packet

HI

93711-0

Total Cl2

DPD Powder Reagent

1

packet

PRECISION AND ACCURACY

Precision is how closely repeated measurements agree with each other. Precision is usually expressed as standard deviation (SD).

Accuracy is defined as the nearness of a test result to the true value.

Although good precision suggests good accuracy, precise results can be inaccurate. The figure explains these definitions. In a laboratory using a standard solution of 1.00 mg/L chlorine and a representative lot of reagent, an operator obtained with a single instrument a standard deviation of 0.02 mg/L.

6

PRINCIPLE OF OPERATION

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or crystal lattices.

If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the matter and on the physical-chemical characteristics of the substance according to the Lambert-Beer Law:

-log I/Io = ελ c d

or

A = ελ c d

Where:

-log I/Io =

Absorbance (A)

Io

=

intensity of incident light beam

I

=

intensity of light beam after absorption

ελ

=

molar extinction coefficient at wavelength λ

c

=

molar concentration of the substance

d

=

optical path through the substance

Therefore, the concentration «c» can be calculated from the absorbance of the substance as the other factors are known.

Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents. Given that the absorption of a compound strictly depends on the wavelength of the incident light beam, a narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize measurements. The optical system of Hanna’s HI 96 series colorimeters is based on special subminiature tungsten lamps and narrow-band interference filters to guarantee both high performance and reliable results.

HI 96 series block diagram (optical layout)

7

A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette. The optical path is fixed by the diameter of the cuvette. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity Io or I.

The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an electric current, producing a potential in the mV range.

The microprocessor uses this potential to convert the incoming value into the desired measuring unit and to display it on the LCD.

The measurement process is carried out in two phases: first the meter is zeroed and then the actual measurement is performed.

The cuvette has a very important role because it is an optical element and thus requires particular attention. It is important that both, the measurement and the calibration (zeroing) cuvettes, are optically identical to provide the same measurement conditions. Whenever possible use the same cuvette for both. It is necessary that the surface of the cuvette is clean and not scratched. This to avoid measurement interference due to unwanted reflection and absorption of light. It is recommended not to touch the cuvette walls with hands.

Furthermore, in order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the cuvette to prevent any contamination.

FUNCTIONAL DESCRIPTION

INSTRUMENT DESCRIPTION

1)

RANGE/GLP/

key

2)

CAL CHECK key

3)

ZERO/CFM key

4)

READ/

/TIMER

key

5)

ON/OFF key

6)

Liquid

Crystal Display (LCD)

7)

Cuvette

alignment indicator

8)

Cuvette holder

KEYPAD DESCRIPTION

ON/OFF: to turn the meter on and off.

ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to measurement, or confirm edited values. In calibration mode press to confirm factory calibration restore.

READ//TIMER: this is a multi-functional key. In measurement mode, press to make a measurement, or press and hold for three seconds to start a pre-programmed countdown prior to measurement. In GLP mode press to view the next screen.

CAL CHECK: this is a bi-functional key. Just press to perform the validation of the meter, or press and hold for three seconds to enter calibration mode.

RANGE/GLP/: this is a multi-functional key. Just press to change the parameter. Press and hold for three seconds to enter GLP mode. In calibration mode press to edit the date and time.

OPERATING MODES

Measurement mode: default operation mode, enables both validation and measurement.

Calibration mode: may be entered by keeping CAL CHECK pressed for three seconds (the “CAL” tag appears), it enables calibration of the instrument.

GLP mode: may be entered by keeping RANGE/GLP/ pressed for three seconds (“GLP” appears), it enables consulting of user calibration date or restore factory calibration.

Loading…

You can only view or download manuals with

Sign Up and get 5 for free

Upload your files to the site. You get 1 for each file you add

Get 1 for every time someone downloads your manual

Buy as many  as you need

Detail Specifications:

941/941343-hi_96711.pdf file (04 Apr 2023)

Accompanying Data:

Hanna HI 96711 Multimeter PDF Instruction Manual (Updated: Tuesday 4th of April 2023 05:45:55 AM)

Rating: 4.4 (rated by 45 users)

Compatible devices: 8505A, ND 2365-1, 7D13, 57030, 211101, HI 96734C, L34410A, HI 99556.

Recommended Documentation:

Text Version of Instruction Manual

(Ocr-Read Summary of Contents, UPD: 04 April 2023)

  • 28, Hanna HI 96711 28 SALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTS SALES AND TECHNICAL SERVICE CONTACTSSALES AND TECHNICAL SERVICE CONTACTS SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (0785…

  • 1, 1 HI 96711 Free & Total Chlorine Fast Tracker Ion Specific Meter www.hannainst.com Instruction Manual

  • 13, 13 Measurement date: year (yyyy), month (mm), day (dd). Measurement time: hour (hh) and minutes (mm). Delete the last record. The blinking “DEL” indicates that the last record will be deleted if the /DEL key is pressed. Note:To delete a record, a second confirmation is required. Delete all records. The blinking “DEL” indicates that all records will be deleted if the /DEL key is …

  • 24, 24 RECALL DATA The log records are organized in chronological order. The first displayed record is the last one stored. • To view logged data, press RCL/SETUP. • Use the UP or DOWN keys to scroll the log memory, record by record. • Press ZERO/ key to scroll through the record contents. The record panels are displayed one by one. Each record contains several panels…

  • 23, Hanna HI 96711 23 LOGGINGLOGGING LOGGINGLOGGING LOGGING LOG DATA The log on demand feature is active after a valid measurement is obtained (no errors). • To log a value, press LOG/CFM when a measurement result is displayed. The meter asks to “READ TAG” for identification of the sampling location. • To give the ID code for the identification of the sampling location, simpl…

  • 5, Hanna HI 96711 5 ABBREVIATIONSABBREVIATIONS ABBREVIATIONSABBREVIATIONS ABBREVIATIONS °C degree Celsius USEPA US Environmental Protection Agency °F degree Fahrenheit LCD Liquid Crystal Display mg/L milligrams per liter. mg/L is equivalent to ppm (parts per million) mL milliliter RTC Real Time Clock RH Relative Humidity iButton ®® ®® ® is registered Trademark of “MAXIM/DALLAS semiconductor Co…

  • 20, 20 • Turn the meter on by pressing ON/OFF. • Enter setup by holding the RCL/SETUP key for three seconds. The current date appears on the LCD. • Use the UP or DOWN keys to switch between time and date. SETUPSETUP SETUPSETUP SETUP • The instrument will show for two seconds the CAL CHECK standard value and then “Stor” to confirm that the new calibration data has be…

  • 10, 10 DISPLAY DESCRIPTION 1) Four digit main display. 2) Measurement unit. 3) Three and a half digit secondary display. 4) “TIME” and “DATE”: appear with time (hh:mm) and date (yyyy.mm.dd) presentation. 5) The hourglass icon: appears when the instrument is making an internal check-up. 6) Status information. 7) Battery icon: appears when the battery voltage is getting low. BE…

  • 15, 15 • Turn the meter on by pressing ON/OFF. When the LCD displays “—-”, the meter is ready. On the secondary LCD “P1” or “P2” will appear, indicating that the free or total chlorine parameter is selected. The display code that appears is the one of the last selected parameter. • Select the parameter program for Free Chlorine “P1”, or Total Chlorine “P2”, by pressi…

  • 18, 18 CALIBRATION PROCEDURECALIBRATION PROCEDURE CALIBRATION PROCEDURECALIBRATION PROCEDURE CALIBRATION PROCEDURE CALIBRATION PROCEDURE • Turn the meter on by pressing ON/OFF. When the LCD displays “—-”, the meter is ready. On the secondary LCD “P1” or “P2” will appear, indicating that the free or total chlorine parameter is selected. The display code…

Recommended Instructions:

DCR-IP45, CDX-R3300S, 0804, FASG7073LA, 200

  • ModelR5010TRMS Digital MultimeterInstructionManualPantone 534 BluePantone 123 YellowPantone 485 RedPantone 123 YellowPantone 534 BlueBlackRich Black — 20/20/20/100Blue — 100/80/30/5Yellow — 0/27/100/0Red — 10/100/100/5Yellow — 0/27/100/0Blue — 100/80/30/5Only if you REALLY need them:Pantone 534 Blue — 100/80/30/5Pantone 485 Red — 10/100/100/5Pantone 123 Ye …

    R5010 20

  • 1■■■■■ Pince multimètre■■■■■ Clamp multimeter■■■■■ Vielfachmesszange■■■■■ Pinza multimetro■■■■■ Pinza multimetricaF 07FRANÇAISENGLISHDEUTSCHITALIANOESPANOLNotice de fonctionnementUser’s manualBedienungsanleitungLibretto d’IstruzioniManual de Instrucciones …

    F07 24

  • Instruction ManualHI 93729Fluoride ISMhttp://www.hannainst.comThis Instrument is inCompliance with the CE DirectivesHanna Instruments reserves the right to modify the design, constructionand appearance of its products without advance notice.WARRANTYWARRANTYWARRANTYWARRANTYWARRANTYHI 93729 is warranted for two years against defects in workman-ship and materials when used for it …