Service
Manual
DAQSTATION CX1000/CX2000
SM 04L31A01-01E
SM 04L31A01-01E
2nd Edition
Yokogawa Electric Corporation
Important Notice to the User
This manual contains information for servicing YOKOGAWA’s DAQSTATION CX1000/ CX2000. Check the serial number to confirm that this is the correct service manual for the instrument to be serviced. Do not use the wrong manual.
Before any maintenance and servicing, read all safety precautions carefully.
Only properly trained personnel may carry out the maintenance and servicing described in this service manual.
Do not disassemble the instrument or its parts, unless otherwise clearly permitted by this service manual.
Do not replace any part or assembly, unless otherwise clearly permitted by this service manual.
In principle, Yokogawa Electric Corporation (YOKOGAWA) does not supply parts other than those listed in the customer maintenance parts list in this service manual (mainly modules and assemblies). Therefore if an assembly fails, the user should replace the whole assembly and not components within the assembly (see “Note”). If the user attempts to repair the instrument by replacing individual components within the assembly, YOKOGAWA assumes no responsibility for any consequences such as defects in instrument accuracy, functionality, reliability, or user safety hazards.
YOKOGAWA does not offer more detailed maintenance and service information than that contained in this service manual.
All reasonable efforts have been made to assure the accuracy of the content of this service manual. However, there may still be errors such as clerical errors or omissions. YOKOGAWA assumes no responsibility of any kind concerning the accuracy or contents of this service manual, nor for the consequences of any errors.
All rights reserved. No part of this service manual may be reproduced in any form or by any means without the express written prior permission of YOKOGAWA. The contents of this manual are subject to change without notice.
Note
YOKOGAWA instruments have been designed in a way that the replacement of electronic parts can be done on an assembly (module) basis by the user. YOKOGAWA instruments have also been designed in a way that troubleshooting and replacement of any faulty assembly can be done easily and quickly. Therefore, YOKOGAWA strongly recommends replacing the entire assembly over replacing parts or components within the assembly. The reasons are as follows:
•The instruments use high-performance microprocessors, large scale CMOS gate arrays, and surface-mount components to provide state-of-the-art performance and functions.
•Repair of components can only be performed by specially trained and qualified maintenance personnel with special highly-accurate tools, including costly ones.
•When taking the service life and cost of the instruments into consideration, the replacement of assemblies offers the user the possibility to use YOKOGAWA instruments more effectively and economically with a minimum in downtime.
•Zip is a trademark or registerd trademark of Iomega Corporation in the United States and/or other countries.
•Adobe and Acrobat are trademarks of Adobe System incorporated.
Disk No. SM13
2nd Edition : August 2002 (YK)
All Rights Reserved, Copyright © 2002, Yokogawa Electric Corporation
Introduction
This manual contains information for servicing YOKOGAWA’s DAQSTATION CX1000/ CX2000.
Note
This is the second edition of the manual, dated August 2002.
WARNING
This service manual is to be used by properly trained personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to the safety precautions prior to performing any service. Even if servicing is carried out according to this service manual, or by qualified personnel, YOKOGAWA assumes no responsibility for any result occurring from this servicing.
Safety Precautions
The following general safety precautions must be taken during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific WARNINGS given elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument.
Yokogawa Electric Corporation assumes no liability for the customer’s failure to comply with these requirements.
WARNING
Use the Correct Power Supply
Ensure the source voltage matches the voltage of the power supply before turning ON the power.
Use the Correct Power Cord and Plug
To prevent an electric shock or fire, be sure to use the power supply cord recommend by YOKOGAWA. The main power plug must be plugged in an outlet with a protective grounding terminal. Do not invalidate protection by using an extension cord without protective grounding.
Connect the Protective Grounding Terminal
The protective grounding terminal must be connected to ground to prevent an electric shock before turning ON the power.
Do Not Impair the of Protective Grounding
Never cut off the internal or external protective grounding wire or disconnect the wiring of the protective grounding terminal. Doing so creates a potential shock hazard.
Do Not Operate with Defective Protective Grounding
Do not operate the instrument if you suspect the protective grounding might be defective.
Do Not Operate Near Flammable Materials
Do not operate the instrument in the presence of flammable liquids or vapors. Operation of any electrical instrument in such an environment constitutes a safety hazard.
Do Not Remove Any Covers
There are some areas components inside the instrument containing high voltage. Do not remove any cover, if the power supply is connected. The cover should be removed by qualified personnel only.
Ground the Instrument before Making External Connections
Connect the protective grounding before connecting the instrument to a measurement or control unit.
Firnish a switch (double-pole type) to separate the CX2000 from the main power supply in the power supply line. In addition, make sure to indicate that the switch is a power control for the CX2000 on the switch.
Switch Specifications
Steady-state current ranting:1 A or more, inrush current rating:60 A or more
Connect a fuse between 2 A and 15 A in the power supply line.
Safety Symbols Used on Equipment and in Manuals
To avoid injury, death of personnel or damage to the instrument, the operator must refer to an explanation in the user’s manual.
High temperature. To avoid injury caused by hot surfaces, the operator must not touch the heatsink.
Protective grounding terminal, to protect against electrical shock.
This symbol indicates that the terminal must be connected to ground before operation of equipment.
This symbol represents a functional grounding terminal. Such terminals should not be used as a protective grounding terminal.
A WARNING sign calls attention to a procedure, practice, or condition, that could result in the injury or death of personnel if not correctly performed or adhered to.
A CAUTION sign calls attention to a procedure, practice, or condition, that could result in damage to or the destruction of part of the instrument if not correctly performed or adhered to.
Overview of This Manual
This manual is meant to be used by qualified personnel only. Make sure to read the safety precautions at the beginning of this manual as well as the warnings and cautions contained in the chapters relevant to any servicing you may be carrying out.
This manual contains the following chapters.
1Principles of Operation
Provides an introduction and safety considerations.
2Testing
Explains the tests for checking the performance of the instrument.
3Adjustments
Explains the adjustments which can be performed by users.
4Replacing Parts
Describes maintenance which can be performed by users.
5Troubleshooting
Presents procedures for troubleshooting and how to proceed in case parts need to be replaced.
6Schematic Diagram
Provides a system configuration diagram.
7Customer Maintenance Parts List
Contains exploded views and a list of replaceable parts.
8Procedures for Disassembly (CX1000)
Lists the steps required to remove parts from the instrument.
Specifications are not included in this manual. For specifications, refer to IM 04L31A01-01E or IM 04L31A01-03E.
Contents
Important Notice to the User ………………………………………………………………………………….. |
1 |
||
Introduction …………………………………………………………………………………………………………. |
2 |
||
Safety Precautions ……………………………………………………………………………………………….. |
2 |
||
Safety Symbols Used on Equipment and in Manuals ………………………………………………… |
3 |
||
Overview of This Manual ………………………………………………………………………………………. |
4 |
||
Chapter 1 Principles of Operation |
|||
1.1 |
Block Diagram of the CX1000/CX2000 …………………………………………………………………. |
1-1 |
|
1.2 |
Input Section …………………………………………………………………………………………………….. |
1-2 |
|
Chapter 2 |
Testing |
||
2.1 |
Overview of Tests ………………………………………………………………………………………………. |
2-1 |
|
2.2 |
Test Procedures ………………………………………………………………………………………………… |
2-3 |
|
2.3 |
CTRL Module Assembly Tests …………………………………………………………………………… |
2-10 |
|
2.4 |
DIO Module Assembly Test ……………………………………………………………………………….. |
2-11 |
|
Chapter 3 |
Adjustments |
||
3.1 |
Calibration of the Measuring Instrument’s Input……………………………………………………… |
3-1 |
|
3.2 |
Control Output Calibration ………………………………………………………………………………….. |
3-6 |
|
Chapter 4 |
Replacing Parts |
||
4.1 |
Replacement of the Control Output Terminal Block (Module) …………………………………… |
4-1 |
|
Chapter 5 |
Troubleshooting |
||
5.1 |
Procedure ………………………………………………………………………………………………………… |
5-1 |
|
5.2 |
Flow Chart ………………………………………………………………………………………………………… |
5-2 |
|
5.3 |
Troubleshooting Checklist …………………………………………………………………………………… |
5-3 |
|
Chapter 6 |
Schematic Diagram |
||
6.1 |
CX1000 Schematic Diagram ……………………………………………………………………………….. |
6-1 |
|
6.2 |
CX2000 Schematic Diagram ……………………………………………………………………………….. |
6-2 |
|
Chapter 7 Customer Maintenance Parts List |
|||
7.1 |
CX1000 Customer Maintenance Parts List ……………………………………………………………. |
7-1 |
|
7.2 |
CX1000 Standard Accessories ……………………………………………………………………………. |
7-8 |
|
7.3 |
CX2000 Customer Maintenance Parts List ……………………………………………………………. |
7-9 |
|
7.4 |
CX2000 Standard Accessories ………………………………………………………………………….. |
7-16 |
|
Chapter 8 Procedures for Disassembly (CX1000) |
|||
8.1 |
Removing the Bezel …………………………………………………………………………………………… |
8-1 |
|
8.2 |
Removing the Board Assemblies and the Drive ……………………………………………………… |
8-4 |
|
8.3 |
Removing the Keyboard …………………………………………………………………………………….. |
8-6 |
|
8.4 |
Replacing the Backlight ……………………………………………………………………………………… |
8-8 |
|
8.5 |
Removing the Terminal Assembly ………………………………………………………………………. |
8-10 |
Chapter 1 Principles of Operation
1.1Block Diagram of the CX1000/CX2000
Block Diagram of the CX1000
CPU |
|||||||
Control/Measurement |
Data storage functions |
||||||
Input section |
|||||||
Input terminal |
External storage |
||||||
Internal memory |
|||||||
Scanner assembly |
|||||||
A/D assembly |
|||||||
Display unit |
|||||||
Calculation function |
|||||||
Control Output section |
|||||||
Measurement Alarm function |
|||||||
Control Output |
|||||||
Control Alarm function |
|||||||
Contact Input |
|||||||
Communication function |
|||||||
Contact Output |
(optional) |
||||||
Remote function |
|||||||
(optional) |
|||||||
Measurement Alarm Relay |
|||||||
Block Diagram of the CX2000
CPU
Data storage functions
Measurement Input section
Input terminal |
External storage |
Scanner assembly |
Internal memory |
A/D assembly |
|||||||||
Display unit |
VGA out function |
(optional) |
|||||||
Control Input section |
Calculation function |
||||||||
Input terminal |
|||||||||
Measurement Alarm function |
|||||||||
Scanner assembly |
|||||||||
Control Alarm function |
|||||||||
A/D assembly |
|||||||||
Communication function |
|||||||||
Control Output section |
Remote function |
(optional) |
|||||||
Control Output |
(optional) |
||||||||
Measurement Alarm Relay |
|||||||||
Contact Input |
(optional) |
||||||||
Transmitter power supply |
|||||||||
Contact Output |
(optional) |
||||||||
Control — purpose DIO |
|||||||||
Refer for details see schematic diagram page 6-1 and 6-2.
1
of Principles
Operation
1.2Input Section
A/D Assembly
The A/D assembly has items such as a programmable gain amp, voltage reference, PWM modulator, current source for RTD measurements, differential amp, voltage source for RJC, serial parallel converter, control logic, and an occurred scanner SSR control signal.
The A/D assembly uses a sinewave oscillating type self-resonant switching power supply (DC/DC converter), and noise filtering is achieved by signal integration.
The A/D assembly detects the frequency of the power while it is ON and the integrated time becomes 20 ms or 16.67 ms. Therefore it carries a very high rate of noise rejection for the power frequency (in auto mode).
In case the power frequency of the instrument and of the measured object are different, the appropriate integrated time is manually selectable. In case of the CX2000, the selection of 100 ms for 50/60 Hz is also available. A 16 bit resolution is achieved regardless of the integrated time.
Input Terminal
The internal printboard is isothermal because a print board with a metal core is being used. Therefore, stable reference junction compensation is realized.
Scanner Assembly
An in-house SSR (solid state relay) is being used for the scanner. The SSR, having a semiconductor switch, has a withstand voltage as high as 1500 V and a leakage current of only 1 nA. For that reason, it has the following features.
•Semi-infinite life due to the absence of mechanical contacts
•Silent operation
•No occurrence of thermoelectric power.
On the other hand, compared to a mechanical relay, the SSR has, the disadvantage of a bigger ON resistance and OFF capacity. As a result, RTD measurement and noise resistance characteristics are affected. Regarding RTD measurements, a differential amp was inserted into the previously mentioned analog circuit without increasing the number of parts, so that it would receive no influence from ON resistance.
For RTD measurements in Measurement input section there is generally no insulation between channels.
For RTD measurements in Control input section there is generally insulation between channels.
Data Storage Functions
For storing data, the CX1000/CX2000 has 1.2 MB of internal memory and is equipped with a Zip drive, or an ATA flash memory card drive. The measured data can also be saved to external storage media such as floppy disks, Zip disks, and ATA flash memory cards.
Display Unit
The CX has a 5.5-inch (CX1000) or 10.4-inch (CX2000) TFT color LCD on which it displays the measured results (240 (vertical) × 320 (horizontal) pixels for the CX1000 or 480 (vertical) × 640 (horizontal) pixels for the CX2000).
1.2 Input Section
Calculation Function
The CX1000/CX2000 performs differential computation, linear scaling, and square roots using a microprocessor on the CPU board.
Measurement Alarm Function
The following eight alarm types can be set.
High limit (H), low limit (L), differential high limit (h), differential low limit (l), rate-of- change on increase (R), rate-of-change on decrease (r), alarm delay upper limit alarm (T), or alarm delay lower limit alarm (t).
Control Alarm Function
The following nine alarm types can be set.
PV high-limit, PV low-limit, Deviation high-limit, Deviation low-limit, Deviation high & low limit, SP high-limit, SP low-limit, Output high-limit, Output low-limit
Other Functions
•Communication Function:
Ethernet (standard)
RS-232/RS-422A interface added (optional).
•Remote Function:
The trigger, start/stop, time adjustment, and other functions can be controlled remotely (optional).
•Measurement Alarm Relay:
Measurement alarm output and memory end/fail output (option for CX1006 and CX2000).
•Transmitter Power Supply:
DC24 V output for transmitter (only CX2000’s optional).
•VGA Out Function (Option for CX2000 Only):
The instrument’s screen can be displayed on an external monitor via VGA output.
•Control — Purpose DIO (CX2000 Only):
Contact input
Input for designated operations such as start/stop. Activate using a no-voltage contact or open collector signal. This is a 12-point input.
Contact output
Control Alarm output consists of 12 points of transistor contact output.
Control Output Section
Control Output
The following types are available for universal control output.
•Current output (continuous PID control output)
Continuously output a current (analog signal) proportional to the calculated PID values.
•Time proportional PID voltage pulse output
Output an ON/OFF signal, having a pulse width proportional to the time, as a voltage per the calculated PID values.
•On/Off control relay contact output
•Output an ON/OFF signal, having a pulse width proportional to the time, at a relay contact point per the calculated PID values.
•Output an ON/OFF signal to a relay contact point corresponding to the sign (+/–) of the deviation in the measured value from the specified target value.
Contact Input
Input for designated operations such as start/stop. Activate using a no-voltage contact or open collector signal.
Contact Output
Control Alarm output consists of relay contact output and transistor contact output.
1
of Principles
Operation
Chapter 2 Testing
2.1Overview of Tests
The following describes general testing procedures for DAQSTATION CX1000/CX2000 series instruments. For tests on specific modules or assemblies, see sections 2.3 and later.
Operating Conditions
Ambient Temperature: |
23 |
± 2°C |
Relative Humidity: |
55 |
± 20% |
Test Instruments
Instrument |
Specifications |
|
DC voltage generator |
Accuracy: 0.005% of setting + 1 mV |
|
DMM |
Accuracy: 0.005% of rdg + 1 mV |
|
Resistors |
Accuracy: 0.01% or better |
|
Insulation tester |
500 VDC |
|
Withstanding voltage tester |
AC 1 to 3 kV, 500 VDC |
|
External monitor (for test of /D5 option) |
VGA monitor (H: 33.3 kHz, V: 60.168 Hz) |
|
Oscilloscope |
200 kS/s or more, isolated input |
|
Thermostatic chamber |
ZC-114 (Coper Electronics Co., Ltd.) or equivalent |
|
Testing Conditions |
The tests cover all included A/D converters.
The unit’s analog input is in analog multiplexer format.
For the CX1000, channel group 1–6 of the measurement input section and channel group 7–11 of the control input section are each assigned to one A/D converter. For the CX2000, channel groups 1–5 and 6–10 of the control input section (slot 1) and channel groups 1–10 and 11-20 of the measurement input section (slots 2 and 3) are each assigned to one A/D converter.
Therefore, except for when specifying inputs linked to the same A/D converter, only one arbitrary channel within a group need be tested (for example, not channels 1—5 but only channel 1).
2.1 Overview of Tests
Tests
Insulation Resistance Test
Withstand Voltage Test
Measurement Accuracy Test
Error between Channels Test
Excessive Input Test
Burnout Test
Reference Junction Compensation Accuracy Test
Display Function Test
VGA Output Function Test (For /D5 Option Only)
Serial Communications Function Test (Only When The -1 Or -2 Suffix Code Is
Specified For The Communications Port).
Battery Backup Function Test
Continuous Operation Test
Current Output Accuracy Test
Voltage Pulse Output Test
Control Output Relay Test
DIO Test
KEY Function Test
Media Function Test
Alarm Relay Output Function Test (Only If The /A6, /A6R, /A4F, Or /A4FR Option Is Installed) Remote Function Test (Only If The /A6R Or /A4FR Option Is Installed) Communications Function Test (Ethernet)
Test of 24 VDC Transmitter Power Output (Only If The /TPS4 Is Installed)
Consult your nearest Yokogawa representative regarding the following tests.
Power Supply Frequency Detection Function Test
Memory Test
2.2 Test Procedures
Insulation Resistance Test
Perform this test using a DC 500 V insulation resistance meter and confirm that the
results meet the criteria below.
Terminals |
Reference Values |
Notes |
Power terminal to earth terminal |
100 MΩ or higher |
Short all channels prior to test |
Measurement input terminal to |
||
earth terminal |
||
Control input terminal to earth |
||
terminal |
||
RS-422-A/485 SG terminal to |
100 MΩ or higher |
Test only if the basic |
RS-422-A/485 FG terminal |
specification code for the |
|
communications port is -2. |
||
Ethernet input/output terminal |
100 MΩ or higher |
Short all terminals prior to test |
to earth terminal |
||
Withstand Voltage Test
Perform the test using a withstanding voltage tester and confirm that the results meet the
criteria below, and that the instrument does not malfunction.
Terminal |
Reference Values |
|
AC power terminal to earth terminal* |
Leakage current of 10 mA or less at 1.5 kV AC for 1 minute |
|
DC power terminal to earth terminal (/P1)* |
Leakage current of |
10 mA or less at 0.5 kV AC for 1 minute |
Measurement input terminal to earth terminal† |
Leakage current of |
2 mA or less at 1.5 kV AC for 1 minute |
Control input terminal to earth terminal† |
Leakage current of |
2 mA or less at 1.5 kV AC for 1 minute |
Between measuring input terminals‡ |
Leakage current of |
1 mA or less at 1 kV AC for 1 minute |
Between control input terminals‡ |
Leakage current of |
1 mA or less at 1 kV AC for 1 minute |
Current and voltage pulse to earth terminal§ |
Leakage current of |
2 mA or less at 1 kV AC for 1 minute |
Control relay terminal to earth terminal |
Leakage current of |
2 mA or less at 1.5 kV AC for 1 minute |
(2, 4, and 6 loop models)|| |
||
DO relay terminal to earth terminal |
Leakage current of |
2 mA or less at 1.5 kV AC for 1 minute |
(2, 4, and 6 loop models)# |
||
DO(Tr) terminal to earth terminal |
Leakage current of |
2 mA or less at 0.5 kV AC for 1 minute |
(2, 4, and 6 loop models or /CST1)** |
||
D1 terminal to earth terminal |
Leakage current of |
2 mA or less at 0.5 kV AC for 1 minute |
(2, 4, and 6 loop models or /CST1)†† |
||
Alarm relay terminal to earth terminal |
Leakage current of |
2 mA or less at 1.5 kV AC for 1 minute |
(/A6, /A6R, /A4F, /A4FR)‡‡ |
||
Remote terminal to earth terminal (/A6R, /A4FR)§§ |
Leakage current of |
2 mA or less at 0.5 kV AC for 1 minute |
24 V transmitter power supply output to earth terminal Leakage current of |
10 mA or less at 0.5 kV AC for 1 minute |
|
(/TPS4)|||| |
*Short L and N (or +/– with the /P1 option)
† Short all channels
‡ Short the + and – input terminals (except for the RTD b terminal)
Short the even and odd channels (for example, short channels 1-3-5 and 2-4-6 on the CX1000 or channels 1-3- 5-7-9 and channels 2-4-6-8-10 for the CX2000), and test the withstanding voltage between them.
§ Short mA, PULS, and C.
|| Short NO, NC, and C on CTRL OUT.
# Short 1NO, 1C, and 2NO on DO.
**Short DO3—DO6 and C.
†† Short all DI terminals.
‡‡ Short all alarm relay terminals. §§ Short all remote terminals.
|||| Short all transmitter power supply output terminals (+/–).
2.2 Test Procedures
Measurement Accuracy Test (Measurement Input and Control Input)
Check that the specifications below are met given the following operating conditions: 23
± 2°C, 55 ± 10% RH, and a warm up time of 30 minutes.
Range |
Input Value |
Criterion |
Specifications |
20 mV |
+20.00 mV |
±3 digits |
±(0.1% of rdg + 2 digits) |
0.00 mV |
±2 digits |
||
–20.00 mV |
±3 digits |
||
60 mV |
+60.00 mV |
±6 digits |
±(0.1% of rdg + 2 digits) |
0.00 mV |
±2 digits |
||
–60.00 mV |
±6 digits |
||
200 mV |
+200.0 mV |
±3 digits |
±(0.1% of rdg + 2 digits) |
0.0 mV |
±1 digits |
||
–200.0 mV |
±3 digits |
||
2 V |
+2.000 V |
±3 digits |
±(0.1% of rdg + 2 digits) |
0.000 V |
±1 digits |
||
–2.000 V |
±3 digits |
||
6 V |
+6.000 V |
±6 digits |
±(0.1% of rdg + 2 digits) |
0.000 V |
±1 digits |
||
–6.000 V |
±6 digits |
||
20 V |
+20.00 V |
±3 digits |
±(0.1% of rdg + 2 digits) |
0.00 V |
±1 digits |
||
–20.00 V |
±3 digits |
||
50 V |
+30.00 V |
±5 digits |
±(0.1% of rdg + 3 digits) |
0.00 V |
±2 digits |
||
–30.00 V |
±5 digits |
||
Pt100 |
18.52Ω/–200°C |
±0.4°C |
±(0.15% of rdg + 0.3°C) |
100.00Ω/0°C |
±0.2°C |
||
313.71Ω/600°C |
±0.9°C |
Error between Channels Test
Connect the (+), (–), and (b) terminals on all measurement and control input channels using thick leads, then check whether the following criteria are met.
Range |
Input Value |
Criterion |
20 mV |
Short (+) (–) |
0.00 mV ±2 digits |
Pt100 |
103.9 Ω |
Measured value: 10°C ± 0.2°C,difference between channels: |
0.2°C or less |
Note
•If the measurement error does not meet the criterion (10°C ± 0.2°C), and unless you determine the cause to be the effect of the resistance in the test facility’s RTD wiring, use the following measurement criteria.
Measured value: 10°C ± 0.5°C
Difference between channels: 0.2°C or less
•See your nearest Yokogawa representative for the error between channels in the control input section.
Excessive Input Test
1.Connect the plus (+) and minus (–) sides of all measurement and control input channels.
2.Set all channels to the 20 mV range.
3.Apply 10 V between the plus and minus sides for 1 minute, and confirm that there is no malfunction.
4.Set all channels to the Pt100 range.
5.Apply ±10 V between (A) and (B) for 1 minute, and confirm that there is no malfunction.
2.2 Test Procedures
Burnout Test
1.Connect a 2 kΩ resistor between the plus (+) and minus (–) sides of an arbitrary channel.
2.Connect 56 kΩ resistor and a 0.1 µF capacitor in parallel between the plus (+) and minus (–) sides of an arbitrary channel other than the one in step 1.
3.Set the range for the 2 connected channels to TC — TYPE K, and the Burnout to ON.
4.Check that the channel to which 2 kΩ was connected displays a temperature close to the room temperature.
5.Check that the channel to which the 56 kΩ resistor and the 0.1 µF capacitor were connected in parallel displays Burnout (overrange).
6.Set Burnout to OFF.
7.Check that both channels in steps 1 and 2 do not display Overrange.
Reference Junction Compensation Accuracy Test
Perform 0°C measurement for TYPE — T on all control and measurement channels, and confirm that the result is below the following reference value.
Reference value: 0 ± 0.5°C
CAUTION
•Use a calibrated non-insulated thermocouple without a terminal tip, having a Φ0.5 mm or narrower strand. Also, be sure to take the level of calibration error into consideration during testing.
•Continuously monitor the thermostatic chamber, and check that the temperature remains at 0°C ± 0.01°C.
•When using a 0°C thermostatic chamber (for example the ZC-114 by Coper Electronics Co. Ltd.) raise the tip of the thermocouple up 10 mm from the bottom.
•Conduct all tests in a stable environment.
•Install a windbreak if necessary.
•Allow a 15 minute warm up after wiring the thermocouple.
•Always use the terminal covers.
Display Function
Check the adjustment of the display color and backlight intensity using the following procedure.
1.Press the MENU key twice to enter Set mode. The Set mode screen is displayed.
2.Press the Display soft key.
3.Select Color.
4.The default color and color name is displayed for each channel; compare the colors and their names to make sure there are no discrepancies.
2.2 Test Procedures
5.Press ESC to return to the previous screen.
6.Press the View, Direction, LCD soft key.
7.Move the cursor to Brightness.
8.For the CX1000, select 1 through 8 and confirm that each brightness level is brighter than the one before it. For the CX2000, select 1 through 4 and confirm that each brightness level is brighter than the one before it.
9.During other phases of testing as well, always be checking for any abnormalities in the display that may appear.
10.Set Brightness to 3.
VGA Output Function Test (for the /D5 Option Only)
Connect a VGA monitor to the VGA output terminal on the back of the CX2000, then look at the screen to check the VGA output. Perform steps 1—10 above for the display function and check the results.
Serial Communications Function Test (Only If the -1 or -2 Suffix Code is Specified for the Communications Port)
During this test, actual communications are performed to check whether the RS-232, RS-422A/485, or other functions are operating properly. However, if you used serial communications successfully to carry out other tests, you don’t need to perform this test. Perform the tests using the procedure below.
1.Connect a cable between the PC and the CX1000/CX2000.
If communications cannot be carried out via RS-422A/485, use an RS-232 adapter.
2.Send an arbitrary command from the PC using application software for CX1000/ CX2000 series instruments or other software, and confirm that the expected result occurs.
Battery Backup Function Test
1.While the CX1000/CX2000 is ON, set the date and time (see section 3.3, “Setting the Date and Time” in manual IM 04L31A01-01E or IM 04L31A01-03E).
2.Turn the power OFF.
3.Wait at least one minute, then turn the power back ON.
4.Confirm that the date and time set in step 1 are correct.
Continuous Operation Test
1.Enter the following settings.
•Enter AUTO SAVE to the media (see section 9.1 of the manual, IM 04L31A0101).
•Set all channels to a range of 20 V (see section 4.6 of the manual, IM 04L31A0101).
•Set input to open.
•Set display type to waveform display (see section 1.16 of the manual, IM 04L31A0101).
2.Press the START key.
3.Run the instrument continuously for 24 hours or more.
4.Press the STOP key.
5.Check the following:
•That a data file was created in the MEDIA INFO screen (see section 4.6 of the manual, IM 04L31A01-01E).
•That there is no abnormal variation in the straight line waveform.
•That there is no other strange sound or odor coming from the instrument.
2.2 Test Procedures
Current Output Accuracy
1.Change the mode to setup and choose Control, Control action, and then Input setting. Set the control mode to Single. Press the ESC key to return to the preceding menu, and choose Output processing.
2.Set the type of control output to Current-output and analog output to 4-20 mA for all the loops, save the settings, and change the mode to normal (the cycle time can be set freely).
3.Press the DISP/ENTER key and choose Control and then Control groups to display the control display.
4.Choose a loop number to measure with a blue arrow using the arrow keys.
5.Press the MODE softkey, and change the operation mode to manual by choosing MAN with the up and down arrow keys and pressing DISP/ENTER key to confirm it.
6.Press the RUN/STP softkey, and start control operation by choosing RUN with the up and down arrow keys and pressing DISP/ENTER key to confirm it.
7.Press the OUT softkey. Set the output ratio to 0.0% and 100.0% with the up and down arrow keys and measure each output current with an ammeter or the like. Verify that the measured currents are within the ranges as shown in the table below.
Output Ratio |
Reference Value |
Allowable Range |
0.0% |
4 mA |
3.984 to 4.016 mA |
100.0% |
20 mA |
19.984 to 20.016 mA |
(Load condition: 250 Ω ±1%)
Voltage Pulse Output
1.Choose Output processing by following the procedures in step 1 of “Current Output Accuracy.”
2.Set the type of control output to Voltage-pulse and the cycle time to 1 second for all the loops, save the settings, and change the mode to normal (the type of analog output can be set freely).
3.Change the operation mode to manual and start control operation by following the procedures of steps 5 and 6 of “Current Output Accuracy.”
4.Press the OUT softkey. Set the output ratio to 20.0% with the up and down arrow keys and observe the output voltage waveform on an oscilloscope or the
like. Verify that a rectangular waveform is displayed at a 1-second interval (0.2- second high and 0.8-second low) and 12 V ± 5%.
Control Relay Output
1.Choose Output processing by following the procedures in step 1 of “Current Output Accuracy.”
2.Set the type of control output to On/Off-control for all the loops, save the settings, and change the mode to normal (the cycle time and type of analog output can be set freely).
3.Change the operation mode to manual and start control operation by following the procedures of steps 5 and 6 of “Current Output Accuracy.”
4.Press the OUT softkey. Set the output ratio to 0.0% and 100.0% with the up and down arrow keys and verify that the relay outputs are as follows:
Output Ratio |
NO-C |
NC-C |
0.0% |
Break |
Make |
100.0% |
Make |
Break |
2.2 Test Procedures
Test of Digital Input/Output (not applicable for models CXx0xx)
Digital Input
1.Press the DISP/ENTER key and choose CONTROL and then DI/DO STATUS.
2.The digital input terminals are divided into blocks. Short-circuit between the common terminal and the terminals, and check the DI status display. The
indicator is green for an open circuit and red for a short circuit.
Digital Output
1.Short-circuit input terminals of all channels and adjust the range and alarm setting as follows:
Mode |
Voltage |
||
Range |
|||
Range |
2 V |
||
Span |
–2.000 to 2.000 |
||
Alarm |
On |
||
Type |
H |
||
(Level 1) |
|||
Value |
–1.000 |
||
2.Prepare the same input terminals step 1, and adjust the alarm relay output as follows:
Relay Output |
On |
|
Alarm |
DOxxx and ROxxx |
|
(Level 1) |
No. |
where xxx: digital |
output number to test |
||
3.Verify that a digital output turns on using an appropriate tester with the following
circuit:
I ≤ 50 mA
R
DO terminal
LED
V ≤ 24 V DC
Common
Test of Power Supply Voltage Switching
Change the power supply voltage to 240 V AC and verify that the UUT display does not malfunction.
Note that this test does not apply to the 24 V AC/DC-driven model (option code: /P1).
Test of Ethernet Interface
Use a computer in which the standard software “DAQSTANDARD for CX” is installed and that has an Ethernet communication function, for the following test.
Run the launcher program. Select Ethernet in the network settings and make the communication settings. Then, verify that the settings of the UUT, such as the ranges, can be set up correctly as well as those settings can be read from the computer using the settings software.
Test of Storage Drive
1.Insert a formatted disk into the drive of the UUT.
2.In the set mode*1, choose Save/Load, Clear data, and then Save settings. Save the current panel conditions under a desired filename.
3.In the set mode, choose Save/Load, Clear data, and then File list. Verify that the file saved in step 2 exists in the list.
2.2 Test Procedures
Test of Storage Drive
1.Insert a formatted disk into the drive of the UUT.
2.In the set mode*1, choose Save/Load, Clear data, and then Save settings. Save the current panel conditions under a desired filename.
3.In the set mode, choose Save/Load, Clear data, and then File list. Verify that the file saved in step 2 exists in the list.
Test of Alarm Relay Contact Outputs (applicable to option codes /A6, /A6R, /A4F and / A4FR)
Insulation Resistance and Withstanding Voltage Tests
Item |
Measured Point |
Specification |
Insulation resistance |
Between relay output terminals and |
No less than 20 MΩ at 500 V DC |
grounding terminal |
||
Withstanding voltage |
As above |
Free from damage after applying 1500 V AC, |
50/60 Hz for 1 minute (with breaking leakage |
||
current set to 2 mA) |
||
Alarm Actions
1.Prepare the same input terminals and settings as step 1 in “Digital Output” (page 2-8), and also adjust the alarm relay output as follows:
Relay Output |
On/off |
|
Alarm |
Ixx |
|
(Level 1) |
No. |
where xx: alarm |
output number to test |
||
2.Verify that an alarm contact works as follows upon turning on/off the corresponding relay output.
Terminals |
Normal |
During Alarm |
Remarks |
NO–C |
Break |
Make |
Where the output relay action is set to |
NC–C |
Make |
Break |
normally de-energized (factory set) |
Test of Remote Control (applicable to option codes /A6R and /A4FR)
Insulation Resistance and Withstanding Voltage Tests
Item |
Measured Point |
Specification |
Insulation resistance |
Between remote control terminals and |
No less than 20 MΩ at 500 V DC |
grounding terminal |
||
Withstanding voltage |
As above |
Free from damage after applying 500 V DC for 2 |
minute (with breaking leakage current set to 2 mA) |
Remote Control Actions
Assign individual functions to 8 remote control inputs, then short-circuit each of those
inputs in turn and verify that the CX1000/CX2000 is controlled as specified.
Test of 24 VDC Transmitter Power Output (applicable to option code /TPS4)
Test of Insuration Resistance and Withstanding Voltage
Insulation resistance |
Between 24 VDC output terminals and |
No less than 20 MΩ at 500 VDC |
grounding terminal |
||
As above |
Free from damage after applying 500 VAC, 50/60 Hz for |
|
1 minute (with breaking leakage current set to 10 mA) |
||
Withstanding voltage |
Between 24VDC output terminals |
Free from damage after applying 500 VAC, 50/60 Hz for |
1 minute (with breaking leakage current set to 10 mA) |
||
2.3 |
CTRL Module Assembly Tests |
||||
This section describes the test procedure for CTRL MODULE ASSY (B8700CL, |
|||||
B8700CM, B8700CN, and B8700FT) which is used on the CX2000. This test is not |
|||||
necessary if you will perform the general tests in section 2.2. Perform this test on the |
|||||
module by itself. |
|||||
Test Instruments |
|||||
Instrument |
Specifications |
||||
DC current meter |
Accuracy: 0.01% of rdg |
||||
Withstanding voltage tester |
AC 1 to 3 kV, 500 VDC |
||||
Oscilloscope |
200 kS/s or more, isolated input |
||||
Resistor |
250 Ω ±1% |
||||
Jig |
A measurement instrument having the same functions and |
||||
characteristics as the CX2000 or CX1000. |
|||||
Tests
• Withstand Voltage Test
Consult your nearest Yokogawa representative regarding the following tests.
•Output Accuracy Test
•Voltage Pulse Output Test
•Control Relay Output Test
•DI Test
•DO Test
Testing Environment
Ambient temperature of 23±5°C, relative humidity of 55 ±20%
CAUTION
•Before starting the test, allow the instrument to warm up for 30 minutes or more.
•Make sure the module is installed in the CX1000/CX2000.
Test Procedure
Perform a withstand voltage test on the module installed in the CX1000/CX2000 at the points listed below.
Test Points |
Reference Values |
|
Current and voltage pulse output to earth terminal on |
Leakage current of 2 mA or less at 1 kV AC for 1 minute |
|
the CX1000/CX2000* |
||
Control relay terminal to CX1000/CX2000 earth terminal† |
Leakage current of 2 mA or less at 1.5 kV AC for 1 minute |
|
DO relay terminal to CX1000/CX2000 earth terminal‡ |
Leakage current of 2 mA or less at 1.5 kV AC for 1 minute |
|
DO (Tr) relay terminal to CX1000/CX2000 earth terminal§ |
Leakage current of 2 mA or less at 0.5 kV AC for 1 minute |
|
DI terminal to CX1000/CX2000 earth terminal|| |
Leakage current of 2 mA or less at 0.5 kV AC for 1 minute |
|
* |
Short mA, PULS, and C. |
|
† |
Short NO, NC, and C. |
|
‡ |
Short NO and C. |
|
§ |
Short DO3—6 and C. |
|
|| |
Short all DIs. |
Loading…
*
4
L
3
1 A 1
7
1 E 0
CX2000 Installation and
Connection Guide
IM 04L31A01-71E 4th Edition
4thEdition: May 2007 (YK)
All Right Reserved, Copyright © 2001
Yokogawa Electric Corporation
Y okogawa Electric Corporation
Thank you for purchasing the CX2000.
This manual contains simple explanations about how to install and connect
the CX2000. For more information about the procedures described herein,
safety precautions, and the CX2000 functions and operation, please refer to
the PDF manual found on the provided CD-ROM.
Installation Procedure
1. Cut the instrument panel according to the diagram below.
Panel Cut Diagram
Units : mm
360 min.
+2
281
0
2. Insert the CX2000 into the front of the panel.
3. Using the mounting brackets, attach the CX2000 to the panel as shown
in the following figure.
First, attach the two mounting brackets and temporarily fasten the
attachment screws. Next, fix the CX2000 in place by tightening the
attachment screws with the appropriate torque (0.7 to 0.9 N-m.). As you
fasten the screws, press the mounting bracket against the case so that
they are in contact with each other.
Screw temporarily
Panel
Mounting
bracket
Attachment
screw
Fix in place
Case
Panel
Mounting bracket
In contact
Front
with each other
Torque driver
(The figure shows the case when
(flat blade)
the mounting brackets are used
Attachment screw
on the top and bottom of the case.)
For details about the CX2000 external dimensions, installation environment,
and more, please refer to the CX2000 User’s Manual (IM 04L31A01-01E)
found on the provided CD-ROM.
4
*
Connection Types and Procedures
There are various terminals and connectors on the rear panel of the CX2000. Connecting them to peripheral devices allows you to perform control and measurement operations. Below are the names of each connector and terminal, as
well as connection procedures.
Connecting the Ethernet Interface
•
When only Connecting to a Hub
Connect the CX2000 and the PC through a HUB as shown in the
following figure.
10 Base-T straight cable
Ethernet
interface connector
(Multiple devices can be
CX
connected by using a hub.)
HUB
PC
Ethernet
interface connector
CX
Ethernet cable
10 Base-T straight cable
•
When Connecting to a Preexisting Network
The following figure illustrates an example in which the CX2000 and
a PC are connected to the network. When connecting the CX2000 or
the PC to a preexisting network such things as the transfer rate and
connector type must match. For details, consult your system or
network administrator.
Network
10 Base-T straight cable
(Adapters that support
Ethernet
the 10 Base-T (hubs and
interface connector
routers, for example)
CX
Ethernet cable
Connecting a Monitor to the VGA Output Terminal
(Option /D5)
Supported monitors: VGA monitors and multisync monitors with VGA
display capability.
1. Turn off the CX2000 and the monitor.
2. Connect the CX2000 and the monitor using an analog RGB cable (D-
Sub connector).
3. Turn on the CX2000 and the monitor. The screen of the CX2000 is
displayed on the monitor.
Loop 1 and 2 control output terminal block
Magnet
switch
Loop 3 and 4 control output terminal block
Loop 5 and 6 control output terminal block
Option terminal block (Options /A6, /A4F, /A4FR,
/A6R, /CST1,and /TPS4)
Connecting the Serial Interface
•
RS-232 (When Connecting to a Computer or Other Such Devices)
Verify that the CX2000 has an RS-232 connector, and then connect a serial cable to it. Connect the other end of the serial cable to the other device.
1
2
3 4
5
Pin No. Signal Name
2
3
5
7
6
7
8
9
8
* Pins 1, 4, 6, and 9 are not used.
•
RS-422/485 (When Connecting to a PLC, Temperature Controller or
Other Such Devices)
Verify that the CX2000 has an RS-422/485 connector, and then connect the crimp
connectors (for 4-mm screws) to the terminal strip as illustrated on the right. Do not
expose more than 5 cm of the cable surface from the shield.
With an RS-422/485
PC
CX2000 Rear Panel
VIDEO OUT(VGA)
R1
SSR
Contact
Contact
input (6)
output (6)
R2
Connecting Input/Output Connectors
Magnet
Contact
Contact
switch
input (6)
output (6)
When connecting various input/output connectors for control or
measurement, proceed as described below.
Connect the crimp connectors (for 4-mm screws) to the terminal
strip.
R3
Contact
Contact
1. Turn off the CX2000 and remove the terminal cover.
input (6)
output (6)
2. Wire the signal wires to the terminals.
Attach the terminal cover and secure it with the screws.
Signal Meaning
RD (Received Data)
Received data from the connected device. Input signal.
SD (Send Data)
Send data to the connected device. Output signal.
SG (Signal Ground)
Signal ground.
RS (Request to Send) Handshaking signal used when receiving data from the connected device. Output signal.
CS (Clear to Send)
Handshaking signal used when sending data to the connected device. Input signal.
Four-wire
Shield potential
Connecting the Power Supply
When using electrical wiring, be certain to follow the safety
recommendations prescribed in the CX2000 User’s Manual.
Use a power supply that meets the following conditions:
Rated supply voltage:
Supply voltage range used:
With an RS-232
Rated supply voltage frequency:
Permitted supply voltage frequency range:
L N
Maximum power consumption:
1. Turn OFF the CX2000 and open the cover (transparent) for the
power supply wires.
2. Connect the power cord and the protective ground cord to the power
supply terminals.
Power cord
3. Close the cover (transparent) for the power supply wires and secure
it in place with screws.
For the power supply specifications and connection method of the 24VDC/AC
power supply option, please refer to the CX2000 User’s Manual.
Two-wire
FG
SG SDB SDA RDB RDA
FG
SG SDB SDA RDB RDA
Shield
Shield potential
Shield
100 to 240 VAC
90 to 132, 180 to 264 VAC
50/60 Hz
50/60 Hz ± 2%
75 VA (100 V), 106 VA (240 V)
100-240V
AC~
50/60Hz
106VA MAX
L
N
Protective grounding cord
Control input terminal block (loops 1, 2, 3, 4, 5,
and 6)
Measurement input terminal block (CH1 to 10)
Measurement input terminal block (CH11 to 20)
Document Information:
|
|
Pages Preview: |
Note for Owners:
Guidesimo.com webproject is not a service center of YOKOGAWA trademark and does not carries out works for diagnosis and repair of faulty YOKOGAWA CX2000 equipment. For quality services, please contact an official service center of YOKOGAWA company. On our website you can read and download documentation for your YOKOGAWA CX2000 device for free and familiarize yourself with the technical specifications of device.
More Controller Devices:
-
SEDEMAC GC1100
MANUAL FOR GC1100: GENSET CONTROLLER Product Manual Version #0.00 ABSTRACT This manual is intended as an information guide for operating SEDEMAC’s GC1100 genset controller. Doc #SED-MAN-GC1100-002 Date: 20-Apr-2018 …
GC1100 Controller, 70
-
Green Heron RT-21
1RT-21 Az-El Controller Manual addendum to RT-21 — September 25, 2018 Overview: The RT-21 Az-El controller consists of two RT-21 units with a shared power supply and shared chassis. The unit features a pair of new OLED displays for sharp, crisp viewing, two main RT-21 PCB each with their own RS232 port, USB port, and 8 position terminal strip on the rear panel. When teamed with the free GH Trac …
RT-21 Controller, 24
-
Hunter XC Hybrid
XC HYBRIDPROGRAMMING GUIDESETTING THE DATE AND TIME1. Turn the dial to the CURRENT TIME/DAY position.2. Use the or buttons to change the year. Press the button to proceed to setting the month.3. Use the or buttons to change the month. Press the button to proceed to setting the day.4. Use the or buttons to change the day. Press the button to proceed to setting the time.5. …
XC Hybrid Controller, 2
-
Wyrestorm NHD-SW-0501
5×1 IP Streaming Presentation Switcher/Scaler with ScreenLink™Wireless Collaboration, Mic Input and CEC controlNHD-SW-0501 Quickstart Guide IMPORTANT! Installation Requirements!• Visit the product page to download the latest rmware, document versions, and the WyreStorm Maintenance Suite conguration tools.• Install the latest rmware onto all NetworkHD devices in the system.• Net …
NHD-SW-0501 Switch, 4
-
Daikin RDT
Installation and Maintenance Manual IM 893-10Group: Applied Air SystemsPart Number: IM 893Date: July 2017RoofPak® Singlezone Heating and Cooling UnitsRPS/RDT/RFS/RCS 015D—140D with MicroTech® III Unit Controller and Refrigerant R-410A …
RDT Air Conditioner, 152
Recommended Documentation:
User’s
Manual
Yokogawa Electric Corporation
Model CX2000/CX2010/CX2020/CX2210/CX2200/CX2410/
CX2420/CX2610/CX2620
DAQSTATION CX2000
IM 04L31A01-01E
8th Edition
Service Manual DAQSTATION CX1000/CX2000 SM 04L31A01-01E SM 04L31A01-01E Yokogawa Electric Corporation 2nd Edition Important Notice to the User This manual contains information for servicing YOKOGAWA’s DAQSTATION CX1000/ CX2000. Check the serial number to confirm that this is the correct service manual for the instrument to be serviced. Do not use the wrong manual. Before any maintenance and servicing, read all safety precautions carefully. Only properly trained personnel may carry out the maintenance and servicing described in this service manual. Do not disassemble the instrument or its parts, unless otherwise clearly permitted by this service manual. Do not replace any part or assembly, unless otherwise clearly permitted by this service manual. In principle, Yokogawa Electric Corporation (YOKOGAWA) does not supply parts other than those listed in the customer maintenance parts list in this service manual (mainly modules and assemblies). Therefore if an assembly fails, the user should replace the whole assembly and not components within the assembly (see “Note”). If the user attempts to repair the instrument by replacing individual components within the assembly, YOKOGAWA assumes no responsibility for any consequences such as defects in instrument accuracy, functionality, reliability, or user safety hazards. YOKOGAWA does not offer more detailed maintenance and service information than that contained in this service manual. All reasonable efforts have been made to assure the accuracy of the content of this service manual. However, there may still be errors such as clerical errors or omissions. YOKOGAWA assumes no responsibility of any kind concerning the accuracy or contents of this service manual, nor for the consequences of any errors. All rights reserved. No part of this service manual may be reproduced in any form or by any means without the express written prior permission of YOKOGAWA. The contents of this manual are subject to change without notice. Note YOKOGAWA instruments have been designed in a way that the replacement of electronic parts can be done on an assembly (module) basis by the user. YOKOGAWA instruments have also been designed in a way that troubleshooting and replacement of any faulty assembly can be done easily and quickly. Therefore, YOKOGAWA strongly recommends replacing the entire assembly over replacing parts or components within the assembly. The reasons are as follows: • The instruments use high-performance microprocessors, large scale CMOS gate arrays, and surface-mount components to provide state-of-the-art performance and functions. • Repair of components can only be performed by specially trained and qualified maintenance personnel with special highly-accurate tools, including costly ones. • When taking the service life and cost of the instruments into consideration, the replacement of assemblies offers the user the possibility to use YOKOGAWA instruments more effectively and economically with a minimum in downtime. • Zip is a trademark or registerd trademark of Iomega Corporation in the United States and/or other countries. • Adobe and Acrobat are trademarks of Adobe System incorporated. Disk No. SM13 2nd Edition : August 2002 (YK) All Rights Reserved, Copyright © 2002, Yokogawa Electric Corporation SM 04L31A01-01E 1 Introduction This manual contains information for servicing YOKOGAWA’s DAQSTATION CX1000/ CX2000. Note This is the second edition of the manual, dated August 2002. WARNING This service manual is to be used by properly trained personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to the safety precautions prior to performing any service. Even if servicing is carried out according to this service manual, or by qualified personnel, YOKOGAWA assumes no responsibility for any result occurring from this servicing. Safety Precautions The following general safety precautions must be taken during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific WARNINGS given elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Yokogawa Electric Corporation assumes no liability for the customer’s failure to comply with these requirements. WARNING Use the Correct Power Supply Ensure the source voltage matches the voltage of the power supply before turning ON the power. Use the Correct Power Cord and Plug To prevent an electric shock or fire, be sure to use the power supply cord recommend by YOKOGAWA. The main power plug must be plugged in an outlet with a protective grounding terminal. Do not invalidate protection by using an extension cord without protective grounding. Connect the Protective Grounding Terminal The protective grounding terminal must be connected to ground to prevent an electric shock before turning ON the power. Do Not Impair the of Protective Grounding Never cut off the internal or external protective grounding wire or disconnect the wiring of the protective grounding terminal. Doing so creates a potential shock hazard. Do Not Operate with Defective Protective Grounding Do not operate the instrument if you suspect the protective grounding might be defective. 2 SM 04L31A01-01E Do Not Operate Near Flammable Materials Do not operate the instrument in the presence of flammable liquids or vapors. Operation of any electrical instrument in such an environment constitutes a safety hazard. Do Not Remove Any Covers There are some areas components inside the instrument containing high voltage. Do not remove any cover, if the power supply is connected. The cover should be removed by qualified personnel only. Ground the Instrument before Making External Connections Connect the protective grounding before connecting the instrument to a measurement or control unit. Firnish a switch (double-pole type) to separate the CX2000 from the main power supply in the power supply line. In addition, make sure to indicate that the switch is a power control for the CX2000 on the switch. Switch Specifications Steady-state current ranting:1 A or more, inrush current rating:60 A or more Connect a fuse between 2 A and 15 A in the power supply line. Safety Symbols Used on Equipment and in Manuals To avoid injury, death of personnel or damage to the instrument, the operator must refer to an explanation in the user’s manual. High temperature. To avoid injury caused by hot surfaces, the operator must not touch the heatsink. Protective grounding terminal, to protect against electrical shock. This symbol indicates that the terminal must be connected to ground before operation of equipment. This symbol represents a functional grounding terminal. Such terminals should not be used as a protective grounding terminal. WARNING A WARNING sign calls attention to a procedure, practice, or condition, that could result in the injury or death of personnel if not correctly performed or adhered to. CAUTION A CAUTION sign calls attention to a procedure, practice, or condition, that could result in damage to or the destruction of part of the instrument if not correctly performed or adhered to. SM 04L31A01-01E 3 Overview of This Manual This manual is meant to be used by qualified personnel only. Make sure to read the safety precautions at the beginning of this manual as well as the warnings and cautions contained in the chapters relevant to any servicing you may be carrying out. This manual contains the following chapters. 1 Principles of Operation Provides an introduction and safety considerations. 2 Testing Explains the tests for checking the performance of the instrument. 3 Adjustments Explains the adjustments which can be performed by users. 4 Replacing Parts Describes maintenance which can be performed by users. 5 Troubleshooting Presents procedures for troubleshooting and how to proceed in case parts need to be replaced. 6 Schematic Diagram Provides a system configuration diagram. 7 Customer Maintenance Parts List Contains exploded views and a list of replaceable parts. 8 Procedures for Disassembly (CX1000) Lists the steps required to remove parts from the instrument. Specifications are not included in this manual. For specifications, refer to IM 04L31A01-01E or IM 04L31A01-03E. 4 SM 04L31A01-01E 1 Contents Important Notice to the User ............................................................................................... 1 Introduction ......................................................................................................................... 2 2 Safety Precautions .............................................................................................................. 2 Safety Symbols Used on Equipment and in Manuals ......................................................... 3 Overview of This Manual .................................................................................................... 4 3 Chapter 1 Principles of Operation 1.1 Block Diagram of the CX1000/CX2000 ............................................................................ 1-1 1.2 Input Section .................................................................................................................... 1-2 4 Chapter 2 Testing 2.1 Overview of Tests ............................................................................................................. 2-1 2.2 2.3 Test Procedures ............................................................................................................... 2-3 CTRL Module Assembly Tests ....................................................................................... 2-10 2.4 DIO Module Assembly Test ............................................................................................ 2-11 5 6 Chapter 3 Adjustments 3.1 Calibration of the Measuring Instrument’s Input ............................................................... 3-1 3.2 Control Output Calibration ............................................................................................... 3-6 7 Chapter 4 Replacing Parts 4.1 Replacement of the Control Output Terminal Block (Module) .......................................... 4-1 8 Chapter 5 Troubleshooting 5.1 5.2 Procedure ........................................................................................................................ 5-1 Flow Chart ........................................................................................................................ 5-2 5.3 Troubleshooting Checklist ................................................................................................ 5-3 Chapter 6 Schematic Diagram 6.1 CX1000 Schematic Diagram ............................................................................................ 6-1 6.2 CX2000 Schematic Diagram ............................................................................................ 6-2 Chapter 7 Customer Maintenance Parts List 7.1 CX1000 Customer Maintenance Parts List ...................................................................... 7-1 7.2 7.3 CX1000 Standard Accessories ........................................................................................ 7-8 CX2000 Customer Maintenance Parts List ...................................................................... 7-9 7.4 CX2000 Standard Accessories ...................................................................................... 7-16 Chapter 8 Procedures for Disassembly (CX1000) SM 04L31A01-01E 8.1 Removing the Bezel ......................................................................................................... 8-1 8.2 8.3 Removing the Board Assemblies and the Drive ............................................................... 8-4 Removing the Keyboard .................................................................................................. 8-6 8.4 8.5 Replacing the Backlight ................................................................................................... 8-8 Removing the Terminal Assembly .................................................................................. 8-10 5 Chapter 1 Principles of Operation 1.1 1 Block Diagram of the CX1000/CX2000 Principles of Operation Block Diagram of the CX1000 CPU Data storage functions Control/Measurement Input section External storage Input terminal Internal memory Scanner assembly A/D assembly Display unit Calculation function Control Output section Measurement Alarm function Control Output Control Alarm function Contact Input Communication function Contact Output Remote function (optional) Measurement Alarm Relay (optional) Block Diagram of the CX2000 CPU Measurement Input section Data storage functions External storage Input terminal Scanner assembly A/D assembly Internal memory Display unit Control Input section VGA out function (optional) Calculation function Input terminal Measurement Alarm function Scanner assembly A/D assembly Control Alarm function Communication function Control Output section Remote function (optional) Measurement Alarm Relay (optional) Transmitter power supply (optional) Control - purpose DIO (optional) Control Output Contact Input Contact Output Refer for details see schematic diagram page 6-1 and 6-2. IM 04L31A01-01E 1-1 1.2 Input Section A/D Assembly The A/D assembly has items such as a programmable gain amp, voltage reference, PWM modulator, current source for RTD measurements, differential amp, voltage source for RJC, serial parallel converter, control logic, and an occurred scanner SSR control signal. The A/D assembly uses a sinewave oscillating type self-resonant switching power supply (DC/DC converter), and noise filtering is achieved by signal integration. The A/D assembly detects the frequency of the power while it is ON and the integrated time becomes 20 ms or 16.67 ms. Therefore it carries a very high rate of noise rejection for the power frequency (in auto mode). In case the power frequency of the instrument and of the measured object are different, the appropriate integrated time is manually selectable. In case of the CX2000, the selection of 100 ms for 50/60 Hz is also available. A 16 bit resolution is achieved regardless of the integrated time. Input Terminal The internal printboard is isothermal because a print board with a metal core is being used. Therefore, stable reference junction compensation is realized. Scanner Assembly An in-house SSR (solid state relay) is being used for the scanner. The SSR, having a semiconductor switch, has a withstand voltage as high as 1500 V and a leakage current of only 1 nA. For that reason, it has the following features. • Semi-infinite life due to the absence of mechanical contacts • Silent operation • No occurrence of thermoelectric power. On the other hand, compared to a mechanical relay, the SSR has, the disadvantage of a bigger ON resistance and OFF capacity. As a result, RTD measurement and noise resistance characteristics are affected. Regarding RTD measurements, a differential amp was inserted into the previously mentioned analog circuit without increasing the number of parts, so that it would receive no influence from ON resistance. For RTD measurements in Measurement input section there is generally no insulation between channels. For RTD measurements in Control input section there is generally insulation between channels. Data Storage Functions For storing data, the CX1000/CX2000 has 1.2 MB of internal memory and is equipped with a Zip drive, or an ATA flash memory card drive. The measured data can also be saved to external storage media such as floppy disks, Zip disks, and ATA flash memory cards. Display Unit The CX has a 5.5-inch (CX1000) or 10.4-inch (CX2000) TFT color LCD on which it displays the measured results (240 (vertical) × 320 (horizontal) pixels for the CX1000 or 480 (vertical) × 640 (horizontal) pixels for the CX2000). 1-2 IM 04L31A01-01E 1.2 Input Section 1 Calculation Function Measurement Alarm Function The following eight alarm types can be set. High limit (H), low limit (L), differential high limit (h), differential low limit (l), rate-ofchange on increase (R), rate-of-change on decrease (r), alarm delay upper limit alarm (T), or alarm delay lower limit alarm (t). Control Alarm Function The following nine alarm types can be set. PV high-limit, PV low-limit, Deviation high-limit, Deviation low-limit, Deviation high & low limit, SP high-limit, SP low-limit, Output high-limit, Output low-limit Other Functions • Communication Function: Ethernet (standard) RS-232/RS-422A interface added (optional). • Remote Function: The trigger, start/stop, time adjustment, and other functions can be controlled remotely (optional). • Measurement Alarm Relay: Measurement alarm output and memory end/fail output (option for CX1006 and CX2000). • Transmitter Power Supply: DC24 V output for transmitter (only CX2000’s optional). • VGA Out Function (Option for CX2000 Only): The instrument's screen can be displayed on an external monitor via VGA output. • Control - Purpose DIO (CX2000 Only): Contact input Input for designated operations such as start/stop. Activate using a no-voltage contact or open collector signal. This is a 12-point input. Contact output Control Alarm output consists of 12 points of transistor contact output. Control Output Section Control Output The following types are available for universal control output. • Current output (continuous PID control output) Continuously output a current (analog signal) proportional to the calculated PID values. • Time proportional PID voltage pulse output Output an ON/OFF signal, having a pulse width proportional to the time, as a voltage per the calculated PID values. • On/Off control relay contact output • Output an ON/OFF signal, having a pulse width proportional to the time, at a relay contact point per the calculated PID values. • Output an ON/OFF signal to a relay contact point corresponding to the sign (+/–) of the deviation in the measured value from the specified target value. Contact Input Input for designated operations such as start/stop. Activate using a no-voltage contact or open collector signal. Contact Output Control Alarm output consists of relay contact output and transistor contact output. IM 04L31A01-01E 1-3 Principles of Operation The CX1000/CX2000 performs differential computation, linear scaling, and square roots using a microprocessor on the CPU board. Chapter 2 Testing 2.1 Overview of Tests Operating Conditions Ambient Temperature: Relative Humidity: 23 ± 2°C 55 ± 20% Test Instruments Instrument Specifications DC voltage generator Accuracy: 0.005% of setting + 1 mV DMM Accuracy: 0.005% of rdg + 1 mV Resistors Accuracy: 0.01% or better Insulation tester 500 VDC Withstanding voltage tester AC 1 to 3 kV, 500 VDC External monitor (for test of /D5 option) VGA monitor (H: 33.3 kHz, V: 60.168 Hz) Oscilloscope 200 kS/s or more, isolated input Thermostatic chamber ZC-114 (Coper Electronics Co., Ltd.) or equivalent Testing Conditions The tests cover all included A/D converters. The unit's analog input is in analog multiplexer format. For the CX1000, channel group 1–6 of the measurement input section and channel group 7–11 of the control input section are each assigned to one A/D converter. For the CX2000, channel groups 1–5 and 6–10 of the control input section (slot 1) and channel groups 1–10 and 11-20 of the measurement input section (slots 2 and 3) are each assigned to one A/D converter. Therefore, except for when specifying inputs linked to the same A/D converter, only one arbitrary channel within a group need be tested (for example, not channels 1—5 but only channel 1). SM 04L31A01-01E 2-1 2 Testing The following describes general testing procedures for DAQSTATION CX1000/CX2000 series instruments. For tests on specific modules or assemblies, see sections 2.3 and later. 2.1 Overview of Tests Tests Insulation Resistance Test Withstand Voltage Test Measurement Accuracy Test Error between Channels Test Excessive Input Test Burnout Test Reference Junction Compensation Accuracy Test Display Function Test VGA Output Function Test (For /D5 Option Only) Serial Communications Function Test (Only When The -1 Or -2 Suffix Code Is Specified For The Communications Port). Battery Backup Function Test Continuous Operation Test Current Output Accuracy Test Voltage Pulse Output Test Control Output Relay Test DIO Test KEY Function Test Media Function Test Alarm Relay Output Function Test (Only If The /A6, /A6R, /A4F, Or /A4FR Option Is Installed) Remote Function Test (Only If The /A6R Or /A4FR Option Is Installed) Communications Function Test (Ethernet) Test of 24 VDC Transmitter Power Output (Only If The /TPS4 Is Installed) Consult your nearest Yokogawa representative regarding the following tests. Power Supply Frequency Detection Function Test Memory Test 2-2 SM 04L31A01-01E 2.2 Test Procedures Insulation Resistance Test 2 Terminals Reference Values Notes Power terminal to earth terminal Measurement input terminal to earth terminal Control input terminal to earth terminal 100 MΩ or higher Short all channels prior to test RS-422-A/485 SG terminal to RS-422-A/485 FG terminal 100 MΩ or higher Test only if the basic specification code for the communications port is -2. Ethernet input/output terminal to earth terminal 100 MΩ or higher Short all terminals prior to test Withstand Voltage Test Perform the test using a withstanding voltage tester and confirm that the results meet the criteria below, and that the instrument does not malfunction. Terminal Reference Values AC power terminal to earth terminal* Leakage current of 10 mA or less at 1.5 kV AC for 1 minute DC power terminal to earth terminal (/P1)* Leakage current of 10 mA or less at 0.5 kV AC for 1 minute Measurement input terminal to earth terminal† Leakage current of 2 mA or less at 1.5 kV AC for 1 minute Control input terminal to earth terminal† Leakage current of 2 mA or less at 1.5 kV AC for 1 minute Between measuring input terminals‡ Leakage current of 1 mA or less at 1 kV AC for 1 minute Between control input terminals‡ Leakage current of 1 mA or less at 1 kV AC for 1 minute § Current and voltage pulse to earth terminal Leakage current of 2 mA or less at 1 kV AC for 1 minute Control relay terminal to earth terminal (2, 4, and 6 loop models)|| Leakage current of 2 mA or less at 1.5 kV AC for 1 minute DO relay terminal to earth terminal (2, 4, and 6 loop models)# Leakage current of 2 mA or less at 1.5 kV AC for 1 minute DO(Tr) terminal to earth terminal (2, 4, and 6 loop models or /CST1)** Leakage current of 2 mA or less at 0.5 kV AC for 1 minute D1 terminal to earth terminal (2, 4, and 6 loop models or /CST1)†† Leakage current of 2 mA or less at 0.5 kV AC for 1 minute Alarm relay terminal to earth terminal (/A6, /A6R, /A4F, /A4FR)‡‡ Leakage current of 2 mA or less at 1.5 kV AC for 1 minute Remote terminal to earth terminal (/A6R, /A4FR)§§ Leakage current of 2 mA or less at 0.5 kV AC for 1 minute 24 V transmitter power supply output to earth terminal Leakage current of 10 mA or less at 0.5 kV AC for 1 minute (/TPS4)|||| * Short L and N (or +/– with the /P1 option) † Short all channels ‡ Short the + and – input terminals (except for the RTD b terminal) Short the even and odd channels (for example, short channels 1-3-5 and 2-4-6 on the CX1000 or channels 1-35-7-9 and channels 2-4-6-8-10 for the CX2000), and test the withstanding voltage between them. § Short mA, PULS, and C. || Short NO, NC, and C on CTRL OUT. # Short 1NO, 1C, and 2NO on DO. ** Short DO3—DO6 and C. †† Short all DI terminals. ‡‡ Short all alarm relay terminals. §§ Short all remote terminals. |||| Short all transmitter power supply output terminals (+/–). SM 04L31A01-01E 2-3 Testing Perform this test using a DC 500 V insulation resistance meter and confirm that the results meet the criteria below. 2.2 Test Procedures Measurement Accuracy Test (Measurement Input and Control Input) Check that the specifications below are met given the following operating conditions: 23 ± 2°C, 55 ± 10% RH, and a warm up time of 30 minutes. Range Input Value Criterion Specifications 20 mV +20.00 mV 0.00 mV –20.00 mV ±3 digits ±2 digits ±3 digits ±(0.1% of rdg + 2 digits) 60 mV +60.00 mV 0.00 mV –60.00 mV ±6 digits ±2 digits ±6 digits ±(0.1% of rdg + 2 digits) 200 mV +200.0 mV 0.0 mV –200.0 mV ±3 digits ±1 digits ±3 digits ±(0.1% of rdg + 2 digits) 2V +2.000 V 0.000 V –2.000 V ±3 digits ±1 digits ±3 digits ±(0.1% of rdg + 2 digits) 6V +6.000 V 0.000 V –6.000 V ±6 digits ±1 digits ±6 digits ±(0.1% of rdg + 2 digits) 20 V +20.00 V 0.00 V –20.00 V ±3 digits ±1 digits ±3 digits ±(0.1% of rdg + 2 digits) 50 V +30.00 V 0.00 V –30.00 V ±5 digits ±2 digits ±5 digits ±(0.1% of rdg + 3 digits) Pt100 18.52Ω/–200°C 100.00Ω/0°C 313.71Ω/600°C ±0.4°C ±0.2°C ±0.9°C ±(0.15% of rdg + 0.3°C) Error between Channels Test Connect the (+), (–), and (b) terminals on all measurement and control input channels using thick leads, then check whether the following criteria are met. Range Input Value Criterion 20 mV Short (+) (–) 0.00 mV ±2 digits Pt100 103.9 Ω Measured value: 10°C ± 0.2°C,difference between channels: 0.2°C or less Note • • If the measurement error does not meet the criterion (10°C ± 0.2°C), and unless you determine the cause to be the effect of the resistance in the test facility's RTD wiring, use the following measurement criteria. Measured value: 10°C ± 0.5°C Difference between channels: 0.2°C or less See your nearest Yokogawa representative for the error between channels in the control input section. Excessive Input Test 1. 2. 3. 4. 5. 2-4 Connect the plus (+) and minus (–) sides of all measurement and control input channels. Set all channels to the 20 mV range. Apply 10 V between the plus and minus sides for 1 minute, and confirm that there is no malfunction. Set all channels to the Pt100 range. Apply ±10 V between (A) and (B) for 1 minute, and confirm that there is no malfunction. SM 04L31A01-01E 2.2 Test Procedures Burnout Test 1. 2. 3. 4. 5. 6. 7. Reference Junction Compensation Accuracy Test Perform 0°C measurement for TYPE - T on all control and measurement channels, and confirm that the result is below the following reference value. Reference value: 0 ± 0.5°C CAUTION • • • • • • • Use a calibrated non-insulated thermocouple without a terminal tip, having a Φ0.5 mm or narrower strand. Also, be sure to take the level of calibration error into consideration during testing. Continuously monitor the thermostatic chamber, and check that the temperature remains at 0°C ± 0.01°C. When using a 0°C thermostatic chamber (for example the ZC-114 by Coper Electronics Co. Ltd.) raise the tip of the thermocouple up 10 mm from the bottom. Conduct all tests in a stable environment. Install a windbreak if necessary. Allow a 15 minute warm up after wiring the thermocouple. Always use the terminal covers. Display Function Check the adjustment of the display color and backlight intensity using the following procedure. 1. Press the MENU key twice to enter Set mode. The Set mode screen is displayed. 2. 3. 4. SM 04L31A01-01E Press the Display soft key. Select Color. The default color and color name is displayed for each channel; compare the colors and their names to make sure there are no discrepancies. 2-5 2 Testing Connect a 2 kΩ resistor between the plus (+) and minus (–) sides of an arbitrary channel. Connect 56 kΩ resistor and a 0.1 µF capacitor in parallel between the plus (+) and minus (–) sides of an arbitrary channel other than the one in step 1. Set the range for the 2 connected channels to TC - TYPE K, and the Burnout to ON. Check that the channel to which 2 kΩ was connected displays a temperature close to the room temperature. Check that the channel to which the 56 kΩ resistor and the 0.1 µF capacitor were connected in parallel displays Burnout (overrange). Set Burnout to OFF. Check that both channels in steps 1 and 2 do not display Overrange. 2.2 Test Procedures 5. 6. 7. 8. 9. 10. Press ESC to return to the previous screen. Press the View, Direction, LCD soft key. Move the cursor to Brightness. For the CX1000, select 1 through 8 and confirm that each brightness level is brighter than the one before it. For the CX2000, select 1 through 4 and confirm that each brightness level is brighter than the one before it. During other phases of testing as well, always be checking for any abnormalities in the display that may appear. Set Brightness to 3. VGA Output Function Test (for the /D5 Option Only) Connect a VGA monitor to the VGA output terminal on the back of the CX2000, then look at the screen to check the VGA output. Perform steps 1—10 above for the display function and check the results. Serial Communications Function Test (Only If the -1 or -2 Suffix Code is Specified for the Communications Port) During this test, actual communications are performed to check whether the RS-232, RS-422A/485, or other functions are operating properly. However, if you used serial communications successfully to carry out other tests, you don't need to perform this test. Perform the tests using the procedure below. 1. Connect a cable between the PC and the CX1000/CX2000. If communications cannot be carried out via RS-422A/485, use an RS-232 adapter. 2. Send an arbitrary command from the PC using application software for CX1000/ CX2000 series instruments or other software, and confirm that the expected result occurs. Battery Backup Function Test 1. 2. 3. 4. While the CX1000/CX2000 is ON, set the date and time (see section 3.3, “Setting the Date and Time” in manual IM 04L31A01-01E or IM 04L31A01-03E). Turn the power OFF. Wait at least one minute, then turn the power back ON. Confirm that the date and time set in step 1 are correct. Continuous Operation Test 1. 2. 3. 4. 5. 2-6 Enter the following settings. • Enter AUTO SAVE to the media (see section 9.1 of the manual, IM 04L31A0101). • Set all channels to a range of 20 V (see section 4.6 of the manual, IM 04L31A0101). • Set input to open. • Set display type to waveform display (see section 1.16 of the manual, IM 04L31A0101). Press the START key. Run the instrument continuously for 24 hours or more. Press the STOP key. Check the following: • That a data file was created in the MEDIA INFO screen (see section 4.6 of the manual, IM 04L31A01-01E). • That there is no abnormal variation in the straight line waveform. • That there is no other strange sound or odor coming from the instrument. SM 04L31A01-01E 2.2 Test Procedures Current Output Accuracy 1. 2. 4. 5. 6. 7. Output Ratio Reference Value 0.0% 4 mA 100.0% 20 mA Allowable Range 3.984 to 4.016 mA 19.984 to 20.016 mA (Load condition: 250 Ω ±1%) Voltage Pulse Output 1. 2. 3. 4. Choose Output processing by following the procedures in step 1 of “Current Output Accuracy.” Set the type of control output to Voltage-pulse and the cycle time to 1 second for all the loops, save the settings, and change the mode to normal (the type of analog output can be set freely). Change the operation mode to manual and start control operation by following the procedures of steps 5 and 6 of “Current Output Accuracy.” Press the OUT softkey. Set the output ratio to 20.0% with the up and down arrow keys and observe the output voltage waveform on an oscilloscope or the like. Verify that a rectangular waveform is displayed at a 1-second interval (0.2second high and 0.8-second low) and 12 V ± 5%. Control Relay Output 1. 2. 3. 4. Choose Output processing by following the procedures in step 1 of “Current Output Accuracy.” Set the type of control output to On/Off-control for all the loops, save the settings, and change the mode to normal (the cycle time and type of analog output can be set freely). Change the operation mode to manual and start control operation by following the procedures of steps 5 and 6 of “Current Output Accuracy.” Press the OUT softkey. Set the output ratio to 0.0% and 100.0% with the up and down arrow keys and verify that the relay outputs are as follows: Output Ratio SM 04L31A01-01E NO-C NC-C 0.0% Break Make 100.0% Make Break 2-7 2 Testing 3. Change the mode to setup and choose Control, Control action, and then Input setting. Set the control mode to Single. Press the ESC key to return to the preceding menu, and choose Output processing. Set the type of control output to Current-output and analog output to 4-20 mA for all the loops, save the settings, and change the mode to normal (the cycle time can be set freely). Press the DISP/ENTER key and choose Control and then Control groups to display the control display. Choose a loop number to measure with a blue arrow using the arrow keys. Press the MODE softkey, and change the operation mode to manual by choosing MAN with the up and down arrow keys and pressing DISP/ENTER key to confirm it. Press the RUN/STP softkey, and start control operation by choosing RUN with the up and down arrow keys and pressing DISP/ENTER key to confirm it. Press the OUT softkey. Set the output ratio to 0.0% and 100.0% with the up and down arrow keys and measure each output current with an ammeter or the like. Verify that the measured currents are within the ranges as shown in the table below. 2.2 Test Procedures Test of Digital Input/Output (not applicable for models CXx0xx) Digital Input 1. Press the DISP/ENTER key and choose CONTROL and then DI/DO STATUS. 2. The digital input terminals are divided into blocks. Short-circuit between the common terminal and the terminals, and check the DI status display. The indicator is green for an open circuit and red for a short circuit. Digital Output 1. Short-circuit input terminals of all channels and adjust the range and alarm setting as follows: Range Mode Voltage Range 2V Span –2.000 to 2.000 On Alarm (Level 1) 2. Type H Value –1.000 Prepare the same input terminals step 1, and adjust the alarm relay output as follows: On Relay Output Alarm (Level 1) 3. No. DOxxx and ROxxx where xxx: digital output number to test Verify that a digital output turns on using an appropriate tester with the following circuit: I ≤ 50 mA R DO terminal LED V ≤ 24 V DC Common Test of Power Supply Voltage Switching Change the power supply voltage to 240 V AC and verify that the UUT display does not malfunction. Note that this test does not apply to the 24 V AC/DC-driven model (option code: /P1). Test of Ethernet Interface Use a computer in which the standard software “DAQSTANDARD for CX” is installed and that has an Ethernet communication function, for the following test. Run the launcher program. Select Ethernet in the network settings and make the communication settings. Then, verify that the settings of the UUT, such as the ranges, can be set up correctly as well as those settings can be read from the computer using the settings software. Test of Storage Drive 1. 2. 3. 2-8 Insert a formatted disk into the drive of the UUT. In the set mode*1, choose Save/Load, Clear data, and then Save settings. Save the current panel conditions under a desired filename. In the set mode, choose Save/Load, Clear data, and then File list. Verify that the file saved in step 2 exists in the list. SM 04L31A01-01E 2.2 Test Procedures Test of Storage Drive 1. 2. 3. Test of Alarm Relay Contact Outputs (applicable to option codes /A6, /A6R, /A4F and / A4FR) Insulation Resistance and Withstanding Voltage Tests Item Measured Point Specification Insulation resistance Between relay output terminals and grounding terminal No less than 20 MΩ at 500 V DC Withstanding voltage As above Free from damage after applying 1500 V AC, 50/60 Hz for 1 minute (with breaking leakage current set to 2 mA) Alarm Actions 1. Prepare the same input terminals and settings as step 1 in “Digital Output” (page 2-8), and also adjust the alarm relay output as follows: Relay Output On/off No. Ixx where xx: alarm output number to test Alarm (Level 1) 2. Verify that an alarm contact works as follows upon turning on/off the corresponding relay output. Terminals Normal During Alarm NO–C Break Make Where the output relay action is set to Remarks NC–C Make Break normally de-energized (factory set) Test of Remote Control (applicable to option codes /A6R and /A4FR) Insulation Resistance and Withstanding Voltage Tests Item Insulation resistance Withstanding voltage Measured Point Between remote control terminals and grounding terminal As above Specification No less than 20 MΩ at 500 V DC Free from damage after applying 500 V DC for 2 minute (with breaking leakage current set to 2 mA) Remote Control Actions Assign individual functions to 8 remote control inputs, then short-circuit each of those inputs in turn and verify that the CX1000/CX2000 is controlled as specified. Test of 24 VDC Transmitter Power Output (applicable to option code /TPS4) Test of Insuration Resistance and Withstanding Voltage Insulation resistance Between 24 VDC output terminals and No less than 20 MΩ at 500 VDC grounding terminal As above Free from damage after applying 500 VAC, 50/60 Hz for 1 minute (with breaking leakage current set to 10 mA) Withstanding voltage Between 24VDC output terminals Free from damage after applying 500 VAC, 50/60 Hz for 1 minute (with breaking leakage current set to 10 mA) SM 04L31A01-01E 2-9 2 Testing Insert a formatted disk into the drive of the UUT. In the set mode*1, choose Save/Load, Clear data, and then Save settings. Save the current panel conditions under a desired filename. In the set mode, choose Save/Load, Clear data, and then File list. Verify that the file saved in step 2 exists in the list. 2.3 CTRL Module Assembly Tests This section describes the test procedure for CTRL MODULE ASSY (B8700CL, B8700CM, B8700CN, and B8700FT) which is used on the CX2000. This test is not necessary if you will perform the general tests in section 2.2. Perform this test on the module by itself. Test Instruments Instrument Specifications DC current meter Accuracy: 0.01% of rdg Withstanding voltage tester AC 1 to 3 kV, 500 VDC Oscilloscope 200 kS/s or more, isolated input Resistor 250 Ω ±1% Jig A measurement instrument having the same functions and characteristics as the CX2000 or CX1000. Tests • Withstand Voltage Test Consult your nearest Yokogawa representative regarding the following tests. • Output Accuracy Test • Voltage Pulse Output Test • Control Relay Output Test • DI Test • DO Test Testing Environment Ambient temperature of 23±5°C, relative humidity of 55 ±20% CAUTION • Before starting the test, allow the instrument to warm up for 30 minutes or more. • Make sure the module is installed in the CX1000/CX2000. Test Procedure Perform a withstand voltage test on the module installed in the CX1000/CX2000 at the points listed below. Test Points Reference Values Current and voltage pulse output to earth terminal on the CX1000/CX2000* Leakage current of 2 mA or less at 1 kV AC for 1 minute Control relay terminal to CX1000/CX2000 earth terminal† Leakage current of 2 mA or less at 1.5 kV AC for 1 minute DO relay terminal to CX1000/CX2000 earth terminal‡ Leakage current of 2 mA or less at 1.5 kV AC for 1 minute § 2-10 DO (Tr) relay terminal to CX1000/CX2000 earth terminal Leakage current of 2 mA or less at 0.5 kV AC for 1 minute DI terminal to CX1000/CX2000 earth terminal|| * Short mA, PULS, and C. † Short NO, NC, and C. ‡ Short NO and C. § Short DO3—6 and C. || Short all DIs. Leakage current of 2 mA or less at 0.5 kV AC for 1 minute SM 04L31A01-01E 2.4 DIO Module Assembly Test Test Instruments Instrument Specifications Withstanding voltage tester AC 1 to 3 kV, 500 VDC Jig A measurement instrument having the same functions and characteristics as the CX1000/CX2000 Tests • Withstand Voltage Test Consult your nearest Yokogawa representative regarding the following tests. • DI Test • DIO Test Testing Environment Ambient temperature of 23 ±5°C, relative humidity of 55 ±20% CAUTION • The following assumes that the test will be performed with the module installed in the CX1000/CX2000. • The test cannot be performed on models having no internal control loops with no DIO module installed. Test Procedure Perform a withstand voltage test on the module installed in the CX1000/CX2000 at the points listed below. Test Points Reference Values DO terminal to CX1000/CX2000 earth terminal* † DI terminal to CX1000/CX2000 earth terminal * Short DO1—12 and C. † Short all DIs. SM 04L31A01-01E Leakage current of 2 mA or less at 0.5 kV AC for 1 minute Leakage current of 2 mA or less at 0.5 kV AC for 1 minute 2-11 2 Testing This section describes the test procedure for CTRL MODULE ASSY (B8700CY) which is used on the CX1000/CX2000. This test is not necessary if you perform the general tests in section 2.2. Perform this test when testing the module by itself. Chapter 3 Adjustments 3.1 Calibration of the Measuring Instrument’s Input Allow the instrument to warm up for at least 30 minutes prior to calibration. Instruments for Calibration Accuracy: 0.005% of setting +1 µV 100 Ω, 300 Ω (accuracy: 0.01% or less) DC voltage generator Resistor 3 Adjustments Overview The structure of the A/D converter in the CX2000 differs depending on the model. Multiple channels (groups) share one A/D converter. All A/D converters must be calibrated, so calibration must be performed on each group. CX1000 Monitor No. of Inputs for Meas. Measurement Interval Total Number of Channels No. of CH/AD No. of A/D 6ch 1s 6 6 1 * Group: a classification for all channels that share a single A/D converter Group (1—6) Control (1st Slot) Number of Internal Control Loops Measurement interval Total Number of Channels No. of CH/AD No. of A/D Group 0 loops — 0 0 0 — 2 loops 250 mS 5 5 1 (1—5) Measurement interval Total Number of Channels No. of CH/AD No. of A/D Group* CX2000 Control (1st Slot) Number of Internal Control Loops 0 loops — 0 0 0 — 2 loops 250 mS 5 5 1 (1—5) 4 or 6 loops 250 mS 10 5 2 * Group: a classification for all channels that share a single A/D converter (1—5) (6—10) Monitor (2nd, 3rd Slot) No. of Inputs for Meas. Measurement Interval Total Number of Channels No. of CH/AD No. of A/D Group 10 ch 1s 10 10 1 (1—10) 20 ch 1s 20 10 2 (1—10) (11—20) Ranges to Be Calibrated and Their Parameters Range Zero Full Scale 20 mV 0 mV 20 mV 60 mV 0 mV 60 mV 200 mV 0 mV 200 mV 1V 0V 1V 2V 0V 2V 6V 0V 6V 20 V 0V 20 V Pt100† 100 Ω 300 Ω † You must calibrate the 200 mV range before calibrating the RTD range. SM 04L31A01-01E 3-1 3.1 Calibration of the Measuring Instrument's Input Calibrating the Input Range For each group, input the zero and full scale for the input range to the channels below and perform calibration. CX1000 • Voltage Range Control Number of Internal Group 2 (1—5ch) Control Loops Zero Full Scale 0 loops — — 2 loops CH1* CH2* * The expression channel is not used for control input, but for expediency the inputs are referred to (from right to left) as CH1 and CH2 respectively. Monitor No. of Inputs for Meas. Group 1 (1—6ch) Zero Full Scale 6ch CH1 CH2 Zero Full Scale • RTD Range Control Number of Internal Group 2 (1—5ch) Control Loops 100 Ω 300 Ω 0 loops — — 2 loops CH1* CH2* * The expression channel is not used for control input, but for expediency the inputs are referred to (from right to left) as CH1 and CH2 respectively. Monitor No. of Measurement Group 1 (1—6ch) Input Channels 100 Ω 300 Ω 6ch CH1 CH2 CX2000 • Voltage Range Control (1st Slot) Number of Internal Group 1 Group 2 Control Loops Zero Full Scale Zero Full Scale 0 loops — — — — 2 loops CH1 CH2 — — Loops 4,6 CH1 CH2 CH6 CH7 Monitor (2nd, 3rd Slot) No. of Inputs for Meas. 3-2 Group 1 Group 2 Zero Full Scale Zero 10 ch CH1 CH2 — Full Scale — 20 ch CH1 CH2 CH11 CH12 SM 04L31A01-01E 3.1 Calibration of the Measuring Instrument's Input • RTD Range Control (1st Slot) Number of Internal Group 1 Group 2 Control Loops 100 Ω 0 loops — — — — 2 loops CH1 CH2 — — Loops 4, 6 CH1 CH2 CH6 CH7 300 Ω 100 Ω 300 Ω 3 Monitor (2nd, 3rd Slot) Group 1 Adjustments Model Group 2 100 Ω 300 Ω 100 Ω 300 Ω 10 ch CH1 CH2 — — 20 ch CH1 CH2 CH11 CH12 Calibration Procedure The calibration procedure can be carried out using keys on the instrument or communications commands. Calibrating Using Keys 1. Turn ON the unit while holding down the UP arrow key. The calibration screen appears. 2. Press the A/D Input Adjust soft key. The A/D adjust screen is displayed. 01-05, 06-10 shows control input, and 01-10, 11-20 shows measurement input. 3. SM 04L31A01-01E Select a group to calibrate (01—05, 06—10, 01—10, 11—20). 3-3 3.1 Calibration of the Measuring Instrument's Input 4. Press the ENTER key. The A/D adjust screen is displayed. 5. Input the zero and full scale value to the calibration channels shown in "Calibrating the Input Range" (page 3-2). Press the Cal/Exec soft key. The A/D adjust (Cal/Exe) screen is displayed. 6. CAUTION If you select Display in the A/D adjust screen, you can confirm the calibrated value and manually input values using keys. Key-based manual calibration has a large effect on measurement accuracy, so only perform the procedure if you think it is absolutely necessary. 7. 8. 9. 10. 11. 12. 13. 14. 3-4 Press the soft key corresponding to the range to be calibrated (20 mV, 60 mV, 200 mV etc.). Press the ENTER key. The calibration begins. Repeat steps 5—8 and calibrate all ranges. Press ESC. You are returned to the A/D adjust screen. Press the End soft key. A dialog box appears asking you whether or not to save and exit. Select Yes. You are returned to the A/D adjust (Cal/Exe) screen. Repeat steps 3—12 to calibrate and save calibrated values for each group. When you are finished with all procedures, turn the power OFF. SM 04L31A01-01E 3.1 Calibration of the Measuring Instrument's Input Calibration Using Communications The calibration process proceeds as follows: 1. Enter Calibration mode 2. Perform calibration 3. Store the calibrated values 4. Return to operation mode The commands necessary for calibration are listed below. For the input connection method, see "Input Range Calibration" on page 3-2. DSp1 Switch to setting mode p1 0 Switch to operation mode 1 Switch to Set up mode 2 Switch to A/D calibration mode p2 p3 Perform A/D calibration (Cal/Exe) Group For control input: 01—05 for the CX1000, and 01—05 or 06—10 for the CX2000 For measurement input: 01—06 for the CX1000, and 01—10 or 11—20 for the CX2000 Type of operation Cal/Exec Calibration range 20 mV, 60 mV, 200 mV, 1 V, 2 V, 6 V, 20 V, Pt100 XZp1,p2,p3,p4,p5 Manual correction of A/D calibration value (Display) p1 Group For control input: 01—05 for the CX1000, and 01—05 or 06—10 for the CX2000 For measurement input: 01—06 for the CX1000, and 01—10 or 11—20 for the CX2000 p2 Type of operation Display p3 Calibration range 20 mV, 60 mV, 200 mV, 1 V, 2 V, 6 V, 20 V, Pt100 p4 Zero input value p5 Full scale input value XZp1,p2,p3 p1 p2 p3 Store the calibrated values (exit) Group For control input: 01—05 for the CX1000, and 01—05 or 06—10 for the CX2000 For measurement input: 01—06 for the CX1000, and 01—10 or 11—20 for the CX2000 Type of operation: End Data save selection STORE Save to a file ABORT Don’t save XZp1,DISPLAY? Get information for calibrated values p1 Group For control input: 01—05 for the CX1000, and 01—05 or 06—10 for the CX2000 For measurement input: 01—06 for the CX1000, and 01—10 or 11—20 for the CX2000 SM 04L31A01-01E 3-5 Adjustments XZp1,p2,p3 p1 3 3.2 Control Output Calibration Allow the instrument to warm up for at least 30 minutes prior to calibration. Instruments for Calibration DC current meter Resistor Accuracy: 0.01% of rdg 250 Ω ± 1% Overview of Calibration During calibration of the control output (current output terminal), each output is connected to one D/A converter, so you must calibrate the outputs separately. The module installed in the CX1000 is 01. In the CX2000, the modules are numbered in order (01, 02, 03) from top to bottom. Also within each module, output number 01 is assigned to loop 1, and output 02 to loop 2. Calibration Procedure The calibration procedure can be carried out using keys or communications commands. Calibration Using Keys 1. Connect a 250 Ω ±1% resistor and DC current meter in serial to the current output terminal. 2. Turn ON the unit while holding down the UP arrow key. The calibration screen appears. 3. Press the Pid Analog Output Adjust soft key. The Pid Analog Output Adjust screen is displayed. 4. Use the arrow keys and soft keys to select the module number and output number to be calibrated. While monitoring with a current meter, change the calibration value at each Output Value (4 mA and 20 mA) so that it falls within the calibration range shown in the following chart. The first digit of the calibration value corresponds to 0.366211 µA. 5. 3-6 Output Value Calibration Range 4 mA 3.99 mA—4.01 mA 20mA 19.99 mA—20.01 mA SM 04L31A01-01E 3.2 Control Output Calibration 6. Press the Write function key for each output number. The calibration values are saved. 3 Adjustments 7. 8. 9. Repeat steps 4—6 and calibrate all module and output numbers. Press ESC. You are returned to the calibration screen. Turn OFF the power Calibration Using Communications The commands required for control output calibration are listed below. For other necessary commands, see Calibration Using Communications (page 3-5) in section 3.1, “Calibrating the Measuring Instrument Input.” Perform step 1 of Calibration Using Keys before using these commands for calibration. For OUTPUT ZZp1,p2,p3,p4 p1 Action OUTPUT p2 Module number (1 for the CX1000, or 1—3 for the CX2000) p3 Output channels (1 and 2) p4 4 mA/20 mA p5 Calibration value (For 4 mA, 9285—12561) (For 20 mA, 46421—62804) For WRITE ZZp1,p2,p3,p4 p1 p2 p3 p4 p5 Action (OUTPUT/WRITE) Module number (1 for the CX1000, or 1—3 for the CX2000) Output channels (1 and 2) (Calibration value for 4 mA, 9285—12561) (Calibration value for 20 mA, 46421—12561) Note When specifying WRITE, you can omit p4 and p5. If you do so, the currently set calibration value is written. SM 04L31A01-01E 3-7 Chapter 4 Replacing Parts 4.1 Replacement of the Control Output Terminal Block (Module) WARNING Replacement of terminal blocks and control relays are to be carried out by engineers authorized by YOKOGAWA. When replacement becomes necessary, please contact your nearest dealer or YOKOGAWA representative. 4 Replacing Parts Parts List CX1000 Part No. Name Option Code, Model B8700GA Alarm Module /A6 B8700GB Alarm Remote Module /A6R B8700GC Alarm Fail Module /A4F B8700GD Alarm Remote Fail Module /A4FR B8700FT CTRL Module 2 Loop CX1206 B8700FP CTRL MEAS 0 Loop Module Assembly CX1006 not /N2 B8700FQ CTRL MEAS 2 Loop Module Assembly CX1206 not /N2 B8700FR CTRL MEAS 0 RTD Module Assembly CX1006 /N2 B8700FS CTRL MEAS 2 RTD Module Assembly CX1206 /N2 CX2000 SM 04L31A01-01E Part No. Name Option Code, Model B8700CL CTRL 2 Loop Module Assembly CX22xx, CX24xx, CX26xx B8700CM CTRL 4 Loop Module Assembly CX24xx, CX26xx B8700CN CTRL 6 Loop Module Assembly CX24xx, CX26xx B8700CY DIO Module Assembly /CST1 B8700CX Alarm Terminal Assembly /A6 B8700CZ Alarm Terminal Assembly /A6R B8700CJ Alarm Terminal Assembly /TPS4 B8700DX Alarm Terminal Assembly /A4F B8700DZ Alarm Terminal Assembly /A4FR B8700CP CTRL Input 2 Loop Module Assembly CX22xx B8700CQ CTRL Input 4 Loop Module Assembly CX24xx B8700CR CTRL Input 6 Loop Module Assembly CX26xx B8700CS 10ch Input Module Assembly CX2x10, CX2x20 B8700CU 10ch 3RTD Input Assembly CX2x10, CX2x20 /N2 B8700CT 20ch Input Module Assembly CX2x20 B8700CV 20ch 3RTD Input Assembly CX2x20 /N3 4-1 4.1 Replacement of the Control Output Terminal Block (Module) Replacement of Each Terminal Block (Module) WARNING • To prevent electric shock, cut the power to the main unit and disconnect any wiring that may be connected to it before replacing the terminal block. • To prevent electric shock when disconnecting wires, ensure the main power supply is turned OFF. Follow the procedures below to replace the blocks. 1. Loosen (but do not completely remove) the 2 terminal cover screws, remove the cover, then remove the wiring from the terminal block. Ignore this step if the terminal block is not wired. Connector Connector Terminal cover screws Terminal cover screws Terminal block attachment screws 2. 3. 4. 5. Terminal block attachment screws Loosen the 2 terminal block attachment screws. Pull the terminal block straight out and away from the unit. Be careful not to bend the connector pins while removing the terminal block. Check the angle of the terminal block, install it on the main unit, and then fasten by tightening the terminal block attachment screws. Install the terminal cover, and fasten with the terminal cover screws. When Replacing Only the Control Relay WARNING • To prevent electric shock, unplug the main power cord before replacing the control relay. • To prevent electric shock when disconnecting wires, ensure the main power supply is turned OFF. • To prevent electric shock, use a withstanding voltage tester to check the withstanding voltage between the relay and the protective grounding after replacing the relay. Withstand voltage: 1500 VAC, 1 minute • You must perform an insulation resistance test and wishstand voltage test after replacing the control relay. Follow the procedures below to replace the blocks. 1. Loosen (but do not completely remove) the 2 terminal cover screws, remove the cover, then remove the wiring from the terminal block. Ignore this step if the terminal block is not wired. 4-2 SM 04L31A01-01E 4.1 Replacement of the Control Output Terminal Block (Module) Connector Connector Terminal cover screws Terminal cover screws Terminal block attachment screws Terminal block attachment screws 4. Loosen the 2 terminal block attachment screws. Pull the terminal block straight out and away from the unit. Be careful not to bend the connector pins while removing the terminal block. Open the back lid of the terminal block. The letters printed on the relays indicate the loops they correspond to: RL1 is for loops 1, 3, and 5, and RL2 is for loops 2, 4, and 6. Back lid 5. Using a tool such as a pair of needle-nosed pliers, grasp the relay and pull it directly outward. If you are not replacing the other relay, be careful not to damage it when removing the relay to be replaced. Push the Relay Lock Latch to the Left Control relay Needle-nosed pliers or a similar tool Replace the relay with the following: Matsushita Electronic Works, Ltd. Control Relay Part No.: DSP1-DC12V Web Site: http://www.nais-e.com/ SM 04L31A01-01E 4-3 4 Replacing Parts 2. 3. 4.1 Replacement of the Control Output Terminal Block (Module) 6. Insert the relay into the socket by hand, with the line on the top of the relay facing to the left. This line faces left 7. 8. Close the back lid of the terminal block. Check the angle of the terminal block, install it on the main unit, and then fasten by tightening the terminal block attachment screws. 9. Install the terminal cover, and fasten with the terminal cover screws. 1 0 . Confirm that the instrument functions properly, then perform a withstanding voltage test on the contact output terminal to protective ground of each control relay. 4-4 SM 04L31A01-01E Chapter 5 Troubleshooting 5.1 Procedure 1 2 3 4 Determine the type of problem. Check for possible user error. Check the connections and the settings of equipment to determine whether there was a handling mistake. Execute the self diagnostic test by turning the power ON, and identify any problem items. Analyze the cause of the problem according to the troubleshooting flow chart. Do not touch the circuit or parts with live voltage because the power unit contains a highvoltage electrical circuit. The p ower unit is furnished with a dedicated cover to prevent electric shock. Do not remove this cover. Never touch any part not subject to adjustment. 5 SM 04L05A01-01E Troubleshooting Make sure to connect input terminals (voltage or current) correctly. The internal circuit may be damaged when wrongly connected. 5-1 5.2 Flow Chart This flow chart consists of general service operations when a fault occurs. This chart is not always suitable for every kind of fault. However, it is recommended to perform operations according to the flow chart. START Phenomenon check NO Cause check YES Assembly exchange and adjustment Performance test execution Repaired? NO YES END 5-2 SM 04L05A01-01E Troubleshooting Checklist Operational Power cable connection Fuse is blown (/P1 option only) Power ass’y CPU ass’y Memory ass’y Display ass’y FAIL state CPU ass’y Memory ass’y Display ass’y Optional terminal ass’y Memory cannot be backed up Battery connector is disconnected? Battery voltage is low (less than +3.0V) CPU ass’y Memory ass’y Display ass’y LCD is not normal Measured value incorrect 5 Troubleshooting Power is not turned ON Panel key operation is not normal SM 04L05A01-01E Check Item Exchange Check Trouble Adjust 5.3 FFC ass’y of the keyboard is disconnected/broken Keyboard ass’y CPU ass’y Memory ass’y Display ass’y FFC ass’y of the LCD is disconnected/ broken CPU ass’y Memory ass’y Display ass’y LCD ass’y Input wiring is disconnected Noise A/D ass’y Scanner ass’y Measured temperature is incorrect Input is disconnected Noise Terminal cover is removed RJC INT/EXT setting A/D board ass’y Input terminal Scanner board ass’y Measured value fluctuates Power frequency setting is incorrect Noise External storage media is not normal Floppy disk/Zip disk/PC card drive unit 5-3 SM 04L31A01-01A CN8 CN101 CN7 CN1 CN3 CN13 CN4 CN15 CN2 B9967TF (AF/DC24V) A1485UP (STD) POWER SUPPLY CN1 B9967MC CN3 B9968MG CN2 B9967UD (AC/DC24V) B9967TD (STD) SUB POW BOARD ASSY CN3 CN4 POWER CABLE CN41 CN1 POWER CABLE CN42 SCREW CN3 CN1 CN16 OUTPUT MAIN BOARD ASSY B8700RG CN2 CN2 OUTPUT SUB BOARD ASSY B8700SH CN1 CN2 A1053VA LCD & INVERTER SW BOARD ASSY B9967TC CN1 CN3 KEY FPC B9967MB B8700RK DISPLAY BOARD ASSY CN3 CN1 B8700RA CN2 CN1 B9900BR BATTERY ASSY CN2 B8700UZ MEMORY BOARD ASSY CPU BOARD ASSY CN1 CN1 CN17 CN1 FDD DRIVE A1092UN IDE FFC B9968MB ZIP DRIVE A1131UN CN2 ZIP CONN BOARD ASSY B9968SU CN1 CN2 IDE BOARD ASSY B9968ST CN1 +RS-232 +RS-422 B9968TP TEST CN4 CN2 PC-CARD BOARD ASSY B9968SV CN1 MOTHER BOARD ASSY B9967TB B9968RK +RS-422 +RS-232 B9968RG STD B9968RJ B9968TQ B9968SQ STD FDD FFC B9968MA CN2 RS-422 TM1 COMM TERM BOARD ASSY ETHERNET RS-232 CN2 CN3 COMM BOARD ASSY CN10 CN1 FDD BOARD ASSY B9968SW ARARM BOARD ASSY B9968US (ALM) B9968UT (REM/ALM) SCANNER ASSY B8700RM (MON) B8700QM (ISO MON) B8700RS (CTRL. &MON) B8700QS (ISO CTRL. & MON) CN2 CN1 DISP FPC B9967MB AD BOARD ASSY B8700RP(MON) B8700QP (ISO MON) B8700RD(CTRL. &MON) B8700QD (ISO CTRL. & MON) CN2 CN14 CN1 CN12 B8700RN (MON) B8700QN (ISO MON) B8700RQ (CTRL. & MON) B8700QQ (ISO CTRL. & MON) TERMINAL ASSY CN1 CN5 B9968SF 6.1 Schematic Diagram RJC BOARD ASSY Chapter 6 Schematic Diagram CX1000 Schematic Diagram 6 6-1 6-2 CN2 CN14 CN4 CN13 CN15 CN3 CN11 CN41 CN1 VGA OUT CN3 CN2 CN1 B9968UX VGA BOARD ASSY CN2 DISP CABLE B9968MD AD BOARD ASSY B9968QA CN2 LCD A1067VA INVERTER UNIT A1527UP CN2 CN1 CN35 B9900BR BATTERY ASSY CN2 B8700UZ ZIP DRIVE A1131UN CN2 ZIP CONN BOARD ASSY B9968SU B9968MB B9968MA FDD DRIVE A1092UN IDE FFC CN1 CN2 FDD FFC CN2 +RS-232 CN1 CN2 PC-CARD BOARD ASSY B9968SV TM1 RS-422 B9968RK +RS-422 B9968RJ CN2 CN3 ETHERNET RS-232 IDE BOARD ASSY B9968ST CN1 B9968TP FDD BOARD ASSY B9968SW CN1 CN1 +RS-232 +RS-422 B9968TQ CN1 ALARM BOARD ASSY B9968US ALM B9968UT REM/ALM CN1 TPS4CH BOARD ASSY B9968QW TPS4CH ASSY MOTHER BOARD B9968SP(SCREW) TEST B9968RG STD CN4 B9968SQ STD CN5 CN1 CN1 CN9 DIO MAIN BOARD ASSY B8700RE DIO CN1 CN2 DIO SUB BOARD ASSY B8700RF DIO COMM BOARD ASSY CN1 CN8 OUTPUT MAIN BOARD ASSY B8700RG CN2 CN2 OUTPUT SUB BOARD ASSY B8700RH COMM TERM BOARD ASSY CN4 CN1 CN1 CN7 OUTPUT MAIN BOARD ASSY B8700RG OUTPUT MAIN BOARD ASSY B8700RG CN1 CN6 CN2 CN2 CN2 CN2 SCANER ASSY B9968SD B9968TY(ISO) CN4 CN33 OUTPUT SUB BOARD ASSY B8700RH OUTPUT SUB BOARD ASSY B8700RH MEMORY BOARD ASSY SW BOARD ASSY B9968SK CN3 INV CABLE B8700QK CN4 B8700RK KEY FFC B9968MC CN11 B8700RA DISPLAY BOARD ASSY CN3 CN1 DISP CABLE B9968MD DISP-VGA CABLE B9968MM A1484UP(STD) POWER SUPPLY CN1 B9968MH B9968MG CN3 CN3 CN31 CN36 CN37 CN1 CN101 CN2 CN34 TERMINAL ASSY B9968UF B9968UM(ISO) CPU BOARD ASSY CN1 CN32 CN1 CN5 B9968SF RJC BOARD ASSY SCANER ASSY B9968SD B9968TY(ISO) CN25 CN4 CN23 CN2 CN1 CN3 CN21 AD BOARD ASSY B9968QA POW CABLE CN2 B8700SL SCREW(CN1) SUB POW BOARD ASSY CN3 CN4 CN2 CN24 TERMINAL ASSY B9968UF B9968UM(ISO) CN26 CN27 CN1 CN101 CN1 CN22 POW CABLE VGA CABLE B9968MN CN42 SCREW SCANER ASSY AD BOARD ASSY B8700SJ(2LOOP) B8700RC(2LOOP) B8700RB(4,6LOOP) B8700RJ(4,6LOOP) CN16 CN17 CN1 CN101 CN1 CN12 TERMINAL ASSY B8700RR(2LOOP) B9968UM(4,6LOOP) CN1 CN5 B9968SF B9968SF CN1 CN5 RJC BOARD ASSY RJC BOARD ASSY 6.2 CX2000 Schematic Diagram SM 04L31A01-01A Chapter 7 Customer Maintenance Parts List 7.1 CX1000 Customer Maintenance Parts List 7 Customer Maintenance Parts List Note: Parts marked with a SM 04L31A01-01A symbol are CMPL (Customer Maintenance Parts List) parts. 7-1 7.1 CX1000 Customer Maintenance Parts List Complete Set 5 6 5 7 1 7-2 2 3 4 Item 1 2 3 4 5 Part No. B9967AM B8700EN Y9308LB B8700FA B9968AT Qty 1 1 2 1 4 6 7 B8700DR B9967AX 2 1 Description Tag Cover Tag Plate B.H. Screw, M3x8 Bezel Assembly (see page 7-5) Sheet Sheet Packing Note: CMPL Parts SM 04L31A01-01A 7.1 CX1000 Customer Maintenance Parts List 17 21 20 19 18 3 4 5 6 8 12 11 10 9 7 14 15 16 13 7 Customer Maintenance Parts List 22 1 2 35 36 37 38 39 40 41 42 27 28 29 30 26 24 25 23 32 33 34 31 SM 04L31A01-01A 7-3 7.1 CX1000 Customer Maintenance Parts List 2 Part No. Qty B8700RP 1 B8700RD 1 B8700QP 1 B8700QD 1 B8700RM 1 Description MON AD Assembly (CX1006 not /N2) C&M AD Assembly (CX1206 not /N2) (select) ISO MON AD Assembly (CX1006 /N2) ISO C&M AD Assembly (CX1206 /N2) MON Scanner Assembly (CX1006 not /N2) 3 4 B8700RS B8700QM B8700QS B8700FF B8700RA 1 1 1 1 1 C&M Scanner Assembly (CX1206 not /N2) (select) ISO MON Scanner Assembly (CX1006 /N2) ISO C&M Scanner Assembly (CX1206 /N2) CPU Board Assembly CPU Board Assembly 5 6 7 8 9 B9967TA B8700NP B8700EE B8700QZ B8700UZ 1 1 1 1 1 10 11 12 13 B9900BR B9968EM A9069KY B9967DA B9967DB 1 1 1 1 1 Battery Assembly Rivet Clamp COMM Board Assembly (CX1 COMM Board Assembly (CX1 B9967DD B9967DC B9968TQ B9968TP B9968SQ 1 1 1 1 1 COMM Board Assembly (CX1 6COMM Board Bracket COMM Board Assembly (CX1 COMM Board Assembly (CX1 COMM Board Assembly (CX1 B9968CZ B9968CZ B8700FG B9967DV B9967CW 1 2 1 1 1 B9967DW B8700SL B9967UD A1485UP B9967TF 1 1 1 1 1 Gel Sheet (CX1 6- -0), (CX1 6- -1) (select) Gel Sheet (CX1 6- -2) S-Power Assembly (not /P1) (select) S-Power Assembly (/P1) Power Bracket Cover (not /P1) (select) Power Bracket Cover (/P1) Sub Pow Board Assembly (not /P1) 24V Power Board Assembly (/P1) Power Supply (not /P1) (select) DC Power Board Assembly (/P1) 21 22 23 24 25 B9967CX B9967TB B9967GA B9967DM B9967DN 1 1 1 1 1 Power Bracket Base Mother Board Assembly FDD Drive Assembly (CX1 06-1) Sheet Sheet 26 27 28 29 30 B9968GA A1092UN Y9204LS B9968SW B9968MA 1 1 3 1 1 31 32 33 34 35 B9968GL A1492JS Y9208LB B9968SV B9967GD 1 1 2 1 1 ATA Drive Assembly (CX1 06-3) Socket Screw PC-Card Assembly Zip Drive Assembly (CX1 06-2) 36 37 38 39 40 B9967DM B9968GD A1150UN Y9203LB B9968SU 1 1 1 3 1 Sheet Zip Drive Assembly Memory System Screw Zip Conn Board Assembly 41 42 B9968ST B9968MB 1 1 Item 1 14 15 16 17 18 19 20 7-4 Display Board Assembly Memory Assembly Name Plate Memory Board & Battery Assembly Memory Board Assembly 6- -1) 6- -2) (select) -0) 6- -1) 6- -2) (select) 6- -0) FDD Drive Assembly Memory System Screw FDD Board Assembly FDD FCC IDE Board Assembly IDE FFC Note: CMPL Parts SM 04L31A01-01A 7.1 CX1000 Customer Maintenance Parts List Bezel Assembly 18c 20 18a 18 19 21 22 18b 7 Customer Maintenance Parts List 17 16 15 14 13 12 10 1 11 9 8 2 3 4 5 6 7 Item 1 2 3 4 5 Part No. Qty Description B8700FH 1 Key Case Assembly B9967BN 1 Hinge Pin B9567AQ 1 Spring B9967BP 1 Front Plate E9655AL 1 Spring Item 12 13 14 15 16 Part No. Qty Description B9967BU 2 Screw B9967AY 1 Packing B9967MA 1 Key FPC B9967BT 1 FPC Guard B9967BX 1 Rivet 6 7 8 9 10 B9967BM B9967BJ B8700FL B9967TC B9967BZ 1 1 1 1 1 Door Knob Front Cover Key Top SW Board Assembly Micro SW Pin 17 18 18a 18b 18c B8700FB B9967BF A1053VA B9967AQ A1039VZ 1 1 1 1 1 Sub Bezel Assembly LCD Assembly LCD Name Plate Back Light Module 11 B9967BK 1 Front Case 19 20 21 22 B9968BM B9967MB B9967BY B9968EN 2 1 1 1 Hinge Arm Display FFC Stay Bracket Bushing SM 04L31A01-01A 7-5 7.1 CX1000 Customer Maintenance Parts List 20 21 25 24 28 22 27 23 18 26 19 7 4 2 3 1 5 6 16 13 8 11 12 10 14 15 17 9 7-6 SM 04L31A01-01A 7.1 CX1000 Customer Maintenance Parts List Item Part No. Qty Description 1 B8700GA 1 Alarm Module (/A6) B8700GB 1 Alarm Remote Module (/A6R) (select) B8700GC 1 Alarm Fail Module (/A4F) B8700GD 1 Alarm Remote Fail Module (/A4FR) B8700FT 1 CTRL Module 2 Loop (CX1206) 9 2 B9968DN B8700CK B9968DJ B9968DF B9900SG 1 2 2 1 2 B9968DG B8700GF B8700GH B8700GK B8700GM 1 1 1 1 1 Cover Name Plate (/A6) Name Plate (/A6R) Name Plate (/A4F) Name Plate (/A4FR) B8700EW B8700GE B8700GG B8700GJ B8700GL 1 1 1 1 1 Name Plate (CX1206) Name Plate (/A6) Name Plate (/A6R) Name Plate (/A4F) (select) Name Plate (/A4FR) B8700EV B8700HJ B8700HK B8700HG B8700HH 1 1 1 1 1 Name Plate (CX1206) Name Plate (/A6) Name Plate (/A6R) Name Plate (/A4F) (select) Name Plate (/A4FR) B8700HL B8700FP B8700FQ B8700FR B8700FS 1 1 1 1 1 B9968DJ B9968DF B9900SG B9968DG B8700EH 2 1 2 1 1 Screw Cover Assembly Screw Cover Name Plate (CX1006 not /N2) B8700EL B8700EQ B8700ET B8700EG B8700EK 1 1 1 1 1 B8700EP B8700ES B8700EJ B8700EM B8700ER 1 1 1 1 1 Name Plate (CX1206 not /N2) (select) Name Plate (CX1006 /N2) Name Plate (CX1206 /N2) Name Plate (CX1006 not /N2) Name Plate (CX1206 not /N2) (select) Name Plate (CX1006 /N2) Name Plate (CX1206 /N2) Name Plate (CX1006 not /N2) Name Plate (CX1206 not /N2) (select) Name Plate (CX1006 /N2) B8700EU B8700ED B8700EF B9968EG B9968RJ 1 1 1 1 1 Name Plate (CX1206 /N2) Name Plate (not /P1) (select) Name Plate (/P1) Power Plate COMM Term Board Assembly (CX1 B9968RK B9968RG Y9308LB B9968AJ B9968AG 1 1 2 1 1 COMM Term Board Assembly (CX1 COMM Term Board Assembly (CX1 B.H. Screw, M3x8 Sheet (CX1 6- -0) Name Plate (CX1 6- -1) (select) 25 B9968AH Y9305LB B9968EH B9968EJ Y9305TS 1 1 1 1 1 Name Plate (CX1 6- -2) B.H. Screw, M3x5 Blind Bracket (CX1 6- -0) (select) RS-232 Bracket (CX1 6- -1) Tapping Screw (CX1 6- -1, -0) 26 27 28 Y9308LB B9968EE A1447JZ 1 1 1 B.H. Screw, M3x8 Terminal Modular Cover 3 4 5 6 7 8 10 16 17 18 19 20 21 22 23 24 SM 04L31A01-01A (select) Name Plate (CX1206) CTRL MEAS 0 Loop Module Assembly (CX1006 not /N2) CTRL MEAS 2 Loop Module Assembly (CX1206 not /N2) (select) CTRL MEAS 0 Loop RTD Module Assembly (CX1006 /N2) CTRL MEAS 2 Loop RTD Module Assembly (CX1206 /N2) 7 Customer Maintenance Parts List 11 12 13 14 15 Conn Cover Assembly Screw (CX1206) (select) Screw (/A6, /A6R, /A4F, /A4FR) Cover Assembly Screw 6- -1) 6- -2) 6- -0) (select) Note: CMPL Parts 7-7 7.2 CX1000 Standard Accessories 1 2 3 4 6 5 Item Part No. 1 IM04L31A01-03E IM04L31A01-04E IM04L31A01-72E IM04L31A01-73E 2 B9900BX 3 4 5 6 7 7-8 E9655FX A1053MP B9968PK CXA100-01 B8700MA Qty 1 1 1 1 2 5 1 1 1 1 7 Description DAQSTATION CX1000 User’s Manual DAQSTATION CX1000 Operation Guide Precautions on the Use of the CX1000/CX2000 CX1000 Installation and Connection Guide Bracket Assembly B.H. Screw, M4x6 ( ) Mag Memory :Zip 100MB disk (DX1 06-2) Memory System :ATA flash memory card (DX1 06-3) DAQ Standard for CX Note: CX1000/CX2000 electric manual CMPL Parts SM 04L31A01-01A 7.3 CX2000 Customer Maintenance Parts List 7 Customer Maintenance Parts List Note: Parts marked with a SM 04L31A01-01A symbol are CMPL (Customer Maintenance Parts List) parts. 7-9 7.3 CX2000 Customer Maintenance Parts List Complete Set 1 4 3 2 6 5 7 2 3 4 Part No. B8700BA B8700BE Y9414LB B8700AF B9968AM Qty 1 1 3 1 1 5 6 7 B9968AT B8700DR B9968AX 4 4 1 Item 1 7-10 Description Bezel Assembly† (select) (see page 7-13) Bezel Assembly* B.H. Screw, M4x14 Tag Plate Tag Cover Note: * CX2 0- - -1 Sheet † CX2 0- - -2 Sheet CMPL parts Packing SM 04L31A01-01A 7.3 CX2000 Customer Maintenance Parts List 1 2 3 4 5 29 28 6 7 8 9 10 11 43 7 44 21 20 19 22 45 1314 15 16 12 23 17 18 Customer Maintenance Parts List 24 25 26 27 34 39 38 37 36 35 41 42 40 33 32 30 31 SM 04L31A01-01A 7-11 7.3 CX2000 Customer Maintenance Parts List Part No. B8700SL B9967UD B8700RC B8700RB B8700SJ Qty 1 1 1 1 1 B8700RJ B9968QA B9968SD B9968TY B9968QA 1 1 1 1 1 Description Sub Power Board Assembly (not /P1) (select) S24-Sub Power Board Assembly (/P1) 2Loop AD Board Assembly (CX2210, 2220) (select) 4Loop AD Board Assembly (CX2410, 2420, 2610, 2620) 2Loop Scanner Assembly (CX2210, 2220) (select) 4Loop Scanner Assembly (CX2410, 2420, 2610, 2620) Slow AD (PT) Assembly 10ch Scanner Assembly (select) 10ch ISO Scanner Assembly (/N2) Slow AD (PT) Assembly (CX2 20) B9968SD B9968TY Y9414LB B9968EL A1193MN 1 1 1 1 1 10ch Scanner Assembly (CX2 10ch Scanner Assembly (CX2 B.H. Screw, M4x14 Clamp Magnetic Parts 20) (select) 20 /N2) 1 1 1 1 1 Comm Board Assembly (CX2 Comm Board Assembly (CX2 Comm Board Assembly (CX2 CPU Board Assembly CPU Board Assembly 0-1) 0-2) (select) 0-0) 12 13 B9968TQ B9968TP B9968SQ B8700CW B8700RA 14 15 16 17 18 B9968CX B8700RK B8700NA B8700AE B8700QZ 1 1 1 1 1 19 20 21 22 23 B9900BR A9069KY B9968EM B8700UZ B9968GA 1 1 1 1 1 Battery Assembly Clamp Rivet Memory Board Assembly FDD Drive Assembly (CX2 0-1) 24 25 26 27 28 A1092UN Y9204LS B9968MA B9968SW A1484UP 1 3 1 1 1 Memory System Screw FDD FCC FDD Board Assembly Power Supply (not /P1) 29 30 31 32 B9968SZ B9968SP B9968GL A1492JS Y9208LB 1 1 1 1 2 (select) DC24 Power Assembly (/P1) Mother Board Assembly ATA Drive Assembly (CX2 0-3) Socket Screw 33 34 35 36 37 B9968SV B9968GD A1150UN Y9203LB B9968MB 1 1 1 3 1 PC-Card Board Assembly Zip Drive Assembly (CX2 0-2) Memory System Screw IDE FFC 38 39 40 41 42 B9968SU B9968ST B9968HQ B9968HR B9968UX 1 1 1 1 1 Zip Conn Board Assembly IDE Board Assembly VGA Board Assembly (/D5) Bracket VGA Board Assembly 43 44 45 A1193MN Y9414LB B9968EL 1 1 1 Magnetic Parts (/D5) B.H. Screw, M4x14 (/D5) Clamp (/D5) Item 1 2 3 4 5 6 7 8 9 10 11 7-12 CPU Bracket Display Board Assembly Memory Board Assembly Name Plate Memory Board & Battery Assembly SM 04L31A01-01A 7.3 CX2000 Customer Maintenance Parts List Bezel Assembly 25 26 24 23 22 29 28 27 21 20 19 Customer Maintenance Parts List 18 17 16 15 14 13 11 12 10 4 2 3 4 5 6 7 8 9 10 11 12 13 Part No. B8700BR B8700BV B9968BX B9900NQ B9968BW Model B8700BA B8700BE Qty Item 1 1 1 1 1 1 1 1 1 B9968BS B9567AQ B9968BY B8700BW B8700BY 1 1 1 1 1 1 1 1 B9968SK B9968BT B9968BZ B9968AY B9968GZ 1 1 1 1 5 SM 04L31A01-01A 9 5 1 1 1 1 5 1 Description Key Case Assembly (for English) (select) Key Case Assembly (for Japanese) Door Knob Spring Front Plate Front Cover Spring Hinge Pin Key Top (for English) Key Top (for Japanese) (select) Switch Board Assembly Front Case Micro Switch Pin Packing Screw 8 6 7 Item 14 15 16 17 29 Part No. B9968BU B9968MC B8700BB A1067VA A1046VZ 18 19 20 21 22 Qty Model B8700BA B8700BE 2 3 7 1 1 1 1 1 1 1 1 1 1 Description FPC Guard Key FFC Sub Bezel Assembly LCD Back Light B9968ML B9968BL A1527UP B9968EN B9988DL 1 1 1 1 2 1 1 1 1 2 Inverter Cover Stay Bracket Power Supply Bushing Screw 23 24 25 26 27 B9968BM B9968BQ B9543SQ B9968BP B8700DD 2 1 5 1 1 2 1 5 1 1 Hinge Arm Hinge Bracket NRP-345 Hinge Bracket Blind Sheet 28 B8700DC 1 1 Blind Sheet 7-13 7.3 CX2000 Customer Maintenance Parts List 59 58 60 71 65 64 70 68 61 57 62 63 66 69 31 27 26 25 29 28 30 32 39 35 34 33 37 36 38 40 47 43 42 41 45 44 46 48 55 51 50 49 53 52 54 56 7 2 4 3 5 6 8 1 15 10 12 11 9 13 14 16 23 18 20 19 21 22 24 17 67 7-14 SM 04L31A01-01A 7.3 CX2000 Customer Maintenance Parts List Qty Item Part No. 1 B8700CP B8700CQ B8700CR B9968DN B9968DJ 1 1 1 1 2 B9968DF B9900SG B9968DG B8700AP B8700AQ 1 2 1 1 1 Cover Assembly Screw Cover Name Plate (for B8700CP) Name Plate (for B8700CQ) (select) B8700AR B8700AV B8700AW B8700AX B8700HP 1 1 1 1 1 Name Plate (for B8700CR) Name Plate (for B8700CP) Name Plate (for B8700CQ) (select) Name Plate (for B8700CR) Name Plate (for B8700CP) 10 B8700HQ B8700HR B8700CS B8700CU B9968DJ 1 1 1 1 2 11 12 13 14 B9968DF B9900SG B9968DG B8700DS 1 2 1 1 67 2 3 4 5 6 7 8 9 Description CTRL IN MO. AS. 2L (CX22 ) CTRL IN MO. AS. 4L (CX24 ) CTRL IN MO. AS. 6L (CX26 ) Conn Cover Assembly (CX20 ) Screw Part No. 49 B8700CY B8700CX B8700CZ B8700CJ B8700DX 1 1 1 1 1 DIO Module Assembly (/CST1) Alarm Terminal Assembly (/A6) Alarm Terminal Assembly (/A6R) Alarm Terminal Assembly (/TPS4) Alarm Terminal Assembly (/A4F) B8700DZ B9968DN B9968DF B9968DG B9900SG 1 1 1 1 2 Alarm Terminal Assembly (/A4FR) Conn Cover Assembly Cover Assembly Cover Screw B8700CK B9968DJ B8700AY B8700DJ B8700DK 2 2 1 1 1 B8700HW B8700DG B8700DH B8700AZ B8700BH 1 1 1 1 1 Screw (/CST1) Screw (/A6 /A6R /TPS4 /A4F /A4FR) Name Plate (/CST1) Name Plate (/A6) Name Plate (/A6R) (select) Name Plate (/TPS4) Name Plate (/A4F) Name Plate (/A4FR) Name Plate (/CST1) Name Plate (/A6) B8700BJ B8700HZ B8700DE B8700DF B8700HY 1 1 1 1 1 Name Plate (/A6R) Name Plate (/TPS4) Name Plate (/A4F) Name Plate (/A4FR) Name Plate (/CST1) B8700HJ B8700HK B8700HX B8700HG B8700HH 1 1 1 1 1 Name Plate (/A6) Name Plate (/A6R) Name Plate (/TPS4) Name Plate (/A4F) Name Plate (/A4FR) B9968EE B9968AE B9968HM B9968RJ B9968RK 1 1 1 1 1 Terminal Name Plate (not /P1) (select) Name Plate (/P1) † COMM Term Board Assembly ‡ COMM Term Board Assembly B9968RG Y9308LB B9968AG B9968AH B9968AJ 1 2 1 1 1 COMM Term Board Assembly * B.H. Screw, M3x8 Name Plate † (select) Name Plate ‡ Sheet * 65 Y9305LB B9968EG B9968EH B9968EJ Y9305TS 1 1 1 1 1 B.H. Screw, M3x5 Power Plate Blind Bracket * † RS-232 Bracket Tapping Screw 66 68 69 70 71 Y9308LB A1447JZ B9968MN B9968EK B8700AG 1 1 1 1 1 B.H. Screw, M3x8 Modular Cover VGA Cable (/D5) (select) Cap (not /D5) Name Plate (select) 67 50 51 52 53 54 (select) Name Plate (for B8700CQ) Name Plate (for B8700CR) INPUT MO. AS. 10ch (CX2 10, 2 20) (select) 3RTD INP. AS. 10ch (CX2 10, 2 20 /N2) Screw 55 Cover Assembly Screw Cover Name Plate (for B8700CS) Qty Item 56 Description (select) (select) (select) (select) Name Plate (for B8700CU) Name Plate (for B8700CS) (select) Name Plate (for B8700CU) Name Plate (for B8700CS) (select) Name Plate (for B8700CU) B8700CT B8700CV B9968DN B9968DJ B9968DF 1 1 1 2 1 INPUT MO. AS. 20ch (CX2 20) 3RTD INP. AS. 20ch (CX2 20 /N3) (select) Conn Cover Assembly (CX2 10) Screw Cover Assembly B9900SG B9968DG B8700DT B8700DV B8700AK 2 1 1 1 1 25 67 B8700BG B8700HT B8700HV B8700CL B9968DN 1 1 1 1 1 26 27 28 29 30 B9968DF B9968DG B9900SG B8700CK B8700AL 1 1 2 2 1 31 32 33 67 34 B8700AS B8700HL B8700CM B9968DN B9968DF 1 35 36 37 38 39 B9968DG B9900SG B8700CK B8700AM B8700AT 1 2 2 1 1 16 17 67 18 19 20 21 22 23 24 1 1 1 40 41 67 42 43 B8700HM B8700CN B9968DN B9968DF B9968DG 1 1 1 1 1 44 45 46 47 48 B9900SG B8700CK B8700AN B8700AU B8700HN 2 2 1 1 1 SM 04L31A01-01A Screw Cover Name Plate (for B8700CT) Name Plate (for B8700CV) Name Plate (for B8700CT) 57 58 59 60 61 (select) (select) Name Plate (for B8700CV) Name Plate (for B8700CT) (select) Name Plate (for B8700CV) CTRL Module AS. 2L (CX22 , CX24 , CX26 Conn Cover Assembly (CX20 ) 62 63 64 ) (select) Cover Assembly Cover Screw Screw Name Plate (select) 7 (select) 1 1 1 1 1 Customer Maintenance Parts List B8700DU B8700AJ B8700BF B8700HS B8700HU 15 (select) Name Plate Name Plate CTRL Module AS. 4L (CX24 , CX26 ) (select) Conn Cover Assembly (CX20 , CX22 ) Cover Assembly Cover Screw Screw Name Plate Name Plate Name Plate CTRL Module AS. 6L (CX26 ) Conn Cover Assembly (CX20 , CX22 Cover Assembly Cover Note: , CX24 ) * CX2 0- † CX2 0- -0 -1 ‡ CX2 0- -2 (select) Screw Screw Name Plate Name Plate Name Plate 7-15 7.4 CX2000 Standard Accessories 2 1 4 3 6 5 7 Item Part No. 1 IM 04L31A01-01E IM 04L31A01-02E IM 04L31A01-71E IM 04L31A01-72E 2 B8700MA 3 4 5 6 7 B9900BX E9655FX A1053MP B9968PK CXA100-01 Qty 1 1 1 1 1 Description DAQSTATION CX2000 User’s Manual DAQSTATION CX2000 Operation Guide CX2000 Installation and Connection Guide Precaution on the Use of the CX1000/CX2000 CD for Manual for CX 2 5 1 1 1 Bracket Assembly B.H. Screw, M4x6 ( ) Mag Memory :Zip 100MB disk (CX2 0-2) Memory System :ATA flash memory card (CX2 DAQ Standard for CX Note: 7-16 0-3) CMPL parts SM 04L31A01-01A Chapter 8 Procedures for Disassembly (CX1000) 8.1 Removing the Bezel WARNING The service manual is to be used by properly trained personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to the safety precautions prior to performing any servicing. Even if servicing is carried out according to this service manual, or by qualified personnel, YOKOGAWA assumes no responsibility for any result occurring from that servicing. CAUTION If removing the assembly, be sure to disconnect the power cable from the outlet before doing so. Opening the Bezel 1. Open the Key Case Assembly then remove the two screws as shown. 8 IM 04L31A01-01E The Bezel hooks on to holes in the top of the case. Pull the bottom of the bezel outward slightly, lift the entire bezel upwards to release it from the case, then swing the bottom of the bezel upwards to open it. 8-1 Procedures for Disassembly (CX1000) 2. 8.1 Removing the Bezel Note • A gasket is attached to the edge of the case. When reinstalling the bezel, be sure that the gasket is correctly fastened. • A lever is attached to the right side of the bezel. By hooking the lever onto the side of the case, you can lock the bezel in the open position. Lever • The power switch is located in the upper left as you are facing the instrument. When this switch is turned OFF, it prevents power from being supplied to the instrument even when the power cord is connected. Therefore be sure to turn the power switch ON before you reassemble the instrument. Power Switch 8-2 IM 04L31A01-01E 8.1 Removing the Bezel Removing the Bezel 3. Remove the two flexible cables that connect the bezel to the main unit. The connector utilizes a latch. Pull outwards on the left and right sides of the latch to release it, then pull out the cable. Connector 4. Release the bezel by pushing the latches on the left and right hinge arms using a long pointed tools such as a Phillips head screw driver. With the latches pushed in, pull the bezel out and remove it from the main unit. Latch Hinge Arm 8 Procedures for Disassembly (CX1000) CX with the front bezel removed IM 04L31A01-01E 8-3 8.2 Removing the Board Assemblies and the Drive 1. Remove the bezel (see section 8.1). Removing the Board Assemblies 2. Remove the three bracket screws. 3. Slide the bracket down to pop it out of position, then remove it. Removing the bracket allows removal of all board assemblies. CX with the bracket removed. S-Power Assembly Comm Board Assembly CPU Board Assembly AD Assembly MON Scanner Assembly 4. Pull out and remove the S-Power assembly, Comm Board assembly, and CPU Board assembly. You may have to pull forcefully to remove the MON Scanner assembly and the AD assembly. You can pry them out gradually by inserting tools into the holes on either side of the board, then rocking the board side to side while pulling. 5. Note • • 8-4 The connection where the AD assembly is inserted into the mother board may be tight. When reinserting the boards, make sure that they are completely inserted into the connectors. If it is difficult to grasp the MON Scanner assembly, first remove the drive to provide extra space (see next page). IM 04L31A01-01E 8.2 Removing the Board Assemblies and the Drive Removing the Drive 1. Remove the screw as shown in the figure below. 2. Remove the drive by pulling it outwards. CAUTION The ZIP drive has a thin transparent plate that can be sharp. Take care not to cut your fingers when removing the drive. CX with all boards and the drive removed. 8 Procedures for Disassembly (CX1000) IM 04L31A01-01E 8-5 8.3 Removing the Keyboard 1. 2. 3. Remove the bezel (see section 8.1). Remove the two screws attaching the aluminum sheet to the back of the bezel. Remove the keyboard flat cable rivet. When removing the rivet, first pull out the pin. Aluminum Sheet Rivet 4. Use the tip of a pen to push in the hinge pin on the left side of the Key Case assembly, then pull the assembly toward you to remove it. Be careful not to damage the flat cable when removing the assembly. 5. Remove the two front case screws from the back of the key board. Tightening torque: 0.5—0.6 Nm (5—6 kgf cm) The front case is snapped on to the front cover. Use a tool such as a flathead screwdriver to pry open the front case and remove it. 6. 8-6 IM 04L31A01-01E 8.3 Removing the Keyboard Disassembled Key Case Assembly 8 Procedures for Disassembly (CX1000) IM 04L31A01-01E 8-7 8.4 Replacing the Backlight 1. 2. Remove the bezel (see section 8.1). Remove the two screws attaching the aluminum sheet on the back of the bezel, and the rivet that fastens the flat cable (see section 8.3). Aluminum Sheet Rivet Display Bracket Earth Bracket 3. 4. 5. Remove screw numbers 1?3 as shown, then remove the Earth bracket. Remove screw number 4, then remove the Display bracket. Remove the LCD screen from the plastic screen cover, and set the LCD screen down on a static-proof surface. Note Be careful not to scratch the LCD screen when handling it. 6. Remove the two screws as shown, then remove the rear cover. Note Only use a screw driver that precisely fits the screws. 7. 8-8 Remove the four screws from the sides of the LCD screen as shown. IM 04L31A01-01E 8.4 Replacing the Backlight 8. 9. Remove the four screws from the inside corners of the LCD as shown. Remove the two contact segments one at a time by grasping them with tweezers. Contact segment Lump Cable Bracket Connector FFC Contact segment 10. 11. Remove the FFC and lump cable from their connectors. Slowly lift the bracket upward to remove the bracket assembly. Note • • Take care not to damage any components on the inverter or signal processing boards. Be careful not to damage the FFC. LCD assembly with the bracket assembly removed Contact segment 8 Being careful not to damage any surrounding parts, grasp the backlight and slowly pull it out. Note • If the backlight is broken, be careful not to touch any glass shards when removing it. Also, check for glass shards that may have fallen into the lamp holder or other areas, and remove them. • Dirt and grime on the backlight and LCDs can cause unevenness in display intensity. Be sure to keep these devices clean as you handle them. LCD assembly with the backlight removed IM 04L31A01-01E 8-9 Procedures for Disassembly (CX1000) 12. 8.5 Removing the Terminal Assembly WARNING The service manual is to be used by properly trained personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to the safety precautions prior to performing any servicing. Even if servicing is carried out according to this service manual, or by qualified personnel, YOKOGAWA assumes no responsibility for any result occurring from that servicing. CAUTION When removing the assembly, be sure to disconnect the power cable from the outlet before doing so. 8-10 1. Remove the screws from the upper right and lower left of the two terminal assemblies. 2. Pull out the terminal assemblies. IM 04L31A01-01E