SPOTCHEM EZ

SP — 4430

Operating Manual

b

loading

Related Manuals for Arkray SPOTCHEM EZ SP-4430

Summary of Contents for Arkray SPOTCHEM EZ SP-4430

  • Page 1
    SPOTCHEM EZ SP — 4430 Operating Manual…
  • Page 2
    This manual contains important information on the functions of the SPOTCHEM EZ SP-4430. supplied by : ARKRAY, Inc. This manual is issued by: ARKRAY, Inc. Read carefully prior to starting up the unit. It is recommended to retain this manual for future use.
  • Page 3
    Introduction The SP-4430 is an automated biochemical analyzer using a dry chemistry system. This manual contains outlines, instructions for the operation, maintenance and troubleshooting for the SPOTCHEM SP-4430. Before operating the SP-4430, read this manual carefully. It is recommended to retain this manual for future use. WARNING Always be careful when handling blood samples or waste reagent ●…
  • Page 4
    Caution Marks Pay careful attention to these caution marks in this Operating Manual. ■Regarding accident resulting in injury or death. To prevent infection of yourself or others from WARNING pathogenic microbes, follow the instructions given herein. To prevent injuries to yourself or others, or material CAUTION damages, follow the instructions described.
  • Page 5
    Caution Labels Caution Stickers are adhered to the SP-4430 to prevent accidents as below. The descriptions are given on the following pages. Biohazard cautions. Wear protective gloves to prevent exposure to pathogenic microbes. The centrifuge turn round at high speed. Keep your hands off when mesurement is in process. Biohazard cautions.
  • Page 6: Table Of Contents

    Contents Automated biochemical analyzer, SP-4430 Premise …………………………………………………………………………………… Introduction …………………………………………………………………………… Caution Marks ……………………………………………………………………… Caution Labels …………………………………………………………………… Contents ……………………………………………………………………………… 1-1 Outline Chapter 1 …………………………………………………………………………… 1-02 1-1-1 Features INTRODUCTION ………………………………………………………………… 1-02 1-1-2 Measurement principles ………………………………………… 1-03 1-1-3 Specifications ……………………………………………………… 1-04 1-2 Shipping Carton ……………………………………………………………

  • Page 7
    2-5 Calibration …………………………………………………………………… 2-28 2-5-1 Overview of calibration ………………………………………… 2-28 2-5-2 Calibration by magnetic card ………………………………… 2-29 2-5-3 Calibration by calibrator kit ………………………………… 2-31 2-6 Interpreting measurement results ……………………… 2-39 2-6-1 Printing normal measurement results …………………… 2-39 2-6-2 Printing calibration results ………………………………………
  • Page 8
    07 ● SP — 4430…
  • Page 10: Outline

    Chapter 1 INTRODUCTION The SP-4430 is an automated biochemical analyzer using a dry chemistry system. This chapter contains the information on the functions and measurement principles of the SP-4430. 1-1 Outline 1-1-1 Features 1-1-2 Measurement principles 1-1-3 Specifications 1-2 Shipping Carton 1-2-1 Analyzer 1-2-2 Accessories 1-3 Parts Description and Function…

  • Page 11
    1-1 Outline Chapter 1 INTRODUCTION 1-1-1 Features The footprint of the SP-4430 is as small as that of a ■Compact and lightweight notebook computer. It weighs only 5 kg and can be design easily carried by one person. The small unit contains various components such as a display, printer, 1- sample centrifuge and automatic tip disposal mechanism.
  • Page 12
    SP — 4430の概要 Chapter 1 INTRODUCTION 1-1-2 Measurement principles The light emitted from LED becomes monochromatic light of a certain wavelength after passing through an optical fiber (Five different types of optical filters are provided, and the optimum wavelength is selected for each test item). Monochromatic light is separated and transmitted to the photometry section of each channel by ten optical fibers.
  • Page 13
    SP — 4430の概要 Chapter 1 INTRODUCTION 1-1-3 Specifications Sample Serum, Plasma, Whole blood (Only Hb) Measurement item General biochemical measurement items, 22 items Measurement wavelength 5 wavelengths (405, 550, 575, 610 and 820 nm) Measurement principle Optical measurement of reflection intensity of reagent color reaction Measurement Range Set for sach measurement item…
  • Page 14
    Shipping Carton Chapter 1 INTRODUCTION 1-2-1 Analyzer Open the package and confirm that all items are included. 1・Analyzer (S-4430) 2・Power cord 3・AC adapter 4・Accessory case 5・Operating 6・Warranty card with case,   manual Delivery label Item Description Qty. SP-4430 Analyzer Power cord AC adapter Accessory case Operating manual…
  • Page 15
    梱包品の確認 Chapter 1 INTRODUCTION 1-2-2 Accessories Open the package and confirm that all items are included. 1・Tip 2・Cleaning wire 3・Cleaner set 4・Nozzle set 5・Wrench set for nozzle replacement 6・Centrifuge tube 7・Tip waste case 8・Protective cover 9・Thermal printer paper Item Description Qty.
  • Page 16
    Parts Description and Function Chapter 1 INTRODUCTION 1-3-1 Front of the analyzer 12・Built-in printer 11・Magnetic card reader 10・Display 9・Operator panel 7・Black & white plates 13・Maintenance cover 1・Reagent table 3・Port 5・Tip hole 6・Tip waste case 4・Centrifuge 14・Protective cover 8・Table cover 2・Centrifuge-equipped Multi Rack Function Item Reagent table…
  • Page 17
    各部の名称とはたらき Chapter 1 INTRODUCTION 1-3-2 Operator panel Item Function START Starts measurement. Select Yes from the Yes/No option. STOP Stops measurement. Select No from the Yes/No option. Cancels entry. FEED Feeds the built-in printer with paper while pressed. MENU Switches the page on each menu display. 0~9 (ten key) Selects the menu number.
  • Page 18
    SP — 4430の概要 Chapter 1 INTRODUCTION 1-3-3 Rear of the analyzer 1・Power switch 3・Cooling Fan 5・B.C.R.  7・Paper holder 2・Power input terminal  4・COM.  6・CONT. Item Function Power switch For turning on and off the power supply to the analyzer. Power input terminal For the supplied AC adapter. Cooling Fan For Ventilating heated air in the unit to prevent overheating.
  • Page 19
    1-4 Setting up the Analyzer Chapter 1 INTRODUCTION 1-4-1 Cautions CAUTION 10cm or more AC100-120V 50 / 60Hz Before setting up the analyzer, read the following notes and always take proper AC220-240V safety precautions. Allow a space of 10 cm or more between Operate the analyzer with power of the ●…
  • Page 20
    SP — 4430の概要 Chapter 1 INTRODUCTION CAUTION Before setting up the analyzer, read the following notes and always take proper safety precautions. Do NOT set up the analyzer where Avoid exposing the analyzer to direct ● ● chemicals are stored nearby, or where sunlight, condensation and wind.
  • Page 21
    装置の設置 Chapter 1 INTRODUCTION 1-4-2 Setting up the analyzer The parts in the analyzer are secured firmly in order to prevent scratches and/or damage due to transportation. Remove the fixing tapes before setting up the analyzer. Read 1-4-1 “Cautions on installation” carefully before setting up the analyzer.
  • Page 22
    SP — 4430の概要 Chapter 1 INTRODUCTION 1-4-3 First operation after setting up This section explains the way to load thermal printer paper, and set data and time after turning on the power switch. Turn on the power. Turn on the power switch on the rear ●…
  • Page 23: Notes At Transportation

    装置の設置 Chapter 1 INTRODUCTION 1-4-4 Notes at transportation Remove the tape fixing the ● protective cover. Set the maintenance cover and the ● tip waste case. Press [STOP] to close the table ● C o v e r i s c l o s i n g . /…

  • Page 24
    装置の設置 Chapter 1 INTRODUCTION 1-4-4 Notes at transportation When carrying the analyzer in cars, use an exclusive carrying case. The carrying case is optional. Follow the following procedures to prevent exposure to pathogenic microbes and damage to the analyzer. Wear protective gloves to prevent exposure to pathogenic microbes. WARNING Inspect the inside of the analyzer.
  • Page 25
    1 — 16 ● SP — 4430…
  • Page 27
    Chapter 2 MEASUREMENT The normal measurement and two types of calibration can be done with the SP-4430. This chapter describes the outline and operating procedures of each measurement. 2-1 Outline of Measurement 2-1-1 SPOTCHEMⅡ Reagent Strips 2-1-2 Normal measurement 2-1-3 Calibration 2-2 Cautions 2-2-1 Cautions 2-2-2 Handling samples…
  • Page 28: Outline Of Measurement

    2-1 Outline of Measurement Chapter 2 MEASUREMENT 2-1-1 SPOTCHEMⅡ Reagent Strips SPOTCHEMⅡ Reagent Strips (optional) must be used with this analyzer. There are two types of SPOTCHEM Reagent Strips. Use the correct type of strips for the measurement purpose. Multi Reagent Strip ●…

  • Page 29: Normal Measurement

    測定を始める前に Chapter 2 MEASUREMENT 2-1-2 Normal measurement In normal measurement, more than one item can be measured at a time for one sample. The Reagent Strip Table can hold three Single Reagent Strips and one Multi Reagent Strip, so that simultaneous measurement of up to nine items is available.

  • Page 30: Calibration

    装置の設置 Chapter 2 MEASUREMENT 2-1-3 Calibration This analyzer provides two types of calibration: Calibration by magnetic card (Reagent Card) and Calibration by Calibrator. Calibration must be performed for all fields of Reagent Strips (see “2-5-1 Outline of calibration” for details). For calibration by Calibrator, the following two types of Calibrator Kit and Magnetic Card (Calibration Card) are used.

  • Page 31
    装置の設置 Chapter 2 MEASUREMENT 2-1-4 校正について The flowchart of calibration. SP — 4430 ● 2 — 05…
  • Page 32: Cautions

    2-2 Cautions Chapter 2 MEASUREMENT 2-2-1 Cautions temperature:20 〜 25℃ :30 〜 60% humidity Before turning ON the power, reconfirm The analyzer has a temperature control If there is anything wrong with the ● ● ● “1-4-1 Cautions” so that the analyzer is function in order to yield correct analyzer, or in case of odor or smoke, always…

  • Page 33: Handling Samples

    装置の設置 Chapter 2 MEASUREMENT 2-2-2 Handling samples Wear protective gloves Blood is used as a measurement sample Handling samples slightly differs ● ● with this analyzer. Be aware that blood may depending on test items. Follow the be contaminated with pathogenic microbes instructions on the package insert of that may cause infectious disease.

  • Page 34: Handling Magnetic Cards

    測定時の注意 Chapter 2 MEASUREMENT 2-2-4 Handling magnetic cards Do NOT place a magnetic card close Do NOT scratch the magnetic surface (supplied Do NOT use a magnetic card ● ● ● to a magnetic object (magnet, TV set, (stripe). Information on the magnetic with Reagent Strips or Calibrator) etc.) Information on the magnetic card card may become unreadable.

  • Page 35: Handling Centrifuges

    装置の設置 Chapter 2 MEASUREMENT 2-2-6 Handling centrifuges Uncap the tube Samples with extremely high hematocrit levels Heparin is contained as an anticoagulant. Samples with extremely high hematocrit Pipette 250 μL  to 270 μL  of  whole blood ● ● ● Close the cap tightly after setting a levels may yield incorrect measurement into the centrifuge tube.

  • Page 36: Preparation For Measurement

    2-3 Preparation for Measurement Chapter 2 MEASUREMENT 2-3-1 Preparation Prepare necessary instruments as described in the following table before starting Normal Measurement, Calibration by Magnetic Card or Calibration by Calibrator. IMPORTANT Take the Reagent Strips out of the refrigerator and leave them at room temperature (10〜 Single Reagent Strips 30℃) for 20 minutes.

  • Page 37: Startup

    装置の設置 Chapter 2 MEASUREMENT 2-3-2 Startup The operating procedures from turning ON the power to the MAIN MENU are described below. NOTE Before turning ON the power, make sure that the maintenance cover is in position. If light enters into the analyzer during initialization, an error will occur.

  • Page 38: Checks Before Measurement

    測定の準備 Chapter 2 MEASUREMENT 2-3-3 Checks before measurement Check each part before performing Normal Measurement or Calibration. WARNING Wear protective gloves to prevent exposure to pathogenic microbes. CAUTION Discard used samples, tips and protective gloves separately from general waste according to the local regulations on biohazardous waste. Thermal printer paper.

  • Page 39: Preparation Of Samples

    装置の設置 Chapter 2 MEASUREMENT 2-3-4 Preparation of samples Prepare samples as described below except for Hb and HDL-C, which require different procedures (see pages 2-16 and 2-17 for sample preparation for Hb and HDL-C). Test items may be added or changed in the future.

  • Page 40
    測定の準備 Chapter 2 MEASUREMENT 2-3-3 測定前の確認 Even if the sample stays at the Inverting a sample to mix bottom after being inverted, heparin is well mixed with the Gently invert a sample tube. 1・ sample. Do NOT shake the tube vigorously. In 3 seconds, invert the sample 2・…
  • Page 41
    Chapter 2 MEASUREMENT Using a centrifuge Prepare a centrifuge. ● A Centrifuge tube has 1 line Uncap the tube and pipette 250 μL of ● marked on it for 250μL. heparinized whole blood or plain blood into the centrifuge tube. The centrifuge tube includes Heparin as an ●…
  • Page 42
    Chapter 2 MEASUREMENT Measuring the Hb When measuring Hb, dilute whole blood Orange cap samples 6 times with distilled water. Prepare a Whole blood Sample Tube (orange ● Whole blood sample tube cap). Uncap the tube and pipette the whole blood ●…
  • Page 43
    Chapter 2 MEASUREMENT If HDL-C reagent(direct method) Using HDL-C Kits is used for measurement, this procedure is unnecessary. Prepare a Serum Sample Tube (blue cap). ● Serum or plasma (100 ) Pipette 100μL of serum or plasma. ● Gently pipette 100 μL of HDL Separation ●…
  • Page 44: Measurement

    2-4 Measurement Chapter 2 MEASUREMENT 2-4-1 Normal measurement This section describes the operating procedure of Normal Measurement. In Normal Measurement, simultaneous measurement of up to nine items is possible for one sample, using Single and Multi Reagent Strips. Before measurement, make necessary preparation as instructed in “2-3 Preparation for Measurement”.

  • Page 45
    測定操作 Chapter 2 MEASUREMENT 2-4-1 通常測定 Check the lot number of Reagent Strips. Press [2] key on the Standby screen. ● 【 Sー01:G G T 】 ( 1/30 ) 【 X X X X X X C A R D 】 The Confirm screen is displayed, ●…
  • Page 46
    測定結果 Chapter 2 MEASUREMENT 2-4-1 通常測定 The number of the previous Set measurement No. or ID (if necessary). measurement No + 1 is displayed ID of up to 13 characters can be set for each measurement. If ID setting is not as the initial number.
  • Page 47
    測定結果 Chapter 2 MEASUREMENT 2-4-1 通常測定 No set-up is available where Set the sample type (when necessary). sample type is not registered. Sample type can be set for each measurement. When sample type is not The registration can be made on the Parameter Menu.
  • Page 48
    装置の設置 Chapter 2 MEASUREMENT 2-3-4 検体の準備 Set the samples. Measurement of uncentrifuged whole blood. ■ Open the protective cover and place ● the uncapped centrifuge tube in position. Close the cover. ● Make sure the sample tubes are NOTE uncapped, otherwise it may cause breakage of centrifuge tube and the Nozzle.
  • Page 49
    測定結果 Chapter 2 MEASUREMENT 2-4-1 通常測定 ■ Remeasurement (or additional measurement) with a centrifuge. When measuring the once-measured sample with a centrifuge, such measurement can be done at a reduced centrifugal time. IMPORTANT When performing remeasurement or additional measurement , sample quantity may be short.
  • Page 50
    装置の設置 Chapter 2 MEASUREMENT 2-3-4 検体の準備 Remove a Reagent Strip from the ● aluminum foil package without touching the reagent field with the fingers. reagent field Place the Reagent Strips on the ● Reagent Table by holding the right side of the strip. Be sure to insert the end of a IMPORTANT Reagent Strip into the groove of the…
  • Page 51
    測定結果 Chapter 2 MEASUREMENT 2-4-1 通常測定 Remove the Reagent Strip from the ● package without touching the reagent field with fingers. reagent field Insert Reagent Strips on the ● Reagent Table as shown at right. Be sure to insert the end of a IMPORTANT Reagent Strip into the groove of the Reagent Table so that the Reagent…
  • Page 52
    装置の設置 Chapter 2 MEASUREMENT 2-3-4 検体の準備 As the measurement proceeds, the ● M e a s u r i n g ・・・ 01:23 display of approximate remaining S t o p( S T O P ) time changes to the ordinary time indication.
  • Page 53
    測定結果 Chapter 2 MEASUREMENT 2-4-1 通常測定 Remove any used Reagent Strips ● and discard them. When the message “DISCARD TIPS” is displayed, remove the Tip Waste Case and discard used tips. Reattach the Tip Waste Case. ● To discontinue measurement, press ●…
  • Page 54: Calibration

    2-5 Calibration Chapter 2 MEASUREMENT 2-5-1 Overview of calibration Calibration is necessary to maintain measurement accuracy. Calibration reduces unfavorable effects of reagent pack that changes with time, eliminates gaps among lots or facilities, and maintains measurement accuracy to a certain level. Basically, calibration should be performed using the SPOTCHEM Calibrator Kit for each type of Reagent Strips, but it requires a lot of time to do so for all types of Reagent Strips.

  • Page 55: Calibration By Magnetic Card

    装置の設置 Chapter 2 MEASUREMENT 2-5-2 Calibration by magnetic card By inserting “Reagent Card” attached to Single Reagent Strips or another one attached to Multi Reagent Strips into a magnetic-card reader, differences among lots and changes with time of reagent pack are automatically calibrated.

  • Page 56
    校正 Chapter 2 MEASUREMENT 2-5-1 校正の概要 When the stripe is inserted, the ● I n s e r t a c a r d . screen shown at the right is T h e s a m e s t r i p e displayed.
  • Page 57: Calibration By Calibrator Kit

    装置の設置 Chapter 2 MEASUREMENT 2-5-3 Calibration by calibrator kit Use the SPOTCHEM Calibrator Kit for calibration. For Hb items, use the SPOTCHEM Calibrator Hb Kit. WARNING Wear protective gloves to prevent exposure to pathogenic microbes. CAUTION Discard used samples, tips and protective gloves separately from general waste according to local regulations on biohazardous waste.

  • Page 58
    校正 Chapter 2 MEASUREMENT 2-5-1 校正の概要 To check the calibration mode and the ● 【 Sー01:G G T 】 number of calibration times, press [3] ( 1/30 ) 【 L & H N=3 】 key on the CALIBRATION MENU to display the Calibration information checking screen.
  • Page 59
    校正 Chapter 2 MEASUREMENT 2-5-1 校正の概要 Insert calibration card. Press [2] key on the CALIBRATION To stop calibration, press ● I n s e r t a c a r d . [STOP] key. MENU. T h e s a m e s t r i p e The Magnetic card entry screen is displayed.
  • Page 60
    装置の設置 Chapter 2 MEASUREMENT 2-5-3 キャリブレータ校正 When performing 1-point Set the Tip. calibration (High), proceed to Set the tips in the tip hole of the procedure 9. ● Centrifuge-equipped Multi Rack using tweezers. Do NOT touch the end of the Tip IMPORTANT with bear hands.
  • Page 61
    校正 Chapter 2 MEASUREMENT 2-5-1 校正の概要 Insert the Reagent Strips with the ● same items, starting from the S1 channel on the Reagent Table. If the number of calibration is set as more than 3 for 2-point calibration, calibration should be sperately performed twice.
  • Page 62
    装置の設置 Chapter 2 MEASUREMENT 2-5-3 キャリブレータ校正 Start measurement of the “Low” calibrator. ● Press [START] key. The Reagent Table Press [STOP] key to stop the M e a s u r i n g ・・・ and Centrifuge equipped Multi Rack calibration, the Calibration S t o p(…
  • Page 63
    校正 Chapter 2 MEASUREMENT 2-5-1 校正の概要 When performing 1-point Set the tip. calibration for «Low» only, proceed to procedure 13. Set the tip by following procedure 5. ● Set the “High” Calibrator. Press [STOP] key to stop Set the “High” Calibrator by following procedure 6. ●…
  • Page 64
    校正 Chapter 2 MEASUREMENT 2-5-1 校正の概要 When a message, “Remove used ● tips.” is displayed, remove the Tip Waste Case and discard used tips. Set the Tip Waste Case again. ● R e m o v e u s e d t i p s . O K(…
  • Page 65: Interpreting Measurement Results

    2-6 Interpreting measurement results Chapter 2 MEASUREMENT 2-6-1 Printing normal measurement results To print out the measurement results, Normal printing and Survey Mode printing are available. The Survey Mode has the following additional items to the print out of the Normal mode.

  • Page 66
    測定結果のみかた Chapter 2 MEASUREMENT 2-6-1 通常測定結果 Normal printing ( 1 ) ( Man ) When “Printing of measurement Version. 1・ results” is performed on the Measurement date and time : Date and time when [START] key is pressed. 2・ Sub Menu, the results are Measurement number : The number is counted starting from 0001 (When the power is on.) 3・…
  • Page 67
    装置の設置 Chapter 2 MEASUREMENT 2-5-3 キャリブレータ校正 Survey Mode printing (for 1 item) ( 2 ) Measurement value which temperature conversion, unit conversion or correlation 1・ correction is not applied. Temperature. Printed out as ….. for the items other than enzyme items. 2・…
  • Page 68: Printing Calibration Results

    測定結果のみかた Chapter 2 MEASUREMENT 2-6-2 Printing calibration results The calibration results are printed when calibration by calibrator kit is completed. The formats of the Single Reagent Strips and the Multi Reagent Strips are slightly different. Even when calibration errors (E 30 : Abnormal data) occur during the calibration the calibration results are printed.

  • Page 69
    測定結果のみかた Chapter 2 MEASUREMENT 2-6-2 キャリブレータ校正結果 Calibration Result of Multi Reagent Strip ( 2 ) From the left, name of Multi Reagent Strip, item symbol, lot number and expiry date . 1・ From the left, item name and item symbol. 2・…
  • Page 70
    2 — 44 ● SP — 4430…
  • Page 72: Overview Of Sub Menu

    Chapter 3 SUB MENU With the SUB MENU, auxiliary operations other than normal measurement and calibration can be performed. This chapter describes the SUB MENU functions and operating procedure. 3-1 Overview of SUB MENU 3-1-1 Composition of SUB MENU 3-2 Measurement Results Menu 3-2-1 Print measurement results 3-2-2 Transmit measurement results 3-2-3 Delete measurement results…

  • Page 73: Overview Of Sub Menu

    3-1 Overview of SUB MENU Chapter 3 SUB MENU 3-1-1 Composition of SUB MENU ■ MAIN MENU After the power is ON and warm-up is completed, the MAIN MENU is displayed. In the MAIN MENU, normal measurement, calibration, and the SUB MENU are available.

  • Page 74
    装置の設置 Chapter 3 SUB MENU 1-4-3 設設置後の始動と終了 SP — 4430 ● 3 — 03…
  • Page 75: Measurement Results Menu

    3-2 Measurement Results Menu Chapter 3 SUB MENU 3-2-1 Print measurement results If [STOP] key is pressed during Print measurement results stored in the memory (Max. 100 measurements). operation, the previous screen The following three options for printing are available. is restored.

  • Page 76
    装置の設置 Chapter 3 SUB MENU 2-3-4 測定の準備 Enter search condition. If designation of the range of Enter the dates of starting and ● D a t e <99ー01ー01>Y MD measurement dates (the dates finishing measurements by sliding of starting and finishing t o <00ー12ー31>…
  • Page 77: Transmit Measurement Results

    測定結果メニュー Chapter 3 SUB MENU 3-2-2 Transmit measurement results If [STOP] key is pressed during Transmit measurement results (Max. 100 measurements) stored in the operation, the previous screen memory. is restored. The following three types for transmission are available. In case of retransmission, 1.

  • Page 78
    装置の設置 Chapter 3 SUB MENU 2-3-4 測定の準備 Enter searching condition. If designation of the range of Enter the dates of starting and ● D a t e <99ー01ー01>Y MD measurement dates (the dates finishing measurements by sliding of starting and finishing t o <00ー12ー31>…
  • Page 79: Delete Measurement Results

    測定結果メニュー Chapter 3 SUB MENU 3-2-3 Delete measurement results Delete all the measurement results stored in memory. Display setting screen. Press [2] key on the MAIN MENU. ● 1 . M e a s u r e 2 . S u b m e n u SUB MENU1/2 is displayed.

  • Page 80
    測定結果メニュー Chapter 3 SUB MENU 3-2-2 測定結果の送信 Wildcards ■ Wildcards can be used for ID search. Wildcards have two types of characters, “? ” and “*”, indicating a single or any number of characters. “ ? ” indicates a single character. ●…
  • Page 81: Parameter Menu

    3-3 Parameter Menu Chapter 3 SUB MENU 3-3-1 Print parameters Print the present parameter settings by each measurement item for checking. Display setting screen. Press [2] key on the MAIN MENU. ● 1 . M e a s u r e 2 . S u b m e n u SUB MENU 1/2 is displayed.

  • Page 82
    測定結果のみかた Chapter 3 SUB MENU Parameters can be printed by executing “Printing parameters”. The formats are slightly different between Single Reagent Strips and Multi Reagent Strips. (1) Printing of single parameters Version. 1・ Printed date and time: The date and time when “printing of parameters” is executed. 2・…
  • Page 83
    測定結果のみかた Chapter 3 SUB MENU 2-6-2 キャリブレータ校正結果 (2) Printing of multi parameters From the left, name of Multi Reagent Strip, item symbol, lot number and expiry 1・ date. Calibration type 2・ CARD means calibration by magnetic card and Cal. means calibration by calibrator kit.
  • Page 84: Enter Parameters

    装置の設置 Chapter 3 SUB MENU 3-3-2 Enter parameters If [STOP] key is pressed during Set parameters (measurement conditions) for temperature, coefficient entry, the entry is canceled and the cursor returns to the of correlation correction, normal value range, and calibration previous “[ ]”…

  • Page 85
    パラメータメニュー Chapter 3 SUB MENU 3-3-1 パラメータ設定内容の印字 Select measurement items. Besides [hyphen (-)] key, [0], [2], Select measurement items for ● [ Sー01:G G T ] [4], [5], [6], and [8] keys can be parameters by pressing [hyphen (-)] used in selecting items. (…
  • Page 86
    装置の設置 Chapter 3 SUB MENU 3-3-2 パラメータ設定 Press [ENTER] key. ● C a l . T y p e[ C A R D ] The cursor moves to the entry Sー01 <2> position for the number of times of calibration. Any number from 2 to 6 (times) can be entered.
  • Page 87
    装置の設置 Chapter 3 SUB MENU 3-3-2 パラメータ設定 Store all the parameter settings. Press [START] or [STOP]. ● W r i t i n g ・・・ / If [START] key is pressed, the parameter settings are saved and the Measurement item selecting screen is restored.
  • Page 88: Initialize Parameters

    パラメータメニュー Chapter 3 SUB MENU 3-3-3 Initialize parameters Initialize parameter settings to the factory setting values. Display setting screen. Press [2] key on the MAIN MENU. ● 1 . M e a s u r e 2 . S u b m e n u SUB MENU 1/2 is displayed.

  • Page 89
    装置の設置 Chapter 3 SUB MENU 3-3-2 パラメータ設定 End Initialization. If the initialization is finished, press ● 1 . M e a s u r e 2 . S u b m e n u [STOP] key three times to return to 3…
  • Page 90: Enter Sample Type

    パラメータメニュー Chapter 3 SUB MENU 3-3-4 Enter sample type Enter sample types to measure. The entered types are printed out with the measurement results. Maximum 5 different types can be set. Display the screen. Press [2] key on the MAIN MENU. ●…

  • Page 91
    パラメータメニュー Chapter 3 SUB MENU 3-3-4 Input of identifiers for sample classifications Setting the sample type to use on the basic setting. In the end, set the sample type to ● REF value base type use on the basic setting. <Man…
  • Page 92: Copy Standard Range Setting

    パラメータメニュー Chapter 3 SUB MENU 3-3-5 Copy standard range setting The standard range setting for a certain sample type can be copied in all items to one for another sample type. Copy is available only when two or more sample types are input. Display setting screen.

  • Page 93
    パラメータメニュー Chapter 3 SUB MENU 3-3-4 Input of identifiers for sample classifications Selecting the standard range of the duplicate screen. Press [-] key to select sample type ● PRES SPC>>Next SPC on the duplicate screen for which [ Man ] […
  • Page 94: Maintenance

    3-4 Maintenance Chapter 3 SUB MENU 3-3-3 パラメータの初期化 Proper maintenance is required to maintain the accuracy of the analyzer. Select “Maintenance” on the SUB MENU to perform either daily maintenance or periodical maintenance. Preparation for the maintenance. Press [2] key on the MAIN MENU. ●…

  • Page 95: Mode Menu

    3-5 Mode Menu Chapter 3 SUB MENU 3-3-3 パラメータの初期化 The Survey Mode is canceled Switch to the Survey Mode. When measurement is performed in the automatically when the power is Survey Mode, raw data as well as normal measurement results can be turned off.

  • Page 96: Built-In Clock Adjustment

    3-6 Built-in Clock Adjustment Chapter 3 SUB MENU 3-3-2 パラメータ設定 Set the Date and time of the built-in clock. Once the date and time are set, resetting is not necessary. However, resetting may be necessary in a long time of use. Display setting screen.

  • Page 97
    3 — 26 ● SP — 4430…
  • Page 99: Outline Of Maintenance

    Chapter 4 MAINTENANCE Proper maintenance is required to maintain satisfactory measurement. This chapter describes maintenance and replacement of consumables. 4-1 Outline of Maintenance 4-1-1 Frequency of Maintenance 4-2 Daily Maintenance 4-2-1 Cleaning the Reagent Table 4-2-2 Cleaning the Tip Waste Case and Protective Cover 4-3 Periodic Maintenance 4-3-1 Replacement of Thermal Printer Paper 4-3-2 Cleaning the Optical Window…

  • Page 100: Outline Of Maintenance

    4-1 Outline of Maintenance Chapter 4 MAINTENANCE 4-1-1 Frequency of Maintenance The following table shows the parts requiring maintenance and the frequency of maintenance. Perform daily or periodical maintenance according to the table. Cleaning part Frequency Page ★ Cleaning of Reagent Table Daily 4-03 Cleaning of Tip Waste Case…

  • Page 101: Daily Maintenance

    4-2 Daily Maintenance Chapter 4 MAINTENANCE 4-2-1 Cleaning the Reagent Table With numerous measurements, residue of sample and Reagent Strips stay on the Reagent Table. When they are adhered to a new Reagent Strip, correct measurement results may not be obtained or Reagent Strips may be jammed inside.

  • Page 102
    お手入れの概要 Chapter 4 MAINTENANCE 4-1-1 お手入れの目安 Moisten a cotton swab with distilled ● water and wipe out stains or dust adhered to the Reagent Table. Clean the groove well carefully, not to break the lugs on the tip of the groove.
  • Page 103
    毎日のお手入れ Chapter 4 MAINTENANCE Moisten a cotton swab with distilled ● water and wipe out stains or dust adhered to the Rubber Plate. If the Rubber Plate is wet, use a dry ● cotton swab. Attach the top cover by sliding to ●…
  • Page 104: Cleaning The Tip Waste Case And Protective Cover

    毎日のお手入れ Chapter 4 MAINTENANCE 4-2-2 Cleaning the Tip Waste Case and Protective Cover The Tip Waste Case holds used tips. (※) Its capacity is for 5 measurements .Discard used tips and wash the Tip Waste Case. Clean the Protective Cover if necessary. A message suggesting the disposal of used tips is displayed every 5 (※)…

  • Page 105
    お手入れの概要 Chapter 4 MAINTENANCE 4-1-1 お手入れの目安 Disinfect and wash the Tip Waste Case. Disinfect the Tip Waste Case with ● alcohol and wash it with tap water. Wipe it dry with a cloth. ● Reattach the Tip Waste Case. Put the Tip Waste Case into the ●…
  • Page 106: Periodic Maintenance

    4-3 Periodic Maintenance Chapter 4 MAINTENANCE 4-3-1 Replacement of Thermal Printer Paper When the printer paper runs out, a red line appears on both sides of the printer paper. If the line is seen, replace it with a new roll. Approximately 500 measurements can be printed out on one roll of paper.

  • Page 107
    お手入れの概要 Chapter 4 MAINTENANCE 4-1-1 お手入れの目安 Prepare new printer paper. Cut off a single turn of the paper of ● new roll. Cut the top end straight to avoid a paper jam. Set new printer paper. Place a new roll of paper in the ●…
  • Page 108: Cleaning The Optical Window

    定期的なお手入れ Chapter 4 MAINTENANCE 4-3-2 Cleaning the Optical Window If dust is adhered to the Optical Window, correct measurement result may not be obtained due to improper detection of reflect on light caused by fluctuation of wavelength. Clean the Optical Window once every 2,000 measurements.

  • Page 109
    お手入れの概要 Chapter 4 MAINTENANCE 4-1-1 お手入れの目安 Turn over the analyzer carefully so ● that the right side (to which small legs are attached) faces down. Do NOT hold the Table Cover while turning the analyzer. Do NOT damage the connecting NOTE part of the Table Cover.
  • Page 110
    定期的なお手入れ Chapter 4 MAINTENANCE 4-3-2 測光窓の清掃 Restore the analyzer. Reattach the rubber caps firmly to ● the bottom of the analyzer. Carefully turn over the analyzer to ● the original position, without holding the Table Cover. NOTE Do NOT damage the connecting part of the Table Cover.
  • Page 111: Cleaning The Nozzle

    お手入れの概要 Chapter 4 MAINTENANCE 4-3-3 Cleaning the Nozzle With numerous measurements, sample blood residue may stay inside the Nozzle and cause clogging. Clean the Nozzle once every 10,000 measurements or once a year. Nozzle cleaning wire, Tweezers, Requirements Tissue paper and Protective gloves WARNING Wear protective gloves to prevent exposure to pathogenic microbes.

  • Page 112
    定期的なお手入れ Chapter 4 MAINTENANCE 4-3-2 測光窓の清掃 Remove the side cover. ● Remove the Nozzle Tube. Remove the Nozzle Tube from the ● Nozzle using tweezers. NOTE Do NOT scratch or damage the Nozzle Tube. Clean the Nozzle. Insert the nozzle cleaning wire into ●…
  • Page 113
    お手入れの概要 Chapter 4 MAINTENANCE 4-3-3 ノズルの清掃 Insert the Nozzle Tube. Remove the nozzle cleaning wire ● from the Nozzle. Pinch the Nozzle Tube with tweezers ● and insert it into the tube joint. Reattach the covers. Put the side cover to the analyzer. ●…
  • Page 114: Replacement Of Nozzle

    定期的なお手入れ Chapter 4 MAINTENANCE 4-3-4 Replacement of Nozzle With numerous measurements, the O-ring attached to the Nozzle becomes deteriorated. Requirements Nozzle, Nozzle replacement tool and Protective gloves WARNING Wear protective gloves to prevent exposure to pathogenic microbes. CAUTION Discard used samples, tips and protective gloves separately from general waste according to local regulations on biohazardous waste.

  • Page 115
    お手入れの概要 Chapter 4 MAINTENANCE 4-3-3 ノズルの清掃 Remove the side cover. ● Remove the Tip Waste Case. Loosen the fixing screws and ● remove the tip ejector. Remove the Tip Waste Case. ● Replace the Nozzle. Fasten the upper part using the ●…
  • Page 116
    定期的なお手入れ Chapter 4 MAINTENANCE 4-3-4 ノズルの交換 Reattach the Tip Waste Case. Put the Tip Waste Case by tightening ● the screws. Reattach the covers. Put the side cover to the analyzer. ● Put the top cover to the analyzer by ●…
  • Page 118
    Chapter 5 TROUBLESHOOTING When there is a failure in operation or a problem with the analyzer, an error or trouble occurs. This chapter describes, the kinds of errors / troubles, causes and remedy. 5-1 Error Messages 5-2 Trouble Messages 5-3 After sales service SP — 4430 ● 5 — 01…
  • Page 119: Error Messages

    Error Messages Chapter 5 TROUBLESHOOTING 4-1-1 お手入れの目安 When an error occurs, an alarm sounds and the error message appears ●The following errors are not on the screen. printed when the magnetic The error message is automatically printed in order to preserve the card is inserted.

  • Page 120
    毎日のお手入れ Chapter 5 TROUBLESHOOTING 4-2-1 試験片テーブルの清掃 Description and error message Conditions and causes Remedy When calibrating by calibrator kit, a Set the Reagent Strip correctly. ● ● many Reagent Strip is set. Return to the menu because the ● The bar code was not correctly read changing item is not allowed during ●…
  • Page 121
    毎日のお手入れ Chapter 5 TROUBLESHOOTING 4-2-1 試験片テーブルのお手入れ Description and error message Conditions and causes Remedy A sample vessel is misplaced. Set the sample correctly. ● ● There is no sample or an insufficient Confirm the amount of the sample. ● ● amount of sample.
  • Page 122
    お手入れの概要 Chapter 5 TROUBLESHOOTING 4-1-1 お手入れの目安 Description and error message Conditions and causes Remedy The clock is not correctly adjusted. Readjust the date and time. ● ● The power was not connected for a long (see “3-6 Built-in Clock Adjustment”) ●…
  • Page 123
    毎日のお手入れ Chapter 5 TROUBLESHOOTING 4-2-2 チップ廃却ケースの清掃 Description and error message Conditions and causes Remedy When sampling, air bubbles or fibrin Remove air bubbles or fibrin. ● ● was aspirated. Confirm the sample amount. ● When sampling, the sample amount ● E37 Sampling error…
  • Page 124: Trouble Messages

    5-2 Trouble Messages Chapter 5 TROUBLESHOOTING 4-1-1 お手入れの目安 When a trouble occurs, an alarm sounds, and a trouble message is displayed. The trouble message is automatically printed in order to preserve the trouble record. When a trouble occurs during measurement, perform the measurement again. It may IMPORTANT have influenced the result obtained before the trouble occurred.

  • Page 125
    トラブルメッセージ Chapter 5 TROUBLESHOOTING 4-2-2 チップ廃却ケースの清掃 Description and trouble message Conditions and causes Remedy Trouble occurred in the pump-driving Turn on the power again. ● ● mechanism. If the same trouble occurs, contact ● your distributor. T09 Pump ■■■■■■■■■■■■OK (ENTER) The optimal gain could not be set. Turn on the power again.
  • Page 126
    お手入れの概要 Chapter 5 TROUBLESHOOTING 4-1-1 お手入れの目安 Description and trouble message Conditions and causes Remedy Failure to discard tip. Turn off the power, and remove the ● ● tip from the nozzle. T21 Tip ejecting ■■■■■■■■■■■■OK (ENTER) White plate or black plate is dirty. Clean the white plate or black plate.
  • Page 127
    トラブルメッセージ Chapter 5 TROUBLESHOOTING 4-2-2 チップ廃却ケースの清掃 Description and trouble message Conditions and causes Remedy Trouble with parameter memory storage. If the same trouble occurs, contact ● ● Power was cut while writing in memory your distributor. ● or deleting. T93 Memory:parameter (S-03■■■■■■)■OK(ENTER)…
  • Page 128: After Sales Service

    5-3 After sales service Chapter 5 TROUBLESHOOTING 4-1-1 お手入れの目安 A warranty is included in this packaging box. This warranty is necessary when the analyzer ■Warranty : requires repair. After filling in the necessary items and confirming the described contents, keep the warranty in a safe place. ■Repairs : When the analyzer does not function Contact your distributor.

  • Page 129
    Chapter 5 TROUBLESHOOTING 5-12 ● SP — 4430…
  • Page 131
    Chapter 6 APPENDIX 6-1 Transmission Specifications 6-1-1 Protocol 6-1-2 Format 6-1-3 Format for measurement result (format 1) 6-1-4 Format for measurement result (format 2) SP — 4430 ● 6 — 01…
  • Page 132: Transmission Specifications

    6-1 Transmission Specifications Chapter 6 APPENDIX 6-1-1 Protocol Transmission format Start-stop system (asynchronous) unidirectional transmission by serial transmission format (in compliance with JIS X5101). Data format One character consists of the following 11 bits. Start bit: 1 bit Data bit: 7 bits (ASCII code) Parity bit: 1 bit (even) Stop bit: 2 bits Baud rate…

  • Page 133: Format

    毎日のお手入れ Chapter 6 APPENDIX 6-1-2 Format Block structure is regular. One block consists of start, data and end. This is illustrated below in the following descriptions. Start Data Start (S) ● Start of each block is < STX >. Start of block is expressed as S in the following illustrations. Data ●…

  • Page 134: Format For Measurement Results (Format 1)

    Chapter 6 APPENDIX 6-1-3 Format for measurement results (format 1) The measurement result (format 1) is the same as the “regular format” in SP-4410 or SP-4420. The receiving program designed to receive the measurement results of SP-4420 (regular format) can normally receive the measurement results (format 1) of SP-4430.

  • Page 135
    お手入れの概要 Chapter 6 APPENDIX 4-1-1 お手入れの目安 ● Format of Multi Reagent Strip measurement results M U L T I : ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■…
  • Page 136
    毎日のお手入れ Chapter 6 APPENDIX ● Format of Single Reagent Strip measurement results ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■…
  • Page 137
    お手入れの概要 Chapter 6 APPENDIX 4-1-1 お手入れの目安 ● Data errors In the Multi Reagent Strip and Single Reagent Strip measurement results, data parts for each item (22 bytes including end CR/LF) are output as follows when measurement errors occur. ●Error in range or prozone (OVER) 028 029 036 037 ■…
  • Page 138
    Chapter 6 APPENDIX 6-1-4 Format of measurement results (format 2) The measurement result (format 2) is the same as the “extended format” in SP-4410 or SP-4420. The receiving program designed to receive the measurement results of SP-4420 (extended format) can normally receive the measurement results of SP-4430 (format 2). ■…
  • Page 139
    83-00139F…

83-00140F

Printed in Japan

DOC.No.OM-28573 REV.6 03/02

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Premise

This manual contains important information on the functions of the SPOTCHEM EZ SP-4430

Distributed by : A.Menarini I.F.R. srl

Diagnostics Division

This manual is issued by: ARKRAY, Inc.

Read carefully prior to starting up the unit.

It is recommended to retain this manual for future use.

This product conforms to European Directive 98/79/EC.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Introduction

The SP-4430 is an automated biochemical analyzer using a dry chemistry system.

This manual contains outlines, instructions for the operation, maintenance and troubleshooting for the SPOTCHEM SP-4430. Before operating the SP-4430, read this manual carefully.

It is recommended to retain this manual for future use.

Always be careful when handling blood samples or waste reagent strips. Incorrect or imprecise procedures may result in exposure to pathogenic microbes.

This analyzer must only be operated by those trained in proper procedures for clinical testing and handling of hazardous waste. Anyone who operates the analyzer for the first time must be assisted by a trained person.

Never touch the Reagent Table, Centrifuge-equipped Multi Rack, or other places where sample residue may collect with bare hands. When performing maintenance, always wear protective gloves to prevent exposure to pathogenic microbes.

Separate used samples, tips and protective gloves from general waste and discard them according to local regulation on biohazardous waste.

All rights reserved. Reproduction of this manual is forbidden.

The contents of this manual are subject to change without further notice.

Although we take all possible measures to ensure the contents of this manual, please notify your distributorwhen you have questions, or find errors or omission.

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Caution Marks

Pay careful attention to these caution marks in this Operating Manual.

Regarding accident resulting in injury or death.

To prevent infection of yourself or others from pathogenic microbes, follow the instructions given herein.

To prevent injuries to yourself or others, or material damages, follow the instructions described.

Regarding damage and performance of products.

Failure to follow the instructions may lead to incorrect results.

To prevent damage to the unit and assist you in making best use of the capabilities of the analyzer, additional explanations and notes are provided herein.

Reference information on operation, additional explanations and related functions are provided herein.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

2-5

Calibration

2-28

2-5-1

Overview of calibration

2-28

2-5-2

Calibration by magnetic card

2-29

2-5-3 Calibration by calibrator kit

2-31

2-6

Interpreting measurement results

2-39

2-6-1

Printing normal measurement results

2-39

2-6-2

Printing calibration results

2-42

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 1

INTRODUCTION

The SP-4430 is an automated biochemical analyzer using a dry chemistry system. This chapter contains the information on the functions and measurement principles of the SP-4430.

1-1 Outline

1-1-1 Features

1-1-2 Measurement principles

1-1-3 Specifications

1-2 Shipping Carton

1-2-1 Analyzer

1-2-2 Accessories

1-3 Parts Description and Function

1-3-1 Front of the analyzer

1-3-2 Operator panel

1-3-3 Rear of the analyzer

1-4 Setting up the Analyzer

1-4-1 Cautions

1-4-2 Setting up the analyzer

1-4-3 First operation after setting up

1-4-4 Notes at transportation

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

1-1 Outline

Chapter 1 INTRODUCTION

1-1-1 Features

■Compact and lightweight design

The footprint of the SP-4430 is as small as that of a notebook computer. It weighs only 5 kg and can be easily carried by one person. The small unit contains various components such as a display, printer, 1- sample centrifuge and automatic tip disposal mechanism. No water supply or drainage system is required, therefore it is suitable for examinations in schools or bedside in hospitals. Power consumption is greatly reduced.

■Energy-saving design

■Automatic Sampling

■Built-in centrifuge for 1 sample

■Simple calibration using magnetic cards

■Sample Control by Bar code

■Online Network through two-way Transmission

Power consumption per 1 hour is approximately 50 W on average (AC 100V, 60Hz when continuous measurements are conducted).

The analyzer automatically absorbs samples set on the Centrifuge-equipped Multi Rack and drops them on reagent strips. Manual operation is not necessary, so that the sampling amount becomes constant and stable measurement results can be obtained.

The analyzer has a built-in centrifuge for 1 sample. By placing the whole blood sample in the centrifuge tube and setting it on the Centrifuge-equipped Multi Rack, the operation, centrifugation suction sampling is conducted automatically. This saves centrifugation of whole blood samples before measurement.

Calibration using magnetic cardsReagent Card are provided with the Reagent Stripsis possible, without using the SPOTCHEM Calibrator Kit. By inserting magnetic cards into the magnetic card reader, differences between reagent strip lot number and daily deviations are automatically calibrated.

Optional Hand-held Bar Code Reader is available. By reading the bar code of each sample, the bar code is automatically allocated as patient ID.

With the optional Data Management Software and external devices, remote control of measurement results and remote update of parameter settings are possible.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

1-1 SP-4430

Chapter 1 INTRODUCTION

1-1-2 Measurement principles

The light emitted from LED becomes monochromatic light of a certain wavelength after passing through an optical fiber (Five different types of optical filters are provided, and the optimum wavelength is selected for each test item). Monochromatic light is separated and transmitted to the photometry section of each channel by ten optical fibers.

In each photometry section, the monochromatic light transmitted by the optical fibers is irradiated on the reagent fields, which has a color reaction after sampling. Its reflected light is read with 2 photodiodes, and the system calculates measurement results by end-point assay (EPA) or reaction-rate assay (RRA).

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

1-1 SP-4430

Chapter 1 INTRODUCTION

1-1-3 Specifications

Sample

Serum, Plasma, Whole blood (Only Hb)

Measurement item

General biochemical measurement items, 22 items

Measurement wavelength

5 wavelengths (405, 550, 575, 610 and 820 nm)

Measurement principle

Optical measurement of reflection intensity of

reagent color reaction

Measurement Range

Set for sach measurement item

Reagent Strip

SPOTCHEM Reagent Strip

Processing speed

63 samples per hour

Minimum sample volume

6 × the number of measurement items + 38μL:

serum, plasma

Sample consumption

4 6μL(per 1 measurement item): serum, plasma

Sample container

Whole blood: exclusive whole blood sample tube

(Orange cap)

Centrifuge tube

Serum, plasma: exclusive serum sample tube

(Blue cap)

Simultaneous measurement

3 Single Reagent Strips or 1 Multi Reagent Strip

Maximum of 9 test items of continuous

measurement using Single Reagent Strips and a

Multi Reagent Strip is available.

Reagent reaction temperature

37

Light source

LED and Interference filter

Calibration method

Calibration by magnetic card (Reagent Card)

Calibration by Calibrator Kit

Data storage volume

100 tests

Display

20 digits

2 lines LCD

Built-in Printer

36-character thermal printer (58 mm width)

External Output

RS-232C interface

Transmission Method

Single or Two-way Transmission

Transmission Rate

300, 600, 1200, 2400, 4800, 9600bps

Measurement conditions

Temperature: 10 30

Humidity: 20

80 % (Non-condensing)

Power supply

AC100

120V, 220 240V

50/60Hz AC adapter method

Power input

Max. 100VA

Dimensions

338 mm(W)

203 mm(D) 167 mm(H)

Weight

Approximately 5.4 kg

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

1-2 Shipping Carton

Chapter 1 INTRODUCTION

1-2-1 Analyzer

Open the package and confirm that all items are included.

NO.

Item

Description

Qty.

1

Analyzer

SP-4430

1

2

Power cord

1

3

AC adapter

1

4

Accessory case

1

5

Operating manual

This book

1

6

Warranty card

Warranty card with case, Delivery label

1set

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Open the package and confirm that all items are included.

NO.

Item

Description

Qty.

1

Tip

10 tips

1 Set

2

Cleaning wire

Nozzle cleaning

1 Set

3

Cleaner set

Brush, Cotton swab

1 Set

4

Nozzle set (EZ)

Nozzle with O-ring

1 Set

5

Wrench set for nozzle replacement

2 Wrench, Adapter

1 Set

6

Centrifuge tube

10 pieces

1 Set

7

Tip waste case

2 pieces

1 Set

8

Protective cover

2 pieces

1 Set

9

Thermal printer paper

width 58 mm

1 Set

SPOTCHEM EZ SP-4430 OPERATING MANUAL DOC.No.OM-28573 REV.6 03/02

1-3 Parts Description and Function Chapter 1 INTRODUCTION

1-3-1 Front of the analyzer

NO.

Item

1

Reagent table

For setting the Reagent Strip. The temperature is kept at 37 for the

reaction conditions of Reagent.

2

Centrifuge-equipped Multi Rack

For setting the samples and tips. The centrifuge for 1 sample is built in.

3

Port

For setting the samples.

4

Centrifuge

For setting the whole blood samples that have not been centrifuged.

5

Tip hole

For setting tips.

6

Tip waste case

The container for used tips automatically discarded after sampling.

Filled with 5 measurements.

7

Black & white plates

The standard reflection plates used for measurement of reflectivity.

8

Table cover

Prevents entering the external light.

Slide forward the Reagent Table and Centrifuge-equipped Multi Rack to

open.

9

Operator panel

For starting or stopping measurement and entering IDs.

10

Display

Displays information such as operating state of the unit and error messages.

11

Magnetic card reader

For inserting magnetic cards: Reagent Card and Calibration Cards.

12

Built-in printer

Thermal-type printer. Prints out measurement results and setting conditions.

13

Maintenance cover

Protects the Nozzle Driving Units. Also prevents the operator from contacting

the nozzle during measurement. There are top and side covers.

14

Protective cover

Prevents the operator from contacting the Centrifuge Tube.

Also protects spilling of samples.

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Item

Function

START

Starts measurement. SelectYesfrom the Yes/No option.

STOP

Stops measurement. SelectNofrom the Yes/No option.

Cancels entry.

FEED

Feeds the built-in printer with paper while pressed.

MENU

Switches the page on each menu display.

0~9 (ten key)

Selects the menu number. Enter numerical values and ID.

-/.(hyphen/period)

Selects item, move cursor , switch the page on the display,

and enter minus sign and decimal point.

ENTER

Determines the entry.

Check message to proceed to the operation.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

1-1 SP-4430

Chapter 1 INTRODUCTION

1-3-3 Rear of the analyzer

NO.

Item

Function

1

Power switch

For turning on and off the power supply to the analyzer.

2

Power input terminal

For the supplied AC adapter.

3

Cooling Fan

For Ventilating heated air in the unit to prevent overheating.

4

COM.

For the cable of the optional external device.

5

B.C.R.

Jack for the bar-code reader (option).

6

CONT.

For adjusting screen contrast. Turning clockwise to make

the screen darker, counterclockwise makes it brighter.

7

Paper holder

For setting thermal printer paper.

DOC.No.OM-28573 REV.6 03/02 SPOTCHEM EZ SP-4430 OPERATING MANUAL

1-4 Setting up the Analyzer

Chapter 1 INTRODUCTION

1-4-1 Cautions

Before setting up the analyzer, read the following notes and always take proper safety precautions.

Allow a space of 10 cm or more between the rear of the analyzer and the wall. Failure to do so may cause overheating. Excessive load on the cable connection may cause fire or incorrect measurement results.

To avoid electric shock and/or fire, use the attached power cord to connect with a power outlet. For details, contact your distributer.

Operate the analyzer with power of the correct voltage and frequency. Otherwise fire or damage may result.

Connect the power plug directly to an outlet, not via an extension cord or power tap. The power supply for the analyzer is 100 VA. Before turning on the power switch, make sure that the total input of devices connected to a receptacle of the same circuit doesn’t exceed 1500 VA (100V, 15A).

Do NOT unnecessarily disassemble or modify the analyzer. Such actions may invite danger of exposure to pathogenic microbes, and cause fire or damage.

Place the analyzer on a stable and level surface free of vibration. Failure to do so may damage the analyzer, correct measurement results may not be obtained, and injury may result. Do NOT place the analyzer where there is a risk of falling.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

1-1 SP-4430

Chapter 1 INTRODUCTION

Before setting up the analyzer, read the following notes and always take proper safety precautions.

Do NOT set up the analyzer where chemicals are stored nearby, or where corrosive gases or electrical noise are generated. They may damage the analyzer, and malfunctions and/or injury may occur. Correct measurement results may not be obtained.

Avoid exposing the analyzer to direct sunlight, condensation and wind. Otherwise, correct measurement results may not be obtained, and may cause deformation or malfunction of the analyzer.

To connect the analyzer to external devices, be sure to use proper cables to avoid electric shock and/or fire. For details, contact your distributor.

Place the analyzer in a room at temperatures between 10 and 30 with humidity between 20% and 80%. Otherwise, correct measurement results may not be obtained.

Make sure that the room is well ventilated when a large amount of carbon dioxide is generated (when gas stoves, oil heaters and instantaneous water heaters are used in the room). This is because the pH of the Reagent Strips, which use an alkaline reaction reagent, decreases under the influence of carbon dioxide and correct measurement data may not be obtained.

Be careful not to be put hands under the analyzer.

DOC.No.OM-28573 REV.6 03/02 SPOTCHEM EZ SP-4430 OPERATING MANUAL

1-4 Chapter 1 INTRODUCTION

1-4-2 Setting up the analyzer

The parts in the analyzer are secured firmly in order to prevent scratches and/or damage due to transportation. Remove the fixing tapes before setting up the analyzer. Read 1-4-1 “Cautions on installation” carefully before setting up the analyzer.

1. Remove the fixing tape.

Remove the tape fixing the table cover.

Open the maintenance cover and remove the tape fixing the nozzle.

Reattach the maintenance cover.

2. Connect the power cord.

Make sure that the power switch on the rear of the analyzer is OFF.

Connect the AC adapter to the receptacle on the rear panel of the analyzer, and plug the other end of the cord into the AC outlet.

3. Connecting an external device (if necessary)

When using an external device, connect the exclusive connecting

cable to COM on the rear panel of

the analyzer.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

1-1 SP-4430

Chapter 1 INTRODUCTION

1-4-3 First operation after setting up

For error correction or troubleshooting, “see Chapter

5 Troubleshooting”.

IMPORTANT

This section explains the way to load thermal printer paper, and set data and time after turning on the power switch.

1. Turn on the power.

Turn on the power switch on the rear panel of the analyzer.

“|” is displayed.

After the analyzer name and system version are displayed, warm-up starts.

S PO TCHEM

E Z

S P V X X X X

Wa r m i n g u p

About 10 minutes after (room

temperature at 25 ), warm-up is

. Me a s u r e . S u bme n u

. C a l i b r a t e

completed and the MAIN MENU is

displayed as shown at right.

2. Set the thermal printer paper.

Set the attached thermal printer paper (See 4-3-1 “Thermal printer

paper replacement”).

3. Confirm the time and date.

Set the date and time (see “3-6

Built-in Clock Adjustment”).

If the date is not set correctly, correct measurement results may not be

obtained and an error may occur.

4. Remove the fixing tape on the protective cover.

Press [1] on the MAIN MENU.

The Table Cover opens and the Reagent Table and Centrifugeequipped Multi Rack slide forward.

DOC.No.OM-28573 REV.6 03/02 SPOTCHEM EZ SP-4430 OPERATING MANUAL

1-4 Chapter 1 INTRODUCTION

1-4-4 Notes at transportation

If no key entry is made within 3 minutes after the Table Cover is opened, an alarm beeps and the Table Cover is closed.

If [STOP] key is pressed while the message “CANCEL” is displayed, the Table Cover remains open and the Standby screen is restored.

After the Table Cover is closed, a message shown at the right is displayed.

To restore the MAIN MENU, press [STOP] key.

When [ENTER] key is pressed, the Table Cover opens and the Standby screen is restored.

Remove the tape fixing the protective cover.

Set the maintenance cover and the tip waste case.

Press [STOP] to close the table cover.

5. Turn off the power.

When terminating the operation or setting, turn off the power after making sure that the MAIN MENU is displayed.

C o v e r

i s

c l o s i n g .

C o v e r

i s

c l o s i n g .

S t o p S TOP

B a c k

t o

MENU S TOP

B a c k

t o

ME A S EN T ER

. Me a s u r e . S u bme n u

. C a l i b r a t e

SPOTCHEM EZ SP-4430 OPERATING MANUAL DOC.No.OM-28573 REV.6 03/02

1-4 Chapter 1 INTRODUCTION

1-4-4 Notes at transportation

When carrying the analyzer in cars, use an exclusive carrying case. The carrying case is optional. Follow the following procedures to prevent exposure to pathogenic microbes and damage to the analyzer.

Wear protective gloves to prevent exposure to pathogenic microbes.

1. Inspect the inside of the analyzer.

Press [1] on the MAIN MENU. The Table Cover opens and the Reagent Table and Centrifuge-equipped Multi Rack slide forward.

The Standby screen is displayed.

Make sure that there are no used Reagent Strips or samples. If any, discard them.

When there are used tips in the Tip Waste Case, discard them.

Press [STOP] to return to the MAIN MENU and turn off the power switch.

2. Placing analyzer into carrying case.

Place the analyzer, AC adapter and power code in the carrying case.

. Me a s u r e . S u bme n u

. C a l i b r a t e

S t a n d b y

I D

I N FO

. Me a s u r e . S u bme n u. C a l i b r a t e

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 2

MEASUREMENT

The normal measurement and two types of calibration can be done with the SP-4430. This chapter describes the outline and operating procedures of each measurement.

2-1 Outline of Measurement

2-1-1 SPOTCHEM Reagent Strips

2-1-2 Normal measurement

2-1-3 Calibration

2-2 Cautions

2-2-1 Cautions

2-2-2 Handling samples

2-2-3 Handling Reagent Strips

2-2-4 Handling magnetic cards

2-2-5 Handling tips

2-2-6 Handling centrifuges

2-3 Preparation for Measurement

2-3-1 Preparation

2-3-2 Startup

2-3-3 Checks before measurement

2-3-4 Preparing samples

2-4 Measurement

2-4-1 Normal measurement

2-5 Calibration

2-5-1 Overview of calibration

2-5-2 Calibration by magnetic card

2-5-3 Calibration by calibrator kit

2-6 Interpreting measurement results

2-6-1 Printing normal measurement results

2-6-2 Printing calibration results

DOC.No.OM-28573 REV.6 03/02 SPOTCHEM EZ SP-4430 OPERATING MANUAL

2-1 Outline of Measurement Chapter 2 MEASUREMENT

2-1-1 SPOTCHEM Reagent Strips

SPOTCHEM Reagent Strips (optional) must be used with this analyzer. There are two types of SPOTCHEM Reagent Strips. Use the correct type of strips for the measurement purpose.

Multi Reagent Strip

A Multi Reagent Strip has 5~6 reagent fields for analyzing various items including normal screening, and liver or renal function screening.

Single Reagent Strip

A Single Reagent Strip has only one reagent field on it. Several different reagent items of Single Reagent Strips can be measured for one sample.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 2 MEASUREMENT

2-1-2 Normal measurement

In normal measurement, more than one item can be measured at a time for one sample. The Reagent Strip Table can hold three Single Reagent Strips and one Multi Reagent Strip, so that simultaneous measurement of up to nine items is available. The operating procedure of normal measurement is described as shown below.

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Chapter 2 MEASUREMENT

2-1-3 Calibration

This analyzer provides two types of calibration: Calibration by magnetic card (Reagent Card) and Calibration by Calibrator. Calibration must be performed for all fields of Reagent Strips (see “2-5-1 Outline of calibration” for details).

For calibration by Calibrator, the following two types of Calibrator Kit and Magnetic Card (Calibration Card) are used.

SPOTCHEM Calibrator Kit

For use with all items except Hemoglobin

SPOTCHEM Calibrator Hb Kit

For use with Hemoglobin

Each Calibrator Kit contains Low and High lyophilized serum and a Calibration Card. For detailed procedures, see the package insert of the Calibrator Kit.

Arkray Spotchem SP-4430 User manual

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 2 MEASUREMENT

The flowchart of calibration.

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

2-2 Cautions

2-2-1 Cautions

Chapter 2 MEASUREMENT

Before turning ON the power, reconfirm “1-4-1 Cautions” so that the analyzer is always operated under proper conditions.

If the analyzer is out of order, contact your distributor for repair. Do NOT attempt to repair or modify it by yourself. It may damage the analyzer or get injured.

The analyzer has a temperature control function in order to yield correct measurement results at the room temperature of 10~30 . For more accurate measurement, measurement at the room temperature of 20~25and at humidity of 30~60% is recommended.

Do NOT put a container with samples in it on the analyzer. If samples spill on the device, it may damage.

If there is anything wrong with the analyzer, or in case of odor or smoke, turn OFF the power immediately and unplug. Otherwise the analyzer may be damaged and cause fire.

Do NOT move the analyzer during measurement. Such movement may cause malfunction, resulting in incorrect measurement results.

Perform proper maintenance following the instruction to ensure high accuracy.

Do NOT place anything in front of the Table Cover. It automatically opens during operation.

Use SPOTCHEM Reagent Strips only for the analyzer. The other types of Reagent Strips are unacceptable.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 2 MEASUREMENT

2-2-2 Handling samples

Blood is used as a measurement sample with this analyzer. Be aware that blood may be contaminated with pathogenic microbes that may cause infectious disease. Exercise utmost care whenever handling blood. Incorrect or imprecise procedures may result in exposure to pathogenic microbes.

Handling samples slightly differs depending on test items. Follow the instructions on the package insert of SPOTCHEM Reagent Strips.

2-2-3 Handling Reagent Strips

Do NOT use expired Reagent Strips or deteriorated strips of which the reagent fields show any sign of discoloration or deformation, even if they are before the expiration date. Correct measurement results may not be obtained.

Take Reagent Strips out of the refrigerator 20 minutes before measurement and allow them to reach room temperature (10 to 30 ). Otherwise, correct measurement results may not be obtained.

Open a reagent strip package immediately before use. If the opened reagent strip is left for a while, it will absorb water from the air or gather dust, which may result in incorrect measurement results.

Do NOT touch the reagent field on the strip with fingers. Sebum on the reagent field may result in incorrect measurement results.

When a new reagent strip box is opened, perform calibration by magnetic card using the supplied Reagent Card (see 2-5-2).

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Chapter 2 MEASUREMENT

2-2-4 Handling magnetic cards

Do NOT place a magnetic card close to a magnetic object (magnet, TV set, etc.) Information on the magnetic card may become unreadable.

Do NOT scratch the magnetic surface (stripe). Information on the magnetic card may become unreadable.

Do NOT use a magnetic card (supplied with Reagent Strips or Calibrator) for SP-4420 or SP-4430 with other devices. The card may become jammed in the device.

2-2-5 Handling tips

Do NOT touch the pointed end of a Tip with bare hands. If it becomes soiled, correct sampling may not be possible, resulting in incorrect measurement results.

Do NOT reuse a tip. Water-repellent treatment is applied to each tip so that the sample can be pipetted correctly. If it is rinsed, the coating will come off and correct measurement results may not be obtained.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 2 MEASUREMENT

2-2-6 Handling centrifuges

Pipette 250 to 270 whole blood into the centrifuge tube. Insufficient sample volume may result in incorrect measurement results. Excessive volume may cause blood to stain the cover.

When it takes time before blood samples collected are brought into the centrifuge, the samples may begin to coagulate. The samples, which show heavy coagulation or deposition of fibrin, may not be subjected to measurement.

A syringe containing heparin can be used to prevent coagulation when it takes time before the samples are brought into the centrifuge, although the machinecontains heparin.

Heparin is contained as an anticoagulant. Close the cap tightly after setting a sample, and invert it and use the sample after removing the cap.

Remove air bubbles or skin on the surface of the sample, if any. Otherwise they may cause incorrect measurement results.

Wipe out the sample attached to the outside of centrifuge tube before placing it to the centrifuge, otherwise it may cause spattering the sample.

Samples with extremely high hematocrit levels may yield incorrect measurement results.

DOC.No.OM-28573 REV.6 03/02 SPOTCHEM EZ SP-4430 OPERATING MANUAL

2-3 Preparation for Measurement Chapter 2 MEASUREMENT

2-3-1 Preparation

Prepare necessary instruments as described in the following table before starting Normal Measurement, Calibration by Magnetic Card or Calibration by Calibrator.

IMPORTANT

Take the Reagent Strips out of the

and leave them at room temperature

30 ) for 20 minutes.

IMPORTANT Using Reagent Strips without allowing them to reach room temperature may

result in incorrect measurement results.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 2 MEASUREMENT

2-3-2 Startup

NOTE

To obtain the appropriate contrast on the screen, rotate the contrast adjustment knob (“CONT.”) on the rear of the analyzer.

Warm-up time depends on the room temperature.

The operating procedures from turning ON the power to the MAIN MENU are described below.

Before turning ON the power, make sure that the maintenance cover is in position. If light enters into the analyzer during initialization, an error will occur.

Before turning ON the power, make sure that nothing is placed in front of the Table Cover. The Table Cover opens during initialization. If there is an obstruction, a problem may occur.

1. Turn ON the power.

Turn ON the power at the rear of the analyzer. When the power is ON, “|” is displayed

The name of the analyzer and the system version (“VXXXX” shown at the right) are displayed and warm-up starts.

The screen appears during warm-up as shown at the right, and initialization and self-check of each function are performed.

About 10 minutes after (when at 25 ), warm-up is completed and the MAIN MENU is displayed.

S PO TCHEM E Z

S P V X X X X

. Me a s u r e . S u bme n u

. C a l i b r a t e

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Chapter 2 MEASUREMENT

2-3-3 Checks before measurement

Check each part before performing Normal Measurement or Calibration.

Wear protective gloves to prevent exposure to pathogenic microbes.

Discard used samples, tips and protective gloves separately from general waste according to the local regulations on biohazardous waste.

1. Thermal printer paper.

If a red line appears on both sides of the printer paper, replace it with a new roll (see“4-3-1 Thermal printer paper replacement”).

2. Reagent Table.

Check the Reagent Table and clean it if necessary (see “4-2-1 Cleaning the Reagent Table”).

3. Centrifuge-equipped Multi Rack.

Make sure that the Centrifugeequipped Multi Rack has no used samples in it. If any, discard them.

Install the Tip Waste Case.

4. Setting the parameters.

Set the parameters as necessary. If the same parameter settings are continuously used, resetting is not necessary. See “Chapter 3 SUB MENU” for details.

IMPORTANT To check the present parameter settings, print them out (see “3-3-1 Printing parameters”).

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 2 MEASUREMENT

2-3-4 Preparation of samples

The required volume of a serum or plasma sample is“6x (number of test items)+38 ”.For example, the required volume for 5 test items is : 6 x 5 + 38 = 68 .

A Serum or Whole blood Sample Tube has 2 lines marked on it for 100 and 500 .

IMPORTANT

The Whole blood Sample Tube contains heparin as an

anticoagulant.

Prepare samples as described below except for Hb and HDL-C, which require different procedures (see pages 2-16 and 2-17 for sample preparation for Hb and HDL-C). Test items may be added or changed in the future. Read the package insert of SPOTCHEM Reagent Strips carefully for details on handling samples.

Wear protective gloves to prevent exposure to pathogenic microbes.

Serum or plasma sample.

Prepare a Serum Sample Tube (blue cap).

Pipette serum or plasma. If measurement is not performed immediately, cap the tube to prevent contamination or evaporation.

Remove any air bubbles or skin on the surface of the sample.

They may cause incorrect measurement results.

Whole blood.

Prepare a Whole blood Sample Tube (orange cap).

Uncap the tube and pipette whole blood.

Cap the tube tightly.

Invert the sample 5 to 6 times (see

“Inverting a sample” on the next page).

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Chapter 2 MEASUREMENT

Even if the sample stays at the bottom after being inverted, heparin is well mixed with the sample. Do NOT shake the tube vigorously.

NOTE

IMPORTANT

Inverting a sample to mix

1 Gently invert a sample tube.

2 In 3 seconds, invert the sample again.

3 Wait for 3 seconds, and invert it again.

4 Repeat the above procedure several times.

To use a whole blood sample tube, prepare a centrifuge.

When using an optionally available

centrifuge , carefully read the

operating manual of the centrifuge.

Place the samples in the centrifuge taking rotor balance into consideration (see the block below).

Centrifuge the samples for 3 minutes or more. It is not necessary to remove the supernatant from the centrifuged samples. Uncap the tubes and place them directly in the sample rack.

Remove any air bubbles or skin on the surface of the sample.

They may cause incorrect measurement results.

Ensuring rotor balance

Place sample tubes symmetrically against the rotation axis as illustrated at right. For centrifugation of 1 or 5 samples, use a dummy tube filled with an equal amount of water to make a total of 2 or 6 sample tubes.

SPOTCHEM EZ SP-4430 OPERATING MANUAL

DOC.No.OM-28573 REV.6 03/02

Chapter 2 MEASUREMENT

A Centrifuge tube has 1 line marked on it for 250 .

Using a centrifuge

Prepare a centrifuge.

Uncap the tube and pipette 250 of heparinized whole blood or plain blood into the centrifuge tube.

The centrifuge tube includes Heparin as an anticoagulant.

Close the cap tightly after setting a sample, and invert it.

Use the sample after removing the cap.

Wipe out the sample attached to the outside of centrifuge tube before placing it to the centrifuge, otherwise it may spattering the sample.

Place samples in the centrifuge ( see“ 2-4-1

Normal measurement”).

IMPORTANT

Remove any air bubbles or skin on the surface of

the sample. They may cause incorrect measurement

results.

Wipe out the sample attached to the outside of

centrifuge tube before placing it to the centrifuge,

otherwise it may cause spattering the sample .

IMPORTANT

The samples may coagulate when it takes time

before the samples collected are brought into the

machine. Those showing heavy coagulation or

deposition of fibrin sometimes can not be measure.

A syringe containing heparin can be used to

prevent coagulation when it takes time before the

samples are brought into the centrifuge, although

the machine contains heparin.

DOC.No.OM-28573 REV.6 03/02

SPOTCHEM EZ SP-4430 OPERATING MANUAL

Chapter 2 MEASUREMENT

Measuring the Hb

When measuring Hb, dilute whole blood samples 6 times with distilled water.

Prepare a Whole blood Sample Tube (orange cap).

Uncap the tube and pipette the whole blood sample into the tube.

Cap the tube tight.

Invert the sample several times ( see “2-14

Inverting a sample”).

Prepare a Serum Sample Tube(blue cap). Pipette 200 of distilled water.

Pipette 40 of the prepared whole blood sample into the above Serum Sample Tube.

Cap the tube tight.

Invert the sample several times again.

IMPORTANT Remove any air bubbles or skin on the surface of the sample.

They may cause incorrect measurement results.

Loading…

Premise

01 ● SP — 4430

This manual contains important information on the functions of the

SPOTCHEM EZ SP-4430.

supplied by : ARKRAY, Inc.

This manual is issued by: ARKRAY, Inc.

Read carefully prior to starting up the unit.

It is recommended to retain this manual for future use.

This product conforms to Directive 89/336/EEC, as amended by

Directive 92/31/EEC, and to Directive 73/23/EEC. 89/392/EEC.

E

Detail Specifications:

1467/1467218-spotchem_ez_sp4430.pdf file (02 Feb 2023)

Accompanying Data:

Arkray SPOTCHEM EZ SP-4430 Measuring Instruments PDF Operating Manual (Updated: Thursday 2nd of February 2023 09:02:17 AM)

Rating: 4.8 (rated by 43 users)

Compatible devices: Lab 004, 740, AUTION ELEVEN AE-4022, SMARTRAIL-X835-MID, TSP 303 Plus, SPOTCHEM EZ, RS-232, Industrial Oval Gear Series.

Recommended Documentation:

Text Version of Operating Manual

(Ocr-Read Summary of Contents, UPD: 02 February 2023)

  • 33, f SP — 4430 ● 2 — 07 1-4 装置の設置 Chapter 2 MEASUREMENT 2-2-2 Handling samples Handling samples slightly differs depending on test items. Follow the instructions on the package insert of SPOTCHEM Reagent Strips. Wear protective gloves Blood is used as a measurement sample with this analyzer. Be aware that blood may be contaminated with pathogenic microbes that may cause inf…

  • 100, C Cleaning part Cleaning of Reagent Table Cleaning of Tip Waste Case Cleaning of Protective Cover Replacement of Thermal Printer Paper Cleaning of Optical Window Cleaning of Nozzle Replacement of Nozzle 4 — 02 ● SP — 4430 4-1 Outline of Maintenance Chapter 4 MAINTENANCE 4-1-1 Frequency of Maintenance The following table shows the parts requiring maintenance and the frequency of maintena…

  • 18, C SP — 4430 ● 1 — 09 1-1 SP — 4430の概要 Chapter 1 INTRODUCTION 1-3-3 Rear of the analyzer 1・Power switch 3・Cooling Fan 5・B.C.R.  7・Paper holder 2・Power input terminal  4・COM.  6・CONT. Function For turning on and off the power supply to the analyzer. For the supplied AC adapter. For Ventilating heated air in the unit to preven…

  • 2, E01 ● SP — 4430 Premise This manual contains important information on the functions of the SPOTCHEM EZ SP-4430. supplied by : ARKRAY, Inc. This manual is issued by: ARKRAY, Inc. Read carefully prior to starting up the unit. It is recommended to retain this manual for future use. This product conforms to Directive 89/336/EEC, as amended by Directive 92/31/EE…

  • 124, C SP — 4430 ● 5 — 07 5-2 Trouble Messages Chapter 5 TROUBLESHOOTING 4-1-1 お手入れの目安 Description and trouble message Conditions and causes Remedy The pressure does not increase since the rubber plate is dirty or deformed. The tube or pipe is clogged. The tube connecting the nozzle with the pump is disconnected. The tube connecting the pump with the pressure-detecting sensor …

  • 107, Arkray SPOTCHEM EZ SP-4430 b SP — 4430 ● 4 — 09 4-1 お手入れの概要 Chapter 4 MAINTENANCE 4-1-1 お手入れの目安 Prepare new printer paper.3. Cut off a single turn of the paper of new roll. Cut the top end straight to avoid a paper jam. ● Set new printer paper.4. Place a new roll of paper in the paper holder, with the paper end facing up. Insert the top of the paper end into the slot. The paper…

  • 80, Arkray SPOTCHEM EZ SP-4430 C SP — 4430 ● 3 — 09 Wildcards ■ Wildcards can be used for ID search. Wildcards have two types of characters, “? ” and “*”, indicating a single or any number of characters. ● “ ? ” indicates a single character. ● “*” indicates any number (including zero) of characters. For instance, if “????” key is entered, 4-character IDs are retrieved. If “A*”…

  • 36, C2 — 10 ● SP — 4430 2-3 Preparation for Measurement Chapter 2 MEASUREMENT 2-3-1 Prepare necessary instruments as described in the following table before starting Normal Measurement, Calibration by Magnetic Card or Calibration by Calibrator. Preparation IMPORTANT Take the Reagent Strips out of the refrigerator and leave them at room temperature (10∼ 30℃) for 20 minutes. Singl…

  • 105, Arkray SPOTCHEM EZ SP-4430 C SP — 4430 ● 4 — 07 4-1 お手入れの概要 Chapter 4 MAINTENANCE 4-1-1 お手入れの目安 Disinfect and wash the Tip Waste Case.3. Disinfect the Tip Waste Case with alcohol and wash it with tap water. Wipe it dry with a cloth. ● ● Reattach the Tip Waste Case.4. Put the Tip Waste Case into the Centrifuge-equipped Multi Rack. ● End cleaning.6. Turn ON the power. War…

  • 27, C Chapter 2 MEASUREMENT 2-1 Outline of Measurement 2-1-1 SPOTCHEMⅡ Reagent Strips 2-1-2 Normal measurement 2-1-3 Calibration 2-2 Cautions 2-2-1 Cautions 2-2-2 Handling samples 2-2-3 Handling Reagent Strips 2-2-4 Handling magnetic cards 2-2-5 Handling tips 2-2-6 Handling centrifuges 2-3 Preparation for Measurement 2-3-1 Preparation 2-3-2 Startup 2-3-3 Checks before measurement 2-3-4…

Recommended Instructions:

SMC7204BRA, AC 10 FT, 30PW6341/37B, GraffitiPad

  • SBS-HT-70CBEDIENUNGSANLEITUNG THERMO-HYGROMETERUSER MANUALTHERMO-HYGROMETER INSTRUKCJA OBSŁUGIREJESTRATOR WILGOTNOŚCI I TEMPERATURYNÁVOD K POUŽITÍMĚŘIČ TEPLOTY A VLHKOSTIMANUEL D´UTILISATIONTHERMOMÈTRE HYGROMÈTREISTRUZIONI PER L’USOTERMOMETRO IGROMETROMANUAL DE INSTRUCCIONESTERMÓMETRO HIGRÓMETRODE | EN | PL | CZ | FR | IT | ES …

    SBS-HT-70C 4

  • E-NLS Noise Level Sensor Installation Instruction The E-NLS noise level sensor that can be used to detect sounds and report occurrences through the user interface of an E-2D,-5D or -16D Enterprise Environment Monitoring System (SYSTEM). Connect the sensor to an available RJ45 Sensor port on a SYSTEM and configure the SYSTEM to report and send alerts when nois …

    ENVIROMUX Series 2

  • Operating instructionsvers 1.23 CW20160601SWEMA ABPepparv. 27123 56 FARSTA, SWEDENTel: +46 8 94 00 90 Fax: +46 8 93 44 93E-mail: [email protected] Web site: www.swema.comContents:1. Introduction…………………… 12. Technical data…………………. 13. Function……………………….. 24. Charging……………………….. 75. Driver Insta …

    SwemaFlow 125 12

Popular Right Now:
Operating Impressions, Questions and Answers:

Table of Contents for Arkray SPOTCHEM EZ SP-4430:

  • C SP — 4430 ● 04 1 2 3 1 Biohazard cautions. Wear protective gloves to prevent exposure to pathogenic microbes. Caution Stickers are adhered to the SP-4430 to prevent accidents as below. The descriptions are given on the following pages. 2 The centrifuge turn round at high speed. Keep your hands off when mesurement is in process. Biohazard cautions. Wear protective gloves to prevent exposure to pathogenic microbes. 3 Biohazar

  • C4 — 18 ● SP — 4430 End replacement.7. 4-3 定期的なお手入れ Chapter 4 MAINTENANCE 4-3-4 ノズルの交換 Reattach the covers.6. Put the side cover to the analyzer. Put the top cover to the analyzer by sliding it to the right. ● ● Reattach the Tip Waste Case.5. Put the Tip Waste Case by tightening the screws. ● Turn ON the power. Warm-up starts. The Reagent Table and Centrifuge-equipped Multi Rack slide back to the original p

  • e 1-1 Outline Chapter 1 INTRODUCTION ■Compact and lightweight design ■Energy-saving design ■Automatic Sampling ■Built-in centrifuge for 1 sample ■Simple calibration using magnetic cards ■Sample Control by Bar code ■Online Network through two-way Transmission 1-1-1 Features The footprint of the SP-4430 is as small as that of a notebook computer. It weighs only 5 kg and can be easily carrie

  • C SP — 4430 ● 3 — 15 Chapter 3 SUB MENU 1-4 装置の設置 3-3-2 パラメータ設定 Press [ENTER] key. The cursor moves to the entry position for the number of times of calibration. Any number from 2 to 6 (times) can be entered. Press [ENTER] key. The Standard range setting screen is displayed. ● ● Set the standard range.6. Enter the lower limit of the standard range with the numeric keys. Press [Enter] key. The cursor moves to the entry position for the upper limit value. Enter the

  • C SP — 4430 ● 3 — 17 3-3 パラメータメニュー Chapter 3 SUB MENU 3-3-3 Initialize parameters Display setting screen.1. Press [2] key on the MAIN MENU. SUB MENU 1/2 is displayed. Press [2] key. The PARAMETER MENU 1/2 is displayed. Press [3] key. The Password entering screen is displayed. Enter password “99”. ** is displayed. Press [ENTER] key. The Measurement item selecting screen is restored. ● ● �

  • b6 — 06 ● SP — 4430 4-2 毎日のお手入れ Chapter 6 APPENDIX ●Format of Single Reagent Strip measurement results 001 ■ 023 ■ 045 ■ 067 ■ 002 ■ 024 ■ 046 ■ 068 ■ 003 ■ 025 ■ 047 ■ 069 ■ 004 ■ 026 ■ 048 ■ 070 ■ 005 ■ 027 ■ 049 ■ 071 ■ 006 ■ 028  050 �

  • C4 — 04 ● SP — 4430 4-1 お手入れの概要 Chapter 4 MAINTENANCE 4-1-1 お手入れの目安 Moisten a cotton swab with distilled water and wipe out stains or dust adhered to the Reagent Table. Clean the groove well carefully, not to break the lugs on the tip of the groove. If the Reagent Table is wet, wipe using a dry cotton swab. If any fiber is remained on the Reagent Table, brush it off with the cleaning brush again. Do not spill water on the analyzer to avoid damage. Wipe out dust adhered to the table pins

  • f SP — 4430 ● 2 — 07 1-4 装置の設置 Chapter 2 MEASUREMENT 2-2-2 Handling samples Handling samples slightly differs depending on test items. Follow the instructions on the package insert of SPOTCHEM Reagent Strips. Wear protective gloves Blood is used as a measurement sample with this analyzer. Be aware that blood may be contaminated with pathogenic microbes that may cause infectious disease. Exercise utmost care whenever handling blood. Incorrect or imprecise procedures may result in exposure to pathogenic microbes. 2-2-3 Handling Reagent Strips ●●

  • C SP — 4430 ● 2 — 37 Set the tip.9. Set the tip by following procedure 5. ● When performing 1-point calibration for «Low» only, proceed to procedure 13. ▲ 2-5 校正 Chapter 2 MEASUREMENT 2-5-1 校正の概要 End the calibration. 13. When calibration is completed, the measurement results are printed. The Table Cover opens, and the Reagent Table and Centrifuge- equipped Multi Rack slide forward. The Calibration standby screen is restored. Remove the used Reagent Strips and samples, an

  • C2 — 06 ● SP — 4430 2-2 Cautions Chapter 2 MEASUREMENT 2-2-1 Cautions Before turning ON the power, reconfirm “1-4-1 Cautions” so that the analyzer is always operated under proper conditions. ● temperature:20 ∼ 25℃ humidity :30 ∼ 60% The analyzer has a temperature control function in order to yield correct measurement results at the room temperature of 10~30℃. For more accurate measurement, measurement at the room temperature of 20~25℃ and at humidity of 30~60

  • C2 — 02 ● SP — 4430 2-1 Outline of Measurement Chapter 2 MEASUREMENT 2-1-1 SPOTCHEMⅡ Reagent Strips (optional) must be used with this analyzer. There are two types of SPOTCHEM Reagent Strips. Use the correct type of strips for the measurement purpose. Multi Reagent Strip A Multi Reagent Strip has 5~6 reagent fields for analyzing various items including normal screening, and liver or renal function screening. ● Multi Reagent Strip ××××× Item nameBar code Single Reagent Str

  • C SP — 4430 ● 2 — 17 Prepare a Serum Sample Tube (blue cap). Pipette 100μL of serum or plasma. Gently pipette 100μL of HDL Separation Reagent into the tube. Cap the tube tight. Invert the sample more than 10 times following to “ 2-14. Inverting a sample” and leave the sample about 5 minutes. Centrifuge the samples for 3 minutes, taking rotor balance into consideration(see page 2- 14). Check if the supernatant is impure or there is any skin or not. Removing the supernatant is not necessary. Uncap the tube and

  • b Caution Marks Pay careful attention to these caution marks in this Operating Manual. ■Regarding accident resulting in injury or death. To prevent infection of yourself or others from pathogenic microbes, follow the instructions given herein. WARNING CAUTION To prevent injuries to yourself or others, or material damages, follow the instructions described. ■Regarding damage and performance of products. Failure to follow the instructions may lead to incorrect results. To prevent damage to the unit and assist you in making best use of the capabilities of the

  • C SP — 4430 ● 3 — 05 Chapter 3 SUB MENU 1-4 装置の設置 2-3-4 測定の準備 If designation of the range of measurement dates (the dates of starting and finishing measurements) is unnecessary, press [ENTER] key twice to proceed to the next setting screen. Be sure to enter correct dates to avoid inconsistencies. The 2-digit number of the “year” section of the date indicates the last 2 digits of year and is interpreted as follows: 00~89 -> 2000~2089 90~99 -> 1990~1999 See

Questions, Opinions and Exploitation Impressions:

You can ask a question, express your opinion or share our experience of Arkray SPOTCHEM EZ SP-4430 device using right now.

Понравилась статья? Поделить с друзьями:
  • Воздухоочиститель дайсон инструкция на русском языке
  • Холодильник liebherr premium biofresh инструкция на русском языке
  • Ацетилсалициловая кислота это аспирин или нет инструкция по применению
  • Солита чистящее средство инструкция по применению для котлов
  • Nissan x trail руководство по эксплуатации купить