Xmt868i руководство по эксплуатации

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

Flow

DigitalFlow™ XMT868i

Ultrasonic Flow Transmitter for Liquids

(1 & 2-Channel)

Service Manual

910-171S

panametrics.com

August 2021

loading

Related Manuals for Panametrics DigitalFlow XMT868i

Summary of Contents for Panametrics DigitalFlow XMT868i

  • Page 1
    Flow DigitalFlow™ XMT868i Ultrasonic Flow Transmitter for Liquids (1 & 2-Channel) Service Manual 910-171S panametrics.com August 2021…
  • Page 3
    Service Manual 910-171S August 2021 panametrics.com Copyright 2021 Baker Hughes company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and company names are trademarks of their respective…
  • Page 4
    [no content intended for this page]…
  • Page 5: Table Of Contents

    Contents Chapter 1. Calibration Introduction ………………… .1 Accessing the Calibration Menu .

  • Page 6
    Contents Appendix B. Calibration and Testing with PanaView Introduction ………………..19 Accessing the Calibration/Test Menu.
  • Page 7
    Preface Information Paragraphs • Note paragraphs provide information that provides a deeper understanding of the situation, but is not essential to the proper completion of the instructions. • Important paragraphs provide information that emphasizes instructions that are essential to proper setup of the equipment.
  • Page 8
    Preface Environmental Compliance Waste Electrical and Electronic Equipment (WEEE) Directive Baker Hughes is an active participant in Europe’s Waste Electrical and Electronic Equipment (WEEE) take-back initiative, directive 2002/96/EC. The equipment that you bought has required the extraction and use of natural resources for its production. It may contain hazardous substances that could impact health and the environment.
  • Page 9: Chapter 1. Calibration

    Chapter 1. Calibration Chapter 1. Calibration Introduction Calibrating the Model XMT868i’s analog outputs and inputs is explained in this chapter. In addition, testing the optional totalizer, frequency and alarm relay outputs is discussed. The chapter includes the following specific topics: •…

  • Page 10: Calibrating And Testing Analog Outputs

    Chapter 1. Calibration Calibrating and Testing Analog Outputs Every Model XMT868i flowmeter includes two built-in analog outputs (A and B) at terminal block J1, which is designated as Slot 0. Additional analog outputs may be added to the Model XMT868i by installing an appropriate option card in Slot 1.

  • Page 11: Calibrating The Low End Of The Output Range

    Chapter 1. Calibration 1.3.2 Calibrating the Low End of the Output Range From the Keypad Program, press the right arrow key to scroll to the CALIB menu. Press [Enter]. In the Cal menu, scroll to the Slot 0 or Slot 1 option, and press [Enter]. Note: The Slot 1 option only appears in the menu if an option card has been installed in this slot.

  • Page 12: Calibrating Analog Inputs

    Chapter 1. Calibration 1.3.4 Testing Output Linearity (cont.) Table 1 lists the expected ammeter readings at various % Full Scale settings, for both 4-20 mA and 0-20 mA scales. Refer to this table to verify the accuracy of the ammeter readings taken above. Table 1: Expected Ammeter Readings % Full Scale 4-20 mA Scale*…

  • Page 13: Preparing For Calibration

    Chapter 1. Calibration 1.4.1 Preparing for Calibration Prepare for the calibration procedure by connecting a Slot 0 analog output (or an independent calibrated current source) to the desired input on the option card. See Figure 4 on page 12 to identify the IN(+) and RTN(-) pins for the desired input on terminal block J2.

  • Page 14: Calibrating Rtd Inputs

    Chapter 1. Calibration Calibrating RTD Inputs RTD analog inputs may be added to the XMT868i by installing an appropriate option card in Slot 1. The option card contains two or four RTD inputs that are designated as A, B, C and D. Both the set point and slope point values for each input must be specified prior to use.

  • Page 15: Testing Alarm Relays

    Chapter 1. Calibration Testing Alarm Relays Alarm relays may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card contains two or four alarm relays, which are designated A, B, C and D. 1.6.1 Preparing for Testing Prepare for the testing procedure by connecting an ohmmeter to the NO and COM pins of the desired alarm on the…

  • Page 16: Testing Totalizer Outputs

    Chapter 1. Calibration Testing Totalizer Outputs Totalizer outputs may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card contains two or four totalizer outputs, which are designated A, B, C and D. 1.7.1 Preparing for Testing Prepare for the testing procedure by connecting a pulse counter to the desired output of the option card, as shown in…

  • Page 17
    Chapter 1. Calibration 1.7.2 Testing the Output (cont.) IMPORTANT: The procedure for testing all the outputs is the same. However, when testing another output, be sure to reconnect the pulse counter at terminal block J2. See Figure 4 on page 12 for the correct pins. Use the arrow keys to enter a value for the minimum pulse on time (between 1 sec and 10,000 sec) for the frequency of the totalizer pulses and press [Enter].
  • Page 18: Testing The Frequency Outputs

    Chapter 1. Calibration Testing the Frequency Outputs Frequency outputs may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card contains two or four frequency outputs, which are designated A, B, C and D. 1.8.1 Preparing for Testing Prepare for the testing procedure by connecting a frequency counter to the desired output on the option card, as…

  • Page 19
    Chapter 1: Calibration Escape Escape Enter KEYPAD PROGRAM PROG RESET CALIB CNTRS (See Programming Manual) Reset Totals DARKN LITEN SLOT0 SLOT1 Slot 0 Analog Outputs Slot 1 Option Card Analog Inputs Alarm Relays Slot 1 Slot 1 ANALOG OUTPUT press [ESC] key ALARMS 4 mA 20 mA…
  • Page 20
    Chapter 1: Calibration -01 (AA,HH) -02 (FF,TT,FT,CT,CF) -03 (FO,TO,CO) -04 (FA,FH,TA,TH,CA,CH) -05 (CI,TI,FI) -06 (CR,FR,TR) Pin 1 Pin 1 Pin 1 Pin 1 Pin 1 Pin 1 -07 (CIR) -08 (AI,HI) -09 (OI) -10 (OR) -11 (AR,HR) -12 (II) Pin 1 Pin 1 Pin 1 Pin 1…
  • Page 21: Chapter 2. Error Codes

    Chapter 2. Error Codes Chapter 2. Error Codes Introduction The Model XMT868i ultrasonic flow transmitter is a reliable, easy to maintain instrument. When properly installed and operated, as described in the Startup Guide, the meter provides accurate flow rate measurements with minimal user intervention.

  • Page 22
    Chapter 2. Error Codes E0: No Error Problem: No error condition currently exists. Cause: This message appears briefly to confirm that the response to another error message has corrected the problem. Action: No action is required. E1: Low Signal Problem: Poor ultrasonic signal strength or the signal exceeds the limits entered using the User Program.
  • Page 23
    Chapter 2. Error Codes E5: Amplitude Error Problem: The signal amplitude exceeds the limits programmed in the SETUP-SIGNAL submenu of the User Program. Cause: Solid or liquid particulates may be present in the flowcell. Poor coupling for clamp-on transducers. Action: Refer to Chapter 3, Diagnostics, to correct any flowcell problems.
  • Page 24
    Chapter 2. Error Codes DigitalFlow™ XMT868i Service Manual…
  • Page 25: Chapter 3. Diagnostics

    Chapter 3. Diagnostics Chapter 3. Diagnostics Introduction This chapter explains how to troubleshoot the Model XMT868i if problems arise with the electronics enclosure, the flowcell, or the transducers. Indications of a possible problem include: • display of an error message on the LCD display screen or the PanaView text display •…

  • Page 26: Viewing Diagnostic Parameters Using The Keypad Program

    Chapter 3. Diagnostics 3.2.1 Viewing Diagnostic Parameters Using the Keypad Program (cont.) Scroll to the desired Channel option, as listed in Table 2. Table 2: Channel Options Option Description Channel 1 Channel 2 CH1+CH2 CH1-CH2 (CH1+CH2)/2 For each channel, select the desired Measurement Parameter, as shown in Table 3. Note: The measurement units that appear in these prompts are those selected in the GLOBL-SYSTM menu earlier in this section.

  • Page 27: Viewing Diagnostic Parameters Using Panaview

    Chapter 3. Diagnostics 3.2.2 Viewing Diagnostic Parameters Using PanaView Power up the XMT868i and wait until it has initialized. Launch PanaView and wait until it initializes the XMT868i and displays the basic PanaView window. As described in Appendix C, Programming the XMT868i with PanaView, of the Programming Manual, open the Meter Browser window and highlight the meter to be used.

  • Page 28
    Chapter 3. Diagnostics Table 3: Available Diagnostic Parameters Keypad Parameter PanaView Choice Parameter Choice Description Good SS up UP Sig Strength Signal strength for the upstream transducer. 50–75 <50 or >75 SS do DN Sig Strength Signal strength for the downstream transducer. 50–75 <50 or >75 Soundspeed…
  • Page 29: Diagnostic Record

    Chapter 3. Diagnostics Table 3: Available Diagnostic Parameters Keypad Parameter PanaView Choice Parameter Choice Description Good DENSr Return Dens Displays the density for the return input (energy N.A. N.A. measurement). DELTh Delta h Displays the delta enthalpy which is the difference N.A.

  • Page 30: Flowcell Problems

    Chapter 3. Diagnostics Flowcell Problems If preliminary troubleshooting with the Error Code Messages and/or the Diagnostic Parameters indicates a possible flowcell problem, proceed with this section. Flowcell problems fall into two categories: fluid problems or pipe problems. Read the following sections carefully to determine if the problem is indeed related to the flowcell. If the instructions in this section fail to resolve the problem, contact GE for assistance.

  • Page 31: Pipe Problems

    Chapter 3. Diagnostics 3.4.2 Pipe Problems Pipe-related problems may result either from a failure to observe the installation instructions, as described in the Startup Guide, or from improper programming of the meter. By far, the most common pipe problems are the following: The collection of material at the transducer location(s).

  • Page 32: Transducer Problems

    Chapter 3. Diagnostics Transducer Problems Ultrasonic transducers are rugged, reliable devices. However, they are subject to physical damage from mishandling and chemical attack. Clamp-on transducers are also subject to installation variables such as physical misalignment and faulty coupling to the pipe on which they are mounted. Because transducers problems are largely dependent on the type of transducers used, wetted or clamp-on, the following list of potential problems is grouped according to transducer type.

  • Page 33: Clamp-On Transducer Problems

    Chapter 3. Diagnostics 3.5.2 Clamp-on Transducer Problems POOR COUPLING TO PIPE: Clamp-on transducers must be in close contact with the pipe. Make sure the pipe wall is smooth and generally free of paint. The couplant material must fill voids between the transducer and the pipe, and must be firmly coupled or bonded to both the pipe and the transducer.

  • Page 34
    Chapter 3. Diagnostics DigitalFlow™ XMT868i Service Manual…
  • Page 35: Chapter 4. Parts Replacement

    Chapter 4. Parts Replacement Chapter 4. Parts Replacement Introduction The Model XMT868i has been designed to permit easy on-site upgrades and parts replacement. See Figure 7 on page 11 for details of the standard XMT868i electronics enclosure assembly. The instructions in this chapter, along with a few common tools, are all that is required to perform the following tasks: •…

  • Page 36: Removing The Circuit Board Assembly

    Chapter 4. Parts Replacement Removing the Circuit Board Assembly All parts replacement procedures for the Model XMT868i require the removal of the circuit board assembly from the electronics enclosure. Therefore, complete the instructions in this section before attempting any other service tasks. The Model XMT868i circuit board assembly consists of circular front and rear circuit boards with several rectangular circuit boards sandwiched between the circular end boards and oriented perpendicular to them.

  • Page 37: Replacing The Lcd Display/Interconnect Board Subassembly

    Chapter 4. Parts Replacement Replacing the LCD Display/Interconnect Board Subassembly The LCD display normally provides years of dependable service, but it is easily field-replaceable when necessary. To replace the LCD display, see Figure 8 on page 12 for the component locations, and complete the following steps: Complete the steps in Removing the Circuit Board Assembly on page 2.

  • Page 38: Replacing The Fuse

    Chapter 4. Parts Replacement Replacing the Fuse Note: This procedure applies only to the fuse on the DC power supply. The fuse on the AC power supply is not field-replaceable; please consult the factory for replacement. If it has been determined that the fuse in the XMT868i has been blown, complete the following steps to install a new fuse: Complete the steps in Removing the Circuit Board Assembly on page 2.

  • Page 39: Replacing The User Program

    Chapter 4. Parts Replacement Replacing the User Program The Model XMT868i’s User Program is stored on an erasable programmable read only memory (EPROM) chip. The EPROM, which is designated as component U6, is mounted in a socket on the front of the main circuit board. EPROM replacement may be required to replace a defective chip or to upgrade to a newer software version.

  • Page 40: Installing An Option Card

    Chapter 4. Parts Replacement Installing an Option Card The Model XMT868i flowmeter can accommodate up to two option cards. A data logging or communication card may be installed in Slot 2 and a variety of I/O option cards are available for installation in Slot 1. To install an option card, refer to Illustrations 9, 10 and 11 in Figure 9 on page 13 and complete the following steps: Complete the steps in the Removing the Circuit Board Assembly section on page 2.

  • Page 41: Attaching An Optional Heat Sink To An I/O Option Card

    Chapter 4. Parts Replacement 4.6.1 Attaching an Optional Heat Sink to an I/O Option Card If you need to attach an optional heat sink to an I/O option card (the TI, TR, FI or FR options in Appendix B of the Startup Guide), refer to Illustration 9 in Figure 9 on page 13 and complete the following steps: Orient the thermal heat pad(s) to the heat sink as shown in Figure 9 on page 13 and apply it to the heat sink.

  • Page 42: Assembling/Installing The Circuit Board Assembly

    Chapter 4. Parts Replacement Assembling/Installing the Circuit Board Assembly Most of the parts replacement procedures described in this chapter require removal of the circuit board assembly from the electronics enclosure. Furthermore, most of the procedures require varying degrees of disassembly of the circuit board assembly.

  • Page 43: Installing The Circuit Board Assembly

    Chapter 4. Parts Replacement 4.7.2 Installing the Circuit Board Assembly Be sure that the circuit board assembly is completely assembled, as described in the previous section. Then, refer to Illustrations 1, 2 and 3 in Figure 7 on page 11 and install the circuit board assembly into the electronics enclosure as follows: Slide the terminal board end of the circuit board assembly into the front of the electronics enclosure.

  • Page 44
    Chapter 4. Parts Replacement DigitalFlow™ XMT868i Service Manual…
  • Page 45
    Chapter 4: Parts Replacement Set Screw Set Screw Rear Cover (solid) Front Cover with Window Illustration 1 Electronics Assembly Enclosure Assembly Mounting Screw with Washer Assembly (4 places) Terminal Block on Slot 2 Option Card Ground Screw Label Plate with Ground Symbol Grounding Jumper Wire Illustration 2…
  • Page 46
    Chapter 4: Parts Replacement Mounting Screw and Fuse Cover DC Power Supply Board Washer Assembly (3 places) P9 Connector on Terminal Board LCD Display, connects to Keypad Board and J10 Connector on Main Board Interconnect Board J2 Connector Sub-Assembly P1 Connector Bracket Screw Green Grounding Jumper P4 Connector…
  • Page 47
    Chapter 4: Parts Replacement Screw with Washer Assembly and Mating Nut (not shown) I/O Heatsink Bridge (2 places) Gap Filler (white surface Power Supply Heatsink facing the Bridge) Screw, Lock Washer and Flat Washer (2 places) Screw with Washer Assembly (3 places) Gap Filler Power Supply Board (white surface…
  • Page 48
    Chapter 4: Parts Replacement DigitalFlow™ XMT868i Service Manual…
  • Page 49: Appendix A. Service Record

    Appendix A. Service Record Appendix A. Service Record Introduction Whenever any service procedure is performed on the Model XMT868i flow transmitter, the details of the service should be recorded in this appendix. An accurate service history of the meter can prove very helpful in troubleshooting any future problems.

  • Page 50
    Appendix A. Service Record Table 6: Service Record Date Description of Service Performed By DigitalFlow™ XMT868i Service Manual…
  • Page 51: Diagnostic Parameters

    Appendix A. Service Record A.2.1 Diagnostic Parameters After a successful initial installation of the Model XMT868i and whenever any system malfunction is noticed, the values for the diagnostic parameters should be entered in Table 7. Table 7: Diagnostic Parameters Parameter Initial Current Parameter…

  • Page 52
    Appendix A. Service Record DigitalFlow™ XMT868i Service Manual…
  • Page 53: Appendix B. Calibration And Testing With Panaview

    Appendix B. Calibration and Testing with PanaView Appendix B. Calibration and Testing with PanaView Introduction Calibrating the Model XMT868i’s analog outputs and inputs is explained in this chapter. In addition, testing the optional totalizer, frequency and alarm relay outputs, and XMT hardware and software is discussed. The chapter includes the following specific topics: •…

  • Page 54: Calibrating Slot 0 Analog Outputs

    Appendix B. Calibration and Testing with PanaView To enter data into the user program, double-click on the Calibration/Test menu. The window appears similar to Figure 11. Figure 11: The Calibration/Test Menu Proceed to the appropriate sections of this chapter to calibrate and/or test any of the installed inputs and/or outputs.

  • Page 55
    Appendix B. Calibration and Testing with PanaView From the Calibration/Test menu (see Figure 11 on page 20), double-click on the Calibration entry to open the option. In the center pane, double-click on Slot 0. Note: The Slot 1 option only appears in the center pane if an option card has been installed in this slot.Double-click on either Output A or Output B to open the Analog Output menu.
  • Page 56: Calibrating Slot 1 Option Cards

    Appendix B. Calibration and Testing with PanaView Calibrating Slot 1 Option Cards To calibrate any option card installed in Slot 1 of the XMT868i, enter the Calibration/Test menu as shown in Figure 11 on page 20. In the center pane, double-click on Slot 1. Note: The Slot 1 option only appears in the center pane if an option card has been installed in this slot.

  • Page 57: Analog Outputs

    Appendix B. Calibration and Testing with PanaView B.4.1 Analog Outputs Additional analog outputs may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card contains two or four analog outputs, which are designated as A, B, C and D. Both the zero-point and full-scale values for each output must be calibrated.

  • Page 58: Analog Inputs

    Appendix B. Calibration and Testing with PanaView B.4.2 Analog Inputs Analog inputs may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card contains two or four analog inputs, which are designated as A, B, C and D. Both the zero-point and full-scale values for each input must be calibrated.

  • Page 59: Rtd Inputs

    Appendix B. Calibration and Testing with PanaView B.4.3 RTD Inputs RTD inputs may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card contains two or four RTD inputs, which are designated as A, B, C and D. Both the set and slope temperature values for each input must be calibrated.

  • Page 60: Testing Alarm Relays

    Appendix B. Calibration and Testing with PanaView B.4.4 Testing Alarm Relays Alarm relays may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card contains two or four alarm relays, which are designated A, B, C and D. Prepare for the testing procedure by connecting an ohmmeter to Alarm A of the option card.

  • Page 61: Testing Totalizer Outputs

    Appendix B. Calibration and Testing with PanaView B.4.5 Testing Totalizer Outputs Totalizer outputs may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card contains two or four totalizer outputs, which are designated A, B, C and D. Prepare for the testing procedure by connecting a pulse counter to Output A of the option card, as shown in Figure 2 on page 8.

  • Page 62: Testing Xmt868I Software And Hardware

    Appendix B. Calibration and Testing with PanaView Testing XMT868i Software and Hardware In addition to calibration procedures for input and output cards, the Calibration/Test menu offers tests for XMT868i hardware and procedures for uploading signal array data and XMT memory contents into PC files for examination. To test XMT868i hardware and software, Open the Meter Browser window and highlight the XMT868i.

  • Page 63
    Appendix B. Calibration and Testing with PanaView B.5.1 Uploading Signal Array Data (cont.) PanaView now offers two options for Uploading the array data. • If you double-click on To Display, the right pane of the window appears similar to Figure 13, and the data is available for immediate evaluation.
  • Page 64: Uploading Xmt868I Memory

    Appendix B. Calibration and Testing with PanaView B.5.2 Uploading XMT868i Memory Note: This option is for factory use only. The Upload XMT Memory option is another diagnostic tool. IMPORTANT: Consult GE before using this data to change any parameters. In the Calibration/Test menu, double-click on the Upload XMT Memory option. The first prompt shows the Hex Start Address.

  • Page 65: Testing Xmt868I Hardware

    Appendix B. Calibration and Testing with PanaView B.5.3 Testing XMT868i Hardware Using PanaView, you can check crucial elements of the XMT868i hardware: the EPROM and the RAM. You can also set the number of channels (either 1 or 2). IMPORTANT: Contact GE if the XMT868i fails any of the tests covered below.

  • Page 66
    Appendix B. Calibration and Testing with PanaView DigitalFlow™ XMT868i Service Manual…
  • Page 67: Appendix C. Factory Tests

    Appendix C. Factory Tests Appendix C. Factory Tests C.6 Introduction For user security, the XMT868i offers two special functions in a separate menu: selection of single- or dual-channel operation, and the factory-default reset. While the first function enables you to reset the channel configurations set during manufacturing, the reset function, by restoring the meter to its original state, allows you to purge potential errors in program memory.

  • Page 68
    Appendix C. Factory Tests The previous two prompts repeat until all of the specified # of LCD Parameters have been set up. When all of the display parameters have been set up, the meter returns to the Global I/O window. To leave the Keypad Program, press the [Escape] key three times.
  • Page 69
    Index Numerics EPROM 20 mA Option, Analog Inputs 5, 24 Replacing 5 Test 31 4 mA Option, Analog Inputs 5, 24 User Program 5 Error Codes Alarm Option Card, Testing 7 E13 — Volumetric Overflow 3 E0 — No Error 2 Alarm Relays, Testing 26 E1 — Low Signal 2 Ammeter, in Calibration 2…
  • Page 70
    Index Transducers Option Card Corrosion 8 Cycle Skip 9 Alarm Relay 7 Internal Damage 8, 9 Analog Inputs 4 Leaks 8 Analog Outputs 2, 23, 24, 25 Misalignment 9 Bracket 6 Physical Damage 8, 9 Calibrating 22 Placement 7 Expansion Slots 1, 19 Poor Coupling 9 Frequency Outputs 10 Problems 8…
  • Page 71
    Notify Panametrics Sensing, giving full details of the problem, and provide the model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, Panametrics will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to a service center will be provided.
  • Page 72
    Warranty [no content intended for this page] DigitalFlow™ XMT868i Service Manual…
  • Page 74
    Customer Support Centers U.S.A. The Boston Center 1100 Technology Park Drive Billerica, MA 01821 U.S.A. Tel: 800 833 9438 (toll-free) 978 437 1000 E-mail: mstechsupport@bakerhughes.com Ireland Sensing House Shannon Free Zone East Shannon, County Clare Ireland Tel: +353 (0)61 470291 E-mail: mstechsupport@bakerhughes.com Copyright 2021 Baker Hughes company.

GE Sensing XMT868i Operating Manual

GE

Measurement & Control

Flow

DigitalFlow™ XMT868i

Ultrasonic Flow Transmitter for Liquids (1 & 2-Channel)

Programming Manual

910-171P Rev. G

January 2014

DigitalFlow™ XMT868i

Ultrasonic Flow Transmitter for Liquids (1 & 2-Channel)

Programming Manual

910-171P Rev. G

January 2014

www.ge-mcs.com

©2014 General Electric Company. All rights reserved. Technical content subject to change without notice.

[no content intended for this page]

Preface

Information Paragraphs

Note paragraphs provide information that provides a deeper understanding of the situation, but is not essential to the proper completion of the instructions.

Important paragraphs provide information that emphasizes instructions that are essential to proper setup of the equipment. Failure to follow these instructions carefully may cause unreliable performance.

Caution! paragraphs provide information that alerts the operator to a hazardous situation that can cause damage to property or equipment.

Warning! paragraphs provide information that alerts the operator to a hazardous situation that can cause injury to personnel. Cautionary information is also included, when applicable.

Safety Issues

WARNING! It is the responsibility of the user to make sure all local, county, state and national codes, regulations, rules and laws related to safety and safe operating conditions are met for each installation.

Auxiliary Equipment

Local Safety Standards

The user must make sure that he operates all auxiliary equipment in accordance with local codes, standards, regulations, or laws applicable to safety.

Working Area

WARNING! Auxiliary equipment may have both manual and automatic modes of operation. As equipment can move suddenly and without warning, do not enter the work cell of this equipment during automatic operation, and do not enter the work envelope of this equipment during manual operation. If you do, serious injury can result.

WARNING! Make sure that power to the auxiliary equipment is turned OFF and locked out before you perform maintenance procedures on the equipment.

Qualification of Personnel

Make sure that all personnel have manufacturer-approved training applicable to the auxiliary equipment.

Personal Safety Equipment

Make sure that operators and maintenance personnel have all safety equipment applicable to the auxiliary equipment. Examples include safety glasses, protective headgear, safety shoes, etc.

Unauthorized Operation

Make sure that unauthorized personnel cannot gain access to the operation of the equipment.

DigitalFlow™ XMT868i Programming Manual

iii

Preface

Environmental Compliance

Waste Electrical and Electronic Equipment (WEEE) Directive

GE Measurement & Control is an active participant in Europe’s Waste Electrical and Electronic Equipment (WEEE) take-back initiative, directive 2002/96/EC.

The equipment that you bought has required the extraction and use of natural resources for its production. It may contain hazardous substances that could impact health and the environment.

In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources, we encourage you to use the appropriate take-back systems. Those systems will reuse or recycle most of the materials of your end life equipment in a sound way.

The crossed-out wheeled bin symbol invites you to use those systems.

If you need more information on the collection, reuse and recycling systems, please contact your local or regional waste administration.

Visit http://www.ge-mcs.com/en/about-us/environmental-health-and-safety/1741-weee-req.html for take-back instructions and more information about this initiative.

iv

DigitalFlow™ XMT868i Programming Manual

Contents

Chapter 1. Programming Site Data

1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

1.2 Programming Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

• Keypad Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2

1.2.1 Accessing the Keypad Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3

1.3 Entering Data in the Channel Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

1.3.1 Selecting the Channel Measurement Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4

1.3.2 Entering Data in the Channel System Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5

1.3.2a Programming the Mass Flow Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7

1.3.2b Programming the Energy Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8

1.3.3 Entering Pipe Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9

1.3.3a Special Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9

1.3.3b Pipe Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

1.3.3c Pipe OD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

1.3.3d Path and Axial Lengths. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

1.3.3e Pipe Lining. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

1.3.3f Tracking Windows. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

1.3.3g Fluid Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

1.3.3h Reynolds Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

1.3.3i Number of Traverses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

1.3.4 Entering Input/Output Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

1.3.4a Zero Cutoff Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

1.3.4b Temperature Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

1.3.5 Entering Setup Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

1.3.5a The SIGNL Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

1.3.5b The AVRG Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

1.3.5c The INIT Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

1.3.5d The ADVAN Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

1.3.5e The KV/xx Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

1.3.5f The MULTK Factors Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

1.3.5g The Mass Flow Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

1.3.5h The Code Length Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

1.4 Entering Data in the GLOBL Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

1.4.1 Entering GLOBL-SYSTM Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

1.4.1a Volumetric Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

1.4.1b Totalizer Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

1.4.1c Programming Mass Flow Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

1.4.1d Programming the Energy Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

1.4.2 Setting Up Inputs and Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

1.4.2a Setting Up Error Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

DigitalFlow™ XMT868i Programming Manual

v

Contents

1.4.2b Setting Up Option Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29 1.4.3 Entering Communications Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37 1.4.3a Setting Up the Serial Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37 1.4.3b Setting Up MODBUS (RS485) and MODBUS/TCP Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 1.4.3c Setting Up a MODBUS/TCP Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39 1.4.3d Requesting Parameters Using MODBUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40

1.5 Exiting the User Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42

Chapter 2. Displaying Data

2.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43 2.2 Displaying Data with the LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43 2.2.1 Adjusting LCD Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43 2.2.2 Programming the LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44 2.3 Displaying Data on a Computer Terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46 2.3.1 Preparing for PanaView Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46 2.3.2 The Text Display Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47 2.3.3 The Text Display Output (cont.). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48 2.3.3a Displaying Multiple Process Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48 2.3.3b Displaying Multiple Text Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49 2.3.4 The Graphing Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49 2.3.4a Setting Up a Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49 2.3.4b Graphing Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 2.3.5 Displaying Transducer Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51 2.3.5a Reading Transducer Signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52 2.3.5b Plotting Transducer Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52 2.3.5c Saving Transducer Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53

Chapter 3. Logging Data

3.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55 3.2 The Data Logging Option Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55 3.3 Logging Using PanaView . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55 3.4 Creating Meter Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56 3.4.1 Creating a Standard Meter Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57 3.4.2 Creating an Error Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59 3.4.3 Checking Log Memory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 3.4.4 Stop Logging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 3.5 Creating PC Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61

3.6 Viewing Meter Log Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64 3.7 Viewing PC Log Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66

Chapter 4. Printing Data

4.1 Data Types for Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69

vi

DigitalFlow™ XMT868i Programming Manual

Contents

Chapter 5. Clearing Data

5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 5.2 Clearing the XMT868i’s Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 5.2.1 Clearing Site Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 5.2.2 Clearing Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 5.2.3 Clearing the Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

Appendix A. Menu Maps

Appendix B. Data Records

B.1 Available Option Cards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 B.2 Option Cards Installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 B.3 Site Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

Appendix C. Programming the XMT868i Using PanaView™

C.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 C.2 Programming Using PanaView™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 C.2.1 Preparing for PanaView Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 C.2.2 Setting Up the Serial Communications Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 C.2.3 Setting up Ethernet Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 C.2.4 Modifying Ethernet Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 C.3 Adding the XMT868i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94

C.4 Entering the User Program With PanaView . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 C.5 Entering Data in the Channel Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 C.5.1 Selecting the Channel Measurement Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 C.5.2 Entering Data in the Channel System Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 C.5.2a Programming the Mass Flow Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102 C.5.2b Programming the Energy Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 C.5.3 Entering Pipe Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104 C.5.3a Special Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104 C.5.3b Pipe Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 C.5.3c Pipe OD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 C.5.3d Path and Axial Lengths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 C.5.3e Pipe Lining . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 C.5.3f Tracking Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 C.5.3g Fluid Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 C.5.3h Reynolds Correction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 C.5.3i Number of Traverses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108 C.5.4 Entering Input/Output Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 C.5.4a Zero Cutoff Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 C.5.4b Temperature Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 C.5.5 Entering Setup Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110

DigitalFlow™ XMT868i Programming Manual

vii

Contents

C.5.5a The Signal Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110 C.5.5b The V Averaging Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 C.5.5c The Default Setup Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

C.6 Entering Data in the Global Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117 C.6.0a Entering Global-System Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118 C.6.0b Volumetric Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 C.6.0c Totalizer Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 C.6.0d Programming Mass Flow Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 C.6.0e Programming the Energy Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 C.6.0f Setting Up Inputs and Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 C.6.0g Setting Up Error Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 C.6.0h Setting Up Option Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125 C.6.0i Changing the LCD Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133 C.6.0j Entering Communications Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134 C.6.0k Setting Up the Serial Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134 C.6.0l Requesting Parameters Using MODBUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136

C.7 Exiting the Site Edit Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138 C.8 Saving Site Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139 C.8.1 Saving Current Site Data to the Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140 C.8.2 Saving New Site Data to the XMT868i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141 C.8.3 Saving a Site to the PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141 C.8.4 Saving Site Data in Text Form. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142

Appendix D. PanaView Menu Maps for the XMT868i Appendix E. Foundation Fieldbus Communications

E.1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151 E.2 Configuration Utility Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152 E.3 Selecting the Desired Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152 E.4 Selecting Units for AI Blocks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154 E.5 Resetting Instrument Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155 E.6 Function Block Application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156

viii

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

Chapter 1. Programming Site Data

1.1Introduction

The Model XMT868i flow transmitter must be properly installed and programmed, as described in the Startup Guide, before it can provide accurate flow rate measurements. After completing the installation and initial setup, use this chapter to program the advanced features of the Model XMT868i’s Keypad Program.

The Keypad Program provides access to the various programmable features of the Model XMT868i. Step-by-step programming instructions are presented in this chapter. Refer to the appropriate section for a discussion of the following menu features:

Channel-ACTIV — activate one or both channels and select the desired measurement method

Channel-SYSTM — enter the individual channel parameters

Channel-PIPE — enter the pipe parameters

Channel-I/O — set up the inputs and outputs

Channel-SET UP — set the signal limits, response times and activating mass flow.

GLOBL-SYSTM — enter the system units

GLOBL-I/O — set up error handling, option cards and display

GLOBL-COMM — set the serial port and MODBUS parameters

As a programming aid, a complete set of menu maps for the PROG menu is included in Appendix A, Menu Maps. The specific figure numbers will be referenced throughout this chapter, as required.

1.2Programming Methods

You can use two methods for programming the XMT868i flowmeter:

The keypad (shown in Figure 1 on page 2), which permits direct communication with the XMT868i

IMPORTANT: The XMT868i’s keypad permits programming of the instrument through the glass faceplate without removing the cover. Thus, all programming procedures may be performed while the unit is installed in a hazardous area.

PanaView™, a PC-based software program that communicates with the XMT868i using its RS232/RS485 serial Ethernet port.

Although the displays differ, the general procedures are the same for the keypad and PanaView methods. This chapter provides detailed keypad programming instructions. If you are using PanaView, please refer to Appendix C,

Programming with PanaView, and/or the PanaView User’s Manual for complete instructions.

DigitalFlow™ XMT868i Programming Manual

1

Chapter 1. Programming Site Data

Keypad Program

Along with the 2-line, 16-character LCD, the XMT868i includes a 6-key magnetic keypad. The decal cutout for each key contains a hall effect sensor, pushbutton switch and visible red LED. The magnetic wand used to activate a magnetic key is found attached to the meter chassis below the front panel. An operator activates the key by pressing the magnetic wand up to the glass lid over the desired key. The LED will light indicating a successful key press.

Note: The pushbutton switch will also act as a key press but is done with the glass lid open. Do not use the pushbutton switch in a hazardous area where the lid needs to be installed.

Use the magnetic keypad to navigate through the user program. The menu map may be followed in sequence, or the four arrow keys may be used to scroll through the prompt screens. Figure 1 shows the front of the XMT868i, with magnetic keypad and magnetic wand.

Magnetic Wand

Figure 1: XMT868i Magnetic Keypad and Wand

Programming of the ACTIV, SYSTM, and PIPE submenus of the CHx menus and the GLOBL-SYSTM menu are required for basic operation of the Model XMT868i. Failure to accurately enter all of the necessary information will result in unreliable flow data. Therefore, be sure to complete at least the sections of this chapter pertaining to those three submenus.

Note: Because it is so essential, instructions for programming the ACTIV, SYSTM, and PIPE submenus of the CHx menus and the GLOBL-SYSTM submenus are also included in the Startup Guide. If that programming has already been completed, skip those sections in this chapter.

Except for the three submenus noted above, it is not necessary to program the Model XMT868i flowmeter in any particular order. Therefore, the sections of this chapter need not be completed in sequence. Enter the keypad program as described in Accessing the Keypad Program on the next page and proceed immediately to any section of interest.

Note: Be sure to record all the programming data entered in this chapter in Appendix B, Data Records.

2

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.2.1 Accessing the Keypad Program

Six keys on the keypad enable users to program the XMT868i:

[Enter] — confirms choice of a specific option and data entry within the option

[Escape] — allows users to exit from a specific option without entering unconfirmed data

[] and []- enable users to highlight a specific window in the display option or to scroll through a list of options (parameters, letters, numbers) in a menu

[] and [] — enable users to scroll to a specific option, among choices in an option, or to a character in a text entry. Also, if the [] key is pressed when the cursor position (underlined) is at the first digit of the bottom LCD line, the previous parameter is displayed and can be edited.

When you power up the XMT868i, the initial screen shows the model and software version:

GE

XMT868i X3A.STD

The meter then starts to display measured parameters.

10.00Ft/s

To enter the Keypad Program, press the [Escape] key, followed by the [Enter] key, and the [Escape] key again. Each successive key must be pressed within 10 seconds of the prior key. The Keypad Program window appears:

KEYPAD PROGRAM

PROG

1

Press [Enter] to begin programming in the Channel or Global menus.

IMPORTANT: If the keypad has not been pressed for 10 minutes, the XMT868i exits the Keypad Program and returns to displaying measurements. The meter retains any configuration changes that were confirmed with the [Enter] key, and restarts as if the operator had completed the programming cycle.

DigitalFlow™ XMT868i Programming Manual

3

Chapter 1. Programming Site Data

1.3Entering Data in the Channel Menu

The Channel menu is used to enter data specific to each channel. Refer to Figure 28 on page 77 and remember to record all programming data in Appendix B, Data Records.

Note: In this manual, only the programming of Channel 1 will be described. To program Channel 2 of a 2-Channel meter, simply repeat the same procedures presented for Channel 1.

1.3.1 Selecting the Channel Measurement Method

The ACTIV submenu permits selection of the desired measurement method.

1.In the Keypad Program, scroll to PROG and press [Enter].

2.In the PROG menu, scroll to the desired channel (CH1 or CH2) and press [Enter].

Note: In the 1-Channel XMT868i, CH2 does not appear.

3.In the Channel PROGRAM menu, scroll to ACTIV and press [Enter].

4.The option offers two choices: OFF and TRANS (Transit-Time). Scroll to the desired option for this channel and press [Enter].

Note: In the 1-Channel XMT868i, OFF does not appear.

The XMT868i returns to the Channel PROGRAM menu. Scroll to the SYSTM option.

4

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.2 Entering Data in the Channel System Option

1.From the Channel PROGRAM menu, scroll to SYSTM and press [Enter].

2.The first prompt asks for the Channel (Site) Label.Use the four arrow keys to enter the desired label (in any numeric or text combination up to five characters), and press [Enter].

3.The next prompt asks for the Channel (Site) Message. Use the four arrow keys to enter up to 15 characters, and press [Enter].

4.The program now asks if you want the Energy Option, which enables flow measurement in a heating or cooling system. Scroll to Off or On, and press [Enter].

5.The next prompt asks for the Volumetric Units. Scroll through the units (from the list shown in Table 1) and press [Enter] when you reach the desired unit.

Note: Refer to the SYSTM option of the GLOBL menu to choose between English and metric measurements.

Table 1: Available Volumetric Units

English Volumetric Units

Metric Volumetric Units

GAL/S — U.S. Gallons per Second

L/S = Liters per Second

GAL/M — U.S. Gallons per Minute

L/M = Liters per Minute

GAL/H — U.S. Gallons per Hour

L/H = Liters per Hour

MGD — Mega U.S. Gallons per Day

ML/D = Mega Liters per Day

ft3/s — Cubic Feet per Second

m^3/s = Cubic Meters per Second

ft3/m — Cubic Feet per Minute

m^3/m = Cubic Meters per Minute

ft3/h- Cubic Feet per Hour

m^3/h = Cubic Meters per Hour

ft3/d- Cubic Feet per Day

m^3/d = Cubic Meters per Day

Mf3/d- Mega Cubic Feet per Day

Mm3/d = Mega Cubic Meters per Day

BBL/S — Barrels per Second

BBL/S = Barrels per Second

BBL/M — Barrels per Minute

BBL/M = Barrels per Minute

BBL/H — Barrels per Hour

BBL/H = Barrels per Hour

BBL/D — Barrels per Day

BBL/D = Barrels per Day

MBL/D — Mega Barrels per Day

MBL/D = Mega Barrels per Day

A-I/S-Acre-inches per second

N/A

A-I/M -Acre-inches per minute

N/A

A-I/H -Acre-inches per hour

N/A

A-I/D -Acre-inches per day

N/A

A-F/S -Acre-feet per second

N/A

A-F/M-Acre-feet per minute

N/A

A-F/H -Acre-feet per hour

N/A

A-F/D -Acre-feet per day

N/A

DigitalFlow™ XMT868i Programming Manual

5

Chapter 1. Programming Site Data

1.3.2 Entering Data in the Channel System Option (cont.)

6.Scroll to the desired number of Vol Decimal Digits (digits to the right of the decimal point) in the volumetric flow display, and press [Enter].

7.Scroll to the desired Totalizer Units for the totalized flow rate display (listed in Table 2), and press [Enter].

.

Table 2: Totalizer Units

English

Metric

GAL — U.S. Gallons

L — Liters

MGAL — Mega U.S. gallons

ML — Megaliters

ft^3 — Cubic Feet

m^3- Cubic Meters

Mft^3 — Mega Cubic Feet

Mm^3- Mega Cubic Meters

BBL — Barrels

BBL — Barrels

MBBL — Megabarrels

MBBL — Megabarrels

AC-IN — Acre-inches

AC-FT — Acre-feet

8.Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the totalized flow rate display, and press [Enter].

The program now varies, depending on whether you have activated Mass Flow (see the Programming Manual for details) or the Energy Option.

If you have activated Mass Flow, continue to step 1 on the next page.

If you have activated the Energy Option (but not Mass Flow), skip to step 1 on page 7.

If you have activated neither option, the XMT868i returns to the Channel PROGRAM window.

6

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.2a Programming the Mass Flow Option

1. Scroll to the desired Mass Flow Units for flow rate display (listed in Table 3), and press [Enter].

Table 3: Available Mass Flow Units

English

Metric

LB = Pounds

Kilograms

KLB = KiloPounds (Thousands of Pounds)

Tonnes = Metric Tons (1000 KG)

MMLB = MegaPounds

TONS (2000 LB)

2.Scroll to the desired Mass Flow Time Units (from seconds to days), and press [Enter].

3.Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the mass flow rate display, and press [Enter].

4.Scroll to the desired Mass Totals (units for the totalized mass flow rate display, listed in Table 3), and press [Enter].

5.Scroll to the desired number of Mass Decimal Digits (digits to the right of the decimal point) in the totalized mass flow rate display, and press [Enter].

The program varies, depending on whether or not you have enabled the Energy Option.

If you have not enabled the option, the XMT868i returns to the Channel PROGRAM window. Proceed to the PIPE option.

If you have enabled the option, continue to the next page.

DigitalFlow™ XMT868i Programming Manual

7

Chapter 1. Programming Site Data

1.3.2 Entering Data in the Channel System Option (cont.)

1.3.2b Programming the Energy Option

1. Scroll to the desired Power Units (listed in Table 4) and press [Enter].

Table 4: Available Power Units

English

Metric

kBTUh = Thousands of BTUs/hour

kCALs = Kilocalories/second

MMBTU = Mega BTUs/hour

kCALm = Kilocalories/minute

kW = Kilowatt

kCALh = Kilocalories/hour

TONS = Tons (2000 LB)

kCALd = Kilocalories/day

MCALs = Megacalories/second

MCALm = Megacalories/minute

MCALh = Megacalories/hour

MCALd = Megacalories/day

kW = Kilowatt

MW — Megawatt

2.Scroll to the desired number of Power Decimal Digits, and press [Enter].

3.Scroll to the desired Energy (Total) units (shown in Table 5), and press [Enter].

Table 5: Available Total Energy Units

English

Metric

kBTU = Thousands of BTUs

kCAL = Kilocalories

MMBTU = Mega BTUs/hour

MCAL = Megacalories

kWHr = Kilowatt-hour

kWHr = Kilowatt-hour

TONS = Tons (2000 LB)

MWHr — Megawatt-hour

4.Scroll to the desired number of Energy Decimal Digits, and press [Enter].

5.If you are using a heating or cooling system, scroll to the appropriate selection (Heat or Cool) and press [Enter].

6.The final prompt asks whether you want to conduct Flow Measurement at the point of RTN (Return, where the liquid exits) or of SPPLY (Supply, where the liquid enters). Scroll to the appropriate choice, and press [Enter].

You have completed entering data in the SYSTM option. The meter returns to the Channel PROGRAM menu.

8

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.3 Entering Pipe Parameters

Enter the transducer and pipe parameters using the PIPE submenu. While following the programming instructions, refer to Figure 29 on page 78.

1.From the Channel PROGRAM menu, scroll to the PIPE option and press [Enter].

2.The first prompt asks for the Transducer Number.

For a standard transducer, enter the number engraved on the transducer head in the right pane, and press [Enter].

If there is no number engraved on the transducer head, press the right arrow key to scroll to the Special option, enter an assigned number (from 91 to 99), and press [Enter].

IMPORTANT: Special transducers, which have no engraved number on the head, are rarely used. Examine the transducer head carefully for a number.

The menu now varies, depending on your selection in step 2.

If you entered the number for a standard clamp-on transducer, proceed to the Pipe Material prompt in step 4.

If you entered the number for a standard wetted transducer, proceed to the Pipe OD prompt in step 5.

If you entered the number for a special transducer, proceed to step 3 below.

1.3.3a Special Transducers

Note: For special transducers, GE will supply a transducer data sheet with programming information.

3. For special transducers:

a.The program first asks for the Wedge Type. Press the appropriate transducer type (Rayleigh clamp-on, shear clamp-on, or wetted).

Note: GE will supply the information required for steps b through f with the transducers.

b.Press the appropriate Frequency (from 500 kHz to 5 MHz).

c.Enter the Time Delay (Tw) value supplied by GE and press [Enter].

If you have selected a special wetted transducer, the program now goes to step 5. But if you have selected a clamp-on transducer, three more prompts appear.

d.Enter the Wedge Angle of the transducer (the angle of the ultrasonic transmission, in degrees) and press [Enter].

e.Enter the Wedge Temperature and press [Enter].

f.Enter the Wedge Soundspeed and press [Enter].

DigitalFlow™ XMT868i Programming Manual

9

Chapter 1. Programming Site Data

1.3.3 Entering Pipe Parameters (cont.)

1.3.3b Pipe Material

4.Scroll to the appropriate Pipe Material from the supplied list. For some materials (shown in Table 6), the program displays additional selections. At the desired choice, press [Enter].

Table 6: Pipe Material Choices

Material

Types

Steel

Carbon Steel

Stainless Steel

Iron

Ductile Iron

Cast Iron

Copper

no additional selection required

Aluminum

no additional selection required

Brass

no additional selection required

Copper Nickel

70%Cu/30%Ni — 70/30 Copper/Nickel

90%Cu/10%Ni — 90/10 Copper/Nickel

Glass

Pyrex

Heavy Silicate Flint

Light Borate Crown

Plastic

Nylon

Polyethylene

Polypropylene

PVC (Polyvinyl Chloride), CPVC

Acrylic

OTHER

Enter the soundspeed of the pipe material

and press [Enter]. If the soundspeed is

unknown, refer to the Sound Speeds and Pipe

Size Data manual (914-004).

10

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.3c Pipe OD

5.To select the appropriate Pipe OD unit type from the list shown in Table 7, scroll to the right side of the screen, and use the up and down arrow keys to step through the list. Press [Enter]. Then enter the known pipe outside diameter or circumference on the left side and press [Enter].

Obtain the required information by measuring either the pipe outside diameter (OD) or circumference at the transducer installation site. The data may also be obtained from standard pipe size tables found in Sound Speeds and Pipe Size Data (914-004).

Table 7: Available Pipe OD Units

English

Metric

inch

mm = millimeters

feet

m = meters

in/PI = pipe circumference in inches

mm/PI = pipe circumference in millimeters

ft/PI = pipe circumference in feet

m/PI = pipe circumference in meters

6. Enter the known Pipe Wall thickness (in inches or mm) in the right pane and press [Enter].

If you are using clamp-on transducers, skip to step 9 on the next page.

1.3.3d Path and Axial Lengths

7.To enter the Path Length:

a.Use the [] arrow key to highlight the path length unit type at the right of the screen.Then use the [] and [] arrow keys to scroll to the desired unit type (inches or feet for English measurements, or millimeters or meters for metric).

b.Use the [] arrow key to return to the numeric entry on the left, and enter the path length of the ultrasonic signal. Press [Enter].

Note: If a spoolpiece was ordered with the meter, the transducer signal path length (P) and the transducer signal axial length (L) are engraved on the flowcell and/or are included in the documentation supplied with the meter. For on-site transducer installations, refer to Appendix C, Measuring P and L Dimensions, for instructions.

8.In the same manner, enter the appropriate Axial Length L unit type and axial length of the ultrasonic signal, and press [Enter]. Skip to step 10 on the next page.

1.3.3e Pipe Lining

9.Scroll to the appropriate choice to indicate whether you have a Pipe Lining.

a.If you have selected Yes, then select the Lining Material from the supplied list.

If OTHER is selected, enter the lining sound speed, and press [Enter].

b.Then enter the lining thickness and press [Enter].

DigitalFlow™ XMT868i Programming Manual

11

Chapter 1. Programming Site Data

1.3.3 Entering Pipe Parameters (cont.)

1.3.3f Tracking Windows

10.Press the appropriate choice to indicate whether you want Tracking Windows. These windows are used to detect the receive signal when you are unsure of the fluid sound speed.

1.3.3g Fluid Type

11.Press the appropriate Fluid Type. The selections for fluid type vary depending on whether:

the Energy Option is ON or OFF; and

the Tracking Window is enabled or disabled.

Refer to Table 8 if Energy Option is OFF, or to Table 9 on page 13 if Energy Option is ON.

Note: Some of the fluid types may require additional selections, as shown in the following tables.

Table 8: Fluid Types for ENERGY OFF

Tracking Windows =

NO

Additional Selections

YES

Additional Selections

Water

Normal Water

Water,

No additional selections required.

Sea Water

0-100C

If Normal Water is selected, enter the Water

Temperature and press [Enter].

Oil

Lube Oil

Water,

No additional selections required.

Crude Oil

0-260C

Methanol

No additional selections required.

Oil,

No additional selections required.

(25 C)

Tracking

Ethanol

No additional selections required.

OTHER

Enter the Minimum Soundspeed and press

(25C)

[Enter]. Then enter the Maximum Soundspeed

and press [Enter].

LN2

No additional selections required.

Freon

No additional selections required.

OTHER

Enter the fluid soundspeed and press [Enter].

12

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.3g Fluid Type (cont.)

Table 9: Fluid Types for ENERGY ON

Tracking Windows =

NO

Additional Selections

YES

Additional Selections

Water

Enter the Water Temperature and press [Enter].

Water,

No additional selections required.

0-260C

Water,

Enter the Fluid Soundspeed and press [Enter].

Water,

Enter the Percentage of Water and press [Enter].

Glycol

Then enter the Percentage of Water and press

Glycol

Mixture

[Enter].

Mixture

OTHER

Enter the Fluid Soundspeed and press [Enter].

OTHER

Enter the Minimum Soundspeed and press

[Enter]. Then enter the Maximum Soundspeed

and press [Enter].

1.3.3h Reynolds Correction

12.Scroll to the appropriate choice to indicate whether you want Reynolds Correction.

If Off is selected, enter the Calibration Factor and press [Enter].

If Activ is selected, the program asks for KV Input Selection. Scroll to SIGS (Signals), SOUND (Soundspeed) or STATC.

If SIGS or SOUND is selected, enter the Calibration Factor and press [Enter].

If STATC is selected, enter the Kinematic Viscosity and press [Enter]. Then enter the Calibration Factor and press [Enter].

For wetted transducers, the program returns to the Channel PROGRAM menu. You have completed entering PIPE parameters.

For clamp-on transducers, proceed to step 13 below.

1.3.3i Number of Traverses

13.Scroll to the appropriate Number of Traverses (from 1 to 5) and press [Enter].

14.Accept the value for the Transducer Spacing by pressing [Enter].

Note: The XMT868i automatically calculates the transducer spacing. You should not enter a different value unless specifically instructed by the factory.

The XMT868i returns to the Channel PROGRAM menu. You have completed entering PIPE parameters.

Note: The site data programmed in this chapter is automatically stored in the meter’s non-volatile RAM (NVR). If the power to the meter is disconnected, the same site data will be applied when the unit is next powered on.

DigitalFlow™ XMT868i Programming Manual

13

Chapter 1. Programming Site Data

1.3.4 Entering Input/Output Parameters

Enter the zero cutoff value and set up the temperature inputs using the I/O submenu. While programming these parameters, see Figure 30 on page 79.

IMPORTANT: If an option card in Slot 1 fails to appear in this menu, it may be turned OFF. See the GLOBL-I/O-OPTN section on page 29 for setup instructions.

1.3.4a Zero Cutoff Value

Near a zero flow rate, the Model XMT868i’s readings may fluctuate due to small offsets caused by thermal drift or similar factors. To force a zero display reading when there is minimal flow, enter a zero cutoff value as described in the following steps:

1.From the Channel PROGRAM menu, scroll to the I/O option in the center pane.

2.The program asks for the Zero Cutoff. Enter a value from 0 to 1 ft/sec (0 to 0.30 m/sec) for the zero cutoff and press

[Enter].

The recommended setting is 0.1 ft/sec (0.03 m/sec).

Proceed with one of the following:

If the ENERGY OPTION is OFF, the display now returns to the Channel menu. If the ENERGY OPTION is ON, refer to Temperature Input below.

1.3.4b Temperature Input

The Model XMT868i can use either a fixed temperature value or a live temperature input to calculate energy. Complete the following steps to configure this input:

Note: The site data programmed in this chapter is automatically stored in the meter’s non-volatile RAM (NVR). If the power to the meter is disconnected, the same site data will be applied when the unit is next powered on.

3.At the Temp Supply prompt, scroll to Fixed to enter a constant temperature value or on Slot1 to select the output in Slot 1 that will provide the live temperature input for the supply.

Note: If Slot 1 contains an activated option card with an analog input assigned to TEMP or an RTD input, Slot 1 appears as an option at the above prompt. If the process temperature is stable, a fixed value may be used, but most applications require a live temperature input.

14

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.4b Temperature Input (cont.)

4.One of the following two prompts will appear, depending on your selection in Step 3.

If Fixed was chosen, enter the known fixed process temperature and press [Enter]. The meter will only accept values from –328° to 1832°F (–200o to 1000oC).

If Slot 1 was chosen, scroll to the desired temperature input (from A to D) and press [Enter].

5.At the Temp Return prompt, scroll to Fixed to enter a constant temperature value or on Slot1 to select the output in Slot 1 that will provide the live temperature input for the return.

Note: If Slot 1 contains an activated option card with an analog input assigned to TEMP or an RTD input, Slot 1 appears as an option at the above prompt. If the process temperature is stable, a fixed value may be used, but most applications require a live temperature input.

6.One of the following two prompts will appear, depending on your selection in Step 3.

If Fixed was chosen, enter the known fixed process temperature and press [Enter]. The meter will only accept values from –328° to 1832°F (–200o to 1000oC).

If Slot 1 was chosen, scroll to the desired temperature input (from A to D) and press [Enter].

The program now returns to the Channel PROGRAM menu.

1.3.5 Entering Setup Parameters

The signal limits and response times for the Model XMT868i are specified using the SET UP submenu. While following the programming instructions, refer to Figure 30 on page 79. This submenu includes four options:

SIGNL — set the parameters related to the transducer signal

AVRG — specify the response of the meter to step changes

INIT — initialize all parameters to default values

ADVAN — enable mass flow, edit kinematic viscosity vs. sound speed table, activate K factors, and select transmit code length.

To enter the SET UP submenu, scroll to the SET UP option in the Channel PROGRAM menu and press [Enter]. Remember to record all programmed data in Appendix B, Data Records.

DigitalFlow™ XMT868i Programming Manual

15

Chapter 1. Programming Site Data

1.3.5 Entering Setup Parameters (cont.)

1.3.5a The SIGNL Option

Use this option to set the limits for the incoming signal and other parameters affecting the transducer signal. For example, the programmed signal strength low limit may be used to determine the trigger point for an alarm.

CAUTION! The SIGNL default settings are suitable for most applications. Consult the factory before changing any of these parameters.

1.From the SET UP option menu, scroll to SIGNL and press [Enter].

2.Press [Enter] to accept the current Signal Low Limit value or enter a new value and press [Enter].

The default value for the above parameter is 40, and values from –20 to 100 are acceptable. The E1: LOW SIGNAL error message appears when the signal strength falls below the programmed Signal Low Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.

3. Press [Enter] to accept the current Corr. Peak Limit value or enter a new value and press [Enter].

The default value for the above parameter is 100, and values from 0 to 500 are acceptable. The E4: SIGNAL QUALITY error message appears when the signal quality falls below the programmed Cor. Peak Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.

4. Press [Enter] to accept the current Sndspeed +– Lim. value or enter a new value and press [Enter].

The default value for the above parameter is 20%, and values from 1% to 50% are acceptable. The E2: SOUNDSPEED error message appears when the calculated fluid soundspeed differs from the fluid soundspeed entered in the CHx-SYSTM menu by more than the programmed Soundspeed +– Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.

5. Press [Enter] to accept the current Vel Low Limit value or enter a new value and press [Enter].

The default value for the above parameter is –150.0 ft/sec. (–46 m/sec) and values from –500 to 500 ft/sec (–150 to 150 m/sec) are acceptable. The E3: VELOCITY RANGE error message appears when the calculated fluid velocity is less than the programmed Velocity Low Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.

6. Press [Enter] to accept the current Vel High Limit value or enter a new value and press [Enter].

The default value for the above parameter is 40.0 ft/sec (12.142 m/sec) and –100 to 100 ft/sec (-30.48 to 30.48 m/sec) are acceptable values. The E3: VELOCITY RANGE error message appears when the calculated fluid velocity exceeds the programmed Vel High Limit value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.

16

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.5a The SIGNL Option (cont.)

7. Press [Enter] to accept the current Accel. Limit value or enter a new value and press [Enter].

The default value for the above parameter is 15.0 ft/sec2 (4.6 m/sec2) and values from 0 to 100 ft/sec2 (0 to 30 m/sec2) are acceptable. The E6: CYCLE SKIP error message appears when the calculated fluid velocity changes by more than the programmed Acceleration Limit value from one reading to the next. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.

8. Press [Enter] to accept the current Amp. Discrim Low value or enter a new value and press [Enter].

The amplitude discriminator measures the transducer signal received by the Model XMT868i. The default value for the above parameter is 14, and values from 0 to 100 are acceptable. The E5: AMPLITUDE error message appears when the amplitude discriminator falls below the programmed Amp. Discrim Low value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.

9. Press [Enter] to accept the current Amp. Discrim. High value or enter a new value and press [Enter].

The amplitude discriminator measures the transducer signal received by the Model XMT868i. The default value for the above parameter is 34, and values from 0 to 100 are acceptable. The E5: AMPLITUDE error message appears when the amplitude discriminator exceeds the programmed AMP. DISCRIM HIGH value. See Chapter 2, Error Codes, in the Service Manual for a discussion of error codes.

10. Press [Enter] to accept the current Delta T Offset value or enter a new value and press [Enter].

An offset between the upstream and downstream transit times is specified at this prompt. The default value for the above parameter is 0 sec, and values from –1000 to 1000 sec are acceptable.

11. Press [Enter] to accept the current % of Peak value or enter a new value and press [Enter].

The percentage of peak used to calculate the transit times and Delta T is specified at this prompt. The default value for the above parameter is -40%, and values from -100 to 100% are acceptable.

Note: This setting is a starting point for detecting the signal. The meter will automatically adjust this value if the calculated transit time is unacceptable. You can set the limits for this value using the MIN. PEAK% and MAX. PEAK% discussed on the next page.

12. Scroll to the desired XMIT Sample Size (from 2 to 32) and press [Enter]

Both the upstream and downstream transducers transmit ultrasonic pulses in bursts, which consist of a series of transmit pulses. XMIT Sample Size determines how many bursts are sent in one direction before sending in the other direction. The default value for the above parameter is 8 and values of 2, 4, 8, 16 and 32 are available.

13. Press [Enter] to accept the current # Errors Allowed value or enter a new value (0 to 16) and press [Enter].

Use this prompt to enter the number of errors the XMT868i can record before it displays an error message. The default value is 8.

DigitalFlow™ XMT868i Programming Manual

17

Chapter 1. Programming Site Data

1.3.5a The SIGNL Option (cont.)

14. Press [Enter] to accept the current Min Peak% Limit value or enter a new value and press [Enter].

Use this prompt to enter the minimum percent of peak that the XMT868i can use to measure transit time. The XMT868i accepts values from -100 to 100.

15. Press [Enter] to accept the current Max Peak% Limit value or enter a new value and press [Enter].

Use this prompt to enter the maximum percent of peak that the XMT868i can use to measure transit time. The XMT868i accepts values from -100 to 100.

After responding to the above prompt, the meter returns to the SET UP option window.

1.3.5b The AVRG Option

Use this option to specify the number of readings that occur before the meter will respond to a step change in flow rate. In general, the smaller the number of readings, the less steady the display will appear. Complete the following steps to set the response time:

1.From the SET UP option menu, scroll to the AVRG option and press [Enter].

2.Scroll to the desired Response Time (from 1 to 60 sec, or STATS (statistics)) and press [Enter].

For best results, select the STATS (statistics) option, as this increases the response time under steady flow conditions while still allowing a rapid response to changes in flow rate.

After responding to the above prompt, the meter returns to the SET UP option window.

1.3.5c The INIT Option

Use this option to initialize (reset) all of the parameters within the SIGNL menu back to their default values. Complete the following steps to reset all of the parameters:

1.From the SET UP option menu, scroll to INIT in the center pane.

2.Scroll to No to keep the current values or on Yes to reset all values to their default settings. After responding to the above prompt, the meter returns to the SET UP window.

1.3.5d The ADVAN Option

This option enables you to enable the more advanced features of the meter. In this option you can do the following:

calculate the kinematic viscosity (KV) based on signal strength or on soundspeed

enter a table of K-factors (based on velocity or reynolds number) that compensates for non-linear flow rates

enable mass flow (calculated for static fluid density)

select the size of the transducer transmission signal.

18

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.5e The KV/xx Option

Note: To select either signal strength or soundspeed for kinematic viscosity calculations, go to Reynolds Correction in the PIPE submenu on page 13.

Use this option to calculate the kinematic viscosity (KV) based on signal strength (SS) or on soundspeed, as selected in the KV Input Selection option of the PIPE option. To use this option, you must enter the KV vs. SS pairs in descending order of KV and ascending order of SS or soundspeed. For example, if pair 1 is KV = 10 centistrokes and SS = 62, then pair 2 must be KV 10 centistrokes and SS 62. The XMT868i accepts 2 to 20 pairs. Complete the following steps to enter KV and signal or soundspeed values:

1.From the ADVAN option menu, scroll to KV/xx and press [Enter].

2.Press [Enter] to accept the current # of KV/SS Pairs or enter a new value (2 to 20) and press [Enter].

3.Press [Enter] to accept the current Sig Strength (or Soundspeed) #01 value or enter a new value and press [Enter]. Enter the signal strength values in ascending order.

Note: For signal strength, the acceptable range is 50.0 to 85.0. For soundspeed, the acceptable range is 125 to 9000.

4.Press [Enter] to accept the current Kinematic Viscosity value or enter a new value (0.050 to 500) and press [Enter]. Enter the kinematic viscosity values in descending order.

The Sig Strength (or Soundspeed) and Kin Viscosity prompts repeat for each pair. After entering all the pairs, the meter returns to the ADVAN option.

DigitalFlow™ XMT868i Programming Manual

19

Chapter 1. Programming Site Data

1.3.5f The MULTK Factors Option

Use this option to enter a table of K-factors. K-factors are used to create a curve for the flow range (based on velocity or reynolds number) that compensates for non-linear flow rates. The meter accepts from 2 to 20 pairs. Complete the following steps to enter multiple K factors for velocity or reynolds values:

1.From the ADVAN option menu, scroll to MULTK and press [Enter].

2.Scroll to Yes to Activate Multi K or on No to disable this option, and press [Enter].

If NO was selected, the meter returns to the ADVAN window. If YES was selected, proceed to Step 3.

3.Scroll to the desired custom type: CstV (velocity) or CstR (reynolds)

4.Scroll to Yes to Edit the K-factor Table or on No to retain the current K-factor table (and return to the ADVAN option) and press [Enter].

Note: If the necessary velocity/reynolds vs. K-factor data was not provided with the Model XMT868i’s documentation, the K-factor table cannot be edited.

If NO was selected, the meter returns to the ADVAN option. If YES was selected, proceed to Step 5.

5. Enter the Number Of K-factors (from 2 to 20) to be entered into the table, and press [Enter].

Note: When editing the K-factor table, the velocities must be entered in increasing order.

6.Enter the Velocity/Reynolds Value for K-factor number “X” and press [Enter].

7.Enter the K-factor corresponding to velocity/reynolds number “X” (0.333 to 3.0) and press [Enter].

The Velocity # and K Factor # prompts repeat for each pair. After entering all the pairs, the meter returns to the ADVAN window.

20

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.3.5g The Mass Flow Option

Use this option to calculate mass flow from a static fluid density. Complete the following steps to enter the static density of the fluid:

1.From the ADVAN menu, scroll to MASS and press [Enter].

2.Scroll to Yes to activate Mass from Static or on No to disable this option.

a.If you select Yes, enter the Fluid Density (0.062 to 624.220 lb/ft3 or 0.001 to 9.9675g/cc).

b.For either option, press [Enter].

After responding to the above prompt, the meter returns to the ADVAN window.

1.3.5h The Code Length Option

Use this option to select the size of the transducer transmission signal. This option is helpful when measuring flow in small pipes. You can choose from the following selections:

AUTO — The meter will automatically determine the code length (short or long) which is based on pipe size and transducer type.

SHORT — is just few pulses with no code pattern built in. In some cases, the diameter of the pipe is so small that it

does not give each transducer the necessary time to send a series of signals before receiving a series of signals. SHORT is also a good selection for pipes made of materials such as PTFE which tend to blur the signal pattern.

LONG — is a series of approximately a dozen pulses that have a distinctive binary pattern, easily recognizable during auto correlation.

Complete the following steps to select the code length:

1.From the ADVAN option menu, scroll to CODEL and press [Enter].

2.Scroll to the desired Xmit Code Length and press [Enter]. AUTO is the default.

After responding to the above prompt, the meter returns to the ADVAN window. Press [Escape] three times to return to the Channel PROGRAM, or four times to return to the Keypad Program.

DigitalFlow™ XMT868i Programming Manual

21

Chapter 1. Programming Site Data

1.4Entering Data in the GLOBL Menu

The GLOBL menu is used to enter information that is not specific to any of the individual channels. Information programmed by using this menu is used to enter several general system parameters (e.g., English or metric units). For meters with 2 channels, this menu is also used to compute parameters such as the sum, difference or average of the channel 1 and channel 2 signals. When calculating the SUM, DIF or AVE readouts, data from the Global-SYSTM submenu is used. Any conflicting data entered in the Channel-SYSTM submenu is overridden.

The following submenus are included in the GLOBL menu:

SYSTM — use to specify the units of measure used in calculations

I/O — used to set up error handling and to configure analog inputs and outputs

COMM — used to set up the serial communications port and MODBUS parameters

To enter the GLOBL menu:

1.From the display screen, press [Escape], [Enter], [Escape] on the keypad.

2.Press [Enter] to enter the PROG menu.

3.In the PROG menu, scroll to GLOBL and press [Enter].

Then proceed to the appropriate section of this chapter for instructions. Refer to Figure 31 on page 80 and Figure 32 on page 81, and remember to record all programming data in Appendix B, Data Records.

1.4.1 Entering GLOBL-SYSTM Data

While completing these instructions, refer to Figure 31 on page 80.

1.From the PROG menu, scroll to GLOBL. Press [Enter].

2.Scroll to the SYSTM option. Press [Enter].

3.The first prompt asks for a Meter Message. Enter a numeric or text message (up to 16 characters) in the right pane and press [Enter].

4.Scroll to the desired System Units selection (either metric or English) and press [Enter]. The XMT868i will display all parameters and measurements in the designated units.

For a single-channel XMT868i, the program returns to the Global PROGRAM menu. For a 2-channel meter, proceed to step 5 on the next page.

22

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.4.1a Volumetric Units

5.Scroll to and scroll to the desired Volumetric Units for the flow rate display. Table 10 lists the available volumetric units.

Table 10: Available Volumetric Units

English

Metric

GAL/S = gallons/second

L/S = Liter/sec

GAL/M = gallons/minute

L/M = Liter/min

GAL/H = gallons/hour

L/H = Liter/hr

MGD = million gallons/day

ML/D = million liters/day

ft3/s = Cubic ft/sec

m^3/s = Cubic m/sec

ft3/m = Cubic ft/min

m^3/m = Cubic m/min

ft3/h = Cubic ft/hr

m^3/h = Cubic m/hr

ft3/d = Cubic ft/day

m^3/d = Cubic m/day

Mf3/d = Mega cubic feet/day

Mm^3/d =Mega cubic meters/day

BBL/S = Barrel/sec

BBL/S = Barrel/sec

BBL/M = Barrel/min

BBL/M = Barrel/min

BBL/H = Barrel/hr

BBL/H = Barrel/hr

BBL/D = Barrel/day

BBL/D = Barrel/day

MBL/D = Mega barrels/day

MBL/D = Mega barrels/day

A-I/S = Acre-inch/sec

A-I/M = Acre-inch/min

A-I/H = Acre-inch/hr

A-I/D = Acre-inch/day

A-F/S = Acre-feet/sec

A-F/M = Acre-feet/min

A-F/H = Acre-feet/hr

A-F/D = Acre-feet/day

6.Scroll to the desired number of Vol Decimal Digits (digits to the right of the decimal point) in the volumetric flow display, and press [Enter].

DigitalFlow™ XMT868i Programming Manual

23

Chapter 1. Programming Site Data

1.4.1b Totalizer Units

7. Scroll to the desired Totalizer Units for the totalized flow rate display (listed in Table 11), and press [Enter].

Table 11: Totalizer Units

English

Metric

GAL = Gallons

L = Liters

MGAL = Mega gallons

ML = Megaliters

ft^3 = Cubic feet

m^3 = Cubic meters

Mft^3 = Mega cubic feet

Mm^3= Mega cubic meters

BBL = Barrels

BBL = Barrels

MBBL = Mega barrels

MBBL = Mega barrels

AC-IN = Acre-inches

AC-FT = Acre-feet

8.Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the totalized flow display, and press [Enter].

The program now varies, depending on whether you have activated Mass Flow (see the Programming Manual for details) or the Energy Option in one of the Channel PROGRAM options.

If you have activated Mass Flow, continue to step 1 on the next page.

If you have activated the Energy Option (but not Mass Flow), skip to step 1 on page 1-30.

If you have activated neither option, the XMT868i returns to the Global PROGRAM window. Press [Escape] to return to the PROG menu.

24

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.4.1c Programming Mass Flow Data

1. Scroll to the desired Mass Flow Units for flow rate display (listed in Table 12), and press [Enter].

Table 12: Available Mass Flow Units

English

Metric

LB = Pounds

KG = Kilograms

KLB = KiloPounds (Thousands of Pounds)

Tonne = Metric Tons (1000 KG)

MMLB = MegaPounds

TONS (2000 LB)

2.Scroll to the desired Mass Flow Time Units (from seconds to days), and press [Enter].

3.Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the mass flow rate display, and press [Enter].

4.Scroll to the desired Mass Totals (units for the totalized mass flow rate display, listed in Table 12), and press [Enter].

5.Scroll to the desired number of Mass Decimal Digits (digits to the right of the decimal point) in the totalized mass flow rate display, and press [Enter].

The program varies, depending on whether or not you have enabled the Energy Option.

If you have not enabled the option, the XMT868i returns to the Global PROGRAM window. Press [Escape] to return to the PROG menu.

If you have enabled the option, continue to the next page.

DigitalFlow™ XMT868i Programming Manual

25

Chapter 1. Programming Site Data

1.4.1d Programming the Energy Option

1. Scroll to the desired Power Units (listed in Table 13) and press [Enter].

Table 13: Available Power Units

English

Metric

kBTUh = Thousands of BTUs/hour

kCALs = Kilocalories/second

MMBTU = Mega BTUs/hour

kCALm = Kilocalories/minute

kW = Kilowatt

kCALh = Kilocalories/hour

TONS = Tons (2000 LB)

kCALd = Kilocalories/day

MCals = Megacalories/second

MCALm = Megacalories/minute

MCALh = Megacalories/hour

MCALd = Megacalories/day

kW = Kilowatt

MW = Megawatt

2.Scroll to the desired number of Power Decimal Digits, and press [Enter].

3.Scroll to the desired Total Energy Units (shown in Table 14), and press [Enter].

Table 14: Available Total Energy Units

English

Metric

kBTU = Thousands of BTUs

kCAL = Kilocalories

MMBTU = Mega BTUs/hour

MCAL = Megacalories

kWHr = Kilowatt-hour

kWHr = Kilowatt-hour

TONS = Tons (2000 LB)

MWHr = Megawatt-hour

4. Scroll to the desired number of Energy Decimal Digits, and press [Enter].

You have completed entering selections in this option.The program returns to the Global PROGRAM menu.

26

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.4.2 Setting Up Inputs and Outputs

Set up the XMT868i’s inputs and outputs using the I/O submenu. While following the programming instructions, refer to Figure 31 on page 80 and Figure 32 on page 81. Remember to record all programmed data in Appendix B, Data Records.The I/O submenu consists of the following options:

Error Handling — program the meter’s response during an error condition

Options — set up any option cards and the Slot 0 analog outputs

LCD — set up the LCD display (refer to Chapter 2, Displaying Data)

To enter the I/O submenu:

1.From the PROGRAM menu, scroll to GLOBL and press [Enter].

2.Scroll to I/O and press [Enter].

Note: In this section, Slot 1 appears as an option only if a suitable option card is installed in Slot 1.

1.4.2a Setting Up Error Handling

This menu option lets you set how the XMT868i will handle the outputs for measurements and average (two-path) measurements during an error condition. See Chapter 2, Error Codes, in the Service Manual for a discussion of the built-in error codes.

1.In the Global I/O option, scroll to ERROR and press [Enter].

2.Scroll to the desired Error Handling selection (as shown in Table 15 and Table 16 on page 28).

a.If you selected Other, enter the 4-20 mA Error Level (the number of milliamps the analog output will put out during a fault condition). Enter a whole number between 0 and 22, and press [Enter].

See Table 15 and Table 16 on page 28 for a description of the error handling options available and how the totalizers and display respond to them for a single and two-channel meter.

Table 15: Error Options and Responses, Single-Channel Meter

Option

Output Response

Totalizer Response

HOLD

Holds the last “good” reading

Holds the last “good” reading and continues to

totalize, based on that reading

LOW

Forces the outputs to the low set point

Stops totalizing

HIGH

Forces the outputs to the high set point

Stops totalizing

HHIGH

Forces the outputs 10% above the high set point

Stops totalizing

OTHER

Allows user to enter 4-20 mA error level in mA.

Sends analog outputs to entered mA level

when a fault occurs.

DigitalFlow™ XMT868i Programming Manual

27

Chapter 1. Programming Site Data

1.4.2 Setting Up Inputs and Outputs (cont.)

1.4.2a Setting Up Error Handling (cont.)

Table 16: Error Options and Responses for a 2-Channel Meter

Totalizer Response When Error Handling is

When Measuring

Display Response

HOLD

LOW, HIGH, HHIGH

CH1 or CH2

Holds last “good” reading.

Holds last “good” reading and continues

Stops totalizing.

(vel, vol, etc.)

to totalize based on that “good” reading.

SUM

Adds two channels using

Holds last “good” reading and continues

Stops totalizing if either or

the last “good” reading.

to totalize based on two channels.

both channels go into error.

DIF

Subtracts two channels

Holds last “good” reading and continues

Stops totalizing if either or

using last “good” reading.

to totalize based on two channels.

both channels go into error.

AVE

See Error Handling for Average Measurements below.

For a one-channel meter, after responding to the above prompt, the meter returns to the Global I/O prompt shown on the previous page. For a two-channel meter, proceed to step 3 below.

Error Handling for Average Measurements — AVE

The 2PATH ERROR HANDLING option is intended for applications where two sets of transducers are installed in the same location in the same pipe to improve accuracy and the meter is operated in AVE mode. With this function enabled, the Model XMT868i performs error handling only if both channels are in error. If this function is disabled, error handling occurs when either channel goes into error.

3. Scroll to Yes to enable Two-path Error Handling, or on No to disable this function, and press [Enter].

Specific responses of the display and the totalizer to the two-path error handling option available at the above prompt are listed in Table 17.

Table 17: 2-Path Error Response Options

Option

Display Response

Totalizer Response

NO

Displays the average of CH1 and CH2, regardless

Outputs the average of CH1 and CH2 totals,

of the error state of either channel.

regardless of the error state of either channel.

YES

1.

If one channel is in error, the other channel’s

1.

If one channel is in error, totalizing continues.

value is displayed as the average.

2.

If both channels are in error, totalizing stops.

2.

If both channels are in error, the last average

reading is held.

After responding to the above prompt, the meter returns to the Global I/O option.

28

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.4.2b Setting Up Option Cards

The Model XMT868i has two built-in analog outputs, which are assigned to Slot 0. Also, a variety of input/output option cards may be installed in Slot 1. See Chapter 1, Installation, of the Startup Guide for a complete description of the available option cards.

To access this submenu, scroll to OPTN in the I/O option. To set up and/or scale the inputs and outputs, complete the following steps:

Scroll to the desired I/O Slot to program the I/O functions in that slot, and press [Enter].

Note: If an option card is not installed in Slot 1, the Slot1 option does not appear at the above prompt.

Proceed to the appropriate section for programming instructions specific to the type of input or output selected above.

Analog Outputs

Complete the following steps to set up the analog outputs for any slot (refer to Figure 31 on page 80):

1. Scroll to the desired Output (A or B), and press [Enter].

Note: The set up of output A is used here as an example. Identical procedures would be used to set up output B.

2.Scroll to Off to disable Output A and return to step 1, or on 0-20 mA or 4-20 mA to specify the desired range for output A. In either case, press [Enter].

Note: For a 1-Channel meter, proceed to step 4.

3.Scroll to the desired Channel option and press [Enter]. See Table 18 for a description of the channel options available.

Table 18: Channel Options

Option

Description

CH1

Channel 1

CH2

Channel 2

SUM

CH1+CH2

DIF

CH1-CH2

AVE

(CH1+CH2)/2

4.Scroll to the desired Measurement Name and press [Enter]. (Table 19 on page 30 provides a description of the available parameters.)

DigitalFlow™ XMT868i Programming Manual

29

Chapter 1. Programming Site Data

1.4.2b Setting Up Option Cards (cont.)

Analog Outputs (cont.)

Note: All the options shown below will only appear if the Mass Flow and Energy Options are activated.

Table 19: Output Measurement Options

Option Bar Choice

Description

VEL

Flow Velocity

VOLUM

Volumetric Flow

+TOTL

Forward Totalized Volume Flow

-TOTL

Reverse Totalized Volume Flow

TIME

Total Flow Measurement Time

MDOT

Mass Flow

+MASS

Forward Totalized Mass Flow

-MASS

Reverse Totalized Mass Flow

POWER

Energy Flow Power

+ENRG

Forward Energy Flow

-ENRG

Reverse Energy Flow

SS up

Signal strength for the upstream transducer.

SS do

Signal strength for the downstream transducer.

DAC up

AGC DAC count for upstream gain setting.

DAC do

AGC DAC count for downstream gain setting.

Soundspeed

Measured speed of sound in the fluid.

T up

Upstream ultrasonic signal transit time.

Tdown

Downstream ultrasonic signal transit time.

DELTAT

Transit time difference between upstream and downstream signals.

REYN#

Current velocity/reynolds number.

K (RE)

K factor, based on the Reynolds number.

PEAK%

Percentage of peak (set to +50 by default).

THETA3

Theta 3 is the angle between the ultrasonic beam and the normal to the pipe wall. It is

calculated from the measured transit time and clamp-on parameters (clamp-ons only).

Qup

Displays the signal quality for the upstream transducer.

Qdown

Displays the signal quality for the downstream transducer.

AMPup

Displays the value for the signal amplitude of the upstream transducer.

AMPdn

Displays the value for the signal amplitude of the downstream transducer.

P#up

Displays signal peaks for the upstream transducer.

P#dn

Displays signal peaks for the downstream transducer.

30

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

Table 19: Output Measurement Options (cont.)

Option Bar Choice

Description

NFup

Displays the normalization factor for the upstream transducer.

NFdn

Displays the normalization factor for the downstream transducer.

Cxdcr

Displays the speed of sound in the transducer (clamp-on only).

TEMPs

Displays the temperature for the supply input (energy measurement).

TEMPr

Displays the temperature for the return input (energy measurement).

Ts-Tr

Difference between supply input and return input measurement (energy measurement).

DENSs

Displays the density for the supply input (energy measurement).

DENSr

Displays the density for the return input (energy measurement).

DELTh

Displays the delta enthalpy which is the difference between the supply and return.

Enthalpy is a measure of energy contained in the fluid (energy measurement).

Note: The measurement units that appear in these prompts are those selected in the Global-System menu earlier in this section.

5.Enter a measurement value for the Base (low) end of the analog output range and press [Enter].

6.Enter a measurement value for the Full (high) end of the analog output range and press [Enter]. After responding to the above prompt, the meter returns to the I/O option.

DigitalFlow™ XMT868i Programming Manual

31

Chapter 1. Programming Site Data

1.4.2b Setting Up Option Cards (cont.)

Option Card Analog Inputs

Complete the following steps to set up the analog inputs of an option card installed in Slot 1 (refer to Figure 32 on page 81):

1. Scroll to the desired Input (A, B, C or D) and press [Enter].

Note: The setup of Input A is used as an example. Identical procedures would be used to set up the remaining inputs.

2.Enter a Label of up to eight characters for input A and press [Enter].

3.Scroll to Off to disable Input A and return to step 1, or on Temp or Spec to designate it as a temperature or special input, and press [Enter].

OFF

If OFF was selected to disable input A, the meter returns to the Input prompt in step 1. Scroll to another input, or press [Escape] to move to another menu.

TEMP

If TEMP was selected to set up input A as a live temperature input, complete steps 4 and 5 below:

4.Enter a temperature value for the Base (low) end of the analog input range and press [Enter].

5.Enter a temperature value for the Full Scale (high) end of the analog input range and press [Enter].

After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.

SPEC

If SPEC was selected to set up Input A as a live special input, complete steps 6 through 10 below:

6.Enter a Name for input A and press [Enter].

7.Enter the Units of measurement for Input A and press [Enter].

8.Enter a value for the Base (low) end of the analog input range and press [Enter].

9.Enter a value for the Full Scale (high) end of the analog input range and press [Enter].

After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.

32

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

Option Card RTD Inputs

Option cards with RTD inputs have a temperature range of -148o to 660oF (–100° to 350°C). Complete the following steps to set up the RTD inputs of an option card installed in Slot 1 (refer to Figure 32 on page 81):

1. Scroll to the desired Input (A, B, C or D) and press [Enter].

Note: The setup of RTD input A is used here as an example. Identical procedures would be used to set up the remaining RTD inputs.

2.Enter a Label of up to eight characters for Input A and press [Enter].

3.Scroll to Off to disable Input A and return to step 1, or on Temp to enable Input A as a live temperature input, and press [Enter].

4.Enter a temperature value for the Base (low) end of the analog input range and press [Enter].

5.Enter a temperature value for the Full (high) end of the analog input range and press [Enter].

After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.

DigitalFlow™ XMT868i Programming Manual

33

Chapter 1. Programming Site Data

1.4.2b Setting Up Option Cards (cont.)

Option Card Alarm Relays

Complete the following steps to set up the alarm relays of an option card installed in Slot 1 (refer to Figure 32 on page 81):

1. Scroll to the desired Output (A, B, C or D), and press [Enter].

Note: The setup of Alarm A is used here as an example. Identical procedures would be used to set up the additional alarms.

2.Scroll to Off to disable Output A and return to step 1, or on High, Low or Fault to select the desired alarm type. Press [Enter].

3.Scroll to No for standard alarm operation, or on Yes for Failsafe operation and press [Enter]. (See Chapter 1, Installation, of the Startup Guide for failsafe wiring instructions.)

For a 1-Channel meter, proceed to step 5.

4.Scroll to the desired Channel option, and press [Enter]. See Table 20 for a description of the channel options available.

Table 20: Channel Options

Option

Description

CH1

Channel 1

CH2

Channel 2

SUM

CH1+CH2

DIF

CH1-CH2

AVE

(CH1+CH2)/2

Note: If FAULT was selected as the alarm type, the next two prompts do not appear. You have completed entering data in this option.

5.Scroll to the desired Measurement Name, and press [Enter]. (Table 19 on page 30 provides a description of the available parameters.)

Note: The measurement units that appear in these prompts are those selected in the Global-SYSTM menu earlier in this section.

6. Enter a value for the Trigger Point of the alarm and press [Enter].

After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.

34

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

Option Card Totalizer Outputs

This type of output issues one pulse per selected volume of flow. The meter produces a pulse each time the programmed amount of flow passes through the pipe. Complete the following steps to set up the totalizer outputs of an option card installed in Slot 1:

1. Scroll to the desired Output (A, B, C or D), and press [Enter].

Note: The set up of output A is used as an example. Identical procedures would be used to set up the other outputs.

2.Scroll to Off to disable Output A and return to step 1, or on TTLZR to set up Output A as a totalizer output. Press

[Enter].

For a 1-Channel meter, skip to step 4.

3.Scroll to the desired Channel option, and press [Enter]. See Table 20 on page 34 for a description of the channel options available at the above prompt.

4.Scroll to the desired Measurement Name, and press [Enter]. (Table 19 on page 30 provides a description of the available parameters.)

Note: The measurement units that appear in these prompts are those selected in the GLOBL-SYSTM menu earlier in this section.

5.Enter a value between 50 sec and 500,000 sec for Min. Pulse On Time (the frequency of the totalizer pulses) and press [Enter].

Note: A complete pulse consists of equal amounts of ON and OFF times. Choose a value that is compatible with the counter to be used.

6. Enter a value for the number of measurement units represented by each pulse, and press [Enter].

After responding to the above prompt, the meter returns to step 1. Press [Escape] to proceed to another menu or option.

DigitalFlow™ XMT868i Programming Manual

35

Chapter 1. Programming Site Data

Option Card Frequency Outputs

This type of output produces a frequency pulse that is proportional to the output measurement. Complete the following steps to set up the frequency outputs of an option card installed in Slot 1 (see Figure 32 on page 81):

1. Scroll to the desired Output (A, B, C or D), and press [Enter].

Note: The set up of Output A is used as an example. Identical procedures would be used to set up the other outputs.

2.Scroll to Off to disable Output A and return to step 1, or on Frequency to set up Output A as a frequency output. Press [Enter].

For a 1-Channel meter, skip to step 4.

3.Scroll to the desired Channel option. See Table 20 on page 34 for a description of the channel options available at the above prompt.

4.Scroll to the desired Measurement Name. (Table 19 on page 30 provides a description of the available parameters.)

Note: The measurement units that appear in these prompts are those selected in the GLOBL-SYSTM menu earlier in this section.

5.Enter a flow rate value for the Base (low) end of the frequency output range and press [Enter].

6.Enter a flow rate value for the Full (high) end of the frequency output range and press [Enter].

7.Enter a value between 1 and 10,000 for the Full Scale Frequency and press [Enter].

After responding to the above prompt, the meter returns to step 1. Press [Escape]:

once to return to the Global I/O option:

twice to return to the Global PROGRAM option: three times to return to the PROG menu:

four times to return to the Keypad Program menu:

five times to exit the Keypad Program and return to the XMT868i display.

36

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.4.3 Entering Communications Data

The Model XMT868i flowmeter is equipped with an RS232 or an RS485 serial interface. MODBUS (RS485), MODBUS/TCP, Ethernet and Foundation-Bus options are available. When the MODBUS, MODBUS/TCP or Foundation Field Bus options are present, the XMT868i may also have the standard RS232/RS485 serial interface. When the Ethernet option is present, the RS232/RS485 serial interface is not present.

Note: To set Fieldbus communications parameters, see Appendix E, Foundation Fieldbus Communications.

The serial port is used to transmit stored data and displayed readings to a personal computer by connecting the meter’s serial interface to the serial port of the PC. In addition, the Model XMT868i can receive and execute remote commands, using PanaView software, using this link.

Use the COMM option to set the communications port and MODBUS parameters. While following the programming instructions, refer to Figure 31 on page 80.

1.4.3a Setting Up the Serial Port

1.To enter this submenu from the Global PROGRAM menu, scroll to COMM and press [Enter].

2.Enter a Meter Address number between 1 and 254 and press [Enter]. The default number is 1.

A meter address is only necessary for communication with the GE PanaView software. See the PanaView User’s Manual for more information.

IMPORTANT: If the meter address or baud rate is changed, communication with PanaView must be re-established with the new address number.

3. Scroll to the desired Baud Rate (300 to 19,200) and press [Enter].

IMPORTANT: The Ethernet option requires that the meter’s baud rate is set to the same rate as the Ethernet option. The factory default is 9600 baud. If a different baud rate is desired, first change the meter’s rate, and then change the Ethernet’s baud rate setting by using the Device Discovery utility.

If you have the RS485 MODBUS option, proceed to step 4 on the next page.

If you have the standard RS232 serial interface or optional Ethernet interface, the meter exits the COMM submenu and returns to the Global PROGRAM prompt.

DigitalFlow™ XMT868i Programming Manual

37

Chapter 1. Programming Site Data

1.4.3b Setting Up MODBUS (RS485) and MODBUS/TCP Communications

When equipped with an optional MODBUS output card, the XMT868i can transmit flow data serially to a flow computer, or SCADA, using a Gould-type RTU protocol. Thus only the MODBUS function command, 3 (read multiple registers), 6 (write multiple registers) is valid. The XMT868i uses the following format for the data exchange:

The Send command (initiated by the host flow computer or controller) comes in the form: [time delimiter]<Addr><3><First Register MSB>

<First Register LSB> <Register Count MSB>

<Register Count LSB><CRC Low><CRC High>[time delimiter]

The Response (initiated by the host flow computer or controller) comes in the form: [time delimiter]<Addr><3><Byte count><Data………..>

<CRC Low><CRC High>[time delimiter]

The format for the returned data types is:

Integer (16 bit Integer) <MSB><LSB>

1Register — 16 bit integer

Integer (32 bit IntegerI) <MSB><LSB><LSB><LSB>

2Register — 32 bit integer

Floating Point (FP)<EXP><MAN><MAN><MAN>

2Registers — 32 bit IEEE floating point number

Use the steps below to configure MODBUS communications:

IMPORTANT: The XMT868i MODBUS communications settings chosen in the next four steps must match those of the MODBUS control system.

4. Scroll to the desired MODBUS Baud Rate and press [Enter].

IMPORTANT: The MODBUS./TCP interface requires a baud rate of 9600.

5.Scroll to the MODBUS Parity and press [Enter].

6.Scroll to the MODBUS Stop Bits and press [Enter].

7.Enter a MODBUS Meter Address (between 1 and 254) and press [Enter]. (The default address is 1.)

The meter returns to the Global PROGRAM prompt. Press [Escape] twice to return to the Keypad Program, and a third time to return to the meter display.

IMPORTANT: You must reboot the XMT868i to load the new settings.

Refer to the section Requesting Parameters Using MODBUS on page 40 to retrieve data from the XMT868i using MODBUS.

38

DigitalFlow™ XMT868i Programming Manual

Chapter 1. Programming Site Data

1.4.3c Setting Up a MODBUS/TCP Connection

A modified XMT868i can use the MODBUS/TCP interface to communicate with an internal network. An optional MODBUS/TCP card with a unique MAC (IP) address (installed only in slot 2) includes an RJ45 connector. To connect the MODBUS/TCP-enabled XMT868i to the network, insert the jack of an RJ45 cable into the RJ45 connector, route the cable through a conduit hole with an appropriate cable clamp, and wire the other end of the cable to the LAN according to the manufacturer’s instructions.

To establish MODBUS/TCP communications with the XMT868i, you must install the Ruiping software utility (available with your modified XMT868i) on a PC connected to the LAN. Once installed and running, the software displays all Ethernet devices currently connected to the subnet. You can identify the XMT868i by its MAC address, supplied as part of customer documentation. The default TCP Port is 502.

The default IP addressing for the XMT868i is DHCP (dynamic). If you need to assign a static IP address to the XMT868i, complete the following steps:

IMPORTANT: To assign a static IP address, the XMT868i must be connected to an Ethernet network with a dynamic address assignment mechanism (DHCP) first.

1. Run Ruiping software, and identify the XMT868i’s current IP address.

Note: “Ruiping-e” will return the found IP addresses as well as the MAC address.

2.From a DOS command prompt, type: telnet “ip address” 10000

3.The “Welcome to Net+Works Configuration Utility” opens. Type in the login name and password.

a.Login: type root.

b.Password: type Netsilicon.

4.In the Main Menu:

a.Select IP Parameters (1). Enter the desired IP changes and return to the Main Menu.

b.Select Enable DHCP Client (3). Select N (2) to disable, then select 4 for Main Menu.

c.At the Main Menu, select 4 for Quit. Power cycle the XMT868i (i.e., turn the meter off and on) to apply the changes.

Refer to the section Requesting Parameters Using MODBUS on page 40 to retrieve data from the XMT868i using MODBUS.

DigitalFlow™ XMT868i Programming Manual

39

Chapter 1. Programming Site Data

1.4.3d Requesting Parameters Using MODBUS

To request specific parameters from the XMT868i using the MODBUS, the control system must access the appropriate register number, as shown in Table 21. Only registers 1–84 are available with the XMT868i for MODBUS communications. Registers 508–512 are used by the XMT868i to store the MODBUS parameters.

:

Table 21: MODBUS Registers

MODBUS

DPR Hex

Scaling

Reg #

Addr

Description

Units

(decimal places)

Size in Bytes

1

1Clear Totalizers

none

2 (16 bit signed)

2

CH1 Velocity

ft/s or m/s

2

4 (32 bit integer)

4

CH1 Volumetric

VOL_U

4 (IEEE 32 bit)

6

CH1 +Totals

TOT_U

Register 10

4 (32 bit integer)

8

CH1 -Totals

TOT_U

Register 10

4 (32 bit integer)

10

CH1 #T Digits

none

0

2 (16 bit integer)

11

CH1 Totalizer Time

sec

2

4 (32 bit integer)

13

2CH1 Error Value

none

0

2 (16 bit integer)

14

CH 1 SSUP

none

1

4 (32 bit integer)

16

CH 1 SSDN

none

1

4 (32 bit integer)

18

CH 1 SNDSP

ft/s or m/s

0

4 (32 bit integer)

20

CH 2 Velocity

ft/s or m/s

2

4 (32 bit integer)

22

CH 2 Volumetric

VOL_U

4 (IEEE 32 bit)

24

CH 2 +Totals

TOT_U

Register 28

4 (32 bit integer)

26

CH 2 -Totals

TOT_U

Register 28

4 (32 bit integer)

28

CH2 # T Digits

none

0

2 (16 bit integer)

29

CH2 Totalizer Time

sec

2

4 (32 bit integer)

31

2CH2 Error Value

none

0

2 (16 bit integer)

32

CH 2 SSUP

none

1

4 (32 bit integer)

34

CH 2 SSDN

none

1

4 (32 bit integer)

36

CH 2 SNDSP

ft/s or m/s

0

4 (32 bit integer)

38

3AVG Velocity

ft/s or m/s

2

4 (32 bit integer)

40

3AVG Volumetric

VOL_U

4 (IEEE 32 bit)

42

3AVG+Totals

TOT_U

Register 46

4 (32 bit integer)

40

DigitalFlow™ XMT868i Programming Manual

Loading…

Panametrics DigitalFlow XMT868i: List of Available Documents

Note for Owners:

Guidesimo.com webproject is not a service center of Panametrics trademark and does not carries out works for diagnosis and repair of faulty Panametrics DigitalFlow XMT868i equipment. For quality services, please contact an official service center of Panametrics company. On our website you can read and download documentation for your Panametrics DigitalFlow XMT868i device for free and familiarize yourself with the technical specifications of device.

More Measuring Instruments Devices:

  • Theben SYN 161 h

    쮕쮕MONTAGEHINWEISESicherheitshinweis: Der Anschluss und die Montage elektrischer Gerätedürfen nur durch eine Elektrofachkraft erfolgen. Eingriffe und Veränderun-gen am Gerät führen zum Erlöschen des Garantieanspruches. Die natio-nalen Vorschriften und die jeweils gültigen Sicherheitsbestimmungen sindzu beachten.INSTRUCTIONS DE MONTAGERèglements de sécurité: Le branchement et le …

    SYN 161 h Timer, 2

  • Texas T Parts ECHO-F1

    Page 1Texas Model A PartsParts & Accessories forModel T Fords1-800-337-6977Texas T PartsTexas T Parts8427 Turkey Creek RoadCollege Station, Texas 77845(979) 260-5433Fax (979) 260-1214www.TexasTParts.comDIGITAL SPEEDOMETER KIT INSTALLATIONINSTRUCTIONS8 Function Digital Speedometer• Speed Functions — 0-120.0 M/H or K/HCurrent SpeedMaximum SpeedAverage Speed• Distance FunctionsOdometer (Up …

    ECHO-F1 Measuring Instruments, 4

  • Zenner zelsius C5-ISF

    Installation and operating manual ……………18Electronic compact heating and cooling meter with single-jet flow sensor ISFM-Bus, wM-Bus and 3 inputs/outputs optionalqp 0,6/1,5/2,5 m³/hManuel de montage et d’installation …………… 34Compteur électronique d’énergie thermique et de frigorie avec mesureur à jet unique (ISF)M-Bus, wM-Bus et 3 entrées/sorties en optionqp 0,6/1,5/2,5 …

    zelsius C5-ISF Measuring Instruments, 84

  • XICOY CG Meter 1.4

    1 CG & Angle Meter 1.4 Users Guide. Plaça Pere Llauger, nau 18 08360, Canet de Mar, Barcelona, Catalonia, Spain E-mail: [email protected]  +34 93 794 27 74 web: www.xicoy.com Xicoy WEEE register number: ES004749 & DE 36558999 © Copyright 2015, Xicoy Electronica SL. All Rights Reserved Manual contents & design: Gaspar Espiell. V1.4 …

    CG Meter 1.4 Measuring Instruments, 9

Recommended Documentation:

Назначение
Описание
Знак утверждения типа
Поверка
Рекомендации к применению

Назначение

Расходомеры-счетчики жидкости ультразвуковые мод. AT868, DF868, XMT868i, PT878, SEN898 предназначены для измерения скорости, объемного расхода и объема жидкостей.

Описание

Расходомеры-счетчики жидкости ультразвуковые мод. AT868, DF868, XMT868i, PT878, SEN898 (далее — расходомеры) состоят из ультразвуковых преобразователей (далее -УП), установленных на измерительном участке трубопровода (далее — ИТ) и электронновычислительного блока (далее — ЭВБ). УП имеют врезное и накладное исполнение. Монтаж УП может осуществляться на ИТ или на специально изготовленный измерительный участок.

В расходомерах реализован время-импульсный метод измерения, при котором пара УП по очереди посылает и принимает кодированные ультразвуковые сигналы. Принцип время импульсного метода измерения заключается в измерении разницы времени прохождения аккустического сигнала по направлению потока и против него. По измеренной разности времен посредством ЭВБ определяется скорость потока жидкости. Значение объемного рахода жидкости расчитывается по скорости потока при известных значениях параметров ИТ (диаметр ИТ, толщина ИТ и т.д.).

Расходомер мод. AT868 (далее — AT868) предназначен для измерения скорости, объемного расхода и объема горячей и холодной воды, воды систем поддержания пластового давления, водонефтяных смесей. ЭВБ AT868 выполнен во влагозащищенном корпусе. AT868 работает в комплекте с врезными или накладными УП.

Расходомер мод. DF868 (далее — DF868) предназначен для измерения скорости, объемного расхода и объема горячей и холодной воды, морской воды, нефти и нефтепродуктов. DF868 работает в комплекте с врезными или накладными УП.

Расходомер мод. XMT868i (далее — XMT868i) предназначен для измерения скорости, объемного расхода и объема горячей и холодной воды, морской воды, сточных вод, нефти и нефтепродуктов, жидких углеводородов, сжиженного природного газа. XMT868i имеет компактный ЭВБ с инфракрасной клавиатурой для конфигурации во взрывоопасных зонах, который может быть смонтирован на измерительном участке трубопровода. XMT868i работает в комплекте с врезными или накладными УП.

Расходомер мод. PT878 (далее — PT878) предназначен для измерения скорости, объемного расхода и объема чистых жидкостей. PT878 имеет портативный ЭВБ, работающий от аккумуляторных батарей. Связь ЭВБ PT878 с инженерным компьютером осуществляется по инфракрасному порту. PT878 работает в комплекте с врезными или накладными УП.

Расходомер мод. SEN898 (далее — SEN898) предназначен для измерения скорости, объемного расхода жидких углеводородов, нефти и нефтепродуктов (исполнение LCT), сжиженного природного газа, криогенных жидкостей (исполнение LNG). SEN898 состоит из ЭВБ и четырех пар врезных УП, установленных на специально изготовленном измерительном участок. Для непосредственного конфигурирования SEN898 используется магнитный стилус.

ЭВБ расходомеров обеспечивают выполнение следующих функций:

—    цифровая обработка сигналов поступающих с УП;

—    измерения и преобразование аналоговых сигналов постоянного тока;

—    хранение измерительной информации и настроечных параметров расходомера;

—    отображение измерительной информации и настроечных параметров расходомера;

—    передача измерительной информации по аналоговым и различным цифровым интерфейсам;

—    защита от преднамеренных и непреднамеренных изменений и несанкционированного доступа.

В комплекте с расходомерами возможна поставка программного пакета PanaView, для установки на операторские и инженерные станции с установленной операционной системой Windows. Программный пакет PanaView позволяем производить следующие операции:

—    загружать в расходомер и сохранять из расходомера конфигурационные данные;

—    формировать протоколы и графики, основываясь на измерительной информации хранящейся в расходомере;

—    отображать и строить графики по текущим значениям измеряемых параметров.

Программный пакет PanaView реализует протоколы связи IDM и PanaLink и поддерживает следующие коммуникационные интерфейсы связи:

—    Последовательный интерфейс (RS232, RS485)

—    Ethernet

—    Инфра красный интерфейс связи (далее — ИК)

Программное обеспечение. Расходомеры имеют встроенное программное обеспечение (далее — ПО). ПО расходомеров имеет уровень защиты С по МИ 3286-2010. Защита ПО расходомеров от несанкционированного доступа с целью изменения параметров, влияющих на метрологические характеристики, осуществляется путем аутентификации (введением пароля администратора), ведения доступного только для чтения журнала событий и ошибок. Возможность внесения преднамеренных и непреднамеренных изменений в ПО расходомеров исключается: наличием в расходомерах функции определения целостности ПО при включении и ограничением свободного доступа к цифровым интерфейсам связи.

Идентификация ПО расходомеров осуществляется путем отображения на дисплее расходомера или подключенного к нему инженерного персонального компьютера структуры идентификационных данных, содержащей номер версии и цифровой идентификатор ПО расходомера. Идентификационные данные ПО расходомеров приведены в таблице 1

Таблица 1 — Идентификационные данные ПО расходомеров

Модель

расходомера

Идентификационное наименование ПО

Номер

версии

ПО

Цифровой

идентификатор

ПО

Алгоритм вычисления цифрового идентификатора ПО

AT868

AT868.BIN

AT2F

5296

CRC-16

DF868

DF868.BIN

F2AJ

02DB

CRC-16

XMT868i

XMT868I.BIN

X3G

D6A4

CRC-16

XMT868HRT.BIN

X2M.

HRT

F32D

CRC-16

PT878

BOOT.004.2

2F

SEN898

IMG878 DSP 001 D.BIN

DSP.1.D

A6E1FAA9

CRC-32

Таблица 2 — Метрологические и технические характеристики AT868 и PT878

2-4.

Наименование характеристики

ф

и

д

о

М

икация

AT868

PT878

Диапазон измерений скорости, м/с (2 направления)

от 0,3 до 12,2

Максимальное количество каналов измерения расхода

2

1

Внешний диаметр ИТ, мм

—    при использовании накладных УП

—    при использовании врезных УП

от 12,7 до 7600 от 25,4 до 5000

от 12,7 до 7600

Толщина стенки ИТ, не более, мм

76,2

Температура измеряемой среды, °С

—    при использовании накладных УП

—    при использовании врезных УП

от минус 40 до 150 от минус 200 до 2301) от минус 40 до 100

от минус 40 до 150 от минус 190до 3001)

Максимальное давление измеряемой среды при использовании врезных УП, МПа

20

Пределы допускаемой относительной погрешности при измерении скорости, объемного расхода и объема, %

—    1 канальное исполнение

■    при поверке проливным методом

■    при поверке имитационным методом

—    2 канальное исполнение

■    при поверке проливным методом

■    при поверке имитационным методом

D2)<0,15 м

D>0,15 м

D<0,15 м

D>0,15 м

±2

±4

±1

±2

±1

±2

±0,5

±1

±2

±4

±1

±2

Температура окружающей среды для ЭВБ °С:

от минус

0 до 55

от минус 20 до 55

Температура хранения для ЭВБ, °С

от минус 40 до 70

от минус 40 до 70

Максимальная длина кабеля от УП к ЭВБ, м

305

Варианты выходного сигнала

аналоговый от 0/4-до 20 мА импульсный/ частотный

аналоговый от 0/4 до 20 мА импульсный/ частотный

Пределы абсолютной погрешности канала аналогового вывода, мА

±0,016

±0,016

Коммуникационный цифровой интерфейс

RS-232

RS-4851)

ИК

Поддерживаемые протоколы связи

IDM

PanaLink, IDM

Длина прямолинейного участка ДО/ПОСЛЕ УП

10D/5D

Защита ЭВБ по ГОСТ 14254-96

IP66

Потребляемая мощность, не более, Вт

20

Электропитание:

—    напряжение постоянного тока

—    напряжение переменного тока

12-28 В 220 В (±10%) 50±1 Гц

питание от аккумулятора

Габаритные размеры ЭВБ, мм

184x150x89

238x138x38

Масса ЭВБ, кг

0,9

1,36

Примечания:

1)    Дополнительная опция.

2)    D — Внутренний диаметр ИТ

Наименование характеристики

Модификация

DF868

XMT868i

Диапазон измерений скорости, м/с (2 направления)

от 0,3 до 12,2

Максимальное количество каналов измерения расхода

2

Внешний диаметр ИТ, мм

—    при использовании накладных УП

—    при использовании врезных УП

от 12,7 до 7600 от 25,4 до 5000

Толщина стенки ИТ, не более, мм

76,2

Температура измеряемой среды, °С

—    при использовании накладных УП

—    при использовании врезных УП

от минус 40 до 150 от минус 190до 3001) от минус 40 до 100 от минус 190до 6001)

от минус 40 до 150 от минус 190до 3001) от минус 40 до 100 от минус 190до 6001)

Максимальное давление измеряемой среды при использовании врезных УП, МПа

20

Пределы допускаемой относительной погрешности при измерении скорости, объемного расхода и объема, %

—    1 канальное исполнение

■    при поверке проливным методом

■    при поверке имитационным методом

—    2 канальное исполнение

■    при поверке проливным методом

■    при поверке имитационным методом

D2)<0,15 м

D>0,15 м

D<0,15 м

D>0,15 м

±2

±4

±1

±2

±1

±2

±0,5

±1

±2

±4

±1

±2

±1

±2

±0,5

±1

Температура окружающей среды для ЭВБ °С:

от минус 20 до 55

от минус 40 до 60

Температура хранения для ЭВБ, °С

от минус 55 до 75

Максимальная длина кабеля от УП к ЭВБ, м

305

Варианты выходного сигнала

аналоговый (от 0/4 до 20 мА) импульсный частотный

аналоговый (от 0/4 до 20 мА) импульсный частотный

Пределы абсолютной погрешности канала аналогового вывода, мА

0,016

0,016

Коммуникационный цифровой интерфейс

RS232/RS4851),

Ethernet^

RS232/RS4851), Ethernet1), Foundation FieldBus1)

Поддерживаемые протоколы связи

IDM, OPC сервер1-*, Modbus RTU1) Modbus TCP1)

HART3), IDM, OPC сервер3), Modbus RTU1), Modbus TCP1) Foundation FieldBus1)

Длина прямолинейного участка ДО/ПОСЛЕ УП

10D/5D

Электропитание:

—    напряжение постоянного тока

—    напряжение переменного тока

12-28 В 220 В (±10%) 50±1 Гц

Потребляемая мощность, не более, Вт

20

Защита ЭВБ по ГОСТ 14254-96

IP66

Маркировка взрывозащиты ЭВБ по ГОСТ Р 51330.13-99

1ExdIICT6/T5

1ExdIICT6/T5

Габаритные размеры ЭВБ, мм

220x210x90

208x208x 168

Наименование характеристики

Модификация

DF868

XMT868i

Масса ЭВБ, кг

5

4,5

Примечания:

1)    Дополнительная опция.

2)    D — Внутренний диаметр ИТ

Таблица 4 — Метрологические и технические характеристики SEN898

Наименование характеристики

ф

и

д

о

М

эикация

SEN898 LCT

SEN898 LNG

Диапазон измерений скорости, м/с (2 направления)

от 0,3 до 12,2

Максимальное кол-во каналов измерения расхода

4

Внешний диаметр трубопровода D, мм (стандартное исполнение)

от 100 до 900

от 100 до 900

Внешний диаметр трубопровода D, мм (специальное исполнение)

от 900 до 1400

от 900 до 1400

Температура измеряемой среды, °С

от минус 40 до 120

от минус 200 до 120

Максимальное давление измеряемой среды, МПа

10

Пределы допускаемой относительной погрешности при измерении скорости, объемного расхода и объема, %

—    при поверке проливным методом

—    при поверке имитационным методом

±0,15

±0,3

±0,25

±0,5

Температура окружающей среды, °С

от минус 40 до 60

Температура хранения, °С

от минус 40 до 70

Варианты выходного сигнала

аналоговый (от 0/4 до 20 мА), импульсный, частотный

Пределы абсолютной погрешности канала аналогового вывода, мА

0,016

Коммуникационный цифровой интерфейс

RS232/RS485, USB

Поддерживаемые протоколы связи

PanaLink, HART, Modbus1)

Длина прямолинейного участка ДО/ПОСЛЕ ультразвуковых преобразователей

10D/5D

Электропитание:

—    напряжение постоянного тока

—    напряжение переменного тока

12-32 В 220 В (±10%) 50±1 Гц

Потребляемая мощность, не более, Вт

20

Защита ЭВБ по ГОСТ 14254-96

IP66

Маркировка взрывозащиты ЭВБ по ГОСТ Р 51330.13-99

1ExdIIB+HT6

Габаритные размеры ЭВБ, мм

330x270x230

Масса ЭВБ, кг

11,5

Примечания

1) Дополнительная опция

Знак утверждения типа

наносится на титульный лист Руководства по эксплуатации типографическим способом и на корпус ЭВБ расходомеров в виде голографической наклейки.

Таблица 5 — Комплектность

Наименование

Количество

ЭВБ расходомера

1 шт.

Комплект УП

1, 2 или 4 комплекта^

Соединительные кабели

1 комплект

Измерительный участок ИТ

1 шт.2)

Программный пакет PanaView для конфигурирования, параметризации и диагностики расходомера

1 шт.2)

Руководство по эксплуатации:

—    Ультразвуковой расходомер жидкости (1 или 2 канала). Модель AT868. Руководство по эксплуатации.

—    Портативный ультразвуковой расходомер жидкости. Модель PT878. Руководство по эксплуатации.

—    Двухканальный ультразвуковой расходомер жидкости. Модель DF868. Руководство по эксплуатации.

—    Двухканальный ультразвуковой расходомер жидкости. Модель XMT868i. Руководство по эксплуатации.

—    Ультразвуковой расходомер жидкости. Модель Sentinel LCT. Руководство по эксплуатации.

—    Ультразвуковой расходомер жидкости. Модель Sentinel LNG. Руководство по эксплуатации.

1 экз.

Расходомеры-счетчики жидкости ультразвуковые мод. AT868, DF868, XMT868i, PT878, SEN898. Методика поверки.

1 экз

Примечания:

1)    Тип и количество УП выбирается в зависимости от требуемой точности измерений, типа измеряемой среды и условий эксплуатации расходомера

2)    Поставляется дополнительно по заказу

Поверка

осуществляется по документу МП 51863-12 «Инструкция. Государственная система обеспечения единства измерений. Расходомеры-счетчики жидкости ультразвуковые мод. AT868, DF868, XMT868i, PT878, SEN898. Методика поверки», утвержденному ГЦИ СИ ООО «Метрологический центр СТП» 15 мая 2012 г.

Перечень основных средств поверки (эталонов):

—    трубопоршневая установка, диапазон воспроизводимого объемного расхода должен соответствовать рабочему диапазону поверяемого расходомера, с пределом основной относительной погрешности ±0,05%;

—    поверочная расходоизмерительная установка (далее — ПУ), диапазон воспроизводимого объемного расхода должен соответствовать рабочему диапазону поверяемого расходомера, с пределом основной относительной погрешности ±0,3%;

—    калибратор многофункциональный MC5-R (далее — калибратор), диапазон измерения токового сигнала от минус 100 до 100 мА, диапазон измерения частотного сигнала от

0,0028 Гц до 50 кГц, счет импульсов до 9999999 имп., пределы допускаемой погрешности в режиме измерения токового сигнала ±0,02 % от показания ±1,5 мкА, пределы допускаемой погрешности в режиме измерения частотного сигнала ±0,1 % от показания, минимальная амплитуда синусоидальных и прямоугольных сигналов: 2 В для частот ниже 10 кГц и длительности импульса более 50 мкс, 3 В для частот от 10 до 50 кГц и длительности импульса от 10 до 50 кГц.

Нормативные документы, устанавливающие требования к ИС

ГОСТ 8.510-2002 Государственная система обеспечения единства измерений. Государственная поверочная схема для средств измерений объема и массы жидкости ГОСТ 14254-96 Степени защиты, обеспечиваемые оболочками (Код IP)

ГОСТ Р 51330.0-99 Электрооборудование взрывозащищенное. Часть 0.Общие требования

ГОСТ Р 51330.1-99 Электрооборудование взрывозащищенное. Часть 1. Взрывозащита вида «взрывонепроницаемая оболочка»

Рекомендации к применению

— осуществление торговли и товарообменных операций.

Table of Contents for Panametrics DigitalFlow XMT868i:

  • DigitalFlow™ XMT868i Startup Guide 3 Chapter 2. Initial Setup 2.2 XMT868i Enclosure Keypad Program (cont.) The meter then starts to display measured parameters. To enter the Keypad Program, press the [Escape] key, followed by the [Enter] key, and the [Escape] key again. Each successive key must be entered within 10 seconds of the prior key. As a guide in following the programming instructions in this chapter, the relevant portions of the Model XMT

  • Chapter 20 DigitalFlow™ XMT868i Startup Guide [no content intended for this page]

  • DigitalFlow™ XMT868i Startup Guide 1 Chapter 4. Specifications Chapter 4. Specifications 4.1 General The general specifications for the Model XMT868i flow transmitter are divided into the following categories: 4.1.1 Hardware Configuration Channel Options: Standard: 1 Channel. Optional: 2 Channel (2 pipes or 2 paths per pipe). Mode Options: Transit-Time. Package Options: Standard: Epoxy-coated aluminum Type 7/4X, IP66 Explos

  • Copyright 2021 Baker Hughes company. This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries. All third-party product and company names are trademarks of their respective holders. 910-171U Rev.G1 Customer Support Centers U.S.A. The Boston Center 1100 Technology Park Drive Billerica, MA 01821 U.S.A. Tel: 800 833 9438 (toll-free) 978 437 1000 E-mail: [email protected] Ireland Sensing

  • DigitalFlow™ XMT868i Startup Guide 11 Chapter 2. Initial Setup 2.4.2.1 Programming the Mass Flow Option 1. Scroll to the desired Mass Flow Units for flow rate display (listed in Table 12), and press [Enter]. 2. Scroll to the desired Mass Flow Time Units (from seconds to days), and press [Enter]. 3. Scroll to the desired number of Decimal Digits (digits to the right of the decimal point) in the mass flow rate display, and press [Enter]. 4. Scroll to the desired Mass Totals (units for the to

  • Chapter 4. Specifications 8 DigitalFlow™ XMT868i Startup Guide 4.5 Flowcell 4.5.1 Tilted Diameter Pipe Sizes: Standard: 2 to 200 in. (50 to 5,000 mm). Optional: >200 in. Material: Standard: Stainless steel; carbon steel; plastics; fiberglass. Optional: Other metals; plastics; etc. Process Connection: Standard: Straight (welding); 150-lb flange; 300-lb flange

  • DigitalFlow™ XMT868i Startup Guide 15 Appendix B. Data Records CHx-SETUP-ADVANCED FEATURES-KV/SS KV/SS (or C3) Pairs KV/SS (or C3) Pairs # Sig. Strength/ Soundspeed Kinematic Viscosity # Sig. Strength/ Soundspeed Kinematic Viscosity 1 1 2 2 3 3 4 4 5 5 6 6 7 7 8 8 9 9 10 10 11 11 12 12 13 13 14 14 15 15 16 16 17 17 18 18 19 19 20 20 CHx-SETUP-ADVANCED FEATURES-MULTIK Custom Type CstV CstR CustomType CstV CstR K-Factor Table K-Factor Table K-Factor # Vel./Reynolds K

  • Chapter 2. Initial Setup 10 DigitalFlow™ XMT868i Startup Guide 2.4.2 Entering Data in the Channel System Option (cont.) 6. Scroll to the desired number of Vol Decimal Digits (digits to the right of the decimal point) in the volumetric flow display, and press [Enter]. 7. Scroll to the desired Totalizer Units for the totalized flow rate display (listed in Table 11), and press [Enter]. 8. Scroll to the desired number of Decimal Digits (digits to the right of the decimal p

  • Contents vi DigitalFlow™ XMT868i Startup Guide 4.3 Transducer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 4.3.1 Wetted Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

  • panametrics.com Flow 910-171U August 2021 DigitalFlow™ XMT868i Ultrasonic Flow Transmitter for Liquids (1 & 2-Channel) Startup Guide

  • Chapter 1. Installation 2 DigitalFlow™ XMT868i Startup Guide 1.3 Site Considerations Because the relative location of the flowcell and the electronics enclosure is important, use the guidelines in this section to plan the XMT868i installation. Figure 1 shows a typical installation. Figure 1: A Typical Model XMT868i System 1.3.1 Electronics Enclosure Location The standard Model XMT868i electronics enclosure is a powder-coated, cast-aluminum, Type-7/4X explosion-proof enclos

  • DigitalFlow™ XMT868i Startup Guide 11 Chapter 1. Installation 1.7.4.2 Wiring the RS485 Interface Use the optional RS485 serial port to network multiple XMT868i flow transmitters to a single computer terminal. As an option, the standard RS232 port on the XMT868i may be configured as a two-wire, half-duplex RS485 interface. IMPORTANT: The XMT868i must be configured at the factory for RS485 operation. The maximum cable length for RS485 is 4000 ft (1200 m). To wire the RS485 serial port, refer to Figure 10 on page

Questions, Opinions and Exploitation Impressions:

You can ask a question, express your opinion or share our experience of Panametrics DigitalFlow XMT868i device using right now.

Понравилась статья? Поделить с друзьями:
  • Чаша менструационная инструкция по применению фото
  • Инструкция по эксплуатации кофемашины delonghi caffe venezia
  • Arluma ru инструкция масло для дерева
  • Мазь синафлан инструкция по применению для чего она нужна взрослым
  • Дюфалак нормазе инструкция по применению цена