Testo 645 руководство по эксплуатации

TESTO 645 Instruction Manual

  1. Manuals
  2. Brands
  3. TESTO Manuals
  4. Test Equipment
  5. 645
  6. Instruction manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

testo 445 · testo 645

Instruction manual

en

Testo-Direct

info@Testo-Direct.com

1.888.610.7664

www.

.com

loading

Related Manuals for TESTO 645

Summary of Contents for TESTO 645

  • Page 1
    445 · testo 645 Instruction manual Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…
  • Page 2: Table Of Contents

    Clearing memory contents ………………..21 According to the conformity certificate, the instruments fulfill 2014/30/EC guidelines. © 1999 Copyright Testo GmbH & Co. The software and software structure included in the product testo 445/645 are protected by copyright laws worldwide. Testo-Direct info@Testo-Direct.com 1.888.610.7664 www.

  • Page 3: Foreword

    Foreword Dear Customer You have made the right decision by choosing a measuring instrument from Testo. Every year, thousands of customers purchase our high-quality products. There are many good reasons for this. 1) We offer value for money. Reliable quality at the right price.

  • Page 4: Initial Operation

    Open the battery compartment at the back of the instrument. Put in block battery. Observe polarisation. Close battery compartment. Refer to “Power supply” Chapter for more information on alternative power supply, charge, battery quality, charging procedure. Observe instrument configuration when using rechargeable batteries. Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 5: First Measurement

    Some things can only be set via PC software (See Ordering data):  Location name (8 characters)   Log head (24 characters), e.g. your company name — is also printed when readings are printed. Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 6: Instrument Description

    (NTC) and velocity location (ball) RS232: PC connection Arrow buttons 12V: Mains connection move the cursor testo 645 Hold Max Min Socket 1: Mean – Temperature probes (T/C type K/J/S or NTC) Socket 2: – Combined probe for humidity…

  • Page 7: Display

    Shows capacity of battery and rechargeable battery. If the inner segment no longer appears (symbol flashes), the battery has to be changed or the rechargeable battery has to be recharged. The instrument switches itself off automatically after 1 minute. Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 8: Overview Of Controls

    Start and end Calculate mean Reset mean Parameter settings Selected The setting functions are activated in accordance with the reading testo testo parameter selected — flashing cursor in display (See 1.): °C System adjustment with EEPROM probe (4.1) Reset — system adjustment Humidity probe (0636.9740 and 0636.9715)/…

  • Page 9
    Print parameters AUTO Unit selection Program automatic saving • Select meas. °C°F cycles • Set m/sfpm [ft/min] (testo 445) measurement rate /min cfm [ft /min](testo 445) mbar Read memory in W [”H O] (testo 445) contents on display with…
  • Page 10: Current Measurement

    A distance 0f 0.5 m should not be exceeded in order to depending on probe(s) guarantee problem-free data transfer. Objects between connected. instrument and printer prevent the data from being transferred. Please observe notes in the instruction manual on the printer. Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 11: Measurement Functions 11-13 Freezing Readings

    34. 8 0 Hold Max/Min Mean Minimum readings since the M.05 N.0017 measurement began 1MIN Save measured 56. 6 2 value Print measured value 1MIN 20. 0 0 Hold Max/Min Mean See multi-point/timed mean calculation Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 12: Multi-Point Mean Calculation

    MEAN page 13) • Select parameter required. New mean calculation possible Update values. Calculate mean M.05 N.0017 MEAN 25. 2 1 Mean ME AN NE W Number of updated values • Save mean Print mean Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 13: Timed Mean Calculation

    55. 2 0 calculate mean. MEAN 00: 1 5  New mean calculation possible Mean M.05 N.0017 Timed duration of MEAN mean calculation 55. 2 1 MEAN NEW 00: 0 5 Save mean. Location  Print mean. Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 14: Location Selection

    The cross-section Use ESC to go Location can only be called up if to selection of volume flow is activated cross-section entry (testo 445 instrument) Ø AxB, m2, Funnel) Increase value 50. 0 Decrease value Select digit box Return to current…

  • Page 15: Parameter Settings

    — flashing cursor in display. M.05 N.0017 System adjustment (only possible in testo 645): testo 645 and probes with EEPROM (Part no. 0614.xxxx) can °C be adjusted to the “zero error adjustment point”. The probe 69. 6 characteristic saved in the measuring instrument is offset at a measurement point.

  • Page 16: Adjusting Humidity Probes

    N.0017 Displaying calculated humidity parameters. If Testo`s combination humidity probe (e.g. 0636.9740) is td°C connected, testo 445 and testo 645 display a calculated humidity parameter in addition to temperature and relative humidity. You 12.9 can select from the following parameters: td°C…

  • Page 17: Initialising Differential Pressure Probe (Testo 445)

    Initializing pressure probe / Activating velocity and volume flow The setting functions are activated in accordance with reading line selected — flashing cursor in display. M.05 N.0017 Differential pressure probe (testo 445 only): /h, in addition to mbar; Activating/deactivating m/s, m mbar initialisation 22.

  • Page 18: Memory Settings

    The symbol corresponding to the selection appears in the top line. 21. 8 0 The function is activated with Location Parameter settings (See Chapter 4). AUTO CLEAR CLEAR AUTO AUTO OU T MEMORY MEMORY MEMORY MEMORY Location Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 19: Manual Saving

    01: 3 series is accepted. M.05 N.0017 AUTO Cancel saving procedure. Location M.06 N.0001 m3/h Starts automatic saving again. An additional measurement series is 19665 AUTO added. Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 20: Reading Or Printing Memory Contents

    The N.0x counter flashes in the display. Location Select the measurement cycle by pressing activates a printout of the selected measurement cycle. M.05 N.0017 °C 79. 7 button enables you to go back one step. Location Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 21: Clearing Memory Contents

    21.2 44.7 22.3 M.001 Readings 21.1 44.6 22.3 31.08.1998 21.8 44.6 22.3 11:20:05 2:m/s 2:td°C 21.4 °C 1.00 2: 44.7 1.66 2: 22.1 °C 2.01 2: 0.55 1.52 2: 9.6 td°C Info : Info : Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 22: Instrument Configuration

    If the mains unit is connected and the charging procedure is confirmed:  Risk of explosion!!!! — Battery operation is set and rechargeable battery is put in: Rechargeable battery will not be charged. See page 24 Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 23: Setting Power Supply, With Rechargeable Battery Or Battery

    Correct battery recharging in the instrument via the mains unit can only be guaranteed if the above mentioned Testo rechargeable batteries are used. If other rechargeable battery types are used, recharging will have to be carried out by an external recharger.

  • Page 24: Setting Date/Time

    The flashing position in the display can be set 1998 = scroll/ = next position — Setting the time TIME 11: 2 5 The blinking position in the display can be set = scroll/ = next position — Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 25: Setting Parameters For Absolute Pressure And Density Compensation

    5. Pitot tube factor (testo 445) Pitot tube factor setting: The factor for Pitot tubes from Testo is constantly at 1.00 and Parameter does not need to be changed. If the Pitot tubes from other firms are used, ask your supplier for the Pitot tube factor and save in instrument.

  • Page 26: Unit Selection

    (m3/min) m3/s cfm (ft3/min) (testo 445 only) The selected parameter flashes.  Select  confirms the selection. Factory reset in W (”H2O) (testo 445 only) mbar The selected parameter flashes  Select  confirms the selection. M.05 N.0017 Factory reset °C…

  • Page 27: Velocity Measurement

    7. Velocity measurement Volume flow funnel (testo 445) A volume flow funnel is needed to measure the volume flow in a ventilation opening (grid or outlet with dual wall clearance). The opening of the funnel must be covered completely by the grid (max.

  • Page 28: Pitot Tube And Pressure Probe (Testo 445)

    The Pitot tube factor for testo Pitot tubes is at a constant 1.00 and does not need to be changed. If non-Testo Pitot tubes are used, please ask your supplier for the factor and save it.

  • Page 29
    CO/CO2 measurement (testo 445) CO measurement (testo 445) The connected probe 0632.1247 is initialised during the switch- on phase (“Count Down” runs in the display). For this reason, the measuring instrument should only be switched on in an atmosphere free of CO. Values will be too low otherwise in subsequent measurements.
  • Page 30: Co/Co Measurement

    (testo 445) The 0632.1240 probe measures concentrations from 0 to 1 vol % CO 2 . The unit can be switched between ppm and vol% as described in “Parameter settings” on page 16. The measuring principle is based on infrared absorption. The probe has a relatively high power consumption on account of its sensor.

  • Page 31: Measuring Absolute Pressure

    Elevation Air pressure Air pressure Air pressure Air pressure Elevation Elevation Elevation M.S.L (mbar) (mbar) (mbar) (mbar) M.S.L M.S.L M.S.L Table: Barometric elevation formula Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 32
    We therefore assume an absolute air pressure in the measurement medium of 1000 mbar. The table on page 31 is also saved in the Testo Comfort software. Absolute pressure can be calculated automatically using this software if the height above mean sea level, barometric pressure and differential pressure are entered.
  • Page 33: Error Messages

    Turn instrument on and off again If the error message returns, please contact your nearest Testo service point. Testo-Direct info@Testo-Direct.com 1.888.610.7664 www. .com…

  • Page 34: Technical Data

    Technical data — Instruments testo 645/445 testo 645/445 Type J (FeCu-Ni) General technical data Measuring range: -200 to +1000 °C Memory space: Up to 3000 readings ±0.3 °C or ±0.5 % of Accuracy* Power supply: Battery / rechargeable batt. reading at 22 °C:…

  • Page 35
    Technical data — Instruments testo 645/445 testo 645 Pressure units can be switched: Temperature measurement mbar, psi; in W(”H Pt100 Calculated velocity values -200 to +800 °C measuring range: density compensated: 0 to 100 m/s Accuracy* ±0.2 °C or ±0.1 % of Calculated volume flow values: at 22 °C:…
  • Page 36: Ordering Data

    System case, aluminium for instrument, probes and additional accessories. Clear layout of case contents by placing 0516 0410 probes in lid. Additional accessories 9 V rechargeable battery for testo 445, testo 645 measuring instruments 0515 0025 PT 100-probes Meas. range / Accuracy Connection cable Part no..

  • Page 37
    Ordering data Fühler für testo 645 / 445 NiCr-Ni probes Meas. range / Accuracy Connection Part no. sec. Quick-action surface probe with sprung thermocouple strip -200…+300°C, Short- Plug-in head 0604 0194 term up to +500 °C, Please order connection cable 0430.0143)
  • Page 38
    Ordering data Probes for testo 645 / 445 Humidity/temperature Measuring range System Connection Part no. probes accuracy* sec. cable Standard indoor air quality probe up to +70 °C 0 to 100 %RH, probe ±2 %rH (2 to 98 %rH) <12…
  • Page 39
    Ordering data Probes for testo 645 / 445 Accessories for humidity/temperature probes, 3-function probe Diameter Part no. Caps for all humidity probes Metal protection cage, material: stainless steel V4A. Quick adjustment time, robust and tempera- 21 mm 0554 0665 12 mm 0554 0755 ture-proof.
  • Page 40
    Ordering data Probes for testo 645 / 445 Velocity probes Sensor Meas. range Accuracy (System) Part no.. Vane/temperature probe, can be attached to 0,6 to 60 m/s ±(0,2 m/s +1 % of m.v.) 0635 9540 handle or telescope -30 to +140 °C…
  • Page 41
    Ordering data Probes for testo 645 / 445 Additional probes Sensor Meas. range Accuracy Part no. 3-function probe for simultaneous measurement Hot bulb 0 to 10 m/s ±(0,03 m/s +5 % v.Mw.) 0635 1540 of Ttemperature, humidity and velocity. With plug-…

This manual is also suitable for:

445

testo 445 · testo 645

Instruction manual

en

Contents

Contents ………………………………………………………………………………………………………………………….

2

Foreword………………………………………………………………………………………………………………………….

3

Initial operation ………………………………………………………………………………………………………………….

4

First measurement……………………………………………………………………………………………………………..

5

Instrument description

-Keypad/ connection assignment………………………………………………………………………………………

6

-Display…………………………………………………………………………………………………………………………

7

Overview of controls………………………………………………………………………………………………………..

8-9

1. Current measurement……………………………………………………………………………………………………

10

Switching on, selecting parameter in display……………………………………………………………….

10

Saving, printing ……………………………………………………………………………………………………..

10

2. Measurement functions11-13

Freezing readings……………………………………………………………………………………………………

11

Maximum readings………………………………………………………………………………………………….

11

Minimum readings…………………………………………………………………………………………………..

11

Multi-point mean calculation……………………………………………………………………………………..

12

Timed mean calculation …………………………………………………………………………………………..

13

3. Location selection ………………………………………………………………………………………………………..

14

4. Parameter settings …………………………………………………………………………………………………..

15-17

System adjustment (testo 645) …………………………………………………………………………………

15

Adjusting humidity probes………………………………………………………………………………………..

16

Displaying calculated humidity parameters (td°C; g/m3; g/kg; J/g)…………………………………..

16

Switches CO2 ppm Vol% (testo 445) …………………………………………………………………….

16

Initialising differential pressure probe (testo 445) ………………………………………………………….

17

Activating m/s in differential pressure probes (testo 445) ………………………………………………

17

Activates volume flow and sets duct cross-sections

for velocity and differential pressure probes (testo 445) ………………………………………………..

17

5. Memory settings………………………………………………………………………………………………………

18-21

Overview ……………………………………………………………………………………………………………….

18

Manual saving ………………………………………………………………………………………………………..

19

Automatic saving…………………………………………………………………………………………………….

19

Reading or printing memory contents ………………………………………………………………………..

20

Clearing memory contents ……………………………………………………………………………………….

21

According to the conformity certificate, the instruments fulfill 2014/30/EC guidelines.

© 1999 Copyright Testo GmbH & Co.

The software and software structure included in the product testo 445/645 are protected by copyright laws worldwide.

2

Contents

6. Instrument configuration……………………………………………………………………………………………

22-26

“AutoOFF” power save function ……………………………………………………………………………….

22

Setting power supply, with rechargeable battery or battery……………………………………………

23

Setting date/time…………………………………………………………………………………………………….

24

Setting parameters for absolute pressure and density compensation………………………………

25

Unit selection …………………………………………………………………………………………………………

26

Factory reset …………..

26

7. Velocity measurement ………………………………………………………………………………………………

27-28

Volume flow funnel (testo 445)………………………………………………………………………………….

27

Pitot tube and pressure probe (testo 445)…………………………………………………………………..

28

Thermal anemometer probes (testo 445) ……………………………………………………………………

28

8. CO/CO2 measurement ……………………………………………………………………………………………..

29-30

Measuring absolute pressure…………………………………………………………………………………………

31-32

Error messages ……………………………………………………………………………………………………………….

33

Technical data…………………………………………………………………………………………………………….

34-35

Ordering data ……………………………………………………………………………………………………………..

36-41

Foreword

Dear Customer

You have made the right decision by choosing a measuring instrument from Testo. Every year, thousands of customers purchase our high-quality products. There are many good reasons for this.

1)We offer value for money. Reliable quality at the right price.

2)Considerably longer guarantee periods of up to 3 years — depending on the instrument.

3)With over 40 years of specialist experience we are optimally equipped to solve your measuring tasks.

4)Our high quality standards are confirmed by ISO 9001 certification.

5)It goes without saying that our instruments bear the CE mark required by the EU.

3

Initial operation

Please read prior to measurement

Do not measure on live parts.

Do not switch instrument on or off with the button

during instrument configuration

during automatic saving

Complete automatic saving before connecting instrument to the Comfort software.

If the instrument does not function remove battery and put it in again. The respective error message is shown in the display. Refer to the information on “Error messages” on page 33 of the Instruction manual.

Observe storage and transport temperature and max. operating temperature (e.g. protect measuring instrument from direct sunlight)

When changing configurations (e.g. changing the probe), the instrument should be switched off because the values specific to the probe can only be read when the instrument is switched on.

The V24 cable (PC connection) can be inserted anytime.

A simultaneous print command is not possible if the PC cable is connected.

Ensure that the contacts are correct in probes with a plug-in head.

The swivel nut in the probe handle should be securely tightened.

Opening the instrument, inexpert handling and use of force cancels your warranty.

Putting in the batteries

9V block battery is included in delivery.

Open the battery compartment at the back of the instrument.

Put in block battery. Observe polarisation.

Close battery compartment.

Refer to “Power supply” Chapter for more information on alternative power supply, charge, battery quality, charging procedure.

Observe instrument configuration when using rechargeable batteries.

4

First measurement

i

Instrument configuration

A description of the instrument and an overview of the controls guarantee a quick introduction.

Note: The measuring instrument must be switched off before a probe is connected.

You will receive up-to-date readings once a probe is inserted and the measuring instrument is switched on. However, you will still need to update or define the data in the instrument:

Date/Time:

Auto Off:

Units:

Some things can only be set via PC software (See Ordering data):

Location name (8 characters)

Log head (24 characters), e.g. your company name — is also printed when readings are printed.

5

Testo 645 Operating Manual

Keypad:

On/Off

Prints

Confirms menu setting/

executes function Saves

Return to current measurement/ location

Arrow buttons move the cursor

Hold Max Min

Mean

Instrument description

Keypad/Connection assignment

Connection assignment:

RS232

2

1

12V

testo 445

Socket 1:

Thermal velocity probes (heated wire or ball)

Vanes (inductive)

Temperature probes (Type K/J/S T/C or NTC)

Differential pressure probes (piezoresistive)

Absolute pressure probes (piezoresistive)

CO2 probe

(2 beam infrared sensor)

CO probe

Socket 2:

Combined probe for humidity (capacitive)

and temperature (NTC or Ni 10000)

3-function probe for humidity (capacitive),

temperature (NTC) and velocity (ball)

RS232: PC connection 12V: Mains connection

testo 645

Socket 1:

Temperature probes (T/C type K/J/S or NTC)

Socket 2:

Combined probe for humidity (capactive)

and temperature (NTC or Ni 10000)

Temperature probes (PT100)

RS232: Connection for PC 12V: Mains unit connection

6

Instrument description

Display

The symbols on the top line are explained below

Name of input socket and parameter

Displays reading in line 1

Name of input socket and parameter

Displays reading in line 2

Displays measurement functions

Explanation of symbols:

Counter for the log number in the memory.

When saving manually: number of a measurement saved. When saving automatically: number of a measurement series.

This counter is needed in order to be able to find single logs or a measurement series when reading the memory.

Counter for saving a measurement cycle (is only required with automatic saving).

The measurement cycle in a measurement series can be found.

Manual saving of a single measurement by pressing the save button .

Automatic saving program has been set up.

Saving is activated by pressing the button.

Symbol for reading contents of memory on display.

Symbol for deleting memory contents

If this symbol appears, the printing function is activated. The symbol flashes while data is being transmitted. You can

print on the desktop printer by pressing the print button .

Shows capacity of battery and rechargeable battery.

If the inner segment no longer appears (symbol flashes), the battery has to be changed or the rechargeable battery has to be recharged. The instrument switches itself off automatically after 1 minute.

7

Overview of controls

ESC

3.

Connect probe

corresponding to

Location

Location

connection assignment

1

Buttons to

(rear of instrument)

select

99

location

2. Measurement functions

1.

Current measurement

Select reading in

HOLD

the first line

Freeze values

Select reading in

MAX

Display maximum values

the second line

Save manually or automatically,

MIN

Manual

depending on setting (5.)

Display minimum values

saving

MEAN

Print

Print readings

readings

Multi-point mean calculation

Update values

Calculate mean

Reset mean

MEAN

Timed mean calculation

Start and end

Calculate mean

Reset mean

4.

Parameter settings

Selected

The setting functions are activated in accordance with the reading

testo testo

parameter

selected — flashing cursor in display (See 1.):

645

445

°C

System adjustment with EEPROM probe

X

(4.1)

Reset — system adjustment

X

%

Humidity probe (0636.9740 and 0636.9715)/

X

X

(4.2)

Checks and adjusts 3-function probe (0635.1540)

td

Select calculated humidity parameters: td°C dew point

X

X

(4.3)

g/m3 — absolute humiidty, g/kg — degree of humidity, J/g — enthalpy

m/s, mbar

Differential pressure probes: initialisation, activation/deactivation

X

m3/h

m/s, m3/h in addition to mbar

(4.4)

Velocity probes: activation/deactivation m3/h in addition to m/s

X

Change cross-section input for an activated volume flow measurement

X

ppm

Switch between ppm and vol% with CO2

X

(4.5)

Initialisation of CO probe

8

Connect probe corresponding to connection assignment (rear of instrument)

1.

Current measurement

5.

Memory settings

MAN

Activate manual saving

AUTO

Program automatic saving

Select meas. cycles

Set measurement

rate

OUT

Read memory contents on display with printing option

CLEAR

Clear memory Note: Everything in the memory is deleted

Overview of controls

Simultaneously

6.

Instrument configuration

Activate — deactivate automatic switch-off function

Select power supply battery/rechargeable battery

Setting: Date

Setting: Time

Parameter setting for absolute pressure and density compensation

°C, %, Abs.mbar,

rho-g/m3, Pitot tube factor

Print parameters

Unit selection

°C °F

m/s fpm [ft/min] (testo 445)

m3/h m3/min m3/s l/s cfm [ft3/min](testo 445)

hPa mbar psi in W [”H2O] (testo 445)

Factory reset

Note:

Flashing display is activated and is confirmed by pressing .

9

1. Current measurement

Switching on / Selecting parameters / Saving / Printing

Connect measurement probes

I/0

Segment test

Line 1

Line 2

Battery / rech.

BAT

V

8.5

battery power

Location

1

Location

Duct

1

M.05

N.0017

MAN

1:

m/s

Line 1

58.06

1:

hPa

Line 2

21.80

Current readings — different, depending on probe(s) connected.

All of the readings, made available by the probes connected to socket 1 and 2, on both lines of the reading display can be called up with the exception of the value selected in the other line.

Scroll with button in line 1:

Select socket.

Select parameter.

If this button in multi-function probes is pressed repeatedly, it is possible to select the parameter (°C, m/s…) to be displayed.

Scroll with button in line 2:

Select socket.

Select parameter.

If this button in multi-function probes is pressed repeatedly, it is possible to select the parameter (°C, m/s…) to be displayed.

There are different processing options available in the menus described below corresponding to the parameters selected.

Example of a reading:

Selected

Parameter

1:

m/s

socket

58.06

Reading

It is possible to activate the following functions during measuring at the touch of a button.

Save readings

Manual or automatic saving is determined by the save setting (Chapter 5)

Print readings.

Note on printing:

A distance 0f 0.5 m should not be exceeded in order to guarantee problem-free data transfer. Objects between instrument and printer prevent the data from being transferred. Please observe notes in the instruction manual on the printer.

10

2. Meaurement functions

Hold/Max/Min

M.05 N.0017 Current measurement

58.06

1:hPa

21.80

Hold

Max/Min

Mean

Freezing values

M.05 N.0017

MAN

1HLD

m/s

Save reading

58.06

Print reading

1HLD

hPa

Hold

21.60

Max/Min

Mean

Maximum readings since the

M.05

N.0017

MAN

measurement began

1MAX

m/s

Save maximum

73.37

measured value

Print maximum

1MAX

hPa

measured value

Hold

34.80

Max/Min

Mean

Minimum readings since the

M.05

N.0017

MAN

measurement began

1MIN

m/s

Save measured

56.62

value

Print measured value

1MIN

hPa

Hold

20.00

Max/Min

Mean

See multi-point/timed mean calculation

11

2. Measurement functions

Multi-point mean calculation Mean

Only for parameter in the 1st line

M.05 N.0017

Current measurement

MAN

1:m/s

58.06

1:hPa

21.80

Hold

Hold

Max/MinHold

Max/Min

MeanMax/MinHold

Mean

Mean

M.05 N.0017

MAN

1:m/s

Multi-point mean calculation

The mean is only calculated for the parameter selected in the first line.

OK

New mean calculation possible

25.21

Selection of volume flow (see note on

Exception:

MEAN

0

page 13)

Select parameter required.

Update values.

Calculate mean

M.05 N.0017

MAN

MEAN

m/s

25.21

Mean

5

Number of updated

values

Save mean

M E A N N E W

Print mean

12

Current measurement

M.05

N.0017

M.05

N.0017

MAN

MAN

1:

m/s

Hold

1:

m/s

Hold

Max/Min

58.06

Mean

55.21

Hold

Max/Min

Hold

Mean

Max/Min

Hold

Mean

1:

hPa

Max/Min

START

Mean

21.80

Max/Min

Mean

00:00

M.05

N.0017

MAN

1:

m/s

Note on saved or printed logs:

1. The log of a multi-point mean

55.17

calculation contains single value,

max. value, min. value and

END

MEAN

mean value.

00:10

2. The log of a timed mean

calculation contains max.

value, min. value and mean

value.

Special solution for volume

flow selected:

3. Any additional velocity values

M.05

N.0017

(m/s) are saved with the volume

MAN

flow values in the memory.

1:

m/s

55.20

GO

MEAN

00:15

New mean

calculation possible

M.05

N.0017

MAN

MEAN

m/s

55.21

MEAN NEW

ESC

or

00:05

Location

2. Measurement functions

Timed mean calculation Mean

Only for parameter in the 1st line

Select required reading.

Start timed mean calculation

Finish timed mean calculation.

Continue timed mean calculation or

calculate mean.

Mean

Timed duration of mean calculation

Save mean.

Print mean.

13

3. Location selection

M.05

N.0017

The location names Location 1 to 99 are specified by the factory.

MAN

1:

Additional location names (8 characters) can only be loaded onto

m/s

1. Current

your instrument using PC software.

58.06

measurement

1:

hPa

ESC

21.80

Location

Location

/

Select location

Duct 101

A cm

50.0

The cross-section can only be called up if volume flow is activated (testo 445 instrument)

B cm

50.0

Return to current measurement

Increase value

Decrease value

Select digit box

Use ESC to go to selection of

cross-section entryØ AxB, m2, Funnel)

Increase value

Decrease value

Select digit box

or:

From now on all measured data which is saved or printed is linked to the selected location or product names.

14

Loading…

Высокоточный терморегисторtesto 645 имеет два разъема для подключения целого ряда зондов:

— Канал 1: зонд температуры т/п Тип K/J/S, NTC

— Канал 2: комбинированный зонд влажности /температуры или зонд температуры Pt 100

Прибор автоматически отображает измеряемые параметры: температуру, относительную влажность, температуру точки росы, температуры точки росы под давлением, абсолютную влажность, уровень влажности и энтальпию.

Кроме того, нажатием кнопки на дисплее отображаются текущие показания, последнее измеренное значение мин./макс./средн. значение.

testo 645 позволяет сохранять данные измерений по месту замера для дальнейшего анализа на пк, или документирования результатов с помощью быстродействующего принтера Testo (опция). Чехол TopSafe (опция) позволяет использовать прибор в суровых промышленных условиях. Транспортировочный кейс(опция) гарантирует безопасность и надежность хранения и транспортировки термогигрометра, всех зондов и пренадлежностей.

Внимание! Для работы прибора необходим зонд, приобретаемый отдельно.

Отличительные особенности testo 645:

  • Высокоточное измерение влажности до + 1%ОВ
  • Промышленное исполнение
  • Большой выбор совместимых датчиков
  • 2 канала для подключения датчиков(температуры и влажности)
  • Встроенная память для сохранения данных (3000 значений)
  • Программное обеспечение для компьютера, для анализа, архивирования и документирования результатов измерений с наименованием мест замеров.
  • Измерение температуры, относительной влажности и температуры точки росы под давлением
  • Функция Hold для фиксирования текущих показаний на дисплее, отображение макс./мин./средн. Значений
  • Удобный анализ данных
  • Заводской протокол калибровки
  • Меню русифицировано

Описание термогигрометра testo 645 (арт: 0560 6450):

Высокоточный терморегистор testo 645 имеет два разъема для подключения целого ряда зондов:

— Канал 1: зонд температуры т/п Тип K/J/S, NTC

— Канал 2: комбинированный зонд влажности /температуры или зонд температуры Pt 100

Прибор автоматически отображает измеряемые параметры: температуру, относительную влажность, температуру точки росы, температуры точки росы под давлением, абсолютную влажность, уровень влажности и энтальпию.

Кроме того, нажатием кнопки на дисплее отображаются текущие показания, последнее измеренное значение мин./макс./средн. значение.

testo 645 позволяет сохранять данные измерений по месту замера для дальнейшего анализа на пк, или документирования результатов с помощью быстродействующего принтера Testo (опция). Чехол TopSafe (опция) позволяет использовать прибор в суровых промышленных условиях. Транспортировочный кейс(опция) гарантирует безопасность и надежность хранения и транспортировки термогигрометра, всех зондов и пренадлежностей.

Внимание! Для работы прибора необходим зонд, приобретаемый отдельно.

Отличительные особенности testo 645:

  • Высокоточное измерение влажности до + 1%ОВ
  • Промышленное исполнение
  • Большой выбор совместимых датчиков
  • 2 канала для подключения датчиков(температуры и влажности)
  • Встроенная память для сохранения данных (3000 значений)
  • Программное обеспечение для компьютера, для анализа, архивирования и документирования результатов измерений с наименованием мест замеров.
  • Измерение температуры, относительной влажности и температуры точки росы под давлением
  • Функция Hold для фиксирования текущих показаний на дисплее, отображение макс./мин./средн. Значений
  • Удобный анализ данных
  • Заводской протокол калибровки
  • Меню русифицировано

Применение testo 645:

1. Локализация плесени

Влажные потолок и стены могут быть результатом структурных повреждений. Однако, в большинстве случаев – это признак неисправной вентиляции. Как только плесень обнаружилась в жилом помещении, жилец предполагает, что она появилась извне. Множество споров произошло по поводу выяснения, кто несет ответственность за появление плесени. На самом деле, довольно сложно установить, что являлось причиной – жилец или структурное повреждение здания.

Testo 645 – идеальный прибор для измерения влажности стен и потолка. Используя подключаемый зонд температурывлажности и зонд температуры поверхности, testo 645 автоматически рассчитывает расстояние до точки росы. Это поможет установить причину появления плесени, появилась ли она из-за влаги извне в связи со структурными повреждениями или неисправностью изоляции.

2. Измерение влажности материалов

Измерение влажности строительных материалов и различных пород древесины, является важной частью повседневной работы в самых различных областях. Для большинства пользователей важно иметь возможность измерить материал быстро и без повреждений.

Неважно применяется ли прибор для оценки сушильного процесса цементной стяжки или горючей древесины,гигрометр testo 645 идеально подходит для контроля содержания влаги в строительных материалах и древесине.

3. Измерение влажности древисины

К прибору testo 645 можно подключить зонд рассеянного поля чтобы быстро и без повреждений измерить влажность древесины. Характеристические кривые для твердых и мягких пород древесины и ДСП зашиты в память прибора.

Измеренные результаты актуальны на 5 см в глубину и соответствуют содержанию влаги в процентном соотношении веса материала в отношении к весу сухого продукта.

Измеренные значения можно сохранить в памяти прибора или передать на ПК для дальнейшего хранения и обработки, используя специальное ПО.

4. Мониторинг сушильных процессов

Будь то свежая стяжка, или контроль ущерба высыхания влаги, своевременное определение уровня влаги позволит Вам не терять времени и сохранить деньги.

Как правило, укладка стяжки следует после штукатурных работ. Далее следуют укладка полов и отделочные работы. Поэтому расписание сушильных процессов крайне важно.

Измерение влажности позволит оценить настало ли время укладки полов.

Testo 645 производит быстрые и неразрушающие измерения равновесной влажности. Прибор так же облегчает работу при сушильных процессах полов и стен.

5. Мониторинг производственных процессов и условий хранения и транспортировки

Для измерения температуры и влажности повсеместно используются решения testo на производствах и складах, в серверных, в музеях и архивах, в охлаждаемых складских зонах, в рефрижераторных контейнерах и витринах.

Производители, перевозчики, или дилеры чувствительных продуктов, должны быть уверены в правильности соблюдения параметров микроклимата на всех этапах. Testo 645 предлагает возможность выборочных замеров в соответствующих местах.

Прибор для измерения температуры/влажности testo 645 предлагает широкий выбор зондов влажности. С помощью прибора можно измерять отностиельную влажность воздуха, влажность материалов. В дополнение к стандартным зондам, к прибору можно подключить до 3 радиозондов. Измеренные значения можно сохранить в памяти прибора или передать на ПК для дальнейшего хранения и обработки при помощи специального ПО.

Техничские характеристики

Параметры Значения
Измерение температуры (сенсор Pt100)
Диапазон измерений -200 … +800 °C
Погрешность ±0,1 % от изм. знач. (200,1° … 800 °C)
±0,2 °C (-200 … 200 °C)
Разрешение 0,1 °C
Измерение температуры (сенсор NTC)
Диапазон измерений -50 … +150 °C
Разрешение 0,1 °C
Измерение температуры (термопара тип K (NiCr-Ni))
Диапазон измерений -200 … +1370 °C
Погрешность ±0,5 % от изм. знач. (-200 … -60,1 °C)
±0,5 % от изм. знач. (+60,1 … +1370 °C)
±0,3 °C (-60 … +60 °C)
Разрешение 0,1 °C
Измерение температуры (термопара тип J (Fe-CuNi))
Диапазон измерений -40 … +750 °C
Погрешность ±0,5 % от изм. знач. (-200 … -60,1 °C)
±0,5 % от изм. знач. (+60,1 … +750 °C)
±0,3 °C (-60 … +60 °C)
Разрешение 0,1 °C
Измерение температуры (термопара тип S (Pt10Rh-Pt))
Диапазон измерений -50 … +1700 °C
Измерение влажности (емкостный сенсор)
Диапазон измерений 0 … 100 % ОВ
Погрешность see probe data
Разрешение 0,1 % ОВ (-50 … +150 °C)
Общие технические данные
Размеры 215 x 68 x 47 мм
Рабочая температура 0 … +50 °C
Корпус ABS
Количество каналов 2-х канальный
Гарантия 2 года
Тип батареи 9 В блочная батарейка
Ресурс батареи 45 h
Тип дисплея LCD
Размер дисплея Размер дисплея: четыре строки
Температура хранения -20 … +70 °C
Вес 255 г

Внешний вид и основные элементы термогигрометра

Комплект поставки

Наименование Количество
Промышленный термогигрометр testo 645 1
Чехол 1
Батарейки 1
Заводской протокол калибровки 1

Понравилась статья? Поделить с друзьями:
  • Багрос 4000 руководство программиста
  • Цефтриаксон инструкция по применению цена уколы детям дозировка
  • Мануал шкода рапид 2015
  • Ацц 600 мг в порошке инструкция по применению
  • Honor инструкция на русском для чайников