- Manuals
- Brands
- Megger Manuals
- Measuring Instruments
- IDAX 300
Manuals and User Guides for Megger IDAX 300. We have 1 Megger IDAX 300 manual available for free PDF download: User Manual
- Manuals
- Brands
- Megger Manuals
- Measuring Instruments
- IDAX 300
- User manual
-
Contents
-
Table of Contents
-
Bookmarks
Quick Links
IDAX 300/322/350
IDAX 5 Software
Insulation diagnostic analyzers
User guide
Art No. ZP-AG03E
Doc. AG033586EE V06a 2020
Related Manuals for Megger IDAX 300
Summary of Contents for Megger IDAX 300
-
Page 1
IDAX 300/322/350 IDAX 5 Software Insulation diagnostic analyzers User guide Art No. ZP-AG03E Doc. AG033586EE V06a 2020… -
Page 3
© 2013-2020, Megger Sweden AB. All rights reserved. The contents of this manual are the property of Megger Sweden AB. No part of this work may be reproduced or transmitted in any form or by any means, except as permitted in written license agreement with Megger Sweden AB. Megger Sweden AB has made every reasonable attempt to ensure the completeness and accuracy of this document. -
Page 4: Table Of Contents
(366) Specimen capacitance above limit ….37 …………. 12 (367) Measured DC current > MaxDCCurrent …. 37 3.1 Front panel IDAX 300 ……..12 (368) Measured hum current > MaxHumCurrent ..37 7 Appendix B 3.2 Front panel IDAX 322 ……..13 Measurement technique ….
-
Page 5
AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 6: Introduction
PC. ▪ Reliable measurements in high-interference IDAX 300/322/350 provides an accurate and reliable environments condition assessment of insulation in transformers, ▪ Multi-function test set for transformer bushings, generators and cables.
-
Page 7: Warranty
If you need to ship the unit back to your local this user manual. representative or to Megger Sweden, make sure that the unit is well protected. Megger cannot take ▪ Check the equipment received against the packing list to ensure that all materials are present. Notify responsibility for any transport damages.
-
Page 8: Safety
47 / 63 Hz. The unit can also be returned to Before making connection to the power Megger at any time at no charge for source, determine that the instrument rating the disposal. matches the voltage of the power source and…
-
Page 9: Safety Instructions
The AC power cord is the disconnecting device. Maintenance Always disconnect test leads from test specimen before attempting to disconnect them at the DISCONNECT the MAINS plug before any test set. cleaning or maintenance. AG033586EE ZP-AG03E IDAX 300/322/350…
-
Page 10
Polish with a soft, dry cloth, taking care not to scratch the display screen cover. The cables and mating panel receptacles can be cleaned with isopropyl or denatured alcohol applied with a clean cloth. IDAX 300/322/350 ZP-AG03E AG033586EE… -
Page 11
2 SAFETY AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 12: Instrument Description
3 INSTRuMENT DESCRIPTION Instrument description 3.1 Front panel IDAX 300 MAINS To turn the instrument On/Off Protective Earth/Ground connector Mains connector Connector for connection to test object ground. Fuse This is also used as measurement input. Use small screwdriver to gently pry out to change Always connect the instrument to station earth/ the fuse, F1, 2AT slow blow.
-
Page 13: Front Panel Idax 322
This is also used as measurement input. Always connect the instrument to station earth/ground using the separate earth/ground cable. The separate earth/ ground wire should be the first connection made and the last removed. AG033586EE ZP-AG03E IDAX 300/322/350…
-
Page 14: Front Panel Idax 350
USB memory to the internal PC USB Port For connecting to an external computer. Put the switch in EXT position. To use the internal PC, put the switch in INT position. IDAX 300/322/350 ZP-AG03E AG033586EE…
-
Page 15: Standard Accessories
▪ Test & Calibration Certificate and can be used together with IDAX 4 and 5. Be ▪ uSB cable, A — B, 2 m (5 ft) (IDAX 300/322 only) careful if installing other manufacturer’s software ▪ Transport case IDAX 300 (IDAX 300 only) on the IDAX unit.
-
Page 16: Preparing Instrument
AC,50/60Hz) to the IDAX instrument and switch it on. For IDAX 322 connect external interlock and strobe (optional). For IDAX 300/322, connect a communication cable, either an Ethernet cable or USB cable, between IDAX 300/322 and the computer. If you use Ethernet, you may connect it to company network.
-
Page 17: Idax 322 Interlock Positions
GSTg-B Ground Blue GSTg-RB Ground Red and Blue Connection examples – Power transformers IDAX transformer measurements/connections are organized as follows: IDAX 300 (two winding transformer) Test No. Measure Test mode Output Input 1 Input 2 (Yellow) (Red) (blue) GSTg-RB Varning uST-R ACTIVE output –…
-
Page 18
2 and test 3. In case of a three winding transformer the IDAX 300 second input channel is utilized. The test leads need to be moved between test 2 and test 3 and between test 4 and test 5. -
Page 19
3 INSTRuMENT DESCRIPTION AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 20: Idax 5 Software
Then choose suitable test object template, Note At any time you may access Nameplate and e.g. 2W-transformer (two-winding change/add information by pressing “Edit transformer). Nameplate” button in the program main Enter mandatory temperature data window. for ambient and object temperature (highlighted). IDAX 300/322/350 ZP-AG03E AG033586EE…
-
Page 21: Measurement
There are two pictures showing the two versions of Note From 2012, IDAX has single input terminals IDAX 300, IDAX 300 delivered 2011 or earlier (to the and utilizing a new methodology where stray left) and IDAX 300 delivered 2012 or later.
-
Page 22: Disconnect/Re-Connect
In Moisture Test, limits for insulation geometry matching are based on the selected test object. For power transformer, limits are pending if the design is shell or core type, as described in the application note “Dielectric Frequency Response (DFR) Measurement IDAX 300/322/350 ZP-AG03E AG033586EE…
-
Page 23: Measurement Using Individual Temperature Correction For Single Material (Itc1)
There are two pictures showing the two versions of Insulation assessment guidelines IDAX 300, IDAX 300 delivered 2011 or earlier (to the left) and IDAX 300 delivered 2012 or later. The default limits of %DF at 20°C are based on…
-
Page 24: Left-Section Of The Main Window
4 IDAX 5 SOFTWARE 4.3 Operating IDAX 5 Bounded Paper (RBP). The evaluation of results at 1Hz is based on Megger experience. Left-section of the main window Disconnect/re-connect It is possible to disconnect the computer and let the IDAX unit run standalone.
-
Page 25: Middle Section Of The Main Window
▪ At the bottom there is a button for deleting the evaluation of the measurement after correction to selected sweep as well as the button for opening a 20°C. The assessment is based on material selection new file. and object temperature during the measurement. AG033586EE ZP-AG03E IDAX 300/322/350…
-
Page 26
The graph tab will let you view and compare multiple measurements. You can look at both %DF/%PF or capacitance at the same time. A Y-axis will appear for display. It is also possible to view the measurement points in a table. IDAX 300/322/350 ZP-AG03E AG033586EE… -
Page 27
Will set frequency for presentation in ers, or in situations with very high interfer- Test tabs, for Line Frequency %DF/%PF ence, standard single-frequency measure- measurements and power frequency ments are recommended. interference measurements (in the initialization procedure). AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 28
Default materials are OIP, RIP and Kraft paper but more materials can be added. Press “Edit Assessment Limits” to show a dialogue window where modifications of assessment limits for selected materials and frequencies can be set. IDAX 300/322/350 ZP-AG03E AG033586EE… -
Page 29: Right-Section Of The Main Window
Edit Analysis window will pop up and Oscilloscope modifications to default setting values are possible. The Oscilloscope displays the IDAX 300 internal signals in Red, White and Green. The red signal is Note Each type of liquid has its own limits.
-
Page 30
USB or Ethernet, they will pop-up as a button labelled e.g. “Connect to unit p23”. Select all Selects all measurements in dialog window. The measurements not needed in report can be deselected. IDAX 300/322/350 ZP-AG03E AG033586EE… -
Page 31: Analysis Window
Parameters for insulation modelling. Start/Stop/Remake Press button“Start”to start a test, and press the button “Stop” to stop a test. If you have marked an executed test in the legend, you are not able to override this test AG033586EE ZP-AG03E IDAX 300/322/350…
-
Page 32: New File
Insulation Resistance (IR) New File at 60 seconds, Dielectric Absorption Ratio (DAR) and Polarization Index (PI). Note The measurement time needs to be sufficient (10 minutes) to display all values. IDAX 300/322/350 ZP-AG03E AG033586EE…
-
Page 33
4 IDAX 5 SOFTWARE AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 34: Calibration
It is recommended to calibrate the IDAX at least once a year. There are two options regarding calibration of IDAX: ▪ The IDAX unit can be sent to Megger for calibration ▪ Calibration can be performed by the customer using the optional Calibration Set, Cat. No. AG-90010.
-
Page 35
5 CALIBRATION AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 36: Appendix A Error Messages
▪ change limit set by MinSpecimenC to an for IDAX software 4.0 or newer, the IDAX software approximately 10% lower value than measured do not understand the incapability and it usually capacitance result in error 347. IDAX 300/322/350 ZP-AG03E AG033586EE…
-
Page 37: Specimen Capacitance Above Limit
(e.g. measure to ground without use of guard) ▪ It is possible increase limit level for DC current in Measurement Template, but only when difference is very small and all other possibilities excluded AG033586EE ZP-AG03E IDAX 300/322/350…
-
Page 38: Appendix B Measurement Technique
The analogue signals (voltages) are then converted to digital samples of the signals that are used in subsequent calculations. IDAX 300/322/350 ZP-AG03E AG033586EE…
-
Page 39: Modelling Principles
20 e where W — activation energy of pressboard 0,9 eV k — Boltzman constant, T — object temperature, T reference temperature (20°C) in Kelvin degrees. ω — angular frequency at 20°C. AG033586EE ZP-AG03E IDAX 300/322/350…
-
Page 40: Sample Modelling
PF = cosφ = Series: tanδ = – Re[Z] Z = R jωC Im[Z] = – ωIm[Z] If tanδ and cosφ (PF) are small, then tanδ » cosφ = PF (E.g. tan =0.1 corresponds to PF=0.0995) IDAX 300/322/350 ZP-AG03E AG033586EE…
-
Page 41
(or resistivity). The dielectric and resistive models are derived as follows: jωC C = C (ε’ – jε’’) Dielectric: = Re� � ε’ jωC Δ ε’ ε’ = –Im� � ε’’ jωC Resistive: = Re� � ε’ jωC AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 42: Specifications
Ambient temperature GSTg-R: GST: Guard Red, Ground Blue GSTg-B: GST: Guard Blue, Ground Red Operating IDAX 300/322: -20°C to +55°C (-4°F to GSTg-RB GST: Guard Red and Blue +131°F) *IDAX 300 can measure mutiple test IDAX 350: -10°C to +55°C (14°F to +131°F) modes in an automatic sequence.
-
Page 43
AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 44: Index
Safety ………….. 8 Safety instructions ……….. 9 Sample modelling ………. 40 Service and support……… 7 Specifications ……….42 Standard Accessories ……..15 Start the IDAX 5 software ……20 Symbols …………8 Test modes ………… 17 IDAX 300/322/350 ZP-AG03E AG033586EE…
-
Page 45
8 SPECIFICATIONS AG033586EE ZP-AG03E IDAX 300/322/350… -
Page 46
Your “One Stop” Source for all your Megger Group Limited UNITED KINGDOM electrical test equipment needs Dover, Kent CT17 9EN ▪ Battery Test Equipment ENGLAND ▪ ▪ ▪ Cable Fault Locating Equipment AUSTRALIA POLAND ▪ ▪ BULGARIA ROMANIA ▪ Circuit Breaker Test Equipment ▪…
- Госреестр 48036-11
- Межповерочный интервал 24 месяца
- Гарантия 1 год
- Производитель Megger
- Производство: Великобритания
Особенности:
- Megger IDAX 300 — быстрая и точная оценка влаги в силовых трансформаторах
- Хорошо зарекомендовавшая себя технология, приборы Megger IDAX 300 работают в полевых условиях более 10 лет
- Надежные результаты при любой температуре
- Автоматизированный анализ содержания влаги и проводимости масла — нажатием пальца
- Точное измерение во всей области частот, максимальная помехоустойчивость
- Выполняет проверку изоляции трансформаторов, вводов, кабелей и генераторов
Megger IDAX 300 обеспечивает точную и надежную оценку состояния силовых трансформаторов, вводов, генераторов и кабелей. Система Megger IDAX 300 повышает эффективность технического обслуживания, обеспечивая оптимизацию нагрузки и срока службы оборудования. IDAX-300 в три раза меньше и легче своего предшественника IDAX-206, но обладает той же способностью предоставлять надежные данные.
Усовершенствованное программное обеспечение позволяет проводить испытания и проще, и быстрее.
Megger IDAX 300 измеряет емкость и потери в изоляции между витками в множестве контрольных точек области частот. Получаемая уникальная информация позволяет оценить состояние масла и твердой изоляции, и ее можно использовать для разграничения между влажностью и другими потенциальными проблемами. Измерение возможно при любой температуре.
Применение
Из-за старения целого поколения силовых трансформаторов сегодняшняя электроэнергетика столкнулась с серьезными проблемами, так как отказы трансформаторов, их ремонт и соответствующие затраты приводят к значительным экономическим потерям. Трансформаторы стали наиболее проблемными элементами электрических сетей.
Потребность в надежных методах контроля и диагностики привела ведущих экспертов в этой области к выбору новых технологий, которые позволяют значительно увеличить надежность и оптимизировать работу каждого из элементов сети [1].
Megger IDAX 300 – один из самых современных приборов диагностики изоляции, основанный на методе диэлектрической спектроскопии. Этот метод анализа до настоящего времени использовался только в лабораторных условиях и IDA/IDAX – это первый прибор, который позволил реализовать данный метод в полевых условиях (1997). Прибор Megger IDAX 300 и использованная в нем методика подтверждена эксплуатацией в промышленности различных стран мира в последние десять лет.
Одно из самых важных применений IDAX – это определение старения или содержания влаги в изоляции трансформаторов, так как наличие влаги в изоляции значительно ускоряет процесс ее старения. Так же влага может вызывать образование пузырьков между обмотками, что приводит к аварийным отказам. IDAX обеспечивает возможность достоверной оценки содержания влаги за одно испытание. Это испытание может быть выполнено при любой температуре и занимает менее 18 минут.
Решения по ремонту и/или замене должно быть основано на знании состояния изоляции и ожидаемой нагрузки оборудования. Добавление нескольких лет к нормативному периоду эксплуатации трансформатора, генератора и кабеля за счет оптимизации условий работы (загрузки) на основе надежных диагностических данных позволит владельцу оборудования существенно экономить на техническом облуживании.
Методика также может использоваться для оценки состояния и старение изоляции вводов, трансформаторов тока и выключателей.
Многочисленные текущие проекты исследований в институтах и университетах по всему миру добавляют опыта и ценности для пользователей Megger IDAX 300.
Общие характеристики | |
Питание (номинал) | 90 – 265 В АС, 50/60 Гц |
Потребляемая мощность | 250 ВА |
Размеры | Прибор 335 x 300 x 99 мм Кейс для переноски 520 x 430 x 210 мм |
Масса | Прибора 5 кг Масса с кейсом 10 кг Масса аксессуаров 8,5 кг в мягком кейсе |
Порты связи | USB 2.0 и LAN |
Измеряемые параметры | |
Каналы | 1 канал стандартно, 2 канала опция |
Погрешность | 0.5% + 1 pФ |
Тангенс угла потерь | Диапазон 0 – 10 (с точностью измерения емкости или точнее) |
Погрешность | >1% +0.0003, 1 мГц – 100 Гц, C > 1 нФ >2% +0.0005, 100 Гц – 1 кГц, C > 1 нФ |
Выходы | |
Диапазоны по напряжению/току 10 В | 0 – 10 Vпик / 0 – 50 мА |
Диапазоны по напряжению/току 200 В | 0 – 200 Vпик / 0 – 50 мА |
Частота | 0,1 мГц – 5 кГц |
Требования к ПК | |
Операционная система | Windows 2000 / XP / Vista / 7 |
Процессор | Pentium 500 МГц |
Оперативная память | 512 Мб ОЗУ или более |
Интерфейс | USB 2.0 |
1. Анализатор состояния изоляции трансформаторов Megger IDAX 300 – 1 шт.;
2. Кабели для измерений (18м синий и красный) – 1 комп.;
3. Кабель питания – 1 шт.;
4. Кабель заземления – 1 шт.;
5. Кабель для генератора (18м) – 1 шт.;
6. USB-кабель – 1 шт.;
7. CD-диск c ПО IDAX for Windows – 1 шт.;
8. Кейс для транспортировки – 1 шт.;
9. Сумка для кабелей – 1 шт.;
10. Руководство по эксплуатации – 1 экз.;
11. Методика поверки – 1 экз.
Рекомендуем посмотреть также
User Manual: IDAX300-and-IDAX350-User-Manual-Brochure
Open the PDF directly: View PDF .
Page Count: 20
IDAX Monitoring SW to monitor processes using Dielectric Frequency Response User’s Manual WWW.MEGGER.COM Art No. ZP-AG02E Doc. AG033573AE V01a 2017 IDAX Monitoring SW to monitor processes using Dielectric Frequency Response User’s Manual NOTICE OF COPYRIGHT & PROPRIETARY RIGHTS © 2016, Megger Sweden AB. All rights reserved. The contents of this manual are the property of Megger Sweden AB. No part of this work may be reproduced or transmitted in any form or by any means, except as permitted in written license agreement with Megger Sweden AB. Megger Sweden AB has made every reasonable attempt to ensure the completeness and accuracy of this document. However, the information contained in this manual is subject to change without notice, and does not represent a commitment on the part of Megger Sweden AB. Any attached hardware schematics and technical descriptions, or software listings that disclose source code, are for informational purposes only. Reproduction in whole or in part to create working hardware or software for other than Megger Sweden AB products is strictly prohibited, except as permitted by written license agreement with Megger Sweden AB. TRADEMARK NOTICES Megger® and Programma® are trademarks registered in the U.S. and other countries. All other brand and product names mentioned in this document are trademarks or registered trademarks of their respective companies. Megger Sweden AB is certified according to ISO 9001 and 14001. Postal address: Megger Sweden AB Box 724 SE-182 17 DANDERYD SWEDEN T +46 8 510 195 00 F +46 8 510 195 95 AG033573AE Visiting address: Megger Sweden AB Rinkebyvägen 19 SE-182 36 DANDERYD SWEDEN seinfo@megger.com www.megger.com ZP-AG02E IDAX Monitoring 3 Contents 1 Introduction ................................................................ 6 Software Installation...........................................6 2 Getting Started ................................................................ 8 3 Sample connection .............................................................. 10 Oven bushings and high temperature coaxial cables... 10 4 Sweep and temperature settings....... 12 4.1 Overview....................................................12 4.2 Sweep........................................................13 4.3 Temperature...............................................13 Automatic entry......................................................... 13 Manual entry............................................................. 13 Loading a temperature profile.................................... 14 5 Export and Report .............................................................. 16 5.1 Export and report........................................16 Design a personalized report...................................... 16 5.2 Settings......................................................17 5 Appendix A .............................................................. 18 Individual Temperature Correction (ITC).............18 4 IDAX Monitoring ZP-AG02E AG033573AE AG033573AE ZP-AG02E IDAX Monitoring 5 1 Introduction 1 Introduction IDAX Monitoring is an additional feature to IDAX 5.0 SW to make it possible use a IDAX 300 to monitor processes using DFR (Dielectric Frequency Response). Examples of processes to be monitored are: ▪▪Dry-outs, reduction of moisture content, of unimpregnated as well as impregnated cellulose in e.g. power transformers. For such processes ITC (Individual Temperature Corrected) dissipation factor are most commonly used Software Installation Monitoring functionality is installed during installation of IDAX 5.0 software, but it is disabled by default. In order to enable this functionality a licence code must be obtained. Monitoring functionality can be activated by entering user name and corresponding activation code. ▪▪Impregnation, e.g. impregnation of dry cellulose with insulation liquids or resins. For such processes ITC (individual Temperature Corrected) capacitance may be used; however, be aware of that temperature change the geometry and the effect of geometry is not corrected in the ITC. ▪▪Curing of resins. Curing status strongly affect the dielectric properties of the insulation material and therefore monitor of such process is possible. See e.g. “Analysis of Dielectric Response Measurement Methods and Dielectric Properties of Resin-Rich Insulation During Processing” by Anders Helgeson IDAX Monitoring can be used in factory environment as well as in the field. The connection to a PC or other intelligent system is the same as for IDAX via the USB port and/or the Ethernet port on the IDAX front panel. The IDAX may be connected to the PC or other intelligent system operation, however, you may disconnect and reconnect at a later occasion during the monitoring process. Then IDAX Monitoring SW automatically collect all data points collected. 6 IDAX Monitoring ZP-AG02E AG033573AE 1 Introduction AG033573AE ZP-AG02E IDAX Monitoring 7 2 Getting Started 2 Getting Started This section provides a step-by-step instruction. 1] Start the IDAX 5.0 SW. SW will appear similar to picture below. 2] All data will always be saved in a file, therefore, you should press the “New File” button (right-hand side). Important “Monitor” has inherited most of functionality from “Moisture” and therefore “Test Objects” available is limited to those currently available in “Moisture”, and only one “Test Mode” per test is allowed. For monitoring, UST configuration is strongly recommended and therefore the “Two Winding Transformer” with “Test Mode” CHL (UST-R) can be used for almost all samples (if a GST measurement is required, select e.g. “Auto Transformer Without Tertiary”). 4] Select “Test Object” (normally Two Winding Transformer even if you intend monitoring a current transformer, bushing etc.). 5] Enter temperatures into compulsory fields 3] Then the “Create a New File” window will appear, defaulted to “Moisture”. Press the “Monitor” tab. For more information by “Moisture”/”DFR”/”%DF/%PF/”DC” and the tabs “Single Test”/”Multiple Tests/”Use Measurement as Template”, see IDAX User’s manual. “Ambient Temperature” and “Object Temperature”, e.g. 22C. These two fields are there as inherited from “Moisture”, these values will not be used in the Monitoring function. 6] Enter other fields if you need information (or add information to nameplate later). 7] Press “Create”. 8] Save the “File” at location of your choice 8 IDAX Monitoring ZP-AG02E AG033573AE 2 Getting Started Act. Energy 9] Now, you have a “Test” (labeled CHL, upper, left-hand side of window) that can be run (after connected to IDAX and start by press “Start” button in lower right-hand side corner Activation Energy used for ITC temperature correction. Incorrect values scale data slightly incorrect but has limited impact as long as Reference temperature is not too far from actual measurement temperature. Typical values (Cellulose means pressboard, Kraft paper or Crepe paper): ▪▪Oil-impregnated cellulose, -20-80°C: 0.9 eV ▪▪Non-impregnated cellulose, >80°C: 1.0 eV Temperature Reference temperature for ITC Moisture Select a database (Default or other, to be modified in Settings tab), and Moisture is plotted. Note that in case Moisture is selected, Activation energy and (reference) Temperature is fixed to value determined by the selected database In lower part of right-hand side, it is a section labeled “Settings” that display part of settings for the test: Stop frequency Lowest frequency to be used in DFR sweep of the full sweep defined on settings tab (see section Settings tab). You may change “Stop frequency” in drop down list. Test mode By clicking this button you see the basic connection diagram. Voltage (RMS) Applied voltage level. You may use any voltage level, however, if you use voltage level below 7 Vrms (10 V peak), you preferable change from 200 V peak amplifier to 10 V peak amplifier on Settings page (may be advisable measuring under Vapor Phase conditions). AG033573AE ZP-AG02E IDAX Monitoring 9 3 Sample connection 3 Sample connection This section provides a step-by-step instruction on how to connect the test specimen. 1] Start the IDAX 5.0 SW. 2] All data will always be saved in a file, therefore, you start by press “New File” button. Oven bushings and high temperature coaxial cables A common application of IDAX Monitoring is monitor sample during processing in a climate chamber, e.g. oven drying or Vapor Phase drying. In this case, bushings for bringing cables into the chamber are necessary and often also special high-temperature cables, which does not get damaged by high temperatures and kerosene. There are different types of connectors and cables on the market that will work for this application, description below will act as example. BNC-connectors: ▪▪BNC Adapter, female/female: Rosenberger 51K501K00N4 ▪▪High temperature cable: PRO POWER, RG142 Coaxial Cable, RG142, 50 ohm, order code 2064894 ▪▪BNC connector (fit to cable above): Rosenberger 51S107-108N5 10 IDAX Monitoring ZP-AG02E AG033573AE 3 Sample connection AG033573AE ZP-AG02E IDAX Monitoring 11 4 Sweep and temperature settings 4.1 Overview Moisture In the Overview tab, the measurement results as function of time are presented. Note One or more parameters (e.g. %DF) to be plotted need to be activated; otherwise the graph will be empty. By select a database, you activate Moisture assessment based on table database selected (see Settings page). Inactivate the function by selecting “None”. If a database is selected, Frequency (Hz), Act Energy (eV) and Temperature (°C) are set to fixed values determined by database selected. The parameters buttons (below the graph) are: %M The moisture content in % based on Moisture assessment table: Left-hand side axis °C The temperature in degree Celsius; Left-hand side axis C Capacitance in Farad; Right-hand side Y-axis C’’ Complex part of complex capacitance (Impedance Z = 1/(j*2*pi*frequency*(C-jC’’)) %DF Dissipation factor*100 = Loss current*100 / Capacitive current = C’’*100/C %PF Power factor*100 = Loss current*100/Total current = C’’*100/(sqrt(C’’^2+C^2)) Based on settings of “Settings” (lower part of righthand section of window) the frequency sweep data are plotted differently: Frequency (Hz) Frequency for data to be plotted Act. Energy (eV) Activation energy if ITC temperature correction is enabled Temperature By select a temperature you activate (°C) ITC temperature correction to selected temperature (otherwise, select None) 12 IDAX Monitoring ZP-AG02E AG033573AE 4 Sweep and temperature settings 4.2 Sweep 4.3 Temperature In the Sweep tab, the measurement sweeps are presented. At left-hand side the frequency sweeps are listed labeled with Date+Time. In the temperature tab, the measurement temperatures are presented. At left-hand side the temperature labeled with Date+Time is listed (format according to computer settings) at the right hand the temperature at the specific time. Note One or more parameters (e.g. C and %DF) to be plotted need to be activated; otherwise the graph will be empty. The parameters buttons (below the graph) are: C Capacitance in Farad; Right-hand side Y-axis C’’ Complex part of complex capacitance (Impedance Z = 1/(j*2*pi*frequency*(C-jC’’)) %DF Dissipation factor*100 = Loss current*100 / Capacitive current = C’’*100/C %PF Automatic entry Measurement temperatures may be read using either automatic logging device (default) or logfile. Automatic logger use Modbus data interface and is connected via LAN (Ethernet). In order to use it the IP address or hostname must be entered in the field “IP / Hostname” Power factor*100 = Loss current*100/Total current = C’’*100/(sqrt(C’’^2+C^2)) You may change interval at which data are read and saved. It must be verified with specifications of logging device if other parameters need to be changed in case different logger than Querx-PT100 is used. Use any compatible temperature logger with Modbus TCP interface. Manual entry ▪▪Press “Add” button to enter temperature manually. ▪▪To delete a temperature data point, press “Del”. ▪▪Delete all data points by pressing “Clear”. ▪▪Load temperature data points by pressing the “Load” button (more information about “Load” see section “Loading a temperature profile” below). ▪▪The “Edit” function is activated by double-clicking on a temperature in the list. The data points are automatically sorted by date/time and extrapolated before after covered times (using AG033573AE ZP-AG02E IDAX Monitoring 13 4 Sweep and temperature settings two last data points if more than one data point). If you need two constant temperatures, e.g. 100°C and 120°C use 4 data points (two at 100°C and 2 at 120°C). Note If no temperature is entered “Ambient Temperature” from Nameplate is used. Loading a temperature profile It is also possible to load a complete temperature profile. The file should be semicolon separated *.csv with the following format: YYYY-MM-DD HH:MM;Temperature Example YYYY-MM-DD HH:MM;Temperature can be: 2013-09-19 09:42;21,5 The first row means 19th (19) September (09) 2013, at 09:42 (24h syntax). Temperature in degree Celsius is shown after the semicolon. 1] By pressing “Load” button, an Open dialog window will appear showing all files with extension “.csv”. The “Load” function accepts both point and comma as decimal separator. Note 14 By changing view to show all files “*.*” it is possible to show files created e.g. as *.txt IDAX Monitoring ZP-AG02E AG033573AE 4 Sweep and temperature settings AG033573AE ZP-AG02E IDAX Monitoring 15 5 Export and Report 5 Export and Report 5.1 Export and report 1] Press the “Export” button to export all measurement data into a *.txt-file 2] Pressing the “Report” button connects data to Microsoft Word and opens a report dialog where pre-defined report templates are available. 3] Select “Monitor_Example_A4” and press “Generate”. Note 3] Below you see the data “Overview (picture)” that earlier was placed under the Overview name. By delete the square (the overview picture from template) and then drag-anddrop e.g. Sweep (Picture) to same location, you have changed the template. All graphs/pictures will be in report as in SW, if you have enabled only %DF in Overview graph, only %DF will be present in Overview graph in report. 4] Save it under a new name that will appear under Template next time you press “Report” button. Design a personalized report The most simple way of designing a personalized report template is to select an existing template and then modify it. Example 1] Select “Monitor_Example_A4”. 2] Press “Modify”. The “Monitor_Example_A4” report can now be modified under “File”. You may design the report using normal Word functionality and “drag-and-drop” data from the legend on left-hand side of report generator. 16 IDAX Monitoring ZP-AG02E AG033573AE 5 Export and Report 5.2 Settings ▪▪At left hand-side, bottom section, you may add a new table “New”, delete an exciting table “Remove” or “Import” a table into the SW (e.g. created by another SW installation). Export is available from bottom part of table (righthand side). ▪▪“Moisture Table” is the actual table selected • Name is simply the name. • Frequency affects the relation between dissipation factor and moisture, therefore the frequency for the table need to be stated. • Temperature affects the relation between dissipation The Settings Tab looks similar to standard IDAX SW, described in IDAX User’s manual. In bottom, middle, part of the window a section labeled Monitoring is added with the following functions: Sleep Time: (Default 15 minutes) After a measurement is completed, the system waits (sleep) for the selected time before next measurement is started. If a frequency sweep takes about 5 minutes to perform (1 kHz to 10 mHz, single frequency measurements) and sleep time is 15 minutes you will get a new sweep every 20 minutes. Selectable alternatives are none, 5, 10, 15, 20, 30, 60 or 120 minutes. factor and moisture, therefore the frequency for the table need to be stated. • Activation energy for the material assessed affects the ITC correction and therefore it need to be stated (estimated). • The table is the relation between %DF (dissipation factor/100) and moisture content. Values between points are interpolated using Log(%DF) and Log(Moisture%) and extrapolated from two last points outside covered range. Delete data pair by pressing the red “X” Add new data pair by pressing the green “Add” button. Press “Save” to save the data table. Press“Export” of table, to be stored or imported by other used. Time Format: (Default “Process Time”) Edit assessment limits Each DFR sweep has a Date+Time stamp (when it starts) and each temperature also has a date+Time stamp. You can select if data should be plotted with X-axis set as Date+Time (Computer time) or if time = 0 is set at start of first measurement sweep (Process time). Assessment limits for %DF, %PF and Moisture can be adjusted to suit particular requirements. These limits are applicable only to measurements during monitoring process. Edit Ranges (screen shot below) A moisture table may be selected which maps an ITC corrected dissipation factor at a certain frequency to a moisture content. ▪▪At left hand side you have available tables. The “Default” table comes with installation. AG033573AE ZP-AG02E IDAX Monitoring 17 5 Appendix A 5 Appendix A Individual Temperature Correction (ITC) An important property on most insulation materials is the possibility makes a master curve from a set of DFR data obtained at different insulation temperatures [1]. 10 DF at 20C DF at 40C DF at 60C DF at 80C Dissipation factor 1 Figure 2. Master curve at 20°C, dry Kraft paper 0,1 Since the master curve now describe the relation between frequency and temperature, dielectric parameters such as the dissipation factor (DF) at 50 Hz can now be plotted as function of temperature [2]. See Figure 3. 0,01 0,001 0,001 0,01 0,1 1 10 Frequency, [Hz] 100 1000 Figure 1. DFR of dry mass-impregnated Kraft paper at different temperatures. To create a master curve, you prepare the insulation sample and measure the sample at different temperatures while making sure that other entities such as moisture content is kept constant. See Figure 1. The temperatures in Figure 1 are 20°C (blue, lowest dissipation factor at 1 Hz), 40°C (green), 60°C (orange) and 80°C (red, highest dissipation factor at 1 Hz). This technique is used to establish individual temperature correction factors (ITC) [3] to correct measurements performed at a higher, or lower, than a reference temperature of e.g. 20°C. For example if measure the DF at 50 Hz of the insulation material in Figure 1-3 at 40°C, the DF is about 0,004 or 0,4%, see figure 1 (or Figure 3). By measure a frequency sweep and plot it as function of temperature, the estimated dissipation factor at 20°C is about 0,005 (0,5%). The curves are then shifted in frequency by using Arrhenius equation with activation energy as a single parameter and temperature as variable. Figure 2 show the master curve created from measurement data in Figure 1. The activation energy is 0.9 eV. Figure 3. DF at 50 Hz as function of temperature (°C) A similar application is to estimate the DF at a high operating temperature based on a measurement at ambient temperature. In Figure 4, the DF as function of frequency is measured on an about 100 m oil-filled 400 kV cable at about 20°C; the cable connects a Generator Step-Up transformer to the substation. 18 IDAX Monitoring ZP-AG02E AG033573AE 5 Appendix A REFERENCES [1] U Gafvert, “Condition Assessment of Insulation Systems, Analysis of Dielectric Response Methods, NORD-IS 1996 [2] ITC patent. US2010/0106435 A1 [3] M Ohlen et al, Dielectric frequency response and temperature dependence of dissipation factor, TechCon Asia Pacific, Australia, 2009 Figure 4. DF as function of frequency for an oil-filled 400 kV cable The DF is high, about 0,01 (1%) at 50 Hz but is it critical or not? By looking at the 50 Hz DF as function of temperature, see Figure 5, heat dissipation calculations showed that the if DF at 60-80°C is above about 10%, there is a large risk of a thermal runaway breakdown. The cable was replaced prior any failure. Figure 5. DF at 50 Hz as function of temperature. Data from figure 4. Other examples are when evaluating new of modified insulation materials. A manufacturer of Kraft papers for bushings was evaluating different modifications in order to lower the DF. The new material had significantly lower DF at 20 °C, however, DFR measurement showed increase loss at lower frequencies and consequently, the 50 Hz DF was higher at 90°C compared to the original bushing paper. AG033573AE ZP-AG02E IDAX Monitoring 19 Your “One Stop” Source for all your electrical test equipment needs ▪▪Battery Test Equipment ▪▪Cable Fault Locating Equipment ▪▪Circuit Breaker Test Equipment ▪▪Data Communications Test Equipment ▪▪Fiber Optic Test Equipment ▪▪Ground Resistance Test Equipment ▪▪Insulation Power Factor (C&DF) Test Equipment ▪▪Insulation Resistance Test Equipment ▪▪Line Testing Equipment Megger is a leading global manufacturer and supplier of test and measurement instruments used within the electric power, building wiring and telecommunication industries. With research, engineering and manufacturing facilities in the USA, UK, Germany and Sweden, combined with sales and technical support in most countries, Megger is uniquely placed to meet the needs of its customers worldwide. Megger is certified according to ISO 9001 and 14001. Megger is a registered trademark. ▪▪Low Resistance Ohmmeters ▪▪Motor & Phase Rotation Test Equipment ▪▪Multimeters ▪▪Oil Test Equipment ▪▪Portable Appliance & Tool Testers ▪▪Power Quality Instruments ▪▪Recloser Test Equipment ▪▪Relay Test Equipment ▪▪T1 Network Test Equipment ▪▪Tachometers & Speed Measuring Instruments ▪▪TDR Test Equipment ▪▪Transformer Test Equipment ▪▪Transmission Impairment Test Equipment ▪▪Watthour Meter Test Equipment ▪▪STATES® Terminal Blocks & Test Switches ▪▪Professional Hands-On Technical and Safety Training Programs T +46 8 510 195 00 F +46 8 510 195 95 ▪▪AUSTRALIA ▪▪BULGARIA ▪▪CANADA ▪▪CZECH REPUBLIC ▪▪CHINA ▪▪FRANCE ▪▪GERMANY ▪▪HUNGARY ▪▪INDIA ▪▪INDONESIA ▪▪KINGDOM OF BAHRAIN ▪▪KOREA ▪▪MALAYSIA ▪▪PAKISTAN ▪▪PHILIPPINES ▪▪POLAND ▪▪ROMANIA ▪▪RUSSIA ▪▪SINGAPORE ▪▪SLOVAK REPUBLIC ▪▪SOUTH AFRICA ▪▪SPAIN ▪▪SWEDEN ▪▪SWITZERLAND ▪▪TAIWAN ▪▪THAILAND ▪▪UNITED ARAB EMIRATES ▪▪USA ▪▪VIETNAM WWW.MEGGER.COM Postal address: Megger Sweden AB Box 724 SE-182 17 DANDERYD SWEDEN Megger Group Limited UNITED KINGDOM Dover, Kent CT17 9EN ENGLAND Visiting address: Megger Sweden AB Rinkebyvägen 19 SE-182 36 DANDERYD SWEDEN seinfo@megger.com www.megger.com Subject to change without notice. Printed matter No. ZP-AG02E Doc. AG033573AE V01a 2017
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.7 Linearized : No XMP Toolkit : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:08:04 Create Date : 2017:09:05 13:45:49+02:00 Metadata Date : 2017:09:08 16:00:46+02:00 Modify Date : 2017:09:08 16:00:46+02:00 Creator Tool : Adobe InDesign CS6 (Windows) Instance ID : uuid:ec0c5320-4c75-47bb-9176-41d15198cd35 Original Document ID : xmp.did:BD9802DA5C87E411A41ED7CCE3D136AD Document ID : xmp.id:CBD84CBB2F92E71185C7901C5157D6CA Rendition Class : proof:pdf Derived From Instance ID : xmp.iid:CAD84CBB2F92E71185C7901C5157D6CA Derived From Document ID : xmp.did:07DCD661D3E5E511A5098DC8AFBFD6C6 Derived From Original Document ID: xmp.did:BD9802DA5C87E411A41ED7CCE3D136AD Derived From Rendition Class : default History Action : converted History Parameters : from application/x-indesign to application/pdf History Software Agent : Adobe InDesign CS6 (Windows) History Changed : / History When : 2017:09:05 13:45:49+02:00 Format : application/pdf Producer : Adobe PDF Library 10.0.1 Trapped : False Page Count : 20 Creator : Adobe InDesign CS6 (Windows)
EXIF Metadata provided by EXIF.tools
User Manuals, Guides and Specifications for your Megger IDAX-300 Measuring Instruments. Database contains 1 Megger IDAX-300 Manuals (available for free online viewing or downloading in PDF): Operation & user’s manual .
© 2009 Megger AB
IDAX User’s Manual
IDAX User’s Manual2
© 2009 Megger AB
Table of ContentsForeword 0
Part I IDAX 3
………………………………………………………………………………………………………………….. 31 Overview
………………………………………………………………………………………………………………….. 72 Instrument Panels
………………………………………………………………………………………………………………….. 93 Software Installation
………………………………………………………………………………………………………………….. 94 Measurement Technique
………………………………………………………………………………………………………………….. 125 Getting Started with IDAX
………………………………………………………………………………………………………………….. 216 IDAX System Control
Part II Results User Interface 25
………………………………………………………………………………………………………………….. 261 Results Menus and Commands
………………………………………………………………………………………………………………….. 632 Test Browser
Part III IDAX Error Messages 66
………………………………………………………………………………………………………………….. 701 (347) Output voltage is not within specified limits
………………………………………………………………………………………………………………….. 702 (361) Overvoltage
………………………………………………………………………………………………………………….. 703 (364) Measured capacitances don’t match
………………………………………………………………………………………………………………….. 704 (365) Specimen capacitance below limit
………………………………………………………………………………………………………………….. 715 (366) Specimen capacitance above limit
………………………………………………………………………………………………………………….. 716 (367) Measured DC current > MaxDCCurrent
………………………………………………………………………………………………………………….. 717 (368) Measured hum current > MaxHumCurrent
Index 72
IDAX 3
© 2009 Megger AB
1 IDAX
1.1 Overview
IDAX is an insulation diagnostic system for investigations/analysis of dielectric materials,normally insulators. The measurement method used is dielectric spectroscopy, i.e. measurementof the dielectric properties of the material as a function of the frequency, and in some casesalso as a function of the voltage. By studying the dielectric material properties as functions offrequency it is possible to make a distinction between different types of phenomena. Forexample, it is possible to separate polarization loss from leakage currents.
The system applies a sinusoidal voltage with desired frequency over the sample. This voltagewill generate a current in the sample. By accurately measuring the voltage and the current, thesample impedance can be calculated. Depending upon the actual sample model, variousparameters can be calculated from the impedance, such as capacitance, loss, resistance etc. Thisprocedure can then be repeated at the specified frequencies and voltage levels and even moreinformation on the sample can be revealed. More information can be found in the chapter»Measurement technique».
The system is fully computerized and most of the measurements can be performedautomatically.
More technical and safety details are given in sections:SpecificationsSystem and accessoriesSafety precautions
1.1.1 Specifications
The Insulation Diagnostic System IDAX measures the insulation impedance, at differentfrequencies, for insulation materials present in high and medium voltage transmission anddistribution networks. The measurement results are used for insulation diagnosis.
The IDAX uses voltages up to 200 Vpeak (~140 V RMS). The voltage range can be increased byadding an external high voltage unit (the high voltage unit is specified separately).
1.1.1.1 IDAX-300 Specifications
GeneralThe Insulation Diagnostic System IDAX-300 measures the insulation impedance, at differentfrequencies, for insulation materials present in high and medium voltage transmission anddistribution networks. The measurement results are used for insulation diagnosis.
The IDAX-300 uses voltages up to 200 Vpeak (~140 V RMS). The voltage range can be increasedby adding an external high voltage unit (the high voltage unit is specified separately). TheIDAX-300 is operated using an external laptop/PC.
The IDAX-300 uses a three electrode set-up, which allows for measurements of non-groundedas well as grounded objects with and without guard.
The IDAX-300 system applies sinusoidal shaped voltages at different frequencies across the testobject and simultaneously measures the current through it. The output voltage is either takenfrom a 10 Vpeak arbitrary waveform generator or from a 200 Vpeak arbitrary waveformgenerator.
IDAX User’s Manual4
© 2009 Megger AB
OutputVoltage/current range,10V:
0 — 10 Vpeak, 0 — 50 mApeak
Voltage/current range,200V:
0 — 200 Vpeak, 0 — 50 mApeak
Frequency range: 0.1 mHz — 10 kHz
MeasurementsInputs: Channel 1, channel 2, groundCapacitance Range: 10 pF — 100 uFAccuracy: 0.5% + 1 pFDissipation Factor Range: 0 — 10 (with retained accuracy of capacitance; otherwise higher)Accuracy: >1% +0.0003, 1 mHz — 100 Hz, C > 1 nF
>2% +0.0005, 100 Hz — 1 kHz, C > 1 nFNoise Level: Max 500 uA at 50 Hz/60 HzTest Modes, 2 Channels: UST-1, UST-2, UST-1+2, GST, GST-Guard-1, GST-Guard-2,
GST-Guard-1+2With 2-ch measurement option additionally UST-1+UST-2,UST-1+GST-Guard-2, UST-2+GST-Guard-1, UST-1+2+GST
Calibration: Calibration box allows field calibration, recommended interval 2years
GeneralMains Input (nominal): 90 — 265 V ac, 50/60 HzPower Consumption(max):
250 VA
Communication Ports: USB 2.0 and LAN
PhysicalInstrument Weight: 4.9 kg/11 lbsCase and InstrumentWeight:
9.9 kg/22 lbs
Accessories Weight: 8.5 kg/18 lbs (soft bag)Dimensions: 335 x 300 x 99 mm
17.7 x 6.3 x 16.1 inDimensions with Case: 520 x 430 x 220 mm
20.5 x 17 x 8.7 in.
EnvironmentalOperating Ambient Temp: 0°C to +50°C /-32°F to +122°FStorage Ambient Temp: -40°C to 70°C / -40°F to +158°FHumidity: 20% — 95% RH, non-condensingCE Standards: IEC61010 (LVD) EN61326 (EMC)
PC RequirementsOperating System: Windows 2000/ XP / VistaCPU/RAM: Pentium 500 MHz/512 Mb or betterInterface: USB 2.0
IDAX 5
© 2009 Megger AB
1.1.2 System and Accessories
1.1.2.1 IDAX-300 System and Accessories
IDAX-300 Transport case
Cables
IDAX User’s Manual6
© 2009 Megger AB
1.1.3 Safety Precautions
A » Warning» statement denotes hazardous conditions and actions that can cause bodilyharm or death. A «Caution!» statement denotes conditions and actions that can damage the IDAX or theequipment being tested.
Warning!
Always follow local safety regulations when working on high voltage equipment.
Make certain all personnel working with IDAX have been trained in its correct useand that all applicable safety precautions are taken.
BEFORE connecting this unit to the mains outlet user must verify that only a singleprotective ground systems exists AND no measurable voltage potential existsbetween the mains outlet protective ground and the test object protective ground. Ifa voltage potential is found between mains outlet protective ground and the testobject protective ground then additional measures described in local safety standardsshould be applied to ensure personal safety.
The measurement system can generate hazardous and even lethal voltages.
Do not attempt to service the IDAX yourself as opening or removing covers mayexpose you to dangerous voltage.
Do not use any accessories that are not intended for use together with the IDAX.
Unplug the IDAX from the mains supply when it is left unattended or not in use.
Before cleaning, unplug the IDAX from the mains supply. Use a damp cloth forcleaning. Do not use liquid or aerosol cleaners.
Caution!
Make sure that the mains voltage selected on the selector switch located on the backpanel corresponds to the voltage level before connecting the mains.
Refer all servicing to qualified service personnel.
If you need to return your IDAX, please use either the original crate or one ofequivalent strength.
IDAX 7
© 2009 Megger AB
1.2 Instrument Panels
1.2.1 IDAX-300 Front Panel
Overview
Powera) Power ON/OFF switch
Turns unit on and off.b) Mains connector
For connecting to mains outlet.c) Fuse
Use small screwdriver to gently pry outto change the fuse, F1, 2A slow blow.
Interfaced) USB Port
for connecting computere) ETHERNET
RJ45 connector for Ethernet connectionvia twisted pair cable.
IDAX User’s Manual8
© 2009 Megger AB
Outputf) Enable switch
1 for activating the output of voltage.0 for deactivating the output of voltage.
g) GeneratorGenerator output connector.
h) ActiveRed high intensity LED. Is lit when theoutput is switched on, dark when theoutput is switched off.
Externali) AMPLIFIER
Connector for external amplifier
Inputj) Input 1 and Sense 1
First input channels.k) Input 2 and Sense 2
Second input channels.l) Protective ground
Connector for connection to stationground.
m) TestTest inputs.
IDAX 9
© 2009 Megger AB
1.3 Software Installation
The IDAX-206 come with the measurement and display software already installed on theinternal computer. On the IDAX-300 and if you want to use the IDAX-206 with an external PCyou have to install the software.
To start installing the software insert the USB memory stick and run the program IDAXInstallXXXX.exe from it.
When first starting the IDAX software after installing it make sure the IDAX206/300 isconnected and turned on. For the IDAX-206 also check that the switch in the rear panel is set to EXT USB. When the software starts it will look for the IDAX hardware and connect to itautomatically.
1.4 Measurement Technique
Short overview of measurement principle is given in this section.
How impedance is measured Gives basic principle of IDAX operationSample modeling Describes theoretical models used for presenting measurement
results
1.4.1 How Impedance is Measured
IDAX measures impedance. By measuring the impedance at one point, i,e, at a specificfrequency and amplitude, parameters such as resistance, capacitance and loss can be calculated. The impedance of a sample is measured by applying a voltage across the sample. This voltagewill generate a current through the sample. By accurately measuring the voltage and thecurrent, the impedance can be calculated, see illustration below.
AV
ZU
IElectrometerVoltmeter
Computer withDSP-board
Voltage source
Sample
Measured voltage
Measured current
Control voltage
Fig.1. Measurement of electrical impedance.
The impedance is calculated using Ohm’s law:
where Z, U and I are complex entities.
The voltage (and the current) is generated by a voltage source. There are currently two internalvoltage sources available in the IDAX system, which can deliver a maximum peak output of10 Vpeak and 200 Vpeak, respectively. The voltage is measured by means of a voltmeter and
IDAX User’s Manual10
© 2009 Megger AB
the current is measured by an ammeter or electrometer which acts as a current-to-voltageconverter. The analogue signals (voltages) are then converted to digital samples of the signalsthat are used in subsequent calculations.
1.4.2 Sample Modelling
The impedance, Z, can be presented directly or by using different impedance models. Two waysof presenting Z directly are the polar and the rectangular, as follows:
Polar:
Rectangular:
Two simple models which are usually used in circuit analysis, although more seldom ininsulation analysis, are capacitance, C, and resistance, R. The equivalent RC circuit modelsavailable are series and parallel models (see Fig.1) calculated as follows:
a) b)
Fig.1. Equivalent circuit models: a) series RC circuit, b) parallel RC circuit.
Parallel:
Series:
where w = 2pf and f is frequency.
Another model, more often used in insulation diagnostics, is the complex capacitance modeldescribing the insulation impedance as a complex capacitance, where the imaginary part of thecapacitance represents the losses. The complex capacitance model is defined as follows:
where
IDAX 11
© 2009 Megger AB
Complex C:
The DC’ is defined as the capacitance, C’, with an arbitrary constant k (usually negative) added.The aim of this parameter is to make it possible to distinguish between small changes incapacitance in graphical presentation.
A model, very often used in insulation diagnostics, is a description of the insulation impedanceas a capacitance combined with a dissipation factor, tand, or a power factor (PF or cosj). Thecapacitance, tand and cosj/PF are defined as follows:
C, PF, Tan-Delta:
If tand and cosj (PF) are small, then tand » cosj = PF (E.g. tand=0.1 corresponds to PF=0.0995)
Insulation diagnostics is based on material characterization and therefore material models areoften used. To be able to define material parameters from measured impedance Z thegeometry of the sample, described in terms of the geometrical capacitance C0, has to bedefined. In the illustration below, a vacuum (or air-filled) capacitor of defined geometry isshown. Since no «material» is between the electrodes, the capacitance of a) is the geometricalcapacitance.
I
U C0
I
U C
I
U Ce e’& ’’ e s,r’&
a) b) c)Material parameter models based on a geometrical capacitance C0 and material parameters.
In the above illustration b) and c) a material is inserted between the electrodes and it willinfluence the current, I, flowing in the circuit. The influence of the material can be described bydifferent parameters using either a dielectric model or a conductive model. In the dielectricmodel the «material capacitance», the permittivity, is a complex function describing both thecapacitance and the loss. Whereas in the conductive model the capacitance is described by apermittivity and the loss by a conductivity (or resistivity). The dielectric and resistive models arederived as follows:
IDAX User’s Manual12
© 2009 Megger AB
Dielectric:
Resistive:
If geometrical capacitance, C0, is unknown it can be set by the user by entering a permittivity
(dielectric constant), e’, for the material. Entering a permittivity, e’, will let the system calculatean approximate C0 making use of the material models available. Yet, one must be aware thatthe accuracy of the absolute values are limited by the accuracy of the entered permittivity. Ifthe geometrical capacitance is unknown and an approximate permittivity is given, thegeometrical capacitance is calculated as follows:
The impedance value, Z, used in the calculation is the first measurement point in the actualmeasurement.
1.5 Getting Started with IDAX
This describes how to perform a diagnostic measurement on a specific test specimen. Moredetails about the software functions can be found in IDAX System Control and IDAX Results. It is recommended to start using IDAX by going through the procedure in this section in orderto become familiar with the instrument and the software. This procedure is also a good test toperform whenever a confirmation of the proper functioning of the instrument is required. Thistest is executed using the standard equipment delivered with IDAX.
Before starting to use IDAX certain preparatory procedures must be followed.
IDAX 13
© 2009 Megger AB
Preparatory Procedure: IDAX instrument1. Before connecting IDAX instrument power cord to the mains outlet:§ read the safety precautions§ connect the transparent ground lead to a protective earth (ground) close to measurement
specimen and to panel of instrument.
2. Connect IDAX instrument to a mains outlet which meets the requirements in User’s Manualand local safety regulations, and switch on IDAX instrument using the mains switch POWERon the front panel.
Preparatory Procedure: Specimen1. All specimen capacitances must be discharged before connecting IDAX leads. This means that
all specimen electrodes must be short circuited and grounded. The electrodes must remaingrounded until IDAX leads are connected. Below a general picture of ground connections isshown.
2. In the case of Test Capacitor, keeping the electrodes grounded while connecting IDAX leadsis complicated. Therefore, in this particular case with Test Capacitor, it is permitted to removethe ground connections before connecting IDAX.
1.5.1 Measurement
1. Make sure that IDAX is properly connected to mains outlet.
2. Turn on IDAX using the mains switch POWER on the front panel. The IDAX performs aself-test and start-up procedure.
3. Activate the Results window if not active after start-up.
4. In Results create new object selecting Object from menu File / New. Click «Browse» andselect !!Test Capacitor.obj template and enter arbitrary name as Object Name, for example»Test» and click OK.
5. By default all measurement templates listed as Associated Measurement Templates will beavailable for performing the measurements. It is possible to deselect one or moremeasurement templates from Associated Measurement Templates. As shown in picturebelow, four Measurement Templates !!C10 (GST-Guard), !!C12 (UST), !!C12, C10, C10+C12and !!C20 (GST-Guard) are associated with the object «Test» for IDAX-206, similar procedureas IDAX-300.
6. Fill in available fields with relevant information.
7. Create new measurement sequence selecting Measurement Sequence from menu File /New. Arbitrary name can be entered as Measurement Sequence, for example «GettingStarted» and press OK. New measurement sequence with corresponding Measurement Planfiles will appear in Test Browser.
IDAX User’s Manual14
© 2009 Megger AB
8. Fill in available fields with relevant information.
1.5.1.1 Test Sequence for IDAX-300
This is Getting Started test sequence when using IDAX-300
The specimen to be measured using the built-in capacitance box built into the system thatterminates at the front panel. The respective capacitances are:
C10 = ~2.5 nFC12 = ~4.7 nF
Warning!
The measurement system can generate hazardous and even lethal voltages.Carefully read «Safety precautions».
Ungrounded Specimen Test (UST) Measurement
IDAX 15
© 2009 Megger AB
In Test Browser select Measurement Plan file !!C12 (UST), carefully read and follow theinstructions. Or follow the instructions below.
1. Make sure that preparation procedures (both IDAX and Specimen) have been followed.
2. Connect the output to the lower capacitance connector.
3. Connect Current 1 and Sense 1 to the top capacitance connector.
4. Remove possible ground connections from somewhere.
5. Turn on the output switch activate the output and then start the measurement.
6. Wait until measurement is finished or abort earlier by stopping the measurement.
Grounded Specimen Test (GST) Measurement
IDAX User’s Manual16
© 2009 Megger AB
In Test Browser select Measurement Plan file !!C10 (GST-Guard), carefully read and follow theinstructions. Alternatively, follow the instructions below.
1. Make sure that preparation procedures have been followed (see also «Ungrounded SpecimenTest (UST) Measurement»).
2. Connect the generator output to the upper capacitance connector.
3. Connect Current 1 and Sense 1 to the lower capacitance connector.
4. Remove possible ground connections from the connection pins.
5. Turn on the output switch activate the output and then start the measurement.
6. Wait until measurement is finished or abort earlier by stopping the measurement.
1.5.1.2 Comparing 1-ch IDAX-206 and 2-ch IDAX-300
IDAX-300 (two winding transformer)
Test no. Measure Configuration Energize(Gen,Yellow)
Channel 1(Red)
Channel 2(blue)
1 CH GST-Guard-1+2 H L NC
2 CHL UST-1 H L NC
3 CL GST-Guard-1+2 L H NC
IDAX-206 (two winding transformer)
Test no. Measure Configuration Energize(Hi, Red)
Channel 1(Lo, Blue)
1 CH GST-Guard H L
2 CHL UST H L
3 CL GST-Guard L H
IDAX-300 (three winding transformer)
Test no. Measure Configuration Energize Channel 1 Channel 2
IDAX 17
© 2009 Megger AB
(Gen,Yellow)
(Red) (blue)
1 CH GST-Guard-1+2 H L T
2 CHL UST-1 H L T
3 CL GST-Guard-1+2 L H T
4 CLT UST-2 L H T
5 CT GST-Guard-1+2 T H L
6 CTH UST-1 T H L
IDAX-206 (three winding transformer)
Test no. Measure Configuration
Energize(Hi, Red)
Channel 1(Lo, Blue)
1 CH GST-Guard H L T shorted to L
2 CHL UST H L T shorted to tank
3 CL GST-Guard L H T shorted to H
4 CLT UST L H T shorted to tank
5 CT GST-Guard T H L shorted to H
6 CTH UST T H L shorted to tank
In case of a two winding transformer the set-up of IDAX-206 (IDA 200) and IDAX-300 are verysimilar; the leads need to be moved between test 2 and test 3.
In case of a three winding transformer the IDAX-300 second input channel is utilized (standard);the leads need to be moved between test 2 and test 3, between test 4 and test 5. If useIDAX-206 (IDA 200) on a three winding transformer, the leads need to be moved between everytest.
If using IDAX-300 with two electrometers (option), the two set-ups using same cable set-up canbe measured simultaneously, e.g. test 1 and test 2 (test 3 and test4; test 5 and test 6).
Note that Test 6, CTH (Energize T and measure H), in many cases are replaced by CHT (EnergizeH and Measure T):· IDAX-300: CHT, UST-2, H, L, T (same cable set-up as Test 1 and Test 2)· IDAX-206: CHT, UST, H, T, L shorted to tank
1.5.2 During Measurement
When the measurement has been started the measurement file icon changes to an iconcorresponding to the data file.
1. By pressing <F4> on the keyboard, Results window can be activated if not already active andthe measurement progress can be viewed when selecting Graph tab. The graph shows thedissipation factor for the Test Capacitor in the frequency range 0.1 — 1000 Hz .
IDAX User’s Manual18
© 2009 Megger AB
2. By pressing <F5> on the keyboard, an oscilloscope will be displayed on the screen. The Ch 0(red) signal in the oscilloscope displays the applied voltage over the sample, Ch 1 (white) isthe output voltage from the first electrometer and Ch 2 (blue) is the output voltage from thesecond electrometer. The output voltage from the electrometer is a function of the currentthrough the sample.
3. The instrument stops automatically after the last measurement point is completed, however,the measurement can be interrupted earlier by a stop command. You can press stop buttonor, on the IDAX-206, the OUTPUT OFF button. On the IDAX-206 a lit green READY LED onthe front panel indicates that the measurement is stopped or finished.
IDAX 19
© 2009 Megger AB
1.5.3 Presenting Results
By default the measurement data is presented using a model defined in the measurement plan.However, additional models can be added while inspecting the results (see Selecting Models fordetails).
Depending on which parameters are of interest to the user, the measurement data can bepresented using different models. E.g. in the tand model the dissipation and/or power factor aswell as capacitance can be displayed in the graph as a function of frequency. See also how toselect models in sections «Configuration / Models» and «Sample Modelling». The results can be viewed using three additional tabs: Graph, Table and 50/60 Hz data asshown in pictures below.
Graph tab is active
IDAX User’s Manual20
© 2009 Megger AB
Table tab is active
50/60 Hz Data tab is active
1.5.4 Selecting Models
1. Select Configuration / Models in Results window and all available models will be presentedas shown below. In this picture, both the «Dielectric» and «C, PF, Tan-delta» models areticked. Select the desired models and click OK.
IDAX 21
© 2009 Megger AB
2. In the Graph tab you can select the desired parameters. For example, the dielectricparameter e’.
3. In Results window menu the choices File / Export, File / Print and File / Report allow forexporting the measurement data and creation of reports.
All measurements are stored on the IDAX 206 system hard disk or the measurement computer’shard disk on an IDAX-300 system and can be opened and analysed at any time. For comparisonpurposes, many measurements can be loaded into the same Results window.
1.6 IDAX System Control
The IDAX System Control performs a measurement according to commands pre-arranged in aMeasurement plan or command file (C-file), using programming syntax described in IDAX Command and Variable References. The Measurement plan used is available in Results tabMeasurement plan.
IDAX User’s Manual22
© 2009 Megger AB
In addition to the IDAX System Control program window, a few more windows are accessiblefrom here.
Window FunctionsResults Editing of measurement plans and displaying the results.
Window is activated after up-start by defaultComments Adding comments to the actual measurement.Messages Displaying messages from the software during a
measurement.Oscilloscope Viewing the actual voltage and current curve forms.
1.6.1 Message Window
Messages from IDAX system to the user will be displayed in the Messages window. For example,if a measurement was aborted prematurely by the program, the reason for it will be displayedhere.
1.6.2 Connect and Disconnect
In the File menu you have the choice to connect (if not connected) and disconnect (ifconnected) to the IDAX unit. If you are not connected and the software needs to connect to theIDAX unit it will be done automatically.
1.6.3 Oscilloscope
The Oscilloscope displays the signals present on channel 0 (white, Ch0) and channel 1 (red, Ch1)inputs of the analogue IO unit.
IDAX 23
© 2009 Megger AB
Ch0 and Ch1 and their check boxesThe Ch0 and Ch1 check boxes select the channels that are displayed. Trace 0, white, isproportional to the sample voltage. Trace 1, red, shows the electrometer output voltage, whichis a function of the input current.
Horizontal controlThe oscilloscope time base can be set either on auto scale or manual. On auto scale, the timebase is set from the generator signal from the DSP. In the manual mode, the time base is set bythe Time/div slide bar. The time above the slide bar refers to the time per division, the fulloscilloscope window width consists of ten divisions.
RollThe Roll check box can be used for slow signals that the oscilloscope may otherwise havedifficulties to accurately lock.
Vertical controlThe gain of the input channels can be adjusted with the two vertical slide bars next to the Gridcheck box at the oscilloscope window. The voltage range is displayed beside the Ch0 and Ch1check boxes. The offset of the displayed signals can be adjusted by placing the mouse pointerinside the oscilloscope window and dragging it up and down while holding the mouse button.
GridThe Grid check box displays or hides the grid.
1.6.4 Changing Language
The language in which IDAX user interface (IDAX software) is displayed can be changed. Thefollowing languages are available:§ English§ Swedish§ German§ French§ Spanish
Language is changed selecting menu Configuration / Change Language in IDAX SystemControl, which opens a file opening dialog. The selected language file, *.lng, is loaded and itwill be used after restarting the IDAX program. Each language has its own language file, *.lng,and all language files are located in the directory D:IDALanguage files.
IDAX User’s Manual24
© 2009 Megger AB
In addition to IDAX interface language, the language of Measurement Templates is changed ifa translation is available. The translation files are stored in directory D:IDALanguage files.
1.6.5 Test Leads
The Test Leads dialog is used to select and, possible edit the settings for different cable sets.The presets drop-down is used to select what cable set you are to use for a measurement. If thecable set you are planning to use is not included you can use the “Add” button to add a newpreset.
The settings for a cable set can be edited. What the different settings means is explained in thistable.
Generator Input 1 Input 2 Ground
R(Ohm)
Resistance of centreconductor
Resistance ofcentreconductor
Resistance ofcentreconductor
Resistance ofearth/groundconductor
L(Henry)
—— Inductance ofinput 1 cable
Inductance ofinput 1 cable
Inductance ofearth/groundcable
C12(Farad)
Capacitance betweencentre conductor andmiddle shield
Cap. betweenconductor andshield
Cap. betweenconductor andshield
——
C23(Farad)
Capacitance betweenmiddle shield and outershield
—— —— ——
1.6.6 About
In «About» window the software versions of installed IDAX system, its program componentsand information about the operating system are displayed.
IDAX 25
© 2009 Megger AB
1.6.7 Calibration
The IDAX system presents its measurement results derived from measured voltage and current.Therefore, the IDAX system should mainly be calibrated with voltage and current standards.
Calibration of the IDAX-206 system is performed using an external calibration box containingvoltage, current and impedance references. The system can be calibrated against the calibrationbox every year or every 6 months. On the other hand, if the system is used and stored outsidecertain temperature/moisture limits, more frequent calibration is recommended.
The system may also require more frequent calibration to compensate for exposure totemperature gradients during measurement work, or if the calibration temperature differs toomuch from the ambient temperature. Moreover, one may recommend more frequentcalibration if tighter accuracy specifications are to be met. Using the external calibration box forcalibration of the system makes it possible to always have a calibrated system which eliminatesthe need to send the whole system away for calibration. The only thing that needs to be sentaway is the external calibration box. It is recommended to calibrate the box every two years.
1.6.8 Calibration Procedure IDAX-300
It is recommended to calibrate the IDAX-300 at least once a year. There are two optionsregarding calibration of IDAX-300:§ The IDAX unit can be sent to Pax for calibration§ Calibration can be performed by the customer using the optional Calibration Set
The second option means that the IDAX instrument is available since it does not need to be sentaway for calibration. The calibration procedure is easy to perform using the optionalCalibration Set, which consists of:§ Calibration Box§ Calibration Software§ Cables and connectors§ User’s Manual
The only thing that needs to be sent away is the Calibration Box and calibration of theCalibration Box can be performed by any local accredited laboratory or testing facility handlingtest and measure instruments.
2 Results User Interface
Results program is the window where all data from the measurement are displayed in agraphical or text format. Earlier performed measurements can also be opened for comparisonpurposes. Moreover, a measurement plan for the measurement to be performed can be editedhere.
Interaction with the program is done via windows and toolbar menu, the functions aredescribed in «Results Menus and Commands»
Results window is divided into two sections:§ Test Browser (left-side)§ Measurement Plan and presentation of the measurement results in graphical and table
format (right-side)
IDAX User’s Manual26
© 2009 Megger AB
2.1 Results Menus and Commands
Windows menu
Menu DescriptionFile Menu Operations with measurement objects, measurement files,
etc.Edit Menu Editing and changing properties of opened measurements.Diagram Menu Zooming featuresView Menu Selection of presentation formConfiguration Menu Configuration of Results window settingsHelp Menu On-line Help
Icon menuIcon menu is designed for easier navigation and operations with Test Browser.
Open measurement
Close measurement
Save
Results User Interface 27
© 2009 Megger AB
New Measurement Sequence
Toggle between Show and Hide sweep
Show/Hide Test Browser
Switch between Measurement Sequence and Measurement Plan or data file
Go one Measurement Plan or data file Down in Measurement Sequence if there are morethan one
Go one Measurement Plan or data file UP in Measurement Sequence if there are morethan one
2.1.1 Results File Menu
Menu Command DescriptionNew Create new objects and Measurement SequencesAdd/Remove Templates Add new or remove existing Measurement templates
Open Measurement Open selected measurement(s)Close Measurement Close selected measurementsRemake Measurement Repeat a measurement
Save Save current template or Measurement PlanSave As… Save the current template under a new name.
Load D-file… Temporarily load a measurement file into ResultsImport Import measurement dataExport Export measurement data
Print Print short summary of measurementReport Create detailed report about measurementSend to… Send data to an external analysis program
Rename File/Folder Rename individual file, test sequence or object folderDelete File/Folder Delete individual file(s), test sequence(s) or object folder(s)
Exit Close Results program
2.1.1.1 File / New
Depending on selected file or folder in Test Browser user can create:
New Object Create new Object is created if the object to be measured is notamong those in Test Browser
New Test Sequence Create new Measurement Sequence for a selected Object
2.1.1.1.1 File / New / Object
New object is created if the object to be measured is not among those in Test Browser.
1. Select New / Object from File menu which opens New Object dialog
IDAX User’s Manual28
© 2009 Megger AB
2. Select appropriate object template by using Browse button and press Open
3. Enter name for the new object, for example, «Test capacitor.obj» and press OK.
Results User Interface 29
© 2009 Megger AB
4. Newly created object appears in Test Browser as shown below.
2.1.1.1.2 File / New / Measurement Template
1. From View menu select View / Browse Measurement Templates
2. From File menu select File / New / Measurement Template, enter a name, e.g. template.icf, for new template and press Save. New empty template will appear in Text View mode forediting. See also section «IDAX 206 Commands and Variables».
2.1.1.1.3 File / New / Object Template
1. From View menu select View / Browse Object Templates
2. From File menu select File / New / Object Template, enter a name for new template and
IDAX User’s Manual30
© 2009 Megger AB
press Save. New empty template will appear in Text View mode for editing. See also section»IDAX 206 Commands and Variables».
2.1.1.1.4 File / New / Measurement Sequence
New Measurement Sequence is created when:· new measurement is to be performed on new object· new measurement is to be performed on an existing object· individual measurement file(s) is imported
Note!Before creating new Measurement Sequence, at least one measurement templatemust be associated with the test object.
1. To create new test sequence:· select New / Measurement Sequence from File menu
or· right-click on chosen test object and from menu selecting New Measurement
Sequence (see picture below)
2. New Measurement Sequence will be automatically assigned a name in form of date as shownbelow (date has the format according to International Standard ISO 8601). The name forMeasurement Sequence (i.e. «First Test» as in picture below) is optional.
Results User Interface 31
© 2009 Megger AB
2.1.1.2 File / Add/Remove Templates
User can add or remove Measurement Templates, (C-files) for selected measurement objectusing this command. It invokes a dialog window where user can manipulate with availableMeasurement Templates (see picture below).
This dialog window can be also invoked by right mouse click on selected object and choosing Add/Remove Templates from the menu which appears as shown below.
IDAX User’s Manual32
© 2009 Megger AB
2.1.1.3 File / Open Measurement
One or more measurements can be opened by this command depending on selected item inTest Browser.
Selected item Data to be openedSingle data file All sweeps in selected data fileMeasurement Sequence All sweeps in all data files under selected Measurement SequenceObject All sweeps in all data files in all Measurement Sequences under
selected Object
2.1.1.4 File / Close Measurement
One or more measurements can be closed by this command depending on selected item in TestBrowser.
Selected item Data to be closedSingle data file All sweeps in selected data fileMeasurement Sequence All sweeps in all data files under selected Measurement SequenceObject All sweeps in all data files in all Measurement Sequences under
selected Object
2.1.1.5 File / Remake Measurement
In cases when user is dissatisfied with a certain measurement it is possible to repeat thismeasurement with the same settings using Remake Measurement as follows:
1. Connect IDAX to the test object if not connected.
2. Select the measurement file, (D-file) to be repeated (obligatory).
3. Choose Remake Measurement. It will create a Measurement Plan, (C-file) from the selecteddata file, i.e. all measurement settings from this data file are copied into the newMeasurement Plan, (C-file).
4. Make corrections in Measurement Plan if needed.
5. Press ON on IDAX front panel to start the measurement.
2.1.1.6 File / Save
Save any changes made to the current template or Measurement Plan.
2.1.1.7 File / Save As
1. Select template which is to be saved under new name.
2. Invoke Save As dialog from menu File / Save As…
Results User Interface 33
© 2009 Megger AB
3. Enter new name for the template and press OK.
Note: The Object Templates are saved in Object Templates directory and MeasurementTemplates under Measurement Templates directory
2.1.1.8 File / Load D-file
Shortcut keys: Ctrl+L
This command allows for temporary loading of single measurement files, D-files. Duringloading a Temporary Object and Temporary Sequence are created as shown in picture below.All subsequent data files are opened under the same Temporary Sequence.
The opened files can be then imported under an existing object by using File / Import File / ToStructure.
If Results program is opened as an independent program and the data files are temporaryopened, then closing Results program, user is asked if the opened files should be imported intothe existing data structure.
IDAX User’s Manual34
© 2009 Megger AB
2.1.1.9 File / Import
When sharing measurement data between different IDAX instruments or computers, Importfunction facilitates the process of insertion of individual measurements into existing datastructure. Data are imported in a fashion which corresponds that they can be appropriatelyimported back into the same or other data structure.
Depending on data to be imported type and structure the following options are available:
IDA Zip File Import archived measurement dataFile to Structure Import selected file into existing structureMeasurement Templates Import selected Measurement TemplatesObject Templates Import selected Object Templates
2.1.1.9.1 File / Import / IDA Zip File
Short-cut keys: Ctrl+Alt+I
Import the data in from archived (*.izf) format. Data are inserted among existingmeasurements preserving the structure of the archived file.
1. Choose File / Import / IDA Zip File to invoke regular Open dialog.
2. Select file to be imported and press OK.
2.1.1.9.2 File / Import / File to Structure
Command is used to import individual measurement files into existing data structure, i.e. itrequires an Object and a Measurement Sequence. New Object and Measurement Sequenceneed to be defined in case there is no object and/or sequence to import the data.
1. From menu File / Import / File to Structure invoke import dialog.
2. Select the directory where measurement(s) to be imported are located.
3. Select Object under which the measurement must be imported.
Results User Interface 35
© 2009 Megger AB
4. Select the Measurement Sequence.
5. Select measurement(s) to be imported. By default all measurements located in givendirectory are imported. Deselect those which not to import by unchecking appropriate checkboxes.
IDAX User’s Manual36
© 2009 Megger AB
6. Press OK button and the measurement will appear in the Test Browser.
2.1.1.9.3 File / Import / Measurement Templates
Measurement Templates can be imported from a selected location on the IDAX hard drive oruser’s computer. It is done by invoking a dialog window shown below from menu File / Import/ Measurement Templates:
1. Choose directory where template(s) to be imported are located by typing in the pathmanually or using Browse button.
2. Select template(s) to be imported from available templates in the left hand side.
Results User Interface 37
© 2009 Megger AB
3. Press Add button or double click by left mouse button on chosen template.
4. Press OK button.
2.1.1.9.4 File / Import / Object Templates
Object Templates can be imported from a selected location on the IDAX hard drive or user’scomputer. It is done by invoking a dialog window shown below from menu File / Import /Object Templates:
1. Choose directory where the template(s) to be imported are located by typing in the pathmanually or using Browse button.
2. Select the template(s) to be imported from available templates in the left hand side.
3. Press Add button or double click by left mouse button on chosen template.
IDAX User’s Manual38
© 2009 Megger AB
4. Press OK button.
2.1.1.10 File / Export
When sharing measurement data between different IDAX instruments or computers, Exportfunction facilitates the process of extraction of individual measurements or templates fromexisting data structure. Data are exported in such a fashion that they can be appropriatelyimported back into the same or other data structure.
Depending on objective the Export function can be used in differently:
IDA Zip File Export data structure to a single fileData Export tab-delimited data or graph into a fileMeasurement Templates Export selected Measurement TemplatesObject Templates Export selected Object Templates
2.1.1.10.1 File / Export / IDA Zip File
Short-cut keys: Ctrl+Alt+E
Export of the data in an archived (*.izf) format. By exporting data into an IDA Zip File, the datastructure is preserved. It is possible to export:
· individual measurements — Object and Measurement Sequence information are included.· Measurement Sequences — Object information and all measurements in this sequence are
included.· Objects — all Measurement Sequences with respective measurements are included.
2.1.1.10.2 File / Export / Data
Short-cut keys: Ctrl+E
The «Export» command invokes an export-dialog with two different file type options, export toa file as a graph or the measurement points:
· WMF (Windows metafile)· TXT (Tab delimited format)
If the measurement data is exported as a TXT-file one can also select appropriate Format andDecimal separator. The Format option allows user to remove the header of the file, whichmakes it easier to import the data into another program. Decimal numbers are written in thefile using chosen Decimal separator option — point or comma.
Results User Interface 39
© 2009 Megger AB
Moreover, it is possible to export also the measurement file itself by checking checkbox «Include D-File», which may be favourable when exporting to floppy disk.
2.1.1.10.3 Header
Export with header
IDAX User’s Manual40
© 2009 Megger AB
Export without header
2.1.1.10.4 Point/Comma
Point as a decimal separator
Comma as a decimal separator
Results User Interface 41
© 2009 Megger AB
2.1.1.10.5 File / Export / Measurement Templates
Measurement Templates can be exported to a selected location on the IDAX hard drive or user’scomputer. It is done by invoking a dialog window shown below from menu File / Export /Measurement Templates:
1. Select templates to be exported side from available templates in the left hand.
2. Press Add button or double click by left mouse button.
3. Choose destination directory typing in the path or using Browse button.
4. Press OK button.
To remove a template:
1. Select it from the right hand side from selected templates.
2. Press Remove button or double click by left mouse button.
IDAX User’s Manual42
© 2009 Megger AB
2.1.1.10.6 File / Export / ObjectTemplates
Object Templates can be exported to a selected location on the IDAX hard drive or user’scomputer. It is done by invoking a dialog window shown below from menu File / Export /Object Templates:
1. Select templates to be exported side from available templates in the left hand.
2. Press Add button or double click by left mouse button.
3. Choose destination directory typing in the path or using Browse button.
4. Press OK button.
To remove a template:
1. Select it from the right hand side from selected templates.
2. Press Remove button or double click by left mouse button.
2.1.1.11 File / Print
Shortcut keys: Ctrl+P
The Print command allows user to print out a simple report of the measurement results in atext format.
Graph, Table or 50/60 Hz Data viewPressing Print first invokes a special window for Print Setup, where user can compose theappearance of the report. By checking in given options one can print out a report consisting of:
· calculated values such as equivalent current, dissipated power, power factor andcapacitance at 10 kV and at 50/60 Hz,
· table containing the data, which are selected in the Results window,· graphical presentation of the measurement.
Results User Interface 43
© 2009 Megger AB
Test Plan viewWhen in Measurement Plan view and in Text View mode user can print test plan text. If notin Text View mode, the Print command is inactive.
2.1.1.12 File / Report
Shortcut keys: Ctrl+Alt+P
This command invokes a dialog window where two report generators are available.
Report Generator DescriptionList&Label Create report using special set of drivers (only option available
when creating report on IDAX)MS Word Create Report using features Microsoft Word text editor
(Requires MS Word application installed on computer)
2.1.1.12.1 List&Label Report
List&Label is a set of special drivers provided for designing, viewing and printing reports.
When invoking List&Label report generator, a dialog window allows for choosing betweenvarious options as shown in the picture below. If there are already prepared report file(s) theywill be listed in the left side of the dialog window and can be selected to work with.
IDAX User’s Manual44
© 2009 Megger AB
Menu functions DescriptionNew report file Create new report file. See also separate manual for
List&Label.Add report Add a report to the list of available reportsDelete Delete the selected report fileModify Modify in selected report file can be done. It means that
the layout, appearance, etc. can be changed. (See separatemanual for List&Label).
Preview Preview selected report file with the users dataPrint Print the report according to previously designed layout
from selected report file, for example, Two windingtransformer + bushings.lst
Close Close Report dialog
2.1.1.12.2 MS Word Report
MS Word report generator allows for designing, viewing and printing reports using MicrosoftWord features.
When invoking MS Word report generator, a dialog window allows for choosing betweenvarious options as shown in the picture below. If there are already prepared report file(s) theywill be listed in the left side of the dialog window and can be selected to work with.
Menu functions DescriptionNew report file Create new report file using MS WordAdd report Add a report to the list of available reportsDelete Delete the selected report fileModify Modify in selected report file can be done. It means that
the layout, appearance, etc. can be changed.Compile Preview selected report file with the users data. Printing is
done using MS Word print options.Close Close Report dialog
Results User Interface 45
© 2009 Megger AB
2.1.1.13 File / Send to
Shortcut keys: Ctrl+D
In addition to different data export options, the measurement data file, (D-file) can be sent toother programs directly for further analysis. For this purpose Results are configured byproviding search path to the external program. The configuration is made from menu Configuration / Program… where the appropriate executable file is selected in dialog box.
Now the data file loaded into Results can be sent to the chosen external program from File /Send to… as shown below. If the configuration step has been omitted, the dialog shown aboveappears and user is prompted to enter the search path at this stage.
When a executable application is selected, the name of the executable is added to the Sendto… menu as shown below, where MODS has been chosen as application where data are to besent.
IDAX User’s Manual46
© 2009 Megger AB
The data can also be send to selected application by right mouse click and selecting Send to ..MODS as shown in picture below.
2.1.1.14 File / Rename File/Folder
1. Select the file or folder to be renamed
2. Choose Rename File/Folder from menu or right-click on the file or folder
3. Type new file or folder name and press Enter
2.1.1.15 File / Delete File/Folder
1. Select the file or folder to be deleted
2. Choose Delete File/Folder from menu or right-click on the file or folder. A pop-up windowwill appear:
Results User Interface 47
© 2009 Megger AB
3. To confirm, select Yes and press Enter or left-click, otherwise select No.
2.1.1.16 File / Exit
Shortcut keys: Alt+F4
1. Closes Results program, without quitting IDAX System Control program. Results can berestored by pressing F4 or selecting from menu Windows / Results in IDAX System Controlwindow.
2. If Results were launched as stand-alone application — File / Exit quits Results program withoutsaving changes.
2.1.2 Results Edit Menu
Commands for handling text and objects
Menu Command DescriptionUndo Undo last operation
Copy Copy object(s) or text to clipboardCopy All Copy all selected objects or text to clipboardCut Cut object(s) or text to clipboardPaste Paste object(s) or text from clipboard
Select All Select all object(s) or text from Results window
Commands for handling graphical output
Menu Command DescriptionToggle Sweep Toggles between show and hide measurement sweepMark Sweep Shows measurement sweepUnmark Sweep Hides measurement sweep
Change Sweep Color Changes colour appearance of measurement sweep
2.1.2.1 Edit / Undo
Shortcut keys: Ctrl+Z
Reverse the effect of the last action performed, such as typing, deleting or formatting text, etc.
2.1.2.2 Edit / Copy
Shortcut keys: Ctrl+C
Place a duplicate of the selected text or object(s) onto the Clipboard where it can be pastedelsewhere.
IDAX User’s Manual48
© 2009 Megger AB
2.1.2.3 Edit / Copy All
Shortcut keys: Ctrl+Alt+C
GraphCopy Results window to Clipboard. i.e. make screen dump as using Alt+Print Screen.
TableCopy content of the table to Clipboard
50/60 Hz ValuesCopy measured values at 50/60 Hz to Clipboard.
Test PlanHighlight all text in the current text frame (Text View) and place a duplicate of it onto theClipboard where it can be pasted elsewhere.
2.1.2.4 Edit / Cut
Shortcut keys: Ctrl+X
Delete the selected text from Measurement Plan or Template and place it on the Clipboardwhere it can be pasted elsewhere.
2.1.2.5 Edit / Paste
Shortcut keys: Ctrl+V
Place the contents of the clipboard (the last text cut or copied) back into Test Plan.
2.1.2.6 Edit / Select All
Shortcut keys: Ctrl+A
Highlight all text in the current text frame (Text View) for subsequent copying.
2.1.2.7 Edit / Toggle Sweep
Toggles between Mark Sweep and Unmark Sweep, i.e. displays or hides the selectedsweep(s) in the graph, table and the values at power frequency. The effect of Toggle Sweepdepends on position in Test Browser.
Position in Test Browser Displayed sweeps Object All opened measurements under this object are displayed/hiddenMeasurement Sequence All opened measurements in this sequence are displayed/hiddenMeasurement file All sweeps in this file are displayed/hiddenA sweep selected Only selected sweep is displayed/hidden
2.1.2.8 Edit / Mark Sweep
Displays the selected sweep in the graph, table and shows values at power frequency. In orderto have an effect of Mark Sweep:
· the sweeps must be unmarked (hidden)· the sweeps to be displayed also depends on selected position in Test Browser.
Position in Test Browser Displayed sweeps Object All opened measurements under this object are displayedMeasurement sequence All opened measurements in this sequence are displayedMeasurement file All sweeps in this file are displayed
Results User Interface 49
© 2009 Megger AB
A sweep selected Only selected sweep is displayed
2.1.2.9 Edit / Unmark Sweep
Hides the selected sweep in the graph, table and the values at power frequency. In order tohave an effect of Unmark Sweep:
· the sweeps must be marked (displayed)· the sweeps to be hidden also depends on selected position in Test Browser.
Position in Test Browser Displayed sweeps Object All opened measurements under this object are hiddenMeasurement sequence All opened measurements in this sequence are hiddenMeasurement file All sweeps in this file are hiddenA sweep selected Only selected sweep is hidden
2.1.2.10 Edit / Change Sweep Color
Changes colour appearance of selected sweep. Change Sweep Color invokes a colour dialogwindow from which a desired colour for sweep can be selected. This command can also beinvoked right-click on selected sweep.
2.1.2.11 Edit / Read Only
Allows for changes in those fields which are read only by default:· object data when measurement sequence is selected· Measurement plan for data files (D-files)
By selecting Edit / Read Only this feature is removed and user can change the content of allfields.
2.1.3 Results Diagram Menu
Menu command DescriptionZoom in Zooming in into a selected areaZoom out Returning to full graph
2.1.3.1 Diagram / Zoom In
Shortcut keys: Ctrl+I
1. The Zoom In command allows user to select a rectangular area of the by dragging arectangle with the mouse.
IDAX User’s Manual50
© 2009 Megger AB
2. The marked area will the be magnified to cover the entire graph area as shown below.
Results User Interface 51
© 2009 Megger AB
2.1.3.2 Diagram / Zoom Out
Shortcut keys: Ctrl+O
Zoom out command fully displays all active curves similarly as turning on Autoscale in axissettings. The axis settings in the axis settings, however, remain unchanged.
2.1.4 Results View Menu
Menu Command DescriptionBrowser Show/Hide Test Browser.
Browse Object Tree Allow to select which items are displayed Test Browser.Browse Measurement Templates Show only Measurement TemplatesBrowse Object Templates Show only Object Templates
Information Display information about Object, Measurement Sequence,Measurement Plan etc.
Graph Show graphical representation of measurementsTable Show table view of the measurement results50/60 Hz Data Present the values at power frequency
Text View Open Measurement Plan or Template programmingcommands for its interface and measurement settings forreview and editing
Show Measurement Templates Show/Hide Measurement Templates
2.1.4.1 View / Browser
Toggles between hide and show Test Browser. It may be convenient to hide it in order tomaximise the area for graphical output as well as for table and when editing measurementplans.
2.1.4.2 View / Browse Object Tree
Displays all objects inclusive associated Measurement Templates, Measurement Sequences,Measurement Plans, (C-files) and measurement data files, (D-files).
2.1.4.3 View / Browse Measurement Templates
Displays all available Measurement Templates and it is possible to edit any of them as well ascreate new ones.
2.1.4.4 View / Browse ObjectTemplates
Displays all available Object Templates and it is possible to edit any of them as well as createnew ones.
2.1.4.5 View / Information
Measurements performed by IDAX are described in a command file, called Measurement Plan orC-file (*.icf). This text file contains all information specific for the measurement in question, and is inits entirety saved in the data file (*.idf) before the measurement data. Displayed content of Information tab depends on selected place in Test Browser. There areseveral possibilities:
· Object selected· Measurement Sequence selected
IDAX User’s Manual52
© 2009 Megger AB
· Measurement Template selected· Measurement Plan selected· Measurement Data selected
Information — ObjectDisplays information about object corresponding to the chosen Object Template. Objectinformation can be edited using also Text view mode. The changes made here will, however,be unique for the given object. Permanent changes must be made in Object Template(s).
Information — Measurement SequenceMeasurement Sequence displays information relevant to given measurement occasion.Measurement Sequence can be edited using also Text view mode. The changes made here will,however, be unique for the given Measurement Sequence only. Permanent changes must bemade in Object Template(s).
Information — Measurement Template1. Selected Measurement Template is associated with an object
Displays the appearance of Measurement Plan for specific conditions (see also MeasurementTemplates) and given object. Measurement Template(s) can be edited using also Text viewmode. The changes made here will, however, be unique for the given object only. Permanentchanges must be made in Measurement Template(s).
Results User Interface 53
© 2009 Megger AB
2. When using View / Browse Measurement TemplatesSimilar as above, but the changes made in Text view mode will be permanent and will haveeffect when creating next new object or adding given Measurement Template to existingobject(s).
Information — Measurement PlanMeasurement Plan displays the same information as Measurement Template and can be editedalso using Text view mode. In the Text view mode appearance of the Measurement Plan viewcan be specified, thus hiding irrelevant information from the user. The changes made here will,however, be unique for the given measurement. Permanent changes must be made inMeasurement Template(s).
Information — Measurement Data
IDAX User’s Manual54
© 2009 Megger AB
Measurement Data displays the same information as Measurement Plan. The data itself is notvisible, yet can be viewed in Text mode.
Caution: The data files are Read-only and even the editing is possible after disabling Read-onlyattribute, it is strongly discouraged. Any data loss or damage caused by editing of data files isuser’s responsibility.
The IDAX System Control reads the Measurement Plan and each line is interpreted insubsequent order. Normally, every variable or command is contained in a separate row.However, multiple variables and commands can be entered in the same row if they areseparated with semicolons. When the measurement is started, current Measurement Plan iscopied to a file LastC.icf in D:IDA catalogue, which is then read and executed by the IDAXSystem Control.
2.1.4.6 View / Graph
Shortcut keys: Ctrl+G
Displays graphical output of an ongoing measurement and/or previously made and openedmeasurements. It is possible to display graphical representation of the measurements usingdifferent parameters, i.e. loss tangent, power factor, PF, capacitance, C and others (see also Configuration / Models).
By clicking right mouse button on a measurement point the properties of this point are shownin Properties box in left lower corner of Results window.
Results User Interface 55
© 2009 Megger AB
2.1.4.7 View / Table
Shortcut keys: Ctrl+T
Ongoing measurement and/or previously made and opened measurements are displayed in atext mode. In addition to measured values loss tangent, power factor, PF, capacitance, C andothers, in Table mode also hum current, DC offset and object’s capacitance are shown forspecified configurations.
2.1.4.8 View / 50/60 Hz Data
Measured valuesWhen a frequency sweep is performed, the measurements are done at certain pre-definedfrequencies. Due to possible interferences at power frequency (50 or 60 Hz) it is generally
IDAX User’s Manual56
© 2009 Megger AB
recommended to avoid this frequency. Instead, two measurements near the power frequencyare performed and the values at power frequency are obtained by interpolation.
10 kV equivalentsThe 10 kV equivalents are values that are calculated from the measured values. Using lowvoltage, e.g. 200 Vpeak, will result in a certain current through the sample. A linear insulatingmaterial, such as oil and/or paper, will have the same dissipation factor and capacitance also ata 10 kV test voltage. Therefore, it is possible to calculate the current and power at 10 kV fromthe measured values.
2.1.4.9 View / Text View
Displays text behind the Measurement Plan or Template. In Text View all formatting andmeasurement settings are shown for user and editing is enabled. It is also possible to view datafiles (D-files), but their editing is disabled.
2.1.4.10 View / Show Measurement Templates
Shows or hides Measurement Templates associated to objects in Test Browser. Hiding templatessaves space in Test Browser, which may be recommended.
2.1.5 Results Configuration Menu
There are commands under this menu which allow user to configure output of measurementresults, display desired measured values in both graphical and text format as well as editmeasurement plan(s).
Menu command DescriptionInformation Configure appearance and Template and Measurement Plan
editing toolsGraph Configure appearance of graphical outputTable Configure Table view Power Frequency Data Display values of losses and capacitance at power (network)
frequency (50 or 60 Hz)
Send to Select an external program for data analysis
Models Display available models for presentation of results
2.1.5.1 Configuration / Information
Editing of Measurement and Object Templates and Measurement Plans in Text View mode canbe facilitated by means of this dialog window.
Results User Interface 57
© 2009 Megger AB
1. Checking/unchecking Gutter displays/hides line numbering in the left side.
2. Advanced help enables automatic appearance of alternatives for variables. When typing avariable name followed by «=» sign, the alternatives appear as a drop-down menu (seebelow) and user can select the appropriate one.
3. Highlighting settings allows user to change the color scheme for appearance of text editor(see example below).
4. Font… allows user to select the fonts to be used in Text view of Measurement or ObjectTemplates and Measurement Plan.
2.1.5.2 Configuration / Graph
Allows user to configure appearance of graphical output
Menu commands Description
IDAX User’s Manual58
© 2009 Megger AB
Display Title Toggle ON/OFF display titleAxis Configure Axis settingsBackground Change background color
2.1.5.2.1 Configuration / Graph / Display Title
Toggles ON/OFF the graph title.
2.1.5.2.2 Configuration / Graph / Axis
The Axis command invokes the axis dialog window. This window allows the names, scales, grid,prefixes, etc. on the axes to be modified. The dialog is divided into three parts:
§ Title§ X-axis§ Y-axis.
2.1.5.2.2.1 Autoscale
Autoscale automatically sets the scale so that all active curves are displayed in the graph.
Note!Negative values are not possible to display in logarithmic scale.
2.1.5.2.2.2 Grid
For grid settings, three options are available:
§ None, no grid is shown.§ Small, also called tics. Only an indication at the axis, no lines in the graph.§ Full, all lines are displayed in the graph.
Results User Interface 59
© 2009 Megger AB
2.1.5.2.2.3 Label
Allows user to enter a label text for both X- and Y- axis.
2.1.5.2.2.4 Log/Lin
Makes it possible to change between logarithmic and linear scale for the X and/or Y axis.
Note!Negative values are not possible to display in logarithmic scale.
2.1.5.2.2.5 Precision
Precision sets how many digits of the point is presented in the Graph/Table.
2.1.5.2.2.6 Prefix/Scientific
Changes between Prefix or Scientific notation in the Graph/Table.
2.1.5.2.2.7 Title
Here the title for the graph in Results can be written. If Auto is checked, the title of themeasurement will be displayed for the measurement. If more than one measurements are
IDAX User’s Manual60
© 2009 Megger AB
loaded in the results window, with different titles, no title will be displayed.
2.1.5.2.2.8 Unit
Units for both X- and Y- axis.
2.1.5.2.3 Configuration / Graph / Background
Shortcut keys: Ctrl+B
Using this command the background of the graph can be changed. The selected color is used ondisplay and in files exported in *.wmf format, however, when printing the background color isset to white.
2.1.5.3 Configuration / Table
Allows user to choose between displaying or hiding the disturbance data when in Table mode.
2.1.5.4 Configuration / Power Data
Configures display of desired values at power frequency (50 or 60 Hz). This command invokes adialog window shown below. The options are as follows:
1) showing either loss tangent (tand) or Power Factor (PF)
2) select power frequency§ 50 Hz, e.g. Europe, Australia, most part of Africa and Asia§ 60 Hz, e.g. USA, Canada, Mexico, certain countries in South America
3) showing interpolated 10 kV equivalents for current, dissipated power, Power Factor andcapacitance
Results User Interface 61
© 2009 Megger AB
2.1.5.5 Configuration / Program…
The dialog window, which appears when selecting this menu command, allow user to choosean external application for further data analysis. This is done by providing the search path tothis program as shown below.
2.1.5.6 Configuration / Models…
Shortcut keys: Ctrl+M
Command Configuration / Models… invokes a window where available models forpresentation of measurement results are shown. Different models can be selected and thecorresponding results in graphical or text format will be simultaneously displayed.
IDAX User’s Manual62
© 2009 Megger AB
Selected models appear at the bottom of Results window, where one can choose whichparameters are to be shown. Below the Dielectric and C, PF, Tan-Delta models are chosen to beused in Results and e¢ as a parameter is displayed in the graph.
2.1.6 Impedance models
The following impedance models and their parameters are available:
Results User Interface 63
© 2009 Megger AB
2.1.7 Results Help Menu
Menu command DescriptionContents Show content of all IDAX HelpResults Help Display help for Results onlyWhat’s this? Pop-up help for unclear fields or menu
2.1.7.1 Help / Contents
IDAX Help contents displays available Help topic:
Topic DescriptionIDAX Overview of IDAX operating principle and functionsResults User Interface Preparation of measurements and presentation of resultsCommand and Variable Reference Reference of programming directivesIDAX Error Messages Error messages generated by IDAX and possible
countermeasures
2.2 Test Browser
Test Browser is located at the left hand side of the Results window and it works similarly toMicrosoft Windows Explorer. The main purpose of the Test Browser is to facilitate:
§ navigation between already performed measurements, Object and MeasurementTemplates§ organisation of new measurements on the existing or new objects§ creation and editing Measurement and Object Templates
IDAX User’s Manual64
© 2009 Megger AB
2.2.1 Data Structure
Test Browser is arranged as a tree-like structure consisting of:§ Objects§ Measurement Sequences§ Measurement Plans and data files
Objects Each object is represented by an object template specified in a file Object.obj. This file containsthe information about the object which is constant such as owner, manufacturer, serial number,etc. There are several pre-defined Object Templates for power transformers, bushings andcables. Depending on user’s needs these files can also be modified as well as new templates canbe created.
When creating new objects it is recommended to select Measurement Templates which can beassociated to those objects.
Measurement SequenceMeasurement Sequence is created in order to make a measurement and therefore at least oneMeasurement Template must be associated with the Object.
Measurement structureIn order to perform a measurement a Measurement Sequence must be available or a new onemust be created. New sequence is created by selecting desired object and press File / New /Measurement Sequence. This will result in:
· A dialog will appear where a test date and an optional name of Measurement Sequencecan be defined
· All Measurement Templates associated with the object will be copied to the folder of newMeasurement Sequence (inclusive the object information) as command or C-files (*.icf).
· In order to run a measurement, a command file from those under Measurement Sequenceis chosen and IDAX is started by pressing ON.
· When measurement is started, the command file is replaced by a data file containing bothinformation from the command file and measurement data. Finished measurement can be
Results User Interface 65
© 2009 Megger AB
repeated by selecting data file and choosing File / Remake Measurement. This will createa new command file copy from data file, which is possible to start again. (The name ofcommand file will remain the same as original data file with additional number at the end,for example xxxxx-2 or yyyy-3, where number indicates times the measurement has beenrepeated.
· When measurement is finished the next command file in given Measurement Sequence canbe selected and run until the entire Measurement Sequence is done. Unwanted commandfiles can be deleted or simply left as they are.
2.2.2 Measurement Templates
Measurement Templates are templates which are recommended to use when performingvarious measurements on power and distribution transformers, bushings and cables. The formview of Measurement Template is pre-specified, however, can be modified to satisfy differentrequirements. For information about the commands in the IDAX C file see the Commands andVariable Reference in the program help file or included pdf manual.
In all Measurement Templates it is provided that user can choose to perform either:1) frequency sweep down to a specified stop frequency, or
2) measurement only at power frequency (50 or 60 Hz). This last option is not available for XLPEcable measurements and in limited extent in general measurement plans.
Two winding transformerAvailable Measurement Templates are:§ 2W-CH — measurement of insulation between high voltage winding and ground§ 2W-CHL — measurement of insulation between high and low voltage windings§ 2W-CL — measurement of insulation between low voltage winding and ground
Three winding transformerAvailable Measurement Templates are:§ 3W-CH — measurement of insulation between high voltage winding and ground§ 3W-CHL — measurement of insulation between high and low voltage windings§ 3W-CHT — measurement of insulation between high voltage and tertiary winding§ 3W-CLT — measurement of insulation between low voltage and tertiary winding§ 3W-CL — measurement of insulation between low voltage winding and ground§ 3W-CT — measurement of insulation between tertiary winding and ground
Autotransformer without tertiary windingAvailable Measurement Templates are:§ 2WA-CH — measurement of insulation between high/low voltage winding and ground
Autotransformer with tertiary windingAvailable Measurement Templates are:§ 3WA-CH — measurement of insulation between high/low voltage winding and ground§ 3WA-CHT — measurement of insulation between high/low voltage and tertiary windings§ 3WA-CT — measurement of insulation between tertiary winding and ground
BushingsAvailable Measurement Templates are (for each phase):§ C1, H1 — measurement of bushing H1* main insulation, C1§ C1, H2 — measurement of bushing H2 main insulation, C1§ C1, H3 — measurement of bushing H3 main insulation, C1§ C2, H1 — measurement of bushing H1 insulation between test tap and ground sleeve, C2§ C2, H2 — measurement of bushing H2 insulation between test tap and ground sleeve, C2§ C2, H3 — measurement of bushing H3 insulation between test tap and ground sleeve, C2
IDAX User’s Manual66
© 2009 Megger AB
* H1, H2 and H3 are default labels for bushings of different phases.
PILC Cables§ PILC Cable, A — measurement of insulation of phase A*§ PILC Cable, B — measurement of insulation of phase B§ PILC Cable, C — measurement of insulation of phase C
* A, B and C are default labels for different phases of the cable.
XLPE Cables§ XLPE Cable, A — measurement of insulation of phase A*§ XLPE Cable, B — measurement of insulation of phase B§ XLPE Cable, C — measurement of insulation of phase C
* A, B and C are default labels for different phases of the cable.
General measurementsAvailable Measurement Templates are:§ General (PF Data) — generic measurement plan for measuring only at power frequency (50/60
Hz)§ General (Sweep) — generic measurement plan for measuring of frequency sweep§ General HV — generic measurement plan for measurements using IDA High Voltage Unit
2.2.3 Object Templates
Object Templates are templates which are recommended to use when creating new objects.Available pre-specified templates are:
§ !!Test Capacitor § !General Specimen§ Autotransformer with Tertiary + Bushings§ Autotransformer with Tertiary§ Autotransformer without Tertiary + Bushings§ Autotransformer with Tertiary§ Bushings§ PILC (Paper Insulated Lead Covered) Cable§ Three Winding Transformer + Bushings§ Three Winding Transformer§ Two Winding Transformer + Bushings§ Two Winding Transformer§ XLPE Cable
The form view of Object Template is pre-specified, however, it can be modified to satisfydifferent requirements. For information about the commands in the IDAX C file see theCommands and Variable Reference in the program help file or included pdf manual.
3 IDAX Error Messages
System messages
No Message Explanation101 Filename missing File name not specified file dialog. Specify a file
name.103 Not string A string variable is expected.104 Could not open file: Wrong file name, file is missing or path is incorrect.105 Could not open or create
keyProblem accessing Windows Registry.
IDAX Error Messages 67
© 2009 Megger AB
No Message Explanation106 Key could not be deleted Problem accessing Windows Registry.107 Redeclaration of variable A variable is declared two or more times in the C–
file.108 Error writing to file The log file from a Calibration or a Self test could
not be created.109 Incorrect data type The Data type specified does not exist.112 Cannot start measurement Displayed when IDAX is not properly set up in order
to perform a measurement. Usually accompanied bya more detailed message.
113 Error reading Windowsregistry
Problem accessing Windows Registry.
Errors associated with C-file
No Message Explanation201 Not a complex number Variable is expected to be complex number in «X+Yi»
format.202 Illegal assignment Wrong type has been assigned to variable.203 Unknown directive Incorrect command. Most likely a typing error.204 No «Data type» specified Type must be specified before data field. Example: $
Type = ‘R’205 No «Category» specified Category must be specified before data field.
Example $ Cat = ‘C’206 Incorrect «Category» Wrong variable in category (Cat) has been used.208 Error on line Indicates the number of line in C-file, where error
was detected.209 «VoltageSource» incorrect «VoltageSource» has to be of a defined type. Most
likely a typing error.210 Boolean is expected for
check box («CB»)Check boxes can have only Boolean values.
211 Not an integer Parameter values must be of the same type as in theparameter declaration.
212 Not a real Parameter values must be of the same type as in theparameter declaration.
213 Not a Boolean Parameter values must be of the same type as in theparameter declaration.
214 Unknown «Data type» «Data type» must be of a defined type. Most likely atyping error.
216 Wrong number of arguments Too few or too many arguments in a variable.217 Syntax error Undefined error in the syntax. Usually accompanied
by additional information.218 To few radio buttons in
group («RB»)There must be at least two buttons in a radio group
219 Not a radio buttons item(«RB»)
The function is assigned to a value that does notmatch any radio button. Most likely a typing error.
220 Not an array of integers Parameter values have to be of the same type as inthe parameter declaration.
221 Invalid bitmap «Image» Image is not a *.bmp file or the file is corrupted.222 Variable not declared All new variables must be declared with a «New»
function.223 «AmplitudeScale» not
assigned»AmplitudeScale» must be assigned. See User’sManual for options.
224 «VoltageSource» notassigned
«VoltageSource» must be assigned. See User’sManual for options.
225 «FileType» not assigned «FileType» must be assigned. See User’s Manual foroptions.
226 «SpecVersion» not assigned «SpecVersion» must be assigned. See User’s Manualfor options
IDAX User’s Manual68
© 2009 Megger AB
No Message Explanation227 «DraftSpecVersion» not
assigned»DraftSpecVersion» must be assigned. See User’sManual for options.
228 «Configuration» notassigned
«Configuration» must be assigned if using aTermination Box with relays. See User’s Manual foroptions.
229 Value not assigned If cat ‘P’ is assigned, the field can not be left blank.
Errors related to IDAX 206 system and measurement settings
No Message Explanation301 Wrong argument in
«LinSweep» functionWrong number/type of argument. See User’s ManualAppendix A.6 for options
302 Wrong argument in»LogSweep» function
Wrong number/type of argument. See’s ManualAppendix A.6 for options.
303 Settings could not besaved
User settings could not be saved.
306 No DSP device name The name of the DSP instrument is not valid, contacta service supplier.
307 Could not open DSP The DSP instrument did not connect properly.308 No geometric capacitance309 Error initializing DSP.
DSP not present orinactive.
The DSP instrument did not connect properly.
311 Neither the file LastC.icfnor specified file couldbe found
The C-file is not found in the system. Probablyproblem with reading/writing to disc or Windowsregistry.
312 Unknown «C-File» command The specified command does not exist. Most likely atyping error.
313 Wrong number of arguments Too few or too many arguments in a command.314 No «Amplitude» «Amplitude» must be assigned. See User’s Manual
Appendix A.5 for options.315 No «Frequency» «Frequency» must be assigned. See User’s Manual
Appendix A.5 for options.319 No measurement to comment Comments are possible during a measurement or
directly after the measurement.320 No current measurement No measurement which has been started.321 DSP variables wrong The DSP-variables (e.g. MinTime, MinCycles, etc.) are
wrong.322 DSP code not found No or wrong DSP code found in the system.323 Index does not exist
in NT registryProblems accessing Windows Registry.
324 Could not interpolatecalibration data
Could not interpolate transfer functions.
325 No voltage source selected A voltage source must be selected before measuring.327 Could not get info from
DSPProblems with the connection to the DSP instrument.
329 «MeasurementPerPoints»must be greater than zero.
One measurement at each frequency must beperformed.
330 DSP reset failed Problems with the connection to the DSP instrument331 «Else» without «if» Syntax error in C-file.332 «End» without «if» Syntax error in C-file.333 Only variables or
constants are allowedOnly variables or constants are allowed in an if–elsestatement.
334 Types does not match Wrong type of data (e.g. string assigned to aninteger).
339 Error updating «D-File» Measured data was not properly added to the D-file.340 Invalid voltage The desired voltage is out of the specified limits.
Possible typing error in the C-file.341 Invalid current The applied settings results in a current that is too
IDAX Error Messages 69
© 2009 Megger AB
No Message Explanationhigh.
342 Invalid frequency The desired frequency is out of the specified limits.Possible typing error in the C-file.
343 Cable parameter not found: The specified parameter was not found in the samplecable.
344 Cable parameters not found No cable data was found.345 «Temination box» parameter
not found:The specified parameter was not found in the»Termination Box».
346 «Termination box»parameters not found
No «Termination Box» data were found.
347 Output voltage is notwithin specified levels.
The applied voltage differs from the desired.
351 «InitMeasureZ» settings: Not possible to perform InitMeasureZ measurement.This message is followed by one of 340-342.
352 No «Amplifier type»selected
High Voltage amplifier not selected.
356 Electrometer Overload Overload in Electrometer instrument during ameasurement.Measurement point is probablyinaccurate. Message also written in point property.
357 Error reading backplaneserial
Problem accessing instrument backplane.
358 Cannot find the Nameoption in file
Syntax error in Calibration or Self test files.
359 Cannot find the Connectionoption in file
Syntax error in Calibration or Selftest files.
360 Error in file … atline …
Syntax error in Calibration or Selftest files.
361 Overvoltage An overvoltage occurred.364 Measured capacitances
don’t matchDisagreement between values of capacitancemeasured for different configurations
365 Specimen capacitance belowlimit.
Measured capacitance below specified limit.
366 Specimen capacitance abovelimit.
Measured capacitance above specified limit.
367 Measured DCcurrent>MaxDCCurrent
Measured DC current exceeds limits set byMaxDCCurrent variable.
368 Measured humcurrent>MaxHumCurrent
Measured interference (hum) current exceeds limitsset by MaxHumCurrent variable.
Errors associated with D-file
No Message Explanation400 No sweep Error in D-file syntax.401 No points in sweep Error in D-file syntax.403 Wrong model Error in D-file syntax.404 Wrong parameter Error in D-file syntax.405 Printer error: Displays error from printer routines.406 Illegal value for x-min Illegal settings for axis values.407 Illegal value for y-min Illegal settings for axis values.408 Error: y-max is less than
y-minIllegal settings for axis values.
409 Error: x-max is less thanx-min
Illegal settings for axis values.
410 Error: Table not exported The D-file table was not exported properly.411 No information to show The information dialog could not display the
information.412 Cannot read from A:
(floppy)Results cannot read from floppy disk. Possiblydamaged floppy.
413 Cannot read from drive Results cannot read from specified drive.
IDAX User’s Manual70
© 2009 Megger AB
3.1 (347) Output voltage is not within specified limits
Measured output voltage is outside specified limits.
Possible reasons and countermeasures
1) Voltage electrode, Hi, is grounded:§ check measurement set-up and disconnect ground§ change measurement configuration if terminal of test object cannot be disconnected from
ground
2) Voltage electrode, Hi, is connected to measuring electrode, Lo or Ground:§ check measurement set-up and disconnect measuring or guard electrodes from the voltage
electrode. Voltage electrode, Hi, must not be connected to either measuring or guardelectrode
3) High stray capacitances to ground or high capacitance of the test object:§ lower the highest frequency used in C-file (see also Measurement variables)§ Lower the InitFrequency in the C-file (see also Measurement variables)
3.2 (361) Overvoltage
Measurement aborted due to overvoltage detected on measuring electrode, Lo. Potentialdifference between signal ground and real ground exceeds limiting value.
Possible reasons and countermeasures
1) Ground electrode is not connected to true ground:§ connect Ground electrode to true station/substation ground
2) Transients caused by accidentally disconnected ground connection:§ check ground connection
3.3 (364) Measured capacitances don’t match
Values of capacitance measured for different configurations, the UST, GST-Guard andGST-Ground are in disagreement.
Possible reasons and countermeasures:
1) When performing UST measurement, measuring electrode, Lo, is connected together withGround electrode or Lo is connected to ground:§ check measurement set-up and make sure that measuring electrode, Lo, is connected to a
non-grounded terminal of the test object and that Ground electrode is connected to ground
3.4 (365) Specimen capacitance below limit
Measured capacitance below limit specified in C-file by MinSpecimenC.
Possible reasons and countermeasures
1) Measured capacitance higher than 10 pF.
IDAX Error Messages 71
© 2009 Megger AB
Specimen size is very small which results in low value of capacitance:§ change limit set by MinSpecimenC to an approximately 10% lower value than measured
capacitance§ select another measurement configuration if possible§2) Measured capacitance lower than 10 pF.Most likely no contact with specimen:§ check connections with specimen for loose contacts§ check measurement cables for damage
3.5 (366) Specimen capacitance above limit
Measured capacitance above limit specified in C-file by MaxSpecimenC.
Possible reasons and countermeasures
Large size of test object results in high values of capacitance:§ change limit set by MaxSpecimenC to an approximately 10% higher value than measured
capacitance§ select another measurement configuration if possible§ decrease in test voltage allows for measuring at higher frequencies
3.6 (367) Measured DC current > MaxDCCurrent
Measured DC current exceeds the limits set in C-file by MaxDCCurrent.
Possible reasons and countermeasures
Features of the test object as well as certain configurations can result in high values of DCcurrent, use caution when analyzing the data.§ increase the limit for DC current set by MaxDCCurrent
3.7 (368) Measured hum current > MaxHumCurrent
Measured interference or hum current exceeds the limits set in C-file by MaxHumCurrent.
Possible reasons and countermeasures
1) Level of interferences is very high:§ try to reduce the influence of interferences by placing measuring cables close to each other
and avoiding cable loops§ increase the limit for hum current set by MaxHumCurrent
IDAX User’s Manual72
© 2009 Megger AB
Index- 3 -347 70361 70365 70366 71367 71368 71
— 5 -50/60 Hz Data Tab 19
— A -About 24Autoscale 58Axis
Autoscale 58Grid 58Label 59Scale 59Settings 58Title 59Unit 60
— C -Calibration 25
Calibration Procedure 25EM0 Calibration 25Full Calibration 25
Capacitance Geometrical 10
Conductivity 10Configuration
Graph 56, 57Information 56Models 56, 61Power Data 60Power Frequency Data 56Send to… 56Table 56, 60
Current Range 3
— D -Diagram
Zoom In 49Zoom Out 49, 51
Directives Emmision 3Immunity 3LVD 3Mechanics 3
Dissipation factor 10
— E -Edit
Change Sweep Color 47, 49Copy 47Copy All 47, 48Cut 47, 48Mark Sweep 47, 48Paste 47, 48Read Only 47, 49Select All 47, 48Toggle Sweep 47, 48Undo 47Unmark Sweep 47, 49
Equivalent Circuit Paralell 10Series 10
Error Messages 66Export 38
Data 38IDA Zip File 38Measurement Templates 41Object Templates 42Point or Comma 40TXT format 38With or without header 39WMF format 38
— F -File
Add Templates 31Add/Remove Templates 27Close Measurement 27, 32Delete File/Folder 27, 46Exit 27, 47Export 27Import 27
Index 73
© 2009 Megger AB
File Load D-file 27, 33New 27New Measurement Sequence 30New Measurement Template 29New Object 27New Object Template 29Open Measurement 27, 32Print 27, 42Remake Measurement 27, 32Remove Templates 31Rename File/Folder 27, 46Report 27, 43Save 27, 32Save As… 27, 32Send To… 27, 45
Frequency Range 3
— G -Getting Started
Data File 17Performing Measurement 17
Graph 57Axis 58Background 60Display Title 58
Graph Tab 19Grid 58
— H -Header 39Help
Contents 63Results Help 63What’s this? 63
— I -IDAX-206 System and Accessories 5IDAX-206 System Control
Comments 21Message Window 21Oscilloscope 21Selftest 21
Impedance Measurement 9Models 10
Impedance Models 62
Import 34File to Structure 34IDA Zip File 34Measurement Templates 36Object Templates 37
Instrument Panel Front 7Interface 7Output 7
— L -Label 59Language
Change Language 23Lin 59Linear scale 59Log 59Logarithmic scale 59
— M -MaxDCCurrent 71MaxHumCurrent 71MaxSpecimenC 71Measurement Templates 65Messages
Error messages 22Message window 22
MinSpecimenC 70Models 20, 61
— N -New 27
Measurement Sequence 30Measurement Template 29Object 27Object Template 29
— O -Object Templates 66Oscilloscope
Chanell checkboxes 22Grid 22Horizontal control 22Roll 22Vertical control 22
Output voltage 70
IDAX User’s Manual74
© 2009 Megger AB
Overvoltage 70
— P -Permittivity 10Power factor 10Precision 59Prefix 59Presenting Results
Getting Started 19On screen 19
— R -Report 43
List&Label Report 43MSD Word Report 44
Results Configuration Menu 56Diagram Menu 49Edit Menu 47File Menu 27Help 63Menus and Commands 26Test Browser 63User Interface 25View Menu 51
— S -Safety 6Scale 59Scientific 59Settings for Graph 58Specifications 3
— T -Table Tab 19Temperature 3
Operating 3Storage 3
Test Browser 63Data Structure 64Measurement Templates 65Object Templates 66
— U -Unit 60
— V -Version 24View
10 kV Equivalents 5550 Hz data 5550/60 Hz Data 5160 Hz data 55Browse Measurement Templates 51Browse Object Templates 51Browse Object Tree 51Browser 51Graph 51, 54Information 51Show Measurement Templates 51, 56Table 51, 55Text View 51, 56
Voltage 3Range 3