Инфузомат фмс руководство по эксплуатации

B. Braun Infusomat fmS Instructions For Use Manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

Infusomat

fmS

®

Instructions for Use

Software IFME/IFMe

loading

Related Manuals for B. Braun Infusomat fmS

Summary of Contents for B. Braun Infusomat fmS

  • Page 1
    Infusomat ® Instructions for Use Software IFME/IFMe…
  • Page 2
    Patient Safety Attention: Consult accompanying documents! Read Instructions for Use prior to use. Application only under regularly supervision by specially trained staff. ➤ Operation Connected electrical equipment must comply ➤ Ensure the unit is properly positioned and with the relevant IEC/EN-publications secured.
  • Page 3: Table Of Contents

    Infusomat ® Contents Infusomat® fmS / Overview Page 4 Operation Page 6 Additional Settings Page 7 Special Functions Page 8 Alarm Conditions Page 13 Start-up Graphs and Trumpet Curves Page 15 Technical Data Page 16 Warranty / TSC* / Service / Cleaning Page 18 Ordering Page 19…

  • Page 4: Infusomat® Fms / Overview

    Overview Handle For easy transport. Display All important information in plain text. Green background illumination only if connected to mains or a key is pushed at battery operation. Keypad for Input Correct Input/CLEAR Display reset to 000.0 Red LED indicates alarm condition Additionally «AAA.A”…

  • Page 5
    Guide for Short Infusion Pole Multi-Function- Connector (MFC) Mains/Power Connection Connection for staff (protect against ingress of moisture). In case of call, ambulances mains/power failure, the pump switches to (12 V DC) and inter- battery operation. Battery operation time: face for fm anaes- >…
  • Page 6: Operation

    Operation Infusion upper clip 1. Ensure reliable installation Never position infusion bottle below pump level. ➤ Connect staff call. ➤ Insert spike vertically into infusion bottle. Fill lower part of drop chamber to max. 2/3. ➤ Open roller clamp. Original 2.

  • Page 7: Additional Settings

    Additional Settings To Change the Delivery Rate volume has to be set. ➤ Press START/STOP. Green operating indicator goes out. ➤ Press C. Target Time (Time Pre-selection) Display reset to 000.0 ➤ Press key below TIME. ➤ Enter new delivery rate. (No setting possible, Enter target time via keypad (50 min = 50;…

  • Page 8: Special Functions

    Recording the data: In connection with the software “IFME” the The protocol of events only can be transferred Infusomat fmS is equipped with a history func- to a computer via interface (MFC interface tion (memory of events). This permanent lead). For this a terminal programme installed in…

  • Page 9
    Special Functions Select Special Mode SF Depending on the version, several functions may Interval bolus be deactivated. Contact service. Automatic bolus in set time interval. In case of ➤ To set special functions press key below SF manual bolus administration the interval bolus repeatedly until desired special function is is skipped.
  • Page 10
    Special Functions ➤ Drug Display Exit from selection: Press key below END. 9 selectable drug names can be stored (input via service program only). Check Battery Capacity Remaining battery life time is displayed, ➤ Select with key below (+). Drug is displayed, e.g.
  • Page 11
    Dosage Calculation (Overview) values (rate or dosage) are displayed. The dosage calculation automatically calculates ➤ First check rate displayed for plausibility, the delivery rate in ml/h. then confirm RATE. Value is displayed. Setting parameters: ➤ Start infusion. 1. Concentration Note: Dosage value with unit is displayed (down — per ml or on the right).
  • Page 12
    Special Functions ➤ Check entered or calculated rate for plausi- Secondary infusion bility and confirm with key below RATE. Rate is e.g. 100 ml Primary Infusion bag volume with displayed. e.g. 1000 ml delivery rate 10 ml/h bag volume with ➤…
  • Page 13: Alarm Conditions

    Alarms Operating alarms — audible alarm every 4 sec., — operating indicator off, — staff call. Remedy failure and restart infusion. ➤ Switch off device. Drop Alarm / Pressure Alarm ➤ Connect to mains/power or 12 V DC power Infusion bottle empty? source.

  • Page 14
    The delivery rate and a drug can be entered in matically will be switched to integrated the Infusomat® fmS as «proposal” via an rechargeable battery. external computer. Both items of data must be checked on the Infusomat fmS and acknow- Charge Battery ledged. ➤ Charge battery in case of: ➤…
  • Page 15: Start-Up Graphs And Trumpet Curves

    Start-up Graphs and Trumpet Curves Start-up Graphs Trumpet Curves 2 (ml/h) flow deviation Delivery rate = 1 ml/h Delivery rate = 1 ml/h Epmax Epmin t (min) 120 p∆t (min) 31 50 (ml/h) flow 10 deviation Delivery rate = 25 ml/h Delivery rate = 25 ml/h 37,5 Epmax…

  • Page 16: Technical Data

    Technical Data Type of unit Volumetric infusion pump Classification (acc. to IEC/EN 60601-1) defibrillator-proof, CF equipment Protection Class I; IP 22 (Moisture protection: drip protected) Class (acc. to Directive 93/42 EEC) II b Rated voltage / current 230 V AC~ (0.06 A), 50/60 Hz Mains fuse T 0.16 A or 200 V/230 V/240 V AC~ (0.06 A), 50/60 Hz…

  • Page 17
    Air detector Technical sensitivity Air bubbles > 0.01 ml Alarm triggering: With air bubble size of typ. 0.3 ml (limit value 0.4 ml) or 1.5 ml/h (cumulative value of 1 h as of air bubble volume 0.01 ml) Can be set from 0.01 to 0.3 ml via service program only Can be set from 0.5 to 3.5 ml/h via service program only Accuracy of set delivery rate typ.
  • Page 18: Warranty / Tsc* ) / Service / Cleaning

    Warranty provided concerning waste disposal and hygiene for batteries and disposables. B. Braun provides as from the date of delivery a warranty of 2 years for every Infusomat® fmS. This covers repair or replacement of parts dama- ged as a result of design/manufacturing errors or material defects.

  • Page 19: Ordering

    Ordering Art.-Nr. Infusomat® fmS 230 V 871 5548 Infusomat® fmS 200 — 240 V 871 5440 Infusomat® fmS 100 – 120 V 871 5416 Recommended accessories for Infusomat® fmS Connecting lead for potential equalisation 870 1628 MFC-Connecting lead for staff call 871 1682 MFC-Connecting lead for ambulance cars (12 V DC) 871 1674…

  • Page 20
    HOSPITAL CARE B. Braun Melsungen AG P.O.Box 11 20 D-34209 Melsungen Tel. +49 (0) 56 61 – 71-0 Fax + 49 (0) 56 61 – 71-20 44 Material-No. 3891 2341, Drawing No. M671010100F04 05/01 Printed on pulp bleached 100 % chlorine-free…

Patient Safety

Read Instructions for Use prior to use. Application only under regularly supervision by specially trained staff.

Operation

Ensure the unit is properly positioned and

secured.

Prior to use check audible and visual alarms

during self test. Also check the device for possible

damage.

If staff call is used we recommend to check the

equipment once after connecting the pump.

Connect to patient only after switching on the

device. Interrupt the connection during changing

administration set(s) to prevent incorrect dose

delivery.

Select cannula/catheter suitable for use with the

intended medical application.

Position the infusion line free of kinks.

Recommended change of disposables after 24 h

(consider national hygiene regulations).

Compare displayed value with entered value.

Start infusion only if values are corresponding.

Installation in medically used rooms must comply

with the appropriate regulations (e.g. VDE 0100,

VDE 0107 or IEC-publications.

Possible explosion hazard if used in presence of

flammable anaesthetics!

Air in line cannot be detected by the air detector

at stop-cocks, infusion ports and additional

administration set components.

Other components

Variations in pressure (e.g. as caused by change

of level) can affect the accuracy of the device.

Where several infusion lines are connected on

one single vascular access the possibility of their

exerting a mutual influence vice-versa cannot be

excluded.

Refer to respective manufacturer’s information

for possible incompatibilities of equipment resp.

drugs.

Use only compatible combinations of equipment,

accessories, working parts and disposables.

It is recommended to use original Infusomat®

infusion lines only.

The use of not recommended resp. incompatible

disposables may influence the technical specification.

Attention: Consult accompanying documents!

Connected electrical equipment must comply

with the relevant IEC/EN-publications

(e.g. IEC/EN 60950 for data-processing equipment).

The user/operator is responsible for the system

configuration if additional equipment is connected.

The international standard IEC/EN 60601-1-1 has to

be taken into account.

Safety Standards

The Infusomat® fmS satisfies all safety standards for

medical electrical devices in compliance with

IEC/EN 60601-1 and IEC/EN 60601-2-24.

The EMC-limits (electro-mangnetic compatibility)

according to IEC/EN 60601-1-2 and IEC/EN

60601-2-24 are maintained. If the equipment is

operated in the vicinity of other equipment which

may cause high levels of interference (e.g. HF surgical

equipment, nuclear spin tomography units, mobile

telephones etc.) maintain the recommended protec-

tive distances for these devices. Under certain

conditions malfunctions may occur which lead to a

device alarm with permanent alarm tone (see also

alarm conditions, page 13). Interferences may occur

e.g. at electro-magnetic fields > 10 V/m resp.

electro-magnetic discharges > 8 kV.

Special Function «without drip control»,

see page 10.

Page 1: Infusomat FmS Service Manual

Service Manual

Infusomat® fmS

0-10

0 —

Version 2.3 English

Page 2: Infusomat FmS Service Manual

0

This Service Manual is valid for Voltage 230 V: Ord. No.

Infusomat® fmS, German. . . . . . . . . . . . . . . . . . . . . . . . . 871 5424

Infusomat® fmS, French. . . . . . . . . . . . . . . . . . . . . . . . . . 871 5521

Infusomat® fmS, Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5530

Infusomat® fmS, Italian . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5564

Infusomat® fmS, Danish . . . . . . . . . . . . . . . . . . . . . . . . . 871 5432

Infusomat® fmS, Norwegian . . . . . . . . . . . . . . . . . . . . . . 871 5491

Infusomat® fmS, Swedish . . . . . . . . . . . . . . . . . . . . . . . . 871 5505

Infusomat® fmS, Finnish . . . . . . . . . . . . . . . . . . . . . . . . . 871 5513

Infusomat® fmS, Czech . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5580

Infusomat® fmS, Polish . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5599

Voltages 200 V / 230 V / 240 V, switchable:

Infusomat® fmS, English . . . . . . . . . . . . . . . . . . . . . . . . . 871 5440

Infusomat® fmS, English . . . . . . . . . . . . . . . . . . . . . . . . . 871 5548

Infusomat® fmS, Spanish. . . . . . . . . . . . . . . . . . . . . . . . . 871 5459

Infusomat® fmS, Portuguese . . . . . . . . . . . . . . . . . . . . . . 871 5467

Infusomat® fmS, Turkish . . . . . . . . . . . . . . . . . . . . . . . . . 871 5572

Voltages 100 V / 110 V / 120 V, switchable:

Infusomat® fmS, English . . . . . . . . . . . . . . . . . . . . . . . . . 871 5416

Infusomat® fmS, Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . 871 5475

Infusomat® fmS, Spanish. . . . . . . . . . . . . . . . . . . . . . . . . 871 5483

Infusomat® fmS, Portuguese . . . . . . . . . . . . . . . . . . . . . . 871 5394

This Service Manual is available under the following part number:

Designation Part No.

Service Manual Infusomat® fmS, English . . . . . . . . . . 8713 9124

Languages of this Manual Designation Part No.Service Manual Infusomat® fmS, German . . . . . . . . . . 8713 9123

The complete Service Manual contains the following pages:

Page 0-1 to page 0-10Page 1-1 to page 1-12Page 2-1 to page 2-8Page 3-1 to page 3-10Page 4-1 to page 4-20Page 5-1 to page 5-2Page 6-1 to page 6-2Page 7-1 to page 7-4Page 8-1 to page 8-8Page 9-1 to page 9-2Page 10-1 to page 10-4Page 11-1 to page 11-2Page A-1 to page A-2

0 — 2 Infusomat® fmS 2.3 gb

Page 3: Infusomat FmS Service Manual

0

0 — Table of Contents

Important Preliminary Remarks Service Work Page 0 — 5Technical Safety Checks Page 0 — 5Current Versions Page 0 — 5Revision Service Page 0 — 5Quality Management Page 0 — 6Checks and Repair Page 0 — 6Notes on ESD Page 0 — 6Spare Parts and Test Equipment Page 0 — 7Setting Off Page 0 — 7List of Abbreviations Page 0 — 8

Contact Persons Technical Training Page 0 — 9Entry for Technical Training Page 0 — 9Ordering of Spare Parts and Test Equipment Page 0 — 9Service Hotline Page 0 — 9Return of Spare Parts and Test Equipment Page 0 — 9Safety Officer

(§ 30 MPG) Page 0 — 9Translation Page 0 — 9

System Overview Physical Construction Page 1 — 1Operation Flow Chart Page 1 — 3Function Page 1 — 4Voltage Supply Page 1 — 5Signal Table Page 1 — 6Mains Operation Page 1 — 7Battery Operation Page 1 — 8Alarm Circuit Page 1 — 8Pump Unit Page 1 — 8Computer Interface Page 1 — 9Braun fluid manager system (fm system) Page 1 — 10Internal Assignment Page 1 — 10Accessories Page 1 — 11

Software Software Update Page 2 — 1Approved Software Versions Page 2 — 2Error Messages and Alarms Page 2 — 4Software Default Values Page 2 — 6

Service Program Structure of the Service Program Page 3 — 1Additional Functions with

Plugged Service Connector Page 3 — 2Start / Quit the Service Program Page 3 — 3Unit Data Page 3 — 3History Data Page 3 — 4Test Page 3 — 5Unit Modifications Page 3 — 6Calibration Page 3 — 9

Unit Elements Mains Fuses Page 4 — 1Battery Page 4 — 1Door Lock Page 4 — 2Pump Cover Page 4 — 3Housing and Handle Page 4 — 4

Infusomat® fmS 2.3 gb 0 — 3

Page 4: Infusomat FmS Service Manual

Table of Contents0

Controller Board Page 4 — 5Rear Panel Page 4 — 6Front Frame Page 4 — 11Pump Unit Page 4 — 12Pressure Sensor Page 4 — 14Air Inline Sensor Page 4 — 17Operating Unit Page 4 — 18Barcode Label Page 4 — 19Frame with Seal Page 4 — 19

Checks after Repair Page 5 — 1

Maintenance Page 6 — 1

Technical Safety Check TSC Page 7 — 1

Procedural Instructions on the TSC Visual Inspection Page 8 — 1Electrical Safety

according to IEC / EN 60601-1 or VDE 0750 and VDE 0751 Page 8 — 2

Functional Inspection Page 8 — 3

Test Equipment and Special Tools Page 9 — 1

Spare Parts List Unit Elements Page 10 — 1Colours Page 10 — 2Miscellaneous Page 10 — 2Software Update Page 10 — 2

Index Page 11 — 1

Revision Documentation Revision Service-Documentation Page A — 1Current Information Page A — 1

0 — 4 Infusomat® fmS 2.3 gb

Page 5: Infusomat FmS Service Manual

0

0 — Important Preliminary Remarks

Service Work The present manual is for your information only. The possession of this manual does not authorize the performance of service work. Service tasks may only be executed by persons, who

— have received appropriate training on the system from B. Braun

— are included in the revision service

— possess the necessary test equipment and mechanical aids, and

— fulfill the personal requirements (training and knowledge).

Technical Safety Checks The user is obliged to perform or to have performed the Technical Safety Checks on those medial products for which these checks have been prescribed by the manufacturer and to carry them out according to the indications of the manufacturer as well as the generally approved technical standards while adhering to the pe-riods stated (§ 6 MP BetreibV).

B. Braun also recommends training on the Technical Safety Checks, or to perform at least the steps indicated in the current version of the manual, as:

— the TSC requires that the instructions in the manuals are ob-served

— the manuals are a reference for measurements

— depending on the unit type, the Service Program must be called which may lead to a dangerous unit condition in case of inappropriate operation. Furthermore, a special service connector may be necessary.

Current Versions This manual version corresponds to the state when the manual was written. B Braun reserves the right to make technical modifi-cations. The state of the revision is indicated by the index number in the footer of every page.

Revision Service The possession of this manual does not automatically mean inclu-sion in the revision service. You will be included in the revision service after:

— technical training by B. Braun Melsungen or

— a written order placed with the sales department of B. Braun (fee required).

Responsibility of the Manufacturer The manufacturer, person who assembles, installs or imports the device can only be held responsible for safety, reliability and per-formance if

— mounting, enhancements, new settings, changes or repairs are carried out by duly authorized persons,

Infusomat® fmS 2.3 gb 0 — 5

Page 6: Infusomat FmS Service Manual

Important Preliminary Remarks0

— the electrical installation in the corresponding room meets the requirements of the VDE 0107, VDE 0100 part 710 or IEC 60364-7-710 and the national standards,

— the device is used in accordance with the instructions for use and the Service Manual,

— the Technical Safety Checks are performed at regular inter-vals,

— a current manual which corresponds to the revision state is used when carrying out maintenance, repair and service,

— the service technician takes part in the revision service,

— the technician has participated in a technical training course for the specific B. Braun unit.

Quality Management B. Braun is certified in accordance with DIN EN ISO 9001 and ISO 13485. This certification also includes maintenance and serv-ice.

The unit has the CE label. The CE label confirms that the device corresponds to the “Directive of the Council for Medical Products 93/42/EC” of June 14, 1993.

Checks and Repair Training may only be performed by B. Braun. The possession of the manual does not authorize the performance of repairs. The in-structions on electrostatic sensitive components (ESD standards) must be observed.

After repair a device check or diagnosis is to be carried out.

Notes on ESD Semiconductors can be destroyed by electrostatic discharge. Es-pecially MOS components can be damaged by interference from electrostatic fields, even without discharge via contact. This type of damage is not immediately recognizable. Unit malfunctions can even occur after a longer period of operation.

Each workstation must be equipped according to the recommen-dations with the necessary static protective measures, if ESD components or boards are handled.

Each workstation must be equipped with a conductive table sur-face. The conductive surface, the soldering iron or the soldering stations must be grounded via protective resistors.

Chairs must be of antistatic design. The floor or floor mats should be of electrically conductive material.

Personnel must wear conductive wristbands which are connected to a central ground potential via protective resistors, e.g. the ground contact of a wall outlet. Furthermore it is recommended that personnel wear cotton clothing and electrically conductive shoes to prevent electrostatic charge.Fig.: 0 — 1

0 — 6 Infusomat® fmS 2.3 gb

Page 7: Infusomat FmS Service Manual

Important Preliminary Remarks 0

Spare Parts and Test Equipment Only use original spare parts from the manufacturer. Do not tamper with assembly groups which can only be exchanged com-pletely. The spare parts required are listed in Section 9.

Service personnel are responsible for the calibration of their test equipment. Original test equipment can be calibrated at the works of B. Braun. Further information is available upon request.

Setting Off Additional notes and warnings are set off as follows:

Note

Is used for additional or special notes concerning information and working steps.

CAUTION

Is used for working steps which may result in damage to the unit, system or to a connected device.

WARNING

IS USED FOR WORKING STEPS WHICH MAY RESULT IN PERSONAL INJURY.

References to chapters are shown as follows (see “Setting Off“ ➨ pg. 0 — 8)

References to figures and tables are shown as follows Fig.: 2 — 3 or Table 2 — 1

References to item numbers in figures are shown as follows (Fig.: 1 — 1 / Item 1) In this case “Fig.: 1 – 1“ is the figure number and “Item 1“ the item number within the figure.

When the Service Manual is stored as pdf-file, these references are displayed green. Click with the mouse button on a reference to jump to the corresponding source.

Menu commands are described as: Menu File.

Infusomat® fmS 2.3 gb 0 — 7

Page 8: Infusomat FmS Service Manual

Important Preliminary Remarks0

List of Abbreviations Abbreviations which are not generally known, but are used in this manual, are listed below.

CC Computer Controlled

CLR Clear

Dig Digit

ESD Electrostatic Discharge

UA Unit Alarm

FuP Function Microprocessor

KuP Control Microprocessor

LCD Liquid Crystal Display

OIL Original Infusomat Line

PCA Patient Controlled Analgesia

TSC Technical Safety Check

TEMP Temperature

0 — 8 Infusomat® fmS 2.3 gb

Page 9: Infusomat FmS Service Manual

0

pg:0 — 10

0 — Contact Persons

Technical Training Via local representative.

Entry for Technical Training Application for a technical training course must be made via the responsible representative.

Ordering of Spare Parts and Test Equipment Please contact your local B. Braun subsidary.

International Technicians (Intercompany)Nadja Machal

Fax: +49 5661 / 75 -47 89 e-mail: [email protected]

Service Hotline Karl Tippel, Tanja Kördel Phone: +49 5661 / 71 — 35 25Fax: +49 5661 / 71 — 35 26e-mail: [email protected]: [email protected]

Return of Spare Parts and Test Equipment B. Braun Melsungen AG Schwarzenberger Weg 73-79 Wareneingang Werk C 34 212 Melsungen Germany

Safety Officer (§ 30 MPG)

Dr. Ludwig Schütz e-mail: [email protected]

Translation Brückner GmbH, Germany

Infusomat® fmS 2.3 gb 0 — 9

Page 10: Infusomat FmS Service Manual

Contact Persons0

For your notes:

0 — 10 Infusomat® fmS 2.3 gb

Page 11: Infusomat FmS Service Manual

11 — 12

1 — System Overview

Physical Construction The Infusomat® fmS is a compact volumetric peristaltic infusion pump.

Standard delivery rate range 0.1 to 999.9 ml/h

The unit is operated via a membrane keyboard. It is equipped with an LCD-display (liquid crystal display) for the display of the deliv-ery rate and the operating support of the user. Two control LEDs display alarms, and the running of the infusion pump.

Barcode

A barcode label is attached to the left front side of new Infuso-mat® fmS unit versions which can be retrofitted on previous de-vices. This barcode label is used to read the serial and DIANET type number via a scanner when the Infusomat® fmS is operated in an fm-system.

Fig.: 1 — 1 Unit design Infusomat® fmS

Handle

Door lock

Control LEDs

Keyboard

LCD-display

Softkeys

Guide for short pole clamp

Drop sensor socket

Flow inhibitor

Air inline sensor

Drop sensor

Mains connection

Potential equalization

fm Recessed plug

Optical

MFC-socket

interface

Peristaltic pump

Barcode label

Infusomat® fmS 2.3 gb 1 — 1

Page 12: Infusomat FmS Service Manual

System Overview1

Infusion Lines

The Infusomat® fmS can be operated with the Original Infuso-mat® Line (OIL) and the Infusomat® Space Line. The lines are dis-tinguished by the different silicone pump segments. They have to be inserted in a different way; in addition the TSC differs.

WARNING

OPERATING THE UNIT ALTERNATELY WITH THE ORIGINAL INFU-SOMAT® LINE AND THE INFUSOMAT® SPACE LINE IS NOT RECOM-MENDED.

Note

Please observe the “Instruction for Technical Service — Adjustment to Space Line“ when using the Infusomat® Space Line for the first time.

1 — 2 Infusomat® fmS 2.3 gb

Page 13: Infusomat FmS Service Manual

System Overview 1

Operation Flow Chart

See instructions for use for detailed information.

Fig.: 1 — 2

Switch On

Standard operation Service connectorplugged … 1

Special functions … 3

Set delivery rate Volume preselection Time preselection

Rate calculation

Start infusion

Stop infusion

Recall info

Change delivery rate …

Recall Bolus

1 See Service Program diagram2 Only when preselected time or preselected volume has expired3 Please pay attention to activation in Service Program4 Only available with IFMC5 From software IFMe, IFME on6 From software IFME on

KOR Mode 2

— Dose calculation 5

— Bolus function

— Standby function

— Drug change

— CC Mode 4

— Switch-off pres-sure

— Drop control

— Piggyback 5

— Battery capacity

— Data lock

— Alarm tone 6

— Contrast setting

— Clock5

Infusomat® fmS 2.3 gb 1 — 3

Page 14: Infusomat FmS Service Manual

System Overview1

Function Two independent software-controlled microprocessor systems control and monitor the hardware. On the basis of their functions, they are defined respectively as a control and a function proces-sor. Both systems work with independent clock frequencies and have access to different program and data memories. All safety-relevant functions are handled by both microprocessors and the results are counter checked (CF- and FC-latch).

The input via the keyboard is fed to both processors. Additionally the acknowledgement signal of the ON/OFF key is fed to the mains power supply logic (voltage E/A-TAS). The function proces-sor has also access to this logic via E/A-INT.

Description of the voltage signals (see „Signal Table“ ➪

p. 1 — 6).

Fig.: 1 — 3 Block diagram

1 — 4 Infusomat® fmS 2.3 gb

Page 15: Infusomat FmS Service Manual

System Overview 1

Voltage Supply The voltage supply is generated either directly from mains, via the FM connector (14 V connection to the fluid manager system), or via the MFC-connector (11 to 16 V) and as an internal supply via the internal 7.2 V NiCD battery. The mains module is available in three versions: 230 V, 220 / 230 / 240 V and 100 / 110 / 120 V. The rated voltage has a tolerance of + 10% to – 15 %.

A voltage of 11 V to 18 V is available after transformation and rectification. This voltage is fed to the battery charge circuit and the unit supply. This is also valid for an external 12 V supply from the MFC- or FM-connector. The FET V10 switches between the ex-ternal and the internal voltage supply. The transistor V10b works as an ON/OFF switch for the operating voltages UPS, UMOT and +5V. The +5V supplies the complete electronics including the double channel microprocessor system. A window comparator constantly monitors the +5V for undervoltage or overvoltage. The function is checked during switch-on. The operating voltage UPS supplies the stepper motor and the UMOT, the stepper motor drive.

The transistor V47 switches the operating voltage UMOT. In case of an alarm the motor is switched off by V47. Additionally the switching function of the transistor is checked during the switch-on test.

The circuit has two separate assembly groups with separate sup-ply voltages UBA and UBB. The ON/OFF circuit has a retriggerable delay switch-off. A follow-up charging circuit drives the transis-tor V10b.

The alarm logic (operating voltage UBB) is an RS latch. This is set when the unit is running and activates the alarm circuit. The alarm buzzer and driver are also driven by UBB. The ON/OFF circuit is activated and the voltage supply is switched on by pressing the ON/OFF key. The alarm latch is reset simultaneously. A function test of the voltage monitoring, motor circuit and alarm activation is performed. The voltage supply is maintained by cyclic self-hold-ing pulses fed to the logic. The alarm latch is also activated.

Infusomat® fmS 2.3 gb 1 — 5

Page 16: Infusomat FmS Service Manual

System Overview1

Signal Table

Fig.: 1 — 4

Signal Meaning Signal Meaning

+5V Voltage supply electronic PKS2 Pump Head Sensor 2

5V-HT Overvoltage test PKSS Pump Head Sensor Control

5V-LT Undervoltage Test PRS Staff Call Relay Control

AK-I Battery Charge and Discharge Current PRS-F Staff Call Relay Function Channel

AK-LAD Battery Capacity ON/OFF PKSS Pump Head Sensor Control

AK-Test Battery Test PRS.RUF Staff Call Relay Control

ALA-UB Operating Voltage Alarm RDE Rate Display Enable

CS Chip Select RES Power on Reset

DI Data Input RES-F Reset Function Channel

DO Data Output RES-K Reset Control Channel

E/A-INT ON/OFF by Microprocessor RTS Return to Send (DIANET)

E/A-ST ON/OFF Status Rx Receive Data

E/A-TAS ON/OFF Key SCK Serial Data Clock

Table 1 — 1 Signal table (Part 1 of 2)

1 — 6 Infusomat® fmS 2.3 gb

Page 17: Infusomat FmS Service Manual

System Overview 1

Mains Operation When the unit is connected to mains the unit supply voltage is switched on for the duration of the switch-off delay time. If the microprocessor recognizes a sufficient mains voltage for charg-ing, the voltage supply is maintained. In this case only a battery balance is carried out, because a key was not pressed. The current battery capacity and the battery operting hours are displayed in the LCD. The unit is switched off when the ON/OFF key is pressed for at least 2 seconds. Thereby the self-holding is triggered and the alarm latch is reset with a delay. After another 20 seconds the unit is switched off, because the pulses are missing. If the Infuso-mat® fmS is switched off in mains operation with the ON/OFF key, the internal mains voltage is still present.

EDB Electronic Occlusion Pressure TD-A0 Text Display Address 0

FMC-F FM Connection Function Channel TD-A1 Text Display Address 1

FMC-K FM Connection Control Channel TD-A2 Text Display Address 2

UEXT-N External 12V Supply (-) TD-A3 Text Display Address 3

UEXT-P External 12V Supply (+) TD-E Text Display Enable

UFM-P External 12V Supply (FM) TD-R/W Text Display Read/Write

LFCL Air Sensor Clock TD4 Text Display Data 4

LFDA Air Sensor Data TD5 Text Display Data 5

LFS Air Sensor Signal TD6 Text Display Data 6

LFSEL Air Sensor Selection TD7 Text Display Data 7

LFT Air Sensor Reset TSA Drop Sensor Output

MISO Serial Data Output Interface TSCL Drop Sensor Clock

MOSI Serial Data Input Interface TSE Drop Sensor Receiver

MOTEIN Motor ON TSR Drop Sensor Regulation

MS Motor Control TSS Drop Sensor Control

P-ENA Port Enable Tx Transmit Data

PH0 Phase 0 UBA, UBB Supply Voltage for Alarm, On/Off Logic, RTC

PH1 Phase 1 UPS Switched Operating Voltage UB

PH2 Phase 2 UPS-M UPS Measurement Line

PH3 Phase 3 UMOT Supply Voltage of Motor Drive

PKS Pump Cover Sensor UMOT-M UMOT Measurement Line

PKS1 Pump Head Sensor 1 URTC Supply Voltage Clock Module

Signal Meaning Signal Meaning

Table 1 — 1 Signal table (Part 2 of 2)

Infusomat® fmS 2.3 gb 1 — 7

Page 18: Infusomat FmS Service Manual

System Overview1

In mains operation battery function is checked during the switch-on test. Therefor the charge and discharge current are measured and the charging of the battery is interrupted for the duration of measurement.

Battery Operation The battery function is monitored by the following data: charge current, discharge current and time, and self-discharge time. In battery operation the battery function is checked during switch-on test. The theoretical load condition is read from the clock mod-ule of the battery. Then the battery is connected to UPS and the voltage is measured. If the minimum requirements are not reached a battery alarm is activated.

Alarm Circuit The alarm generation consists of:

— Standstill of pump due to switch-off of MS (motor circuit) and UMOT (motor operating voltage)

— Audible alarm due to the drive via ALA-AK (control channel) or via ALA-UB. The alarm volume is about 65dBA.

— Optical alarm. Is displayed in the LCD- and a separate LED-display. Additionally the set rate flashes with AAA.A.

— Staff call via the MFC staff call cable.

The user must check the optical and audible alarm during the switch-on test. An alarm must be activated to test the staff call alarm of the Infusomat® fmS, e.g. open pump cover during oper-ation.

Pump Unit The pump head is driven by a stepper motor. Each full step of the motor is realized by 5 microsteps. The motor is driven by an FET output stage. The function processor controls the motor with the MS signal. A slot disc which is mounted on the pump head axle is scanned by two light barriers (PKS1 and PKS2 signal). Thereby the control microprocessor monitors the direction of rotation and speed of the pump head.

The pump head position is also determined with the PKS2 signal. The motor can therefore be accelerated during the withdrawal phase. Thus a nearly pulse-free flow is realized in the lower deliv-ery range (<100 ml/h). The total pump head cycles and running time are available in the Service Program under history data.

1 — 8 Infusomat® fmS 2.3 gb

Page 19: Infusomat FmS Service Manual

System Overview 1

Mechanical Occlusion Pressure:

The Infusomat® fmS has a linear peristaltic pump. This pump has 12 slides which are driven by a camshaft.

When the pump cover is closed, the pump tube is squeezed (oc-clusion) by at least one of the slides, independent of the pump head position. The complete pump unit is mounted behind the front panel in the cabinet frame. The hinges and the locking bow for the pump cover are led through the front panel. The pump cov-er is automatically closed when the operating unit door is closed. The slides are pressed against the pump cover by a spring system in the pump unit. Thereby a delivery pressure is realized and me-chanically limited by the springs.

If the pressure limit is exceeded there is no volume delivery. The drop sensor activates an alarm. If one of the springs fails, the spring system will ensure that an unsafe condition cannot occur (free flow). The two remaining springs ensure an appropriately high occlusion pressure.

Electronic Occlusion Pressure:

The electronic occlusion sensor is mounted on the output side of the pump. A spring pressure loaded slide is seated on the infusion line. An increase of pressure in the infusion line leads to a deflec-tion of the coil core via the pressure slide. The depth of immersion is measured inductively. When a preset pressure threshold is reached the pump drive is switched off, and an alarm is activated. The electronic occlusion pressure is a single channel circuit. In case of a failure, the mechanically limited maximum pressure can be reached.

Motor Switch-Off by Both Processors:

Function processor: MS signal to switch off the motor drive. — Control processor: MOTEIN signal to switch off the drive of the motor operating voltage.

Computer Interface The Infusomat® fmS is equipped with a computer interface. It can be connected to the optical interface or via the MFC service con-nector. To activate the computer operation please ask for a de-tailed description from B. Braun.

Up to software version IFMC: DIANET From software version IFMe, IFME on: DianetStar

Infusomat® fmS 2.3 gb 1 — 9

Page 20: Infusomat FmS Service Manual

System Overview1

Braun fluid manager system (fm system) The Infusomat® fmS can be operated as a stand-alone unit or in-tegrated in an intensive care unit, e.g. the B. Braun fluid manager system (fm system). It is integrated by simply snapping the unit into the system.

Mains supply and data communication are automatically con-nected. Thereby data acquisition and transmission to higher com-puter system levels are possible.

Internal Assignment

Fig.: 1 — 5

1 — 10 Infusomat® fmS 2.3 gb

Page 21: Infusomat FmS Service Manual

System Overview 1

Accessories General Designation Ord. No.

Mounting clip for drop chamber ”TK 2000” . . . . . . . . 3477 3223

Mounting clip for drop chamber ”Intrafix air” . . . . . . 3477 3215

Drop sensor, complete . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 578A

Short pole clamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5873

Drop chamber holder . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3088

Mains lead (200-240V~) . . . . . . . . . . . . . . . . . . . . . . . . 3450 2718

Mains lead (100-120V~) . . . . . . . . . . . . . . . . . . . . . . . . 3450 5423

Mains lead USA / CAN . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5393

Universal clamp, complete . . . . . . . . . . . . . . . . . . . . . . 3450 5857

Pole clamp (universal clamp, rotating) . . . . . . . . . . . . 3450 9054

Infusomat® fmS 2.3 gb 1 — 11

Page 22: Infusomat FmS Service Manual

System Overview1

For your notes:

1 — 12 Infusomat® fmS 2.3 gb

Page 23: Infusomat FmS Service Manual

22 — 8

2 — Software

Software Update Designation Ord. No.

Update kit IFMC02003 . . . . . . . . . . . . . . . . . . . . . . . . . 3450 645A

Update kit IFMe02004 / IFME03004 . . . . . . . . . . . . . . 3450 645C

Update kit IFMe02005 / IFME03005 . . . . . . . . . . . . . . 3450 645D

Update kit IFMe02006 / IFME03006 . . . . . . . . . . . . . . 3450 645E

Update kit IFMe02007 / IFME03007 . . . . . . . . . . . . . . 3450 645F

MFC interface line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 1661

The higher digit always replaces the lower digit for the revision level, e.g. IFME02007 replaces IFME02006.

Units with an old software version (e.g. IFMC02001) can be up-dated to the new software version IFME03007.

When the software group changes the unit functions are changed, too. Therefore unit users must be informed (e.g. instruct the user and exchange the instructions for use – software coding (e.g. IFME) is to be found on the cover page of the instructions for use.)

Note

Do not use an update program on Windows NT systems.

Note

Mark the unit after having updated the software! The new soft-ware version must be clearly recognizable.

Only update from old to new software versions, never in reverse order (e.g. never update from IFMC02003 to IFMC02001!).

All units used in one ward should have the same software status and basic setup to avoid operator mistakes.

Note

Software updates must be reported to B. Braun for registration. Observe the notes of the update program and the supplements!

Fig.: 2 — 1 Coding

Position 1 2 3 4 5 6 7 8 9

Digit I F M E 0 3 0 0 7

Revision level

Hardware identification

Software group

Infusomat® fmS

Infusomat® fmS 2.3 gb 2 — 1

Page 24: Infusomat FmS Service Manual

Software2

Approved Software Versions IFMC02001

— Basic software (Must not be used any more. Please contact the Technical Service of B. Braun).

IFMC02002

— Error elimination at Bolus special function at staff call on the fm system

— Optimized air sensor evaluation

IFMC02003

— EMC optimized

— New error code FF16, defective membrane keyboard

IFMe02002

— Preselected volume and time counted down to 0

— Interface changed to DIANET Star (not compatible with Dianet)

— New special function dose calculation

— New special function Piggyback

— New special function clock

— Storage of alarms in case of malfunctions, which can be re-called in the Service Program, function 230

IFME03002

Only controller board with loudspeaker (see „Controller Board“ ➪ p. 4 — 5).

Like IFMe02002 and in addition:

— Alarm volume can be set

— History function

IFMe02003

— Dianet Star corrected

IFME03003

— Additionally History corrected

— Log of volume delivered

— New event: normal mode / piggy mode

IFMe02004 / IFME03004

— Optimized dose calculation

— New language “Hungarian” in language group E

2 — 2 Infusomat® fmS 2.3 gb

Page 25: Infusomat FmS Service Manual

Software 2

IFMe02005 / IFME03005

— Cyclical battery test

— Retaining or deleting last dose calculation when unit is switched off, can be set in the Service Program

IFMe02006 / IFME03006

— Optimized switch-off cycle of Service Program

IFMe02007 / IFME03007

— Reset the battery capacity from 0 mAh to 1 mAH in the event of a negative battery test result.

— Reset the battery voltage acceptance thresholds from 7.55 V to 7.35 V.

Infusomat® fmS 2.3 gb 2 — 3

Page 26: Infusomat FmS Service Manual

Software2

Error Messages and Alarms

Alarms of the function processor 80c535 are displayed in the text box of the LCD-display. Alarms of the control processor 68HC11 are displayed in the 7 segment display. The alarms help to trou-bleshoot unit malfunctions. As not all malfunctions can be con-sidered, unit malfunctions with different messages, which are not listed, can be displayed, or there may be no message.

Detected unit alarms are displayed in the text box as ”Unit Alarms” in the selected language. Additionally the error number is displayed in the text box.

Function Processor 80c535

Fig.: 2 — 2

7 segment display

Text box

Text Box Description

100 defective RAM memory

101 UMOT cannot be switched on

102 UMOT still switched on despite overvoltage

103 UMOT still switched on despite MOTEIN=0

104 UMOT still switched on despite undervoltage

105 ON/OFF key pressed longer than 14 sec

106 defective air sensor (calibration value?)

107 defective program memory

108 defective program flow

109 different number of pump head cycles

110 different keyboard gaps between 80c838 and 68hc11

111 different program versions between 80c535 and 68hc11

112 defective program flow

113 testbit!=0 out of switch-on test

116 defective program memory — text

117 defective program memory — text does not match with program

118 reset during active operation

119 defective ROM

Table 2 — 1

2 — 4 Infusomat® fmS 2.3 gb

Page 27: Infusomat FmS Service Manual

Software 2

Control Microprocessor 68hc11

FFxx is displayed in the 7 segment display with flashing dots. FFxx is the error code.

7 Segment Display Description

FF01 dummy for test

FF02 battery not present / missing battery current

FF03 defective RAM memory

FF04 defective program memory

FF05 defective program memory

FF06 calibration data error from EEPROM

FF07 pump head cycle not plausible

FF08 failure / inaccuracy of system clock

FF09 failure 100msec system clock

FF10 reset during active operation

FF12 no dynamic pressure sensor signal (EDB)

FF16 defective membrane keyboard (from IFMC02003 on)

Table 2 — 2

Infusomat® fmS 2.3 gb 2 — 5

Page 28: Infusomat FmS Service Manual

Software2

Software Default Values Unit No.: _______________________

1) From software IFMe, IFME on2) No longer available in software IFMe, IFME

Menu Item Default Customer Setting

Standard function User language depending on Art. No. _______________________

Alarm type single stage _______________________

Staff call static without OFF Alarm, without switch-on pulse

_______________________

Ward identification ”Ward Identification” _______________________

Drug 0 blank _______________________

Drug 1 … 9 drug 1 … 9 _______________________

Operating alarms 0 _______________________

Minimum delivery rate 0.1 ml/h _______________________

Maximum delivery rate 999.9 ml/h _______________________

Maximum air rate 1.5 ml/h _______________________

Maximum air bubble 0.30 ml _______________________

Special functions Dose calculation deactivated _______________________

Bolus deactivated _______________________

Standby activated _______________________

Drug selection deactivated _______________________

CC Mode 2) deactivated _______________________

Switch-off pressure activated _______________________

Drop control deactivated _______________________

Piggyback1) deactivated _______________________

Battery capacity deactivated _______________________

Data lock deactivated _______________________

Alarm tone 1) deactivated _______________________

Contrast deactivated _______________________

Clock 1) deactivated _______________________

SM menu Interval Bolus dose Off _______________________

Online rate setting On _______________________

Double rate entry Off _______________________

User data Switch-off pressure high _______________________

Contrast optimum contrast _______________________

CC Address2) 1 _______________________

Drug 0 _______________________

Data lock Off _______________________

Standby time 24h 00min _______________________

Drop control On _______________________

2 — 6 Infusomat® fmS 2.3 gb

Page 29: Infusomat FmS Service Manual

Software 2

Bolus key On _______________________

Bolus rate 999.9 ml/h _______________________

Calibration data Air sensor calibration value 182 mV must not be changed

Scale factor 54 _______________________

Unit specific data DIANET type no. depending on unit _______________________

Unit No. depending on unit _______________________

Operating hours depending on unit _______________________

Battery hours depending on unit _______________________

Number of pump head cycles depending on unit _______________________

Menu Item Default Customer Setting

Infusomat® fmS 2.3 gb 2 — 7

Page 30: Infusomat FmS Service Manual

Software2

For your notes:

2 — 8 Infusomat® fmS 2.3 gb

Page 31: Infusomat FmS Service Manual

33 — 10

3 — Service Program

Structure of the Service Program

Fig.: 3 — 1

Service connectorplugged

Switch On

Special function(see operating flow diagram)

Standard operation(see operatingflow diagram)

Short display:software version,

user language

Service Programactivated

Group:Unit data

Group:History data

Group:Test

Group:Unit modification

Group:Calibration

Ward identification120.0

Operating hours200.0

Air inline sensor300.0

Service language400.0

Scale factor510.0

Software version100.0

Drug110.0

Serial number140.0

DIANET type no.150.0

Battery op. hours210.0

Pump headcycles220.0

Alarms230.0

Pressure sensor310.0

User language410.0

Alarm tone420.0

Staff call430.0

Special function440.0

Menu450.0

Delivery rate460.0

Air alarm470.0

Air inline sensor520.0

Pressure stage540.0

Pump data550.0

Delete dose values490

History 2)

560.0

1 From software IFMe, IFME on2 From software IFME on

Dianet modeDisplay 1)

480

Infusomat® fmS 2.3 gb 3 — 1

Page 32: Infusomat FmS Service Manual

Service Program3

Additional Functions with Plugged Service Connector

Software Version and User Language

1. Plug service connector on MFC socket at the rear of the unit.

2. Switch on unit and keep the ON/OFF button pressed (for max. 15 s).

3. The software version and user language are displayed in the LCD-display.

4. The unit is switched on when the ON/OFF button is released.

5. ** appears in the LCD-display if the service connector is plugged.

Two LEDs are integrated in the service connector:

— green = Power supply active

— red = Alarm

The following conditions are activated:

— The operating alarms are muted.

— All special functions are accessible (including the disabled).

— The special functions are slightly modified. (Example: SM battery capacity has keys for 0 min/32 min presetting).

— The battery capacity display switches between the nominal and the actual capacity. If the maximum battery capacity is not reached ”?” will be displayed before the new nominal ca-pacity.

Contrast Setting

1. Select ”Contrast Setting” with the SM key. The softkey sym-bols will flash.

2. Set display contrast with the (+) or (-) key.

3. Return to main menu with END.

Disabling the Pressure Sensor (Occlusion Pressure Mechanical)

The electronic pressure monitoring can be deactivated to check the mechanical occlusion pressure.

1. Select ”Occlusion Pressure” with the SM key.

2. Select ”Mechanical”.

3. Return to main menu with END.

A too low pump speed is indicated in the display by ”Pressure Alarm” with underlined stars.

3 — 2 Infusomat® fmS 2.3 gb

Page 33: Infusomat FmS Service Manual

Service Program 3

Start / Quit the Service Program Activate the Service Program

1. Plug service connector on MFC socket at the rear of the unit. — ** appears in the display.

2. Select ”Service Program” with the SM key. When the service program is activated the red alarm LED flashes. The LED dis-plays the code number of the selected group and function.

FUNCTION END Jumps to the initial function GR+ Selects group FU+ Selects function in the

activated group OK Activates the selected function or if

necessary skips to the sub-functions with NEXT

Quit the Service Program

1. Press END in the main menu. — A data storage query is acti-vated: ”Save changes? Yes / No”. Y / N terminates the Service Program. Press END to jump to the last function.

2. Switch off unit and remove service connector.

Note

Disconnect the unit from mains for at least 30 seconds after ter-mination of the Service Program (memory is deleted). Then the unit can be switched on again.

Unit Data Software Version Function 100.0

1. Select sub-functions with NEXT.

2. The current software version is displayed in the LCD-display: — User program version with date — Language with text version. Further languages with (+). — Service Program version — Service language with text version

3. Return to initial function with END.

Fig.: 3 — 2

Fig.: 3 — 3

Infusomat® fmS 2.3 gb 3 — 3

Page 34: Infusomat FmS Service Manual

Service Program3

Drug Name Function 110.0

Memory for maximum 10 drugs and 20 characters per name.

1. Display stored drug names with NEXT key.

2. Delete displayed entry with CLR.

3. Press YES to modify a drug name: Move cursor to character with NEXT. Select new character from line 3 with << or >>.

4. Repeat procedure for each character.

5. Return to initial function with END.

Ward Identification Function 120.0

Enter and display of a ward specific unit identification. Permanent display if the unit is connected to mains and switched off.

1. Delete displayed entry with CLR. Press YES to enter modifica-tions: Move cursor to character with NEXT. Select new character from line 3 with << or >>.

2. Repeat procedure for each character.

3. Return to initial function with END.

Serial Number Function 140.0

The displayed serial number must correspond with the number on the unit type plate, as this number is used in interface mode.

1. YES activates the entry mode. Enter via the numeric key-board.

2. YES stores the changed or new number.

3. Return to initial function with END.

DIANET Type Number Function 150.0

The displayed serial number must correspond with the number on the unit type plate, as this number is used in the interface mode.

1. YES activates the entry mode. Enter via the numeric key-board.

2. YES stores the changed or new number.

3. Return to initial function with END.

History Data Operating Hour Counter Function 200.0

1. OK activates the display.

2. Return to initial function with END.

Battery Operating Hours Function 210.0

1. OK activates the display.

Fig.: 3 — 4

3 — 4 Infusomat® fmS 2.3 gb

Page 35: Infusomat FmS Service Manual

Service Program 3

2. Return to initial function with END.

Pump Head Cycles Function 220.0

Display of the pump head cycles (delivered volume).

1. OK activates the display.

2. Return to initial function with END.

Operating Alarms Function 230.0

The last 20 operating alarms can be recalled.

They are displayed as 16 bit binary codes and each bit position can be set from 0 to 1.

1. OK activates the alarm display.

2. Display operating alarms -01 to -20 with the (+) and (-) key.

3. Delete operating alarms with CLR.

4. Return to initial function with END.

Unit alarms will also be displayed from software version IFMe, IFME on

Test Air Inline Sensor Function 300.0

After exchange check the function of the air inline sensor.

See TSI-List for permissible check values (see „Technical Safety Check TSC“ ➪ p. 7 — 1).

1. Press OK. The received signal amplitude is displayed as a measured value. (The test value with NEXT is not important).

2. Insert an infusion line filled with air and check the maximum permissible air value.

3. Insert an infusion line filled with fluid and check the mini-mum permissible water value.

4. Return to initial function with END.

Fig.: 3 — 5

0 Battery empty (battery alarm)

1 Pump cover open

2 Drop alarm

3 Air alarm

4 Occlusion alarm

5 Expired standby time

6 CC alarm (interface)

7 KOR end alarm

8 to 14 free

15 Operating alarm

Bit 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Code 1 0 0 0 0 0 0 0 0 0 0 0 0 0 1 0

Example for alarm ”Pump cover open”

Infusomat® fmS 2.3 gb 3 — 5

Page 36: Infusomat FmS Service Manual

Service Program3

Pressure Sensor Function 310.0

Test Equipment

Calibration gauge 4 mm (see „Test Equipment and Special Tools“ ➪ p. 9 — 1)

1. Push in the bottom slide of the finger pump.

2. Activate function with OK button.

3. Open unit door.

4. Note value measured by the pressure sensor (actual value).

5. Insert the 4 mm calibration gauge and close unit door.

6. The measured value read on the pressure sensor must be 5 to 15 digits above the value noted.

7. Quit function with END.

Note

If the 5 to 15 digits are not reached, the pressure sensor unit must be mechanically adjusted (see „Pressure Sensor“ ➪ p. 4 — 14).

Unit Modifications Service Language Function 400.0

English or German can be selected.

1. OK activates the function.

2. Select language with NEXT.

3. Acknowledge with YES.

4. Return to initial function with END.

User Language Function 410.0

Four user languages per language group are available (depending on software).

1. OK activates the function.

2. Select language with NEXT. The language no. and text version are displayed.

3. Acknowledge with YES.

4. Return to initial function with END.

Alarm Tone Function 420.0

Different alarm modes can be selected:

— Single stage

— A ”10 minutes off” alarm can be selected. In this mode the audible alarm is activated with a delay of 10 minutes.

3 — 6 Infusomat® fmS 2.3 gb

Page 37: Infusomat FmS Service Manual

Service Program 3

The activation of the «10 minutes off” alarm is only permissible, if:

— the staff call is connected and

— the Infusomat® fmS has an attention label (label drawing no. M007100000F04).

1. OK activates the function.

2. Select alarm tone with NEXT.

3. Acknowledge with YES.

4. Return to initial function with END.

Staff Call Function 430.0

Different staff call modes can be selected:

— Dynamic with OFF Alarm

— Dynamic without OFF Alarm

— Static without OFF Alarm

For further details see staff call line in the instructions for use.

An additional switch-on pulse (YES/NO) can be activated for each mode to test the staff call unit.

1. OK activates the function.

2. Select staff call type with NEXT.

3. Acknowledge with YES.

4. Return to initial function with END.

Special Functions Function 440.0

Special functions can be activated in the Service Program, which are then available on the user interface. Deactivated special func-tions will not be displayed. The SM softkey will not be displayed in standard operation, if all special functions are deactivated. — Special functions to be selected, see Fig.: 1 — 2.

1. OK activates the function.

2. Select special functions with NEXT.

3. Activate / deactivate the respective function with YES/NO.

4. Return to main menu with END.

Fig.: 3 — 6

Infusomat® fmS 2.3 gb 3 — 7

Page 38: Infusomat FmS Service Manual

Service Program3

Menu Function 450.0

The availability of menus on the user interface can be set.

— Double rate entry

— Online rate entry

— Interval Bolus

1. OK activates the function.

2. Activate / deactivate the decimal function with NEXT.

3. Acknowledge with YES.

4. Return to main menu with END.

Delivery Rate Function 460.0

The maximum and minimum delivery rates can be set. Range of adjustment of the delivery rate: 0.1 to 999.9 ml/h

1. OK activates the function.

2. Select min./max. delivery rate with NEXT.

3. Acknowledge with YES.

4. Enter value with a numeric key.

5. Acknowledge with OK.

6. Return to initial function with END.

Air Alarm Function 470.0

The air inline sensor sensitivity of the air rate alarm in ml/h (total air alarm) and of the maximum air bubble in ml (single bubble) can be adjusted. Setting range air rate: 0.5 to 3.5 ml/h Setting range air bubble: 0.01 to 0.3 ml/h

1. OK activates the function.

2. Select air rate (ml/h) or air bubble (ml) with NEXT.

3. Acknowledge with YES.

4. Enter value with a numeric key.

5. Acknowledge with OK.

6. Return to initial function with END.

3 — 8 Infusomat® fmS 2.3 gb

Page 39: Infusomat FmS Service Manual

Service Program 3

Dianet Mode Display Function 480.0

When operated with DianetStar the respective DianetStar-mode (CA, CC, CD) with address 01, e.g. mode CA and address 01 is dis-played by: „###CA01###“.

The duration of the display after the last data transmission can be set between 0 and 255 seconds.

Deleting Dose Data Function 490.0

When this function is activated operation can be continued with the dose data of the previous therapy.

Calibration

Note

All safety relevant parameters are set by the manufacturer. If these parameters are changed, a new calibration must be per-formed with calibrated test equipment.

Scale Factor Function 510.0

The scale factor can be set in the limits 40 to 99 digits. Every digit step is equivalent to a 0.5 % modification of the delivery rate. An increase of the scale factor reduces the pump speed, and a de-crease increases the pump speed.

A test infusion line (OIL test infusion line with an Infusomat® fmSadjusted for the Original Infusomat® line, and a Space calibration line with an Infusomat® fmS set for the Infusomat® Space Line is to be used for determining the correction value (see „Test Equip-ment and Special Tools“ ➪ p. 9 — 1).

1. OK activates the function.

2. The value can be changed with the entry keyboard (see „De-livery Accuracy“ ➪ p. 8 — 6).

3. Acknowledge with YES.

4. Return to initial function with END.

5. Quit the Service Program and save changes with YES.

6. Switch on unit and check delivery rate (see „Delivery Accu-racy“ ➪ p. 8 — 6). If necessary repeat the delivery rate measurement.

Infusomat® fmS 2.3 gb 3 — 9

Page 40: Infusomat FmS Service Manual

Service Program3

Air Inline Sensor Function 520.0

Alignment or check of the air inline sensor value (alarm threshold) (see „Air Inline Sensor“ ➪ p. 4 — 17).

1. OK activates the function.

2. Press OK again to activate the air inline sensor value.

3. The value can be changed with the entry keyboard.

4. Acknowledge with YES.

5. AIR SENSOR IS SET acknowledges the entry.

6. Return to initial function with END.

7. Quit the Service Program and save changes with YES.

Occlusion Level Function 540.0

Calibrate (see „Pressure Sensor“ ➪ p. 4 — 14).

Pump Data Function 550.0

Note

Compatibility between controller board and pump must be checked before acknowledgement. Only press YES if pump corre-sponds to controller board.

History Function 560.0 *

The history function can be activated or deactivated.

1. Press OK key twice to activate the function.

2. Select the history function with NEXT.

3. ON activates the function. The history protocol memory is in-itialized (the current software version and the serial number are registered). Press OFF key to deactivate the function. The history protocol memory is deleted.

4. Return to initial function with END.

Note

See instructions for use for detailed information.

* Only software IFME

3 — 10 Infusomat® fmS 2.3 gb

Page 41: Infusomat FmS Service Manual

44 — 20

4 — Unit Elements

4.1 Mains Fuses Designation Ord. No.

Fuse T 0.16 A for 200 / 230 / 240 V (10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2847

Fuse T 0.315 A for 100 / 110 / 120 V (10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 0534

Fuse holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 0979

Note

Only use recommended fuses.

Exchange

1. Press the expansion clamps at the fuse holder on the recessed mains plug with a screw driver in direction of the arrows and pull out fuse holder.

2. Replace blown fuses and press in fuse holder. Only use rec-ommended fuses.

Check

Safety check, functional check.

4.2 Battery Designation Ord. No.

Battery incl. connector 1.2 AH / 7.2 V and holder. . . . 3450 2556

Exchange

1. Switch off unit and disconnect from mains.

2. Loosen screw, open battery compartment cover and remove battery.

3. Pull off battery connector.

4. Assembly is done in reverse order.

5. After having exchanged the battery the Infusomat® fmS must be connected to mains, before switching on the unit. Thereby the charge and discharge currents are aligned.

6. Charge battery (16 h).

Note

Defective batteries must be orderly disposed of, e.g. send back to B. Braun Melsungen AG, Wareneingang.

Check

Perform switch-on test in battery operation and check the battery running time, if necessary.

Fig.: 4 — 1

Battery M4x8

Infusomat® fmS 2.3 gb 4 — 1

Page 42: Infusomat FmS Service Manual

Unit Elements4

4.3 Door Lock Designation Ord. No.

Door lock complete with push button . . . . . . . . . . . . . 3450 5601

Spring holder for door lock . . . . . . . . . . . . . . . . . . . . . . 3450 5440

Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790

Exchange

1. Open door and unlatch the spring holder.

2. Remove countersunk screw and press out the mounting by pressing the holder for the door lock.

3. Remove door lock in an upward direction and exchange.

4. Assembly is done in reverse order.

Check

Pump unit check (only mechanical occlusion pressure).

Fig.: 4 — 2 Spring holder

S M 3×6

4 — 2 Infusomat® fmS 2.3 gb

Page 43: Infusomat FmS Service Manual

Unit Elements 4

4.4 Pump Cover Designation Ord. No.

Pump cover with lock . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5717

Blind plug 7.1 mm (10 pcs.) . . . . . . . . . . . . . . . . . . . . . 3477 3207

Torsion spring in lever / pump cover (5 pcs.) . . . . . . . 3477 3363

Torsion spring for pump cover (5 pcs.) . . . . . . . . . . . . 3477 3355

Lever (pump cover). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4092

Hinge pin for pump cover . . . . . . . . . . . . . . . . . . . . . . . 3477 3967

Hinge pin for pump cover lever. . . . . . . . . . . . . . . . . . . 3450 5725

Exchange

Tools: Pin punch 1.8 mm, pin punch 6 mm, 4 mm gauge

1. Open door and remove hinge pin with pin punch (1.8 mm) from below. Do not lose torsion spring.

2. Disassemble pump cover.

3. Insert torsion spring in new pump cover. Insert hinge pin with pin punch (6 mm) from above.

4. Check pressure sensor with 4mm gauge in the Service Pro-gram, and calibrate, if necessary.

Check

Safety check, pump unit check.

Fig.: 4 — 3

Hinge pin forpump cover lever

Hinge pin forpump cover

Torsion springPump cover

Infusomat® fmS 2.3 gb 4 — 3

Page 44: Infusomat FmS Service Manual

Unit Elements4

4.5 Housing and Handle Designation Ord. No.

Housing labelling German . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1843 French . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2130 Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2149 Italian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2157 English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2165 Spanish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2173 Danish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2181 Norwegian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2190 Swedish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2203 Finnish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2211 Portuguese . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2220 Czech . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2238 Polish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2246 Turkish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2459

Foot stand complete with rubber feet . . . . . . . . . . . . . 3450 5415

Rubber feet (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3096

Unit handle with O-rings and PT screws . . . . . . . . . . . 3450 1789

Exchange

1. Remove battery (see „Battery“ ➪ p. 4 — 1).

2. Loosen 4 screws from the foot stands and remove foot stands.

3. Remove safety seal from rear panel, break tamper-proof cap and remove screw.

4. Slidly widen the sides at the bottom of the housing and pull off to the top.

5. Assembly is done in reverse order. To do so, place unit on front side.

6. Safety seal the rear panel screw after functional check.

Check

Safety check.

Fig.: 4 — 4

PT5.0x16

4 x M4x8

O-ring

4 — 4 Infusomat® fmS 2.3 gb

Page 45: Infusomat FmS Service Manual

Unit Elements 4

4.6 Controller Board Designation Ord. No.

Distance sleeve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3366

Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3447

Loudspeaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8848

Controller board with buzzer, raw material no. 3810 7651, soft-ware IFMC, can be updated to software IFMe:

Controller board with loudspeaker, volume control and history function, raw material no. 3810 7996, software IFME:

The controller boards (raw material no. 3810 7651 and 3810 7996) are completely compatible.

This allows older units to be upgraded with the new features (loudspeaker, volume control, and history function) without any problems. In this case the controller board must be ordered as new part.

Exchange

1. Remove battery (see „Battery“ ➪ p. 4 — 1).

2. Dismount cover (see „Housing and Handle“ ➪ p. 4 — 4).

3. Press snap-in pin together at the distance sleeves and remove board carefully.

4. Pull off connector (see figure).

5. Exchange board or buzzer.

Fig.: 4 — 5

Flexible cable LCD module Pressure sensor

Reed sensor

Press snap-in pin togetherLoudspeaker / Buzzer

Battery connec

Rear panel

Air inline sensor connectorCable outlet motorMotor connector

connector

connector

connector

Language Group* New Part Exchange

A Ord. No.:3450 1967 Ord.No.:3488 0844

B Ord. No.:3450 1975 Ord.No.:3488 0852

C Ord. No.:3450 1983 Ord.No.:3488 0860

D Ord. No.:3450 1991 Ord.No.:3488 0879

E Ord. No.:3450 2033 Ord.No.:3488 0887

* Language Group: A German, French, Dutch, Italian B English, Dutch, Spanish, Castellano C Danish, Norwegian, Swedish, Finnish D Spanish, Portuguese, English, Turkish E Czech, Polish, German, English

Language Group* New Part Exchange

A Ord. No.:3450 8759 Ord. No.:3488 1019

B Ord. No.:3450 8708 Ord. No.:3488 1190

C Ord. No.:3450 8716 Ord. No.:3488 1204

D Ord. No.:3450 8724 Ord. No.:3488 1212

E Ord. No.:3450 8732 Ord. No.:3488 1240

Infusomat® fmS 2.3 gb 4 — 5

Page 46: Infusomat FmS Service Manual

Unit Elements4

6. Assembly is done in reverse order. Be careful with the optical components. Insert board in the lower guide parallel to the base plate. (Otherwise problems with the optical interface can occur.)

Note

If ”Calibration Defective” is displayed after having exchanged the controller board, check whether the correct board (risk of mix-up) was assembled.

Check

Safety check, functional check.

4.7 Rear Panel Designation Ord. No.

Rear panel with screws (M3) and seal . . . . . . . . . . . . . 3450 1860

Cover for optical interface. . . . . . . . . . . . . . . . . . . . . . . 3477 3164

Strip seal for rear panel . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3142

MFC connector board . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3374

Potential equalization bolt . . . . . . . . . . . . . . . . . . . . . . 3477 0550

fm recessed plug (3 pin) . . . . . . . . . . . . . . . . . . . . . . . . 3477 3177

Screw 30×8 for fm recessed plug (20 pcs.) . . . . . . . . . 3477 3185

Plain washer 3.2 (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . 3477 3193

Power supply module 200/230/240 V. . . . . . . . . . . . . . 3450 1886

Power supply module 100/110/120 V. . . . . . . . . . . . . . 3450 1894

Power supply module 230 V . . . . . . . . . . . . . . . . . . . . . 3450 1908

Drop sensor socket incl. cable and plug . . . . . . . . . . . . 3450 1878

Fig.: 4 — 6

4 — 6 Infusomat® fmS 2.3 gb

Page 47: Infusomat FmS Service Manual

Unit Elements 4

Fig.: 4 — 7

Pin No.

Signal Name Function

Pin 1 Uext- Input of external supply voltage, connection of shield

Pin 2 not assigned

Pin 3 GND Reference level 0V

Pin 4 Staff call Output open collector or standard 74HC level each with 220 -series resistance

Pin 5 Ub Output supply voltage

Pin 6 MFC-KAD Analog input K

Pin 7 Emergency Off Input interface

Pin 8 MFC-FAD Analog input F

Pin 9 Tx Transmit line interface

Pin 10 STB Output with 470 -series resist-ance, activation of MFC circuit

Pin 11 Rx Receive line interface

Pin 12 Uext+ Input of external supply voltage

Table 4 — 1 MFC pin assignment

Infusomat® fmS 2.3 gb 4 — 7

Page 48: Infusomat FmS Service Manual

Unit Elements4

Exchange Rear Panel

1. Remove battery (see „Battery“ ➪ p. 4 — 1).

2. Dismount cover (see „Housing and Handle“ ➪ p. 4 — 4).

3. Pull off rear panel connectors and loosen both screws on the unit bottom.

Fig.: 4 — 8

Rear panel

M4x8

4 — 8 Infusomat® fmS 2.3 gb

Page 49: Infusomat FmS Service Manual

Unit Elements 4

Exchange MFC Connector Board

Tools: Special socket spanner M18

1. Remove nuts at the fm recessed plug (red/blue).

2. Pull off connector to the mains power supply and drop sensor, see Fig.: 4 — 6.

3. Loosen MFC socket with special socket spanner M18.

4. Assembly is done in reverse order.

Exchange fm Recessed Plug

1. Loosen screws and nuts.

Note

The new recessed plug must be slightly moveable after assembly.

Exchange Potential Equalization Bolt

1. Remove nuts with ring spanner.

2. Exchange bolt.

Fig.: 4 — 9

M18

Fig.: 4 — 10

PT 3×8

2x M4 each2x M6

Cable, red Cable, blue

Infusomat® fmS 2.3 gb 4 — 9

Page 50: Infusomat FmS Service Manual

Unit Elements4

Exchange Mains Module

1. Loosen 3 screws and remove the power supply unit.

Note

The voltage selection for switchable power supplies is at the fuse element.

Exchange Drop Sensor Socket

1. Loosen nut and exchange drop sensor socket.

2. Safety lock with Sicomet 50.

3. Make ground connection.

Check

Safety check, functional check.Fig.: 4 — 11

Power supply

3 X Torx M3x10

4 — 10 Infusomat® fmS 2.3 gb

Page 51: Infusomat FmS Service Manual

Unit Elements 4

4.8 Front Frame Designation Ord. No.

Front frame without flow inhibitor and pressure spring . . . 3450 5822

Circular seal 571 mm / 45 mm . . . . . . . . . . . . . . . . . . . 3477 3126

Flow inhibitor with pressure spring . . . . . . . . . . . . . . . 3477 3258

Pressure spring for flow inhibitor (5 pcs.) . . . . . . . . . . 3477 3266

Tamper-proof caps 10 mm (50 pcs.) . . . . . . . . . . . . . . 3477 3134

Reed sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1754

Exchange

Tools: Pin punch

1. Remove battery (see „Battery“ ➪ p. 4 — 1).

2. Dismount cover (see „Housing and Handle“ ➪ p. 4 — 4).

3. Disassemble pump cover (see „Pump Cover“ ➪ p. 4 — 3).

4. Loosen countersunk screws by piercing a screwdriver through the caps and remove caps.

Note

Do not loosen the assembly screw on the lower pressure sensor board! Otherwise the pump has to be recalibrated with a 4mm gauge.

5. Press Reed sensor out of holder on the rear side, see Fig.: 4 — 13.

6. Press both snap-in pins inwards (rear side top and bottom). Then remove front frame to the front.

7. Remove flow inhibitor with pressure spring and Reed sensors and assemble in new front frame.

8. Insert the seal. Assembly is done in reverse order. Snap-in pins manually. Insert Reed sensor and fasten cables firmly with cable ties.

Check

Safety check, pump unit check.

Fig.: 4 — 12

Snap-in pin(rear side)

Reed sensor(rear side)

Snap-in pin(rear side)

Infusomat® fmS 2.3 gb 4 — 11

Page 52: Infusomat FmS Service Manual

Unit Elements4

4.9 Pump Unit Designation Ord. No.

Finger pump (without motor) including pump, . . . . . 3450 1738 pump cover, seal membrane and boards

Finger pump (without motor and board) . . . . . . . . . . . 3450 9038 incl. pump, pump cover, and seal membrane

Membrane incl. seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5733

Motor with pinion for finger pump. . . . . . . . . . . . . . . . 3450 1924

Exchange

1. Remove battery (see „Battery“ ➪ p. 4 — 1).

2. Dismount cover (see „Housing and Handle“ ➪ p. 4 — 4).

3. Disassemble front frame (see „Front Frame“ ➪ p. 4 — 11).

4. Loosen motor screws, pull off motor connector and remove motor from pump unit.

Note

Do not loosen the assembly screw on the lower pressure sensor board! Otherwise the pump has to be recalibrated with a 4mm gauge.

5. Pull off connector from controller board.

6. Press down snap-in pins (from the inside) and tilt pump for-ward and unhinge.

7. Assembly is done in reverse order.

Note

Be sure that the cable of the Reed sensor and air inline sensor do not hinder the function of the pressure sensor and the monitoring of the motor speed (slot disc).

The complete pump unit (pump without motor, pump cover, membrane, boards) was calibrated by B.Braun. After a complete exchange the unit data and user data must be entered in the Serv-ice Program again as the data memory is on the pump board:

8. Enter unit and user data in the EEPROM.

— Serial number according to the type plate

— DIANET type number according to the type plate

If necessary enter:

— Drug name

— Ward identification

— Alarm tone

— Delivery rate min./max.

Fig.: 4 — 13

Cable tie

Motor

M3x6

Reed sensor

Pump unit

Snap-in pin

4 — 12 Infusomat® fmS 2.3 gb

Page 53: Infusomat FmS Service Manual

Unit Elements 4

— Air alarm: air bubbles in ml and air rate in ml/h respectively

— User language

— Special functions (ON/OFF)

— Menu

— Staff call type

Note

If data is not entered, ”Calibration data faulty” may be displayed after the unit is switched on again.

9. After ending the Service Program save the data.

The counters for operating hours, battery operation and pump head cycles are reset to zero when the pump unit is exchanged. If ”Calibration data faulty” is displayed, select the Service Program and quit with ”Save? Yes”.

Check

Safety check, pump unit check.

Infusomat® fmS 2.3 gb 4 — 13

Page 54: Infusomat FmS Service Manual

Unit Elements4

4.10 Pressure Sensor Test Equipment Ord. No.

Pressure calibration device . . . . . . . . . . . . . . . . . . . . . . 0770 5018

Calibration gauge 4 mm (for adjustment after . . . . . . 0770 1489 exchange of the pump cover)

Designation Ord. No.

Pressure sensor (cmpl.). . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7353 (only delivered when a pressure calibration device is available)

Exchange

CAUTION

The pressure sensor unit is mechanically very sensitive. It must not be dismounted, which means that the pressure sensor board and the spring element must not be detached from the pressure sensor support (plastic part). Make sure that the movement of the coil core inside the coil is not hindered.

The pressure sensor consists of the light barrier board with EEP-ROM and the pressure sensor board with holder. Coil core and bending element are screwed to this holder. Both boards are con-nected via a flat cable.

1. Dismount rear panel (see „Rear Panel“ ➪ p. 4 — 6) or pump unit (see „Pump Unit“ ➪ p. 4 — 12).

2. Pull off the connecting cable to the microprocessor board.

3. Loosen and remove screw (1) of the light barrier board.

4. Loosen screw (2) of the pressure sensor and remove it togeth-er with plain washer and serrated lock washer. Then remove pressure sensor with pressure sensor board and light barrier board.

5. Fit pressure sensor with pressure sensor- and light barrier board.

Note

Make sure that the pressure sensor slide can smoothly run in the guides and that the coil core can move inside the coil.

6. Calibrate pressure sensor:

a) Switch on unit and call in the pressure sensor reading in the Service Program (see „Pressure Sensor Function 310.0“ ➪ p. 3 — 6)

b) Loosen screw (2) on the side of the pressure sensor unit (2.5 mm Allen key).

Fig.: 4 — 14

Pressure sensor with

Light barrier

Screw (1)

Screw (2)

board

pressure sensor board

Serrated lock washer

Plainwasher

Do not loosen screw

4 — 14 Infusomat® fmS 2.3 gb

Page 55: Infusomat FmS Service Manual

Unit Elements 4

c) Push pressure sensor with pressure sensor board to the rear until stopper.

d) Move lower slide of the finger pump rearwards.

e) Note down pressure sensor value.

f) Insert 4 mm calibration gauge.

g) Push pressure sensor with pressure sensor board slightly forward.

h) Tighten screw (2). The measured value read on the pressure sensor must be 5 to 15 digits above the value noted.

Note

Pull nut slightly out of the pressure sensor housing.

i) After calibration return to the initial function with END.

j) Select SAVE NO. (Do not actuate the YES key.)

k) Switch off unit.

7. Assembly is done in reverse order.

8. Check the user data and reenter, if necessary, as data was saved in the EEPROM of the exchanged light barrier board.

9. Calibrate pressure sensor:

Note

Calibration weights are required to calibrate the pressure sensor. An alignment with infusion lines is not permitted.

a) Clean the pump front side.

b) Place the Infusomat® fmS without mains lead and drop sensor and the service connector inserted horizontally (front facing upwards) in the cellular packing of the cal-ibration device.

c) Switch on unit and call in the pressure sensor reading in the Service Program (see „Pressure Sensor Function 310.0“ ➪ p. 3 — 6).

d) Open pump cover.

e) Fasten holders for the calibration weights under the pump cover and let snap in at the locking bow.

f) Position weight 1 (100 g) for 400 mbar carefully in the holder on the pressure sensor. The weight must be placed without any friction in the holder.

g) Note down measured value.

h) Repeat steps f) and g) with weight 2 (128 g) for 800 mbar and weight 3 (166 g) for 1200 mbar.

Infusomat® fmS 2.3 gb 4 — 15

Page 56: Infusomat FmS Service Manual

Unit Elements4

i) Remove weight and holder and do not save data.

j) Call in calibration of pressure stage (function 540) in the Service Program.

k) Enter the calibration values via the keyboard and ac-knowledge with ”yes”.

l) Quit Service Program and save data.

m) Register the changed pressure values in the unit book.

n) Check the electronic occlusion pressure (see „Pressure Cut-Off, Electronic“ ➪ p. 8 — 5).

Check

Safety check, functional check.

4 — 16 Infusomat® fmS 2.3 gb

Page 57: Infusomat FmS Service Manual

Unit Elements 4

4.11 Air Inline Sensor Designation Ord. No.

Air inline sensor incl. connector . . . . . . . . . . . . . . . . . . 3450 193A

Exchange

1. Remove battery (see „Battery“ ➪ p. 4 — 1).

2. Dismount cover (see „Housing and Handle“ ➪ p. 4 — 4).

3. Dismount controller board (see „Controller Board“ ➪ p. 4 — 5).

4. Disassemble front frame (see „Front Frame“ ➪ p. 4 — 11).

5. Press the complete sensor from behind and out of the frame.

6. Insert new sensor and fasten cables firmly with cable ties.

Note

Shielded air inline sensors are installed from serial no. 34504 on.

Note

Be sure that the cables of the Reed sensor and air inline sensor do not hinder the function of the pressure sensor and the monitoring of the motor speed (slot disc).

Check

After exchange of the air inline sensor, please check:

— Air value

— Water value

— Calibration value (alarm threshold), adjust if necessary

Safety check, functional check.

Fig.: 4 — 15

Air inline sensor

Infusomat® fmS 2.3 gb 4 — 17

Page 58: Infusomat FmS Service Manual

Unit Elements4

4.12 Operating Unit Designation Ord. No.

Membrane keyboard with support plate and seal . . . . 3450 1797

LCD module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1819

Frame incl. pressure spring and magnet . . . . . . . . . . . 3450 1827

Round cord 42 mm (5 pcs.) . . . . . . . . . . . . . . . . . . . . . 3477 3347

Circular seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3126

Hinge unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5571

Hinge pin (3 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5580

Magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5849

Flex cable, preformed . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8830

Exchange

1. Remove battery (see „Battery“ ➪ p. 4 — 1).

2. Disassemble door lock.

3. Loosen countersunk screw and bridge.

4. Remove tamper-proof caps (6 pieces) on the door frame by piercing a screwdriver through the caps and loosen counter-sunk screws.

5. Unlatch zero insertion force connector and loosen flex cable.

Note

The position of the flex cable must not be changed, i.e. the pre-formed section must be in the hinge area (pivot). Mark cable, if necessary.

6. Disassemble either LCD module or support plate with mem-brane keyboard or door hinge pins respectively and exchange door frame.

7. Assembly is done in reverse order. Pay attention to the correct direction of the door hinge pin during assembly.

Note

Do not kink either of the flex cables. Push the contacts to the stopper of the zero insertion force connector and lock in same po-sition.

Check

Electrical safety, functional check, pump unit check.

Fig.: 4 — 16

Support plate Zero insertion

Cable is preformed

LCD module

force connector

Fig.: 4 — 17

Frame operating unit

Hinge pin

6 x PT2.5×14

4 — 18 Infusomat® fmS 2.3 gb

Page 59: Infusomat FmS Service Manual

Unit Elements 4

4.13 Barcode Label Designation Ord. No.

Barcode label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 9070

Exchange

1. Remove old barcode label, if existing.

2. Clean adhesion surface with an alcoholic cleaning agent and let dry.

3. Loosen barcode label from the base material and stick it on.

Note

Destroy the type plate delivered.

Check

Check that serial number and pump symbol in the plain text field of the barcode label correspond with the type plate on the pump of the Infusomat® fmS.

4.14 Frame with Seal Designation Ord. No.

Seal plate between frame and front frame (exchange not recommended) . . . . . . . . . . . . . . . . 3477 3240

Frame with seal plate . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1762

Exchange

1. Disassemble all parts as described before and exchange frame with seal plate.

Check

Electrical safety, functional check, pump unit check.

Fig.: 4 — 18

Barcode label

Infusomat® fmS 2.3 gb 4 — 19

Page 60: Infusomat FmS Service Manual

Unit Elements4

For your notes:

4 — 20 Infusomat® fmS 2.3 gb

Page 61: Infusomat FmS Service Manual

55 — 2

5 — Checks after Repair

Depending on the work carried out, perform the relevant check blocks (1., 2. and / or 3.).

1. Visual Inspection 2. Electrical Safetyaccording to IEC / EN 60601-1 or VDE 0750 and VDE 0751

3. Functional Inspection

❒ Infusomat® fmS: Cleanliness, completeness, damage and faults affecting safety Particularly:❒ Damage to and readability of the

labels❒ Completeness, damage and

readability of the additional labels on an Infusomat® fmS set for the Infusomat® Space Line

❒ Pump sealing diaphragm❒ Membrane keyboard❒ Rubber feet❒ Operating unit❒ Locking mechanism of the pump

cover❒ Flow inhibitor❒ Mains plug connector❒ MFC connector incl. MFC cable❒ Check voltage values

100/110/120 V = T 0.315 A 200/230/240 V = T 0.16 A

❒ Mains cable❒ Drop sensor with cable and plug

connector

❒ Mains voltage ______ V AC❒ Protective conductor

resistance incl. mains cable < 0.2 _______

❒ Insulation resistance >> 2 M _________

❒ Earth leakage current at NC 30 A ________ A

Switch on unit❒ Self-test❒ Audible alarm❒ Status displays❒ LCD lighting❒ Display (all symbols and pixels) in the

LCD display

Operation❒ Opening and closing mechanism of the

door❒ Functions of the buttons❒ Enter delivery rate, volume and time.

The values displayed in the LCD display correspond to the values entered.

❒ Staff call Test incl. MFC staff call line, if used

❒ Pump cover alarm❒ Alarm suppression

Pressure cut-off, electronic (Reference values measured with electronic pressure sensor) Delivery rate: 50 ml/h Volume: 250 ml Test without drop sensor

Note

Depending on the setting, the measurement should be performed only with one of the two lines.

❒ With original Infusomat® line❒ Pressure level, low❒ Pressure level, middle❒ Pressure level, high

❒ With Infusomat® Space line❒ Pressure level, low❒ Pressure level, middle ❒ Pressure level, high

(Part 1 of 2)

Infusomat® fmS 2.3 gb 5 — 1

Page 62: Infusomat FmS Service Manual

Checks after Repair5

Observe the procedure information (see „Procedural Instructions on the TSC“ ➪ p. 8 — 1)!

Pressure limitation, mechanical (Reference values measured with electronic pressure sensor)❒ With original Infusomat® line

❒ Pmax ❒ Pmin

❒ With Infusomat® Space line❒ Pmax ❒ Value Pmax by min. 0.10 bar higher

than pressure level, high, electronic❒ Pmin

Safety clamp (flow inhibitor)❒ Pmin

Delivery accuracy Ambient temperature 20 … 25 °C Delivery rate: 250 ml/h Measured volume: 25 ml❒ Divergence

Drop sensor Delivery rate: 400 ml/h❒ Drop sensor alarm, occlusion ❒ Drop sensor alarm, flow

Air inline sensor Delivery rate: 250 ml/h❒ Water value ❒ Air alarm❒ Air value ❒ Threshold value

Battery testOperation mains/battery/mains:❒ Uninterrupted operationSwitch on unit without mains connection❒ Self-test is executed

1. Visual Inspection 2. Electrical Safetyaccording to IEC / EN 60601-1 or VDE 0750 and VDE 0751

3. Functional Inspection

(Part 2 of 2)

5 — 2 Infusomat® fmS 2.3 gb

Page 63: Infusomat FmS Service Manual

66 — 2

6 — Maintenance

It is recommended every 2 years. In addition to the technical safe-ty inspection, the following assemblies/components are to be checked:

1. Check rubber feet and if necessary exchange.

2. Check easy running of the pump cover, lock mechanism and door.

3. Check easy running of the flow inhibitor, clean and if neces-sary exchange pressure springs.

4. Check seal membrane and if necessary exchange.

5. Check the drop sensor optics and spring mechanics and clean, if necessary.

6. Open unit. Internal visual inspection. Clean seal surfaces and if necessary exchange seal strip.

7. Check mechanical occlusion pressure and if necessary cali-brate.

8. Check electronic occlusion pressure and if necessary calibrate (see „Pressure Sensor“ ➪ p. 4 — 14).

9. Assemble and seal unit ready for operation.

Infusomat® fmS 2.3 gb 6 — 1

Page 64: Infusomat FmS Service Manual

Maintenance6

For your notes:

6 — 2 Infusomat® fmS 2.3 gb

Page 65: Infusomat FmS Service Manual

7 — 4

7 — Technical Safety Check TSC

Index d(Master – to be added to the documentation)

Checklist for Technical Safety Check – Every 24 Months

Unit: Infusomat® fmSManufacturer: B. Braun Melsungen AG

User

Observe the Service Manual and the instructions for use. All measured values are to be documented. Accessories used should be included in testing. Make exclusive use of calibrated measuring equip-ment.

Article No. Unit No. Year of Procurement Stock No.

Visual Inspection Electrical Safetyaccording to IEC / EN 60601-1 or VDE 0750 and VDE 0751

Functional Inspection

❒ Infusomat® fmS: Cleanliness, completeness, damage and faults affecting safety Particularly:❒ Damage to and readability of the

labels❒ Completeness, damage and

readability of the additional labels on an Infusomat® fmS set for the Infusomat® Space Line

❒ Pump sealing diaphragm❒ Membrane keyboard❒ Rubber feet❒ Operating unit❒ Locking mechanism of the pump

cover❒ Flow inhibitor❒ Mains plug connector❒ MFC connector incl. MFC cable❒ Check voltage values

100/110/120 V = T 0.315 A 200/230/240 V = T 0.16 A

❒ Mains cable❒ Drop sensor with cable and plug

connector

❒ Mains voltage ______ V AC❒ Protective conductor resistance

incl. mains cable < 0.2 _______

❒ Insulation resistance >> 2 M _________

❒ Earth leakage current at NC 30 A ________ A

Switch on unit❒ Self-test❒ Audible alarm❒ Status displays❒ LCD lighting❒ Display (all symbols and pixels) in the

LCD display

Operation❒ Opening and closing mechanism of the

door❒ Functions of the buttons❒ Enter delivery rate, volume and time.

The values displayed in the LCD display correspond to the values entered.

❒ Staff call Test incl. MFC staff call line, if used

❒ Pump cover alarm❒ Alarm suppression

(Part 1 of 3)

Infusomat® fmS, 2.3 gb 7 — 1Sheet 1 of 4

M670 00 00 92 F04 C 3891 2189 (GB)

Page 66: Infusomat FmS Service Manual

Index d

(Master – to be added to the documentation)

Technical Safety Check TSC

Pressure cut-off, electronic (Reference values measured with electronic pressure sensor) Delivery rate: 50 ml/h Volume: 250 ml Test without drop sensor

Note

Depending on the setting, the measurement should be performed only with one of the two lines.

❒ With original Infusomat® line❒ Pressure level, low

(0,25 … 0.65 bar) ________ bar❒ Pressure level, middle

(0,55 … 0.95 bar) ________ bar❒ Pressure level, high

(0,90 … 1.30 bar) ________ bar❒ With Infusomat® Space line

❒ Pressure level, low (0,21 … 0.61 bar) ________ bar

❒ Pressure level, middle (0,48 … 0.88 bar) ________ bar

❒ Pressure level, high (0,83 … 1.23 bar) ________ bar

Pressure limitation, mechanical (Reference values measured with electronic pressure sensor)❒ With original Infusomat® line

❒ Pmax (1.35 … 1.76 bar) ______ bar❒ Pmin (> 0.95 bar) _______ bar

❒ With Infusomat® Space line❒ Pmax (1.30 … 1.70 bar) ______ bar❒ Value Pmax by min. 0.10 bar higher

than pressure level, high, electronic❒ Pmin (> 0.83 bar) _______ bar

Safety clamp (flow inhibitor)❒ Pmin (> 0.40 bar) _______ bar

Visual Inspection Electrical Safetyaccording to IEC / EN 60601-1 or VDE 0750 and VDE 0751

Functional Inspection

(Part 2 of 3)

M670 00 00 92 F04 C 3891 2189 (GB)

7 — 2 Infusomat® fmS, 2.3 gbSheet 2 of 4

Page 67: Infusomat FmS Service Manual

Technical Safety Check TSC

Index d

(Master – to be added to the documentation)

Delivery accuracy Ambient temperature 20 … 25 °C Delivery rate: 250 ml/h Measured volume: 25 ml❒ Divergence (± 5%) _______ %

Drop sensor Delivery rate: 400 ml/h❒ Drop sensor alarm, occlusion ______ s

(< 5s)❒ Drop sensor alarm, flow

Air inline sensor Delivery rate: 250 ml/h❒ Water value (> 481 mV) _______ mV❒ Air alarm❒ Air value (< 78 mV) _______ mV❒ Threshold value (< 182 mV) _____ mV

Battery testOperation mains/battery/mains:❒ Uninterrupted operationSwitch on unit without mains connection❒ Self-test is executed

Visual Inspection Electrical Safetyaccording to IEC / EN 60601-1 or VDE 0750 and VDE 0751

Functional Inspection

(Part 3 of 3)

Infusomat® fmS, 2.3 gb 7 — 3Sheet 3 of 4

M670 00 00 92 F04 C 3891 2189 (GB)

Page 68: Infusomat FmS Service Manual

Index d

(Master – to be added to the documentation)

Technical Safety Check TSC

Lines Used Accessories Used

❒ Original Infusomat line (OIL)❒ Infusomat® Space Line❒ Manufacturer: __________________

Type: __________________________❒ Manufacturer: __________________

Type: __________________________

❒ MFC staff call line❒ ____________________________❒ ____________________________❒ ____________________________❒ ____________________________

Calibrated for usage of:❒ Original Infusomat Line (OIL)❒ Infusomat® Space Line

(Note the stickers and retrofit, if necessary)

Test result: Defects found which could endanger patients, users or third parties: Yes No

Inspection performed by:

Measures to be taken: Repair ___________________

Unit handed over on:

Special features / documentation:

To:

Date / Signature:

Next deadline for TSC:

M670 00 00 92 F04 C 3891 2189 (GB)

7 — 4 Infusomat® fmS, 2.3 gbSheet 4 of 4

Page 69: Infusomat FmS Service Manual

88 — 8

8 — Procedural Instructions on the TSC

Visual Inspection 1. Check the Infusomat® fmS and accessories for cleanliness.

2. Check the Infusomat® fmS and accessories for completeness and check configuration. Pay special attention to the follow-ing parts:

a) Check the special additional labels of the Infusomat® fmS adjusted for the Infusomat® Space Line, please see Fig.: 8 — 1.

3. Check the Infusomat® fmS and its accessories for damage and the labels for readability. Pay special attention to the fol-lowing parts:

a) Pump sealing diaphragm

b) Membrane keyboard

c) Rubber feet

d) Operating unit

e) Lock mechanism of the pump cover

f) Flow inhibitor

g) Mains plug connector

h) MFC plug connector

i) Check fuse values A fuse with T 0.315 A must be installed for devices for 100/110/120 V. A fuse with T 0.16 A must be installed for devices for 200/230/240 V.

Fig.: 8 — 1

Infusomat® fmS 2.3 gb 8 — 1

Page 70: Infusomat FmS Service Manual

Procedural Instructions on the TSC8

Electrical Safety according to IEC / EN 60601-1 or VDE 0750 and VDE 0751

The values to be measured are specified in the TSC (see „Techni-cal Safety Check TSC“ ➪ p. 7 — 1). The measured values are to be recorded.

1. Disconnect all connection and interface cables from the de-vice.

Protective Conductor Resistance

1. Measure between the protective conductor of the mains ca-ble and the following parts:

— Potential equalization bolt

— Bolt for door lock

— Left-hand contact (when looking at the connector) of the FM connector

— Unit housing:

a) If the unit is not sealed countersunk screw at the rear of the unit.

b) If the unit is sealed, remove lacquer from one of the holes in the foot stands.

c) Document largest value.

Note

Do not use the foot stand assembly screws as alternative meas-urement points.

Insulation Resistance

1. Measurement with 500 V between the short-circuited mains connectors and the potential equalization bolt.

Earth Leakage Current

1. Measure the earth leakage current without single fault con-dition (NC = Normal Condition) incl. mains cable.

2. Measure the earth leakage current without single fault con-dition (NC = Normal Condition) incl. mains cable with re-versed polarity.

3. Document largest value.

8 — 2 Infusomat® fmS 2.3 gb

Page 71: Infusomat FmS Service Manual

Procedural Instructions on the TSC 8

Functional Inspection Switch on Unit

1. Switch on unit and check the following details:

— Self-test

— Audible alarm One short and afterwards two short tones

— Status displays LEDs (red, green) light up for a short moment

— LCD lighting

— Display on the LC display All symbols and display pixels flash for a short moment Fig.: 8 — 2. Then „000.0ml/h“ is displayed

Operation

1. Open door. Pump cover must automatically open when the unit door is opened.

2. Insert the Original Infusomat Line (OIL). If the unit is designed for an Infusomat® Space Line, then insert an Infusomat® Space Line.

3. Close door. The door must lock smoothly and close correctly at the top and bottom.

4. Carry out infusion with intermediate bolus and press all but-tons at least once. Infusion and bolus are performed and all buttons trigger the function desired.

5. Plug MFC service connector on the MFC socket.

6. Trigger a staff call (e.g. open pump cover during operation). The red LED in the MFC service connector lights up. — if ”dynamic” is set, 1 sec. — if ”static” is set, until the alarm is acknowledged. Check the MFC staff call line if installed.

7. Press the alarm button. The current alarm is suppressed for 2 minutes.

Fig.: 8 — 2

Infusomat® fmS 2.3 gb 8 — 3

Page 72: Infusomat FmS Service Manual

Procedural Instructions on the TSC8

Test Setup

Perform test setup with the subassemblies listed below, please see also Fig.: 8 — 3:

An electronic manometer should be used for the measurement described hereafter.

If a mechanic manometer is used instead values which are approx. 100 mbar lower are to be expected. Proceed as follows:

Note

The liquid level in the container must be approx. 80 cm above the lower edge of the device.

— Original Infusomat® Line (new, can be used for the complete TSC incl. functional check) (1 piece)

or

— Infusomat® Space Line (new, can be used for the complete TSC incl. functional check) (1 piece)

— Infusion bag or bottle, min. 500 ml (1 piece)

— Three-way valve (2 pieces for measurement with electronic manometer)

— 10 ml syringe (air buffer for measurement with electronic manometer) (syringe drawn up with air to 10 ml and piston fixed mechan-ically) (1 piece)

— 1 ml syringe for air bubble injection (1 piece)

— Electronic manometer with peak value recognition (1 piece)

— Pressure cell (option of electronic manometer) Diameter 100 mm / 160 mm, measuring range 2.5 bar / 4 bar (1 piece)

— Graduated cylinder 25 ml, +/- 0.4 ml (1 piece)

Fig.: 8 — 3 12

10 ml air

8 — 4 Infusomat® fmS 2.3 gb

Page 73: Infusomat FmS Service Manual

Procedural Instructions on the TSC 8

Pressure Cut-Off, Electronic

The MFC service connector must not be plugged. The drop sensor must not be connected.

1. Insert the line of the test setup in the device.

2. Enter a delivery rate according to the TSC.

3. Select pressure stage according to the TSC.

4. Vent test setup, position of the three-way cocks please see Fig.: 8 — 4.

5. Start infusion and deliver first of all in to an open system (without manometer).

6. Switch over three-way cocks, please see Fig.: 8 — 5, and de-liver against the manometer. Read off maximum value on the pressure gauge upon an alarm and before an automatic pressure reduction and com-pare with the specifications in the TSC.

7. Check all pressure stages listed in the TSC and document val-ues.

Pressure Limitation, Mechanical

The drop sensor must not be connected.

1. Plug MFC service connector on the MFC plug connector.

2. Activate the occlusion pressure mechanical (see „Disabling the Pressure Sensor (Occlusion Pressure Mechanical)“ ➪ p. 3 — 2).

3. Enter a delivery rate according to the TSC.

4. Start infusion and deliver first of all in to an open system (without manometer).

5. Deliver against the manometer. Read off the corresponding value on the manometer and compare with the specifications of the TSC for Pmax and Pmin.

6. Document values. Do not reduce pressure.

Safety Clamp (Flow Inhibitor)

1. Stop infusion from the pressure limitation test mechanically and unlock the operating unit, read off value on the manometer and compare with the spec-ifications in the TSC.

Fig.: 8 — 4 12

Fig.: 8 — 5

12

Infusomat® fmS 2.3 gb 8 — 5

Page 74: Infusomat FmS Service Manual

Procedural Instructions on the TSC8

2. Document value.

3. Pull off the MFC service connector.

WARNING

SWITCH THE DEVICE OFF AFTER THE SAFETY CLAMP WAS CHECKED SO THAT THE ELECTRONIC PRESSURE CONTROL IS RE-ACTIVATED AFTER A RESTART.

4. Switch device off.

Delivery Accuracy

Requirements:

— Mount test setup as shown in Fig.: 8 — 6.

— Ambient temperature and delivery rate according to the TSC.

1. Insert line of the test setup in the device.

2. Vent test setup. Bottom part of drop chamber must be 2/3 filled.

3. Insert cannula in empty graduated cylinder.

4. Set delivery rate according to the TSC.

5. Simultaneously start infusion and stop watch.

6. Stop the stop watch when the measured volume defined in the TSC is reached.

7. Evaluate deviation in percent according to Table 8 — 1 and record.

Fig.: 8 — 6

25 ml ± 0.4 ml

Reference height80 ± 2 cm

Measuring Time Deviation%

6 min 40.0 sec -10

6 min 18.9 sec -5

6 min 15.0 sec -4

6 min 11.1 sec -3

6 min 7.3 sec -2

6 min 3.6 sec -1

6 min 0.0 sec 0

5 min 56.4 sec 1

5 min 52.9 sec 2

5 min 49.5 sec 3

5 min 46.2 sec 4

5 min 42.9 sec 5

5 min 27.3 sec 10

Table 8 — 1

8 — 6 Infusomat® fmS 2.3 gb

Page 75: Infusomat FmS Service Manual

Procedural Instructions on the TSC 8

The accuracy of the delivery rate can be checked using a normal infusion line when the TSC is carried out.

A test infusion line (OIL test infusion line with an Infusomat® fmSadjusted for the Original Infusomat® line, and a Space calibration line with an Infusomat® fmS set for the Infusomat® Space Line) is to be used for calibration (see „Scale Factor Function 510.0“ ➪p. 3 — 9)! (see „Test Equipment and Special Tools“ ➪ p. 9 — 1).

Drop Sensor

1. Set delivery rate according to the TSC.

2. Connect drop sensor to the device.

3. Start infusion according to the specifications of the TSC.

4. Squeeze off line above the pump. The drop sensor triggers an alarm.

5. Squeeze the drop chamber bottom part, to generate a jet. The drop sensor triggers an alarm immediately.

Air Inline Sensor

1. Plug MFC service connector on the MFC plug connector.

2. Insert a line filled with water in the device and close operat-ing unit.

3. Read off test value of the air inline sensor via the function 300.0 “Air inline sensor” in the Service Program (see „Air In-line Sensor Function 300.0“ ➪ p. 3 — 5).

4. Read off value after approximately 10 seconds and compare with the water value specifications of the TSC.

5. Document values.

6. Enter a delivery rate of 250 ml/h and a volume of 250 ml.

7. Start infusion.

8. Generate an air bubble of approx. 0.4 ml (approx. 56 mm length of line with air) in the supply line to the pump. An alarm is triggered when the air bubble is detected.

9. Insert a line filled with air in the device and close operating unit, or disconnect the line at the container and “vent” in or-der to remove the water out of the line.

10. Read off test value of the air inline sensor via the function 300.0 “Air inline sensor” in the Service Program (see „Air In-line Sensor Function 300.0“ ➪ p. 3 — 5).

11. Read off value after approximately 10 seconds and compare with the air value specifications of the TSC.

12. Read off threshold value via the function 520.0 “Air inline sensor” in the Service Program (see „Air Inline Sensor Func-tion 520.0“ ➪ p. 3 — 10).

Infusomat® fmS 2.3 gb 8 — 7

Page 76: Infusomat FmS Service Manual

Procedural Instructions on the TSC8

13. Compare value with the threshold value specifications of the TSC.

Battery Check

1. Disconnect the device from the mains supply during opera-tion.

2. The symbol for the mains plug on the LC display goes off. An error message is not triggered and operation of the unit is continued.

3. Reconnect unit to the mains.

4. The symbol for the mains plug is displayed on the LC display. An error message is not triggered and operation of the unit is continued.

5. Switch device off.

6. Disconnect unit from mains.

7. Switch on unit. A self-test is carried out.

8 — 8 Infusomat® fmS 2.3 gb

Page 77: Infusomat FmS Service Manual

99 — 2

9 — Test Equipment and Special Tools

For Repair / for TSC Order No.

Test equipment case Infusomat fm (complete) . . . . . 0770 1527 with: Calibration gauge 4 mm (for adjustment after exchange of the pump cover) . . . . . . . . . . . . . . . . . . . . . . . . . 0770 1489 Pin punch 1.8 mm x 160 mm (for hinge pin/ disassembly of the pump cover) . . . . . . . . . . . . . . 0770 1446 Pin punch 6 mm x 125 mm (for hinge pin/ assembly of the pump cover) . . . . . . . . . . . . . . . . . 0770 1454 Flat tool 100 x 20 mm (for assembly/ disassembly of the tube adapter) . . . . . . . . . . . . . 0770 1462 Special socket spanner M18 (for disassembly of the recessed plug). . . . . . . . . . . . . . . . . . . . . . . . . . 0770 1497 Manometer 0 — 4 bar . . . . . . . . . . . . . . . . . . . . . . . 0770 1357 Service connector (red) . . . . . . . . . . . . . . . . . . . . . 0770 0709 MFC service connector . . . . . . . . . . . . . . . . . . . . . . 3450 1215 OIL-test infusion line . . . . . . . . . . . . . . . . . . . . . . . 0770 1500

Space calibration line . . . . . . . . . . . . . . . . . . . . . . . . . . 0770 1610

Pressure calibration device . . . . . . . . . . . . . . . . . . . . . . 0770 5018

Infusomat® fmS 2.3 gb 9 — 1

Page 78: Infusomat FmS Service Manual

Test Equipment and Special Tools9

For your notes:

9 — 2 Infusomat® fmS 2.3 gb

Page 79: Infusomat FmS Service Manual

1010 — 4

10 — Spare Parts List

Unit Elements

Designation Order — No.

Battery

Battery incl. connector plug 1.2 Ah / 7.2 V and holder . . . . . . . . . . . . . . . . . . . . 3450 2556

Mains Fuses

Fuse T 0.16 A for 200 / 230 / 240 V (10 pcs.) . . . . . . . . . . . . . . . 3477 2847

Fuse T 0.315A for 100 / 110 / 120 V (10 pcs.) . . . . . . . . . . . . . . . 3477 0534

Fuse holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 0979

Housing

Housing Labelling German . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1843 French . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2130 Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2149 Italian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2157 English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2165 Spanish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2173 Danish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2181 Norwegian. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2190 Swedish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2203 Finnish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2211 Portuguese . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2220 Czech . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2238 Polish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2246 Turkish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2459

Foot stand complete with rubber feet . . . . . . . . . . . . . 3450 5415

Unit handle with O-rings and PT screws . . . . . . . . . . . 3450 1789

Controller Board

Distance sleeve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3366

Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3447

Loudspeaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8848

Designation Order — No.

Rear Panel

Rear panel with screws (M3) and seal . . . . . . . . . . . . . 3450 1860

Cover for optical interface . . . . . . . . . . . . . . . . . . . . . . 3477 3164

Strip seal for rear panel . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3142

MFC connector board . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3374

Potential equalization bolt . . . . . . . . . . . . . . . . . . . . . . 3477 0550

fm recessed plug (3 pin) . . . . . . . . . . . . . . . . . . . . . . . . 3477 3177

Screw 30×8 for fm recessed plug (20 pcs.) . . . . . . . . . 3477 3185

U Washer 3.2 (20 pcs.). . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3193

Mains module 200 / 230 / 240 V . . . . . . . . . . . . . . . . . 3450 1886

Mains module 100 / 110 / 120 V . . . . . . . . . . . . . . . . . 3450 1894

Mains module 230 V . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1908

Drop sensor socket incl. cable and plug . . . . . . . . . . . . 3450 1878

Pump Cover

Pump cover with lock . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5717

Blind plug 7.1 mm (10 pcs.) . . . . . . . . . . . . . . . . . . . . . 3477 3207

Torsion spring in lever/pump cover (5 pcs.) . . . . . . . . 3477 3363

Torsion spring for pump cover (5 pcs.) . . . . . . . . . . . . 3477 3355

Lever (pump cover). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4092

Hinge pin for pump cover . . . . . . . . . . . . . . . . . . . . . . . 3477 3967

Hinge pin for pump cover lever. . . . . . . . . . . . . . . . . . . 3450 5725

Front Panel

Front panel without flow inhibitor and pressure spring. . . . . . . . . . . . . . . . . . . . . . . . . 3450 5822

Circular seal 570 mm / 45 mm . . . . . . . . . . . . . . . . . . . 3477 3126

Flow inhibitor with pressure spring . . . . . . . . . . . . . . . 3477 3258

Pressure spring for flow inhibitor (5 pcs.) . . . . . . . . . . 3477 3266

Tamper-proof caps 10 mm (50 pcs.) . . . . . . . . . . . . . . 3477 3134

Reed sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1754

Infusomat® fmS 2.3gb 10 — 1

Page 80: Infusomat FmS Service Manual

Spare Parts List10

Designation Order — No.

Pump Unit

Finger pump (without motor) including pump, . . . . . 3450 1738 pump cover, seal membrane and boards

Finger pump (without motor and board) . . . . . . . . . . 3450 9038 incl. pump, pump cover, and seal membrane

Membrane incl. seal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5733

Motor with pinion for finger pump . . . . . . . . . . . . . . . 3450 1924

Pressure Sensor

Pressure sensor (cmpl.) . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7353

Air Sensor

Air sensor incl. connector . . . . . . . . . . . . . . . . . . . . . . . 3450 193A

Door Lock

Door lock complete with push button . . . . . . . . . . . . . 3450 5601

Spring holder for door lock . . . . . . . . . . . . . . . . . . . . . . 3450 5440

Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790

Operating Unit

Membrane keypad with support plate and seal. . . . . . 3450 1797

LCD module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1819

Frame incl. pressure spring and magnet . . . . . . . . . . . 3450 1827

Flexible cable 42 mm (5 pcs.) . . . . . . . . . . . . . . . . . . . . 3477 3347

Hinge unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5571

Hinge pin (3 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5580

Magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5847

Designation Order — No.

Frame with Seal

Seal plate between frame and front panel (exchange not recommended) . . . . . . . . . . . . . . . . 3450 3240

Frame with seal plate . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1762

Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790

Colours

Touch-up pen RAL 7031 (grey) . . . . . . . . . . . . . . . . . . . 3450 6985

Miscellaneous

Assembly screw for display board PT 2.5×14 (10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3100

Screw PT 3×10 Torx (self-forming) . . . . . . . . . . . . . . . . 3450 0960

Rubber feet grey (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . 3477 3096

Attention label fuse T 0.16 A (5 pcs.) . . . . . . . . . . . . . 3450 0898

Screw M 6×8 for fm recessed plug (20 pcs.) . . . . . . . . 3477 3185

U Washer 3.2 (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3193

Software Update

Update kit IFMC02003 . . . . . . . . . . . . . . . . . . . . . . . . 3450 645A

Update kit IFMe02004 / IFME03004 . . . . . . . . . . . . . . 3450 645C

Update kit IFMe02005 / IFME03005 . . . . . . . . . . . . . 3450 645D

Update kit IFMe02006 / IFME03006 . . . . . . . . . . . . . . 3450 645E

MFC interface line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 1661

10 — 2 Infusomat® fmS 2.3gb

Page 81: Infusomat FmS Service Manual

Spare Parts List 10

Designation Order — No.

Poleclamp

Pole clamp (universal clamp, rotating) . . . . . . . . . . . . 3450 9054

Universal Clamp (Poleclamp)

Universal clamp, complete . . . . . . . . . . . . . . . . . . . . . . 3450 5857

Universal clamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8325

Threaded rod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8333

Star handle body . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8384

Safety clip. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8341

Safety hook. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8368

Plate (2 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2610

Connection cap D12/4 (5 pcs.) . . . . . . . . . . . . . . . . . . . 3477 4149

Bellows (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3274

Pressure spring (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4165

Designation Order — No.

Universal Clamp

Universal clamp, complete . . . . . . . . . . . . not available any more

Threaded rod . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 50 5903

Safety hook. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 50 5865

Turning handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 50 5890

Rubber cover (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . 34 77 3290

Bellows (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 77 3274

Connection cap (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . 34 77 3304

Pressure spring for pole fixation (5 pcs.) . . . . . . . . . . . 34 77 3282

Fig.: 10 — 1 Universal Clamp (Poleclamp)

Fig.: 10 — 2 Universal Clamp

Infusomat® fmS 2.3gb 10 — 3

Page 82: Infusomat FmS Service Manual

Spare Parts List10

For your notes:

10 — 4 Infusomat® fmS 2.3 gb

Page 83: Infusomat FmS Service Manual

1111 — 2

11 — Index

A

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 11

Air inline sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 17

Alarm circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 8

Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 — 4

Approved software versions . . . . . . . . . . . . . . . . . . . . . . . . . . 2 — 2

Assignment, internal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 10

B

Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 1

Battery operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 8

Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 4

C

Checks after repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 — 1

Controller board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 5

Current versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 — 5

D

Default values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 — 6

Door lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 2

E

Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 — 2

Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 — 4

F

Frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 19

Front frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 11

Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 4

Functional inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 — 3

H

Handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 4

Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 4

L

List of abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 — 8

M

Mains fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 1

Mains operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 7

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 — 1

O

Occlusion pressure,

electronic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 9

mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 9

Operating unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 18

P

Physical construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 1

Pressure sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 14

Procedural instructions on the TSC . . . . . . . . . . . . . . . . . . . . 8 — 1

Pump cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 3

Pump unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 8, 4 — 12

R

Rear panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 — 6

Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 — 5

Revision documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A — 1

Revision service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 — 5

S

Service Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 — 1

Service Program,

activate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 — 3

quit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 — 3

Software update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 — 1

Spare parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 — 1

Special tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 — 1

T

Technical Safety Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 — 1

Technical Safety Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 — 5

Test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 — 1

TSC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 — 5, 7 — 1

V

Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 — 1

Voltage supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 — 5

Infusomat® fmS 2.3 gb 11 — 1

Page 84: Infusomat FmS Service Manual

Index11

11 — 2 Infusomat® fmS 2.3 gb

Page 85: Infusomat FmS Service Manual

AA — 2

A — Revision Documentation

Revision Service-Documentation Version 2.0

This manual has been completely revised. The most important changes are listed below:

— Changed manual structure

— New TSI list

— New paragraph pressure sensor (exchange / calibration)

— New test equipment for pressure sensor

— New software

— New spare parts

— Total list of spare parts

Version 2.1

— New TSI list

— New controller board

— New software

Version 2.2

This version was approved by B. Braun on 19.04.2006.

The most important changes of this version are listed below:

— New software

— New Service Program functions

— New spare parts

— Changed occlusion pressure data

Version 2.3

This version was approved by B. Braun on 18.02.2008.

The most important changes of this version are listed below:

— New software

— Possible use of the Infusomat® Space line

— New TSC

— Revised procedural instructions on the TSC

Current Information

Infusomat® fmS 2.3 gb A — 1

Page 86: Infusomat FmS Service Manual

Revision DocumentationA

For your notes:

A — 2 Infusomat® fmS 2.3 gb

Что такое инфузионная помпа?

Что такое инфузионная помпа?

Одноразовая эластомерная инфузионная помпа – это медицинское устройство для непрерывного введения лекарственных средств или питания через венозный катетер с постоянной или регулируемой скоростью. Главное преимущество микроинфузионной помпы – амбулаторное лечение. Система непрерывно вводить пациенту обезболивающие средства, препараты для химиотерапии, питательные растворы, парентеральное питание, и длиться это может от 12 минут до 25 суток.

Строение помпы простое и состоит из защитного корпуса, эластомерной мембраны для введения л.с. или парентерального питания, гибких трубок с зажимом и бактериальным фильтром на них, переходником типа луер лок. Некоторые модели укомплектованы регулятором скорости потока подачи лекарства в вену и болюсом для дополнительной дозы препарата, если это требуется пациенту

5 плюсов инфузионных помп

  1. Ускорение постоперационной реабилитации
  • Иммобилизация – одна из причин задержки послеоперационного восстановления пациента. Полная автономность, малый вес и небольшие размеры помпы обеспечивают свободу передвижений. За счет чего возможна выписка пациента домой уже через 1-3 часа после операции.
  1. Индивидуальная настройка
  • Врач подбирает для каждого пациента подходящую скорость введения необходимого лекарственного препарата, учитывая диагноз, индивидуальный болевой порог, особенности проведенной операции и других факторов.
  1. Эффективное лечение
  • Необходимая доза инфузии с постоянной или регулируемой скоростью потока сокращает болевые ощущения и дискомфорт. Во время проведения химиотерапии сокращается возникновение побочных эффектов, таких как головокружение и тошнота.
  1. Надежность и безопасность
  • Это подтверждается медицинскими исследованиями и клинической практикой. Помпа не влияет на частоту осложнений после операций, а в некоторых случаях частота даже снижается – например, число дыхательных осложнений уменьшается в 3,3 раза! (по данным исследования С.В. Авдеева, А.А. Завьялова и др.)
  1. Простота в использовании
  • Пациент не забивает себе голову постоянными мыслями об источниках электропитания, т.к. они просто не требуются. Лечащий врач устанавливает необходимую скорость подачи лекарства, после чего пациент ведет привычный образ жизни в течение всего времени проведения терапии.

Область применения инфузионных помп

У инфузионных пом нет «узкой специализации» — их используют анестезиологи, реаниматологи, кардиологи, онкологи и другие специалисты. В анестезиологии и реаниматологии помпы применяют для интра- и послеоперационного обезболивания. Избавление от боли важно также в ортопедии, после операций на позвоночнике, коррекции сколиоза, и в травматологии, в том числе при эндопротезировании.

Возможности, которые дают помпы для обезболивания пациентов, особенно при длительных перевозках, ценят врачи скорой помощи. Акушеры применяют помпы для анальгезии родов и при лечении гестозов.

Инфузионные помпы

Онкологи и гематологи применяют помпы также для стационарной и амбулаторной химиотерапии и для лечения тромбов. В кардиологии инфузионные помпы тоже помогают справиться с тромботическими осложнениями, а еще – с профилактикой и лечением ишемии. А неврологи, помимо всего перечисленного, используют помпы для введения ноотропов, антисудорожных и анксиолитических препаратов.

Несмотря на широкие возможности микроинфузионных помп количество лекарственных препаратов, которые вводят с их помощью, также велико, и только лечащий врач, в зависимости от поставленного диагноза и состояния пациента, компетентен в подборе необходимых препаратов. Однако присутствуют и некоторые ограничения: помпы предназначены только для жидких препаратов  или очень небольшой густоты, без твердых частиц, т.к. подобные вещества снижают скорость инфузии или засорить помпу.

Также помпы некоторых производителей (например, Woo Young и B.Braun) нельзя наполнять кровью, кровяными препаратами, белками, липидами и жировыми эмульсиями. Выбирая помпу, следует уделить внимание особенностям ее использования, о которых предупреждает производитель.

Например, для химиотерапии чаще используют помпы с объемом 250 — 300 мл и небольшой скоростью потока из-за агрессивности вводимых химиопрепаратов, а для анальгезии – помпы, объем которых составляет 275-400 мл. Однако подбор модели индивидуален для каждого пациента, т.е. при необходимости химиопрепарат может подаваться из помпы в 600 мл, а обезболивающее – из помпы в 100 мл.

SFS-5-25 Инфузионная помпа Nipro Superfuser+ с постоянной скоростью потока 5 мл/час, без болюса, 250 мл

Виды и производители помп

Конечно, есть множество фирм, изготавливающих инфузионные помпы, и потому мы перечислим самых надежных и известных производителей. Помпы делятся на одноразовые и многоразовые, а также – на изделия большого и малого объема, в соответствии с задачей, которую они решают. Помпы большого объема чаще используют для введения питательных растворов, а малого – для лекарственных препаратов.

  1. Микроинфузионные помпы фирмы Vogt Medical (Германия)
  • Фирма производит стерильные одноразовые помпы как с постоянной скоростью потока, так и с регулируемой. Рабочий объем помп с постоянной скоростью выпускаеся двух типов – 100 мл и 275мл. Скорость инфузии 1/2/4/10 мл в час для 100-миллилитровой модели и 5/8 мл в час для 275-миллиметровой.
  • Для помп, снабженных регулятором скорости, при объеме в 100 мл скорость будет переключаться по градациям 1/2/3/4 и 2/4/6/8 мл/час, а при объеме в 275 мл возможны градации — 2/4/6/8 и 4/6/8/10 мл/час.
  • Модели снабженные болюсом позволяют вводить дополнительно от 0,5 до 2 мл лекарственного средства с периодичностью в 15 минут.
  1. Микроинфузионные помпы Accufuser (Южная Корея)
  • Свыше 50 моделей помп представлено на российском рынке. Легко подобрать даже для лечения самых редких форм заболеваний. Accufuser используют для введения препаратов при химиотерапии, анальгезии и антибактериальной терапии.
    Помпы можно подобрать по объему от 50 до 600 мл и по скорости подачи л.с. от 0,5 до 250 мл/ч. Минимальное время инфузии составляет 25 минут, максимальное время инфузии — 25 суток
  1. Инфузионные помпы Easypump II от BBraun (Германия)
  • Особенностью данной помпы является мягкий защитный корпус, который по мере опустошения сжимается.

Действие инфузионной помпы изипамп 2

  • Помпы предназначены для амбулаторного лечения, и позволяют уже через несколько часов после операции отправиться домой. Изипамп 2 носится в специальном кейсе, который можно закрепить на ремне, поясе брюк либо юбки. Объем помп составляет от 60 до 400 мл, а скорость потока варьируется от 2 до 500 мл/ч в зависимости от модели. Время инфузии может занимать от 30 минут до 135 часов.
  1. Микроинфузионные помпы Acti-Fine от M.Schilling (Германия)
  • Продукция компании отличается широким спектром применения: от кардиологии до химиотерапии, от анальгезии до неврологии. Помпы объемом на 100 и 275 мл оснащены регулятором скорости потока с градациями от 1 до 14 мл/ч, в зависимости от модели. При этом регулятор очень прост в использовании. Также в линейке присутствует модель с болюсом для контролируемой пациентом анальгезии.

Главное преимущество инфузионных помпы — амбулаторное лечение

При применении помпы число дыхательных осложнений уменьшается в 3,3 раза

Микроинфузионные помпы не зависят от батареек и электросети

Предостережения:

  • Используйте только стерильные помпы.
  • Не заполняйте помпу больше максимума, заявленного на упаковке.
  • Убедитесь, что лекарственный препарат совместим с используемым устройством.

Рекомендации:

  • Используйте для наполнения помпы шприцы с объемом 50 мл и более.
  • Для определения полного заполнения помпы, откройте зажим и проверьте, с какой скоростью капает жидкость.

SMA-2468-25P Инфузионная помпа Nipro Superfuser+ с регулируемой скоростью потока 2/4/6/8 мл/час, без болюса, 250 мл

Как выбрать помпу?

При выборе инфузионной помпы не стоит ориентироваться исключительно на бренд. В каталоге Стерильно.com представлена продукция компаний, которые давно зарекомендовали себя как надежные производители медицинского оборудования.

Подбирая помпу для инфузионной терапии лучше следовать рекомендациям врача, а также ориентироваться на следующие показатели:

  • объем мембраны, в которую вливают препараты;
  • наличие или отсутствие болюса для контролируемой пациентом анальгезии;
  • наличие или отсутствие регулятора скорости потока;
  • при выборе помпы с постоянной скоростью потока – обратить внимание на то, какова эта скорость: она измеряется в миллилитрах в час (например, 2 мл/час или 4 мл/час и т.д.);
  • время инфузии – этот параметр важно учитывать при подборе помпы для длительного, средне- или краткосрочного введения препарата



Предложения по теме

product

Инфузионная помпа Изипамп 2 (LT 270-27-S)

B.Braun

4550


за штуку

product

Инфузионная помпа Tuoren с регулируемой скоростью потока 2/4/6/8 мл/час, без болюса, 275 мл

TUORen Medical

3220


за штуку

product

Инфузионная помпа Tuoren с регулируемой скоростью потока 2/5/7/10 мл/час, без болюса, 275 мл

TUORen Medical

3360


за штуку

product

Инфузионная помпа Изипамп 2 (LT 100-50-S)

B.Braun

4550


за штуку

product

Инфузионная помпа Tuoren с постоянной скоростью потока 5 мл/час, без болюса, 275 мл

TUORen Medical

2870


за штуку

product

Инфузионная помпа Tuoren с постоянной скоростью потока 10 мл/час, без болюса, 275 мл

TUORen Medical

2870


за штуку

product

Инфузионная помпа Tuoren с регулируемой скоростью потока, без болюса, 100 мл

TUORen Medical

2950


за штуку

product

Инфузионная помпа Изипамп 2 (ST 400-4-S)

B.Braun

4675


за штуку

product

Инфузионная помпа Tuoren с постоянной скоростью потока, без болюса, 100 мл

TUORen Medical

2650


за штуку

product

Инфузионная помпа Изипамп 2 (LT 120-30-S)

B.Braun

5369


за штуку

product

-15%

Микроинфузионная помпа Accufuser CT0040M

Woo Young Medical

3190 2711


за штуку

product

Микроинфузионная помпа Accufuser Continuous C0050L

Woo Young Medical

4990


за штуку

product

Микроинфузионная помпа Accufuser Continuous C0040M

Woo Young Medical

4990


за штуку

product

Микроинфузионная помпа Accufuser Continuous C0020L

Woo Young Medical

4990


за штуку

  • About
  • Blog
  • Projects
  • Help
  • Donate

    Donate icon
    An illustration of a heart shape

  • Contact
  • Jobs
  • Volunteer
  • People

Bookreader Item Preview

texts

B Braun Infusomat fmS Infusion Pump Service Manual

B Braun Infusomat fmS Infusion Pump Service Manual

Addeddate
2020-05-20 04:27:29
Classification
Clinical;Infusion and Syringe Pump;B Braun Infusion and Syringe Pump;B Braun Infusomat fmS
Identifier
manual_B_Braun_Infusomat_fmS_Infusion_Pump_Service_Manual
Identifier-ark
ark:/13960/t0cw38z5k
Ocr
ABBYY FineReader 11.0 (Extended OCR)
Page_number_confidence
54.12
Ppi
300
Scanner
Internet Archive Python library 1.9.0

comment

Reviews

There are no reviews yet. Be the first one to
write a review.

SIMILAR ITEMS (based on metadata)

Понравилась статья? Поделить с друзьями:
  • Паспорт анион 4100 руководство по эксплуатации
  • Инструкция по сбору мочи на сахар
  • Плита hansa сенсорная инструкция по применению
  • Rittal холодильный агрегат sk3304500 руководство
  • Ковер по технологии 2 класс инструкция