EC Declaration of Conformity

We
GOOD WILL INSTRUMENT CO., LTD.

No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan.

GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.

No. 69 Lushan Road, Suzhou New District Jiangsu, China.

declare that the below mentioned product

PEL-300

is herewith confirmed to comply with the requirements set out in the Council Directive on the
approximation of the Law of Member States relating to Electromagnetic Compatibility
(89/336/EEC

,

92/31/EEC

,

93/68/EEC

)

and Low Voltage

Equipment

Directive

(

73/23/EEC,

93/68/EEC

).

For the evaluation regarding the

Electromagnetic Compatibility and Low Voltage

Equipment

Directive, the following standards are applied:

EN 61326-1:

Electrical equipment for measurement, control and laboratory use––EMC requirements

(1997+A1:1998)

Conducted Emission

Electrostatic Discharge

IEC 1000-4-2

(1995)

Radiated Emission

EN 55022 class B (1994)

Radiated Immunity

IEC 1000-4-3

(1995)

Current Harmonics

EN 61000-3-2

(1995)

Electrical Fast Transients

IEC 1000-4-4

(1995)

Voltage Fluctuations

EN 61000-3-3

(1995)

Surge Immunity

IEC 1000-4-5

(1995)

——————

———-

———-

Conducted Susceptibility

EN 61000-4-6

(1996)

——————

———-

———-

Power Frequency Magnetic field

EN 61000-4-8

(1993)

——————

———-

———-

Voltage Dip/Interruption

EN 61000-4-11

(1994)

Low Voltage Equipment Directive 73/23/EEC & amended by 93/68/EEC

IEC/EN 61010-1:2001

1. INTRODUCTION ………………………………………………………………………………………………………………………….. 1

1-1 Features……………………………………………………………………………………………………………………………… 1

2. SPECIFICATIONS………………………………………………………………………………………………………………………… 2

3. OPERATION ……………………………………………………………………………………………………………………………….. 5

3-1 Front and rear panels……………………………………………………………………………………………………………. 5

3-2 Operations ………………………………………………………………………………………………………………………….. 9

(1). Constant Voltage Mode ……………………………………………………………………………………………………. 10

(2). Constant Current Mode ……………………………………………………………………………………………………. 10

(3). Constant Resistance Mode……………………………………………………………………………………………….. 11

(6). Setting of operational duty ………………………………………………………………………………………………… 12

(7). Setting of dynamic mode ………………………………………………………………………………………………….. 13

(8). Setting of LOAD ON/OFF …………………………………………………………………………………………………. 13

(9). Setting of STEP ………………………………………………………………………………………………………………. 14

(10).Setting of MEMORY ………………………………………………………………………………………………………… 14

(12).Setting of repeated executions ………………………………………………………………………………………….. 17

(13).Setting of auto execution ………………………………………………………………………………………………….. 18

4. CALIBRATION PROCEDURES……………………………………………………………………………………………………. 19

2

CONTENTS

Нагрузка электронная программируемая PEL-300: Режимы работы нагрузки: стабилизация напряжения на нагрузке, стабилизация тока нагрузки, стабилизация сопротивления нагрузки; Характер нагрузки: статическая, динамическая

Назначение нагрузки электронной программируемой PEL-300:

  • Режимы работы нагрузки: стабилизация напряжения на нагрузке, стабилизация тока нагрузки, стабилизация сопротивления нагрузки;
  • Характер нагрузки: статическая, динамическая;
  • Входные параметры нагрузки: напряжение от 3 В до 60 В, ток от 6 мА до 60 А, мощность от 1 Вт до 30 Вт;
  • Дискретная установка входных параметров (непосредственным набором на клавиатуре или в пошаговом режиме);
  • Дискретность установки параметров нагрузки 20 мВ/0,2 мА/ 0,33 мОм;
  • Встроенный генератор переходных процессов (от 1 Гц до 1 кГц);
  • Установка ограничения по мощности;
  • Запись/считывание до 100 установленных параметров;
  • Самодиагностика и встроенная программа калибровки;
  • Современный дизайн;
  • Масса 9 кг.

Программируемая электронная нагрузка предназначена для работы в качестве нагрузки при испытании, настройке и регулировке блоков питания, усилителей, звуковоспроизводящей аппаратуры и других радиотехнических устройств с напряжением питания до 60 В, током нагрузки до 60 А, потребляемой мощностью до 300 Вт. Электронная нагрузка обеспечивает испытание в различных статических и динамических режимах работы (контроль напряжения, тока, имитация нагрузочного сопротивления).

Технические характеристики нагрузки электронной программируемой PEL-300:

ВХОДНЫЕ ПАРАМЕТРЫ
Наппряжение на нагрузке: 3 В…60 В
Ток в нагрузке: 6 мА…60 А
Потребляемая мощность: 1 Вт…300 Вт
Минимальное рабочее напряжение: 3 В

СТАБИЛИЗАЦИЯ НАПРЯЖЕНИЯ
Диапазон установки: 3 В…60 В
Погрешность установки: ±(0,1% + 40 мВ)
Дискретность установки: 20 мВ
Шаг установки: 20 мВ.,.6 В

СТАБИЛИЗАЦИЯ ТОКА, СТАТИЧЕСКИЙ РЕЖИМ
Диапазон установки: 6 мА…60 А
Погрешность установки:
±(0,1% + 1 мА) до 0,6 А / ±(0,1% + 10 мА) при 0,6 А…6 А/
±(0,1% + 100 мА) свыше 6 А
Дискретность установки: 0,2 мА (до 0,6 А) / 2 мА (0,6 А.. .6 А) / 20 мА (> 6 А)
Шаг установки: 0,2 мА.. .6 А (до 0,6 А) / 2 мА.. .6 А (0,6 А.. .6 А) / 20 мА.. .6 А (> 6 А)

СТАБИЛИЗАЦИЯ ТОКА, ДИНАМИЧЕСКИЙ РЕЖИМ
Диапазон частот: 1 Гц…1 кГц (±5%)
Дискретность установки: 0,05 Гц (до 10 Гц) / 0,5 Гц (10 Гц. ..100 Гц) / 5 Гц (> 100 Гц)
Коэффициент заполнения импульсов: 10%.. .90% (±10%)
Дискретность установки: 1%

СТАБИЛИЗАЦИЯ СОПРОТИВЛЕНИЯ
Диапазон установки: 50 мОм…1 кОм
Погрешность установки:
±(5% + 1 мОм) до 10 Ом / ±(5% + 100 мОм) при 10 Ом… 100 Ом /
±(5% + 1 Ом) свыше 100 Ом
Дискретность установки:
0,33 мОм (до 1 Ом) / 3,3 мОм (1 Ом…10 Ом) /
33 мОм (10 Ом…100 Ом) / 330 мОм (> 100 Ом)
Шаг установки:
0,33 мОм…1 Ом (до 1 Ом) / 3,3 мОм…Ю Ом (1 Ом…10 Ом) /
33 мОм…Ю0 Ом (10 Ом…100 Ом) / 330 мОм…Ю0 Ом (> 100 Ом)

ОГРАНИЧЕНИЕ ПО МОЩНОСТИ
Диапазон установки: 1 Вт…300 Вт
Погрешность установки: ±(2,0% + 4 Вт)
Дискретность установки: 0,1 Вт

ПАМЯТЬ
Функции: Запись / считывание установленных профилей
Объем: 99 ячеек

ТАЙМЕР
Функции: Задание времени циклического режима работы
Диапазон установки: 1 с…999 мин. 59 с
Дискретность установки: 1 с

ОБЩИЕ ДАННЫЕ
Напряжение питания: 100 В / 120 В / 220 В / 240 В ±10%, 50 / 60 Гц
Габаритные размеры: 255 х 145 х 346 мм
Масса: 9 кг

Комплект поставки PEL-300:

1.Нагрузка электронная программируемая PEL-300 — 1
2.Руководство по эксплуатации — 1

Производство компании GW Instek (Тайвань)

Table of Contents for GW Instek PEL-300:

  • 37 FOR UNITED KINGDOM ONLY NOTE: This lead/appliance must only be wired by competent person. WARNING: THIS APPLIANCE MUST BE EARTHED IMPORTANT: The wires in this lead are coloured in accordance with the following code: Green/ Yellow: Earth Blue: Neutral Brown: Live (Phase) As the colours of the wires in main leads may not correspond with the colours marking identified in your plug/appliance, proceed as follows: The wire which is coloured Green & Yellow must be connected to the Earth terminal marked with the letter E or by the earth

  • 35 5-2 Test Lead The Selection of Output Test Lead and Feedback Test Lead: For safety assurance, please select the adequate output test lead according to the following list: ConductorUL (CSA) Model Wire No. AWG Component pc/mm Cross Section Area (mm) 2 Outer Diameter mm Maximum Conductive Resistor Ω/km Permissible Current A(amp) 24 11/0.16 0.22 0.64 88.6 7.64 22 17/0.16 0.34 0.78 62.5 10.0 20 21/0.18 0.53 0.95 39.5 13.1 18 34/0.18 0.87 1.21 24.4 17.2 16 26/0.254 1.32 1.53 15.6 22.6 14 41/0.254 2.08 2.03 9.90 30.4 12 65/0.254 3.29 2.35 6.24 40.6 1015 T

  • 18 (13). Setting of auto execution: n The light of AUTO is out when auto execution is not functioning. o Press AUTO button, the indicator will light up, and the function will be turned on. Note: Be sure to complete the settings of item 1 to 7 and 10 to 12 before proceeding the setting of auto execution. Example: To set auto execution STEP 1. Press AUTO key (the light of AUTO is on) STEP 2. Press LOAD ON/OFF to proceed auto execution

  • 11 (3). Constant Resistance Mode: n Connect the instrument according to Graph 3-3. o Turn the power switch to “ON” position, then press C.R. MODE button (the light of C.R. MODE is on). p Press LOAD ON/OFF button (the light of LOAD is on), the displays will indicate the value of external voltage and current. Example: To set input resistance at 500Ω in constant resistance mode STEP 1. Press C.R. MODE STEP 2. Press V.I.R. SET STEP 3. Press 500 (unit is specified as “Ω” ) STEP 4. Press ↵ (4). Setting of

  • 2 2. SPECIFICATIONS Functions Modes Ranges Voltage Range 3 ~ 60V Current Range 6mA ~ 60AOperational Limits Power Range 1 ~ 300W Applicable Range 3 ~ 60 V Accuracy ±(0.1% + 40mV) Resolution 20mV Constant Voltage (CV) Mode Step 20mV ~ 6V Applicable Range 6mA ~ 60V 6A ~ 60A ±(0.5% + 100mA) 0.6A ~ 6A ±(0.1% + 10mA)Accuracy 6mA ~ 0.6A ±(0.1% + 1mA) 6A ~ 60A 20mA 0.6A ~ 6A 2mAResolution 6mA ~ 0.6A 0.2mA 6A ~ 60A 20mA ~ 6A 0.6A ~ 6A 2mA ~ 6A Constant Current (CC) Mode Step 6mA ~ 0.6A 0.2mA ~ 6A Range 1Hz ~ 1kHz Accuracy ±5% 100Hz ~ 1kHz 5Hz 10Hz ~ 100Hz 0.5Hz Frequency Reso

  • 1 1. INTRODUCTION 1-1 Features  Simple operation with smart push buttons  High resolution (20mV, 0.2mA, 0.33mΩ)  With voltage, current and power overload prevention functions  Easy access to maximum 100 files  Voltage range: 3 ~ 60V  Current range: 6mA ~ 60A  Maximum power: 300W  Operational modes: constant voltage (CV), constant current (CC), and constant resistance (CR)  With self-testing and auto software calibration functions  Meet UL CSA IEC requirement

  • 6 Graph 3-2 Rear Panel

  • 7 1 POWER : “▃” — power is on; “▉” — power is off. 2 C.C. MODE : To set operation in constant current mode. 3 C.V. MODE : To set operation in constant voltage mode. 4 C.R. MODE : To set operation in constant resistance mode. 5 RCL V : To recall the previous batch of memorized data. 6 RCL W : To recall the next batch of memorized data. 7 STEP V : To add a STEP value to the values of pre-set voltage, current and resistance. 8 STEP W : To subtract a STEP value from the values of pre-set voltage, current and resistance. 9 SW ON/OFF : To sw

  • 33 4-19 D/A Calibrating the SENSE current offset of C.C., C.R. MODE n Set voltage of POWER SUPPLY to 60V and current of it to 1A. o Connect the output terminal of POWER SUPPLY to a DMM in 2A range in series. Hook up the DMM with the input terminal of the load. p Test the input terminal of the load by the 200V range of a DMM. Press SHIFT and C.V. MODE, then key in 3, 8, 1, 8, and ↵ . Adjust the POWER SUPPLY until the DMM display reads 60.00V. q When MEMORY indicates “01” and A indicates “CL18”, key in the current reading (7 digits including deci

  • 3 4-8 Calibrating the D/A, A/D in 0.6A range of C.C. MODE ………………………………………………………………24 4-9 Calibrating the D/A in 0.6A range of W SET…………………………………………………………………………….24 4-10 Calibrating the D/A, A/D in 60A range of C.C. MODE …………………………………………….

  • 20 4-2 Adjusting the reference voltage in 10Ω∼1kΩ range of C.R. MODE n Set voltage of POWER SUPPLY to 60V and current of it to 1A. o Connect the output terminal of POWER SUPPLY to the input terminal of electronic load. p Press SHIFT and C.V. MODE, then key in 3, 8, 0, 2, and ↵. q Test the input terminal of load by the 200V range of a DMM. Adjust the POWER SUPPLY until the DMM display reads 60.00V. Use the 20V range of another DMM to test TP1 of

  • 8 13 V.I.R. SET : To set the values of input voltage, current, and resistance. 14 W SET : To set the maximum of input power. 15 FREQ. SET : To set the frequency in dynamic operational mode. 16 DUTY SET : To set DUTY under dynamic operational mode. 17 0 ∼ 9 : The buttons for input numbers. 18 ↵ : To complete data inputting. 19 DC INPUT : The input terminal of the electronic load (Max.10A ). 20 LOCAL : Not used 21 MEMORY : To display different sets of data memorized. 22 DISPLAY A : To display current

  • 36 6. SAFETY SUMMARY SAFETY TERMS AND SYMBOLS: These terms may appear in this manual or on the product: WARNING. Warning statements identify condition or practices that could result in injury or loss of life. CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property . WARNING: This equipment is not for measurements performed for CAT II, III and IV. Measurement category I is for measurements performed on circuits not directly connec

  • 32 4-18 Calibrating the D/A in 1Hz∼10Hz range of SW MODE n Set voltage of POWER SUPPLY to 60V and current of it to 1A. o Connect the positive output terminal of POWER SUPPLY to a 0.1Ω 5W concrete resistance first, then connect the resistance to the input terminal of the electronic load. Hook up a DMM in 200V range with the input terminal of the load. Connect the input terminal of COUNTER to both ends of the concrete resistance. p Test the input terminal of loa

  • 3 Functions Modes Ranges Applicable Range 50mΩ ~ 1kΩ 100Ω ~ 1kΩ±(5% + 1Ω) 10Ω ~ 100Ω±(5% + 100mΩ) 1Ω ~ 10Ω±(5% + 10mΩ) Accuracy 50mΩ ~ 1Ω±(5% + 1mΩ) 100Ω ~ 1kΩ 330mΩ 10Ω ~ 100Ω 33mΩ 1Ω ~ 10Ω 3.3mΩ Resolution 50mΩ ~ 1Ω 0.33mΩ 100Ω ~ 1kΩ 330mΩ ~ 100Ω 10Ω ~ 100Ω 33mΩ ~ 100Ω 1Ω ~ 10Ω 3.3mΩ ~ 10Ω Constant Resistance (CR) Mode Step 50mΩ ~ 1Ω 0.33mΩ ~ 1Ω Applicable Range 1 ~ 300W Accuracy ±(2% + 4W)Power Setting Resolution 0.1W Accuracy

  • 12 (5). Setting of operational frequency: n Connect the instrument according to Graph 3-3. o Turn the power switch to “ON” position, then press C.C. MODE button (the light of C.C. MODE is on). p Press FREQ. SET button; then input desired FREQ. value after the display indicating the original FREQ. value. Example: To set operational frequency as 1000Hz STEP 1. Press FREQ. SET STEP 2. Press 1000 (unit is specified as “Hz” ) S

Questions, Opinions and Exploitation Impressions:

You can ask a question, express your opinion or share our experience of GW Instek PEL-300 device using right now.

Нагрузка электронная программируемая PEL-300: Режимы работы нагрузки: стабилизация напряжения на нагрузке, стабилизация тока нагрузки, стабилизация сопротивления нагрузки; Характер нагрузки: статическая, динамическая; Входные параметры нагрузки: напряжение от 3 В до 60 В, ток от 6 мА до 60 А, мощность от 1 Вт до 30 Вт; Дискретная установка входных параметров (непосредственным набором на клавиатуре или в пошаговом режиме); Дискретность установки параметров нагрузки 20 мВ/0,2 мА/ 0,33 мОм; Встроенный генератор переходных процессов (от 1 Гц до 1 кГц); Установка ограничения по мощности; Запись/считывание до 100 установленных параметров; Самодиагностика и встроенная программа калибровки; Современный дизайн; Масса 9 кг.

Программируемая электронная нагрузка предназначена для работы в качестве нагрузки при испытании, настройке и регулировке блоков питания, усилителей, звуковоспроизводящей аппаратуры и других радиотехнических устройств с напряжением питания до 60 В, током нагрузки до 60 А, потребляемой мощностью до 300 Вт. Электронная нагрузка обеспечивает испытание в различных статических и динамических режимах работы (контроль напряжения, тока, имитация нагрузочного сопротивления).