Page 1
Patient Monitor MPR6-03 User Manual RM 501.01.000-01-01 UM Version 7, 05/2020… -
Page 3: Table Of Contents
CONTENT 1. DEVICE DESCRIPTION ……………………….6 1.1. I …………………………6 NTRODUCTION 1.1.1. Intended Use …………………………. 6 1.1.2. Basics of Operation and Construction ………………….7 1.1.3. General …………………………. 9 1.1.4. Revision History ……………………….10 1.1.5. Symbols and Marks ……………………… 11 1.1.6. General Specifications ……………………..14 1.2.
-
Page 4
2.2.5. Analysis of Heart Rate Variability (HRV) ………………..91 2.2.6. Synthesis of ECG Leads ……………………..92 2.2.7. Features of the Operation with a Pacemaker (PM) ……………… 92 2.2.8. Features of the Operation in the Operating Rooms ……………… 93 2.2.9. Disinfection of the Patient Cable ………………….93 2.2.10. -
Page 5
3.10.1. Basic Principles of cNIBP Method ………………….. 162 3.10.2. dPWTT Measurement and NIBP Trigger Conditions …………….163 3.10.3. Continuous Arterial Blood Pressure Calculation …………….. 165 3.10.4. Operation of cNIBP Mode ……………………166 3.10.5. Limitations of cNIBP Method ………………….. 167 3.11. MAINSTREAM CO SENSOR …………………… -
Page 6: Device Description
1. Device Description 1.1. Introduction 1.1.1. Intended Use The present operating manual applies to the portable bedside monitor MPR6-03- «Triton» (here- inafter referred to as the device). The device is designed for long-term and continuous monitoring of the following patient parame-…
-
Page 7: Basics Of Operation And Construction
1.1 Introduction to monitor vital functions during operations, in the postoperative period, during inva- sive diagnostic interventions and intensive care; to work in extreme conditions of disaster medicine; for screening. 1.1.2. Basics of Operation and Construction The operation of the pulse oximetry module is based on different spectral absorption of oxy- hemoglobin and reduced blood hemoglobin.
-
Page 8
1.1 Introduction Computer network support allows the device to operate as part of a centralized monitoring com- puter network (CMS), view on a computer and save in its memory the data recorded by the de- vice. The built-in thermal printer provides a printout of the parameters recorded by the device (in- stalled by special order). -
Page 9: General
1.1 Introduction 1.1.3. General The present operating manual is an integral part of the device and is included in its delivery kit. The present document is provided for informational purposes only, it is not allowed to copy, re- produce, translate into another language, save in an information retrieval system, transmit in any form or convert it into a form suitable for storage on electronic media without written per- mission manufacturer Before starting to work with the device, carefully read the present User Manual.
-
Page 10: Revision History
1.1 Introduction To ensure the operability of the device and increase its service life during operation, it is NEC- ESSARY: protect the device from dropping and knocking, especially on the surface of the liquid crystal display; do not allow pressing the touch buttons on the device display with hard objects with sharp corners to avoid damage to the touch panel;…
-
Page 11: Symbols And Marks
1.1 Introduction 1.1.5. Symbols and Marks Symbols, icons and marks used on the device housing Russian Conformity mark Type approval mark (measurement means according legislation of the Russian Fed- eration) ATTENTION, refer to the operating documents! Refer to the operating documents! Serial number Manufacturing date IP21…
-
Page 12
1.1 Introduction Infoport “Infoports” for mainstream CO CAPNO sensor, depth of anesthesia module Ethernet Connector to a local computer network Port for USB Flash memory devices POWER Device power supply LED Charger and internal battery warning LED OPEN Built-in thermal printer cover latch POWER Power supply LED for built-in thermal printer ERROR… -
Page 13
1.1 Introduction Reset to the factory settings Start NIBP measurement Stop NIBP measurement Trends in a graphical form Trends in a table form Curve fragments window Switch the trend scale (figure — current scale) Button for moving along the trends or selecting the serial number of the recorded frag- ment of the curves at the review Button for moving page-by page Start / cancel printing on a thermal printer (in the tabular trend window and in the wave-… -
Page 14: General Specifications
1.1 Introduction 1.1.6. General Specifications Parameter Description General Multiparameter patient monitor for the Intended use long-term and continuous monitoring of the vital functions of the patient. Operation modes (patient groups) Adult, pediatric, neonatal 220 ± 22 V / 50 Hz AC power Operation from a fully charged built-in battery At least 120 min…
-
Page 15
1.1 Introduction Continuous non-invasive blood pressure measurement available (cNIBP)– SysAD, DiaAD, MapAD, PWTT Monitoring of central hemodynamics parameters by the available volume compression oscillometry: non-invasive meas- urement of cardiac output (CO), stroke volume (SV), sys- temic vascular resistance (SVR) with the calculation of a number of related parameters 2.6 Temperature (T), skin and central, 2 channels available… -
Page 16
1.1 Introduction 3.15 Detection range of QRS-complex amplitude 0.05 — 10 mV Pulse oximetry channel Displaying of pulse rate, perfusion, SpO Digital Displaying of the photoplethysmogram Continious Scaling of the photoplethysmogram Automatic indication range 10 — 100 % measurement range 70 — 100 % ±2 % Absolute accuracy of SpО… -
Page 17
1.1 Introduction Non-invasive monitoring of cardiac output (channel of the central hemodynamics) Measurement method — volumetric compression oscil- Available lometry Measured parameters: — SV – stroke volume 10 — 250 ml — СО – cardiac output 1.0 — 20 lpm Calculated parameters: — CI –… -
Page 18
1.1 Introduction ART (Arterial blood pressure), PA (Pulmonary artery pressure), CVP (Central venous pressure), Types (markers) of the measured pressure: ICP (Intracranial pressure) 10.5 RAP (Right atrial pressure), LAP (Left atrial pressure) RVP (Right ventricular pressure) UA (Umbilical arterial pressure) 10.6 Invasive pressure measurement range: -50…… -
Page 19
1.1 Introduction Measurement range: — СО 13.3 concentration 0 — 15 % — СО partial pressure 0 — 115 mmHg Measurement accuracy of СО concentration: ±0.2 % — absolute in the range 0 — 5 % ±4 % — relative in the range 5,1 — 10 % ±6 % — relative in the range 10,1 — 15 % 13.4… -
Page 20
1.1 Introduction Displayed graphic parameters: EEG curve in one measured lead with adjustment of Available the window size (5, 10, 30, 60 s) and EEG scale (10 — 1000 mV) Trends of brain activity index and EMG level with ad- Available 15.3 justable trend length (5, 10, 30, 60 min) -
Page 21
1.1 Introduction 19.4 Temporarily disable the audible alarm for 2 min Available 19.5 System of text messages to the user about alarm Available HR/PR, SpO , SysAD, DiaAD, MapAD, ЧД, Т, StDev. 19.6 Setting of alarm limits for parameters EtCO , FiCO , EtO , FiO… -
Page 22: Getting Started
1.2 Getting Started 1.2. GETTING STARTED 1.2.1. Appearance and Controls Figure 1-1. Appearance of the Patient Monitor MPR6-03 Front panel Screen (color TFT-display) with the touch control panel. The «BATTERY» indicator informs about the operation of the charger and the internal rechargeable battery: off — battery is charged;…
-
Page 23
11 — Manipulator (encoder) to control the device. 12 — Lamp «ALARM» for additional attention, blinks red when an alarm is triggered. Figure 1-2. Patient Monitor MPR6-03, front view The surface of the liquid crystal display of the device shall be protected from shock and other mechanical influences to avoid unevenness, scratch- es, as well as damage to the touch panel and display. -
Page 24
10 — sensor. 11 — Pneumatic connector for the respiratory mechanics module. 12 — Connector for microflow line for sampling. Figure 1-4. Patient Monitor MPR6-03, side view (panel of ports and connectors) Bacterial filter Clips for attaching to the connector panel… -
Page 25: Operating Conditions
“POWER” LED — indicates power supply. Paper exit slot. Cover of the built-in thermal printer. Figure 1-6. Patient Monitor MPR6-03, side view (the built-in thermal printer cover) 1.2.2. Operating Conditions Ambient temperature 10…35 °С. Note. The operating time from the built-in battery is standardized for charging and discharging at tem- peratures from +20 ºС…
-
Page 26: Safety Instructions
1.2 Getting Started 1.2.3. Safety Instructions For safety device meets the requirements of GOST R 50444, GOST R IEC 60601-1, GOST R IEC 60601-1-6, GOST IEC 60601-1-8, GOST R IEC 62304, GOST R IEC 62366, GOST 31515.1, GOST 31515.3, GOST 30324.30, GOST R IEC 60601-2-27, GOST R IEC 60601-2-49, GOST R 50267.34, GOST ISO 9919, GOST ISO 10993-1, GOST R ISO 80601-2-55, GOST R ISO 15223-1, GOST Р…
-
Page 27
1.2 Getting Started fixing equipment and accessories outside the center of gravity. when placing the trolley-mounted device on an inclined plane, the swivel wheels with a brake must be at the bottom to prevent the device from slipping; be careful when transporting the device on the trolley; to prevent overturning when moving, hold the trolley handle with both hands. -
Page 28: Preparing To Turn On The Device
1.2 Getting Started 1.2.4. Preparing to Turn on the Device The device is intended only for visual display and automatic registration of the basic vital physiological parameters of the patient and cannot relieve medical personnel from the obligation of continuous physical observation of the patient.
-
Page 29: Disinfection
1.2 Getting Started Do not try to switch on the device immediately after switch it off. The delay between switching off the device and its subsequent switching on must be at least 5 seconds. During this time pressing the » «…
-
Page 30: Device Control
1.3 Device Control 1.3. DEVICE CONTROL The device is controlled by the touch screen, control buttons “F”, “123”, “CAPNO”, “NIBP”, “FREEZE”, « » and the encoder. 1.3.1. Display (Screen Configuration) Cardiometry window. You can set here display mode (one or three ECG curves) and numerical pa- rameters of the arrhythmia analysis.
-
Page 31: Touch Panel
1.3 Device Control 1.3.2. Touch Panel The device is controlled using the touch screen. Press the screen to change the operation mode of the output window or to change the parameter. A more detailed description of the control of the device using the touch screen is described in the following sections. 1.3.3.
-
Page 32: Configuring Of Curves On The Screen
1.3 Device Control 1.3.5. Configuring of Curves on the Screen To change the measurement channel, press on the name of the curve: You will see at the screen the list of available channels (depending on options in the device): Select the channel you want to display by pressing: Histgrm heart rate variability window PPG SpO (pulse oximetry) window…
-
Page 33: Setting Of Date And Time
1.3 Device Control 1.3.6. Setting of Date and Time Click on the date and time field: You will see the window for setting the date and time: Select the required field for correction by clicking on it, it will be highlighted in a different color. Use the buttons to set the required value.
-
Page 34: Setting Of Sound Alarm Volume
1.3 Device Control 1.3.7. Setting of Sound Alarm Volume Click on the sound control icon: The volume setting window will be displayed: Select the required field for correction by clicking on it, it will be highlighted in a different color. Use buttons to set the required value.
-
Page 35: Setting Of General Parameters, Modes Of Gas Analysis And Displaying Of Hr/ Pr And Rr
1.3 Device Control 1.3.8. Setting of General Parameters, Modes of Gas Analysis and Displaying of HR/ PR and RR When FREEZE mode is activated, it is not possible to enter the general pa- rameters menu. To select the display mode for HR / PR, RR and enter the menu for setting of general parame- ters, click on the corresponding icons on the device screen: HR or PR source selection…
-
Page 36
1.3 Device Control Setting of parameters in accordance with it.1.3.10. Click on the field with the corresponding parameter, select it and set the required value (digital parameters in accordance with p. 1.3.10). The appearance of the general parameters window may change depending on the set of options in the device. -
Page 37: Setting Of Profile
1.3 Device Control 1.3.9. Setting of Profile The device has several groups of profiles (i.e. options for setting parameters by module) stored in non-volatile memory. Profiles define: for the NIBP module — the limit value of the pressure generated in the cuff and the limit time for pressure measurement;…
-
Page 38
1.3 Device Control For all profiles it is possible to set the patient’s weight with an accuracy of 0.1 kg, the minimum value of the patient’s weight is 0.4 kg To restore the factory settings of the profile, click the corresponding icon: If the profile parameters were changed during operation by the user and differ from the factory settings, the display will show the window for returning to factory settings: If you press the «Yes»… -
Page 39
1.3 Device Control Profiles Unit Adult Pediatric Neonatal Bottom threshold RR 1/min Top threshold RR 1/min Bottom threshold Т ºC Top threshold Т ºC Bottom threshold segment Top threshold ST- segment Maximum cuff pressure mmHg Bottom threshold EtСО mmHg Top threshold EtСО mmHg Bottom threshold FiСО… -
Page 40: Setting Of Digital Parameters And Alarm Thresholds
1.3 Device Control 1.3.10. Setting of Digital Parameters and Alarm Thresholds Setting the values of digital parameters and alarm thresholds in the device is performed in the same way. To set the alarm thresholds of all displayed digital parameter, click on the displayed value of the top (or bottom) threshold of the corresponding parameter: You will see the window for setting the alarm threshold (for example, the window for setting the top alarm threshold by the Et value of the multigas module):…
-
Page 41
1.3 Device Control Turn the encoder knob in the required direction. To confirm the changed parameters, press the «OK» button (or the encoder button), to discard changes — «Cancel» Ranges for the alarm thresholds: Setting Symbol Parameter Step Range 350…13 Top threshold of HR/PR, 1/min Bottom threshold of HR/PR, 1/min 12…349… -
Page 42
1.3 Device Control Setting Symbol Parameter Step Range Top threshold of expiratory gas concentration 0.5…10.0 EtAх* Bottom threshold of expiratory gas concentra- 0.1…10.0 tion Top threshold of inspiratory gas concentration 0.5…10.0 FiAх* Bottom threshold of inspiratory gas concentra- 0.1…10.0 tion Top threshold of systolic BP for the channel 1 (2), mmHg 0 … -
Page 43: Setting Of Ecg-Channel Parameters
1.3 Device Control 1.3.11. Setting of ECG-channel Parameters The parameters of the ECG module are controlled using fields and icons located in the cardi- ometry window: To change a parameter, click on the field of its current value and select the required new value in the appeared window.
-
Page 44
1.3 Device Control After switching on the ECG synthesis mode S12, leads V1 — V6 become available for selection to display. S12 lead synthesis mode is available only when using a 5-electrode ECG cable. To control the generation of messages about the heart rhythm deviations, click on the arrhyth- mias list icon. -
Page 45: Setting Of Respiratory Parameters Module
1.3 Device Control 1.3.12. Setting of Respiratory Parameters Module To set the sweep speed of the respiratory parameters module and the required apnea interval, click on the corresponding field of the values of these parameters located in the respirogram window: Press on the required sweep speed and set the apnea interval in accordance with p.1.3.10.
-
Page 46: Settings Of Pulse Oxymetry Module
1.3 Device Control 1.3.13. Settings of Pulse Oxymetry Module Control of the parameters of the pulse oximetry module (the sweep speed and averaging time of digital parameters) is carried out using the fields and icons in ECG module window: To change a parameter, click on the display field of its current value and select the required new value in the window that appears after clicking.
-
Page 47: Settings Of Nibp Module
1.3 Device Control 1.3.14. Settings of NIBP Module Set the units of BP measurement, the measurement period of the NIBP module and select au- tomatic or manual modes of blood pressure measurement by clicking on the corresponding fields in the window of the NIBP channel: Changing of the units of Setting of the BP measure- measurement mmHg / kPa…
-
Page 48: Settings Of The Thermometry Module
1.3 Device Control 1.3.15. Settings of the Thermometry Module In the thermometry channel you can switch the measurement mode on T2 channel to the tem- perature difference measurement mode T1-T2. To switch the mode, click on the mode name in the thermometry channel window: After switching on the T1-T2 mode, the value of the temperature difference between the chan- nels T1 and T2 will be displayed on the T2 display channel:…
-
Page 49: Settings Of The Capnometry Module
1.3 Device Control 1.3.16. Settings of the Capnometry Module The parameters of the capnometry channel are controlled using the fields and icons in the cap- nogram window: Capnometer Setting the apnea zero setting interval and compensation of the O according it. 1.3.10) To change a parameter, click on the display field of its current value and select the required new value in the window that appears after clicking.
-
Page 50: Settings Of Multigas Module
1.3 Device Control 1.3.17. Settings of Multigas Module The parameters of the gas analysis module are controlled using fields and icons in the window of the graphical curves of the multigas channel: Set gas parameters in the field of corresponding gas: To change a parameter, click on the display field of its current value and select the required new value in the window that appears after clicking.
-
Page 51: Settings Of The Ibp Module
1.3 Device Control 1.3.18. Settings of the IBP Module The parameters of the invasive blood pressure measurement (IBP) module are controlled using fields and icons located in the IBP graphic window: To change a parameter, click on the display field of its current value and select the required new value in the window that appears after clicking.
-
Page 52: System Reset
1.3 Device Control 1.3.19. System Reset To make system reset press the button “ ” before turning the device on. Do not release it until the graphs appear on the screen. The screen view and all the device parameters will take their default values.
-
Page 53: Alarms
1.4 Alarms 1.4. ALARMS 1.4.1. Alarm System The alarms generated by the device can be: caused by the patient’s condition (the patient’s parameters measured by the device go out the set thresholds); caused by incorrect operator actions when monitoring a parameter or malfunctions of the device (technical alarms).
-
Page 54: Alarm Messages
1.4 Alarms The sound signal of a technical malfunction can be silenced by pressing the » » button until the next technical malfunction appears. The volume of the technical fault signal can be set in the window for setting the volume of sound signals (see p.
-
Page 55: Silencing The Audible Alarm
1.4 Alarms 1.4.3. Silencing the Audible Alarm The device has the following options to disable the sound alarm: Immediate temporary silencing — press the alarm off symbol on the touch screen (Fig. 1-8, position 19) or the » » button on the front panel of the device. Flashing symbol appears on the screen and all sound alarms turns off for 2 minutes, after which it is automatically re- stored (preschedule unlocking can be done by pressing the same button again),…
-
Page 56: Alarms Of The Pulse Oximetry Module
1.4 Alarms 1.4.4. Alarms of the Pulse Oximetry Module Message Alarm situation Sound alarm Light alarm (output window) « — — — » instead Low (weak) strength of the pulse Weak signal Three short of SрО beeps every 5 s (general) Flashing SрО…
-
Page 57
1.4 Alarms Message Alarm situation Sound alarm Light alarm (output window) sensor failure Sensor failure! Two-tone every 5 (PPG) Technical alarms are highlighted with gray fill. * – this alarm is triggered if the device has previously recorded readings via the pulse oximetry channel. -
Page 58: Alarms Of The Ecg And Respiratory Parameters Modules
1.4 Alarms 1.4.5. Alarms of the ECG and Respiratory Parameters Modules Message Alarm situation (output Sound alarm Light alarm window) « — — — » instead of HR No cardiac activity Asystole Continuous signal (general) Low ECG voltage Low ECG volt- age (general) HR value is less than bottom thresh- Flashing HR value…
-
Page 59: Alarms Of Nibp Module
1.4 Alarms 1.4.6. Alarms of NIBP Module Message (output Alarm situation Sound alarm Light alarm window) SYS value is less than bottom threshold < Flashing SYS value SYS value is less than bottom threshold Two short The same on a on 8…16 % beeps every 5 yellow back-…
-
Page 60
1.4 Alarms Message (output Alarm situation Sound alarm Light alarm window) MAP value is more than top threshold < 4 Flashing MAP value MAP value is more than top threshold on Two short The same on a 4…8 % beeps every yellow back- 10 s ground… -
Page 61: Alarms Of The Thermometry Module
1.4 Alarms Message (output Alarm situation Sound alarm Light alarm window) « — — — » instead Malfunction of the measuring circuit of the Pressure channel Two-tone every NIBP channel failure (NIBP) of BP « — — — » instead Independent NIBP alarm system triggered NIBP module failure! Two-tone every…
-
Page 62: Alarms Of The Capnometry Module
1.4 Alarms 1.4.8. Alarms of the Capnometry Module Message (output Alarm situation Sound alarm Light alarm window) Flashing EtCО EtCО value is less than bottom threshold < 5 value EtCО value is less than bottom threshold on Two short The same on a 5…10 % beeps every 5 yellow back-…
-
Page 63
1.4 Alarms Message (output Alarm situation Sound alarm Light alarm window) There were 3 unsuccessful attempts to purge Off: Occlusion Two-tone every the path (capno) Failed to dry the measuring chamber after 10 Off: Purging Fail Two-tone every (capno) Failure of the measuring chamber emitter Off: Emitter Fail Two-tone every (capno) -
Page 64: Alarms Of The Multigas Module
1.4 Alarms 1.4.9. Alarms of the Multigas Module Message (output Alarm situation Sound alarm Light alarm window) EtXXX value is less than bottom threshold Flashing EtXXX < 5 % value EtXXX value is less than bottom threshold Two short beeps The same on a on 5…10 % every 5 s…
-
Page 65
1.4 Alarms Message (output Alarm situation Sound alarm Light alarm window) Turned off manually by the «CAPNO» button Off by user (multi- or No breath signal at the multigas input gas) more than 1 hour Technical alarms are highlighted with gray fill. XXX –… -
Page 66: Alarms Of The Ibp Module
1.4 Alarms 1.4.10. Alarms of the IBP Module Message (output Alarm situation Sound alarm Light alarm window) SYS value is less than bottom threshold Flashing SYS < 8 % value of IBP1 (IBP2) channel SYS value is less than bottom threshold Two short beeps The same on a on 8…16 %…
-
Page 67: Measurement Of Parameters
2.1 SpO Monitoring 2. MEASUREMENT OF PARAMETERS 2.1. MONITORING OF ARTERIAL OXYGEN SATURATION (SpO Below in it. 2.1.1 — 2.1.3 there is a description if procedure for monitoring SpO using the «Triton» pulse oximetry module (TESM.505008). It. 2.1.4 describes the procedure for monitoring SpO using the Masimo SET pulse oximetry module, which can be optionally installed in the device.
-
Page 68: Application Of Finger Spo Sensors
2.1 SpO Monitoring 2.1.1. Application of Finger SpO Sensors Figure 2-1. Finger sensors (“clip” and rubber) Use a finger on a hand that does not have a pressure cuff or an arterial catheter. Choose a finger with a good pulse strength and the most appropriate sensor in size (if nail is polish, remove the polish before installing the sensor).
-
Page 69
2.1 SpO Monitoring Figure 2-3. Use of the neonatal sensor for newborns When installing a disposable neonatal sensor b) (Fig. 2-2 b, Fig. 2-3), it is necessary to locate the windows of the sensor’s photocells opposite each other. In the case of using the sensor of variant a) in older children or adult patients, the sensor can be located on the finger with fixation by the sensor strap (Fig. -
Page 70: Window Of The Pulse Oxymetry Channel
2.1 SpO Monitoring 2.1.3. Window of the Pulse Oxymetry Channel Display the photoplethysmogram curve in one of the configurable graphical curves display windows (section 1.3.5): filling column; photoplethysmogram (PPG); control icon for SpO sound alarm; top threshold of SpO bottom threshold of SpO measured value of SpO control symbol of HR/PR source.
-
Page 71
2.1 SpO Monitoring Depending on the patient’s mobility, set the most appropriate averaging time (the time after which the measured SрО value is determined): 4, 8 or 16 s (the default value is 16 s, which is most suitable for restless patients) (p. 1.3. 8). Patient movement can distort the sensor signal. -
Page 72: Spo Monitoring Using Masimo Set Technology
2.1 SpO Monitoring ® 2.1.4. SpO Monitoring using Masimo SET technology Description This section provides information on how to operate a device equipped with a pulse oximetry module using Masimo SET® technology (MS-2040) to monitor: saturation of arterial blood with oxygen, SpO photoplethysmogram, PPG heart rate obtained from photoplethysmogram, PR perfusion (filling) index — the numerical value of the pulsating component of the regis-…
-
Page 73
2.1 SpO Monitoring Grounding. Connect the device only to a three-wire, grounded, hospital-grade receptacle. The three-conductor plug must be inserted into a properly wired three-wire receptacle; if a three-wire receptacle is not available, a qualified electrician must install one in accordance with the governing electrical code. -
Page 74
2.1 SpO Monitoring The «Sensitivity» parameter is adjusted in the same way and has 3 values: «Max» (Maximum), «Norm» (Normal), «APOD» (Adaptive Probe Off Detection — adaptive detection of the absence of a sensor). The «Norm» sensitivity mode provides the best combination of sensitivity and sensor discon- nection detection. -
Page 75
2.1 SpO Monitoring (exposure to excessive illumination can be corrected by covering the sensor with a dark or opaque material) Excessive patient movement. Venous pulsations may cause erroneous low readings (e.g. tricuspid value regurgitation). Patient suffers from abnormal pulse rhythm. … -
Page 76
2.1 SpO Monitoring LNSC: DC-IP, YI, Adtx, Pdtx, Inf, NeoPt For sensors LNOP TC-I, Blue, LNCS TC-I, TF-I the error is not standardized 2 % measurement error (at low per- 4, 5 fusion) LNOP: DC-I, DC-IP, Adt, Pdt, Neo L (adult, ped.), Inf-L, LNSC: DC-IP, Adtx, Pdtx, Inf, Neo (adult, ped.), TF-I 3 % LNOP: Neo L (neonatal), NeoPt-L, Blue… -
Page 77
2.1 SpO Monitoring Sensors LNOP Blue Inf, LNOP TF-I, TC-I and LNCS TF-I, TC-I are not used for measurements in the presence of movement. Sensor LNOP YI is not used for measurements at low perfusion. SENSORS Use only Masimo oximetry sensors for SpO2 measurements. … -
Page 78
2.1 SpO Monitoring Sensor Description Application site Weight range 1864 LNCS DC-IP Pediatric reusable finger sensor Finger 10 — 50 kg 1895 LNCS TC-I Reusable ear sensor Ear lobe > 30 kg Foot 1- 3 kg * 2258 LNCS YI Reusable multi-site sensor Great toe >… -
Page 79
2.1 SpO Monitoring Alarm situation Message Sound alarm Light alarm Two short beeps The same on a SрО value is more than top threshold by every 5 s yellow back- 2…4 % ground Three short beeps The same on a SрО… -
Page 80
2.1 SpO Monitoring eliminate sources of interference, high illumination Low Signal IQ The quality of the signal does not allow to obtain reliable measurements Masimo SET End User License Agreement The present agreement is made between the End User (hereinafter referred to as the Purchaser) of the equipment and the Equipment Manufacturer Triton Electronic Systems Ltd. -
Page 81
2.1 SpO Monitoring formance, display or disclosure of the Software (including the related documentation) by the U.S. Gov- ernment or any of its agencies shall be governed solely by the terms of this Agreement and shall be pro- hibited except to the extent expressly permitted by the terms of this Agreement. -
Page 82: Disinfection Of Spo Sensors
2.1 SpO Monitoring 2.1.5. Disinfection of SpO Sensors Disinfection of sensors is carried out in accordance with the requirements of MU-287-113 by wiping them with a 3% solution of hydrogen peroxide with the addition of 0.5% neutral detergent of the «Lotos» type (it is allowed to use other disinfectants of different manufacturers, taking into account their methodology applications).
-
Page 83: Ecg Monitoring
2.2 ECG Monitoring 2.2. ECG MONITORING This icon indicates that the device and patient cable are specially pro- tected against electric shock and defibrillator shock (type CF). To connect the cable connector to the device, grasp the connector housing and firmly insert it into the corresponding connector of the device until it stops.
-
Page 84
2.2 ECG Monitoring Figure 2-9. ECG electrode application scheme Gently squeeze the flaps of the colored patient cable pins and place them on the electrodes in accordance with the selected application pattern. Make sure the clothespins are securely con- nected to the electrodes by gently pulling on the cable wires. After connecting the patient cable, ALWAYS make sure that all leads are applied to the patient and do not touch other conductive parts or ground. -
Page 85: Ecg Channel Window
2.2 ECG Monitoring 2.2.2. ECG Channel Window field of ECG scale; field for setting the lead for the displayed curve; calibration bar; the value of ST-segment offset; field for setting the operating mode of the ECG channel filter; field for setting the type of ECG cable; field for setting the sweep of ECG, PPG, BP curves;…
-
Page 86: Ecg Monitoring Procedure
2.2 ECG Monitoring 2.2.3. ECG Monitoring Procedure Set the appropriate type of ECG-cable (it. 1.3.11). Select the required operating profile (it. 1.3.9). Register the ECG waveform and the appearance of the heart rate value on the device screen. Set the required number of ECG curves (it. 1.3.11). When a cable with three electrodes is selected, only one ECG waveform can be displayed.
-
Page 87
2.2 ECG Monitoring a cutoff frequency of 30 Hz are turned on, these filters prevent strong suppression of QRS complexes by noise, which interferes with their correct assessment; o Diagn. – used for a more detailed analysis of the ECG, in particular, control of the dynamics of the ST segment. -
Page 88: Arrhythmia Analysis
2.2 ECG Monitoring 2.2.4. Arrhythmia Analysis Analysis of arrhythmias allows staff to respond in a timely manner to changes in the patient’s cardiac activity. For the correct operation of the arrhythmia analysis algorithm, it is neces- sary to ensure good contact of the ECG electrodes and minimize the pa- tient’s movements.
-
Page 89
2.2 ECG Monitoring Level/ Message Conditions of formation alarm signal MULTIFORM At least two types of VPCs with a total number of VPCs of at least 3 — in Medium the last minute. The types of VPCs differ in shape, mainly in polarity. MISSED At least 2 of RR-intervals, 2.5 times higher than normal RR-interval — in BEATS… -
Page 90
2.2 ECG Monitoring The arrhythmia analysis option provides automatic recording of waveform fragments for recording and subsequent viewing of alarm situations. During recording, the REC icon appears below the HR value. Fragments are recorded when a message appears or when the message is changed to a higher priority one, which means a new alarm situation. -
Page 91: Analysis Of Heart Rate Variability (Hrv)
2.2 ECG Monitoring 2.2.5. Analysis of Heart Rate Variability (HRV) The device provides the output of graphical and numerical information for the HRV analysis (see Fig. 2-12): Figure 2-12. Window of HRV analysis For correct operation of HRV analysis, it is necessary to select the HR / PR display mode by the ECG channel (it.
-
Page 92: Synthesis Of Ecg Leads
2.2 ECG Monitoring 2.2.6. Synthesis of ECG Leads This feature allows to synthesize ECG in 12 standard leads using a five-electrode configuration. ECG curves in 12 leads and corresponding measurements obtained from lead synthesis are approximations to the standard 12-lead ECG. Since these 12-lead ECGs are not exactly the same electrocardiograms that are read from an electrocardiograph with a standard ECG, they shall not be used for diagnostic interpretation.
-
Page 93: Features Of The Operation In The Operating Rooms
2.2 ECG Monitoring Normal detection of PM is guaranteed only for impulses with the following parameters: duration 1.5 — 15.5 ms, amplitude not less than 0.6 mV. Such impulses must be present in at least one lead. If the parameters of impulses in all leads are outside the specified limits, the device may not recognize the presence of the pacemaker.
-
Page 94: Respiratory Rate Measurement By Respirogram
2.2 ECG Monitoring 2.2.10. Respiratory Rate Measurement by Respirogram Respiratory rate is measured by the impedance method with the same patient cable and set of electrodes used for ECG monitoring. The recommended electrode placement pattern for respiration monitoring is the same as for ECG monitoring (see Figure 2-9).
-
Page 95: Temperature Monitoring
2.3 Temperature Monitoring 2.3. TEMPERATURE MONITORING There are two types of thermistor temperature sensors intended for the temperature monitoring: surface (skin) (see it. 2.3.1); intracavitary (universal) (it. 2.3.2). For a reliable connection with the device, the sensor connectors have a special design that automatically provides mechanical fixation.
-
Page 96: Application Of A Surface Temperature Sensor
2.3 Temperature Monitoring 2.3.1. Application of a Surface Temperature Sensor The surface (skin) temperature sensor is made in the form of a tablet (see Fig. 2-17). It is in- stalled on any part of the patient’s body and is fixed with a strip of adhesive plaster, and can al- so be located without additional fixation in large natural skin folds (armpit, inguinal fold, under the mammary gland in women) or simply placed under the patient’s body.
-
Page 97: Test Of Measurement Accuracy Of The Temperature Sensors
2.3 Temperature Monitoring Figure 2-18. Intracavitary temperature sensor When using the sensor for monitoring the central body temperature, it is introduced into the se- lected cavity, and to facilitate insertion, the surface of the sensor and the cable should be lubri- cated for the required length (sterile glycerin or vaseline oil).
-
Page 98
2.3 Temperature Monitoring neutral detergent of the «Lotos» type (it is allowed to use other disinfectants of different manu- facturers, taking into account their methodology applications). DO NOT ALLOW liquid to get on the sensor connector pins, this can lead to its failure. -
Page 99: Non-Invasive Blood Pressure Measurement (Nibp)
2.4 NIBP Measurement 2.4. NON-INVASIVE BLOOD PRESSURE MEASUREMENT (NIBP) 2.4.1. Cuff Selection and Placement The oscillometric method used for the non-invasive blood pressure (BP) measurement requires reliable registration of the pulse blood pressure fluc- tuations. If they obtained heavily, then the measurement time increases, and the measurement itself becomes unreliable.
-
Page 100
2.4 NIBP Measurement 18…26 Pediatric (18-26 cm) 25…33 Adult (25-40 cm) 34…43 Large adult (34-51 cm) 40…66 On the thigh (40-66 cm)* * — supplied by a special order For example, the «Adult (25 — 40 cm)» cuff (included in the delivery kit of the monitor) with an in- flatable balloon width of 12 cm shall be used on a patient with a shoulder circumference below 33 cm. -
Page 101: Non-Invasive Measurement And Monitoring Of Blood Pressure
2.4 NIBP Measurement 2.4.2. Non-Invasive Measurement and Monitoring of Blood Pressure The non-invasive blood pressure measurement channel has three modes of operation, deter- mined by the type of profile set in the device: neonatal, pediatric and adult. The BP measure- ment ranges for them are the following: adult 0 — 300 mmHg pediatric 10 — 200 mmHg…
-
Page 102
2.4 NIBP Measurement For children of older age groups, it is allowed to take measurements in adult mode in the range of 0 — 300 mmHg using the appropriate cuffs. Make sure that the patient has really high blood pressure. The device has increased resistance to artifacts caused by patient motion, external influences on the cuff, as well as to the effect of arrhythmia, however, take into account that this reduces the measurement accuracy and increases the pumping pressure and measurement time. -
Page 103
2.4 NIBP Measurement During measurements, make sure that the limb where the cuff is installed is always at the level of the heart, otherwise the measurement results may be distorted. You can control the process of measuring blood pressure in the NIBP channel window (see Fig. 2-21), while during the measurement you can observe the current value of the pressure in the cuff (Fig. -
Page 104: Check Of The Cuff Tightness
2.4 NIBP Measurement The device has protection for the maximum pressure in the cuff (< 300 mmHg for adults and 150 mmHg for newborns). The device has protection for the maximum pressure measurement time (<180 s for adults and 90 s for newborns). 2.4.3.
-
Page 105: Malfunctions And Diagnostic Messages
2.4 NIBP Measurement 2.4.5. Malfunctions and Diagnostic Messages In the event of deviations from the normal operation, check the correspondence of the selected operating profile of the device (neonatal / pediatric / adult) to the used cuff, the size and con- stitution of the patient, while also paying attention to the text messages that appear in the win- dow of the NIBP channel: Message…
-
Page 106
2.4 NIBP Measurement Message Description of the situation Troubleshooting No calibration! Loss of calibration constants Contact customer service. in the NIBP channel. Alarm system failure! Malfunction of the autono- Contact customer service. mous safety system of the NIBP channel. Compressor failure! Malfunction of the NIBP Contact customer service. -
Page 107: Measurement Of Со , Оand Anesthetics Concentration
2.5 Measurement of CO and Anesthetics Concentration 2.5. MEASUREMENT OF СО , О AND ANESTHETICS CONCENTRA- TION The device measures and displays: СО concentation at the end of expiration (EtСО СО concentration at the inspiration (FiСО capnogram, respiratory rate (RR) measured by capnogram. By special order, the device additionally measures and displays: О…
-
Page 108
2.5 Measurement of CO and Anesthetics Concentration The outlet nipple (Fig. 1-3, position 6) is located on the side wall of the device and serves to discharge the test sample into the atmosphere (in the presence of volatile anesthetics in the an- alyzed sample, it shall be connected to the ventilation system of a medical institution). -
Page 109: Monitoring The Gas Concentration In The Breathing Mixture
2.5 Measurement of CO and Anesthetics Concentration 2.5.2. Monitoring the Gas Concentration in the Breathing Mixture The respiration curve (capnogram) is displayed in the capnometry channel window (see Fig. 2- 25), the measured values EtCO , FiCO — in the window of gas analysis parameters (Fig. 2-26). Respiration rate (RR) value measured according with the capnogram is displayed in the RR window (Fig.2-15), if the source of RR is selected as “capnogram”…
-
Page 110
2.5 Measurement of CO and Anesthetics Concentration In the absence of a N O measurement channel and in the presence of ni- trous oxide in the breathing mixture, compensation of its influence on the accuracy of CO measuring is performed by manual setting the concentra- tion of N O in the corresponding field of the capnometry channel window (Fig. -
Page 111
2.5 Measurement of CO and Anesthetics Concentration Figure 2-29. Diagram of using a filter without the output for gas analysis If moisture gets on the optical elements of the measuring chamber, a sudden rise in the capno- gram level (see Fig. 2-30) and overestimated digital readings of EtCO and FiCO can occur. -
Page 112: Self-Testing System Of The Capnometer
2.5 Measurement of CO and Anesthetics Concentration 2.5.3. Self-Testing System of the Capnometer The self-diagnostic system monitors the operating conditions of the module and, in necessary cases (occlusion of the path or ingress of moisture into it), automatically takes measures to eliminate the reasons that prevent its normal operation.
-
Page 113: Check Of Operation And Measurement Accuracy Of The Capnometer
2.5 Measurement of CO and Anesthetics Concentration 2.5.4. Check of Operation and Measurement Accuracy of the Capnometer The simplest test of the performance of a capnometer can be carried out by breathing at the end of the sampling line, while the screen displays the breathing cycle (capnogram). EtCO readings for a healthy person usually about 4 — 5%.
-
Page 114: Drying Of The Capnometer Pneumatic Path
2.5 Measurement of CO and Anesthetics Concentration In case of failure, contact the customer service support to find the cause of the leakage and eliminate it. During tightness test, the path should be closed for a short time (< 7 — 8 s and no more than once within 7 — 8 s), otherwise the module will automati- cally shut down.
-
Page 115: Monitoring Of The Inhaled Anesthetics Concentration
2.5 Measurement of CO and Anesthetics Concentration 2.5.8. Monitoring of the Inhaled Anesthetics Concentration By special order, the device can be equipped with a breathing gas analysis module with the function of volatile anesthetics concentration monitoring (hereinafter the gas analysis module). In this case, the device displays: concentration of СО…
-
Page 116
2.5 Measurement of CO and Anesthetics Concentration Preparing for monitoring of the O and inhalation anesthetics concentration is carried out in the same way as for CO described in it 2.5.1. No additional accessories or changes to the patient connection are required. The monitoring procedure is described below. 2.5.8.1 Preparing for the Monitoring Switch on the device. -
Page 117
2.5 Measurement of CO and Anesthetics Concentration o check and, if necessary, set in the general parameters menu (p. 1.3.8) the con- centration factors 1MAC for each gas to calculate the total MAC value (minimum alveolar concentration); o check the tightness of the pneumatic path by short-term clamping of the sampling line or its blockage (see it. -
Page 118
2.5 Measurement of CO and Anesthetics Concentration field for setting the sweep speed, mm/sec; field for setting the apnea interval, sec; field for setting the value of water vapor pressure compensation; field for setting the output channel; scale of O and N O display;… -
Page 119
2.5 Measurement of CO and Anesthetics Concentration When the device is turned on and during operation, the gas analysis module automatically conducts calibration cycles with the message «CALIBRA- TION..» displayed on the screen. Their frequency changes as the tempera- ture mode of the measuring chamber is established. At first they are carried out more often, and about 10 minutes after turning on the gas analysis — every 5 — 10 minutes. -
Page 120
2.5 Measurement of CO and Anesthetics Concentration leakage in the sampling system due to worn parts; incorrect measurement of the zero level due to gas leakage in the area after the pump; malfunction of the zero valve, the valve does not open as a result; too high moisture content of the sample due to aging of the Nafion filter tubes;… -
Page 121
2.5 Measurement of CO and Anesthetics Concentration Check that the concentration is within the limits specified in the manufacturer’s specification for the gas analysis module (see the table at the beginning of the current section). If the data is out of specification, the module must be calibrated or repaired. Note: during the test it is allowed to bring the end of the sampling line to the outlet of the reference gas cylinder so that there was no influence of the ambient atmospheric air. -
Page 122
2.5 Measurement of CO and Anesthetics Concentration D) Calibration of the capnometry module During operation, after replacement of the oxygen sensor or if the measurement error increases, calibration may be required. Calibration is carried out with atmospheric air (approximate oxygen content 21%). -
Page 123: Disinfection Of The Accessories Of Multigas Module
2.5 Measurement of CO and Anesthetics Concentration Message Description of the situation Troubleshooting OCCLUSION Occlusion of the pneumatic path, while an attempt Remove the reason of occlu- is automatically made to purge the path. If the path- sion (replace or purge the line).
-
Page 124: Invasive Blood Pressure Measurement (Ibp)
2.6 IBP Measurement 2.6. INVASIVE BLOOD PRESSURE MEASUREMENT (IBP) Pressure sensors with a sensitivity of 5 mV / mmHg and resistance 300 — 3000 Ohm are used for measurements. Preparing for measurement: Connect the pressure sensor cable to the monitor via the connecting cable. Prepare the flushing solution (saline) and connect the container with the solution to the pressure sensor.
-
Page 125: Ibp Window
2.6 IBP Measurement 2.6.1. IBP Window scale of IBP curves; number of IBP channel; button for enabling calibration; IBP curve label; cutoff frequency of the IBP curve filter; field for selecting the output channel; top alarm threshold for systolic blood pressure; number of IBP channel;…
-
Page 126
2.6 IBP Measurement To perform the calibration, click on the calibration button in the window for displaying the graph- ical pressure curve of the corresponding channel (Fig. 2-37, position 3) and in the dialog box, select the required action: After starting the calibration, wait for the pressure reading to zero (about 20 s). You can assign the type of pressure being measured using a label. -
Page 127: Disinfection Of The Pressure Transducer Cable
2.6 IBP Measurement 2.6.3. Disinfection of the Pressure Transducer Cable Disinfection of cables is carried out in accordance with the requirements of MU-287-113 by wip- ing them with a 3% solution of hydrogen peroxide with the addition of 0.5% neutral detergent of the «Lotos»…
-
Page 128: Additional Features
3.1 Trends 3. ADDITIONAL FEATURES 3.1. TRENDS Trends are data on the monitored parameters of a patient, recorded over a certain period of time and displayed in graphical or tabular form to represent the picture of the development of the patient’s condition over time. The data recorded by the device about the monitored parameters (SpO2, PR, PERF, STDev, RSP with a period of 15 s, the rest with a period of 1 min) automatically are stored in the internal memory from the moment of registration within 240 hours or before erasing (for example, when…
-
Page 129: Graphical Trends
3.1 Trends 3.1.1. Graphical Trends The device can register the following trends: saturation (SрО ) – red line; heart rate (HR) – green line; systolic, mean, diastolic non-invasive blood pressure (NIBP) — yellow line; two channels of temperature (Т1 and Т2) – dark blue line; СО…
-
Page 130
3.1 Trends If the device has additional options, it is possible to register graphical trends of the cardiac index (CI), brain activity index (AI), oxygen consumption VO2, resting energy expenditure REE (it. 3.8 — 3.13). When entering the trend window after turning on the device, the last page is always displayed, and the cursor shows the current time and date. -
Page 131: Tabular Trends
3.1 Trends 3.1.2. Tabular Trends Using the tabular trends, you can view the following parameters recorded by the device: heart rate (HR) – /min; systolic, diastolic and mean non-invasive blood pressure (NIBP); ) – %; saturation (SpO pulse filling (PERF) – %; ST-segment offset (StDv) –…
-
Page 132
3.1 Trends storage of information in the computer memory, as well as the ability to print it out on an external printer. The device has the ability to print tabular trends on the built-in printer (see it. 3.1.4). An example of a printout of tabular trends is shown in Fig. -
Page 133: Recording And Viewing Waveform Fragments
3.1 Trends 3.1.3. Recording and Viewing Waveform Fragments The device can store in memory up to 30 fragments of curves with duration of 10 s each (5 s before the moment of recording and 5 s after). There are 2 modes for recording waveform frag- ments: manual — recording occurs when you press the «FREEZE»…
-
Page 134: Built-In Thermal Printer
3.1 Trends date and time of recording of the printed fragments of curves; device identifier (model, configuration, serial number or MAC address); time scale in seconds relative to the moment of recording the fragments of the curves (5 seconds before the moment of recording and 5 seconds after); the moment of recording a fragment of curves (figure «0»…
-
Page 135: Oxycardiorespirogram
3.1 Trends If the paper color on the printer symbol is red, there is no paper in the printer, or the printer is faulty. The procedure for replacing paper in a thermal printer is described in Annex 1. When operating on battery power, limit printing to the built-in thermal print- er as much as possible to extend the operating time of the device.
-
Page 136: Log Of Alarms And Events
3.1 Trends 3.1.6. Log of Alarms and Events The device optionally can store and display all alarms and some events in chronological order. The latest entries are at the top of the log. When you select the «Event / alarm log» icon, the device allows to view a list of physiological alarms, technical alarms, as well as events for the past time, indicating the name of the alarm, priority (the message is colored in different colors depending on the priority), time and date of occurrence in DD / MM / YY HH: MM format.
-
Page 137
3.1 Trends You can also choose the display mode: — to show alarms and events; — to show only physiological alarms of high and medium priority. -
Page 138: Digital Mode
3.2 Digital mode 3.2. DIGITAL MODE In this mode the display shows only 5 main digital indicators of the patient’s condition — EtCO heart rate, SpO , RR and blood pressure, which allows you to significantly increase the size of the displayed numerical values to facilitate their observation at a distance (see Fig.
-
Page 139: Calculators
3.3 Calculators 3.3. CALCULATORS By special order the device can have an option of calculators. Menu of calculators is unavailable in the FREEZE mode. The device has the following calculators: drug dose calculator and titration table; hemodynamic calculator (calculation of hemodynamic parameters); oxygenation calculator (calculation of parameters of arterial blood oxygenation, oxy- gen delivery and extraction);…
-
Page 140: Dose Calculator And Titration Table
3.3 Calculators 3.3.1. Dose Calculator and Titration Table Setting digital parameters in accordance with it.1.3.10. To calculate the dose of the drug: Enter patient weight. Select a drug. The dose calculator contains a library of commonly used drugs and allows you to configure a drug list on the monitor with predefined dosage recommendations and unit settings.
-
Page 141
3.3 Calculators When specifying a drug that is not listed in the library, correctly indicate the name of the drug (A, B, C, D or E) in accordance with the units of measurement: Drug Unit of measurement Dose change limits Drug А… -
Page 142: Hemodynamic Calculator
3.3 Calculators 3.3.2. Hemodynamic Calculator Setting digital parameters in accordance with Entered manually or it.1.3.10 measured by the device Setting digital parameters in accordance with it.1.3.10 Some of the input calculated parameters can be entered by the user, or measured by the de- vice, as indicated by the highlighted (manual input) or darkened (measured by the device) icon next to the input parameter.
-
Page 143: Oxygenation Calculator
3.3 Calculators 3.3.3. Oxygenation Calculator Some of the input calculated parameters can be entered by the user, or measured by the de- vice, as indicated by the highlighted (manual input) or darkened (measured by the device) icon next to the input parameter. To switch the input mode — click on the icon. Entered manually or measured by…
-
Page 144: Ventilation Calculator
3.3 Calculators alveolar-arterial oxygen difference, AaDO oxygen delivery, DO 3.3.4. Ventilation Calculator Some of the input calculated parameters can be entered by the user, or measured by the de- vice, as indicated by the highlighted (manual input) or darkened (measured by the device) icon next to the input parameter.
-
Page 145: Calculator Of Kidney Function
3.3 Calculators 3.3.5. Calculator of Kidney Function The calculator of the kidney function is intended for evaluation of GFR (glomerular filtration rate, or creatinine clearance) using the Cockroft-Gault formula in the adult profile or the Schwartz formula in the pediatric and neonatal profiles. Setting digital parameters in accordance with…
-
Page 146: Built-In Battery
3.4 Built-in Battery 3.4. BUILT-IN BATTERY The device has a backup power supply (built-in battery with a charger), which ensures autono- mous uninterrupted operation of the device in case of loss of external power supply. The presence of a built-in battery in the device requires taking into account the features of its operation.
-
Page 147: Battery Charging
3.4 Built-in Battery 3.4.2. Battery Charging The battery is charged automatically when there is a mains power, regardless of the operation of the device, however, it must be connected to the mains, that is accompanied with the green «POWER» indicator on the front panel of the device (see Fig. 1-2, position 3). The “BATTERY”…
-
Page 148: Troubleshooting
3.4 Built-in Battery 3.4.4. Troubleshooting Malfunction Possible cause Troubleshooting Time of device operation from Battery is undercharged Fully charge the battery battery decreased Blood pressure measurement is Increase the period between activated too often blood pressure measurements Reducing of the battery capaci- Train the battery, if unsuccess- ful — replace The device does not operate…
-
Page 149: Connecting To On-Board Power Supply
3.5 Connecting to On-Board Power Supply 3.5. CONNECTING TO ON-BOARD POWER SUPPLY By a special reques, the devices can be equipped with equipment for connecting to the on- board power supply of an ambulance car with a voltage of 12 … 15 V. Connection is provided using an additional «Board power»…
-
Page 150: Connecting To The Computer Network
3.6 Connecting to the Computer Network 3.6. CONNECTING TO THE COMPUTER NETWORK At the customer’s request, the device can be equipped with the function of connecting to a com- puter network for operation as a part of the centralized monitoring system CMS-Triton (supplied by separate order).
-
Page 151: Connecting Of The External Usb Flash Memory
3.7 Connecting of the External Usb Flash Memory 3.7. CONNECTING OF THE EXTERNAL USB FLASH MEMORY The device can be equipped with the function of saving data (trends and fragments of curves) on an external USB flash memory device, which is connected to the USB connector located on the rear panel of the device, for the further data viewing on a computer.
-
Page 152: Installation Equipment
3.8 Installation Equipment 3.8. INSTALLATION EQUIPMENT On request the device can be equipped with a set of installation equipment. 3.8.1. Wall Mount Installation Equipment 2. Remove the KROMAX TECHNO-8 Bracket 1. Check the installation kit from the package 3. Remove the plate from the bracket 4.
-
Page 153
3.8 Installation Equipment 6. Mount the bracket on the wall according to 7. Mount the monitor on the bracket the instructions in the kit. 9. Adjust the angle of the monitor and tighten the 8. Fix handle 10. Installation of equipment completed… -
Page 154: Installation Equipment For Trolley Mounting
3.8 Installation Equipment 3.8.2. Installation Equipment for Trolley Mounting 1. Assemble the trolley from the installation 2. For the quick release option, attach the quick kit: release adapter to the monitor with four screws: 3. Place the monitor on the trolley: — fix the trolley with wheel brakes — bring the monitor to the mounting plate on the trolley…
-
Page 155: Non-Invasive Measurement Of Hemodynamic Parameters
3.9 Non-Invasive Measurement of Hemodynamic Parameters 3.9. NON-INVASIVE MEASUREMENT OF HEMODYNAMIC PARAME- TERS This option is supplied by a special order and allows to monitor the following hemodynamic pa- rameters by the method of volume compression oscillometry: 1.0 – 20.0, step 0.1; cardiac output (CO), lpm stroke volume (SV), ml 10 — 250, step 1;…
-
Page 156: Application Rules Of The Channel Of Non-Invasive Measurement Of Hemodynamic Parameters
3.9 Non-Invasive Measurement of Hemodynamic Parameters To provide more favorable conditions in the mode of determining hemody- namic parameters, the pumping rate drops and the maximum pumping pressure rises, so the pump and release time increase. Alarm threshold setting limits: Sign Parameter Setting range…
-
Page 157: Result Reliability Assessment
3.9 Non-Invasive Measurement of Hemodynamic Parameters The patient shall not talk or move during measurement the parameters. Any movements will affect the recorded curve and may cause distortion of the results, so one or more additional measurements of hemodynamic parameters will be required. Cuff size and positioning The width of the cuff should cover at least 40% of the circumference of the shoulder and at least 80% of its length.
-
Page 158
3.9 Non-Invasive Measurement of Hemodynamic Parameters You should take into account that specific oscillograms of the standard form may outwardly dif- fer noticeably from each other — in terms of «density» (the number of oscillations per unit of oscil- logram length), slope steepness and other signs. Differences between specific oscillograms of the standard type are associated with the individu- al characteristics of the subjects (blood pressure value, pulse rate, degree of respiratory waves), as well as with the conditions for using the channel (including the rate of air pumping into the… -
Page 159
3.9 Non-Invasive Measurement of Hemodynamic Parameters Figure 3-14. Oscillograms at atrial fibrillation If you get the following oscillograms (Fig. 3-15) restart the channel: record is not fin- distortion in the ished second section record is not distortion in the finished second section distortion in the forth distortion in the forth… -
Page 160: Measurement Of Hemodynamic Parameters By Ncov Channel
3.9 Non-Invasive Measurement of Hemodynamic Parameters 3.9.4. Measurement of Hemodynamic Parameters by NCOV Channel Cardiac index, l/min/m Cardiac output, l/min Strike volume, ml Strike index, ml/m Systemic vascular resistance, syn·sec·cm Oscillogram Systolic blood pressure, mmHg Diastolic blood pressure, mmHg Mean blood pressre, mmHg Left ventricular stroke power, W 10 — Figure 3-16.
-
Page 161: Alarms Of Non-Invasive Measurement Of Hemodynamic Parameters Channel
3.9 Non-Invasive Measurement of Hemodynamic Parameters 3.9.5. Alarms of Non-invasive Measurement of Hemodynamic Parameters Channel Message Alarm situation (output Sound alarm Light alarm window) CI value is less than bottom CI below Flashing CI value threshold < 4% threshold CI value is less than bottom Two short beeps every 5 s The same on a yel- threshold on 4…8%…
-
Page 162: Continuous Non-Invasive Blood Pressure Measurement ( Cnibp)
3.10 Continuous Non-Invasive Blood Pressure Measurement 3.10. CONTINUOUS NON-INVASIVE BLOOD PRESSURE MEAS- UREMENT (cNIBP) Continuous NIBP (сNIBP) mode allows continuous assessment of changes in blood pressure by a non-invasive method. The main estimated parameters in this mode: systolic arterial pressure; diastolic arterial pressure;…
-
Page 163: Dpwtt Measurement And Nibp Trigger Conditions
3.10 Continuous Non-Invasive Blood Pressure Measurement Figure 3-18. A particular case of PWTT dependence on systolic blood pressure To determine the coefficients of the line, it is necessary to make at least two calibration meas- urements of blood pressure by the classical method with the simultaneous registration of PWTT. The greater the difference between blood pressure and PWTT between the measurement re- sults, the more accurately the coefficients are determined.
-
Page 164
3.10 Continuous Non-Invasive Blood Pressure Measurement After each calibration measurement of BP, the reference value PWTT corresponding to that measurement is calculated. The difference between the current PWTT and the reference value PWTT is taken as an informative parameter of the cNIBP channel: dPWTT = PWTT — PWTT dPWTT is measured in milliseconds. -
Page 165: Continuous Arterial Blood Pressure Calculation
3.10 Continuous Non-Invasive Blood Pressure Measurement Then the dPWTT threshold decreases very slowly, which ensures that blood pressure measurement starts even with a stable PWTT value. Due to this, the maximum interval be- tween blood pressure measurements does not exceed 3-4 hours When setting a fixed threshold dPWTT, an extraordinary start of blood pressure will be per- formed after dPWTT goes beyond the specified limits, regardless of other factors.
-
Page 166: Operation Of Cnibp Mode
3.10 Continuous Non-Invasive Blood Pressure Measurement 3.10.4. Operation of cNIBP Mode The turning on the cNIBP mode is possible only in the automatic blood pres- sure measurement mode. When the cNIBP mode is on, the PWTT icon appears in the classic NIBP window. To enable the cNIBP mode, it is necessary to enable the cNIBP channel window in one of the configurable graphical windows for displaying curves (it.
-
Page 167: Limitations Of Cnibp Method
3.10 Continuous Non-Invasive Blood Pressure Measurement Switching on the cNIBP mode and setting the parameters of its operation is carried out using the fields and icons located in the cNIBP channel window: 3.10.5. Limitations of cNIBP Method The method of continuous measurement of NIBP used in the device has some restrictions on the conditions of its use.
-
Page 168
3.10 Continuous Non-Invasive Blood Pressure Measurement Message Description of the situation Troubleshooting PWTT:No One or more ECG electrodes reset, ECG ca- Connect the ECG cable to the device and ECG signal ble not connected / or to the patient. Check if the ECG cable leads are connected to the ECG elec- trodes. -
Page 169: Mainstream Co Sensor
3.11 Mainstream CO Sensor 3.11. MAINSTREAM CO SENSOR This module is supplied with the device as an option on a separate order and enables to monitor: concentration of СО2 at the expiration end (EtСО2), inspiratory CO2 concentration (FiCO2), capnogram, respiratory rate (RR), determined by capnogram. The operation principle of the mainstream CO2 sensor is based on the method of measuring the absorption of infrared radiation in the absorption spectrum of carbon dioxide.
-
Page 170: Preparing For Co Concentration Measurement
3.11 Mainstream CO Sensor 3.11.1. Preparing for CO Concentration Measurement Before start measuring it is necessary: Select the required ventilation adapter (adult, neonatal). The adapter windows must be dry and clean. If necessary, clean or replace the adapter. Connect the adapter with the capnograph sensor. Align the marks on the adapter and sen- sor and connect until they snap into place.
-
Page 171: Settings Of The Mainstream Co Sensor
3.11 Mainstream CO Sensor 3.11.2. Settings of the Mainstream CO Sensor The parameters of the capnometry channel are controlled using the fields and pictograms locat- ed in the capnogram output window: Fields for setting the apnea interval and the respiratory mixture temperature (input of values is similar…
-
Page 172: Monitoring Of Co Concentration
3.11 Mainstream CO Sensor When returning to the factory settings, the alarm thresholds are set in accordance with it. 1.3.9, and the values of the remaining parameters are set by default in accordance with the profile: Profile Units Adult Pediatric Neonatal Capnogram scale mmHg…
-
Page 173: Zero Calibration Of The Mainstream Co Sensor
3.11 Mainstream CO Sensor field for setting the top alarm threshold by FiCO field for setting concentration units of CO measured value of FiCO measured value of EtCO Figure 3-24. Window of gas analysis parameters Before monitoring make the following: Set the required scale and sweep speed of the capnogram curve.
-
Page 174: Disinfection Of The Mainstream Co Sensor
3.11 Mainstream CO Sensor If it planned to use the module in a closed breathing circuit, where, under certain conditions, the FiCO values may differ from zero, then, it is neces- sary to keep the module switched on for at least 5 — 10 minutes before the first calibration for a more precise zero calibration.
-
Page 175: Disinfection Of The Reusable Airway Adapters
3.11 Mainstream CO Sensor Always keep the optical windows of the sensor clean and dry. It is not allowed to disinfect the module using high temperatures and chem- ical sterilization methods associated with immersing the module in liquid. DO NOT DISINFECT THE MODULE BY IMMERSION IN A DISINFECTING SO- LUTION! Warranty service for the mainstream CO2 sensor, failed as a result of im- proper operation, is not provided.
-
Page 176: Depth Of Anaesthesia And Sedation Module
3.12 Depth of Anesthesia and Sedation Module 3.12. DEPTH OF ANAESTHESIA AND SEDATION MODULE This module is supplied with the device optionally on a separate order and enables to monitor the main parameter determined by the module — the brain activity index AI, calculated by analyz- ing patient`s electroencephalogram (EEG) in time and frequency domains, as well as other ad- ditional parameters: — brain activity index (AI), rel.units.0 — 100, step 1;…
-
Page 177: Preparing For The Monitoring Of The Depth Of Anesthesia
3.12 Depth of Anesthesia and Sedation Module Before analyzing the EEG, its primary processing is carried out, the purpose of which is to re- move motor artifacts, bursts on the EEG caused by the action of various kinds of external elec- trical interference.
-
Page 178: Settings Of The Depth Of Anesthesia Module
3.12 Depth of Anesthesia and Sedation Module Scheme of disposable electrodes location at assessment of anaesthesia depth is given in figure 3.26: yellow black green a) correct cable location b) incorrect cable location Figure 3.26 – Scheme of electrodes location when using the external depth of anaesthesia and sedation module After connection of EEG cable to a patient BE SURE that all its leads are placed on a patient and do not contact other conductive parts or earth…
-
Page 179
3.12 Depth of Anesthesia and Sedation Module The field for setting the alarm thresholds (it. 1.3.10) and the sound alarm mute (it. 1.4.2). The field for setting the alarm thresholds (it. 1.3.10) and the sound alarm mute (it. 1.4.2). To change a parameter, click on the graphic button with displaying its current value and in the appeared window click to select the new required value. -
Page 180: Evaluation Of The Brain Activity
3.12. Depth of Anesthesia and Sedation Module 3.12.4. Evaluation of the Brain Activity size of the EEG display window, sec scale of EEG display, μV EEG curve output field size of trends output window, min output field for the brain activity index trend (yellow, rel. units) and the EMG component trend (red, dB) field for the value of EEG signal quality factor (%) and switching on the window of EEG signal quality indicators output…
-
Page 181: Disinfection Of The External Depth Of Anesthesia And Sedation Module
3.12 Depth of Anesthesia and Sedation Module In the graphic window of the AI channel (p. 3.12.3) set the required scales for displaying the EEG signal and trend window. If necessary, set the brain activity index AI alarm thresholds (p. 3.12.3). It is recommended to evaluate the activity index when the impedance of the electrodes (Z1, Z2, Z3) is not more than 25 kΩ…
-
Page 182: Alarm Of The Depth Of Anesthesia Module
3.12. Depth of Anesthesia and Sedation Module Do not apply excessive tensile forces to the cables when disinfecting the cable parts of the module by pulling through the disinfecting sponge. The use of high temperatures and chemical sterilization methods associated with immersing the module in liquid is NOT ALLOWED for disinfection of the module.
-
Page 183: Calibration
4. Calibration 4. CALIBRATION The device is calibrated by the bodies of the State Metrological Service and the metrological service of legal entities accredited for the right of calibration. Calibration frequency — once a year.
-
Page 184: Maintenance
5. Maintenance 5. MAINTENANCE Before carrying out any maintenance, make sure that the device and its ac- cessories are properly disinfected. Maintenance shall be carried out at the frequency established by the agreement with the service organization, but at least once a year. The maintenance schedule (see below) does not imply disassembly of the electronic unit.
-
Page 185: Troubleshooting
6. Troubleshooting 6. TROUBLESHOOTING Trouble symptom Probable cause Troubleshooting “Power” indicator is off when No mains voltage. Check the mains voltage. device is connected to the The fuse in the mains block has Replace the fule. mains power blown. Replace the power cord. Power cord is damaged.
-
Page 186
6. Troubleshooting Trouble symptom Probable cause Troubleshooting Measured value is below 19 ºС. No temperature reading Put the sensor on the patient in accordance with it.2.3.1, 2.3.2. Malfunction (breakage or leakage) of the temperature sensor. Replace the sensor. Malfuntion of the device. Contact customer service Low or high temperature Poor thermal contact of the sensor… -
Page 187: Storage
6. Troubleshooting Trouble symptom Probable cause Troubleshooting High measurement error of 1. Changes of the oxygen sensor Contact the customer service to the О concentration. parameters during the service life. calibrate or replace the oxygen sensor. 2. Oxygen sensor failure. Cannot print trends or graphs No paper in the built-in thermal Insert paper into thermal printer…
-
Page 188: Transportation
The battery must be charged when storing the device. Storing the device with a discharged battery and not training the battery during storage leads to premature battery failure. 8. TRANSPORTATION For transportation, the device is placed in a polyethylene case in accordance with GOST 10354, and installed in a cardboard box TESM.516400.
-
Page 189: Annex 1. Replacing Paper In The Built-In Thermal Printer
Annex 1. Replacing Paper in the Built-in Thermal Printer Annex 1. Replacing Paper in the Built-in Thermal Printer Open the thermal printer cover. Insert the roll into the slots in the paper compartment of the thermal printer and thread the tape through the paper exit slot (see Figure A1-1).
-
Page 190: Annex 2. Service Menu
Annex 2. Service Menu Annex 2. Service Menu When the FREEZE mode is activated, it is not possible to enter the techno- logical menu. The service menu of the device is used when carrying out adjustment and calibration work with the monitor, as well as for forced switching on of the drying of the measuring path of the cap- nometer.
-
Page 191
Annex 2. Technological Menu Versions of other modules of the device are indicated opposite their designations: pulse oximetry module; cardiometry module; respiratory parameters measurement module; depth of anesthesia and sedation module; NIBP NIBP module; Тerm thermometry channel module; CAPNO SS sidestream capnograph module;… -
Page 192: Annex 3. Calibration Of The Touch Panel
Annex 6. Electromagnetic Environment Annex 3. Calibration of the Touch Panel Calibration of the touch panel is intended for the correct positioning of the touch buttons on the touch panel in accordance with their graphical image. This calibration may be required as a re- sult of the electrical parameters drift of the touch panel during long-term operation of the device.
12 каналов мониторинга
Мониторинг по Гарвардскому стандарту
Капнография в прямом потоке
Анализ анестезиологических газов
Глубина анестезии и другие каналы
Работает с пациентами всех возрастных групп
Экипировка монитора
Аккумулятор на 2 часа работы
Термопринтер
USB
Кабель для подключения к бортсети 12 В
Установка на стойку-тележку либо крепление с помощью кронштейна
Описание
Категории пациентов |
Пациенты всех возрастных групп |
Дисплей сенсорный, цветной | 15 « |
Аккумулятор | 2 часа работы |
Тренды | 240 часов |
Термопринтер, USB, разъём Ethernet, крепление на кронштейн либо установка на тележку | |
Каналы мониторинга |
|
»
Created with Sketch.
Монитор предназначен для эксплуатации в операционных залах
Created with Sketch.
Может работать с новорожденными, в том числе с ЭНМТ от 500 гр
Мультигаз
Анализ анестезиологических газов: изофлюран, севофлюран, десфлюран, галотан; концентрация CO2
Оценка глубины анестезии
Отражает степень угнетения сознания путём анализа EEG
- 01 /
- 02 /
- 03 /
Возможности
Мониторинг глубины анестезии
Мониторинг глубины анестезии
Монитор оценивает параметр AI — индекс активности головного мозга.
Данный вид мониторинга позволяет свести к минимуму неблагоприятные последствия неадекватной седации, обеспечить оптимальный и предсказуемый уровень седации и более быстрый выход пациента из состояния наркоза
Мониторинг анестезиологических газов
Мониторинг анестезиологических газов
Кроме концентрации газов и летучих анестетиков приор отображает расчётный параметр МАК (минимальная альвеолярная концентрация газа)
Неинвазивный мониторинг параметров центральной гемодинамики
Неинвазивный мониторинг параметров центральной гемодинамики
Проводится методом объёмной компрессионной осциллометрии (ОКО). Это косвенный неинвазивный метод определения уровней артериального давления у человека путём регистрации оригинальной измерительной системой объёмных артериальных осциллограмм
Автоматическое внеочередное измерение артериального давления (нНИАД)
Автоматическое внеочередное измерение артериального давления (нНИАД)
Канал позволяет проводить непрерывную оценку измерения АД неинвазивным способом.
В основу метода положена предсказанная теоретически и подтверждённая на практике линейная зависимость времени распространения пульсовой волны от артериального давления
Аксессуары
Документы
Назначение
Мониторы прикроватные реаниматолога и анестезиолога переносные МПР6-03-«Тритон» (в дальнейшем — приборы) предназначены для непрерывного измерений и мониторинга основных параметров жизнедеятельности пациентов:
— электрокардиография (ЭКГ), частота сердечных сокращений (ЧСС),
— насыщение (сатурация) кислородом гемоглобина артериальной крови (8рО2), частота пульса (PR), периферическое кровообращение (фотоплетизмограмма),
— неинвазивное артериальное давление (систолическое, диастолическое, среднее),
— инвазивное давление крови (систолическое, диастолическое, среднее),
— температура тела (Т°С),
— частота дыхания (ЧД), респирограмма,
— концентрация (или парциальное давление) газов во вдыхаемой и выдыхаемой газовой смеси (Fi, Et), кривая дыхания.
Область применения: анестезиология, хирургия, интенсивная терапия и время послеоперационного периода, а также проведение скрининг-тестов при профессиональных осмотрах.
Описание
В зависимости от размеров примененного дисплея приборы выпускаются в следующих исполнениях:
— исполнение 1 (с дисплеем 7″),
— исполнение 2 (с дисплеем 10″),
— исполнение 3 (с дисплеем 12″),
— исполнение 4 (с дисплеем 15″).
в составе следующих измерительных модулей:
— модуль пульсоксиметрии,
— модуль кардиометрии,
— модуль термометрии,
— модуль неинвазивного измерения артериального давления,
— модуль измерения параметров дыхания,
— модуль газоанализа дыхательной смеси,
— модуль инвазивного измерения давления.
Принцип действия модуля пульсоксиметрии основан на различном спектральном поглощении оксигемоглобина и восстановленного гемоглобина крови. Пульсирующая кровь в ткани (пальца или мочки уха) просвечивается источниками излучения в области красного и инфракрасного спектра. Полученные сигналы после соответствующей обработки преобразуются в фотопле-тизмограмму, выводимую на дисплей монитора, и позволяют определить коэффициенты модуляции световых потоков с различными длинами волн и по их соотношению определить насыщение кислородом гемоглобина крови, при этом периодичность модуляции соответствует частоте пульса.
Принцип действия модуля кардиометрии основан на измерении электрического потенциала сердца с помощью электродов, расположенных на поверхности тела пациента. После соответствующей обработки электрические сигналы преобразуются в ЭКГ, которая выводится на дисплей монитора, и используются для расчета ЧСС.
Лист № 2 Всего листов 11
Принцип действия модуля измерения параметров дыхания основан на импедансном методе (изменение сопротивления тела пациента между электродами при вдохе-выдохе). Сигналы с электродов после соответствующей обработки преобразуются в кривую дыхания (респирограмму), выводимую на дисплей монитора, и используются для расчета частоты дыхания.
Принцип действия модуля термометрии основан на измерении сопротивления датчика температуры, который представляет собой терморезистор. Сопротивление терморезистора, зависящее от температуры тела пациента в точке нахождения датчика, измеряется с помощью электрической схемы и преобразуется в значение температуры, отображаемое на дисплее монитора.
Принцип действия модуля неинвазивного измерения артериального давления основан на осциллометрическом методе, при котором пульсации давления в манжете с помощью тензометри-ческого датчика давления преобразуются в сигнал, который после соответствующей обработки используется для расчета величины давления и пульса.
Принцип действия модуля инвазивного измерения давления основан на преобразовании пульсаций давления крови в изменение сопротивления с помощью датчика инвазивного давления. Полученный монитором сигнал используется для расчета значений давления и вывода на дисплей монитора кривой давления.
Принцип действия модуля газоанализа дыхательной смеси в части капнографии основан на методе двухканальной бездисперсионной оптической спектроскопии, при котором луч инфракрасного излучателя проходит через исследуемую пробу газа в измерительной камере и попадает на два датчика-фотоприемника с узкополосными оптическими фильтрами. Фильтр одного канала пропускает только излучение в области максимального оптического поглощения для СО2, а другого — в области, которая не поглощается в СО2. Преобразованные прибором сигналы используются для расчета уровня концентрации СО2 и вывода на дисплей прибора кривой дыхания (капнограм-мы). Принцип действия модуля газоанализа дыхательной смеси в части оксиметрии основан на электрогальваническом эффекте, при котором проходящая через датчик кислорода проба воздуха, взаимодействуя с пластинами датчика, возбуждает на них электрический потенциал, пропорциональный содержанию кислорода.
. UMHF№WA1*M4 t-vA.-i :-ыл ■ и’на шктпкгитпл»
_ ЛЬщ U_
-ЧЛ* 74
‘1 ■ 00’
<« *»1 *
U In
’75
„ Тйз/ (МИ
1 К К IV Л Л i Л , V Лч
УIJ -^J xj О -J О ‘-_i j-
жо^^сШЗбГЗ^ —— —.«ы . (Д/
А Я Я.Ш Ш о
I rt£UI-iivi ш мм BSi dfl Lu •■>,-. 1 ШшЯ ы J ‘
450 0 250 0 0 55 2ЙвО „ 3g. 5
Т U7 2436
■и да i&JHilj
: -141/ 94
Г 9
Т- и
= Г] □! Е2
— W
Рисунок 4 — Общий вид прибора в исполнении 4 (с дисплеем 15″)
Программное обеспечение
Встроенное программное обеспечение (ПО) изделия Мониторы прикроватные реаниматолога и анестезиолога переносные МПР6-03-«Тритон» состоит из модуля ПО, отвечающего за метрологические характеристики для всех видов исполнений.
Программное обеспечение от преднамеренных и непреднамеренных изменений защищено тем, что: а) ПО прошито в микропроцессоре, жестко установленном внутри печатной платы; б) отсутствуют внешние интерфейсы обновления программного обеспечения.
Наименование программного обеспечения |
Идентифика ционное наименование программного обеспечения |
Номер версии (идентификационный номер) программного обеспечения |
Цифровой идентифи катор программ ного обеспечения |
Алгоритм вычисления цифрового идентификатора программного обеспечения |
Модуль ПО, отвечающий за метрологические характеристики |
Metr ID |
не ниже 0214 |
ПО по жесткости испытаний — низкая. Погрешности, вносимые программным обеспечением, отдельно не оцениваются и входят в погрешности измерений, указанных ниже.
№ пп |
Наименование характеристики |
Значение характеристики |
Модуль пульсоксиметрии |
||
1 |
Диапазон измерений отношения индексов модуляции (А) двух синфазномодулированных сигналов, выраженного в единицах сатурации (SpO2), где зависимость между SpO2 и А определяется градуировочной кривой, % Пределы допускаемой абсолютной погрешности измерений, — в диапазоне 70…100, % — в диапазоне 50…69, % — в диапазоне 10…49, % |
10.100 ±2 ±3 ±5 |
2 |
Диапазон измерений частоты модуляции двух синфазномодулиро-ванных сигналов (PR), 1/мин Пределы допускаемой абсолютной погрешности измерений PR, 1/мин |
15.350 ±1 |
Модуль кардиометрии |
||
3 |
Диапазон входных напряжений, размах, мВ |
0,05.10 |
4 |
Входной импеданс, МОм, не менее |
10 |
5 |
Коэффициент ослабления синфазного сигнала, дБ, не менее |
80 |
6 |
Напряжение внутренних шумов, приведенных к входу, мкВ, не более |
30 |
7 |
Диапазон измерений частоты сердечных сокращений (ЧСС), 1/мин Пределы допускаемой абсолютной погрешности измерений ЧСС, 1/мин |
0.350 ±1 |
Модуль термометрии |
||
8 |
Диапазон измерений температуры, °C Пределы допускаемой абсолютной погрешности измерений, °C |
0.50 ±0,1 |
Модуль неинвазивного измерения артериального давления |
||
9 |
Диапазон измерения давлений в манжете, мм рт.ст. Пределы допускаемой абсолютной погрешности измерений, мм рт.ст. |
0.300 ±3 |
10 |
Максимальное давление в манжете: — для взрослых, мм рт.ст., не более — для новорожденных, мм рт.ст., не более |
300 150 |
Модуль инвазивного измерения движения |
||
11 |
Диапазон измерений давления инвазивным способом, мм рт. ст. Пределы допускаемой погрешности измерений: — относительной в диапазоне 101…300 мм рт. ст., % — абсолютной в диапазоне -50.100 мм рт. ст., мм рт. ст. |
-50.300 ±2 ±2 |
12 |
Диапазон компенсации смещения нуля при измерении давления инвазивным способом, мм рт. ст. |
±100 |
Модуль измерения параметров дыхания |
||
13 |
Диапазон измерений частоты дыхания, 1/мин Пределы допускаемой абсолютной погрешности измерений, 1/мин |
0.160 ±3 |
Модуль газоанализа дыхательной смеси |
||
14 |
Диапазон измерений: — уровня концентрации СО2, % (парциального давления СО2), мм рт. ст. — уровня концентрации О2, % |
0.15 (0.115) 0.100 |
15 |
Пределы допускаемой погрешности измерений: уровня концентрации СО2: — абсолютной в диапазоне 0-5%, % — относительной в диапазоне 5,1-10%, % — относительной в диапазоне 10,1-15%, % парциального давления СО2: — абсолютной в диапазоне 0-40 мм рт.ст., мм рт.ст. — относительной в диапазоне 41-80 мм рт.ст., % — относительной в диапазоне 81-115 мм рт.ст.), % Пределы допускаемой абсолютной погрешности измерений уровня концентрации О2, % |
±0,2 ±4 ±6 ±1,5 ±4 ±6 ±2 |
Общие технические характеристики |
||
16 |
Прибор работает — от сети переменного тока напряжением, В частотой, Гц — от встроенного аккумулятора напряжением, В — от внешнего источника постоянного тока, В |
220±22 50 12±1,2 12±1,2 |
17 |
Обеспечивается непрерывная работа в течение, ч, не менее |
24 |
18 |
Время работы от встроенного внутрь и полностью заряженного аккумулятора, мин, не менее от аккумулятора увеличенной емкости, мин, не менее |
30 120 |
19 |
Максимальная потребляемая мощность, В А, не более |
50 |
20 |
Габаритные размеры корпуса прибора (без влагоотделителя), Ш х В х Г: — с дисплеем 7″ (исполнение 1) , мм, не более — с дисплеем 10″ (исполнение 2) , мм, не более — с дисплеем 12″ (исполнение 3) , мм, не более — с дисплеем 15″ (исполнение 4) , мм, не более |
250х220х200 340х230х190 400х290х185 370х360х200 |
21 |
Масса прибора: — с дисплеем 7″ (исполнение 1) , кг, не более — с дисплеем 10″ (исполнение 2) , кг, не более — с дисплеем 12″ (исполнение 3) , кг, не более — с дисплеем 15″ (исполнение 4) , кг, не более |
3.5 6,0 6.5 7.5 |
22 |
Средняя наработка на отказ, ч, не менее |
1000 |
23 |
Средний срок службы, лет |
10 |
24 |
По безопасности прибор соответствует ГОСТ Р 50267.0, ГОСТ Р 50267.27, ГОСТ Р 50267.30, ГОСТ Р 50267.34, ГОСТ Р 50267.49, ГОСТ Р ИСО 9918, ГОСТ Р ИСО 9919, ГОСТ Р 52566 для изделий: — класса I при питании от сети переменного тока; — с внутренним источником питания при питании от встроенного аккумулятора; — класса II при питании от внешнего источника постоянного тока; — с рабочими частями типа CF и защитой от разряда дефибриллятора для рабочих частей модулей: кардиометрии (ЭКГ), неинвазивного измерения артериального давления (NIBP), инвазивного измерения давления (IBP). Режим работы продолжительный. Степень защиты от проникновения влаги IPX0. |
25 |
По электромагнитной совместимости (ЭМС) прибор соответствует требованиям ГОСТ Р 50267.0.2. |
Знак утверждения типа
Знак утверждения типа полиграфическим способом наносится на самоклеящуюся плёнку, плёнка клеится на задней панели прибора. На титульном листе паспорта знак утверждения типа наносится методом печати.
Комплектность
Наименование |
Обозначение |
Колич., шт. |
1 Монитор прикроватный реаниматолога и анестезиолога пе |
||
реносный МПР6-03-«Тритон» |
РМ 501.01.000-01-01 |
|
— электронный блок (исполнение 1)* |
ТЭСМ. 556002 |
— |
— электронный блок (исполнение 2)* |
ТЭСМ.546001 |
— |
— электронный блок (исполнение 3)* |
ТЭСМ.506000 |
— |
— электронный блок (исполнение 4)* |
ТЭСМ. 506000 |
— |
кабель питания |
покупной |
1 |
встроенный аккумулятор * |
покупной |
— |
модуль пульсоксиметрии * |
ТЭСМ.035033 |
— |
модуль кардиометрии * |
ТЭСМ.035029 |
— |
модуль термометрии * |
ТЭСМ. 035034 |
— |
модуль неинвазивного измерения артериального давления * |
ТЭСМ. 555003 |
— |
модуль инвазивного измерения давления* |
ТЭСМ. 53 5015 |
— |
модуль измерения параметров дыхания * |
ТЭСМ. 355003 |
— |
модуль измерения параметров дыхания * |
ТЭСМ. 186004 |
|
модуль газоанализа дыхательной смеси * |
ТЭСМ. 535031 |
— |
модуль газоанализа дыхательной смеси * |
ТЭСМ. 535029 |
— |
модуль газоанализа дыхательной смеси * |
AION 02-31 Multigas Analyzer, AION 03-31 Multigas Analyzer |
|
— модуль принтера* |
ТЭСМ. 396100 |
— |
Периферия: |
||
датчик оптоэлектронный пульсоксиметрических пальцевый |
— |
|
прищепочный ДОПпп -Тритон * |
РМ 501.00.124 |
|
датчик оптоэлектронный пульсоксиметрический пальцевый |
— |
|
прищепочный ДОПпп -Тритон * |
PM 501.00.124-01 |
■ датчик оптоэлектронный пульсоксиметрический неонатальный ДОПн — Тритон * |
РМ 501.00.004 |
— |
— датчик оптоэлектронный пульсоксиметрический неонатальный ДОПн — Тритон * |
РМ 501.00.004-01 |
— |
датчик пульсоксиметрический * |
ТЭСМ. 096008 |
— |
датчик пульсоксиметрический * |
ТЭСМ. 096008-01 |
— |
— датчик пульсоксиметрический педиатрический, одноразовый * |
ТЭСМ. 096009 |
— |
датчик пульсоксиметрический неонатальный, одноразовый * |
ТЭСМ. 096010 |
— |
датчик пульсоксиметрический педиатрический, клипса * |
ТЭСМ. 096011 |
— |
датчик пульсоксиметрический педиатрический, резиновый * |
ТЭСМ. 096012 |
— |
датчик пульсоксиметрический неонатальный, резиновый |
ТЭСМ. 096013 |
— |
датчик пульсоксиметрический неонатальный, одноразовый * |
ТЭСМ. 096014 |
— |
— манжета установочная датчика оптоэлектронного пульсок-симетрического неонатального ДОПн-«Тритон»* |
ТЭСМ. 090002 |
— |
кабель пациента * |
РМ 501.00.120 |
— |
кабель пациента * |
РМ 501.00.120-01 |
— |
кабель пациента * |
РМ 501.00.120-02 |
— |
датчик температуры поверхностный * |
РМ 501.25.000 |
— |
датчик температуры внутриполостной * |
РМ 321.21.000 |
— |
манжета LD-Cuff, производства фирмы «LITTLE DOCTOR INTERNATIONAL (S) PTE. LTD» * |
LD-Cuff «LITTLE DOCTOR INTERNATIONAL», Сингапур |
|
— манжета пневматическая однократного применения для измерения артериального давления у новорожденных * |
ЗАО «Медтехника», Санкт-Петербург |
— |
манжета пневматическая одноразовая для новорожденных* |
SoftCheck «Statcorp Medical», США |
|
— ЭКГ-электроды одноразовые для диагностики * |
«Unomedical Ltd.», Дания |
— |
— устройства и дыхательные системы для наркознодыхательных аппаратов, аэрозольной и кислородной терапии: соединители* |
«ИНТЕРСЕРДЖИКАЛ», Великобритания, Литва |
|
— удлинитель инфузионных насосов, производства фирмы «BALTON Ltd.» |
«BALTON Ltd.», Польша |
— |
— комплект Filter Line * — наборы для контроля давления с преобразователем КОМБИТРАНС* |
«Oridion Medical», Израиль «В.BRAUN Melsungen AG», Г ермания |
— |
— преобразователи КОМБИТРАНС в комплекте с краном ЭКСАДИН* |
«В.BRAUN Melsungen AG», Г ермания |
— |
— пластина для крепления преобразователей КОМБИТРАНС* |
«В.BRAUN Melsungen AG», Г ермания |
— |
— держатель Комби-холдер для преобразователей КОМБИТРАНС* |
«В.BRAUN Melsungen AG», Г ермания |
— |
— многоразовый преобразователь давления СенсоНор 844* |
«В.BRAUN Melsungen AG», Г ермания |
— |
набор для контроля давления КОМБИДИН* |
«B.BRAUN Melsungen AG», Германия |
— |
купол СЕНСОНОР* |
«B.BRAUN Melsungen AG», Г ермания |
— |
кабель питания для подключения к внешнему источнику |
ТЭСМ. 324001 |
— |
постоянного тока (только для приборов с универсальным пи |
ТЭСМ.534028 |
— |
танием) * |
ТЭСМ. 534028-01 |
— |
— переходник датчика пульсоксиметрического * |
ТЭСМ.534014 |
— |
— переходник датчика пульсоксиметрического * |
ТЭСМ.534017 |
— |
— шланг для манжеты * |
ТЭСМ.536128 |
— |
— установочное оборудование * |
ТЭСМ. 533002 |
— |
— установочное оборудование * |
ТЭСМ. 533002-01 |
— |
— установочное оборудование * |
ТЭСМ. 533002-02 |
— |
— установочное оборудование * |
ТЭСМ. 533002-03 |
— |
— модуль импедансной кардиографии* |
ТУ 9441-001-214868342002, Microlux 2Z-CORE™, Россия |
|
— модуль неинвазивного измерения сердечного выброса* |
«ООО ГЛОБУС», Россия |
— |
— модуль оценки глубины анестезии* |
ТУ 9441-018-321193982011, ООО фирма «ТРИТОН- ЭЛЕКТРОНИКС», Россия |
|
— модуль мониторинга параметров респираторной механики* |
ТЭСМ. 520000 |
— |
— модуль газоанализа дыхательной смеси (без отбора пробы) * |
ТЭСМ. 506001 |
— |
2 Запасные части: |
||
предохранитель ВПТ6-7-1А |
АГО 0481.304ТУ |
— |
влагоотделитель * |
ТЭСМ. 039717 |
— |
3 Эксплуатационные документы: |
||
руководство по эксплуатации МПР6-03 * |
РМ 501.01.000-01-01 РЭ |
— |
РМ 501.01.000-01-01 |
— |
|
паспорт МПР6-03 * |
ПС |
* — определяется при заказе
Поверка
осуществляется по методике поверки, изложенной в разделе 4 документа РМ 501.01.000-01-01 РЭ «Монитор прикроватный реаниматолога и анестезиолога переносный МПР6-03-«Тритон». Руководство по эксплуатации», согласованного ГЦИ СИ ООО «Испытательный центр АФК» (с 2010 г. ЗАО «НИИМТ»).
Перечень основных средств поверки (эталонов), применяемых для поверки:
— установка УПОП-Е01М ТУ6680-008-32119398-2000 (Госреестр № 20798-11),
— тест-калибратор ТК-01 ТУ-9441-007-32119398-99 (Госреестр № 19410-10),
— термометр цифровой ТЦ-1200 ТУ 4211-120-56835627-09 (Госреестр № 45039-10)
— измеритель давления цифровой ИДЦ-1 М (Госреестр № 46121-10).
Смесь №1 (ГСО 3795-87) СО2 , остальное — воздух
— концентрация СО2 , %
5, 10, 15, ±0,1
5, 50, 100 ±0,2%
— отклонение, %
Смесь №2 (ГСО 3726-87, 3732-87, 7591-99) О2 , остальное — азот
— концентрация О2,%
— отклонение, %
Сведения о методах измерений
РМ 501.01.000-01-01 РЭ «Монитор прикроватный реаниматолога и анестезиолога переносный МПР6-03-«Тритон». Руководство по эксплуатации»
Нормативные и технические документы, устанавливающие требования к мониторам прикроватный реаниматолога и анестезиолога переносным МПР6-03-«Т ритон»
1) ГОСТ Р 50444-92 «Приборы, аппараты и оборудование медицинские. Общие технические условия».
2) ГОСТ Р 50267.0-92 «Изделия медицинские электрические. Часть 1. Общие требования безопасности».
3) ГОСТ Р 50267.0.2-2005 «Изделия медицинские электрические. Часть 1-2. Общие требования безопасности. Электромагнитная совместимость. Требования и методы испытаний».
4) ГОСТ Р 50267.27-95 «Изделия медицинские электрические. Часть 2. Частные требования к электрокардиологическим мониторам».
5) ГОСТ Р 50267.30-99 «Изделия медицинские электрические. Часть 2. Частные требования безопасности к приборам для автоматического контроля давления крови косвенным методом».
6) ГОСТ Р 50267.34-95 «Изделия медицинские электрические. Часть 2. Частные требования безопасности к приборам для прямого мониторинга кровяного давления».
7) ГОСТ Р 50267.49-2004 «Изделия медицинские электрические. Часть 2. Частные требования безопасности к многофункциональным мониторам пациента».
ГОСТ Р 52566-2006 «Мониторы для контроля содержания кислорода в дыхательной смеси для пациента. Требования безопасности».
9) ГОСТ Р ИСО 9918-99 «Капнометры медицинские. Частные требования безопасности». 10) ГОСТ ISO 9919-2011 «Изделия медицинские электрические. Частные требования безопасности и основные характеристики пульсовых оксиметров».
Рекомендации к применению
— для применения все сферы государственного регулирования обеспечения единства измерений.
Предложите, как улучшить StudyLib
(Для жалоб на нарушения авторских прав, используйте
другую форму
)
Ваш е-мэйл
Заполните, если хотите получить ответ
Оцените наш проект
1
2
3
4
5
Общие характеристики:
Категории пациентов | Взрослые, дети, новорожденные |
Разъемы | Для подключения к компьютерной сети |
Питание | 220 В Аккумулятор не менее 2 часов работы Вход для подключения к бортсети 12 В |
Дисплей | 7″ цветной, сенсорный, жидкокристаллический Кнопки быстрого доступа к параметрам Энкодер |
Одновременное отображение кривых на дисплее | 5 |
Система тревог | 3 уровня приоритетности Настройка границ тревог |
Тренды | Графические и цифровые, до 240 часов |
Масса прибора, кг | Не более 2,5 |
Каналы мониторинга:
Электрокардиография (ЭКГ) | ЭКГ 6 отведений: I, II, III, aVL, aVR, aVF |
Анализ вариабельности сердечного ритма (ВСР) | |
Анализ ST-сегмента | |
Частота сердечных сокращений (ЧСС) | |
Анализ и автоматическое распознавание аритмии | |
Пульсоксиметрия TRITON® | Частота пульса (ЧП) |
Насыщение артериальной крови кислородом (SpO2) | |
Фотоплетизмограмма (ФПГ) | |
Неинвазивное измерение артериального давления (НИАД) | SysAD, DiaAD, MedAD |
Автоматическое внеочередное измерение артериального давления (нНИАД) | SysAD, DiaAD, MedAD, время задержки ФПГ относительно ORS комплекса ЭКГ |
Импедансный метод измерения параметров дыхания | Частота дыхания (ЧД) |
Респирограмма (RESP) | |
Термометрия | 2 канала |
Измерение разности температур |
Вся продукция разрешена к применению на территории Российской Федерации.
Качество и соответствие продукции подтверждено следующими документами, представленными на русском языке:
- Регистрационное удостоверение на продукцию;
- Сертификат соответствия на продукцию;
- Свидетельство об утверждении типа средств измерений (если данное требование предусмотрено действующим законодательством);
- Инструкция по эксплуатации;
- Технический паспорт с гарантийными обязательствами поставщика.
Комплект поставки включает все необходимые принадлежности для обеспечения работоспособности оборудования.
Прикроватный аппарат Тритон МПР6-03
Прикроватный аппарат МПР6-03 – портативный высокотехнологичный аппарат, который предназначен для эффективного отслеживания основных жизненных показателей пациентов. Монитор реаниматолога Тритон МПР6-03 весом 6.4 кг имеет 15-дюйсовый цветной TFT-экран, возможность отображения до 6-ти кривых. В дисплее есть встроенные кнопки для оперативного доступа к нужным параметрам, а также уникальная 3-х уровневая система оповещения.
Особенности переносного монитора пациента Тритон МПР6-03
Монитор пациента Тритон МПР6-03 подойдет как для небольших частных клиник, так и для крупных бюджетных медицинских учреждений. Аппарат дает врачу увидеть полную картину состояния пациента и принять соответствующие меры. Использовать монитор можно в палате и в автомобиле скорой медицинской помощи.
Основные преимущества монитора пациента Тритон МПР6-03:
- удобный USB-порт для передачи информации;
- подходит для всех категорий людей: дети (в том числе новорожденные), взрослые;
- специальные разъемы для сетевых подключений;
- наличие термопринтера;
- ручка для эффективной транспортировки;
- встроенная аккумуляторная батарея.
Варианты дополнительного дооснащения монитора пациента Тритон МПР6-03
Монитор пациента Тритон МПР6-03 подойдет как для небольших частных клиник, так и для крупных бюджетных медицинских учреждений. Аппарат дает врачу увидеть полную картину состояния пациента и принять соответствующие меры. Использовать монитор можно в палате и в автомобиле скорой медицинской помощи.
Фильтр
Антибактериальный фильтр соединен по типу «Луер» – диаметр 28 мм
Одноразовые наборы
Одноразовые наборы для измерения артериального и венозного давления
Кронштейн
Специальный кронштейн для крепления монитора пациента к стене
Тележка
Специальная стойка-тележка для повышения мобильности монитора пациента
Принтер
Термопринтер для печати выбранных параметров
Кран для ИИНД
Многоходовые соединения нескольких инфузионных магистралей, систем
Дополнительные модули для монитора пациента Тритон МПР6-03
Стандартный комплект поставки включает в себя все основные датчики и модули для решения классических задач. Если вашему медицинскому учреждению этого недостаточно, то прикроватный монитор пациента Тритон МПР6-03 можно оснастить дополнительными модулями и опциями, которые решат текущие клинические проблемы
Модули, которые можно установить дополнительно на монитор пациента Тритон МПР6-03:
- Модуль для инвазивного измерения артериального давления
- Модуль измерения насыщения крови кислородом (Nellcor SpO2)
- Модуль капнографии (СО2)
- Модуль мультигаза
- Модуль измерения сердечного выброса
- Модуль мониторинга глубины седации
Цена на монитор пациента Тритон МПР6-03
Наши менеджеры имеют медицинское образование и смогут провести профессиональную консультацию по любым мониторам пациента и дополнительным принадлежностям к ним. CordisMed работает напрямую с производителем и имеет максимальные оптовые скидки.
Свяжитесь с нашими менеджерами по бесплатному телефону 8 (800) 555-30-19 или оставьте заявку
Технические характеристики транспортного монитора пациента Тритон МПР6-03
Общие характеристики Тритон МПР6-03
Категории пациентов |
Взрослые, дети, новорожденные |
Разъемы |
|
Питание |
|
Дисплей |
|
Система тревог |
3 уровня приоритетности, настройка границ тревог |
Тренды |
Графические и цифровые, до 240 часов |
Каналы мониторинга на мониторе пациента Тритон МПР6-03
Электрокардиография (ЭКГ) |
|
Пульсоксиметрия Masimo SET® |
|
Термометрия |
|
Импедансный метод измерения параметров дыхания |
|
Неинвазивное измерение артериального давления (НИАД) |
Сист / Сред / Диаст АД |
На какие параметры нужно обращать внимание при выборе монитора пациента
Автономность
Встроенный аккумулятор нужен для того, чтобы во время отключение электроэнергии монитор пациента продолжал свою работу в штатном режиме. У разных производителей батареи рассчитаны на разное время автономной работы от 1 часа до 20 часов.
Вид монитора пациента
Монитор пациента делятся на две большие категории. Анестезиологические – нужны для контроля параметров жизнедеятельности пациентов во время хирургической операции и глубины анестезии. Прикроватные – нужны для длительного наблюдения пациента в различных отделениях медицинского учреждения.
Базовые параметры
Основные параметры, которые должен отслеживать монитор пациента – сатурация, неинвазивное артериальное давление, частота дыхания, температура тела, частота сердечных сокращений, электрокардиография.
Сенсорный экран
По отзывам специалистов, сенсорные экраны в разы удобнее кнопок. Врачу интуитивно понятно как включить ту или иную функцию и как настроить главный экран с показателями состояния пациента.
Мобильный или стационарный
Прикроватные мониторы нужны в первую очередь для отслеживания состояния пациента в отделении, а транспортные на выезде. Как правило, больше опций в стационарном прикроватном мониторе пациента.
Подключение к ЦСМ
Для удобного отслеживания состояния пациента у монитора пациента должны быть возможность подключения к центральной станции мониторирования, где на одном компьютере объединяются все мониторы пациента.
Аналоги монитора пациента Тритон МПР6-03
Модель |
Класс |
CO2 |
SpO2 |
НИАД |
Температура |
ЧСС |
ЭКГ, отведения |
Доп. параметры |
Диагональ дисплея, дюйм |
Аккумулятор |
Цена |
Bionet BM 1 |
Эконом |
Да |
Да |
Да |
Да |
Да |
Да (3/5) |
— |
4,3 |
Да (до 5 часов) |
от 100000 рублей |
Bionet BM 3 |
Эконом |
Нет |
Да |
Да |
Да |
Да (3) |
ST |
7 |
Да (до 5 часов) |
по запросу |
|
Bionet BM 3 Plus |
Эконом |
Нет |
Да |
Да |
Да |
Да |
Да (3) |
ST |
8,4 |
Да (до 2 часов) |
от 170000 рублей |
Bionet BM 5 |
Эконом |
Да |
Да |
Да |
Да |
Да |
Да (6) |
ИАД, ST |
10,4 |
Да (до 4 часов) |
от 400000 рублей |
Comen Star8000A |
Эконом |
Да |
Да |
Да |
Да |
Да |
Да (5) |
ИАД, EtCO2, FiCO2 |
12,1 |
Да (до 3 часов) |
от 115000 рублей |
Comen Star8000B |
Эконом |
Да |
Да |
Да |
Да |
Да |
Да (3/5) |
ИАД, ST, Nellcor |
8,4 |
Да (до 4 часов) |
от 130000 рублей |
Comen Star8000B-Neo |
Эконом |
Нет |
Да |
Да |
Да |
Да |
Да (3) |
ИАД, Nellcor |
8,4 |
Да (до 4 часов) |
от 250000 рублей |
Votem VP 700 |
Эконом |
Да |
Да |
Да |
Да |
Да |
Да (3) |
ИАД, ST, APG |
7 |
Да |
по запросу |
Монитор пациента G3C Armed |
Эконом |
Да |
Да |
Да |
Да |
Да |
Да |
— |
10,4 |
Да (до 1 часа) |
от 102000 рублей |
Монитор пациента G3H Armed |
Эконом |
Нет |
Да |
Да |
Да |
Да |
Да |
— |
8,4 |
Да (до 1 часа) |
от 103000 рублей |
Armed PC-9000b |
Эконом |
Нет |
Да |
Да |
Да |
Да |
Да (5) |
— |
12,2 |
Да (до 2 часов) |
по запросу |
Armed PC-9000f |
Эконом |
Нет |
Да |
Да |
Да |
Да |
Да (5) |
— |
12,2 |
Да (до 2 часов) |
по запросу |
Armed PC-900s |
Эконом |
Нет |
Да |
Нет |
Нет |
Нет |
Нет |
— |
8,6 |
Да (до 2 часов) |
по запросу |
Реанимационный монитор МИТАР-01-Р-Д |
Эконом |
Да |
Да |
Да |
Да |
Да |
Да (12) |
ИАД |
12,1 |
Да (до 5 часов) |
от 330000 рублей |
Тритон МПР6-03 |
Эконом |
Да |
Да |
Да |
Да |
Да |
Да (12) |
ST, RESP |
15 |
Да (до 2 часов) |
от 320000 рублей |
Характеристика закупаемого товара |
Требования, установленные Заказчиком |
Общие характеристики |
|
Назначение: монитор прикроватный реаниматолога и анестезиолога переносный для длительного и непрерывного наблюдения за жизненно важными функциями пациента. |
Наличие |
Категории пациентов: взрослый, детский, неонатальный |
Наличие |
Работа от сети переменного тока: напряжением 220+22 В частотой 50 Гц |
Наличие |
Масса прибора (со встроенным аккумулятором и блоком питания), кг. |
не более 6,5 |
Работа от встроенного внутрь и полностью заряженного аккумулятора |
не менее 2 часов |
Заряд аккумулятора автоматический |
Наличие |
Дисплей: цветной TFT с сенсорным экраном диагональ не менее 12” |
Наличие |
Транспортировка: ручка на верхней части корпуса |
Наличие |
Технические характеристики |
|
Интерфейс пользователя: Цветной полностью cенсорный TFT дисплей, наличие кнопок быстрого доступа к параметрам, наличие регулятора типа «энкодер» для быстрого изменения параметров |
Наличие |
Одновременное отображение на дисплее кривых |
не менее 6 |
Режим «большие цифры» |
наличие |
Режим «замораживания кривых» |
наличие |
Индикация частоты пульса: звуковая |
наличие |
Система поясняющих текстовых сообщений |
наличие |
Совместимость работы со станцией централизованного мониторинга, поддерживающую одновременное подключение к локальной сети любого количества мониторов, шт |
не менее 32 |
Разъем infoport для подключения внешних модулей |
наличие |
Регистрируемые параметры |
|
сердечная деятельность (ЭКГ/ЧСС) |
наличие |
пульсоксиметрия периферического кровообращения (ФПГ) |
наличие |
оксигенации артериальной крови (SpO2) |
наличие |
частота пульса (ЧП) |
наличие |
неинвазивное измерение артериального давления (НИАД), три параметра – SysAD, DiaAD, MedAD |
наличие |
непрерывное неинвазивное измерение артериального давления (ННИАД), параметры — SysAD, DiaAD, MedAD, PWTT |
наличие |
инвазивное измерение артериального давления (ИАД), не менее 2 каналов |
наличие |
неинвазивный мониторинг центральной гемодинамики: — сердечный выброс (СВ/СО) — ударный объема (УО/SV) — системное сосудистое сопротивление (CCC/SVR) |
Наличие |
температура (Т) – кожная и центральная |
не менее 2 каналов |
дыхание по импедансу (ЧД/Респирограмма) |
наличие |
анализ ST-сегмента |
наличие |
концентрация СО2 на вдохе (FiCO2) и выдохе (EtCO2), капнограмма |
наличие |
концентрация кислорода во вдыхаемом/выдыхаемом воздухе (FiO2, EtO2) |
наличие |
Канал ЭКГ |
Наличие |
Cинтезированная 12 канальная ЭКГ, снимаемая с 5 электродов |
Наличие |
Регистрируемые отведения: I, II, III, aVL, aVR, aVF, V1-V6 |
Наличие |
Одновременное отображение на дисплее любых кривых из 12-ти регистрируемых отведений ЭКГ |
не менее 3 |
Выбор комбинации мониторируемых отведений ЭКГ: пользователем самостоятельно |
Наличие |
Ручное и автоматическое масштабирование отображения кривой ЭКГ |
наличие |
Графическое и цифровое отображение величины смещения ST-сегмента относительно изолинии по всем отведениям в диапазоне, мм |
не менее + 10 |
Скорость отображения кардиокомплекса в масштабе 12.5, 25, 50 мм/сек |
наличие |
Защита от помех электрохирургического инструмента и дефибриллятора |
наличие |
Постоянное отображение кардиоинтервалограммы (ритмокардиограммы) |
наличие |
Отображение гистограммы распределения кардиоинтервалов |
наличие |
Функция анализа и автоматического распознавания аритмии |
не менее 18 |
Отображение цифровых параметров: ЧСС, смещение ST-сегмента, порогов тревог по ЧСС |
наличие |
Диапазон измерения частоты сердечных сокращений (ЧСС), 1/мин нижняя граница: верхняя граница: |
не более 0, не менее 350 |
Канал пульсоксиметрии |
Наличие |
Цифровое Отображение частоты пульса, перфузии, SpO2 |
Наличие |
Отображение фотоплетизмограммы: постоянное |
|
Автоматическое Масштабирование фотоплетизмограммы |
Наличие |
Диапазон измерения SpO2, % нижняя граница верхняя граница |
не более 10, не менее 100 |
Абсолютная погрешность при измерении SpО2 в диапазоне 10-49% в диапазоне 50-69% в диапазоне70-100% |
не более ±5 % не более ±3 % не более ±2 % |
Диапазон измерения частоты пульса (PR), 1/мин нижняя граница верхняя граница |
не более 15, не менее 350 |
Погрешность измерения частоты пульса |
не более ± 1 уд/мин. |
Канал термометрии |
Наличие |
Количество каналов |
не менее 2 |
Метод измерения: терморезисторный |
наличие |
Цифровое отображение температуры, в том числе разность по 2 каналам |
Наличие |
Диапазон измерения, °С нижняя граница верхняя граница |
не более 0, не менее 50 |
Погрешность измерения |
не более + 0,1°С |
Типы датчиков: поверхностный, внутриполостной |
наличие |
Канал капнографии |
Наличие |
Метод капнометрии, не требующий периодических калибровок эталонными газами |
наличие |
Отбор пробы в боковом потоке |
наличие |
Система осушения пробы, не требующая использования магистралей для капнографии с влагопоглотителем |
наличие |
Автоматическая коррекция результата при наличии N2O |
Наличие |
Диапазон измерения: уровня концентрации СО2, % нижняя граница верхняя граница парциального давления СО2, mm.Hg нижняя граница верхняя граница |
не более 0, не менее 15 не более 0, не менее 115 |
Режим отображения: уровня концентрации в % парциального давления в мм.Hg |
Наличие |
Погрешность измерения уровня концентрации СО2: — абсолютная в диапазоне 0-5% — относительная в диапазоне 5,1-10% — относительная в диапазоне 10,1-15% |
не более ±0,2 % не более ±4 % не более ±6 % |
Отображение на дисплее капнограммы, EtCO2, FiCO2, ЧД |
наличие |
Регулировка скорости развертки капнограммы |
не менее 2х скоростей |
Канал оксиметрии |
Наличие |
Диапазон измерения уровня концентрации О2 в диапазоне 0-100% |
Наличие |
Абсолютная погрешность измерения уровня концентрации О2 |
не более ±2% |
Отображение концентрации О2 во вдыхаемом/выдыхаемом воздухе (FiO2/ EtO2) |
наличие |
Канал тонометрии (измерения неинвазивного артериального давления) |
Наличие |
Метод измерения осциллометрический, комбинированный, с предварительным измерением на фазе компрессии |
Наличие |
Цифровое отображение артериального давления: систолическое, диастолическое, среднее |
Наличие |
Диапазон измерения давления: взрослый режим, мм.рт.ст. нижняя граница верхняя граница детский режим, мм.рт.ст. нижняя граница верхняя граница неонатальный режим, мм.рт.ст. нижняя граница верхняя граница |
не более 0, не менее 300 не более 10, не менее 200 не более 10, не менее 135 |
Пределы допускаемой погрешности измерения, мм.рт.ст. |
не более ±3 |
Среднее время измерения: взрослый режим, сек детский режим, сек неонатальный режим, сек |
не более 30 не более 25 не более 25 |
Режимы измерений: взрослый, детский, неонатальный, автоматический, ручной |
наличие |
Регулируемое время цикла измерения в автоматическом режиме, в диапазоне от 1-240 мин, с шагом регулировки не более 1 мин |
Наличие |
Канал ННИАД: непрерывное неинвазивное измерение АД |
Наличие |
Периодическая калибровка осциллометрическим методом, обеспечивающая точность измерения у конкретного пациента. |
наличие |
Автоматический запуск измерения НИАД в случае выхода за установленные пороги тревог |
наличие |
Порог калибровки ННИАД: адаптивный (автоматический), фиксированный (устанавливаемый пользователем) |
Наличие |
Канал измерения инвазивного артериального давления (ИАД) |
Наличие |
Количество каналов |
не менее 2 |
Датчик артериального давления: чувствительность, мВ/В/мм.рт.ст. диапазон импеданса, Ом нижняя граница верхняя граница |
не более 5 не более 300, не менее 3000 |
Функция калибровки нуля |
наличие |
Цифровое отображение артериального давления: систолическое, диастолическое, среднее |
Наличие |
Типы измеряемого давления: ART (системное артериальное давление), PA (давление в легочной артерии), CVP (центральное венозное давление), ICP (внутричерепное давление) RAP (давление в правом предсердии), LAP (давление в левом предсердии) RVP (давление правого желудочка) UA (умбиликальное артериальное давление) |
Наличие |
Диапазон измерения давления, мм.рт.ст. нижняя граница верхняя граница |
не более -50, не менее 300 |
Погрешность измерения: абсолютная в диапазоне -50-100 мм рт. ст. относительная в диапазоне 100-300 мм.рт.ст. |
не более ±2 мм.рт.ст. не более ±2 % |
Отображение частоты пульсаций: по кривой ИАД |
Наличие |
Канал неинвазивного мониторинга центральной гемодинамики |
Наличие |
Неинвазивная технология измерения, не требующая использование одноразовых расходных материалов (датчиков, электродов) |
наличие |
Измеряемые параметры: УО (SV) – ударный объем, мл нижняя граница верхняя граница шаг 1 мл СВ (СО) – сердечный выброс, л/мин нижняя граница верхняя граница шаг 0,1 л/мин |
не более 10, не менее 250 не более 1, не менее 20 |
Расчетные параметры: ССС (SVR) – системное сосудистое сопротивление СИ (CI) – сердечный индекс УИ – ударный индекс, SVRI – индекс системного сосудистого сопротивления SVI – индекс ударного объема LCW – работа левого желудочка LCWI – индекс работы левого желудочка LSW – работа выброса левого желудочка LSP – мощность сокращения левого желудочка LSPI – индекс мощности сокращения левого желудочка VSV – объемная скорость выброса |
Наличие |
Регулируемое время цикла измерения, мин В диапазоне от 1 до 240, с шагом регулировки не более 1 мин |
Наличие |
Канал измерения параметров дыхания |
Наличие |
Метод измерения: импедансный |
|
Отображение респирограммы |
наличие |
Диапазон измерения частоты дыхания, 1/мин нижняя граница верхняя граница |
не более 0, не менее 160 |
Пределы допускаемой погрешности измерения частоты дыхания, 1/мин |
не более ±1 |
Сигналы тревоги |
|
3-х уровневая звуковая и визуальная тревожная сигнализация разных цветов по измеряемым параметрам |
наличие |
Фоновая подсветка цветом, зависимым от приоритета тревоги по всем измеряемым параметрам |
наличие |
Регулирование громкости звуковой тревоги |
не менее 10 градаций |
Возможность выключения подачи звукового сигнала на определенное время, мин |
не менее 2 |
Система текстовых сообщений пользователю о тревожном событии на русском языке |
наличие |
Установка порогов тревоги: ЧСС/ЧП, SpO2, SysAD, DiaAD, ЧД, температура, смещение ST-сегмента, EtCO2, FiCO2 индекс активности мозга (возможность, при подключении дополнительного модуля) |
наличие |
Нерегулируемые сигналы тревоги: асистолия, ошибка измерения давления, не установлен датчик SpO2, ЭКГ, неисправен датчик SpO2, не подстыкован разъем SpO2, низкий вольтаж ЭКГ, слабый сигнал NIBP, апноэ, сброшен датчик ЭКГ |
наличие |
Хранение и отображение данных |
|
Фиксируемые тренды: ЧСС/ ЧП, SpO2%, артериальное давление, температура (2 канала, разность температур), наполнение пульса, ЧД, величина смещения ST-сегмента, сердечный индекс СI EtCO2 |
наличие |
Отображение фиксируемых трендов: графическое, цифровое, табличное |
наличие |
Отображение всех графиков в едином поле с привязкой к действительному времени |
наличие |
Набор отображаемых графиков определяется пользователем |
наличие |
Отображение трендовых данных с дискретностью, сек Нижняя граница: Верхняя граница: |
не более 15 не менее 60 |
Базовый объем памяти регистрации, часов |
не менее 240 |
Автоматический режим запоминания |
Наличие |
Энергонезависимая память |
наличие |
Выведение на печать сохраненных трендов и волновых форм |
наличие |
Запись в память монитора и возможность печати фрагментов кривых |
не менее 30 фрагментов |
Термопринтер |
|
Количество одновременно распечатываемых кривых |
не менее 3 |
Ширина бумаги, мм |
не менее 50 |
Комплект поставки |
|
Блок монитора 1шт |
наличие |
ЭКГ кабель 5-и жильный 1 шт |
наличие |
Одноразовые ЭКГ электроды 50 шт |
наличие |
Датчик пульсоксиметрический (прищепка) 1 шт |
наличие |
Датчик температуры поверхностный 1 шт |
наличие |
Шланг для манжеты НИАД, длина не менее 2,3 м. 1 шт |
наличие |
Манжета для измерения АД, для детей 1 шт |
наличие |
Манжета для измерения АД, для взрослых 1 шт |
наличие |
Комплект для измерения CO2 (устройство соединительное, удлинительная магистраль) 5 компл. |
наличие |
Влагоотделитель 2 шт |
наличие |
Кабель интерфейсный ИАД 1 шт |
наличие |
Датчик инвазивного измерения давления, одноразовый 2 шт |
наличие |
Сетевой шнур, длина не менее 2,3 м 1 шт |
наличие |
Руководство по эксплуатации на русском языке 1 шт |
наличие |