Instruction Manual HI 9142 Portable Waterproof Dissolved Oxygen Meter 1.800.561.8187 information@itm.com www. .com… -
Page 2: Table Of Contents
TABLE OF CONTENTS GENERAL DESCRIPTION HI 9142 is a rugged, waterproof meter that solves the common PRELIMINARY EXAMINATION……….3 problems of field use, such as cold, rain, snow and dust, that can GENERAL DESCRIPTION …………3 damage a meter, rapidly deteriorating its performance and life.
-
Page 3: Preliminary Examination
TABLE OF CONTENTS GENERAL DESCRIPTION HI 9142 is a rugged, waterproof meter that solves the common PRELIMINARY EXAMINATION……….3 problems of field use, such as cold, rain, snow and dust, that can GENERAL DESCRIPTION …………3 damage a meter, rapidly deteriorating its performance and life.
-
Page 4: Probe Functional Description
METER FUNCTIONAL DESCRIPTION PROBE FUNCTIONAL DESCRIPTION 1. D.O. probe 2. Protective cap 3. Shielded cable 4. Polypropylene probe body 5. Temperature sensor 6. O-Ring seal 1. Liquid Crystal Display (LCD) 7. Silver chloride (AgCl) anode 2. On/Off button 8. Platinum cathode (sensor) 3.
-
Page 5
METER FUNCTIONAL DESCRIPTION PROBE FUNCTIONAL DESCRIPTION 1. D.O. probe 2. Protective cap 3. Shielded cable 4. Polypropylene probe body 5. Temperature sensor 6. O-Ring seal 1. Liquid Crystal Display (LCD) 7. Silver chloride (AgCl) anode 2. On/Off button 8. Platinum cathode (sensor) 3. -
Page 6: Specifications
SPECIFICATIONS PROBE INITIAL PREPARATION Probe Preparation All D.O. probes from Hanna Instrument are shipped dry. To Range 0.0 to 19.9 mg/L (ppm) hydrate the probe and prepare it for use, connect it to the -5.0 to 50.0°C (23.0 to 122.0°F) meter and proceed as follows.
-
Page 7
SPECIFICATIONS PROBE INITIAL PREPARATION Probe Preparation All D.O. probes from Hanna Instrument are shipped dry. To Range 0.0 to 19.9 mg/L (ppm) hydrate the probe and prepare it for use, connect it to the -5.0 to 50.0°C (23.0 to 122.0°F) meter and proceed as follows. -
Page 8: Calibration Procedure
PROBE POLARIZATION Connect the probe to the instrument. Turn the instrument ON by pressing The probe is under polarization with a fixed voltage of On/Off. approximately 800 mV. Probe polarization is essential for stable measurements with the At start-up the display will show same recurring degree of accuracy.
-
Page 9
PROBE POLARIZATION Connect the probe to the instrument. Turn the instrument ON by pressing The probe is under polarization with a fixed voltage of On/Off. approximately 800 mV. Probe polarization is essential for stable measurements with the At start-up the display will show same recurring degree of accuracy. -
Page 10: Taking Measurements
(a few minutes • Press CAL to exit zero calibration. for temperature difference of several degrees). The zero calibration of the HI 9142 is very stable, therefore this procedure needs only to be performed whenever the probe is replaced.
-
Page 11
(a few minutes • Press CAL to exit zero calibration. for temperature difference of several degrees). The zero calibration of the HI 9142 is very stable, therefore this procedure needs only to be performed whenever the probe is replaced. -
Page 12: Altitude And Salinity Compensation
D.O. concentration decreasing its value. The following In HI 9142 all the readouts are referred to 0 g/L of salinity value. table reports the oxygen solubility at various temperatures and In fact, salinity affects D.O. concentration decreasing its value.
-
Page 13
D.O. concentration decreasing its value. The following In HI 9142 all the readouts are referred to 0 g/L of salinity value. table reports the oxygen solubility at various temperatures and In fact, salinity affects D.O. concentration decreasing its value. -
Page 14: Probe And Membrane Maintenance
The Platinum cathode (#8 in the Functional Description at page 4) PROBE AND MEMBRANE MAINTENANCE should always be bright and untarnished. If it is tarnished or stained, which could be due to contact with certain gases or The oxygen probe body is made of reinforced plastic for maximum extended use with a loose or damaged membrane, the cathode durability.
-
Page 15
The Platinum cathode (#8 in the Functional Description at page 4) PROBE AND MEMBRANE MAINTENANCE should always be bright and untarnished. If it is tarnished or stained, which could be due to contact with certain gases or The oxygen probe body is made of reinforced plastic for maximum extended use with a loose or damaged membrane, the cathode durability. -
Page 16: Battery Replacement
ACCESSORIES BATTERY REPLACEMENT When the battery level is low, “BATT“ tag is displayed blinking HI 7040L Zero oxygen solution, 500 mL on the LCD to advise the user that approx. 1 hour of working time is left. HI 7041S Refilling electrolyte solution, 30 mL It is recommended to change the batteries as soon as the HI 7041M Refilling electrolyte solution, 230 mL…
-
Page 17: Accessories
ACCESSORIES BATTERY REPLACEMENT When the battery level is low, “BATT“ tag is displayed blinking HI 7040L Zero oxygen solution, 500 mL on the LCD to advise the user that approx. 1 hour of working time is left. HI 7041S Refilling electrolyte solution, 30 mL It is recommended to change the batteries as soon as the HI 7041M Refilling electrolyte solution, 230 mL…
-
Page 18: Warranty
Operation of this instrument in residential area could cause unacceptable HI 9142 is guaranteed for two years against defects in workmanship interferences to radio and TV equipment, requiring the operator to take and materials when used for their intended purpose and maintained all necessary steps to correct interferences.
-
Page 19
Operation of this instrument in residential area could cause unacceptable HI 9142 is guaranteed for two years against defects in workmanship interferences to radio and TV equipment, requiring the operator to take and materials when used for their intended purpose and maintained all necessary steps to correct interferences. -
Page 20
Instruction Manual HI 9142 Portable Waterproof Dissolved Oxygen Meter 1.800.561.8187 information@itm.com www. .com…
Скачать
Instruction Manual
HI 9142
Portable Waterproof
Dissolved Oxygen Meter
This instrument is in Compliance with the CE Directives
http://www.hannainst.com
Range
Resolution
Accuracy
Calibration
Temperature
Compensation
Probe
Battery Type
Life
Auto-off
Environment
Dimensions
Weight
1.800.561.8187
SPECIFICATIONS
0.0 to 19.9 mg/L (ppm)
-5.0 to 50.0°C (23.0 to 122.0°F)
0.1 mg/L
0.1°C (1°F)
±1.5% FS
±0.2°C (±1°F)
excluding probe error
Manual in 100% DO
Automatic in 0% DO
Automatic, 0 to 50°C
(32 to 122°F)
HI 76407/4, polarographic,
with 4 m (13′) cable (included)
3 x 1.5V AAA
approx. 1000 hours of use
After approx. 8 minutes
0 to 50°C (32 to 122°F);
RH max 100%
185 x 72 x 36 mm (7.3 x 2.8 x 1.4″)
300 g (10.6 oz.)
6
www.
PROBE INITIAL PREPARATION
Probe Preparation
All D.O. probes from Hanna Instrument are shipped dry. To
hydrate the probe and prepare it for use, connect it to the
meter and proceed as follows.
1. Remove the red & black plastic cap. This
cap is used for shipping purposes only and
can be thrown away.
2. Wet the sensor by soaking the bottom
(2.5 cm/1″) of the probe in HI 7041S
electrolyte solution for 5 minutes.
3. Rinse the membrane (HI 76407A supplied with the
meter) with some electrolyte
while shaking it gently. Refill
with clean electrolyte.
4. Gently tap the sides of the
membrane with your finger
to ensure that no air bubbles
remain trapped inside.
To avoid damaging the
membrane, do not tap the
membrane directly on the
bottom.
5. Place the rubber O-Ring properly inside the membrane
cap.
6. With the sensor facing down, screw the cap clockwise.
Some electrolyte will overflow.
When not in use and during polarization, protect
the membrane with the supplied cap.
information@itm.com
.com
Shipping
cap
black
red
FILL FIRST
THEN TAP
THEN SCREW
BACK ON
7
Портативный влагозащищенный кислородомер HI 9142 является простым в использовании измерителем содержания растворённого кислорода (DO) с полярографическим датчиком.
Основные особенности:
- автоматическая температурная компенсация;
- доступный и эффективный;
- датчик растворённого кислорода и чемоданчик в комплекте.
При проведении полевых работ приборы могут подвергаться воздействию ненастной погоды. Холод, дождь, пыль и влага могут привести к образованию конденсата на частях корпуса. Влагозащитный корпус прибора HI-9142 успешно противостоит многим трудностям работы в полевых условиях.
Калибровка по нулевой точке выполняется по раствору с нулевым содержанием растворённого кислорода HI-7040, а калибровка в точке 100% производится на воздухе.
Полярографический датчик (HI 76407/4) работает с точностью до 0,3 частей на миллион и поставляется с кабелем длиной 4 м, что позволяет проводить измерения даже в труднодоступных местах.
Технические характеристики | |
---|---|
Диапазон | 0,00…19,9 мг/л |
Разрешение | 0,1 мг/л |
Точность | ± 1,5% полной шкалы |
Калибровка | ручная по одной или двум точкам (ноль и наклон) |
Температурная компенсация | 0…30°C (автоматическая) |
Электрод/зонд | полярографический датчик HI 76407/кабель длиной 4 м |
Тип батареи | 1.5В AA |
Срок службы | примерно 500 часов непрерывной работы |
Температура окружающей среды | 0…50°C |
Относительная влажность | макс 100% |
Габаритные размеры | 196х80х60 мм |
Вес | 500 г |
- Портативный влагозащищенный кислородомер HI 9142
- датчик измерения содержания растворённого кислорода HI76407/4
- запасные мембраны
- раствор электролита HI7041S (30 мл)
- отвёртка для калибровки
- батарейки
- инструкция
- Руководство по эксплуатации HI 9142 — Портативный влагозащищенный кислородомер (English)
1
Instruction Manual
HI 9142
Portable Waterproof
Dissolved Oxygen Meter
w w w . h a n n a i n s t . c o m
HI 9142 — прочный портативный водостойкий измеритель растворенного кислорода (DO), разработанный для обеспечения надежных измерений в суровых условиях окружающей среды. Измеритель имеет ручную калибровку и функцию автоматического отключения. HI9142 поставляется в комплекте с полярографическим датчиком, крышками из PTFE мембраны, раствором электролита, защитным экраном и прочным футляром для переноски. *Прибор внесен в государственный реестр средств измерений * Особенности
Предназначен для проведения ежедневных измерений в процессе биологической обработки питьевой воды. Выполнен в водонепроницаемом корпусе, снабжен автоматической термокомпенсацией до 30°C, очень прост в обращении, не требует применения химических реактивов и относительно дешев. Ручная калибровка по воздуху и нулевой точке. Технические характеристики
Прибор поверяется По отдельной заявке поставляются:
Прибор вначале калибруется по данному нулевому раствору, а затем по воздуху или воде, насыщенной воздухом. Сопутствующие товары: Внимание! Производитель оставляет за собой право изменять конструкцию, технические характеристики, внешний вид, комплектацию товара без предварительного уведомления. |